K Number
K231027

Validate with FDA (Live)

Date Cleared
2023-12-21

(254 days)

Product Code
Regulation Number
866.2390
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KaiBiLi Extended ViralTrans is intended for the collection of clinical specimens containing viruses, chlamydiae, mycoplasmas and ureaplasmas from the collection site to the test site. The KaiBiLi Extended ViralTrans is a culture-based medium that has been validated with multiple sample types and can be used to process clinical specimens using standard laboratory operating procedures for culture of clinical specimens or with other assays that utilize stable recoverable infectious viral particles or bacteria.

Device Description

The KaiBiLi Extended ViralTrans is room temperature stable and can sustain the viability of virus, chlamydiae, mycoplasma and ureaplasma when transported at 2-8°C or 20-25°C. The product can maintain proper pH environment and inhibit the growth of indigenous microbiota.

KaiBiLi Extended ViralTrans consists of modified Hank's balanced salt solution supplemented with bovine serum albumin, cysteine, glutamic acid, sucrose and HEPES. The HEPES buffer protects against pH changes. Phenol red is used as a pH indicator. Sucrose aids in the preservation of organism viability. To minimize the contamination of commensal organisms, Vancomycin, Econazole Nitrate, and Polymyxin B are incorporated into the medium formula.

KaiBiLi Extended ViralTrans is supplied as one plastic flat-bottom vial along with a screw cap for safely containing and transporting biological specimens. The vial is filled with either 1 mL or 3 mL of transport medium and glass beads, or in a kit format together with collection swabs.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Shelf Life (12 months storage at 2-8°C and 20-25°C):All criteria met for 12 months at both temperatures.
- Appearance (clear, pink, transparent liquid media with no turbidity or sedimentation)All lots tested passed.
- Volume (2.7-3.3 mL for 3 mL config, 0.9-1.1 mL for 1 mL config)All lots tested passed.
- pH (7.3 ± 0.2)All lots tested passed.
- Bacteriostatic Performance (no microorganism growth after 48 hours for E. Coli, Staphylococcus aureus, Streptococcus pyogenus, Candida albicans)All lots tested passed.
Recovery Studies (Viability of microorganisms over 48 hours at 2-8°C and 20-25°C):All criteria met.
- Viral, Chlamydial, Mycoplasmal, and Ureaplasmal RecoveryChanges within one log difference (+/- 90%) considered acceptable. All tested microorganisms (HSV-1, HSV-2, RSV, Cytomegalovirus, Adenovirus, Parainfluenza 3, Influenza A, VZV, C. pneumoniae, M. pneumoniae, U. urealyticum) showed acceptable recovery at both 2-8°C and 20-25°C for up to 48 hours.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:

    • Shelf Life: Three lots each of the 1 mL and 3 mL configurations (total of 6 lots) of KaiBiLi Extended ViralTrans medium were evaluated. Each lot was tested at time points T=0, 6, 9, 11, and 12 months.
    • Recovery Studies: Performance evaluation was carried out in three lots of media (newly manufactured, middle-aged, and recently expired media).
    • Microorganisms (each tested in corresponding clinical matrix):
      • Viruses (HSV-1, HSV-2, RSV, Cytomegalovirus, Adenovirus, Parainfluenza 3, Influenza A, VZV): Diluted into two different dilutions and tested in triplicate.
      • Chlamydophila pneumoniae: Tested.
      • Mycoplasma pneumoniae, Ureaplasma urealyticum: Diluted into four different dilutions and tested in duplicate for swab elution method, and in triplicate for roll plate method.
    • Clinical Matrices: Negative clinical matrix pools were contrived from a minimum of five donors for each matrix type (Skin, Genital specimens, Nasopharynx, Throat, Blood, Urine).
  • Data Provenance: The study appears to be a prospective laboratory study conducted by Hangzhou Genesis Biodetection & Biocontrol Co., Ltd. The text does not specify the country of origin of the donors for the negative clinical matrix pools, but the manufacturing company is based in China.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The study describes performance testing of the transport medium's ability to maintain microorganism viability, rather than a diagnostic device that requires expert interpretation of results to establish ground truth. The "ground truth" in this context is the initial microorganism concentration (titer/CFU) at hour zero and its subsequent viability over time, which is measured by laboratory methods (TCID50/mL, IFU/mL, CFU/mL).

