K Number
K230586

Validate with FDA (Live)

Date Cleared
2023-06-09

(99 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Renuvion® Micro Handpiece is intended for the delivery of radiofrequency energy and/or helium plasma where coagulation/contraction of soft tissue is needed. Soft tissue includes subcutaneous tissue.

The Renuvion® Micro Handpiece is compatible with the Apyx One Console Generator, owned by Apyx Medical.

Device Description

The Renuvion® Micro System with all components and accessories is show in the picture below. The system is comprised of a compatible electrosurgical generator (Apyx One Console), a sterile, single use handpiece, and a supply of helium gas. Additional accessories include a digital gas regulator, grounding pad and optional footswitch.

The Renuvion® Micro handpiece is a sterile, single use electrosurgical (monopolar) device intended to be used in conjunction with the Apyx One Console Electrosurgical Generator for the percutaneous delivery of radiofrequency and helium plasma energy to soft tissue. When connected to the Apyx One Generator, the device operates at an adjustable power of up to 12 W (expressed as 0 - 30% where 30% is 12 W) and an adjustable helium gas flow of 1 -1.5 L/Min. Radiofrequency energy is delivered to the handpiece by the generator and used to energize the electrode. When helium gas is passed over the energized electrode, a helium plasma is generated for soft tissue coagulation/contraction.

The handpiece has a non-extendable electrode to generate helium plasma. The handpiece features a low-profile shaft that is 10cm in length. The distal end of the shaft and tip has a 2mm outer diameter that tapers down to 1.5mm for the remaining length of the shaft.

AI/ML Overview

The Renuvion® Micro Handpiece is cleared based on demonstrating substantial equivalence to a predicate device, the Renuvion® APR Handpiece (K223262 and K230272). This type of submission (510(k)) focuses on comparing a new device to an already legally marketed device, rather than proving absolute safety and effectiveness through extensive clinical trials. Therefore, the "acceptance criteria" are primarily based on demonstrating that the new device is as safe and effective as the predicate device by showing similar technological characteristics and performance.

Here's a breakdown of the provided information concerning acceptance criteria and supporting studies:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly list "acceptance criteria" in a separate section with specific numerical targets. Instead, it presents various tests performed and states whether the device "met requirements" or demonstrated equivalent/lesser effects compared to the predicate. The overall acceptance criterion is demonstrating that the Renuvion® Micro Handpiece is as safe and effective as the predicate device.

Test TypeObjectiveReported Device Performance (Result)
Bench Testing
Mechanical Verification & FunctionalityVerify Mechanical functionality of the Renuvion® Micro Handpiece.Mechanical functionality requirements met.
System Compatibility TestingVerify the device, when connected to the generators (with all the components and accessories working together as a system), are working as intended and are compatible with all system components.System compatibility requirements met.
Inspectional VerificationDocument parameters that can be verified through inspection of the Micro Handpiece Labeling and IFU.Inspectional requirements met.
Packaging ValidationEvaluate simulated (and/or real-time) aging and associated package integrity and shelf-life claims. Assess the potential for damage to the device (e.g., drop tests of the instrument in its packaging).Packaging requirements met.
Electrical Safety & EMCCompliance with ANSI/AAMI/IEC 60601-1:2015/(R)2012 and A1:2012, AAMI/ANSI/IEC 60601-2-2:2017, and AAMI/ANSI/IEC 60601-1-2:2014 (4th Edition) standards.The Renuvion® Micro System was determined to be in conformance with these standards.
Biocompatibility TestingCompliance with FDA Guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and AAMI/ANSI/ISO 10993-1:2018 standard.
CytotoxicityEvaluate cytotoxicty.Non-cytotoxic.
SensitizationEvaluate sensitization reaction.No sensitization reaction.
IrritationEvaluate irritation.Non-irritant.
Acute Systemic ToxicityEvaluate acute systemic toxicity.Non-toxic.
Material Mediated Pyrogen/PyrogenicityEvaluate pyrogenicity.Non-pyrogenic.
Pre-Clinical & Ex-Vivo Studies
Thermal Tissue EffectEvaluate the thermal effect of the Renuvion Micro Handpiece against its predicate device on different tissue types at different Renuvion settings.The test device (APYX-10-MICRO) demonstrated an equal or less tissue thermal effect when compared to the predicate device (APYX-15-TP) on five different tissue types at different Renuvion settings.
Tissue Temperature Over 85 Degrees Celsius ComparisonEvaluate the tissue time over 85 degrees Celsius between the Renuvion Micro Handpiece and its predicate device when used at recommended settings for coagulation/contraction of soft tissue.The test device (APYX-10-MICRO) met the acceptance criteria demonstrating that on average the device is at or above 85 degrees for greater than 45 msec. The predicate device showed similar results as a control.

