K Number
K231092

Validate with FDA (Live)

Device Name
T-SHAPE 2
Date Cleared
2023-09-22

(157 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The T-SHAPE 2 and its hand pieces are indicated for the relief of minor muscle aches and pains, relief of muscle spasm, temporary improvement of local blood circulation and temporary reduction in the appearance of cellulite.

Device Description

T-SHAPE 2 treatment is based on the simultaneous application of heat and mechanical manipulation to the tissue, wherein the heat is derived from light energy at a controlled infrared wavelength and from conducted radio frequency (RF) energy and the mechanical manipulation is derived from massage and/or vacuum.

The combined synergistic action between the micro-stimulators (rollers) and a negative pressure (vacuum) created within the hand pieces grasp the skin tissues allowing to achieve the same effects of kneading and stroking tissue by hand. Applications are pre-set by the machine or the operator in relation to intensity, frequency, length of session, degree of tissue suctioning, and allow to address the issues of each individual in an targeted manner.

T-SHAPE 2 is also equipped with:

  • Two light sources type inside some handpieces (see below) with the following wavelengths: 650 and l 980 nm.
  • -Electrodes for bipolar RF energy (500 kHz/ 1 MHz / 1,5 MHz).

Through the display, the light sources and RF energy can be fully deactivated or activated.

The light sources are neither adjustable in intensity (always output at nominal value, as per specifications) nor in frequency (always continuous - CW).

For the RF energy the user can select frequency between 2 MHz, 1 MHz, 500 kHz and intensity between the 10 levels available.

T-SHAPE 2 comprises a main console unit and several handpieces.

A microprocessor-based system controller is used to monitor and direct all the system function and the graphic user interface.

The main console can be connected to the following handpieces (two different hand piece can be connected at the same time):

  • -Large multipole handpiece with 6 diode laser and 6 RF plates;
  • -Medium multipole handpiece with 4 diode laser and 4 RF plates;
  • -Small multipole handpiece 4 RF plates and 1 red LED;
  • -Thermal imaging camera;
  • -Roll handpiece with 4 diode laser and 2 motorized roller with RF;
  • -Mesosphere handpiece with 45 motorized rotating spheres.
AI/ML Overview

The provided text describes a medical device called T-SHAPE 2 and its substantial equivalence to predicate devices, but does not contain information about specific acceptance criteria, quantitative performance metrics, or a study design with sample sizes, expert qualifications, or ground truth establishment.

The document states that "T-SHAPE 2 has been tested in accordance to consensus standard to demonstrate safety and Performance," and lists general performance data categories:

  • Biocompatibility testing: Evaluated according to "biocompatibility flow chart for the selection of toxicity tests for 510(k)s."
  • Safety and electromagnetic compatibility (EMC): Complies with IEC 60601-1, IEC 60825-1, IEC 60601-1-2, and IEC TR 60601-4-2 standards.
  • Software Verification and Validation Testing: Documentation provided per FDA guidance.
  • Performance Bench Testing: "Evaluated to verify the skin tolerance, effect and cosmetic properties of a professional treatment through evaluations and instrumental analysis performed by professionals under medical supervision and consumer self-assessments."

Without quantitative performance data, it is not possible to create a table of acceptance criteria and reported device performance. Also, all other requested details regarding sample sizes, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set information are missing from the provided text.

The closest the document comes to performance metrics for the T-SHAPE 2 is under "Performance Bench Testing," which mentions "evaluations and instrumental analysis performed by professionals under medical supervision and consumer self-assessments" for "skin tolerance, effect and cosmetic properties" and safety mechanisms for RF heating. However, no specific metrics, targets, or results are provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 22, 2023

B&M S.R.L. Marketing Nel Benessere % Chiara Violini Consultant Endo Engineering Srl Via Del Consorzio, 41 Falconara Marittima, Ancona 60015 Italy

Re: K231092

Trade/Device Name: T-shape 2 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: NUV, PBX Dated: August 22, 2023 Received: August 23, 2023

Dear Chiara Violini:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tanisha L. Hithe -S Tanisha L. 2023.09.22 Hithe -S 20:49:01 -04'00'

Tanisha Hithe, MS, MHS Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K231092

Device Name T-SHAPE 2

Indications for Use (Describe)

The T-SHAPE 2 and its hand pieces are indicated for the relief of minor muscle aches and pains, relief of muscle spasm, temporary improvement of local blood circulation and temporary reduction in the appearance of cellulite.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

lite.