4. Adjudication Method for the Test Set

This information is not applicable/provided as the study measures objective quantitative biological viability rather than subjective interpretation requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This document describes the performance of a biological specimen transport medium, not an AI-assisted diagnostic device.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. This document describes the performance of a biological specimen transport medium, not a standalone algorithm.

7. The Type of Ground Truth Used

The ground truth for the performance testing (recovery studies) was established through:

  • Laboratory-measured microbial concentrations: Viral titer (TCID50/mL), Fluorescent Foci Count (IFU/mL) for C. pneumoniae, and Colony Forming Units (CFU/mL or CFU) for M. pneumoniae and U. urealyticum. These quantitative measurements at time zero serve as the baseline "ground truth" to which subsequent measurements are compared.
  • Comparison to initial concentration: Changes in recovery were evaluated relative to the initial concentration (time point 0).

8. The Sample Size for the Training Set

This information is not applicable/provided. The performance studies described are for evaluating a finished medical device (transport medium) rather than training a machine learning model.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/provided as no training set for an algorithm is mentioned.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 21, 2023

Hangzhou Genesis Biodetection & Biocontrol Co., Ltd. Guan Emma, Int'l RA #139 10th Street (East) Hangzhou Economic And Technology Development Zone Hangzhou, 310018, China

Re: K231027

Trade/Device Name: KaiBiLi Extended ViralTrans Regulation Number: 21 CFR 866.2390 Regulation Name: Transport Culture Medium Regulatory Class: Class I, reserved Product Code: JSM Dated: April 11, 2023 Received: April 11, 2023

Dear Guan Emma:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming

{1}------------------------------------------------

product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Natasha Griffin -S

O.B.O. Ribhi Shawar, Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231027

Device Name KaiBiLi Extended ViralTrans

Indications for Use (Describe)

The KaiBiLi Extended ViralTrans is intended for the collection of clinical specimens containing viruses, chlamydiae, mycoplasmas and ureaplasmas from the collection site to the test site. The KaiBiLi Extended ViralTrans is a culture-based medium that has been validated with multiple sample types and can be used to process clinical specimens using standard laboratory operating procedures for culture of clinical specimens or with other assays that utilize stable recoverable infectious viral particles or bacteria.

Type of Use (Select one or both, as applicable)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Dec.18, 2023

The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:

Submitted byHangzhou Genesis Biodetection & Biocontrol Co., Ltd.#139, 10th Street (East), Hangzhou Economic and TechnologyDevelopment Zone, Hangzhou Zhejiang, CN 310018Tel : +86-571-87818163
Contact PersonEmma GuanInt'l RAHangzhou Genesis Biodetection & Biocontrol Co., Ltd.#139, 10th Street (East), Hangzhou Economic and TechnologyDevelopment Zone, Hangzhou Zhejiang, CN 310018Telephone: +86 571-86736901-892Email: yguan@hgb.com.cn
Trade NameKaiBiLi Extended ViralTrans
Common NameTransport culture medium
Review PanelMicrobiology
Regulation21 CFR 866.2390
ClassClass I
Product CodeJSM
Predicate DeviceK042970

1. Device Description

The KaiBiLi Extended ViralTrans is room temperature stable and can sustain the viability of virus, chlamydiae, mycoplasma and ureaplasma when transported at 2-8°C or 20-25°C. The product can maintain proper pH environment and inhibit the growth of indigenous microbiota.

KaiBiLi Extended ViralTrans consists of modified Hank's balanced salt solution supplemented with bovine serum albumin, cysteine, glutamic acid, sucrose and HEPES. The HEPES buffer protects against pH changes. Phenol red is used as a pH indicator. Sucrose aids in the preservation of organism viability. To minimize the contamination of commensal organisms, Vancomycin, Econazole Nitrate, and Polymyxin B are incorporated into the medium formula.

