K Number
K230985

Validate with FDA (Live)

Device Name
Planmeca Viso
Manufacturer
Date Cleared
2023-12-28

(266 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Planmeca Viso is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital X-ray images as well as three-dimensional (3D) optical images of the dento-maxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.

Device Description

The Planmeca Viso -X-ray unit uses cone beam computed tomography (CBCT) to produce three-dimensional (3D) images of the maxillofacial and ENT anatomies. Two dimensional (2D) images are produced with tomosynthesis method (panoramic) imaging) as well as conventional 2D radiography (cephalometric imaging, 2D views). In CBCT a cylindrical volume of data is captured in one imaging procedure. The data consists of several hundred sample images which are taken from different directions to cover a certain pre-programmed target area. These samples are used for 3D reconstruction (using a dedicated 3D reconstruction hardware) that can be viewed in three dimensions using separate workstation and Planmeca Romexis software.

AI/ML Overview

The provided text does not contain detailed acceptance criteria or a comprehensive study report for the Planmeca Viso device that would allow for a complete response to all aspects of your request. Specifically, the document focuses on demonstrating substantial equivalence to a predicate device through technical comparisons and general performance claims, rather than detailing specific quantitative acceptance criteria for image quality or clinical performance and exhaustive evidence proving they are met.

However, based on the information provided, here's a partial response:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Image Quality (General)"suitable for the intended purpose and indications for use of the device"
Clinical Performance"substantially equivalent to the evaluation performed on primary predicate device"
AI Denoising Feature (Endodontic Image Processing)"produce diagnostic image quality in its intended application"
Electrical SafetyComplies with IEC 60601-1+A1:2012+A2:2020
Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2+A1:2020
Basic Safety and Essential Performance of X-ray EquipmentComplies with IEC 60601-1-3+A1:2013+A2:2021
Usability EngineeringComplies with IEC 60601-1-6+A1:2013+A2:2020
Safety of Dental X-ray EquipmentComplies with IEC 60601-2-63+A1:2017+A2:2021
Usability of Medical DevicesComplies with IEC 62366-1+A1:2020
Medical Device Software Life Cycle ProcessesComplies with IEC 62304+A1:2015
Image Quality (Cone Beam CT - CBCT)Similar diagnostic value to predicate device due to similar detector qualities and CT reconstruction algorithm.
Image Quality (Pan / ProCeph)Similar diagnostic value to predicate device
Performance (Bench Testing - Sedentex & DIN 6868 Phantoms)"performance of the device remains substantially similar to that of the primary predicate device" and "performs equally or better in all the testing scenarios."

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: For the AI Denoising feature evaluation, the study included eleven patients.
  • Data Provenance: Not explicitly stated, but the submission is from a Finnish company (Planmeca Oy), suggesting the data could originate from Finland or other European countries. The study appears to be prospective for the AI denoising evaluation, as it explicitly describes "comparing images with no denoising and AI denoising" for 11 patients, implying new data collection. For the general clinical evaluation, it mentions evaluation of "human phantom images," which could be retrospective or simulated, but again, details are lacking.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • Number of Experts: Three dental professionals were involved in the evaluation of the AI Denoising feature.
  • Qualifications of Experts: The specific qualifications (e.g., years of experience, subspecialty) of these dental professionals are not specified in the provided text.

4. Adjudication Method for the Test Set

  • The adjudication method for the AI Denoising study is not explicitly stated. It mentions that a team of dental professionals "evaluated human phantom images" and that the AI denoising was "evaluated in a study performed by three dental professionals." It does not specify if they reached a consensus, if a majority vote was used, or if there was an independent adjudicator.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • A formal MRMC comparative effectiveness study comparing human readers with AI assistance vs. without AI assistance is not explicitly described in the provided text.
  • The clinical evaluation mentions that the device's image quality was evaluated by "product relevant team of professionals who evaluated human phantom images," and this was "substantially equivalent to the evaluation performed on primary predicate device."
  • The AI Denoising feature was evaluated by three dental professionals "by comparing images with no denoising and AI denoising," which is a comparative aspect. However, it doesn't quantify an "effect size of how much human readers improve with AI vs without AI assistance" in terms of diagnostic performance metrics. It only states the AI improved images to "produce diagnostic image quality."

