K Number
K230350

Validate with FDA (Live)

Device Name
VISULAS yag
Date Cleared
2023-09-22

(225 days)

Product Code
Regulation Number
886.4390
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VISULAS yag is intended for use in photodisrupting ocular tissue in the treatment of diseases of the eye, including Posterior capsulotomy, Iridotomy and Posterior Membranectomy.

This device is for Prescription Use (Rx) only.

Device Description

VISULAS yag uses a Q-switched, flashlamp-pumped solid-state laser for photodisruption treatments of diseases of the eye, including posterior capsulotomy, iridotomy and membranectomy. Laser radiation is generated by means of a neodymium-doped yttrium aluminum garnet (Nd:YAG) gain medium inside the laser source. The emitted laser radiation with a near-infrared wavelength of = 1064 nm has a pulse duration of < 4 ns (full-width half-maximum; FWHM) and a focal diameter of 6.5 um ± 20%. The maximum energy output per pulse is 9 to 13 mJ in single-burst mode.

AI/ML Overview

I am sorry, but the provided text is a 510(k) summary for a medical device (VISULAS yag), which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study design with acceptance criteria and device performance results as requested.

The document does not contain:

  • A table of acceptance criteria and reported device performance for a specific study.
  • Information on sample size, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, or ground truth for a clinical study.
  • Details regarding training set size or how ground truth was established for a training set.

Instead, the document primarily compares the subject device's indications for use and technical characteristics to two predicate devices (Ellex YAG Laser K212630 and VISULAS YAG III K042139) to establish substantial equivalence. It mentions "Non-Clinical Performance Testing" and "functional and system level testing showed that the system met the defined specifications," and "Software verification and validation testing were conducted... All testing passed," but these are general statements and do not provide the detailed study information you've asked for.

Therefore, I cannot fulfill your request based on the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 22, 2023

Carl Zeiss Meditec Inc Tanesha Bland Sr. Regulatory Affairs Specialist 5300 Central Parkway Dublin, California 94568

Re: K230350

Trade/Device Name: VISULAS yag Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQF Dated: August 16, 2023 Received: August 16, 2023

Dear Tanesha Bland:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bennett Digitally signed by Bennett N. ﮯ Walker -S Date: 2023.09.22 Walker -S 10:23:07 -04'00'L

for Claudine Krawczyk Acting Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230350

Device Name VISULAS yag

Indications for Use (Describe)

VISULAS yag is intended for use in photodisrupting ocular tissue in the treatment of diseases of the eye, including Posterior capsulotomy, Iridotomy and Posterior Membranectomy.

This device is for Prescription Use (Rx) only.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "ZEISS" in white letters against a blue background. The letters are bold and sans-serif. The background is a solid blue color, and the word "ZEISS" is centered in the image.

In accordance with 21 CFR 807.92 the 510(k) Summary for the VISULAS yag is provided below.

1. SUBMITTER

Applicant:Carl Zeiss Meditec Inc5300 Central ParkwayDublin, CAUSA
Primary CorrespondentTanesha BlandSr. Regulatory Affairs SpecialistCarl Zeiss Meditec, Inc.5300 Central Parkway Dublin, CA 94568(925) 216-7963 PhoneE-mail: tanesha.bland@zeiss.com (preferred)
Secondary CorrespondentLing RenRegulatory Affairs ManagerCarl Zeiss Meditec AGGoeschwitzer Strasse 51-5207745 Jena, Germany

Date Prepared: September 22, 2023

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Zeiss logo. The logo consists of the word "ZEISS" in a bold, sans-serif font. The text is white and is set against a solid blue background. The background is a rectangle with a curved bottom edge.

2. SUBJECT DEVICE

Device Trade Name:VISULAS yag
Classification:21CFR886.4390 laser, Ophthalmic
Regulatory Class:II
Product Code:HQF

3. PREDICATE DEVICE

Predicate Device:Ellex YAG Laser (Ultra Q, Ultra O Reflex) (K212630)
Classification:21CFR886.4390 Ophthalmic Laser
Regulatory Class:ﻟﻤﺴﺎﻋﺪ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮ
Product Code:HOF

4. DEVICE DESCRIPTION

VISULAS yag uses a Q-switched, flashlamp-pumped solid-state laser for photodisruption treatments of diseases of the eye, including posterior capsulotomy, iridotomy and membranectomy. Laser radiation is generated by means of a neodymium-doped yttrium aluminum garnet (Nd:YAG) gain medium inside the laser source. The emitted laser radiation with a near-infrared wavelength of \ = 1064 nm has a pulse duration of < 4 ns (full-width half-maximum; FWHM) and a focal diameter of 6.5 um ± 20%. The maximum energy output per pulse is 9 to 13 mJ in single-burst mode.

