K Number
K230503

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-04-25

(60 days)

Product Code
Regulation Number
870.1220
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Blazer Dx-20 Catheter Cable:

The BSC Cables are intended to be used with the BSC Cardiac Diagnostic Catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials.

Polaris X Catheter Cable:

The BSC Cables are intended to be used with the BSC Cardiac Diagnostic Catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials.

Device Description

Blazer™ Dx-20 Cable:

The subject Blazer Dx-20 Catheter cable (sold separately) is used with the Blazer Dx-20 Bidirectional Duodecapolar Catheter during an electrophysiology procedure. The Blazer Dx-20 Catheter Cable provides an electrical connection between the Blazer Dx-20 Diagnostic Catheter and the Electrophysiology (EP) Recording System. The cable has one connector that directly connects to the diagnostic catheter and multiple connection points that connect to an EP Recording System.

The main trunk consists of a jacketed 20-conductor cable that is split into individual leads by means of an over-molded splitter yoke. The proximal end of the cable has 20 universal 2 mm shrouded male pins that are typical to electrophysiology recording and stimulation systems. The Blazer Dx-20 cable is provided sterile using an EO sterilization process. The cable is reusable and can be reprocessed (i.e. re-sterilized) up to 10 times using either EO or STERRAD™ sterilization methods. The Blazer Dx-20 cable is to be used in a fully-equipped electrophysiology lab by physicians thoroughly trained in invasive cardiac mapping and ablation procedures.

Polaris X™ Cable:

The subject Polaris X Catheter Cable (sold separately) is used with the Polaris X Steerable Decapolar Mapping Catheters during an electrophysiology procedure. The Polaris X cable provides an electrical connection between the Polaris X Catheter and the Electrophysiology (EP) Recording System. The cable has one connector that directly connects to the diagnostic catheter and multiple connection points that connect to an EP Recording System.

The main trunk consists of a jacketed 10-conductor cable that is split into individual leads by means of an over-molded splitter yoke. The proximal end of the cable has 10 universal 2 mm shrouded male pins that are typical to electrophysiology recording and stimulation systems. The Polaris X cable is provided sterile using an EO sterilization process. The cable is reusable and can be reprocessed (i.e. re-sterilized) up to 10 times using EO sterilization method. The Polaris X cable is to be used in a fully-equipped electrophysiology lab by physicians thoroughly trained in invasive cardiac mapping and ablation procedures.

AI/ML Overview

The provided document, a 510(k) summary for the Blazer™ Dx-20 Catheter Cable and Polaris X™ Catheter Cable, does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of AI/ML device performance. The document describes the device, its intended use, comparison with predicate devices, and performance testing related to bioburden, reusability (manual cleaning and resterilization), and general design verification.

Therefore, I cannot provide the detailed information requested in the prompt based on the provided text, as it pertains to medical device regulatory clearance for cables used in electrophysiology procedures, not an AI/ML powered device.

Specifically, the document does not include:

  1. A table of acceptance criteria and reported device performance for an AI/ML algorithm. The performance data section refers to bioburden testing, reusable device testing (cleaning and resterilization), and design verification testing for the physical cables.
  2. Sample sizes used for a test set or data provenance related to AI/ML.
  3. Number of experts or their qualifications for establishing ground truth for an AI/ML test set.
  4. Adjudication methods for an AI/ML test set.
  5. Information about a multi-reader multi-case (MRMC) comparative effectiveness study or human reader improvement with AI assistance.
  6. Standalone (algorithm only) performance data.
  7. Type of ground truth used for AI/ML (expert consensus, pathology, outcomes data, etc.).
  8. Sample size for an AI/ML training set.
  9. How ground truth for an AI/ML training set was established.

The document primarily focuses on demonstrating the substantial equivalence of the new catheter cables to existing predicate devices based on physical and functional characteristics, sterility, and reprocessing capabilities.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

April 25, 2023

Boston Scientific Corporation Stephanie Andre Principal Regulatory Affairs Specialist 4100 Hamline Ave North St. Paul, Minnesota 55112

Re: K230503

Trade/Device Name: Blazer™ Dx-20 Catheter Cable, Polaris X™ Catheter Cable Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: DRF Dated: February 23, 2023 Received: February 24, 2023

Dear Stephanie Andre:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh S. Deoras -S

Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230503

Device Name

Blazer™ Dx-20 Catheter Cable Polaris XTM Catheter Cable

Indications for Use (Describe)

Blazer Dx-20 Catheter Cable:

The BSC Cables are intended to be used with the BSC Cardiac Diagnostic Catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials.

