K Number
K231130

Validate with FDA (Live)

Device Name
TumorSight Viz
Manufacturer
Date Cleared
2023-12-26

(250 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TumorSight Viz is intended to be used in the visualization and analysis of breast magnetic resonance imaging (MRI) studies for patients with biopsy proven early-stage or locally advanced breast cancer. TumorSight Viz supports evaluation of dynamic MR data acquired from breast studies during contrast administration. TumorSight Viz performs processing functions (such as image registration, subtractions, measurements, 3D renderings, and reformats).

TumorSight Viz also includes user-configurable features for visualizing and analyzing findings in breast MRI studies. Patient management decisions should not be made based solely on the results of TumorSight Viz.

Device Description

TumorSight Viz is an image processing system designed to assist in the visualization and analysis of breast DCE-MRI studies.

TumorSight reads DICOM magnetic resonance images. TumorSight processes and displays the results on the TumorSight web application.

Available features support:

  • Visualization (standard image viewing tools, MIPs, and reformats)
  • Analysis (registration, subtractions, kinetic curves, parametric image maps, segmentation and 3D volume rendering)
  • Communication and storage (DICOM import, retrieval, and study storage)

The TumorSight system consists of proprietary software developed by SimBioSys, Inc. hosted on a cloud-based platform and accessed on an off-the-shelf computer.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for TumorSight Viz, based on the provided text:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by demonstrating that the device's performance (Mean Absolute Error and Dice Coefficients) is comparable to inter-radiologist variability and the predicate device, CADstream, and that "all tests met the acceptance criteria".

Measurement DescriptionUnitsAcceptance Criteria (Implied)Validation Testing (Mean Abs. Error ± Std. Dev.)
Tumor Volume (n=157)cubic centimeters (cc)Comparable to inter-radiologist variability6.48 ± 12.67
Tumor-to-breast volume ratio (n=157)%Comparable to inter-radiologist variability0.56 ± 0.93
Tumor longest dimension (n=163)centimeters (cm)Comparable to inter-radiologist variability1.48 ± 1.46
Tumor-to-nipple distance (n=161)centimeters (cm)Comparable to inter-radiologist variability1.00 ± 1.03
Tumor-to-skin distance (n=163)centimeters (cm)Comparable to inter-radiologist variability0.63 ± 0.60
Tumor-to-chest distance (n=163)centimeters (cm)Comparable to inter-radiologist variability0.94 ± 1.34
Tumor center of mass (n=157)centimeters (cm)Comparable to inter-radiologist variability0.735 ± 1.26
Performance MeasurementMetricAcceptance Criteria (Implied)Validation Testing (Mean ± Std. Dev.)
Tumor segmentation (n=157)Volume DiceSufficient for indicating location, volume, surface agreement0.676 ± 0.289
Surface DiceSufficient for indicating location, volume, surface agreement0.873 ± 0.264

Additionally, for the direct comparison with the CADstream predicate device and ground truth:

Performance MeasurementMetricTumorSight Viz / Ground Truth (Mean Abs. Error ± Std. Dev.)CADStream / Ground Truth (Mean Abs. Error ± Std. Dev.)Inter-radiologist Variability (Mean Abs. Error ± Std. Dev.)
Longest Dimension (n=136)Abs. Distance Error1.40 cm ± 1.43 cm1.11 cm ± 1.52 cm1.17 cm ± 1.38 cm
Tumor to Skin (n=136)Abs. Distance Error0.61 cm ± 0.46 cm0.49 cm ± 0.56 cm0.49 cm ± 0.54 cm
Tumor to Chest (n=136)Abs. Distance Error0.77 cm ± 0.90 cm1.37 cm ± 1.01 cm0.79 cm ± 1.01 cm
Tumor to Nipple (n=134)Abs. Distance Error0.98 cm ± 1.06 cm0.80 cm ± 0.86 cm0.82 cm ± 0.98 cm
Tumor Volume (n=134)Abs. Distance Error6.69 cc ± 13.53 cc8.09 cc ± 17.42 ccN/A (not provided for inter-radiologist variability)

