K Number
K230365

Validate with FDA (Live)

Device Name
Sonio Detect
Manufacturer
Date Cleared
2023-07-25

(165 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sonio Detect is intended to analyze fetal ultrasound images and clips using machine learning techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria of the views.
The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.

Device Description

Sonio Detect is a Software as a Service SaaS solution that aims at helping sonographers, OB/GYNs, MFMs and Fetal surgeons (all three designated as healthcare professionals i.e. HCP) to perform their routine fetal ultrasound examinations in real-time. Sonio Detect can be used by Healthcare Professionals HCPs during fetal ultrasound exams for Trimester 1, Trimester 2 and Trimester 3 of the fetus (Gestational Age: from 11 weeks to 37 weeks). The software is intended to assist HCPs in assuring during and after their examination that the examination is complete and all images were collected according to their protocol.
Sonio Detect receives fetal ultrasound images and clips from the ultrasound machine, that are submitted through the edge software by the performing healthcare professional, in real-time and performs the following:

  • . Automatically detect views;
  • Automatically detect anatomical structures within the supported views; .
  • Automatically verify quality criteria of the supported views by checking whether they . conform to standardized quality criteria.
    Quality criteria are related to:
  • the presence or absence of an anatomical structure; ●
  • the zoom level for some views.
    Sonio Detect then automatically associates the image to its detected view. It also highlights in yellow the view and/or the corresponding quality criteria if there are unverified items : quality criteria not verified or view not detected.
    The end user can interact with the software to override the Sonio Detect's outputs (reassign the image to another view or unassign it or assign it if it was not assigned, change the status of a quality criteria from verified to unverified or from unverified to verified). The user has the ability to review and edit/override the matching at any time during or at the end of the exam.
AI/ML Overview

Sonio Detect Acceptance Criteria and Study Details

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for Sonio Detect are implicitly defined by the reported performance metrics, which the FDA has deemed sufficient for substantial equivalence. The reported performance is presented as sensitivities, specificities, and proportions of correctly read annotations.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
3D Fetal Ultrasound Image Detection SensitivityHigh sensitivity0.980 (95% Wilson's CI: 0.930, 0.994)
Doppler Fetal Ultrasound Image Detection SensitivityHigh sensitivity0.963 (95% Wilson's CI: 0.908, 0.985)
Fetal Ultrasound Views Detection Proportion CorrectHigh proportion0.923 (95% Wilson's CI: 0.905, 0.938)
T1 Fetal Ultrasound Views Detection SensitivityHigh sensitivity0.942 (Point estimate)
T2/T3 Fetal Ultrasound Views Detection SensitivityHigh sensitivity0.919 (Point estimate)
T2/T3 Fetal Brain Anatomical Structure Detection SensitivityHigh sensitivity0.857 (Point estimate)
T2/T3 Fetal Brain Anatomical Structure Detection SpecificityHigh specificity0.963 (Point estimate)
T2/T3 Fetal Heart Anatomical Structure Detection SensitivityHigh sensitivity0.900 (Point estimate)
T2/T3 Fetal Heart Anatomical Structure Detection SpecificityHigh specificity0.982 (Point estimate)
Zoom Level Verification Sensitivity (Brain Views)High sensitivity0.952 (95% Wilson's CI: 0.909-0.976)
Zoom Level Verification Specificity (Brain Views)High specificity0.906 (95% Wilson's CI: 0.758-0.968)

2. Sample Size and Data Provenance for Test Set

  • Sample Size: 17,885 fetal ultrasound images.
  • Data Provenance: The data was collected from 7 clinical sites in the United States, France, and Germany. This indicates a multi-national dataset. The data was retrospective as it was "independent of the data used during model development (training/fine tuning/internal validation) and establishment of device operating points."

3. Number of Experts and Qualifications for Ground Truth (Test Set)

The document does not explicitly state the number of experts used or their specific qualifications (e.g., years of experience) for establishing the ground truth of the test set. However, it indicates that the device automatically detects fetal ultrasound views "through reading of annotations on images." This implies that human experts (presumably sonographers, OB/GYNs, MFMs, or Fetal surgeons, as these are the intended users) provided the initial annotations that served as the ground truth.

4. Adjudication Method (Test Set)

The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for establishing the ground truth of the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was performed. The document explicitly states: "Clinical Study: Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device." Therefore, there is no reported effect size of human readers improving with AI vs. without AI assistance.

6. Standalone Performance Study

Yes, a standalone performance study was done. The document states: "Sonio conducted a standalone performance testing on a dataset of 17885 fetal ultrasound images..." This indicates the algorithm's performance was evaluated without human intervention in the loop during the assessment of the test set.