{4}------------------------------------------------

2. Intended Use and Indication for Use

The KaiBiLi Extended ViralTrans is intended for the collection and transportation of clinical specimens containing viruses, chlamydiae, mycoplasmas and ureaplasmas from the collection site to the test site. The KaiBiLi Extended ViralTrans is a culture-based medium that has been validated with multiple sample types and can be used to process clinical specimens using standard laboratory operating procedures for culture of clinical specimens or with other assays that utilize stable recoverable infectious viral particles or bacteria.

3. Device Specification

KaiBiLi Extended ViralTrans is supplied as one plastic flat-bottom vial along with a screw cap for safely containing and transporting biological specimens. The vial is filled with either 1 mL or 3 mL of transport medium and glass beads, or in a kit format together with collection swabs.

The KaiBiLi Extended ViralTrans is offered with one of the following configurations :

Cat. No.Description
M221001KaiBiLi Extended ViralTrans 3 mL-3 mL viral transport medium/vial
M221006KaiBiLi Extended ViralTrans 3 mL with minitip swab-3 mL viral transport medium/vial, with a minitip swab
M221007KaiBiLi Extended ViralTrans 3 mL with regular swab-3 mL viral transport medium/vial, with a regular swab
M221008KaiBiLi Extended ViralTrans 3 mL with regular swab and minitip swab-3 mL viral transport medium/vial, with a regular swab and a minitip swab
M221009KaiBiLi Extended ViralTrans 1 mL-1 mL viral transport medium/vial
M221010KaiBiLi Extended ViralTrans 1 mL with minitip swab-1 mL viral transport medium/vial, with a minitip swab
M221011KaiBiLi Extended ViralTrans 1 mL with regular swab-1 mL viral transport medium/vial, with a regular swab
M221012KaiBiLi Extended ViralTrans 1 mL with regular swab and minitip swab-1 mL viral transport medium/vial, with a regular swab and a minitip swab

Media Formulation:

  • Hank's Balanced Salts
  • HEPES Buffer
  • BSA
  • L-Cysteine .
  • L-Glutamic Acid .
  • · Sucrose
  • Vancomycin
  • Econazole Nitrate
  • Polymyxin B
  • Phenol Red

4. Substantial Equivalence

The KaiBiLi Extended ViralTrans is compared with the predicate device, K042970, in intended use, medium formulation, product configuration, shelf life, packaging and volume, etc. The safety and effectiveness of the KaiBiLi Extended ViralTrans is adequately supported by the substantial equivalence information, materials data, and testing results provided within this Premarket Notification. Below is a summary of comparison table between KaiBiLi Extended ViralTrans and predicate device K042970:

{5}------------------------------------------------

Device & PredicateDevice(s):Device: K231027Predicate: K042970
Device Trade NameKaiBiLi ExtendedViralTransCopan UniversalTransport Medium (UTM-RT) System
General DeviceCharacteristicSimilarities
Intended Use/IndicationsFor UseThe KaiBiLi ExtendedViralTrans is intended forthe collection andtransportation of clinicalspecimens containingviruses, chlamydiae,mycoplasmas andureaplasmas from thecollection site to the testsite. The KaiBiLiExtended ViralTrans is aculture-based mediumthat has been validatedwith multiple sampletypes and can be used toprocess clinicalspecimens usingstandard laboratoryoperating procedures forculture of clinicalspecimens or with otherassays that utilize stablerecoverable infectiousviral particles or bacteria.Copan UniversalTransport Medium (UTM-RT) System is intendedfor the collection andtransport of clinicalspecimens containingviruses, chlamydiae,mycoplasma orureaplasma from thecollection site to thetesting laboratory. UTM-RT can be processedusing standard clinicallaboratory operatingprocedures for viral,chlamydial, mycoplasmaand ureaplasma culture.
Storage Temperature2-8°C and 20-25°CSame
Product configurationMedium tubes; kit withmedium tubes and swaboptionsSame
Single use deviceYesSame
ContainerTube; plastic; self-standing with a screwcap; with glass beadsSame
Shelf Life12 monthsSame
pH7.3 +/-0.2Same
General DeviceCharacteristicDifferences
Media FormulationHank's Balanced SaltsHEPES BufferBSAL-CysteineL-Glutamic AcidSucroseVancomycinEconazole NitratePolymyxin BPhenol RedHank's Balanced SaltsBovine Serum AlbuminL-CysteineGelatinSucroseL-Glutamic AcidHEPES BufferVancomycinAmphotericin BColistinPhenol Red
Supported StrainsAdenovirusCytomegalovirusHerpes Simplex VirusType 1Herpes Simplex VirusType 2Influenza AParainfluenza 3Respiratory SyncytialVirusVaricella Zoster VirusMycoplasmapneumoniaeChlamydia pneumoniaeUreaplasma urealyticumAdenovirusCytomegalovirusEchovirus Type 30Herpes Simplex VirusType 1Herpes Simplex VirusType 2Influenza AParainfluenza 3Respiratory SyncytialVirusVaricella Zoster VirusChlamydia pneumoniaeChlamydia trachomatisMycoplasma hominisMycoplasmapneumoniaeUreaplasma urealyticum
Medium volume1 mL or 3 mL3 mL or 6 mL

