K Number
K230910

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-06-16

(77 days)

Product Code
Regulation Number
886.5916
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The New Vision SC Daily Wear Ortho-K (tisilfocon A) Corneo-Scleral Lens is indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.

Eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. The New Vision SC (tisilfocon A) Scleral Lens is indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:

  1. cannot be adequately corrected with spectacle lenses
  2. requires a rigid gas permeable contact lens surface to improve vision
  3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities

Common causes of corneal distortion include but are not limited to corneal infections, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmam's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy).

The New Vision SC (tisilfocon A) Scleral Lens is indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy. ectodermal dysplasia), neurotrophic keratiis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the New Vision SC (tisilfocon A) Scleral Lens may concurrently provide correction of up refractive error.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

Device Description

The NewVision SC (tisilfocon A) Scleral Lens is a large diameter rigid gas permeable lens designed to vault over the cornea and rest on the conjunctiva overlying the sclera. The New Vision SC Daily Wear Ortho-K (tisilfocon A) Corneo-Scleral Lens partly rests on the cornea, centrally or peripherally, and partly on the limbus or conjunctiva over the sclera. The NewVision SC and New Vision SC Daily Wear Ortho-K (tisilfocon A) Lenses are lathe cut from tisilfocon A, which is an FDA Group #3 fluoro-silicone acrylate material.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device (contact lenses) and does not contain information about an AI/ML-based device or a study involving AI/ML performance. Therefore, I cannot extract the required details such as acceptance criteria, performance metrics, sample sizes for test and training sets, expert qualifications, ground truth establishment, or MRMC studies, as these concepts are not applicable to the content of this document.

The document discusses the substantial equivalence of the NewVision SC and NewVision SC Daily Wear Ortho-K (tisilfocon A) Lenses to a legally marketed predicate device (Optimum Infinite (tisilfocon A) Daily Wear Contact Lenses). The performance data cited are "Non-Clinical Studies" (bench testing and bioburden testing) and "Clinical Studies" based on previous 510(k) clearances, rather than a new clinical study.

Therefore, I cannot provide the requested table or answer the specific questions related to an AI/ML device study.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.

June 16, 2023

Acculens, Inc. % Bret Andre Principal Consultant Eyereg Consulting Inc. 6119 Canter Lane West Linn, OR 97068

Re: K230910

Trade/Device Name: NewVision SC (tisilfocon A) Scleral Lens: NewVision SC Daily Wear Ortho-K (tisilfocon A) Corneo-Scleral Lens Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HQD, MUW Dated: March 31, 2023 Received: March 31, 2023

Dear Bret Andre:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

J Angelo Green -S

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230910

Device Name

NewVision SC (tisilfocon A) Scleral Lens; NewVision SC Daily Wear Ortho-K (tisilfocon A) Corneo-Scleral Lens

Indications for Use (Describe)

The New Vision SC Daily Wear Ortho-K (tisilfocon A) Corneo-Scleral Lens is indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.

Eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. The New Vision SC (tisilfocon A) Scleral Lens is indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:

    1. cannot be adequately corrected with spectacle lenses
    1. requires a rigid gas permeable contact lens surface to improve vision
    1. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities

Common causes of corneal distortion include but are not limited to corneal infections, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmam's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy).

The New Vision SC (tisilfocon A) Scleral Lens is indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy. ectodermal dysplasia), neurotrophic keratiis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the New Vision SC (tisilfocon A) Scleral Lens may concurrently provide correction of up refractive error.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510 (k) SUMMARY

This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:K230910
I.SUBMITTER
Date Prepared:June 6, 2023
Name:AddressAcculens, Inc.5353 West Colfax AvenueLakewood, Colorado 80214United States
Contact Person:Phone number:Troy MillerVP of Operations and Product Development(303) 232-6244
Consultant/Correspondent:Phone numberEyeReg Consulting, Inc.Bret Andre6119 Canter LaneWest Linn, OR 97068(503) 372-5226
II.DEVICE
Trade Name:NewVision SC (tisilfocon A) Scleral Lens;NewVision SC Daily Wear Ortho-K (tisilfocon A)Corneo-Scleral Lens
CommonName:Daily wear rigid gas permeable contact lens
ClassificationName:Rigid gas permeable contact lens. (21 CFR 886.5916)
RegulatoryClass:Class II
Product Code:HQD; MUW

{5}------------------------------------------------

III. PREDICATE DEVICE

The NewVision SC and NewVision SC Daily Wear Ortho-K (tisilfocon A) Lenses are substantially equivalent to the following predicate device:

  • . "Optimum Infinite (tisilfocon A) Daily Wear Contact Lenses" Manufactured by Contamac Ltd. 510(k) number; K212631