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2023

See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "BALDAN" in large, bold, sans-serif font, with a horizontal line directly underneath it. Below the line, the word "GROUP" is printed in a smaller, sans-serif font, with each letter aligned directly under the corresponding letter of "BALDAN". The text is black against a white background.

510(k) Summary

K231092

Introduction:

This document contains the 510(k) Summary for the T-SHAPE 2 device. The content of this summary is based on the requirements of 21 CFR 807.92(c).

Applicant/B&M S.R.L.
ManufacturerMARKETING NEL BENESSERE
Name and Address:Via Leonardo Bruni, 25
20158 Milano (MI)
Italy
510(k) Contact Person:Chiara Violini
Consultant
Email: c.violini@endoengineering.it
Phone: +39-071-9156048
Fax: +39-071-0971883
Date Prepared:30/01/2023
Device Name:T-SHAPE 2
Common or Usual Name:Multifunction platform with LLLT, RF, vacuum and massager.
Classification:Class II
Classification Name:Massager, Vacuum, Light Induced Heating (NUV)
Massager, Vacuum, Radio Frequency Induced Heat (PBX)
Regulation number:21 CFR 878.4810 (NUV)
21 CFR 878.4400 (PBX)
Classification product code:NUV, PBX
Predicate Devices:K122579 – VelaShape – Syneron Medical, Limited
K191528 – Venus Legacy Pro – Venus Concept Ltd.
K211272 – ICOONE Laser Med – I-Tech Industries SRL

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "BALDAN" in a bold, sans-serif font, with a horizontal line underneath. Below the line, the word "GROUP" is printed in a smaller, sans-serif font, with each letter centered under the corresponding letter of "BALDAN". The overall design is simple and clean, likely representing a company or organization name.

Description of the device:

T-SHAPE 2 treatment is based on the simultaneous application of heat and mechanical manipulation to the tissue, wherein the heat is derived from light energy at a controlled infrared wavelength and from conducted radio frequency (RF) energy and the mechanical manipulation is derived from massage and/or vacuum.

The combined synergistic action between the micro-stimulators (rollers) and a negative pressure (vacuum) created within the hand pieces grasp the skin tissues allowing to achieve the same effects of kneading and stroking tissue by hand. Applications are pre-set by the machine or the operator in relation to intensity, frequency, length of session, degree of tissue suctioning, and allow to address the issues of each individual in an targeted manner.

T-SHAPE 2 is also equipped with:

  • Two light sources type inside some handpieces (see below) with the following wavelengths: 650 and l 980 nm.
  • -Electrodes for bipolar RF energy (500 kHz/ 1 MHz / 1,5 MHz).

Through the display, the light sources and RF energy can be fully deactivated or activated.

The light sources are neither adjustable in intensity (always output at nominal value, as per specifications) nor in frequency (always continuous - CW).

For the RF energy the user can select frequency between 2 MHz, 1 MHz, 500 kHz and intensity between the 10 levels available.

T-SHAPE 2 comprises a main console unit and several handpieces.

A microprocessor-based system controller is used to monitor and direct all the system function and the graphic user interface.

The main console can be connected to the following handpieces (two different hand piece can be connected at the same time):

  • -Large multipole handpiece with 6 diode laser and 6 RF plates;
  • -Medium multipole handpiece with 4 diode laser and 4 RF plates;
  • -Small multipole handpiece 4 RF plates and 1 red LED;
  • -Thermal imaging camera;
  • -Roll handpiece with 4 diode laser and 2 motorized roller with RF;
  • -Mesosphere handpiece with 45 motorized rotating spheres.

Indication for Use:

The T-SHAPE 2 and its hand pieces are indicated for the relief of minor muscle aches and pain, relief of muscle spasms, temporary improvement of local blood circulation, and temporary reduction in the appearance of cellulite.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Baldan Group. The word "BALDAN" is in large, bold, sans-serif font, with a horizontal line underneath. Below the line, the word "GROUP" is written in a smaller, sans-serif font, with each letter centered under the corresponding letter of "BALDAN".

Comparison of Technological Characteristics:

T-SHAPE 2 has the same technological characteristics (massager, vacuum, RF energy, light source, control mechanisms) and specifications as its predicate devices.