{6}------------------------------------------------

As shown with the comparison above, KaiBiLi Extended ViralTrans and the predicate's specification, safety and performance are comparable.

{7}------------------------------------------------

5. Shelf Life

The shelf life for the KaiBiLi Extended ViralTrans was determined to be 12 months from the date of manufacture when stored at temperature 2-8℃ and 20-25℃. The shelf life of the KaiBiLi Extended ViralTrans was established using real-time aging performance test at time points T= 0, 6, 9, 11, and 12 months. Three lots each of the KaiBiLi Extended ViralTrans media at 1 mL and 3 mL configurations (a total of 6 lots) were evaluated. At each time point, appearance, volume, pH, and bacteriostatic performance were assessed.

  • (1) Appearance and volume
    To evaluate appearance, stability of the different lots of KaiBiLi Extended ViralTrans were physically or visually examined. The appearance of the product was observed to be a clear, pink, transparent liquid media with no turbidity or sedimentation. Volume range of 2.7-3.3 mL for the 3 mL media configuration or 0.9-1.1 mL for the 1 mL media configuration were observed. All lots tested at each time point passed the criteria for appearance and net volume.

  • (2) pH Stability
    The pH of the media was used as one of the indicators to support product stability. For all the tubes at each time point and with each lot, the pH was within the targeted pH range of 7.3 ± 0.2 after 12 months of storage.

  • (3) Bacteriostatic performance
    KaiBiLi Extended ViralTrans contains Vancomycin, Econazole Nitrate and Polymyxin B to inhibit growth of bacteria or yeast. At each time point, the KaiBiLi Extended ViralTrans were inoculated with the following microorganisms: E. Coli (1.0x103 CFU/mL), Staphylococcus aureus (1.0x103 CFU/mL), Streptococcus pyogenus (1.0x103 CFU/mL), and Candida albicans (1.0x103 CFU/mL). The KaiBiLi Extended ViralTrans had no microorganism growth after 48 hours growth at 37°C. All lots tested at each time point passed the criteria for bacteriostasis.

6. Performance Testing - Recovery Studies

Performance of the KaiBiLi Extended ViralTrans media was evaluated using culture-based recovery studies for viruses and bacteria at different incubation times and temperatures.

For Recovery Studies, virus titer (TCIDso/mL) was quantified to evaluate the recovery of the following viruses in the corresponding matrices listed in Table 1 below: Herpes Simplex Virus Type 1 (ATCC VR-733; HSV-1), Herpes Simplex Virus Type 2 (ATCC VR-734; HSV-2), Respiratory Syncytial Virus (ATCC VR-26; RSV), Cytomegalovirus (ATCC VR-977), Adenovirus (ATCC VR-3), Parainfluenza 3 (ATCC VR-93), Influenza A (ATCC VR-822; Flu A), and Varicella Zoster Virus (ATCC VR-1832; VZV). Recovery of Chlamydophila pneumoniae (ATCC VR-2282) was evaluated using Fluorescent Foci Count method (IFU/mL). Recovery of Mycoplasma pneumoniae (ATCC 15531) and Ureaplasma urealyticum (ATCC 27618) was evaluated using the Swab Elution and Roll Plate methods (CFU/mL). Performance evaluation was carried out in three lots of media that represent newly manufactured, middle-aged, and recently expired media. Negative clinical matrix pools were contrived from a minimum of five donors. Matrix pools were determined to be negative prior to use in the specimen stability recovery studies.