IV. DEVICE DESCRIPTION

The NewVision SC (tisilfocon A) Scleral Lens is a large diameter rigid gas permeable lens designed to vault over the cornea and rest on the conjunctiva overlying the sclera. The New Vision SC Daily Wear Ortho-K (tisilfocon A) Corneo-Scleral Lens partly rests on the cornea, centrally or peripherally, and partly on the limbus or conjunctiva over the sclera. The NewVision SC and New Vision SC Daily Wear Ortho-K (tisilfocon A) Lenses are lathe cut from tisilfocon A, which is an FDA Group #3 fluoro-silicone acrylate material. The physical properties of the New Vision SC and NewVision SC Daily Wear Ortho-K (tisilfocon A) Lenses are as follows:

TISILFOCON A
Refractive Index1.4378
Light Transmission (tinted)>91%
Water Content<1%
Oxygen Permeability(Dk)ISO/FATT Method$180 x 10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg @ 35°C)
Contain one or more of the following color additives conforming to:21 CFR Part 73 & 74, Subpart DD&C Green No. 6, C.I. Solvent Yellow No. 18, D&C Violet No. 2 andD&C Red No. 17
UV Light Blocking(UVB - 280nm – 315nm; UVA 316nm – 380nm)>98% UVB>85% UVA

The NewVision SC (tisilfocon A) Scleral Lens is available in the following parameters:

ParameterRangeTolerance
Base Curve5.5mm to 25.00mm± 0.2mm
Center Thickness0.10mm to 3.00mm± 0.1mm
Diameter12.00mm to 26.00mm± 0.20mm
Spherical Power-35.00 D to +35.00 D(in .12D steps)± 0.12 (0 to = 5D)<br ± 0.18 (5 to = 10.0D)<br ± 0.25 (10 to = 15D)<br ± 0.37 (15 to = 20D)<br ± 0.50 (over 20D)
Cylindrical Power+10.00 D to -10.00 D(in .12 D steps)± 0.25 (0 to = 2D)<br ± 0.37 (2 to = 4D)<br ± 0.50 (over 4D)
Cylindrical Axis1º to 180º (in 1º steps)± 5°
Bifocal Add+.12 D to +6.00 D(in .12 D steps)± 0.25D

{6}------------------------------------------------

The NewVision SC Daily Wear Ortho-K (tisilfocon A) Corneo-Scleral Lens for daily wear orthokeratology is available in following lens parameters:

ParameterRangeTolerance
Base Curve (BC)4.0mm to 12.00mm± 0.05 mm
Center Thickness0.10mm to 0.70mm± 0.02 mm
Diameter6.50mm to 13.50mm± 0.10mm
Secondary Curves0.10mm to 2.00mm (flatter or steeper than BC)± 0.10mm
Peripheral Curves0.10mm to 2.00mm (flatter or steeper than BC)± 0.10mm
Spherical Power-10.00 D to +4.00 D(in 0.25D steps)± 0.12 (0 to = 5D)<br ± 0.18 (5 to = 10.0D)<br ± 0.25 (10 to = 15D)<br ± 0.37 (15 to = 20D)<br ± 0.50 (over 20D)

The New Vision SC and New Vision SC Daily Wear Ortho-K (tisilfocon A) Lenses are shipped "dry" in the Bausch + Lomb Boston® Scleral Lens Case.

V. INDICATIONS FOR USE

The New Vision SC Daily Wear Ortho-K (tisilfocon A) Corneo-Scleral Lens is indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.

Eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. The New Vision SC (tisilfocon A) Scleral Lens is indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:

    1. cannot be adequately corrected with spectacle lenses
    1. requires a rigid gas permeable contact lens surface to improve vision
  • is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface 3. irregularities

Common causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dvstrophy, Cogan's dystrophy).

The NewVision SC (tisilfocon A) Scleral Lens is indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted

{7}------------------------------------------------

cornea or ocular surface disease, the New Vision SC (tisilfocon A) Scleral Lens may concurrently provide correction of refractive error.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE

The NewVision SC and NewVision SC Daily Wear Ortho-K (tisilfocon A) Lenses are

substantially equivalent to the Optimum Infinite (tisilfocon A) Daily Wear Contact Lenses (predicate device - K212631) in the following areas:

  • FDA material group group # 3 fluoro silicone acrylate ● Components/Materials/Formulation (tisilfocon A)
  • Product code (HQD, MUW) ●
  • Classification (Class II) Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916) ●
  • Indications for use (therapeutic - irregular cornea and ocular surface disease, orthokeratology)
  • Lathe cut manufacturing process
  • Scleral (large diameter) design ●
  • Actions and intended use ●

{8}------------------------------------------------

The following table depicts the classification and technical characteristics of the NewVision SC and NewVision SC Daily Wear Ortho-K (tisilfocon A) Lenses in comparison with the predicate devices.

