K Number
K231131

Validate with FDA (Live)

Date Cleared
2023-06-22

(62 days)

Product Code
Regulation Number
878.5400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This product is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.

Device Description

The BL-A10 is a 1060nm Diode Hyperthermic Laser Lipolysis system and is the latest technology for non-invasive body contouring. It applies a 1060nm wavelength laser to target the adipose tissue to reduce fat in areas such as flank and abdomen. The purpose of the device is to breakdown the fat cells in the target area and in doing so contour the body and reduce the number of fat cells in that area. It is designed to meet international safety and performance standards.

AI/ML Overview

The provided text is a 510(k) summary for the "Diode Laser Body Sculpture Systems," a medical device intended for non-invasive lipolysis. While it details the device and its equivalence to a predicate device, it explicitly states, "Clinical Testing: It is not applicable."

Therefore, I cannot provide information on acceptance criteria or a study that proves the device meets those criteria from this document. The manufacturer has determined that clinical testing is not necessary to demonstrate substantial equivalence to the predicate device for this submission.

The document discusses non-clinical testing, including electrical safety, electromagnetic compatibility, and biocompatibility, but these do not involve proving the device meets specific performance criteria for its intended clinical effect through human trials.

To summarize, for your specific request:

  1. A table of acceptance criteria and the reported device performance: Not provided in the document as clinical testing was deemed "not applicable."
  2. Sample sized used for the test set and the data provenance: Not applicable.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable (this device is a physical laser system, not an AI diagnostic tool).
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

The document focuses on demonstrating substantial equivalence to an existing legally marketed device (SculpSure K182741) based on similar indications for use, technological characteristics (laser type, wavelength, lipolysis method), and compliance with electrical safety and biocompatibility standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue rectangle with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 22, 2023

Shanghai Bele Medical Technology Co.,Ltd Jeremy Li General Manager Room 402, 4F, Building 1, No.3255 Shengang Road, Songjiang District Shanghai, Shanghai 201600 China

Re: K231131

Trade/Device Name: Diode Laser Body Sculpture Systems Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System For Aesthetic Use Regulatory Class: Class II Product Code: PKT Dated: March 24, 2023 Received: April 21, 2023

Dear Jeremy Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows the name "Jianting Wang -S" in a large, clear font. The text is horizontally oriented and appears to be the primary focus of the image. The background is plain, ensuring the text is easily readable.

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231131

Device Name Diode Laser Body Sculpture Systems

Indications for Use (Describe)

This product is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 3-510(k) summary K231131

l Submitter

SHANGHAI BELE MEDICAL TECHNOLOGY CO.,LTD

Room 402, 4F, Building 1, NO.3255 Shengang Road, Songjiang District, Shanghai, China

Contact Person

Jeremy Li

General manager

Phone: +86-189 6489 4730

Fax: +86-21 51012882

E-mail: jeremy@belemed.com

II Proposed Device

Trade Name of Device:Diode Laser Body Sculpture Systems
Common name:Low Level Laser System for Aesthetic Use
Regulation Number:21 CFR 878.5400
Regulatory Class:Class II
Product code:PKT
Review Panel:General & Plastic Surgery

III Predicate Devices

510(k) Number:K182741
Trade name:SculpSure
Common name:Low Level Laser System for Aesthetic Use
Classification:Class II
Product Code:PKT
ManufacturerCynosure

IV Device description

The BL-A10 is a 1060nm Diode Hyperthermic Laser Lipolysis system and is the latest technology for non-invasive body contouring. It applies a 1060nm wavelength laser to target the adipose tissue to reduce fat in areas such as flank and abdomen.

The purpose of the device is to breakdown the fat cells in the target area and in doing so contour the body and reduce the number of fat cells in that area. It is designed to meet international safety and performance standards.

V Indication for use

This product is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.

VI Comparison of technological characteristics with the predicate devices

ItemProposed devicePredicate device(K182741)Discussion
Product CodePKTPKTIdentical
Regulation No.21 CFR 878.540021 CFR 878.5400Identical
ClassClass IIClass IIIdentical
Indication for useThis product isintended for non-invasive lipolysis ofthe abdomen, flanks,back, and thighs inindividuals with aBody Mass Index(BMI) of 30 or less.The device isintended to affect theappearance of visiblefat bulges in theabdomen, flanks,back, and thighs.The CynosureSculpSureTM isintended for non-invasive lipolysis ofthe abdomen, flanks,back, and thighs inindividuals with aBody Mass Index(BMI) of 30 or less.In addition, thedevice is intended fornon-invasive lipolysisof the submentalarea in individualswith a BMI of 49 orless. The device isintended to affect theappearance of visiblefat bulges in theabdomen, flanks,back, thighs andsubmental area.When using thepetite mask for non-invasive lipolysis ofthe submental area,the device can alsoaffect theappearance of laxtissue in thesubmental area.Equivalent
Laser typeDiode laserDiode laserIdentical
Wavelength1060nm±20 nm(infrared)1060nm±20 nm(infrared)Identical
Lipolysis methodHeat-assistedHeat-assistedIdentical
Spot size4 * 6 cm² on each ofthe Applicator heads(up to fourapplicators per bodytreatment)4 * 6 cm² on each ofthe Applicator heads(up to fourapplicators per bodytreatment)Equivalent
14.28 cm² (for standard submental mask)10.49 cm2 (for petite submental mask)
Pulse widthCWCWIdentical
Power density0.6~1.6W/cm²Up to 1.4 W/ cm² (body)Up to 3.06 W/ cm² (submental)Similar
Attachment to patientBeltBeltIdentical
VoltageAC110-240V, 16-4A , 50/60Hz.AC200-240V, 50/60Hz, 20AEquivalent
Peak power35W (per applicator)30W (per applicator)Equivalent
CoolingContact coolingContact coolingIdentical

{4}------------------------------------------------

{5}------------------------------------------------

VII Non-Clinical Testing

A series of tests have been performed to verity that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards:

Electrical safety and electromagnetic compatibility

  • IEC 60601-1 : 2005+A1 : 2012+A2 : 2020 / EN 60601-1: 2006+A1: ● 2013+A12:2014+A2:2021 Medical electrical equipment. General requirements for basic safety and essential performance
  • IEC60601-1-2:2014+A1:2020 / EN 60601-1-2:2015+A1:2021 Medical electrical equipment - part 1-2 General requirements for basic - safety - and - performance -Collateral standard: Electromagnetic compatibility - Requirements and tests
  • IEC 60601-2-22:2019/EN 60601-2-22:2020 Medical electrical equipment - part 2-22: Particular requirements for basic essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
  • IEC 60825-1:2014/EN 60825-1:2014 Safety of laser products-Part 1:Equipment classification and requirements

Biocompatibility Evaluation:

Per FDA's Biocompatibility Guidance issued on September 04, 2020 and with regard to Table A.1 Evaluation Tests for consideration in ISO, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" " the following tests performed on the material which contacts with human for Biocompatibility:

  • Cytotoxicity;
  • Skin irritation;

{6}------------------------------------------------

● Skin Sensitization.;

VIII Clinical Testing

lt is not applicable.

IX Conclusion

Base on the performance testing and validation studies that the subject device is substantially equivalent to the predicate device.

§ 878.5400 Low level laser system for aesthetic use

(a)
Identification. A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.