7. Type of Ground Truth Used (Test Set)

The ground truth for the test set was established through "reading of annotations on images." This suggests the ground truth was based on expert annotations or labeling of the ultrasound images, likely by the qualified healthcare professionals who generated the initial data.

8. Sample Size for Training Set

The document does not explicitly state the sample size for the training set. It refers to "data used during model development (training/fine tuning/internal validation)" but does not provide a specific number of images or cases for this phase.

9. How Ground Truth for Training Set was Established

The method for establishing the ground truth for the training set is not explicitly detailed. However, given that the test set's ground truth was based on annotations, it is highly probable that the training set's ground truth was established through a similar process of expert annotation or labeling of the fetal ultrasound images and clips.

{0}------------------------------------------------

July 25, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Sonio % Florian Akpakpa Head of Regulatory Affairs and Quality Assurance 24 Rue du Faubourg Saint Jacques Paris. FR-75014 FRANCE

Re: K230365

Trade/Device Name: Sonio Detect Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, QIH Dated: June 26, 2023 Received: June 27, 2023

Dear Florian Akpakpa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yanna S. Kang -S

Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)K230365
Device NameSonio Detect
Indications for Use (Describe)Sonio Detect is intended to analyze fetal ultrasound images and clips using machine learning techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria of the views.The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.
Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Sonio. The logo consists of a stylized, abstract symbol on the left, resembling a curved shape with a dot above it. To the right of the symbol is the word "sonio" in lowercase letters. The color of both the symbol and the text is a shade of blue.

K230365

510(k) Summary

In accordance with 21 CFR 807.92 the 510(k) summary for Sonio Detect is provided below.

I. Submitter

Applicant:Sonio24 rue du Faubourg Saint Jacques,75014, Paris France
Primary Contact Person:Florian AkpakpaHead of Regulatory Affairs and Quality AssuranceSonioPhone: +33 6 19 38 71 45Email: florian.akpakpa@sonio.ai
Secondary Contact Person:Donna-Bea TillmanSenior ConsultingBiologics ConsultingPhone: +1 (410) 531-6542 - DirectEmail: dtillman@biologicsconsulting.com
Date Prepared:July 25, 2023

II. Device

Device Trade Name:Sonio Detect
Classification Name:21 CFR 892.1550 - accessory to Ultrasonic Pulsed Doppler Imaging System21 CFR 892.1560 - accessory to Ultrasonic Pulsed Echo Imaging System21 CFR 892.2050 - Medical image management and processing system
Regulatory Class:Class II
Product Codes:IYN (Primary)IYO, QIH (Secondary)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Sonio. The logo consists of a stylized blue icon resembling a sound wave or a stylized letter 'S', followed by the word 'sonio' in a sans-serif font, also in blue. The logo is simple and modern, with a clean design.

III. Predicate Device

SonoLyst feature embedded in the GE Medical device Voluson SWIFT, K201828.

This predicate has not been subject to a design-related recall.

No reference devices were used in this submission.

IV. Device Description

Sonio Detect is a Software as a Service SaaS solution that aims at helping sonographers, OB/GYNs, MFMs and Fetal surgeons (all three designated as healthcare professionals i.e. HCP) to perform their routine fetal ultrasound examinations in real-time. Sonio Detect can be used by Healthcare Professionals HCPs during fetal ultrasound exams for Trimester 1, Trimester 2 and Trimester 3 of the fetus (Gestational Age: from 11 weeks to 37 weeks). The software is intended to assist HCPs in assuring during and after their examination that the examination is complete and all images were collected according to their protocol.

Sonio Detect requires the following:

  • Edge Software (described below) to install on a server on the same network as the ● Ultrasound Machine;
  • SaaS accessibility from internet browser.

Sonio's Edge Software is a light-weight application that runs on a server (computer) connected to the same network as the Ultrasound Machine. Sonio Edge Software is installed on the HCP server (computer) and network and the main purpose is to receive DICOM instances from the Ultrasound Machine and upload them to Sonio's Cloud to be used by Sonio Detect.

Sonio Detect receives fetal ultrasound images and clips from the ultrasound machine, that are submitted through the edge software by the performing healthcare professional, in real-time and performs the following:

  • . Automatically detect views;
  • Automatically detect anatomical structures within the supported views; .
  • Automatically verify quality criteria of the supported views by checking whether they . conform to standardized quality criteria.

Quality criteria are related to:

  • the presence or absence of an anatomical structure; ●
  • the zoom level for some views.

Sonio Detect then automatically associates the image to its detected view. It also highlights in yellow the view and/or the corresponding quality criteria if there are unverified items : quality criteria not verified or view not detected.