{6}------------------------------------------------

BALDANGROUPT-SHAPE 2B&M S.R.L.MARKETING NEL BENESSEVia Leonardo Bruni, 25
20158 Milano (MI)Phone: (+39) 02 393 583
Email: info@baldangroup.
SpecificationsPredicate Device(Primary)Predicate DevicePredicate DeviceDevice object of 510(k)Comparison toPredicate
Device Name(K number)VelaShapeK122579Venus Legacy ProK191528 with 4DBody (LB2) and 4DFace (LF2)applicatorsICOONE Laser MedK211272T-SHAPE 2/
SubmitterSyneron Medical,LimitedVenus Concept Ltd.I-Tech Industries SRLB&M S.R.L.MARKETING NELBENESSERE/
Product CodeNUVGEX, PBXNUV, ISANUV, PBXNot considered ISAand GEX becausethey are alreadyconsidered in NUVand PBX
Regulation Number21 CFR 878.481021 CFR 878.440021 CFR 878.481021 CFR 878.4810/
Regulatory ClassIIIIIIII/
Product PictureImage: Velashape deviceImage: Venus Legacy Pro deviceImage: ICOONE Laser Med deviceImage: T-Shape 2 device/

{7}------------------------------------------------

BALDN
GROUP

T-SHAPE 2

Via Leonardo Bruni, 25 20158 Milano (MI) Phone: (+39) 02 393 5831
Email: info@baldangroup.it

SpecificationsPredicate Device(Primary)Predicate DevicePredicate DeviceDevice object of 510(k)Comparison toPredicate
Device Name(K number)VelaShapeK122579Venus Legacy ProK191528 with 4DBody (LB2) and 4DFace (LF2)applicatorsICOONE Laser MedK211272T-SHAPE 2/
Mains110 VAC; 4A; 50 Hz;Single Phase230 VAC; 2.5A; 50 Hz;Single Phase100-120 VAC / 60Hz220-240 VAC / 50Hz240/110 Vac100-120 VAC / 60Hz220-240 VAC / 50HzEquivalent toPredicate Devices
Weight20 kg. / 44.1 lbs40 kg191,80 lb (87 kg)95 kgEquivalent toPredicate Devices
Dimensions38 x 49 x 132 cm /15 x 19 x 51.8 inch40X40X100 cm(DxWxH)37,40x80,71x19,68inch (95x205x50 cm)57 x 86 x 1.70 cmEquivalent toPredicate Devices
BiocompatibilityUnknownMaterials arebiocompatibleBiological EvaluationOf Medical Devices -Part 10: Tests ForIrritation And SkinSensitization /Biological EvaluationOf Medical Device -Part 5: Test forcytotoxicity - Accordingto ISO10993-10:2010,10993-5:2009 and10993-12:2012Biological evaluationfor cytotoxicity,irritation, and skinSensitizationEquivalent toPredicate Devices
SpecificationsPredicate DevicePredicate DevicePredicate DeviceDevice object of 510(k)Comparison to
(Primary)Predicate
Device Name(K number)VelaShapeK122579Venus Legacy ProK191528 with 4DBody (LB2) and 4DFace (LF2)applicatorsICOONE Laser MedK211272T-SHAPE 2/
Indications for useThe VelaShape isindicated for the reliefof minor muscle achesand pain, relief ofmuscle spasm,temporaryimprovement of localblood circulation,temporary reduction inthe appearance ofcellulite, and fortemporary reductionof thighscircumferences.When used with the4D Body (LB2) and 4DFace (LF2) applicators,the Venus Legacy Prodevice is intended forthe delivery of non-thermal RF combinedwith Massage andmagnetic field pulsesfor the treatment ofthe following medicalconditions:• Relief of minormuscles aches andpain, relief of musclespasm• Temporaryimprovement of localblood circulation• Temporary reductionin the appearance ofcelluliteICOONE Laser med(also referred to asICOONE Medical laser)is indicated for therelief of minor muscleaches and pain, reliefof muscle spasms,temporaryimprovement of localblood circulation, andtemporary reduction inthe appearance ofcellulite.The T-SHAPE 2 and itshand pieces areindicated for the reliefof minor muscle achesand pain, relief ofmuscle spasms,temporaryimprovement of localblood circulation, andtemporary reduction inthe appearance ofcellulite.Equivalent toPredicate Devices
System componentLight, RF, Massage,Suction, RollersRF, Massage, Suction,RollersLight, Massage,Suction, RollersLight, RF, Massage,Suction, RollersEquivalent toprimary predicatedevice
Massage