2. Sample Size Used for the Test Set and Data Provenance:

  • The document does not specify numerical sample sizes for the bench, electrical safety, biocompatibility, or ex-vivo tests. It states that tests were conducted according to protocols.
  • Data Provenance:
    • All summarized tests appear to be pre-clinical/bench-top studies and ex-vivo (tissue) tests conducted by the manufacturer, Apyx Medical Corporation.
    • The "Ex-vivo tissue testing" section implies that animal or cadaveric tissue was used, but the specific origin (e.g., species, country) is not mentioned.
    • There is no mention of human subjects or clinical data in this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

  • Since these are primarily bench and ex-vivo tests for a 510(k) submission, the "ground truth" is established by engineering and scientific measurements/observations against predefined specifications and industry standards.
  • The document does not mention the involvement of external medical experts (e.g., radiologists) for establishing ground truth in these specific tests. The experts involved would likely be engineers, scientists, and quality assurance personnel from Apyx Medical performing the tests and comparing results to established criteria. There is no mention of "ground truth" in the context of clinical interpretation, as clinical studies were not performed.

4. Adjudication Method for the Test Set:

  • Adjudication methods like "2+1" or "3+1" are typically used in clinical studies involving interpretation by multiple readers.
  • Given that this submission relies on bench and ex-vivo testing, such adjudication methods are not applicable and are not mentioned in the document. Performance is measured against physical or biological criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence." (Section 6: Clinical Studies).
  • Therefore, there is no effect size reported for human readers improving with AI vs. without AI assistance. The device is not an AI-based diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • This question is not applicable as the Renuvion® Micro Handpiece is an electrosurgical device, not an algorithm or AI-based system. It does not perform diagnostic tasks independently.

7. The Type of Ground Truth Used:

  • For the bench and ex-vivo tests, the "ground truth" is based on:
    • Defined engineering specifications (e.g., mechanical functionality, system compatibility, electrical safety, packaging integrity).
    • Biological responses to known stimuli (e.g., non-cytotoxic, no sensitization reaction for biocompatibility tests).
    • Physical measurements and comparisons (e.g., thermal effect and tissue temperature measurements in ex-vivo tissue).
  • The ground truth is established by the test protocols against established standards and internal design requirements, often compared to the performance of the predicate device.
  • No pathology or outcomes data is mentioned as forming the ground truth for this set of studies.

8. The Sample Size for the Training Set:

  • This question is not applicable for this device. Training sets are relevant for machine learning or AI-based devices. The Renuvion® Micro Handpiece is a physical surgical tool, and its development involves engineering design and testing, not algorithm training.

9. How the Ground Truth for the Training Set Was Established:

  • This question is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 9, 2023

Apyx Medical Corporation Mark Evans Sr. Premarket Regulatory Affairs Specialist (formerly Bovie Medical Corporation) 5115 Ulmerton Road Clearwater, Florida 33760-4004

Re: K230586

Trade/Device Name: Renuvion® Micro Handpiece Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: May 15, 2023 Received: May 15, 2023

Dear Mark Evans:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.06.09 15:33:07 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

Device Name Renuvion® Micro Handpiece

Indications for Use (Describe)

Renuvion® Micro Handpiece is intended for the delivery of radiofrequency energy and/or helium plasma where coagulation/contraction of soft tissue is needed. Soft tissue includes subcutaneous tissue.

