K Number
K230548

Validate with FDA (Live)

Date Cleared
2024-02-02

(339 days)

Product Code
Regulation Number
N/A
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For protection against teeth grinding, and jaw clenching; protections from injury due to bruxism or clenching; relief of bruxism related headaches, migraines and pain; short-term relief from muscle spasm due to occlusal interference; prevention of chronic tension and temporal mandibular joint (TMJ) syndrome that is caused by chronic jaw clenching of the mandibular and maxillary teeth by the temporalis muscles; and temporary treatment of temporal mandibular disorder (TMD) along with the relief of associated headaches and pains in adults 18 years of age or older.

Device Description

The Myoaligner appliance is a customized, removable dental device which is designed to overlay the existing dentition without the need for any alterations done to the receiving dentition. The appliance is a plural piece intra-oral removable appliance, which is designed and delivered to the patients by a licensed practitioner such as a dentist (Rx only). The dentist can prescribe a device that fits over the 4-6 front teeth, back teeth on both sides of the dental arch or all of the teeth in the dental arch both on the upper or lower jaw. The appliance is held in place by mechanical retention. The Myoaligner appliance can be worn on either the top or bottom teeth or both. It is intended for use in adults 18 years of age and older.

AI/ML Overview

The provided text describes the FDA's 510(k) clearance for the Myoaligner Appliance. However, it does not contain the detailed information required to describe acceptance criteria and a study proving the device meets those criteria, as typically found in a clinical study report or a more comprehensive summary of performance data.

Specifically, the document mentions "Performance Testing" but does not provide a table of acceptance criteria and reported device performance for the Myoaligner Appliance itself as typically seen for AI/ML device clearances. It states:

  • "Performance testing confirmed the Myoaligner anterior segment retention is equivalent to reference device NTT-TSS, the Myoaligner posterior segments are equivalent to the predicate Gelb Appliance."
  • "Performance testing was also done to show the effects of repeated use (insertion and removal) and articulating paper to show interdigitation and canine guidance."

This suggests comparative testing against reference/predicate devices for mechanical properties (retention, repeated use endurance, interdigitation/canine guidance), but it does not detail specific quantitative acceptance criteria or the actual results obtained for these tests. The document's primary focus is on establishing substantive equivalence based on shared indications for use, technological characteristics, materials (PMMA), and mechanism of action, rather than demonstrating meeting specific performance metrics through a detailed clinical or standalone study as would be seen for an "AI/ML" device. This device appears to be a physical appliance, not an AI/ML software device, which changes the nature of the expected performance data.

Therefore, I cannot fully complete the requested table or answer all questions based on the provided text. I will complete what is possible and note where information is missing.


Device Name: Myoaligner Appliance

Device Type (based on context): Physical Medical Appliance (mouthguard/splint), not an AI/ML software device. This significantly impacts the expected type of "study" and "acceptance criteria." The performance testing described is mechanical, not related to an algorithm's diagnostic or predictive accuracy.


1. Table of Acceptance Criteria and Reported Device Performance

Given that this is a physical medical appliance, the "acceptance criteria" appear to be related to its mechanical and material properties, demonstrating equivalence to predicates. The document does not provide specific quantitative acceptance criteria or raw performance data in a table format. Instead, it makes qualitative claims of equivalence.

Performance Metric (Acceptance Criteria Implied)Reported Device Performance
Material Conformance (ISO 10477)"The PMMA material conforms to ISO 10477."
Anterior Segment Retention (Equivalent to NTI-TSS)"confirmed the Myoaligner anterior segment retention is equivalent to reference device NTT-TSS"
Posterior Segment Retention (Equivalent to Gelb Appliance)"the Myoaligner posterior segments are equivalent to the predicate Gelb Appliance"
Effects of Repeated Use (Insertion/Removal)"Performance testing was also done to show the effects of repeated use (insertion and removal)" (No specific results provided, implies acceptable performance.)
Interdigitation and Canine Guidance (via articulating paper)"Performance testing was also done to show... articulating paper to show interdigitation and canine guidance." (No specific results provided, implies acceptable performance.)

Note: The document focuses on demonstrating equivalence to existing devices rather than establishing novel quantitative performance targets. The "study" mentioned is performance testing, likely bench testing, not a clinical trial evaluating diagnostic accuracy or patient outcomes against specific statistical endpoints.