The end user can interact with the software to override the Sonio Detect's outputs (reassign the image to another view or unassign it or assign it if it was not assigned, change the status of a

{5}------------------------------------------------

Sonio

510(k) Premarket Notification Submission

quality criteria from verified to unverified or from unverified to verified). The user has the ability to review and edit/override the matching at any time during or at the end of the exam.

The list of views, anatomical structures and quality criteria that can be automatically detected and verified by the software are detailed in the tables 1, 2 and 3 below.

Table 1: List of views that can be automatically detected by the software
ViewGestational Age of the fetus
Transthalamic or Cavum septum Pellucidum or Midline falx /Transventricular or Choroid plexusT1 and T2/T3
Profile or Nuchal translucencyT1 and T2/T3
Crown rump lengthT1
Sagittal spineT2/T3
Abdominal circumferenceT2/T3
Long boneT2/T3
Transcerebellar viewT2/T3
Upper lip, nose and nostrilsT2/T3
Four chambersT2/T3
Left ventricular outflow tractT2/T3
Right ventricular outflow tractT2/T3
Table 2:List of anatomical structures that can be automatically detected by the software
ViewStructure
Brain views and structures at T2/T3
Transthalamic view• Midline falx
Transventricular view• Cavum septum pellucidum
Transcerebellar view• Cerebellum
Heart views and structures at T2/T3
Four Chambers• Aorta
Three vessels• Apex heart
LVOT• Ascending aorta
• Descending aorta
• Interatrial septum
• Interventricular septum
• Left atrium
• Left ventricle
• Pulmonary trunk
• Right atrium
• Right ventricle
• Superior vena cava

{6}------------------------------------------------

Quality criteria of the brain views
For the transthalamic view, SonioDetect automatically evaluates thefollowing criteria:Presence of the cavum septum pellucidum Absence of the cerebellum Brain occupies more than half of the width of theultrasound image
For the transcerebellar view, SonioDetect automatically evaluates thefollowing criteria:Presence of the cerebellum Presence of the cavum septum pellucidum Brain occupies more than half of the width of theultrasound image
For the transventricular view, SonioDetect automatically evaluates thefollowing criteria:Presence of the cavum septum pellucidum Brain occupies more than half of the width of theultrasound image
Quality criteria of the heart views
For the 4 chambers view, SonioDetect automatically evaluates thefollowing criteria:Presence of the right ventricle Presence of the left ventricle Presence of the left atrium Presence of the right atrium Presence of the interventricular septum Presence of the interauricular septum Presence of the apex of the heart Presence of the descending aorta
For the 3 vessels and 3 vessels andtrachea views, Sonio Detectautomatically evaluates the followingcriteria:Presence of the pulmonary trunk Presence of the ascending aorta Presence of the superior vena cava
For the LVOT view, Sonio Detectautomatically evaluates the followingcriteria:Presence of the left ventricle Presence of the aorta Presence of the right ventricle Presence of the left atrium Presence of the apex of the heart

Table 3: List of quality criteria that can be automatically verified by the software

{7}------------------------------------------------

V. Indications for Use

Sonio Detect is intended to analyze fetal ultrasound images and clips using machine learning techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria of the views.

The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.

Sonio Detect and the predicate device have the same intended use for automatic detection of views as well as automatic detection of anatomical structures within the views and verification of quality criteria of the views by comparing them to standardized quality criteria. The Indications for Use statement for Sonio Detect is not identical to the predicate device: however. the difference does not affect the safety and effectiveness of the device relative to the predicate, and so does not constitute a new intended use.

VI. Comparison of Technological Characteristics with the Predicate Device

Table 4 provides a comparison of the Technological Characteristics of Sonio Detect to the predicate SonoLyst.

ItemsPredicate device 1: SonoLyst feature inVoluson SWIFT - K201828Proposed device: Sonio Detect
Manufacturer nameGE MedicalSonio
Device nameSonoLyst feature embedded in the deviceVoluson SWIFTSonio Detect
Regulation Number- Ultrasonic Pulsed Doppler Imaging System. 21 CFR 892.1550, 90-IYN;-Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560, 90-IYO;-Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX- Accessory to Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550- Accessory to Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560- Medical image management and processing system, 21 CFR 892.2050
Product codesIYN (primary)IYOITX (secondary)IYN (Primary)IYO, QIH (Secondary)
Clinical outcome- Images labeled with correct view- Quality criteria are identified as “found” when detected and “not found” when not detected- Images labeled with correct view- Quality criteria are identified as “Verified” when detected and “Not verified” when not detected

Table 4: Comparison of technological characteristics

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Sonio. The logo consists of a blue abstract symbol on the left and the word "sonio" in blue on the right. The symbol appears to be a stylized representation of a sound wave or a stylized letter S.

