K Number
K230899

Validate with FDA (Live)

Device Name
qXR-PTX-PE
Date Cleared
2023-08-22

(144 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

qXR-PTX-PE is a radiological computer-assisted triage and notification software that analyzes adult chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). qXR-PTX-PE uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage.

As a passive notification for prioritization-only software tool within standard of care workflow, qXR-PTX-PE does not send a proactive alert directly to the appropriately trained medical specialists. qXR-PTX-PE is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a stand-alone basis for clinical decision-making.

Device Description

qXR-PTX-PE is a radiological computer aided triage and notification software that analyses adult frontal (AP or PA views) CXR images for the presence of pre-specified suspected target conditions (pleural effusion and/or pneumothorax). The algorithm was trained on training data from across the world. The training dataset consisted of 74% of the data from India, 20.04% from the EU, 3.9% from the US, 1.4% from Brazil and 0.63% from Vietnam. The input for qXR-PTX-PE is a frontal chest X-ray (AP and PA view) in digital imaging and communications in medicine (DICOM) format

Chest X-rays are sent to qXR-PTX-PE by the means of transmission within the user's image storage system (e.g., Picture Archiving and Communication System (PACS)) or other radiological imaging equipment (e.g., X-ray systems) and processed by the qXR-PTX-PE for analysis. Following receipt of chest radiographs, the software device automatically analyses each image to detect features suggestive of pneumothorax and/or pleural effusion.

This would allow the appropriately trained medical specialists to group suspicious exams together that may potentially benefit for their prioritization. Chest radiographs without the suspicious findings are placed in the worklist for routine review, which is the standard of care at present. A secondary capture is available for the information on presence of the suspicious findings.

qXR-PTX-PE does not provide any proactive alerts. qXR-PTX-PE is not intended to direct attention to specific portions of the image. The results are not intended to be used on a standalone basis for clinical decision-making nor is it intended to rule out the target conditions or otherwise preclude clinical assessment of X-Ray cases.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the statement: "The device shows > 95% AUC". The reported performance significantly exceeds this threshold.

MetricAcceptance CriteriaqXR-PTX-PE Performance (Pneumothorax)qXR-PTX-PE Performance (Pleural Effusion)
ROC AUC> 0.950.9894 (95% CI: 0.9829 - 0.9980)0.9890 (95% CI: 0.9847 - 0.9944)
SensitivityNot explicitly defined beyond AUC94.53% (95% CI: 90.42-97.24)96.22% (95% CI: 93.62-97.97)
SpecificityNot explicitly defined beyond AUC96.36% (95% CI: 94.07-97.95)94.90% (95% CI: 93.04-96.39)
Performance Time (Notification)Not explicitly defined, but compared to predicate and other cleared products10 seconds (average)10 seconds (average)

2. Sample Size Used for the Test Set and Data Provenance

  • Pneumothorax Test Set: 613 scans
    • 201 scans with pneumothorax
    • 412 scans without pneumothorax
  • Pleural Effusion Test Set: 1070 scans
    • 344 scans with pleural effusion
    • 726 scans without pleural effusion
  • Data Provenance: Retrospective. All test data was obtained from various hospitals across the US. Specific regions mentioned include Midwest, Northeast, and West. The test set was intentionally obtained from sites different from the training data sites to ensure independence.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: 3
  • Qualifications: ABR (American Board of Radiology) thoracic radiologists with a minimum of 10 years of experience.

4. Adjudication Method for the Test Set

The provided text states that "The ground truth was established by 3 ABR thoracic radiologists with a minimum of 10 years of experience." It does not specify a particular adjudication method (e.g., 2+1, 3+1, or simple consensus). Without further detail, it's assumed to be a consensus among these three experts, but the exact process of resolving discrepancies is not described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not reported. The study focuses purely on the standalone performance of the AI algorithm (qXR-PTX-PE) and compares it to the standalone performance of a predicate device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance study was done. The reported AUC, sensitivity, and specificity metrics are for the qXR-PTX-PE algorithm acting independently.

7. The Type of Ground Truth Used

The ground truth used was expert consensus (3 ABR thoracic radiologists with a minimum of 10 years of experience).