2. Sample Size Used for the Test Set and Data Provenance

The document describes "performance testing" and "biocompatibility assessment" but does not specify sample sizes for these tests. It does not mention a "test set" in the context of a dataset for an AI/ML algorithm.

  • Sample Size: Not specified. (Likely related to number of appliance samples or material samples for mechanical/biocompatibility testing).
  • Data Provenance: Not applicable in the context of patient data for an algorithm. The testing appears to be laboratory-based on the device material/design.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Number of Experts: Not applicable. This is not an AI/ML device relying on expert image annotation or diagnosis for ground truth.
  • Qualifications of Experts: Not applicable. The "ground truth" (or basis for comparison) is conformance to ISO standards and mechanical properties of predicate devices.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. There is no mention of human readers or adjudication processes for interpreting outputs, as this is a physical device and not an AI/ML algorithm.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC Study done? No. This type of study is specifically for evaluating the effectiveness of AI systems (often in medical imaging) on human reader performance. Given that the Myoaligner is a physical appliance, an MRMC study is not relevant to its clearance.
  • Effect Size of Human Reader Improvement: Not applicable.

6. Standalone (Algorithm Only) Performance

  • Was a Standalone Study done? No, because this is not an algorithm. The performance testing refers to the physical properties of the appliance.

7. Type of Ground Truth Used

  • Type of Ground Truth: The "ground truth" for the performance claims appears to be:
    • Material Specifications/Standards: Conformance to ISO 10477 for PMMA.
    • Mechanical Properties of Predicate/Reference Devices: Retention equivalence to NTI-TSS and Gelb Appliance, and observations of repeated use effects and interdigitation/canine guidance (which are functional properties of the device).
    • Biocompatibility: Based on the material composition (PMMA) of the predicates.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. This device does not involve a "training set" as it is a physical appliance, not a machine learning model.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth for Training Set Was Established: Not applicable. No training set for an algorithm is involved. The "ground truth" in the context of device validation relies on established material standards, predicate device characteristics, and engineering principles for mechanical testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

February 2, 2024

Myohealth Technologies, LLC % Colette Cozean, Ph.D. Regulatory Consultant The EyeDeas Company 21581 Midcrest Dr Lake Forest, California 92630

Re: K230548

Trade/Device Name: Myoaligner Appliance Regulatory Class: Unclassified Product Code: MQC, OCO Dated: January 18, 2024 Received: January 19, 2024

Dear Colette Cozean, Ph.D .:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bobak Shirmohammadi -S

For Michael E. Adjodha, M. ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality

{2}------------------------------------------------

Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K230548

Device Name Myoaligner Appliance

Indications for Use (Describe)

For protection against teeth grinding, and jaw clenching; protections from injury due to bruxism or clenching; relief of bruxism related headaches, migraines and pain; short-term relief from muscle spasm due to occlusal interference; prevention of chronic tension and temporal mandibular joint (TMJ) syndrome that is caused by chronic jaw clenching of the mandibular and maxillary teeth by the temporalis muscles; and temporary treatment of temporal mandibular disorder (TMD) along with the relief of associated headaches and pains in adults 18 years of age or older.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

5. 510(k) Summary

Applicant:Myohealth Technologies, INC13765 NW Cornell Rd. Suite 150Portland, Oregon 97229763-363-3336drm@myoaligner.com
Contact Person:Colette Cozean, PhD21581 Midcrest DriveLake Forest, CA 92630(949) 855-2885colettecozean@gmail.com
Date PreparedFebruary 2, 2024
Proprietary NameMyoaligner Appliance
Common NameMouthguard, Prescription
Classification NameUnclassified
Product CodeMQC
Subsequent ProductCodeOCO
Primary PredicateDeviceSmartGuard Night Guard, Original, Night Guard Original and Elite (K123161)
Predicate DevicesSomnobrux Splints including Michigan, ComnBrux B, Gelb and Tanner(K102909), Serena Sleep Night Guard (K230495)
Reference DevicesNTI-TSS (K010876) and The POD (K182820) for comparison in performancetesting, PMMA Blocks from Huge Dental Material Co. (K201683)

Description: The Myoaligner appliance is a customized, removable dental device which is designed to overlay the existing dentition without the need for any alterations done to the receiving dentition. The appliance is a plural piece intra-oral removable appliance, which is designed and delivered to the patients by a licensed practitioner such as a dentist (Rx only). The dentist can prescribe a device that fits over the 4-6 front teeth, back teeth on both sides of the dental arch or all of the teeth in the dental arch both on the upper or lower jaw. The appliance is held in place by mechanical retention. The Myoaligner appliance can be worn on either the top or bottom teeth or both. It is intended for use in adults 18 years of age and older.