The document states: "The mean absolute error and variability between the automated measurements (Validation Testing) and ground truth for tumor volume (measured in cc) and landmark distances (measured in cm) was similar to the variability between device-to-radiologist measurements and inter-radiologist variability. This demonstrates that the error in measurements is consistent to the variability between expert readers." It also notes: "We found that all tests met the acceptance criteria, demonstrating adequate performance for our intended use." And for the comparison to the predicate: "The differences in error between the mean absolute errors (MAE) for the predicate and subject device are clinically acceptable because they are on the order of one to two voxels for the mean voxel size in the dataset. These differences are clinically insignificant."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Validation (Test Set): 161 patients, corresponding to 163 samples (accounting for bilateral disease).
  • Data Provenance: Obtained from six (6) clinical sites in the U.S. All patients had pathologically confirmed invasive, early stage or locally advanced breast cancer. The data was collected to ensure adequate coverage of MRI manufacturer and field strength and similarity with the broader U.S. population for patient age, breast cancer subtype, T stage, histologic subtype, and race/ethnicity. This data is retrospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Seven (7) U.S. Board Certified radiologists.
  • Qualifications of Experts: U.S. Board Certified radiologists. Specific experience level (e.g., years of experience) is not explicitly stated beyond "expert readers."

4. Adjudication Method for the Test Set

  • Adjudication Method: For each case, two radiologists independently measured various characteristics. If the two radiologists did not agree on whether the candidate segmentation was appropriate, a third radiologist provided an additional opinion and established a ground truth by majority consensus (2+1 adjudication).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • The document describes a performance comparison between TumorSight Viz, CADstream (predicate), and ground truth, as well as inter-radiologist variability. However, it does not describe an MRMC comparative effectiveness study directly measuring how much human readers improve with AI vs. without AI assistance. The study focuses on the standalone performance of TumorSight Viz and its comparability to a predicate device and human variability.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

  • Yes, a standalone study was done. The reported performance metrics (Mean Absolute Error, Dice Coefficients) for TumorSight Viz against a radiologist-established ground truth represent the standalone performance of the algorithm. The document explicitly states: "The measurements generated from the device result directly from the segmentation methodology and are an inferred reflection of the performance of the deep learning algorithm."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: Pathologically confirmed breast cancer cases (for patient inclusion) combined with expert consensus of U.S. Board Certified radiologists for specific image measurements and segmentation appropriateness. The ground truth was established by two radiologists measuring characteristics, with a third radiologist adjudicating disagreements by majority consensus.

8. The sample size for the training set

  • Training Set Sample Size: 390 samples (from 736 patients mentioned for training and tuning).
  • Tuning Set Sample Size: 376 samples (from 736 patients mentioned for training and tuning).

9. How the ground truth for the training set was established

  • The document states that 736 patients (766 samples) were used for "training and tuning the device." It explicitly mentions that for the validation set, "Seven (7) U.S. Board Certified radiologists reviewed 163 validation samples to establish the ground truth for the dataset..."
  • The method for establishing ground truth for the training set is not explicitly detailed in the provided text. It is generally implied that such ground truth would also be established by experts, but the specifics are not given for the training portion.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

SimBioSys, Inc. % John J. Smith, M.D., J.D. Official Correspondent Hogan Lovells US LLP 180 North Lasalle Street. Suite 3250 Chicago, Illinois 60601

Re: K231130

Trade/Device Name: TumorSight Viz Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: November 21, 2023 Received: November 21, 2023

Dear John J. Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Wenbo Li for

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K231130

Device Name

TumorSight Viz

Indications for Use (Describe)

TumorSight Viz is intended to be used in the visualization and analysis of breast magnetic resonance imaging (MRI) studies for patients with biopsy proven early-stage or locally advanced breast cancer. TumorSight Viz supports evaluation of dynamic MR data acquired from breast studies during contrast administration. TumorSight Viz performs processing functions (such as image registration, subtractions, measurements, 3D renderings, and reformats).

TumorSight Viz also includes user-configurable features for visualizing and analyzing findings in breast MRI studies. Patient management decisions should not be made based solely on the results of TumorSight Viz.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Dver-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Submitter Details

SimBioSys, Inc. 180 North Lasalle Street, Suite 3250 Chicago IL 60601 United States Contact: Dr. John J. Smith, MD, JD Contact Telephone: (202) 637-3638 Contact Email: john.smith@hoganlovells.com

Details of the Submitted Device

Proprietary Name: TumorSight Viz Common Name: Medical image management and processing system Classification Name: System, Image Processing, Radiological Regulation Number: 892.2050 Product Code: QIH Committee/Panel: Radiology Device Class: II

Type of 510(k) Submission:

Traditional

Identification of the Legally Marketed Predicate Device

Predicate #: K092954

Predicate Trade Name: CADstream Version 5

Product Code: LLZ

Device Description

TumorSight Viz is an image processing system designed to assist in the visualization and analysis of breast DCE-MRI studies.

TumorSight reads DICOM magnetic resonance images. TumorSight processes and displays the results on the TumorSight web application.