8. The Sample Size for the Training Set

The exact total sample size for the training set is not specified. However, the text provides the geographical distribution of the training data:

  • 74% from India
  • 20.04% from the EU
  • 3.9% from the US
  • 1.4% from Brazil
  • 0.63% from Vietnam

9. How the Ground Truth for the Training Set was Established

The document does not explicitly state how the ground truth for the training set was established. It only mentions that the algorithm was "trained on training data from across the world." It can be inferred that expert labeling or clinical diagnoses were likely used, given the nature of a medical imaging AI, but specific details about the process (e.g., number of readers, their qualifications, adjudication) are not provided for the training data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

Qure.ai Technologies % Ayushi Mahendra Senior Regulatory Affairs Specialist Level 7, Commerz II International Business Park Oberoi Garden City, Goregaon (E) Mumbai, Maharashtra 400063 INDIA

Re: K230899

August 22, 2023

Trade/Device Name: qXR-PTX-PE Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QFM Dated: July 24, 2023 Received: July 24, 2023

Dear Ayushi Mahendra:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230899

Device Name qXR-PTX-PE

Indications for Use (Describe)

qXR-PTX-PE is a radiological computer-assisted triage and notification software that analyzes adult chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). qXR-PTX-PE uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage.

As a passive notification for prioritization-only software tool within standard of care workflow, gXR-PTX-PE does not send a proactive alert directly to the appropriately trained medical specialists. qXR-PTX-PE is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a stand-alone basis for clinical decision-making.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Qure.ai's qXR-PTX-PE

1 SUBMITTER

K230899

Qure.ai Technologies Level 7, Commerz II, International Business Park Oberoi Garden City, Goregaon (E), Mumbai 400 063 Phone: +91-9768123013 Primary Contact Person: Ayushi Mahendra Secondary contact person: Sri Anusha Matta

Date Prepared: August 12, 2023

2 DEVICE

Name of Device:qXR-PTX-PE
Common or Usual Name:Radiological Computer Assisted Prioritization Software for Lesions
Classification Name:Radiological Computer Aided Triage and Notification Software
Regulatory Class:Class II
Regulation Number:21 CFR 892.2080
Product Code:QFM

3 PREDICATE DEVICE

Name of Device:Lunit INSIGHT CXR Triage
Manufacturer:Lunit Inc.
510(k) Number:K211733

4 INTENDED USE / INDICATIONS FOR USE:

qXR-PTX-PE is a radiological computer-assisted triage and notification software that analyzes adult chest Xray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). qXR-PTX-PE uses an artificial intelligence algorithm to analyze images for features

{4}------------------------------------------------

suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage.

As a passive notification for prioritization-only software tool within standard of care workflow, qXR-PTX-PE does not send a proactive alert directly to the appropriately trained medical specialists. qXR-PTX-PE is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a stand-alone basis for clinical decisionmaking.

5 DEVICE DESCRIPTION

qXR-PTX-PE is a radiological computer aided triage and notification software that analyses adult frontal (AP or PA views) CXR images for the presence of pre-specified suspected target conditions (pleural effusion and/or pneumothorax). The algorithm was trained on training data from across the world. The training dataset consisted of 74% of the data from India, 20.04% from the EU, 3.9% from the US, 1.4% from Brazil and 0.63% from Vietnam. The input for qXR-PTX-PE is a frontal chest X-ray (AP and PA view) in digital imaging and communications in medicine (DICOM) format

Chest X-rays are sent to qXR-PTX-PE by the means of transmission within the user's image storage system (e.g., Picture Archiving and Communication System (PACS)) or other radiological imaging equipment (e.g., X-ray systems) and processed by the qXR-PTX-PE for analysis. Following receipt of chest radiographs, the software device automatically analyses each image to detect features suggestive of pneumothorax and/or pleural effusion.

This would allow the appropriately trained medical specialists to group suspicious exams together that may potentially benefit for their prioritization. Chest radiographs without the suspicious findings are placed in the worklist for routine review, which is the standard of care at present. A secondary capture is available for the information on presence of the suspicious findings.

qXR-PTX-PE does not provide any proactive alerts. qXR-PTX-PE is not intended to direct attention to specific portions of the image. The results are not intended to be used on a standalone basis for clinical decision-making nor is it intended to rule out the target conditions or otherwise preclude clinical assessment of X-Ray cases.

б COMPARISON OF THE PREDICATE DEVICE

qXR-PTX-PE is technologically similar to the predicate device, Lunit INSIGHT CXR Triage in regards to intended use and technological characteristics. Both are radiological computer-assisted triage and notification software intended to read chest X-rays for the presence of pre-specified target conditions (pneumothorax and pleural effusion). The algorithms function similarly and with the same purpose of workflow notification. There are no notable technological differences between the subject and predicate devices.