INDICATIONS FOR USE .

VISULAS yag is intended for use in photodisrupting ocular tissue in the treatment of diseases of the eye, including Posterior capsulotomy, Iridotomy and Posterior Membranectomy.

This device is for Prescription Use (Rx) only.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "ZEISS" in a bold, sans-serif font. The word is white and is set against a dark blue background. The logo is simple and modern, with a clean design.

SUBSTANTIAL EQUIVALENCE TO PRIMARY PREDICATE 6.

Table 1. Subject to Predicate Device Comparison Table – Indications for Use

DeviceSubject DevicePrimary Predicate Device(K212630)EquivalencyAnalysis
Indicationsfor UseVISULAS yag is intended for use inphotodisrupting ocular tissues in thetreatment of diseases of the eye,including:• Posterior capsulotomy• Iridotomy• Posterior MembranectomyIn the YAG mode (Ultra Q, Ultra QReflex):• Iridotomy and iridectomy• Posterior capsulotomy• Posterior membranectomyThe indication ofVISULAS yag isequivalent to thatof the referencedevice.

Table 2. Subject to Primary Reference Device Comparison Table – Indications for Use

DeviceSubject DevicePrimary Reference Device(K042139)EquivalencyAnalysis
Indicationsfor UseVISULAS yag is intended for use inphotodisrupting ocular tissues in thetreatment of diseases of the eye,including:• Posterior capsulotomy• Iridotomy• Posterior MembranectomyThis device will be used inophthalmic applications, includingposterior capsulotomy and peripheraliridotomy. This device is intended forhealth care workers and may only beused under the supervision of aphysician. This device will not besold to the general public.For posteriorcapsulotomy andiridotomy, theindications for useare equivalent.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Zeiss. The logo consists of the word "ZEISS" in white, bold letters against a solid blue background. The bottom of the blue background is curved, creating a subtle arc shape.

Page 4 of 10

Table 3. Subject to Predicate Device Comparison Table – Technical Characteristics