NewVision SC and NewVision SC DailyWear Ortho-KOptimum InfiniteDaily Wear Contact Lenses
Subject DevicePredicate Device
510(k) NumberK212631
Intended UseDaily WearDaily Wear
Device andClassificationClass IILenses, Rigid Gas Permeable, DailyWearClass IILenses, Rigid Gas Permeable, Daily Wear
Product CodeHQD; MUWHQD; MUW
Production MethodLathe-cutLathe-cut
Material (USAN)Tisilfocon ATisilfocon A
FDA Group #Group # 3 Fluoro Silicone AcrylateGroup # 3 Fluoro Silicone Acrylate
Water Content<1%<1%
UV Absorber/Blocker availableYESYES
Lens DesignsLarge Diameter (Irregular Cornea)Large Diameter (Ocular SurfaceDisease)Daily Orthok (Corneo-Scleral)Large Diameter (Irregular Cornea)Large Diameter (Ocular Surface Disease)Daily Orthok (Corneal)SphericalToricMultifocal/Bifocal
Indications for Use
NewVision SCand New VisionSC Daily WearOrtho-K(SubjectDevice)The NewVision SC Daily Wear Ortho-K (tisilfocon A) Corneo-Scleral Lens is indicated for daily wear inan orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on aprescribed wearing schedule.Eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueoushydrated environment and the saline bath provided by scleral lens designs. The NewVision SC (tisilfocon A)Scleral Lens is indicated for therapeutic use for the management of irregular and distorted corneal surfaceswhere the subject:1. cannot be adequately corrected with spectacle lenses2. requires a rigid gas permeable contact lens surface to improve vision3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularitiesCommon causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or cornealtransplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucidmarginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy,granular corneal dystrophy, Reis-Bucklersdystrophy, Cogan's dystrophy).The NewVision SC (tisilfocon A) Scleral Lens is indicated for therapeutic use in eyes with ocular surfacedisease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary
OptimumInfiniteDaily WearContact Lenses(PredicateDevice:K212631)Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the NewVision SC (tisilfocon A) Scleral Lens may concurrently provide correction of refractive error.Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
The Optimum Infinite (tisilfocon A) SPHERICAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia.
The Optimum Infinite (tisilfocon A) TORIC Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10.00 diopters.
The Optimum Infinite (tisilfocon A) MULTIFOCAL/BIFOCAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
The Optimum Infinite (tisilfocon A) IRREGULAR CORNEA Daily Wear Contact Lens may be prescribed in otherwise non-diseased eyes that require a rigid gas permeable lens for the management of irregular corneal conditions such as; keratoconus, pellucid marginal degeneration or following penetrating keratoplasty or refractive (e.g. LASIK) surgery.
The Optimum Infinite (tisilfocon A) ORTHOKERATOLOGY contact lenses are indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs.Optimum Infinite (tisilfocon A) SCLERAL lenses are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:4. cannot be adequately corrected with spectacle lenses5. requires a rigid gas permeable contact lens surface to improve vision6. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities
Common causes of corneal distortion include but are not limited to corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy).
The Optimum Infinite (tisilfocon A) SCLERAL lenses are indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the Optimum Infinite (tisilfocon A) SCLERAL lenses may concurrently provide correction of refractive error.Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

{9}------------------------------------------------

{10}------------------------------------------------

VII. PERFORMANCE DATA

~ Non-Clinical Studies ~

Non-clinical testing to demonstrate the safety and effectiveness of contact lenses manufactured from tisilfocon A has been addressed by reference to previous 510(k) clearances.

Additionally, the following testing was performed on finished NewVision SC and NewVision SC Daily Wear Ortho-K (tisilfocon A) Lenses:

Bench Testing-manufacturing verification testing was conducted to demonstrate the ability of Acculens, Inc. to manufacture lenses, on a repeatable basis, from supplied lens blanks to a variety of prescribed parameters. All lenses were manufactured to established finished product specifications within the ANSI Z80.20 tolerance.

Bioburden Testing-bioburden testing conducted on rigid gas permeable lenses manufactured at Acculens, Inc. demonstrated that the colony forming units (CFU) per lens was within the established acceptance criteria of less than 100 CFU per lens.

~ Clinical Studies ~

Clinical testing to demonstrate the safety and effectiveness of contact lenses manufactured from tisilfocon A with the labeled indications for use has been addressed through previous 510(k) clearances.

  • ~ Conclusions Drawn from Testing ~
    Results from testing and other information presented in this premarket notification supports the substantial equivalence claim by demonstrating no relevant differences regarding the safety and effectiveness of the NewVision SC and NewVision SC Daily Wear Ortho-K (tisilfocon A) Lenses compared to the predicate devices.

VIII. CONCLUSIONS

Substantial Equivalence

Information presented in this premarket notification establishes that New Vision SC and New Vision SC Daily Wear Ortho-K (tisilfocon A) Lenses for daily wear are as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the proposed indications.

Risks and Benefits

The risks of the subject device are the same as those normally attributed to the wearing of rigid gas permeable (RGP) daily wear contact lenses. The benefits to the patient are the same as those for other RGP contact lenses.

§ 886.5916 Rigid gas permeable contact lens.

(a)
Identification. A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.