In terms of establishing substantial equivalence, the subject and predicate device have intended use, as an image processing tool that triages images for features suggestive of critical findings and produces case-level output. The indications for use proposed for the subject device are similar to those of the predicate device.

{5}------------------------------------------------

Predicate DeviceSubject Device
Lunit INSIGHT CXR TriageqXR-PTX-PE
Device NameLunit INSIGHT CXR TriageqXR-PTX-PE
510(k) NumberK211733
Regulation21 CFR 892.208021 CFR 892.2080
Regulation DescriptionRadiological computer aidedtriage and notification softwareRadiological computer aided triageand notification software
Product CodeQFMQFM
Device typeRadiological Computer-AssistedPrioritization Software For LesionsRadiological Computer-AssistedPrioritization Software For Lesions
ManufacturerLunit Inc.Qure.ai Technologies
Intended use / Indicationsfor UseLunit INSIGHT CXR Triage is aradiological computer-assistedtriage and notification softwarethat analyzes adult chest X-rayimages for the presence of pre-specified suspected criticalfindings (pleural effusions and/orpneumothorax). Lunit INSIGHTCXR Triage uses an artificialintelligence algorithm to analyzeimages for features suggestive ofcritical findings and produce case-level output available in thePACS/workstation for worklistprioritization or triage.As a passive notification forprioritization-only software toolwithin standard of care workflow,Lunit Insight CXR triage doesnot send a proactive alert directlyto appropriately trained medicalspecialists. Lunit INSIGHT CXRTriage is not intended to directattention to specific portions ofan image. Its results are notintended to be used on astandalone basis for clinicaldecision-making.qXR-PTX-PE is a radiologicalcomputer-assisted triage andnotification software that analyzesadult chest X-ray images for thepresence of pre-specified suspectedcritical findings (pleural effusionand/or pneumothorax). qXR-PTX-PEuses an artificial intelligencealgorithm to analyze images forfeatures suggestive of critical findingsand provides case-level outputavailable in the PACS/workstation forworklist prioritization or triage.As a passive notification forprioritization-only software toolwithin standard of care workflow,qXR-PTX-PE does not send a proactivealert directly to the appropriatelytrained medical specialists. qXR-PTX-PE is not intended to direct attentionto specific portions of an image or toanomalies other than pleural effusionand/or pneumothorax. Its results arenot intended to be used on a stand-alone basis for clinical decision-making
Intended UserAppropriately trained medicalspecialists who are qualified tointerpret chest radiographs.Radiologists, clinicians, and otherappropriately trained medicalspecialists qualified to read chestradiographs
ModalityChest X-rayChest X-ray
Predicate DeviceSubject Device
Lunit INSIGHT CXR TriageqXR-PTX-PE
Target clinical conditionsPleural effusion, Pneumothoraxon Chest/Lung Frontal Chest X-rayPneumothorax, Pleural effusion onChest/Lung Frontal Chest X-rays
Algorithm for pre-specified critical findingsdetectionAl algorithm designed to detectpleural effusion andpneumothorax in chest X-rayimages. Lunit INSIGHT CXR Triageuses a vendor agnostic algorithmcompatible with DICOM chest X-ray imagesqXR-PTX-PE uses an AI algorithm todetect pneumothorax and pleuraleffusion on chest X-ray images. qXR-PTX-PE uses a vendor agnosticalgorithm compatible with DICOMchest X-ray images
Notification only/ParallelworkflowYesYes
Input formatDICOMDICOM
Device output in case ofpositive detectionWhen deployed on otherradiological imaging equipment,Lunit INSIGHT CXR Triageautomatically runs after imageacquisition and prioritizes anddisplays the analysis resultthrough the worklist interface ofPACS/workstation.No markup on original image.Secondary capture of the finding.Upon image acquisition fromother radiological imagingequipment (e.g. X-ray systems),an on-device, technologistnotification indicating which caseswere flagged by Lunit INSIGHTCXR Triage in PACS, is generated.The on device notification iscontextual and does not provideany diagnostic information. It isnot intended to inform any clinicaldecision, prioritization, or actionto the technologistWhen deployed on other radiologicalimaging equipment, qXR-PTX-PE willautomatically run after imageacquisition to perform triage. Itdisplays the analysis result throughthe worklist interface ofPACS/workstation.No markup of the conditions will bedone on the original image.Secondary capture of the device willindicate the presence of findingssuspicious of pneumothorax orpleural effusion.Upon image acquisition from otherradiological imaging equipment (e.g.X-ray systems) a passive notificationis generated.
Notification (i.e.,Passive notification. Images withPassive notification. Images with
recipient, timing andmeans of notification)suspicion of pleural effusionand/or pneumothorax are flaggedin PACS/workstation.suspicion of pneumothorax and/orpleural effusion are flagged inPACS/workstation/DICOM viewer.
Where generated results(i.e., DICOM files) arestoredPACS/WorkstationPACS/Workstation/DICOM viewer