{8}------------------------------------------------

Negative Clinicalspecimen TypeMicrobial Testing
SkinHerpes Simplex Virus Type 1
SkinVaricella Zoster Virus
Genital specimensHerpes Simplex Virus Type 2
NasopharynxRespiratory Syncytial Virus
NasopharynxAdenovirus
NasopharynxParainfluenza3
NasopharynxInfluenza A
ThroatChlamydophila pneumoniae
ThroatMycoplasma pneumoniae
BloodCytomegalovirus
UrineCytomegalovirus
UrineUreaplasma urealyticum

Table 1: Negative clinical matrix used for organism validation.

Viral stocks were diluted into two different dilutions into the corresponding pooled negative clinical matrix and 100 µL of each dilution was transferred onto a dry sterile swab and placed into the KaiBiLi Extended ViralTrans media tubes in triplicate and stored at 2-8ºC and 20-25°C for 0, 24, and 48 hours. At each incubation time point, the sample was vortexed and a 200µL aliquot was removed for the recovery study. The recovery study was conducted using suitable host cells and tissue culture media. For tissue culture, host cells were plated in a microwell plate and allowed to adhere for 48-72 hours prior to recovery testing. Hep-2 cells were used for HSV-1, RSV, and C. pneumoniae; Vero cells were used for HSV-2; MRC-5 cells were used for Cytomegalovirus and VZV: A549 cells were used for Adenovirus: LLC-MK2 cells were used for Parainfluenza 3; MDCK cells were used for Flu A.

Bacterial stocks were diluted into four different dilutions into the corresponding pooled neqative clinical matrix and 100 µL of each dilution was transferred onto a dry sterile swab and placed into the KaiBiLi Extended ViralTrans media tubes in duplicate and stored at 2-8°C and 20-25°C for 0, 24, and 48 hours. For the swab elution method, at each incubation time point, each sample was vortexed, serially diluted and a 100µL aliquot was removed for the recovery study. The recovery study was conducted using Mycoplasma pneumoniae culture medium for M. pneumoniae and Ureaplasma urealyticum culture medium for U. urealyticum. For the roll-plate method, a single dilution was tested in triplicate by streaking the swab from the various KaiBiLi Extended ViralTrans media tube incubation time point over the agar media specified above and incubating for CFU enumeration.

Viral titer for the viruses and foci counts for C. pneumoniae were evaluated, and CFU was enumerated for M. pneumoniae and U. urealyticum. The average recovery was calculated

{9}------------------------------------------------

as mean viral titer (TCID50/mL), mean foci count (IFU/mL), or mean CFU/mL, respectively, for each storage temperature and time points. The changes (any increase or decrease) in the recovery between time points (each time point compared to time point 0) were presented in percent values or logio change (negative for decrease and positive for increase). Any change that was within one log difference (+/-90%) was considered acceptable. Results were combined for all the lots, irrespective of age, as all changes were acceptable. The results are presented in Tables 2 and 3 below.