Available features support:

  • Visualization (standard image viewing tools, MIPs, and reformats)
  • Analysis (registration, subtractions, kinetic curves, parametric image maps, segmentation and 3D ● volume rendering)
  • . Communication and storage (DICOM import, retrieval, and study storage)

The TumorSight system consists of proprietary software developed by SimBioSys, Inc. hosted on a cloud-based platform and accessed on an off-the-shelf computer.

{4}------------------------------------------------

Intended Use and Indications for Use

TumorSight Viz is intended to be used in the visualization and analysis of breast magnetic resonance imaging (MRI) studies for patients with biopsy proven early-stage or locally advanced breast cancer. TumorSight Viz supports evaluation of dynamic MR data acquired from breast studies during contrast administration. TumorSight Viz performs processing functions (such as image registration, subtractions, measurements, 3D renderings, and reformats).

TumorSight Viz also includes user-configurable features for visualizing and analyzing findings in breast MRI studies. Patient management decisions should not be made based solely on the results of TumorSight Viz.

Indications for Use Comparison

CADstream is intended to be used in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. CADstream supports evaluation of dynamic MR data acquired during contrast administration. CADstream performs other user selected processing functions (such as image registration, subtractions, measurements, 3D renderings, and reformats). Although the Indication for Use statement for Tumorsight Viz are not identical to that of the predicate, the differences do not alter the intended use as an image visualization device, nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for the visualization and analysis of dynamic magnetic resonance imaging (MRI) studies.

Technological Characteristics

Visualization of dynamic magnetic resonance imaging (MRI) studies is the technological principle for both the subject and predicate devices. It is based on the use of dynamic MRI images in DICOM format which are to be viewed and analyzed by a skilled physician. Both the subject and predicate devices perform the following same technological features:

  • . Standard Image Viewing Tools (zoom, pan, window/level)
  • Image Post Processing (MIPs, reformats, image registration) ●
  • Parametric Maps ●
  • Kinetic Curves
  • Automatic Volume Segmentation
  • Automatic Linear Measurements (distance to nipple, chest, and closest skin surface) ●
  • . DICOM Image Import

The following technological features differ between the subject and predicate devices:

  • Ability to review additional imaging modalities (mammography and ultrasound)
  • Interventional planning ●
  • User created collage of study images ●
  • Serial comparisons ●
  • Customizable reporting ●

{5}------------------------------------------------

Performance Tests

SimBioSys has completed performance testing on an independent dataset to ensure TumorSight Viz meets clinically acceptable levels.

DCE-MRI were obtained from seven hundred thirty-six (736) patients (corresponding to 766 samples when accounting for bilateral disease) were obtained from twelve (12) clinical sites in the U.S. for use in training and tuning the device. DCE-MRI were obtained for one hundred sixty-one (161) patients (corresponding to 163 samples when accounting for bilateral disease) were obtained from six (6) clinical sites in the U.S. for use in validating the device. All patients had pathologically confirmed invasive, early stage or locally advanced breast cancer.

Data was collected to ensure adequate coverage of MRI manufacturer and field strength, and to ensure similarity with the broader population of early-stage and locally advanced breast cancer patients in the U.S. Specifically, patient age at diagnosis, breast cancer subtype, T stage, histologic subtype, and race/ethnicity all reflect the broader U.S. population.

Training Dataset(n=390 samples)Tuning Dataset(n=376 samples)Validation Dataset(n=163 samples)
Age
<3010 (2.6%)9 (2.4%)4 (2.5%)
30-3972 (18.5%)62 (16.5%)33 (20.2%)
40-49103 (26.4%)104 (27.7%)37 (22.7%)
50-59117 (30.0%)109 (29.0%)48 (29.4%)
60-6965 (16.7%)66 (17.6%)32 (19.6%)
>7021 (5.4%)26 (6.9%)9 (5.5%)
Missing2 (0.5%)0 (0.0%)0 (0.0%)
Race/Ethnicity
Black†73 (18.7%)92 (24.5%)19 (11.7%)
Asian and Pacific Islander†20 (5.1%)17 (4.5%)8 (4.9%)
White†267 (68.5%)226 (60.1%)121 (74.2%)
American Indian or AlaskaNative†6 (1.5%)0 (0.0%)0 (0.0%)
Other9 (2.3%)3 (0.8%)2 (1.2%)
Hispanic0 (0.0%)8 (2.1%)6 (3.7%)
Missing/Unknown22 (5.6%)31 (8.2%)7 (4.3%)
  • Non-Hispanic

{6}------------------------------------------------

The following subgroups present in the dataset were comparable to the U.S. population: cancer subtype, grade, histology, T stage, and N stage.

