K Number
K231041

Validate with FDA (Live)

Date Cleared
2023-07-14

(93 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

JiveX is a software only Picture Archiving and Communication System intended to display, process, read, report, communicate, distribute, store, and archive medical data which is available as DICOM or HL7 data, including mammographic images, and bio signals. JiveX also converts case related non-image documents, archives them as DICOM data and serves as a vendor neutral archive.

It supports the physician in diagnosis.

For primary image diagnosis in Mammography only uncompressed or non-lossy compressed images must be used.

Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.

Note: Web-based image distribution and mobile device display of mammographic images are not intended for diagnostic purposes.

For users in the United States of America: Mobile device display is not intended for diagnostic purposes.

Device Description

JiveX is a PACS software, with a Moderate level of concern.

A Communication Server is communicating, storing, converting, and archiving images, documents and signal data via DICOM. HL7 and proprietary interfaces. It also renders images for the web-based image distribution.

The fat clients can be used as workstations for medical reading and reporting. They provide extensive functions for image display and image processing. The reporting of digital mammography images is also supported.

The web-based clients are mainly intended for image distribution on personal computers and mobile devices. They offer less functions than the fat clients. As far as the functions allow for it, the web clients can also be used for reading and reporting on personal computers.

AI/ML Overview

The provided text is a 510(k) Premarket Notification summary for the medical device JiveX, release 5.4. This document primarily focuses on demonstrating substantial equivalence to a predicate device (JiveX 5.3) based on identical indications for use and technological characteristics, rather than presenting a performance study for accepting new functionality or claims.

Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria for new functionality or claims. Specifically:

  • No table of acceptance criteria and reported device performance is provided. The document states that "non-clinical performance testing" was done and the "subject device was found to have a safety and effectiveness profile that is similar to the predicate device," but it does not detail specific performance metrics or acceptance thresholds.
  • No information is provided about sample size for a test set, data provenance, number of experts for ground truth, or adjudication methods.
  • No MRMC comparative effectiveness study is mentioned. The document focuses on demonstrating equivalence to a previous version of the same product.
  • No standalone (algorithm-only) performance is discussed.
  • The type of ground truth used for any performance evaluation is not specified.
  • No sample size for a training set or method for establishing ground truth for a training set is provided. This type of information would typically be relevant for AI/ML-based devices with novel algorithms, which is not the primary focus of this 510(k) submission.

The "Performance Data" section solely describes a typical software verification and validation process (reviews, automated tests, manual test cases, regression testing), which is general software quality assurance, not a specific performance study against defined acceptance criteria for a new clinical claim. The core argument for substantial equivalence relies on the fact that the indications for use and technological characteristics of JiveX 5.4 are identical to its predicate, JiveX 5.3.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

VISUS Health IT GmbH Axel Schreiber Vice President Process & Agile Services Gesundheitscampus-Sued 15 Bochum, 44801 Germany

July 14, 2023

Re: K231041

Trade/Device Name: JiveX (Model Number / Release: 5.4) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: June 9, 2023 Received: June 9, 2023

Dear Axel Schreiber:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231041

Device Name JiveX, Release 5.4

Indications for Use (Describe)

JiveX is a software only Picture Archiving and Communication System intended to display, process, read, report, communicate, distribute, store, and archive medical data which is available as DICOM or HL7 data, including mammographic images, and bio signals. JiveX also converts case related non-image documents, archives them as DICOM data and serves as a vendor neutral archive.

It supports the physician in diagnosis.

For primary image diagnosis in Mammography only uncompressed images must be used.

Typical users of this system are trained professionals, including but not limited to physicians, radioal technicians, and assistants.

Note: Web-based image distribution and mobile device display of mammographic images are not intended for diagnostic purposes.

For users in the United States of America: Mobile device display is not intended for diagnostic purposes.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 "510(k) Summary"

of the Traditional 510(k) Premarket Notification Submission for the medical device JiveX, release 5.4

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92

Submitter

Manufacturer (Owner)VISUS Health IT GmbH,Gesundheitscampus 1544801 Bochum, Germany
Registration Number3007667119
Contact PersonAxel Schreiber, MD, PhDVice President Process & Agile ServicesTelephone +49 234 93693-0Email: schreiber@visus.com
Date PreparedJune 9th, 2023

Device

510(k) NumberK231041
Trade NamesJiveX
Common NamePicture Archiving and Communication System (PACS)
Classification PanelRadiology
CFR Section21 CFR §892.2050
Device ClassClass II
Product CodeLLZ

Predicate Device

Predicate DeviceK212321
Trade NamesJiveX
Common NamePicture Archiving and Communication System (PACS)
Classification PanelRadiology
CFR Section21 CFR §892.2050
Device ClassClass II
Product CodeLLZ

{4}------------------------------------------------

Device Description

JiveX is a PACS software, with a Moderate level of concern.

A Communication Server is communicating, storing, converting, and archiving images, documents and signal data via DICOM. HL7 and proprietary interfaces. It also renders images for the web-based image distribution.

The fat clients can be used as workstations for medical reading and reporting. They provide extensive functions for image display and image processing. The reporting of digital mammography images is also supported.

The web-based clients are mainly intended for image distribution on personal computers and mobile devices. They offer less functions than the fat clients. As far as the functions allow for it, the web clients can also be used for reading and reporting on personal computers.

Indications for Use

JiveX is a software only Picture Archiving and Communication System intended to display, process, read, report, communicate, distribute, store, and archive medical data which is available as DICOM or HL7 data, including mammographic images, and bio signals. JiveX also converts case related non-image documents, archives them as DICOM data and serves as a vendor neutral archive.