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • The AI Denoising feature was evaluated based on its "ability to produce diagnostic image quality in its intended application." While the evaluation involved human professionals comparing images, the goal was to assess the algorithm's output (denoised images). However, a purely standalone assessment without human input or comparison for decision-making is not explicitly detailed. The evaluation seems to be focused on the impact of the AI on image quality for human interpretation.

7. Type of Ground Truth Used

  • For AI Denoising: The ground truth appears to be based on expert assessment/consensus (three dental professionals) regarding the "diagnostic image quality" of images with and without AI denoising.
  • For General Image Quality: The clinical evaluation involved "human phantom images" assessed by "product relevant team of professionals," suggesting a form of expert assessment against an assumed standard of diagnostic suitability.
  • For Bench Testing: The ground truth for bench testing (Sedentex and DIN 6868 Phantoms) would be the physical properties and known characteristics of the phantoms, with the device's performance measured against those.

8. Sample Size for the Training Set

  • The sample size for the training set used for the AI Denoising feature (or any other AI component/algorithm) is not mentioned in the provided text.

9. How the Ground Truth for the Training Set Was Established

  • How the ground truth for the training set (if any was used for AI model development) was established is not mentioned in the provided text.

{0}------------------------------------------------

December 28, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

Planmeca Oy Niina Vuorikallas Director, Quality & Regulatory Affairs Asentajankatu 6 Helsinki. 00880 Finland

Re: K230985

Trade/Device Name: Planmeca Viso Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS Dated: October 17, 2023 Received: November 3, 2023

Dear Niina Vuorikallas:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gabriela M. Rodal Digitally signed by for Gabriela M. Rodal -S -2

Lu Jiang, Ph.D. Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K230985

Device Name

Planmeca Viso

Indications for Use (Describe)

Planmeca Viso is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital X-ray images as well as three-dimensional (3D) optical images of the dento-maxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. SUBMITTER

Manufacturer

Planmeca Oy Asentajankatu 6 00880 Helsinki, Finland Phone: +358 20 7795 500 Fax: +358 20 7795 555 Contact person: Niina Vuorikallas

U.S. designated agent

Planmeca U.S.A. Inc. 2600 Forbs Ave. Hoffman Est, IL 60192 UNITED STATES Contact person: Glen Kendrick

Date Prepared: December 27, 2023

II. DEVICE

Name of Device:Planmeca Viso
Common or Usual Name:Computed Tomography X-ray System
Classification Name:Computed Tomography X-ray System (CT) (21 CFR892.1750)
Regulatory Class:II
Product Code:OAS

III. PREDICATE DEVICE

Primary predicate device, #1, Planmeca Viso, Computed Tomography System 510(k): K181576 Regulation number: 892.1750 Regulatory Class: II Product Code: OAS

Predicate device #2, HDX DENTRIα. Computed Tomography System 510(k): K160140 Regulation number: 892.1750 Regulatory Class: II Product Code: OAS

Predicate device #3, DigiX FDX, Stationary X-ray system 510(k): K223060 Regulation number: 892.1680 Regulatory Class: II Product Code: KPR

{4}------------------------------------------------

Predicate device #2 is only used for X-ray tube comparison. Outside of figure 1, when referred to predicate device, the reference is made to primary predicate device, #1.

DEVICE DESCRIPTION IV.

The Planmeca Viso -X-ray unit uses cone beam computed tomography (CBCT) to produce three-dimensional (3D) images of the maxillofacial and ENT anatomies. Two dimensional (2D) images are produced with tomosynthesis method (panoramic) imaging) as well as conventional 2D radiography (cephalometric imaging, 2D views). In CBCT a cylindrical volume of data is captured in one imaging procedure. The data consists of several hundred sample images which are taken from different directions to cover a certain pre-programmed target area. These samples are used for 3D reconstruction (using a dedicated 3D reconstruction hardware) that can be viewed in three dimensions using separate workstation and Planmeca Romexis software.

INDICATIONS FOR USE V.

Same as predicate device.

Planmeca Viso is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital x-ray images as well as three-dimensional (3D) optical images of the dentomaxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

General

Both devices are 3D CBCT imaging system indicated for imaging dento-maxillofacial-. cervical spine- and ENT-regions and offer a variety of 3D fields of view suitable for each application.

Both systems also provide 2D imaging capabilities for panoramic and cephalometric imaging. In general there are only a few changes in the device.