Images were acquired from sites that utilize standard of care dynamic contrast enhanced MR protocols from GE, Philips, and Siemens scanners with both 1.5T and 3T field strength magnets.

Seven (7) U.S. Board Certified radiologists reviewed 163 validation samples to establish the ground truth for the dataset according to predefined guidelines. For each case, two radiologists measured various characteristics about the cancer including longest dimensions along three axes and tumor to landmark (chest, nipple, skin) distances. Each study was reviewed by two radiologists to determine if the candidate segmentation was appropriate. In cases where the two radiologists did not agree on whether the segmentation was appropriate, a third radiologist provided an additional opinion and established a ground truth by majority consensus.

Independence of validation data from training data was ensured by confirming there was no overlap of patients between training/tuning and validation datasets.

The validation samples were tested using both the TumorSight Viz device and the CADstream device.

The measurements generated from the device result directly from the segmentation methodology and are an inferred reflection of the performance of the deep learning algorithm. For example, the distance from chest or skin is calculated after the deep learning segmentation identifies the region of interest and then the resulting measurement is output.

The mean absolute error and variability between the automated measurements (Validation Testing) and ground truth for tumor volume (measured in cc) and landmark distances (measured in cm) was similar to the variability between device-to-radiologist measurements and inter-radiologist variability. This demonstrates that the error in measurements is consistent to the variability between expert readers. Performance data for the automated measurements is summarized below:

Measurement DescriptionUnitsValidation Testing(Mean Abs. Error ± Std.Dev.)
Tumor Volume (n=157)cubic centimeters (cc)6.48 ± 12.67
Tumor-to-breast volume ratio (n=157)%0.56 ± 0.93
Tumor longest dimension (n=163)centimeters (cm)1.48 ± 1.46
Tumor-to-nipple distance (n=161)centimeters (cm)1.00 ± 1.03
Tumor-to-skin distance (n=163)centimeters (cm)0.63 ± 0.60
Tumor-to-chest distance (n=163)centimeters (cm)0.94 ± 1.34
Tumor center of mass (n=157)centimeters (cm)0.735 ± 1.26

The tumor segmentation was assessed using the Dice coefficient, utilizing both the volumetric and surface Dice coefficients, which together validate the location, volume, and surface agreement with a reference standard.

{7}------------------------------------------------

The surface Dice coefficient is particularly useful as a proxy for the accuracy of 3D rendering and surfaceto-surface distances. Additionally, to further assess the tumor segmentation localization accuracy, we used the distance between the centers of mass of the reference standards and device-generated regions.

Results of Dice and surface Dice are summarized below:

Performance MeasurementMetricValidation Testing(Mean ± Std. Dev.)
Tumor segmentation (n=157)Volume Dice0.676 ± 0.289
Surface Dice0.873 ± 0.264

We found that all tests met the acceptance criteria, demonstrating adequate performance for our intended use.

Risk Management

The device risks were managed and controlled following the requirements of ISO 14971 standard. The device hazards were identified, their risk levels were evaluated and mitigation measures were taken to reduce the risk levels. The benefits of the TumorSight Viz software, outweigh the device residual risks.

Substantial Equivalence

TumorSight Viz is comparable to the predicate in terms of intended use, technological characteristics, and principle of operation.