AttributeSubject DevicePrimary Predicate Device(K212630)Primary Reference Device(K042139)Equivalency Analysis andDiscussion
Device nameVISULAS yagEllex YAG LaserVISULAS YAG III
(Ultra Q, Ultra Q Reflex)
ManufacturerCarl Zeiss Meditec AGEllex Medical Pty LtdCarl Zeiss Meditec AG
510(k)K230350K212630K042139
Classification ProductCodeHQFHQFGEXSubstantially Equivalent.Though the referencedevice was cleared underproduct code 'GEX' inregulation 21 CFR878.4810, current deviceswith these indications foruse and technologicalcharacteristics are classifiedunder product code 'HQF'("Laser, Ophthalmic") inregulation 21 CFR 886.4390("Ophthalmic laser").
Regulation #21CFR886.439021CFR886.439021CFR878.4810Substantially Equivalent.
Ophthalmic LaserOphthalmic LaserLaser surgical instrument foruse in general and plasticsurgery and in dermatology
ApplicationOphthalmic surgeryOphthalmic surgeryOphthalmic surgerySubstantially Equivalent.
Combination DeviceNoNoNoSubstantially Equivalent.
Patient PopulationAdultsNo restrictions given in theoperator manualAdultsSubstantially Equivalent.
AttributeSubject DevicePrimary Predicate Device(K212630)Primary Reference Device(K042139)Equivalency Analysis andDiscussion
User InterfaceGraphical user interface,TouchControl panel, footswitch, laser slit lampRemote control, console,footswitch pedal, laser slitlampGraphical user interface,control, foot switch, laser slitlampSubstantially Equivalent.
Optic PrincipleQ-switched Nd:YAG laserQ-switched Nd:YAG laserQ-switched Nd:YAG laserSubstantially Equivalent.
Wavelength ofTreatment Beam1064 nm1064 nm1064 nmSubstantially Equivalent.
Laser Class inAccordance with IEC60825-143B4Substantially Equivalent.The subject device utilizesstandard IEC 60825-1which contains safetyrequirements that shall befollowed to allow equallysafe usage of the lasers inhumans.
Light HazardClassificationAccording to ISO15004-2Group 2Group 2Group 2Substantially Equivalent.
Mode of Laser BeamSuper GaussianNot specified in the operatormanualSuper GaussianSubstantially Equivalent.
Pulse Length< 4 ns< 4 ns< 4 ns (typically 2 ns to 3 ns)Substantially Equivalent.
Pulse Mode 1 (SinglePulse)9.0 mJ to 13.0 mJ, at max.2.5 Hz0.3 mJ to 10.0 mJ single pulse.Max. 3.0 Hz9.0 mJ to 13.0 mJ, 2.5 Hz(5 shots/ 2s)Substantially Equivalent.
Pulse Mode 2 (DoublePulse)18.0 mJ to 28.0 mJ, at max.1 Hz (burst frequency33 kHz)Not specified. Max. 1.8 Hz18.0 mJ to 28.0 mJ, 1 Hz(1 shots/ 1s) (burst frequency33 kHz)Substantially Equivalent.
AttributeSubject DevicePrimary Predicate Device(K212630)Primary Reference Device(K042139)Equivalency Analysis andDiscussion
Pulse Mode 3 (TriplePulse)29.0 mJ to 45.0 mJ, at max.1 Hz (2 shots/ 2s) (burstfrequency 33 kHz)Not specified. Max. 1.6 Hz29.0 mJ to 45.0 mJ, 0.5 Hz(1 shots/ 2s) (burst frequency 33kHz)Substantially Equivalent.
Energy Attenuation2/4/6/8/10/12/14/16/20/24/28/32/36/40/42/48/56/60/64/70/80/100 %Not specified.2/4/6/8/10/12/14/16/20/24/28/32/36/40/42/48/56/60/64/70/80/100 %Substantially Equivalent.
Applied EnergyResulting from pulse modeand attenuation:0.18 mJ – 13.0 mJ singlepulse in 22 steps.Maximum energy that canbe delivered during clinicaluse is 45 mJ.0.3 mJ to 10.0 mJ single pulse.Continuously variable.Maximum energy that can beselected during clinical use is30 mJ.Average from the last fiveshots, continuously updatedduring the session.Maximum energy output is10 mJ.Substantially Equivalent.
Focus diameter6.5 µm ± 20%8 µm (Full Width HalfMaximum)10 µm in airSubstantially Equivalent.The subject device hasrefined the measuringmethod of the predicatedevice. The focus diameterfor both lasers are identical.
Angle of Exit Aperture(Convergence)14° (round angle)16° (full cone angle)16° (round angle)Substantially Equivalent.The subject device hasrefined the measuringmethod of the predicate
AttributeSubject DevicePrimary Predicate Device(K212630)Primary Reference Device(K042139)Equivalency Analysis andDiscussion
Aiming BeamDiodeWavelength: 660 nm to 680 nmPower: max. 150 µW at thecornea (Class I inaccordance with IEC60825-1)Four-point aiming beamwhich is switchable to 2-point aiming beamLaser diode-pumpedWavelength:Ultra Q: 635 nmUltra Q Reflex: 635 nm or 515 nmClass II in accordance to IEC60825-12-point aiming beamDiodeWavelength: 660 nm to 680 nmPower: max. 150 µW at thecornea (Class I in accordancewith IEC 60825-1)Four-point aiming beamdevice. The convergenceangles are identical.Substantially Equivalent.
Nominal Ocular HazardDistance (NOHD)4 m6.1 m2 mThe NOHD differs due todifferences in the design.This difference does notimpact the safety orperformance of the subjectdevice.
Rated Voltage;Frequency100 V to 240 V (±10 %);50 /60 Hz100 V to 240 VAC; 50/60 Hz100 V to 240 V ±10 %;50 Hz/60 HzSubstantially Equivalent.
Focus Shift betweenAiming Beam andTherapy Beam0/150/225/300 µm anteriorand posterior adjustableTolerance: ±25 µmUltra Q: Posterior only,continuously variable withdetents at 100, 250 and 350 µmUltra Q Reflex: Anterior (A) &posterior (P), continuouslyvariable with detents at 0, 100,200, 300, 400 and 500 µm0/150 µm anterior and posterioradjustableTolerance: ±25 µmSubstantially Equivalent.
IlluminationLED5.6 V, 2 W12 V pre-centered halogenHalogen lamp12 V, 30 WDifferent.
AttributeSubject DevicePrimary Predicate Device(K212630)Primary Reference Device(K042139)Equivalency Analysis andDiscussion
Brightness continuouslyadjustableBrightness continuouslyadjustable
IlluminationOrientationNon-coaxialUltra Q: non-coaxial UltraQ Reflex:coaxialNon-coaxialSubstantially Equivalent.
Illumination Angle0°/±45°/90°180° at horizontal plane(90° for right / left)0°/±45°/90°Substantially Equivalent.
Slit Width0 mm to 14 mm(continuously adjustable)0 - 12 mm0 mm to 14 mm(continuously adjustable)Substantially Equivalent.
Slit Height1/3/5/9/14 mm0.5, 5.0, 8.0 & 12.0 or 0.5, 3.0,8.0 & 12.0 mm1/3/5/9/14 mmSubstantially Equivalent.
Magnification8x/12x/20xUltra Q:Standard: 10×, 16×, 28×Accessory: 6×, 10×, 16×, 28×,45×Ultra Q ReflexStandard: 10×, 16×, 25×Accessory: 6×, 10×, 16×, 28×,45×5x/8x/12x/20x/32xSubstantially Equivalent.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for ZEISS. The logo consists of the word "ZEISS" in white, bold, sans-serif font. The word is placed on a solid blue square. A white curved line is at the bottom of the blue square.