Table 1 Comparison between qXR-PTX-PE and the Predicate Device

{6}------------------------------------------------

{7}------------------------------------------------

Predicate DeviceSubject Device
Lunit INSIGHT CXR TriageqXR-PTX-PE
Performance metrics of the predicate device and qXR-PTX-PE
Performance level -Timing of notificationThe average time taken for thenotification to travel from theLunit INSIGHT CXR Triage to thepoint at which the result isdisplayed in the destinationPACS/RIS/EPR worklist is 14.66seconds.The average time taken for thenotification to travel from qXR-PTX-PE to the point at which the result isdisplayed in the destination PictureArchiving and Communication System(PACS) or workstation/digitalradiographic processing system (ex.digital radiography, digital X-raysystem etc.) is 10 seconds.
Performance level -accuracy of classificationPleural EffusionROC AUC > 0.95AUC: 0.9686 (95% Cl: [0.9547,0.9824])Sensitivity 89.86% (95% CI: [86.72,93.00])Specificity 93.48% (95% CI: [91.06,95.91])PneumothoraxROC AUC > 0.95AUC: 0.9630 (95% Cl: [0.9521,0.9739])Sensitivity 88.92% (95% CI: [85.60,92.24]) Specificity 90.51% (95% CI:[88.18, 92.83])PneumothoraxROC AUC > 0.95AUC: 0.9894 (95% CI: [0.9829,0.9980])Sensitivity 94.53% (95% CI: [90.42,97.24])Specificity 96.36% (95% CI: [94.07,97.95])Pleural EffusionROC AUC > 0.95AUC: 0.989 (95% Cl: [0.9847, 0.9944])Sensitivity 96.22% (95% CI: [93.62,97.97])Specificity 94.90% (95% CI: [93.04,96.39])

7 TESTING

Software

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device has a Moderate level of concern.

Performance Testing - Clinical

The performance of qXR-PTX-PE was validated by clinical tests. All the safety parameters of the device were verified in accordance with the software specifications and applicable performance standards and met the acceptance criteria (Device shows > 95% AUC) (passed), demonstrating that the software fulfills all its requirement specifications.

Clinical studies were conducted on retrospectively collected Chest X-rays to evaluate the performance of qXR-PTX-PE for triaging of pneumothorax and pleural effusion.

{8}------------------------------------------------

The study for pneumothorax included 613 scans (201 scans with pneumothorax and 412 scans without pneumothorax) from various parts of the US. The dataset was obtained from various hospitals across the US in order to generate evidence on the device function in various subgroups. A data collection protocol was set in place to ensure that there were sufficient numbers of important subgroups. The protocol specifies the inclusion and exclusion criteria and the expected numbers of cases and controls to be included. The dataset consisted of 289 males, 287 females and 37 were without this information available. The cases were aged from 22 years to above 85 years. The dataset was obtained from various hospitals across the US (179 from the Midwest, 152 from the Northeast and 12 from the South, region for 145 were not known) in order to generate evidence on the device function in various subgroups. The dataset consisted of clinical confounders that included opacities, presence of hardware, emphysema, scarring, mediastinal widening and pleural thickening. The dataset was also obtained from various X-ray device manufacturers to ensure consistent performance.

The algorithm was trained on training data from across the world. The training dataset consisted of 74% of the data from India, 20.04% from the US, 1.4% from Brazil and 0.63% from Vietnam. The test set was obtained from sites that were different from the training data sites and therefore this ensured the independence of the test data from training data.

The ground truth was established by 3 ABR thoracic radiologists with a minimum of 10 years of experience. The AUC of the device in triaging scans with findings suspicious of pneumothorax exceeded the success criteria with (AUC 98.94 95% Cl (98.28 - 99.82)), Sensitivity 94.53 (90.42-97.24) and Specificity 96.36 (94.07-97.95). The predicate Lunit INSIGHT CXR Triage's performance was ROC AUC 0.9630 (95% Cl: 0.9521 - 0.9739), Sensitivity 88.92% (95% Cl: 85.60 - 92.24) and Specificity 90.51% (95% Cl: 88.18 - 92.83).