The Renuvion® Micro Handpiece is compatible with the Apyx One Console Generator, owned by Apyx Medical.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Apyx Medical. The logo is comprised of the word "Apyx" in orange, with a blue triangle in place of the "A". Below the word "Apyx" is the word "MEDICAL" in blue. The logo is simple and modern, and the colors are bright and eye-catching.

    1. General Information:
Submitted by:Apyx Medical Corporation5115 Ulmerton Road Clearwater, Florida33760-4004United States of America
Establishment Registration #:3007593903
Contact Person:Mark D. Evans, Sr. Premarket Regulatory AffairsSpecialist5115 Ulmerton RoadClearwater, Florida 33760-4004United States of AmericaPhone: (856) 524-5037Email: mark.evans@apyxmedical.com
Date Prepared:March 2, 2023
Trade Names (Model Numbers):Renuvion® Micro Handpiece(APYX-10-Micro)
Common Name:Electrosurgical Handpiece
Classification:Class II per 21CFR 878.4400 - ElectrosurgicalCutting and Coagulation Device and AccessoriesProduct Code GEI
Primary Predicate Device:Secondary Predicate Device:Renuvion® APR Handpiece (K223262)Renuvion® APR Handpiece (K230272)

2. Proposed Indications for Use:

Renuvion® Micro Handpiece is intended for the delivery of radiofrequency energy and/or helium plasma where coagulation/contraction of soft tissue is needed. Soft tissue includes subcutaneous tissue.

The Renuvion® Micro Handpiece is compatible with the Apyx One Console Generator, owned by Apyx Medical.

    1. Device Description and Technological Characteristics:

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for Apyx Medical. The word "Apyx" is in orange, with a blue triangle in the "A". The word "MEDICAL" is in blue and is located below the word "Apyx". The logo is simple and modern.

The Renuvion® Micro System with all components and accessories is show in the picture below. The system is comprised of a compatible electrosurgical generator (Apyx One Console)¹, a sterile, single use handpiece, and a supply of helium gas. Additional accessories include a digital gas regulator2, grounding pad3 and optional footswitch4.

The front view of the entire system with all components and accessories is depicted in Figure 1 below.

Image /page/4/Picture/6 description: The image shows a medical device setup. The main component is a white generator (ESU) with a black screen. Connected to it are a handpiece and cable, a digital regulator attached to a helium tank, foot switches labeled "J-PLASMA" and "COOL COAG", and a grounding pad.

Figure 1: Front View of System

1 510(K) Number: K221830

2 510(K) Number: K221830

3 510(k) Number: K092761

4 510(K) Number: K170188

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for Apyx Medical. The logo consists of the word "Apyx" in orange, with a blue triangle in the "A". The word "MEDICAL" is written in blue below the word "Apyx".

The Renuvion® Micro handpiece is a sterile, single use electrosurgical (monopolar) device intended to be used in conjunction with the Apyx One Console Electrosurgical Generator for the percutaneous delivery of radiofrequency and helium plasma energy to soft tissue. When connected to the Apyx One Generator, the device operates at an adjustable power of up to 12 W (expressed as 0 - 30% where 30% is 12 W) and an adjustable helium gas flow of 1 -1.5 L/Min. Radiofrequency energy is delivered to the handpiece by the generator and used to energize the electrode. When helium gas is passed over the energized electrode, a helium plasma is generated for soft tissue coagulation/contraction.

The handpiece has a non-extendable electrode to generate helium plasma. The handpiece features a low-profile shaft that is 10cm in length. The distal end of the shaft and tip has a 2mm outer diameter that tapers down to 1.5mm for the remaining length of the shaft.