Predicate Device Comparison
CADstream version 5(predicate)TumorSight Viz
510(k)K092954TBD
ManufacturerMerge CAD Inc.SimBioSys Inc.
Regulation Number892.2050892.2050
Regulation NameMedical image management andprocessing systemMedical image management andprocessing system
Classification22
Device Common NameImage Processing SystemImage Processing System
Product CodeLLZQIH
Functions- Extract dynamic contrastenhanced MRI sequence fromMRI images for the 3D displayand visualization of the anatomyof patient's breast- Extract dynamic contrastenhanced MRI sequence fromMRI images for the 3D displayand visualization of the anatomyof patient's breast
CADstream is intended to beused in the visualization,analysis, and reporting ofmagnetic resonance imaging(MRI) studies. CADstreamsupports evaluation of dynamicMR data acquired duringcontrast administration.CADstream performs other userselected processing functions(such as image registration,subtractions, measurements, 3Drenderings, and reformats).CADstream also includes user-configurable features forreporting on findings in breast orgeneral MRI studies.Additionally, CADstream assistsusers in planning MRM guidedinterventional procedures.When interpreted by a skilledphysician, this device providesinformation that may be used forscreening, diagnosis, andinterventional planning. Patientmanagement decisions shouldnot be made based solely ontheresults of CADstream.CADstream may also be used asan image viewer of multi-modality, digital images,including ultrasound andmammography. CADstream isnot intended for primaryinterpretation of digitalmammography images.TumorSight Viz is intended tobe used in the visualization andanalysis of breast magneticresonance imaging (MRI)studies for patients with biopsyproven early-stage or locallyadvanced breast cancer.TumorSight Viz supportsevaluation of dynamic MR dataacquired from breaststudiesduring contrastadministration. TumorSight Vizperforms processing functions(such as image registration,subtractions, measurements, 3Drenderings, and reformats).TumorSight Viz also includesuser-configurable features forvisualizing and analyzingfindings in breast MRI studies.Patient management decisionsshould not be made based solelyon the results of TumorSightViz.
Intended Use
Data Source (Input)MRIMRI
Output/AccessibilityGraphic and text results ofbreast anatomy are accessed viaa device with internetconnectivityGraphic and text results ofbreast anatomy are accessed viaa device with internetconnectivity
Physical Characteristics"-non-invasive software package-DICOM compatible""-non-invasive software package-DICOM compatible"
SafetyClinician review and assessmentof analysis prior to use inplanning MRI guidedinterventional procedures.Clinician review and assessmentof analysis prior to use in pre-operative planning.
Predicate Device Feature Comparison
FeatureCADstream version 5(predicate)TumorSight Viz
Standard image viewing toolsYesYes
MIPsYesYes
ReformatsYesYes
RegistrationYesYes
Subtraction seriesYesYes
View 3D volume renderingYesYes
Kinetic curvesYesYes
Parametric image mapsYesYes
DICOM importYesYes
View finding volumeYesYes
View finding locationYesYes
View finding sizeYesYes
View kinetic curve withhighest uptakeYesYes
View finding distance tonippleYesYes
View finding distance to skinYesYes
View finding distance to chestYesYes
View adjusted finding sizeYesNo - Segmentation is noteditable, but surgical marginsare editable
Interactive rotation of 3Dvolume renderingYesYes

A table comparing the key features of the subject and predicate devices is provided below:

{8}------------------------------------------------

{9}------------------------------------------------

Performance of TumorSight Viz was directly compared to that of CADstream for measurements including tumor longest dimension, tumor to skin distance, tumor to chest distance, and tumor to nipple distance. As summarized in the following table, these were comparable to inter-radiologist variability in the same measurements:

PerformanceMeasurementNMetricTumorSightViz/CADStreamTumorSightViz/GroundTruthCADStream/ GroundTruthInterradiologistVariability
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{10}------------------------------------------------

(Mean± Std. Dev.)(Mean ±Std. Dev.)(Mean ±Std. Dev.)(Mean± Std. Dev.)
LongestDimension136Abs.DistanceError1.48 cm± 1.71 cm1.40 cm ±1.43 cm1.11 cm ±1.52 cm1.17 cm± 1.38 cm
Tumor to Skin136Abs.DistanceError0.94 cm± 0.69 cm0.61 cm ±0.46 cm0.49 cm ±0.56 cm0.49 cm± 0.54 cm
Tumor toChest136Abs.DistanceError1.76 cm± 1.32 cm0.77 cm ±0.90 cm1.37 cm ±1.01 cm0.79 cm± 1.01 cm
Tumor toNipple134Abs.DistanceError0.86 cm± 1.00 cm0.98 cm ±1.06 cm0.80 cm ±0.86 cm0.82 cm± 0.98 cm
Tumor Volume134Abs.DistanceError9.54 cc ±20.89 cc6.69 cc ±13.53 cc8.09 cc ±17.42 ccN/A

The differences in error between the mean absolute errors (MAE) for the predicate and subject device are clinically acceptable because they are on the order of one to two voxels for the mean voxel size in the dataset. These differences are clinically insignificant.

Substantial Equivalence Conclusion

The comparison of the features and non-clinical bench performance testing described above demonstrates that TumorSight Viz is substantially equivalent to the predicate device in function. Furthermore, performance testing in an independent dataset of radiologist measurement ground truth demonstrates adequate performance for the intended use.

Additionally, TumorSight Viz measurement outputs were compared directly to CADstream output for 136 cases, and both sets of measurements were directly compared to radiologist measurements. TumorSight Viz compared equivalently to CADstream on all measurements including tumor longest dimension, tumor to skin distance, tumor to chest distance, and tumor to nipple distance.

Non-clinical bench testing, an independent assessment of device performance to radiologist ground truth, and a direct comparison to CADstream demonstrate that TumorSight Viz is substantially equivalent to CADstream

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).