Average recoveryPercent observed changes
Test strain(TCID50/mL)(-ve indicate reduction)
HSV-13.53x1032.84x1032.23x103-20%-38%
HSV-24.31x1033.54x1032.37x103-18%-46%
RSV1.29x1041.13x1047.14x103-14%-45%
Cytomegalovirus¹5.89x1034.75x1033.32x103-19%-44%
Cytomegalovirus25.87x1034.76x1033.26x103-19%-45%
Adenovirus1.31x1051.06x1058.53x104-20%-36%
Parainfluenza 32.70x1042.19x1041.83x104-19%-33%
Flu A1.46x1041.19x1049.71x103-19%-35%
VZV1.25x1031.08x1037.37x102-14%-41%
Test strainAverage recoveryPercent observed changes
(IFU/mL)(-ve indicate reduction)
C. pneumoniae3.40x1052.72x1052.12x105-21%-39%
Test strainAverage recovery using swab elution methodLog10 observed changes
(CFU/mL)(-ve indicate reduction)
M. pneumoniae6.36x1045.89x1042.43x104-0.03-0.45
U. urealyticum6.74x1045.81x1042.19x104-0.08-0.53
Test strainAverage recovery using roll plate methodLog10 observed changes
(CFU)(-ve indicate reduction)
M. pneumoniae2.97x1022.18x1021.13x102-0.13-0.42
U. urealyticum2.98x1022.11x1029.80x101-0.15-0.48

Table 2: Recovery of viruses and bacteria at 2-8°C storage

1 Cytomegalovirus recovery in blood; 2 Cytomegalovirus recovery in urine

{10}------------------------------------------------

Table 3. Recovery of viruses and bacteria at 20-25°C storage
Test strainAverage recovery (TCID50/mL)Percent observed changes (-ve indicate reduction)
0 hrs24 hrs48 hrs0-24 hrs0-48 hrs
HSV-13.53x10³2.83x10³2.14x10³-21%-41%
HSV-24.31x10³3.49x10³2.30x10³-20%-48%
RSV1.29x10⁴1.02x10⁴6.76x10³-21%-48%
Cytomegalovirus¹5.89x10³4.71x10³3.22x10³-20%-45%
Cytomegalovirus²5.87x10³4.73x10³3.07x10³-20%-48%
Adenovirus1.31x10⁵1.05x10⁵7.89x10⁴-21%-41%
Parainfluenza 32.70x10⁴2.15x10⁴1.78x10⁴-20%-35%
Flu A1.46x10⁴1.17x10⁴9.39x10³-20%-38%
VZV1.25x10³1.04x10³6.86x10²-17%-45%
Test strainAverage recovery (IFU/mL)Percent observed changes (-ve indicate reduction)
0 hrs24 hrs48 hrs0-24 hrs0-48 hrs
C. pneumoniae3.40x10⁵2.69x10⁵2.09x10⁵-21%-39%
Test strainAverage recovery using swab elution method (CFU/mL)Log₁₀ observed changes (-ve indicate reduction)
0 hrs24 hrs48 hrs0-24 hrs0-48 hrs
M. pneumoniae6.36x10⁴5.92x10⁴2.40x10⁴-0.03-0.46
U. urealyticum6.74x10⁴5.60x10⁴2.04x10⁴-0.10-0.56
Test strainAverage recovery using roll plate method (CFU)Log₁₀ observed changes (-ve indicate reduction)
0 hrs24 hrs48 hrs0-24 hrs0-48 hrs
M. pneumoniae2.47x10²2.09x10²1.03x10²-0.15-0.46
U. urealyticum2.98x10²1.74x10²9.40x10¹-0.23-0.50

Table 3: Recovery of viruses and bacteria at 20-25°C storage

1 Cytomegalovirus recovery in blood; 2 Cytomegalovirus recovery in urine

7. Conclusion

The KaiBiLi Extended ViralTrans media demonstrated the recovery of tested viruses (HSV-1, HSV-2, RSV, Cytomegalovirus, Adenovirus, Parainfluenza, Flu A, and VZV) and bacteria (C. pneumoniae, M. pneumoniae, and U. urealyticum) at an acceptable rate when stored at 2-8°C and 20-25°C up to 48 hours.

Based on the indications for use, technological characteristics, safety and performance testing, the subject device, the KaiBiLi Extended ViralTrans, meets the requirements that are considered essential for its intended use and is substantially equivalent to the legally marketed predicate device, Copan Universal Transport Medium (UTM-RT) System , K042970.

§ 866.2390 Transport culture medium.

(a)
Identification. A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.(b)
Classification. Class I (general controls).