Indications for Use: For protection against teeth grinding, bruxism, and jaw clenching; protection of restorations from injury due to bruxism or clenching; relief of bruxism related headaches, migraines and pain; short-term relief from muscle spasm due to occlusal interference; prevention of chronic tension and temporal mandibular joint (TM) syndrome that is caused by chronic jaw clenching of the mandibular and maxillary teeth by the temporalis muscles; and temporary treatment of temporal mandibular disorder (TMD) along with the relief of associated headaches and pains in adults 18 years of age or older.

Technological Characteristics, Components and Materials, Manufacturing Methods: The appliance is digitally designed based on the specific dental records (stone models or digital 3D scans of the patient's maxillary and mandibular arches), which are provided by the dentist. After being digitally designed to the dentists' specifications, the oral appliance is fabricated by an FDA registered milling the process of milling a polymethyl methacrylate (PMMA) block. These precision CAD CAM milling machines are currently used for manufacturing orthotics in the dental industry. A predicate, Serena Sleep Night Guard, is milled in this fashion.

{5}------------------------------------------------

DEVICE ILLUSTRATIONS AND DRAWINGS

Figure 1 below provides an overall view of the Myoaligner device is intended to overlay the dentition. As displayed, the Myoaligner is form-fit to the device is capable of functioning even without use of adhesives. Myoaligner is a plural piece device that is fabricated to overlay one section of the dental arch or several sections of the dental arch depending on the specific prescription that was submitted by the dental provider. In the images below, the Myoaligner is illustrated in 3 individual piece sections (2 posterior and 1 anterior section) to overlay the entire lower dental arch. The designed based on the specific prescription that was submitted by the dental provider and at times may only cover sections of the dental arch or the entire arch.

Image /page/5/Figure/2 description: The image shows different views of the Myoaligner device and its fit to the dentition. There are six images labeled a-f, showing anterior, posterior, superior, and bottom views of the device on a dental model. The image also shows the device in three components and a full arch. The text describes the images as showing the precise fit of the Myoaligner on the dentition model and the different views of the device.

Mechanism of Action: Removable dental appliance therapy is a treatment modality in which the patient wears a removeable appliance that overlays the teeth or dental arches. It prevents the occlusal inteferences from coming into occlusion, which relaxes the jaw muscles and helps to reduce jaw clinching and grinding and reduces associated symptoms such as headaches, migraines, pain and muscle spasming. The same mechanism of action is used with mouthguards and the predicate devices.

{6}------------------------------------------------

Substantial Equivalence:

Subject DevicePrimary Predicate DevicePredicate DevicePredicate Device
Trade NameMyoalignerSmartGuard Night Guard,Original, Night GuardOriginal and EliteSomnoburx Splints includingMichigan, ComnBrux B, Gelband TannerSerena Sleep Night Guard
OwnerMyohealth TechnologiesSmartGuardSomnomedSerena Sleep
510(k) NumberK123161K102909K230495
Image: teeth with yellow substance on themImage: SmartGuard Elite Dental Guard packagingImage: SomnoDent splintImage: Serena Sleep Night Guard
Intended Use
Indication for UseOTC:
> Protection againstbruxism or nighttime teethgrinding
> Intended to reducedamage to the teeth andreduce noise associatedwith bruxing and/orgrinding.
Rx:Rx:Rx:Rx:
> Protection of teeth grinding,bruxism and jaw clenching> Protection of teethgrinding, bruxism and jawclenching>Protection of teeth andrestoration from the forcesof Bruxism>Protect teeth and restorationsagainst the forces of bruxismand alleviatetemporomandibular joint, jawand muscle and tensionheadache pains
>Protection of restoration frominjury due to bruxism orclenching>Protection of restorationfrom injury due to bruxismor clenching
>Relief of bruxism related toheadaches and pains>Relief of bruxism related toheadaches and pains
> Short-term pain relief frommuscle spasm due to occlusalinterference; for the preventionof chronic tension and temporalmandibular joint (TMJ)syndrome that is caused bychronic jaw clenching of themandibular and maxillary teethby the temporalis muscle>Temporary treatment ofTemporal Mandibular Disorder> Short-term pain relieffrom muscle spasm due toocclusal interference; forthe prevention of chronictension and temporalmandibular joint (TMJ)syndrome that is caused bychronic jaw clenching of themandibular and maxillaryteeth by the temporalismuscle>Temporary treatment ofTemporal Mandibular
(TMD) along with the relief ofassociated headaches and painsDisorder (TMD) along withthe relief of associatedheadaches and pains
ClassUnclassifiedUnclassifiedUnclassifiedUnclassified
Product CodeMQC, OCOMQC, OBRMQCMQC, OCO
Intended forsingle patientmultiuseYesYesYesYes
Design
Customized fit foreach patientYes, in the labYes, at home boil-n-biteYes, in the labYes, in the lab
TraysUpper and Lower Tray (full andpartial)Upper Tray Only (6-8 teeth),bottom teeth contacts thebottom of the substrateMichigan and ComnBrux fiton the maxillary arch; Gelband Tanner fit themandibular arch; Gelb fitsthe posterior teethCustom-made intraoral devicethat is fitted to the patient'smouth via computer aideddesign and manufacturing
AdhesivesNoneNoneNoneNone
Cleaned andinspected daily bypatientYesYesYesYes
Composition
MaterialPMMAElvaxPMMAPolyamide 12 (PA 2200)
Mechanism
Mechanism ofActionPrevent teeth grinding andrelax muscles by placingappliance between the teethPrevent teeth grinding andrelax muscles by placingappliance between theteethPrevent teeth grinding andrelax muscles by placingappliance between theteethDisocclusion

{7}------------------------------------------------

Clinical and Non-Clinical Data: A biocompatibility and physical properties assessment was completed based on the material composition of the predicates, which concluded that the subject device was substantially equivalent to the predicate devices. The Company uses the PMMA material from Huge Dental Material Co with no changes to the chemical composition, no surface coating or treatments and no additional dyes or colorants. The tint is selected from a selection of tints provided by Huge Dental. A risk assessment has also been conducted with the subject device, which concluded there are no additional risks as compared to the predicate device(s).

Performance Testing: The PMMA material conforms to ISO 10477. Performance testing confirmed the Myoaligner anterior segment retention is equivalent to reference device NTT-TSS, the Myoaligner posterior segments are equivalent to the predicate Gelb Appliance. Performance testing was also done to show the effects of repeated use (insertion and removal) and articulating paper to show interdigitation and canine guidance.

Manufacturing: The Myoaligner is manufactured in the same manner as the predicate device, Serena Sleep Night Guard. The Myoaligner is manufactured from PMMA blocks manufactured by Huge Dental Material Co (K201683). Each PMMA material manufacturer has met the ISO 10477 guidance and has their own submission on file with the FDA including composition, physical properties and tints.

Differences:

  • Myoaligner offers full or partial upper and lower trays. SmartGuard fits the front 6-8 teeth and NTT 1. (reference device for mechanical testing) fits the front 4-6 teeth as does one Myoaligner version. Somnobrux manufactures 4 versions, 2 upper trays and 2 lower trays. Michigan and Tanner splints cover

{8}------------------------------------------------

the whole arch while the Gelb splint covers the posterior teeth only. One of these predicates cover each of the different Myoaligner designs. The POD (reference device for mechanical testing) also covers all teeth.

    1. The Myoaligner can be prescribed for use during the day or night for 8-10 hours per day, while the predicates are prescribed at night only except for The POD which is prescribed for nighttime use up to 12 hours and the NTI-TSS, which can be prescribed for day or night use.
      Summary: Based on the intended use, technical characteristics, biocompatibility assessment, manufacturing process, labeling and other data provided in this submission, the Myoaligner appliance demonstrates substantial equivalence to the predicate devices and reference devices in both safety and efficacy.

N/A