Integrated detector

The X-ray detectors used are of same size as previously. Viso G5 and ProCeph use new X-ray detectors with similar qualities as the predicate device. Only difference between Viso G7 predicate and subject devices is the interface. Predicate device uses proprietary interface, and subject device uses standard ethernet interface. This has some effects on the device hardware, namely the grabber PCB, which is being replaced with a commercial Nvidia Jetson Nano module and suitable interface PCB. Otherwise they are of similar qualities as before and the produced 3D images provide similar diagnostic value. See fig 2 for technical comparison between devices.

X-ray unit

The dimensions and materials are identical to the predicate device.

X-ray tube

A new x-ray tube has been introduced to the device. It is of similar quality as predicate devices' fixed anode x-ray tubes, and it has maximum tube voltage of 120kV. Previously used fixed anode x-ray tubes may also be used with this device. No other changes for the tube head. Predicate device #2 operates a similar tube manufactured by the same company, the only difference being the target angle and thus the heat characteristics. See fig 1 for technical comparison on X-ray tube properties.

fig 1. X-ray tube comparison

{5}------------------------------------------------

Subject devicePrimary predicatedevice, #1Predicate device #2
Tube typeOPX/105-10SXR 130-10-0.5 SCOPX/105
Nominal voltage120 kV130 kV110 kV
Focal spot0.5 mm0.5 mm0.5mm
Filament characteristics3.5 ÷ 6.0 V2.5 ÷ 4.0 AN/A3.5 ÷ 6.0 V2.5 ÷ 4.0 A
Anode materialTungstenTungstenTungsten
Target angle10°10°
Anode heat storagecapacity40000 J31500 J30000 J
Maximum anode coolingrate500 W310 W250 W
Nominal anode inputpower at 0.1 s (DC)1900 WN/A2000 W
Inherent filtration0.5 mm Al1.1 mm Al0.5 mm Al

X-ray generator

The device uses the same x-ray generator as before.

X-ray collimator

The device uses the same x-ray collimator as before.

Software

Software changes made to accommodate new hardware (see integrated detectors and new xray tube).

CT reconstruction algorithm

Device uses same CT reconstruction algorithm as before, no significant changes are made to the algorithm.