Page 5 of 10

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for ZEISS. The logo consists of the word "ZEISS" in white, bold, sans-serif font. The word is placed on a solid blue square. A white curved line is at the bottom of the blue square.

Page 6 of 10

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Zeiss, a German manufacturer of optical systems and optoelectronics. The logo consists of the word "ZEISS" in a bold, sans-serif font. The text is white and is set against a solid blue background. The blue background is in the shape of a rounded rectangle.

Page 7 of 10

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the word "ZEISS" in a bold, sans-serif font. The word is white and is set against a solid blue background. The background is a square shape with a curved bottom edge.

Page 8 of 10

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "ZEISS" in a bold, sans-serif font. The word is white and is set against a solid blue background. The letters are large and fill most of the frame, with the bottom of the "Z" slightly cut off by a curved white shape.

7. SUMMARY OF STUDIES

Non-Clinical Performance Testing

Non-clinical system testing provided an evaluation of the performance of the system relevant to each of the system specifications. The functional and system level testing showed that the system met the defined specifications. ZEISS demonstrated non-clinical equivalency between the subject device, VISULAS yag, and the predicate device, VISULAS YAG III.

VISULAS yag is not intended for sterilization and no shelf life is specified for VISULAS yag. Therefore, no testing was conducted on VISULAS yag concerning sterilization and shelf life.

Biocompatibility testing was performed on the appropriate components of the subject device. The testing performed aligns with current recognized standards and meets or exceeds testing performed on VISULAS YAG III. Biocompatibility testing demonstrated equivalency between VISULAS yag III.

Software verification and validation testing were conducted, and documentation was provided as recommended in accordance with IEC 62304. There have been no changes in level of concern or software architecture. All testing passed.

EMC and electrical safety are used in comparable standards to the predicate device and the relevant product standards given by the product indication and performance. Electrical safety and EMC testing were performed based on the ILAC-Scheme to include all national deviations for VISULAS yag. All testing passed.

No additional safety or performance concerns have been raised during development or device testing.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "ZEISS" in a bold, sans-serif font. The word is white and is set against a dark blue background. The letters are large and fill most of the frame, and the bottom of the letters are curved.

8. CONCLUSION

The indications for use are identical to the indications for use of the predicate device; and therefore, are deemed to be substantially equivalent.

The technological characteristics and risk profile of the subject device are equivalent to the predicate device; and therefore, are deemed to be substantially equivalent.

Testing methods are equivalent to those of the predicate device; and therefore, are deemed to be substantially equivalent.

Therefore, the subject device meets the requirements for substantial equivalence as compared to the proposed predicate device.

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.