AUC (95% CI)Sensitivity (95% CI), TP/PSpecificity (95% CI), TN/N
98.94 (98.28 - 99.82)94.53 (90.42-97.24), 190/20196.36 (94.07-97.95), 397/412

Table 2 Overall Results of Accuracy Testing of qXR-PTX-PE for Pneumothorax

The device's generalizability was ensured by performing subgroup analyses. The results for pneumothorax were consistent in both genders and the AUC was 98.67 (97.63-100) in male, 99.3(98.64-100) in female. An AUC of 92.47(84.95-100) was observed in the scans with unknown gender. The device's results were also found to be consistent in a wide range of ages: with AUC of 99.12 (98.34-100) in the 22-44 age group, 99.17 (98.37-100) in the 45-64 age group, 98.69 (97.61-100) in the 65-84 age group and 98.62 (97.25-100.00) in the 85 or greather age group age group.

The study for pleural effusion included 1070 scans (344 scans with pleural effusion and 726 scans without pleural effusion) from various parts of the US. A data collection protocol was set in place to ensure that there were sufficient numbers of important subgroups. The protocol specifies the inclusion criteria and the expected numbers of cases and controls to be included. There were 498 scans from females and 551 from males, 21 samples did not have this information available. The ages ranged from 22 years to greater than 85 years. The samples were from various parts of the US (278 from the Midwest, 213 from the Northeast, 6 from the West. For 265, this information was not known). The dataset consisted of clinical confounders that included opacities, presence of hardware, emphysema, scarring, mediastinal widening and pleural thickening. The dataset was also obtained from various X-ray device manufacturers to ensure consistent performance. The algorithm was training data from across the world. The training dataset consisted of 74% of the data from India, 20.04% from the EU, 3.9%

{9}------------------------------------------------

from the US, 1.4% from Brazil and 0.63% from Vietnam. The test set was obtained from sites that were different from the training data sites and therefore this ensured the independence of the test data from training data.

The ground truth was established by 3 ABR thoracic radiologists with a minimum of 10 years of experience. The AUC of the device in triaging scans with findings suspicious of pleural effusion with AUC 98.90 (98.47 -99.44)), sensitivity 96.22 (93.62-97.97) and specificity 94.90 (93.04-96.39). The predicate, Lunit INSIGHT CXR Triage, reported a performance of ROC AUC 0.9686 (95% Cl: 0.9547 - 0.9824), sensitivity 89.86% (95% CI: 86.72 - 93.00) and specificity 93.48% (95% CI: 91.06 - 95.91).

AUC (95% CI)Sensitivity (95% CI), TP/PSpecificity (95% CI), TN/N
98.90 (98.47 - 99.44)96.22 (93.62-97.97), 331/34494.90 (93.04-96.39), 689/726

The device's generalizability for triaging scans with pleural effusion was ensured by performing subgroup analyses. The results for pleural effusion were consistent in both genders and the AUC was 98.73 (98.00-99.74) in female, 99.11(98.62-99.72) in male. An AUC of 96.3(92.59-100) was observed in the scans with unknown gender. The device's results were also found to be consistent in a wide range of ages: with AUC of 99.27 (98.56-100) in the 22-44 age group, 98.1 (96.83-99.95) in the 45-64 age group, 98.78 (98.09-99.62) in the 65-84 age group and 99.2 (98.39-100.00) in the 85 or greater age group. The device's performance time was also assessed and for qXR-PTX-PE. It was the time to analyze the study and send the notification to the worklist. The performance time averaged at 10s. This is comparable to the performance of the predicate performance of 14.66s. This is also comparable to other commercially cleared products with the similar intended use.

8 CONCLUSION

The comparison in Table 1 and the software and performance testing presented above demonstrate that the qXR-PTX-PE device is substantially equivalent to the predicate device. The qXR-PTX-PE is a software only device, similar to the predicate (Lunit INSIGHT CXR Triage). It is as safe and effective as the predicate device. The qXR-PTX-PE has the same intended users and similar indications, technological characteristics, and principles of operation as the predicate device. There are no differences in the indications, therefore there no is risk of its safety and effectiveness being affected when used as labelled. Both devices operate in parallel to the standard of care workflow. The performance testing demonstrates that the qXR-PTX-PE performs as intended and is therefore substantially equivalent to the predicate. Software and Clinical testing supports that the device performs in according with the device requirements.

N/A