{8}------------------------------------------------

B A L D A N
GROUP

T-SHAPE 2

Via Leonardo Bruni, 25 20158 Milano (MI) Phone: (+39) 02 393 5831
Email: info@baldangroup.it

{9}------------------------------------------------

BALDANGROUPT-SHAPE 2B&M S.R.L.MARKETING NEL BENESSEREVia Leonardo Bruni, 2520158 Milano (MI)Phone: (+39) 02 393 5831Email: info@baldangroup.it
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------
SpecificationsPredicate Device(Primary)Predicate DevicePredicate DeviceDevice object of 510(k)Comparison toPredicate
Device Name(K number)VelaShapeK122579Venus Legacy ProK191528 with 4DBody (LB2) and 4DFace (LF2)applicatorsICOONE Laser MedK211272T-SHAPE 2/
MechanicalMassageYesYesYesYesEquivalent toPredicate Devices
Light
Infraredwavelengths850 nmNot present650nm (LED)/ 915 nm(LASER)650/980 nmSee SubstantialEquivalencediscussion
Maximumpower/powerdensity3.3 W/1 W (915 nm)2.69 W/m²2 W/m²/
Suction/vacuum
ModePulsedUnknownFractionedPulsedEquivalent topredicate devices
Maximum Pressure- 390 mbar- 400 mbarUnknown- 400 mbarEquivalent topredicate devices
Radio Frequency
Frequency1 MHz1 MHzNot presentSelectable between2 MHz1,5 MHz1 MHz500 kHzSee SubstantialEquivalencediscussion
Peak RF POWER150 W150 W/150 WEquivalent topredicate devices
ModeBi-polarBi-polar/Bi-polarEquivalent topredicate devices

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Baldan Group. The word "BALDAN" is written in large, bold, serif font, with a horizontal line running underneath it. Below the line, the word "GROUP" is written in a smaller, sans-serif font. The letters are evenly spaced and the overall design is simple and professional.

Performance data:

The following performance data are provided in support of the substantial equivalence determination:

Biocompatibility testing

Materials in contact with patient skin, for a duration less than 24 hours, were evaluated according to the "biocompatibility flow chart for the selection of toxicity tests for 510(k)s", attachment C of "Criteria of Use of International Standard ISO 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing' (blue book memo)".

Safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the T-SHAPE 2 device.

The system complies with the IEC 60601-1, IEC 60825-1 standards for safety and the IEC 60601-1-2 and IEC TR 60601-4-2 standards for EMC.

Software Verification and Validation Testing

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

Performance Bench Testing

T-SHAPE 2 has been evaluated to verify the skin tolerance, effect and cosmetic properties of a professional treatment through evaluations and instrumental analysis performed by professionals under medical supervision and consumer self-assessments.

Comparison of Intended Use:

T-SHAPE 2 device's Intended Use is the same Intended Use of its predicate device.

Substantial Equivalence:

T-SHAPE 2 device is as safe and effective as its predicate devices.

T-SHAPE 2 device has the same intended use and same technological characteristics and specification as its predicate devices, moreover, T-SHAPE 2 has been tested in accordance to consensus standard to demonstrate safety and Performance.

The choice of four frequencies allows you to act more precisely on different layers of the tissue, contrary to a single fixed frequency.

The safety of the treatment is due to the characteristics of the latest generation equipment, which allow you to constantly monitor the parameters set during the session. The system has two safety levels: the first is preset on the skin resistance values with gradual emission of current pulses, the skin resistance have been considered variable between 50 and 500 Ohm and the voltage and power values of the settable 10 level has been calculated considering this range of skin resistance; the second is the intensity by the operator who monitors moment by moment the progress of the treatment and the endogenous heat triggered. Each frequency level respects this operating mode to avoid excessive heating.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the text "BALDAN" in large, bold letters, with a horizontal line running underneath it. Below the line, the word "GROUP" is written in smaller letters, aligned with the letters above. The text appears to be a logo or heading, possibly for a company or organization named Baldan Group.

Thus, T-SHAPE 2 device is substantially equivalent to its predicate devices.

Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, than the legally marketed predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.