It supports the physician in diagnosis.

For primary image diagnosis in Mammography only uncompressed or non-lossy compressed images must be used.

Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.

Note: Web-based image distribution and mobile device display of mammographic images are not intended for diagnostic purposes.

For users in the United States of America: Mobile device display is not intended for diagnostic purposes.

Note: the indications for use have not changed compared to the predicate device.

Comparison of Technological Characteristics

The subject device and the predicate device have identical technological characteristics: JiveX is a software only medical device (SaMD). The communication server and the fat clients for radiology reading are implemented in Java. The web clients are implemented in java script / type script, HTML and CSS.

The software is run on the customer's own hardware. Clients run on personal computers with MS windows operating systems. Alternatively, clients may be run on a server (-farm) and can be distributed using Citrix. The mobile client runs on iPAD. The server also runs on MS Windows operating systems using server hardware either directly or via virtual machines.

{5}------------------------------------------------

Subject Device (K231041)Predicate Device (K212321)
Name, ReleaseJiveX 5.4JiveX 5.3
ManufacturerVISUS Health IT GmbHSame
Indications for useJiveX is a software only PictureArchiving and CommunicationSystem intended to display, process,read, report, communicate,distribute, store, and archivemedical data which is available asDICOM or HL7 data, includingmammographic images, and biosignals. JiveX also converts caserelated non-image documents,archives them as DICOM data andserves as a vendor neutral archive.It supports the physician indiagnosis.For primary image diagnosis inMammography only uncompressedor non-lossy compressed imagesmust be used.Typical users of this system aretrained professionals, including butnot limited to physicians,radiologists, nurses, medicaltechnicians, and assistants.Note: Web-based image distributionand mobile device display ofmammographic images are notintended for diagnostic purposes.For users in the United States ofAmerica: Mobile device display isnot intended for diagnostic purposes.Same
Prescr./OTCPrescriptionSame
Design /Architectureclient / serverSame
Software onlyYes. Software as a Medical Device(SaMD)Same
OperatingsystemsServer: Win. 10/11 Srv. 2012/2016/2019/2022; Client: Win. 8.1/10/11;iOS 14/15/16Server: Win. 10 Srv. 2012/2016/2019Client: Win. 8.1/10; iOS 13, 14
Image anddocumentcommunicationTCP/IP, DICOM, HL7, IHE XDS,WADO-URI, proprietary internaldata transfer protocols, proprietaryinterface to accept JPEG from aniPhone via web interfaceSame
Subject Device (K231041)Predicate Device (K212321)
Accepted ImageFormatsDICOM data + data accepted as nonDICOM and converted to DICOMfor storage: PDF, JPG, TIFF,standard and proprietary ECGformatsSame
Supportedstorage solutionsLocal storage on HDD/RAID/DVD,Network: NAS, SAN, long termstorage solutionsSame
Image datacompressionJPEG 2000 lossless & lossy, ZIP,LZ4, JPEG lossy for web clientsDisplay as received: JPEG lossless& lossy, RLE, MPEG-2Same
Web basedaccessDesktop (not intended formammography reading). Mobiledevices (not intended for reading)Same
Virtualization &Citrix supportYes, VMware. Java and web clientscan be distributed via Citrix.Same
User administr.CentralizedSame
Workflowsupport,RIS/HISintegrationImage Call Up from RIS, PatientInformation Reconciliation,Instance Availability, receivedocuments via HL7 MDM.Supported Standards: HL7, IHE.PACS driven reading workflow.Same
IHE XDSXDS-Consumer, Integrated SourceRepository, XDS RepositorySame
HardwareWindows based, manufacturerindependent server, workstationsand client hardware, iPADSame
ImageProcessingAlgorithms- Zoom, Pan, Rotate, Flip, Magnify- Geometrical Measurements- ROI statistics- Mammography auto shutter- 3D Cross Reference- 4D Navigation- ECG measurements- Interpolation: nearest neighbor, bi-linear, bi-cubic, Lanczos, b-spline- Filters: sharpen, CLAHE- Windowing and LUT mappingSame
Subject Device (K231041)Predicate Device (K212321)
ImageProcessingAlgorithms 3D- MPR, curved MPR- Max. Int. Projection- Min. Int. Projection- Volume Rendering- MIP for tomosynthesis data (notfor diagnostic use)- 3D image registration- fused display, SUV calculationSame
HangingprotocolsYesSame
BookmarksYes: CapturesSame

Tabular comparison of subject and predicate device:

{6}------------------------------------------------

{7}------------------------------------------------

Table 1: Tabular comparison of features and specifications

Performance Data

Summary of Non-Clinical Testing

Verification and validation is done through all development phases and includes

  • review of requirements, software design, code
  • Review and acceptance of newly implemented functionality
  • Daily build of the (intermediate) product and performance of automated tests on unit. . component, x-component and UI level
  • Verification / validation of "off the shelf software" ●
  • . Evaluation of selected software functionality with customers
  • Formal test run of all manual test cases pertaining to new or modified functionality ●
  • Impact testing for all changes that had been introduced ●
  • Extensive regression testing. including full regression testing of all inacceptable hazards. ●

Based on the non-clinical performance testing the subject device was found to have a safety and effectiveness profile that is similar to the predicate device.

Conclusions

While there are some differences between JiveX 5.4 and its predicate device, these differences are very minor and do not affect device substantial equivalence. JiveX 5.4 has the same basic operational principles and technical characteristics as its predicate device and it functions in the same manner. Additionally, it has the same indications for use. It is as safe, as effective, and performs as well as or better than its predicate device. Therefore, JiveX 5.4 is substantially equivalent to the predicate device cited within this submission

N/A