Image /page/5/Figure/6 description: The image shows a Renuvion device with its various components labeled. The device consists of a handle with an activation button, a stainless steel shaft with an energy exit port and tip distance indicators, and a flush port. Additionally, there is a plug connector and a cable attached to the device.

The Renuvion® Micro Handpiece is pictured in Figure 2 below.

Figure 2: Renuvion® Micro Handpiece Major Components

The Renuvion® Micro Handpiece is a modification to the primary predicate device that was cleared in K223262 (Renuvion® APR Handpiece). There are no changes to the principle of operation or mechanism of action.

The Renuvion® Micro Handpiece as compared to the predicate devices has the following differences in technological characteristics to enable overall dimensional profile:

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the logo for Apyx Medical. The logo is comprised of the word "Apyx" in orange, with a blue triangle in place of the letter "A". Below the word "Apyx" is the word "MEDICAL" in blue. The "®" symbol is located to the right of the "X" in "Apyx".

  • front-port ceramic nozzle/tip, rather than predicate side-port configuration ●
  • introduces a sterile saline flush port to assist in removing coagulum build up on the ceramic tip.
  • The subject devices shaft material is stainless steel with an outer polymer layer, ● whereas the predicate used a polymer-based material throughout.
  • o The subject device maximum allowable power setting will lowered compared to the predicate, 30% power and maximum helium flow will be 1.5 L/min

4. Performance Data:

a. Bench Testing

All testing performed on the Renuvion® Micro Handpiece, hereinafter referred to as Micro Handpiece, was derived from the risk assessment in accordance with ISO 14971 which evaluated the safety and effectiveness of the design with Apyx Medical's Design and Development procedures. The test methodology and acceptance criteria were developed from the same standards and internal Design and Development procedures used for clearance of the primary predicate device.

A risk analysis was conducted using the Design Risk Analysis and Control (DRAC) for the subject device (Micro Handpiece). The risk analysis concluded that no new failure modes or risk hazards were identified as a result of the design modifications.

Performance testing to assure that the subject device met performance requirements was conducted in accordance with protocols to verify design specifications. The testing performed is summarized in the following table:

TestObjectiveResult
MechanicalVerification&FunctionalityVerify Mechanical functionality ofthe Renuvion® Micro HandpieceMechanical functionalityrequirements met
SystemcompatibilityTestingVerify the device, when connectedto the generators (with all thecomponents and accessoriesworking together as a system), areworking as intended and arecompatible with all systemcomponents.System compatibility requirementsmet
InspectionalVerificationDocument parameters that can beverified through inspection of theMicro Handpiece Labeling and IFU.Inspectional requirements met

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows the logo for Apyx Medical. The logo consists of the word "Apyx" in orange, with a blue triangle in the negative space of the "A". Below the word "Apyx" is the word "MEDICAL" in blue. The logo is simple and modern, and the colors are bright and eye-catching.

TestObjectiveResult
PackagingValidationEvaluate simulated (and/or real-time) aging and associatedpackage integrity and shelf-lifeclaims. Assess the potential fordamage to the device (e.g., droptests of the instrument in itspackaging).Packaging requirements met

b. Electrical Safety and Electromaqnetic Compatibility (EMC)

Electrical safety and EMC testing were conducted on the Renuvion® Micro System. The handpiece complies with the ANSI/AAMI/IEC 60601-1:2015/(R)2012 and A1:2012, and AAMI/ANSI/IEC 60601-2-2:2017 standards for safety and the AAMI/ANSI/IEC 60601-1-2:2014 (4th Edition) standard for EMC. The Renuvion® Micro System was determined to be in conformance with these standards.

c. Biocompatibility Testing

The biocompatibility evaluation for the Micro Handpiece was conducted in accordance with the June 16, 2016 FDA Guidance, "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and the AAMI/ANSI/ISO 10993-1:2018 standard, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process". A Biological Evaluation Plan within Apyx Medical's risk management process was developed for the Micro Handpiece to determine the testing and risk assessment for the device and to develop a plan for the mitigation of risks identified.