Sonio 510(k) Premarket Notification Submission

ItemsPredicate device 1: SonoLyst feature inVoluson SWIFT - K201828Proposed device: Sonio Detect
IntendedUsersGeneral purpose radiology evaluation andspecialized for OB/GYNQualified and trained healthcare professionalpersonnel in a professional prenatalultrasound (US) imaging environment (thisincludes sonographers, MFMs, OB/GYN,and Fetal surgeons)
ClinicalapplicationsFetal/ObstetricsFetal/Obstetrics
AlgorithmMethodologyArtificial IntelligenceArtificial IntelligenceLecture of biometricsColorimetry for 3D and Doppler
PlatformEmbedded in the ultrasound equipmentSecure cloud-based and stand-alone softwarecompatible with ultrasound system from GEMedical, Samsung and Canon

Sonio Detect's intended users, clinical outcome and clinical applications are similar to those of the predicate device, SonoLyst.

Sonio Detect differ to SonoLyst in the following:

  • Algorithm methodology: Sonio Detect algorithm technology is based on Artificial ● Intelligence, Lecture of biometrics on the image and Colorimetry identification for 3D and Doppler while the Predicate SonoLyst's algorithm technology is only based on Artificial Intelligence:
  • Platform: Sonio Detect is a stand-alone cloud based software that can be used with . different ultrasound systems while the Predicate SonoLyst is embedded in the GE ultrasound system.

However, the differences in algorithm methodology and platform do not raise different questions of safety and effectiveness of the device.

VII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Software Verification and Validation Testing

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

The following quality assurance measures were applied to the development of the system:

  • Risk Analysis
  • Design Reviews ●
  • Software Development Lifecycle ●
  • Algorithm Verification (Algorithm internal validation)
  • Software verification ●
  • Simulated use testing (Validation) ●
  • Performance testing (Verification)

{9}------------------------------------------------

Bench Testing

Sonio conducted a standalone performance testing on a dataset of 17885 fetal ultrasound images from 7 clinical sites in the United States, France and Germany representative of the intended use population. This dataset was independent of the data used during model development (training/fine tuning/internal validation) and establishment of device operating points.

The results of the standalone performance testing demonstrate that Sonio Detect:

  • Automatically detects 3D fetal ultrasound images with high sensitivity (0.980: 95% . Wilson's Confidence Interval: 0.930, 0.994) and Doppler fetal ultrasound images with high sensitivity (0,963; 95% Wilson's Confidence Interval: 0.908, 0.985);
  • . Automatically detects fetal ultrasound views through reading of annotations on images with high proportions of annotations read correctly (0,923; 95% Wilson's Confidence Interval: 0.905, 0.938);
  • Automatically detects some T1 fetal ultrasound views with high sensitivity (0,942; ● Point estimate).
  • Automatically detects some T2/T3 fetal ultrasound views with high sensitivity (0,919; ● Point estimate).
  • Automatically detects some fetal brain anatomical structures in some T2/T3 brain views with high sensitivity (0.857; Point estimate) and high specificity (0.963; Point estimate), and so automatically verifies the corresponding quality criteria;
  • Automatically detects some fetal heart anatomical structures in the some T2/T3 heart views with high sensitivity (0,900; Point estimate) and high specificity (0,982; Point estimate) and so automatically verifies the corresponding quality criteria.
  • . Automatically verifies the zoom level for some brain views with high sensitivity (0.952; 95% Wilson's Confidence Interval: 0.909-0.976) and high specificity (0.906; 95% Wilson's Confidence Interval: 0.758-0.968).

Additionally the performance was also validated for subgroups including: Ultrasound machine manufacturer, BMI, maternal age, gestational age and ethnicity when appropriate.

Sonio Detect was validated only with GE, Canon, and Samsung ultrasound devices and is intended only to be used with these ultrasound vendors.

The results of verification and performance testing demonstrate the safe and effective use of Sonio Detect.

Clinical Study

Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for Sonio. The logo consists of a stylized blue icon resembling a curved shape with a dot, followed by the word "sonio" in blue, lowercase letters. The logo has a clean and modern design.

VIII. Conclusions

Sonio Detect's intended users, clinical outcome and clinical applications are similar to those of the predicate device SonoLyst.

The technological characteristics differences identified and discussed in Section VI do not raise different questions of safety and effectiveness of the device.

Furthermore, results of successful verification and validation activities and additional bench performance testing do not raise any new issue regarding the safety and effectiveness of the device.

Thus, Sonio Detect is substantially equivalent to its predicate device SonoLyst (K201828).

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.