fig 2. technical comparison
Technical specificationSubject devicePrimary predicate device, #1
Classification
FDA product classOAS, class IIOAS, class II
RoHS2011/65/EU2011/65/EU
IEC 60601-1Class I, type BClass I, type B
CISPR 11Class BClass B
IP ClassificationIPX0IPX0
Applied parts(according to IEC 60601-1: 2012)
Patient supportsAs shown in section Patientsupports in user's manualsAs shown in section Patientsupports in user's manuals
Patient handles
Generator(according to IEC 60601-2-7:1998)
Resonant-mode, DSP-controlled, 80 - 160 kHzResonant-mode, DSP-controlled, 80 - 160 kHz
X-ray tube
D-059SBR or SXR 130-10-0.5 SC or OPX 105-10D-059SBR or SXR 130-10-0.5 SC
Focal spot size(according to IEC 60336: 2005)
0.5 x 0.5 mm0.5 x 0.5 mm
Filtration
3DTotal 2.5 mm Al + 0.2mm / 0.5 mm CuTotal 2.5 mm Al + 0.2mm / 0.5 mm Cu
Pan (SmartPan) / ProCephTotal 2.5 mm AlTotal 2.5 mm Al
Tube housing front coverquality equivalent filtration(not included in the specifiedtotal filtration)0.3 mm Al @ 70 kV / HVL 2.6 mm Al0.3 mm Al @ 70 kV / HVL 2.6 mm Al
Anode voltage
3D80 - 120 kV ±5%80 - 120 kV ±5%
Pan (SmartPan)60 - 84 kV ±5%60 - 84 kV ±5%
ProCeph60 - 84 kV ±5%60 - 84 kV ±5%
Anode current
3DD-059SBR: 2-12.5 mA ±10%SXR 130-10-0.5 SC: 2-16mA ±10%OPX 105-10: 2-16mA ±10%D-059SBR: 1-12.5 mA ±10%SXR 130-10-0.5 SC: 1-16mA ±10%
Pan (SmartPan)D-059SBR: 2-14 mA ±10%SXR 130-10-0.5 SC: 2-16mA ±10%OPX 105-10: 2-16mA ±10%D-059SBR: 1-14 mA ±10%SXR 130-10-0.5 SC: 1-16mA ±10%
ProCephD-059SBR: 14 mA ±10%SXR 130-10-0.5 SC: 16mA ±10%OPX 105-10: 14mA ±10%D-059SBR: 14 mA ±10%SXR 130-10-0.5 SC: 16mA ±10%
mAs range
min. / max. as indicated ±(10% + 0.2 mAs)min. / max. as indicated ±(10% + 0.2 mAs)
Dose range and accuracy
Dose range min. / max. as indicated on system user interfaceAccuracy of dosimetric indication (DAP, CTDI): ±40%Dose range min. / max. as indicated on system user interfaceAccuracy of dosimetric indication (DAP, CTDI): ±40%
Linearity of radiation output
< 0.1< 0.1
Exposure time
3DPulsed, effective 1.5 - 36 s asindicated ±10%Pulsed, effective 1.5 - 36 s asindicated ±10%
Pan (SmartPan)2.5 - 15.6 s as indicated ±10%2.5 – 15.6 s as indicated ±10%
ProCeph0.1 - 1.6s as indicated ±10%0.1 - 1.6s as indicated ±10%
SID
3D / Pan (SmartPan)700 mm700 mm
Ceph1700 mm (66.9 in.)1700 mm (66.9 in.)
Magnification
3D1.40 - 1.711.40 - 1.71
Pan (SmartPan)1.321.40
Ceph1.131.13
Duty cycle for height adjustment
25 s ON / 400 s OFF25 s ON / 400 s OFF
Line voltage
100 - 220 V~ / 50 - 60 Hz230 - 240 V~ / 50 Hz100 - 220 V~ / 50 - 60 Hz230 - 240 V~ / 50 Hz
Line current
8 - 17 A8 - 17 A
Input power
Stand by150 VA150 VA
Exposure1800 W1800 W
Line harmonics
Cos better than 0.9Cos better than 0.9
Max. permissible apparentimpedance of supply mains
0.5 Ohms (100 VAC)0.5 Ohms (100 VAC)
Max. continuous heat dissipation
250 W250 W
Internal fuse(s)
User replaceable- 1 fuse on permanentlyinstalled X-ray units100 - 220 V~ / 16A FF H 500 V230 - 240 V~ / 8A FF H 500 V100 - 220 V~ / 16A FF H 500 V230 - 240 V~ / 8A FF H 500 V
Type195100 ELU195100 ELU
External fuse(s)
100 - 220 V ~ / 16A min. - 20Amax. T 250 V230 - 240 V ~ / 10A min. - 20Amax. T 250 V100 - 220 V ~ / 16A min. - 20Amax. T 250 V230 - 240 V ~ / 10A min. - 20Amax. T 250 V
Battery
Lithium battery: 3V, CR2032Panasonic / VartaLithium battery: 3V, CR2032Panasonic / Varta
Max. weight
Base unit165 kg (364 lb)165 kg (364 lb)
ProCeph20 kg (44 lb)20 kg (44 lb)
Environmental requirements
Transport:
Temperature-20°C - +60°C (-4°F - +140°F)-20°C - +60°C (-4°F - +140°F)
Relative humidity10 - 90% RH (non-condensing)10 - 90% RH (non-condensing)
Air pressure700 - 1060 hPa700 - 1060 hPa
Storage:
Temperature-10°C - +50°C (+14°F - +122°F)-10°C - +50°C (+14°F - +122°F)
Relative humidity10 - 90% RH (non-condensing)10 - 90% RH (non-condensing)
Air pressure700 - 1060 hPa700 - 1060 hPa
Operating:
Temperature+10°C - +30°C (+50°F - +86°F)+10°C - +30°C (+50°F - +86°F)
Relative humidity10 - 90% RH (non-condensing)10 - 90% RH (non-condensing)
Air pressure800 - 1060 hPa800 - 1060 hPa
Max. altitude2000 m (1.25 miles)2000 m (1.25 miles)
Image properties
ProCeph:
Panel typeVarex 2530C Gen5Varex 2530P
Flat panel pixel size131 μm139 μm
Flat panel active surface302 x 249 mm (11.89 x 9.80 in.)302 x 249 mm (11.89 x 9.80 in.)
DQE (0)64 %32 %
MTF> 51 % @ 1lp/mm> 48 % @ 1lp/mm
3D:
Panel typeViso G5: Varex 1616ZViso G7: Varex 2530DXViso G5: Varex 1616PTViso G7: Varex 2530PX
Flat panel pixel sizeViso G5: 105 μmViso G7: 139 μmViso G5: 127 μmViso G7: 139 μm
Flat panel active surfaceViso G5: 161.3 x 161.3 (6.35 x 6.35in.)Viso G7: 247.7x 301.1 mm (9.75 x11.85 in.)Viso G5: 157.5 x 157.5 mm (6.20 x6.20 in.)Viso G7: 299.7x 246.3 mm (11.80x 9.70 in.)
DQE (0)Viso G5: 78 % (1x1)Viso G7: 70 % (1x1)Viso G5: 70 % (1x1)Viso G7: 70 % (1x1)
MTFViso G5: > 59 % @ 1lp/mm (1x1)Viso G7: > 48 % @ 1lp/mm (1x1)Viso G5: > 48 % @ 1lp/mm (1x1)Viso G7: > 50 % @ 1lp/mm (1x1)
Voxel sizes75 μm, 150 μm, 300 μm, 450 μm,600 μm75 μm, 150 μm, 300 μm, 450 μm,600 μm
Pan (SmartPan):
Panel typeViso G5: Varex 1616ZViso G7: Varex 2530DXViso G5: Varex 1616PTViso G7: Varex 2530PX
Flat panel pixel sizeViso G5: 105 μmViso G7: 139 μmViso G5: 127 μmViso G7: 139 μm
Flat panel active surfaceViso G5: 8.4 x 161.3 mm (0.33 x6.35 in.)Viso G7: 8.9/17.8 x 166.8 mm(0.35/0.7 x 6.57 in.)8 - 25 x 146 mm (0.31 - 0.98 x5.74 in.)
Operating requirements forProFace program
Optimum colour temperatureApprox. 6500 KelvinApprox. 6500 Kelvin
Even and uniform lighting
No bright lights