Biocompatibility testing performed is summarized below:

Category: External Communicating Device Contact: Tissue / Bone / Dentin Duration of Contact: Limited < 24 hours

TestResult
CytotoxicityNon-cytotoxic
SensitizationNo sensitization reaction

{8}------------------------------------------------

Image /page/8/Picture/2 description: The image shows the logo for Apyx Medical. The logo is orange and blue. The word "MEDICAL" is written in blue below the logo.

IrritationNon-irritant
Acute Systemic ToxicityNon-toxic
Material MediatedPyrogen/PyrogenicityNon-pyrogenic

The design of the subject device is supported by evidence of biocompatibility verification testing to ensure the device is safe for human use. The materials used in the manufacture of the device met biocompatibility test requirements for external communicating device, Tissue/Bone/Dentin, limited duration devices.

5. Pre-Clinical & Ex-Vivo Studies:

Ex-vivo tissue testing was conducted and is summarized in the table below:

TestObjectiveResult
ThermalTissueEffectEvaluate the thermal effect betweenRenuvion Micro Handpiece (APYX-10-MICRO) against its predicate device(Renuvion APR Handpiece, 15 cm, TwinPort, APYX-15-TP).The test device (APYX-10-MICRO) wassubjected to tissue thermal testing anddemonstrated an equal or less tissuethermal effect when compared to thepredicate device (APYX-15-TP) on fivedifferent tissue types at different Renuvionsettings.
TissueTemperatureOver 85DegreesCelsiusComparisonEvaluate the tissue time over 85 degreesCelsius between the Renuvion MicroHandpiece (APYX-10-MICRO) and itspredicate device (Renuvion APRHandpiece, 15 cm, Twin Port, APYX-15-TP) when used at the recommendedsetting per the IFU forcoagulation/contraction of soft tissue suchas subcutaneous tissues.The test device (APYX-10-MICRO) metthe acceptance criteria demonstrating thaton average the device is at or above 85degrees for greater than 45 msec. Thepredicate device showed similar resultsas a control.

{9}------------------------------------------------

Image /page/9/Picture/2 description: The image shows the logo for "Aryx Medical". The logo is primarily orange, with the word "Aryx" in a stylized font. The "A" is shaped like a triangle with a small blue triangle inside, and the "y" is connected to the "x". Below "Aryx" is the word "MEDICAL" in blue, in a sans-serif font.

6. Clinical Studies:

This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

7. Substantial Equivalence:

The primary predicate and the secondary predicate are identical, with the exception of their indications for use.