Polaris X Catheter Cable:

The BSC Cables are intended to be used with the BSC Cardiac Diagnostic Catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary for Blazer Dx-20 Cable and Polaris X Cable K230503

I. Submitter

Boston Scientific Corporation Electrophysiology Division 4100 Hamline Ave North St. Paul, MN 55112

Contact: Stephanie Andre Principal Regulatory Affairs Associate Phone: 651-582-5544 E-mail: stephanie.andre@bsci.com

Date Prepared: February 23, 2023

II. Device

Name of Device:Blazer™ Dx-20 Catheter Cable
Polaris X™ Catheter Cable
Common or Usual Name:Blazer Dx-20 Cable
Polaris X Cable
Classification Name:Electrode recording catheter or electrode recording probe (21 CFR
870.1220)
Regulatory Class:II
Product Code:DRF

III. Predicate Device

Subject DevicePredicate DevicePredicate 510(k)
BlazerTM Dx-20 Catheter CableBlazerTM Dx-20 Catheter CableK081576
Polaris XTM Catheter CableElectrophysiology CableK924163

The predicate cables have not been subject to a design-related recall.

No reference devices were used in this submission.

IV. Device Description

Blazer™ Dx-20 Cable:

The subject Blazer Dx-20 Catheter cable (sold separately) is used with the Blazer Dx-20 Bidirectional Duodecapolar Catheter during an electrophysiology procedure. The Blazer Dx-20

{4}------------------------------------------------

Catheter Cable provides an electrical connection between the Blazer Dx-20 Diagnostic Catheter and the Electrophysiology (EP) Recording System. The cable has one connector that directly connects to the diagnostic catheter and multiple connection points that connect to an EP Recording System.

The main trunk consists of a jacketed 20-conductor cable that is split into individual leads by means of an over-molded splitter yoke. The proximal end of the cable has 20 universal 2 mm shrouded male pins that are typical to electrophysiology recording and stimulation systems. The Blazer Dx-20 cable is provided sterile using an EO sterilization process. The cable is reusable and can be reprocessed (i.e. re-sterilized) up to 10 times using either EO or STERRAD™ sterilization methods. The Blazer Dx-20 cable is to be used in a fully-equipped electrophysiology lab by physicians thoroughly trained in invasive cardiac mapping and ablation procedures.

Polaris X™ Cable:

The subject Polaris X Catheter Cable (sold separately) is used with the Polaris X Steerable Decapolar Mapping Catheters during an electrophysiology procedure. The Polaris X cable provides an electrical connection between the Polaris X Catheter and the Electrophysiology (EP) Recording System. The cable has one connector that directly connects to the diagnostic catheter and multiple connection points that connect to an EP Recording System.

The main trunk consists of a jacketed 10-conductor cable that is split into individual leads by means of an over-molded splitter yoke. The proximal end of the cable has 10 universal 2 mm shrouded male pins that are typical to electrophysiology recording and stimulation systems. The Polaris X cable is provided sterile using an EO sterilization process. The cable is reusable and can be reprocessed (i.e. re-sterilized) up to 10 times using EO sterilization method. The Polaris X cable is to be used in a fully-equipped electrophysiology lab by physicians thoroughly trained in invasive cardiac mapping and ablation procedures.

UPN and Technical Description

Each cable has a single configuration. The UPN and general technical specifications of the Blazer Dx-20 and Polaris X cables are provided in Table 1.

UPNProduct DescriptionLengthNumber of ContactsReprocessing MethodJacket Color
M00420S0Blazer Dx-20Catheter Cable5 ft20EOSTERRADGrey
M0045454S0Polaris XCatheter Cable5 ft10EOBlack

Table 1: Electrophysiology Cable Descriptions and UPNs

{5}------------------------------------------------

V. Intended Use

The BSC cables are intended to be used with the BSC cardiac diagnostic catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials.

VI. Comparison of Technological Characteristics with the Predicate Device

The Blazer Dx-20 Cable and Polaris X Cable incorporate substantially equivalent design, packaging, fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate devices. The non-patient contacting material changes are the primary technological differences between the subject and predicate devices.

Performance Data VII.

The following performance data were provided in support of the substantial equivalence determination:

Bioburden Testing

Bioburden testing, per ISO 11737-1, Microbiological Methods - Part 1: Determination of a Population of Microorganisms on Products, was performed to confirm bioburden levels of the cables were not adversely affected.

Reusable Device Testing

The subject Blazer Dx-20 and Polaris X cables underwent 10x manual cleaning and resterilization qualification. The cables were subjected to 2X EO sterilization in qualified EO cycles and then 10X manual cleaning and resterilization cycles using either EO or STERRAD methods, as described in the Blazer Dx-20 and Polaris X cable IFUs, to confirm the cable design meets its predetermined design input specifications.

Design Verification Testing

Design verification testing was performed on a subset of product specifications to provide objective evidence that after 2X EO sterilization cycles, distribution challenge, climatic conditioning, and 10X manual cleaning and resterilization cycles, the Blazer Dx-20 and Polaris X cables meet the design input specifications in accordance with the applicable product specifications.

{6}------------------------------------------------

VIII. Conclusion

The subject Blazer Dx-20 Catheter Cable and Polaris X Catheter Cable are substantially equivalent in indications for use, design, product function, materials, and sterility to the predicate Blazer Dx-20 Catheter Cable and predicate Electrophysiology Cable, respectively. The performance data, including sterilization, and design verification testing supports that the subject catheters are substantially equivalent to their respective predicate devices.

§ 870.1220 Electrode recording catheter or electrode recording probe.

(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).