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

PERFORMANCE DATA VII.

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

No changes in patient contact parts compared to primary predicate device. Electrical safety and electromagnetic compatibility (EMC)

{10}------------------------------------------------

CB and EMC testing was conducted as per IEC standards 60601-1+A1:2012+A2:2020, 60601-1-2+A1:2020, 60601-1-3+A1:2013+A2:2021, 60601-1-6+A1:2013+A2:2020, 60601-2-63+A1:2017+A2:2021,62366-1+A1:2020 and 62304+A1:2015

Software Verification and Validation Testing

Software verification and validation was performed as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance of Premarket Submissions for Software Contained in Medical Devices.". No issues have been identified during the verification and validation process.

Bench testing

Performance testing in laboratory environment was performed with Viso (primary predicate device) and Viso (subject device) of the Sedentex and DIN 6868 Phantoms. The data was compared in order to see that the performance of the device remains substantially similar to that of the primary predicate device.

Clinical evaluation

The main measure of clinical performance for Planmeca Viso is determined to be image quality. This has been evaluated by the product relevant team of professionals who evaluated human phantom images which were determined to be suitable for the intended purpose and indications for use of the device. These results are substantially equivalent to the evaluation performed on primary predicate device.

AI Denoising feature of the endodontic image processing has been evaluated based on its ability to produce diagnostic image quality in its intended application. This was evaluated in a study performed by three dental professionals in a study conducted with eleven patients, by comparing images with no denoising and AI denoising.

Testing summary

The above testing confirms that Planmeca Viso is safe and effective in its intended use.

VIII. CONCLUSIONS

The performance data of the subject device and the predicate device is equal or close to equal. The performance testing provides data to back this similarity in the devices. The subject device performs equally or better in all the testing scenarios.

The IGZO technology is proven to bring improvements in electronic noise, image lag and noise-equivalent quanta. The IGZO technology has been used in several FDA approved devices at least since 2019.

The tube specifications regarding eg. focus size, anode angle and tube voltage are the same as previously approved Canon D-059SB X-ray tube. The OPX 105-10 X-ray tube is driven with the same hardware as our other previously approved X-ray tubes.

The comparison of characteristics supports substantial equivalence. Planmeca Viso is as safe and effective as the predicate device.

N/A