Subject DevicePrimary PredicateDeviceSecondary PredicateDevice
Feature/CharacteristicRenuvion® MicroHandpieceRenuvion® APRHandpiece (K223262)Renuvion® APRHandpiece (K230272)
ClassificationClass IIClass IIClass II
Regulation Nameand Product Code:Electrosurgical cutting &coagulation device andaccessories, GEIElectrosurgical cutting &coagulation device andaccessories, GEIElectrosurgical cutting &coagulation device andaccessories, GEI
Indications for UseStatementRenuvion® MicroHandpiece is intendedfor the delivery ofradiofrequency energyand/or helium plasmacoagulation/contractionof soft tissue is needed.Soft tissue includessubcutaneous tissue.The Renuvion® MicroHandpiece is compatiblewith the Apyx OneConsole Generator,owned by Apyx Medical.Renuvion® APRHandpiece is intendedfor the delivery ofradiofrequency energyand/or helium plasmacoagulation/contractionof soft tissue is needed.Soft tissue includessubcutaneous tissue.The Renuvion® APRHandpiece is indicatedfor use in subcutaneousdermatological andaesthetic procedures toimprove the appearanceof lax (loose) skin in theneck and submentalregion.The Renuvion® APRHandpiece is intended toRenuvion® APRHandpiece is intendedfor the delivery ofradiofrequency energyand/or helium plasmawherecoagulation/contractionof soft tissue is needed.Soft tissue includessubcutaneous tissue.The Renuvion® APRHandpiece is intendedfor the coagulation ofsubcutaneous softtissues followingliposuction for aestheticbody contouring.The Renuvion® APRHandpiece is indicatedfor use insubcutaneous
Feature/CharacteristicSubject DevicePrimary PredicateDeviceSecondary PredicateDevice
Renuvion® MicroHandpieceRenuvion® APRHandpiece (K223262)Renuvion® APRHandpiece (K230272)
be used with compatibleelectrosurgicalgenerators owned byApyx Medical.dermatological andaesthetic procedures toimprove the(loose) skin in the neckand submental region.The Renuvion®APR Handpiece isintended for thedelivery ofradiofrequencyenergy and/orhelium plasma forcutting, coagulationand ablation of softtissue during opensurgical procedures.The Renuvion® APRHandpiece is intended tobe used with compatibleelectrosurgicalgenerators owned byApyx Medical.
Energy SourceRF Generator, Apyx OneConsole GeneratorRF Generator,Generators owned byApyx MedicalRF Generator,Generators owned byApyx Medical
Energy TypeRadio frequency (RF)Energy and Helium GasRadio frequency (RF)Energy and Helium GasRadio frequency (RF)Energy and Helium Gas
System ComponentsThe Renuvion® MicroSystem consists of:RF GeneratorDisposable HPFoot pedalPower cordGas regulatorThe Apyx Plasma/RFSystem consists of:RF GeneratorDisposable HPFoot pedalPower cordGas regulatorThe Apyx Plasma/RFSystem consists of:RF GeneratorDisposable HPFoot pedalPower cordGas regulator
Feature/CharacteristicSubject DeviceRenuvion® MicroHandpiecePrimary PredicateDeviceRenuvion® APRHandpiece (K223262)Secondary PredicateDeviceRenuvion® APRHandpiece (K230272)
Design & EnergyDeliveryConfiguration• Gas cylinder10cm shaft with front-port configuration andindicator lines on theshaft• Gas cylinder15cm and 27cm longshaft with a side portconfiguration (single ortwin) and indicator lineson the shaft• Gas cylinder15cm and 27cm longshaft with a side portconfiguration (single ortwin) and indicator lineson the shaft
Shaft and Tip OuterDiameter (OD)Tip and Distal end ofshaft OD: 2mmShaft beyond distal endOD: 1.5mm3mm3mm

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for Apyx Medical. The logo is orange and blue. The word "Apyx" is in orange, and the word "Medical" is in blue.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for Apyx Medical. The word "Apyx" is in orange, with a blue triangle in the negative space of the "A". The word "MEDICAL" is in blue and is located below the word "Apyx". The logo is simple and modern.

8. Substantial Equivalence Determination

The subject device design is based on the design of the Renuvion® APR Handpiece. The subject device has the same general intended use, principle of operation and system requirements. Both handpieces share many of the same technological characteristics except for the design modifications described in Section 3 above. The same performance testing completed for the Renuvion® APR Handpiece has also been conducted for the Micro Handpiece. The conclusions drawn from the bench and ex vivo tissue testing demonstrate that the Micro Handpiece is as safe and effective as the legally marketed primary predicate device and therefore is substantially equivalent.

9. Conclusion:

The Renuvion® Micro Handpiece is identical in principle of operation, mechanism of action, clinical use, general intended use, and system requirements to the predicate device, Renuvion APR (K223262 and K230272). Overall, the subject device is a smaller version of the predicate devices with minor design modifications to enable device profile. The different technical characteristics do not raise new questions of safety or effectiveness of the subject device. The data presented in this 510(k) demonstrates that the subject device is as safe and effective as the predicate device and there have been no new questions of safety or effectiveness raised. Thus, supporting a substantial equivalence determination for the subject device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.