K Number
K230286

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-05-17

(104 days)

Product Code
Regulation Number
870.2800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Assert-IQ™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms that may be cardiac-related such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias such as bradycardia, and sinus pauses. The Assert-IQ ICM is also indicated for patients who have been previously diagnosed with atrial fibrillation (AF) or who are susceptible to developing AF. The Assert-IQ ICM is intended to be inserted subcutaneously in the left pectoral region, also described as the left anterior chest wall. The Assert-IQ ICM has not been specifically tested for pediatric use.

Device Description

The Assert-IQ™ ICM system consists of Insertable Cardiac Monitor device models DM5000. DM5300, and DM5500 and is intended to help physicians monitor, diagnose, and document the rhythm in patients who are susceptible to cardiac arrhythmias and unexplained symptoms, as indicated. Specific features include: Patient-initiated triggering of EGM storage using the myMerlin™ mobile application, Automated triggering of EGM storage when tachycardia, bradycardia, or pauses are detected, Automated triggering of EGM storage when atrial fibrillation (AF) is detected, The ability to inhibit EGM storage due to noise and allow for detection and storage of AF and non-AF (pause, bradycardia, and tachycardia) arrhythmias after noise exit, Collection and display of diagnostic trends, including AF burden, PVC burden, and activity trends, Remote monitoring capability, Remote programming capability. The system includes the ICM device, Delivery/Implant Tools, myMerlin™ mobile application, and Merlin.net™ Software.

AI/ML Overview

The provided text is a 510(k) summary for the Abbott Assert-IQ™ Insertable Cardiac Monitor (ICM), seeking substantial equivalence to a predicate device, the Jot Dx™ ICM. The document primarily focuses on demonstrating that the Assert-IQ™ ICM is substantially equivalent to its predicate, rather than providing detailed acceptance criteria and a standalone study proving the device meets those criteria.

Therefore, much of the requested information regarding specific acceptance criteria, detailed device performance metrics, sample sizes for test and training sets, expert qualifications, and adjudication methods for a de novo study is not available in this document. This document acts as a comparison to a previously cleared device.

However, based on the information provided, here's what can be extracted and inferred:

1. Table of Acceptance Criteria and Reported Device Performance

This document does not provide a table of explicit, quantitative acceptance criteria for this device (Assert-IQ™ ICM) and corresponding performance metrics from a dedicated clinical study. Instead, it states that the Assert-IQ™ ICM "leverages the existing market cleared predicate (Jot Dx ICM DM4500 (K212206)) algorithms and functionality, for which the clinical testing is not repeated." This implies that the clinical acceptance criteria and performance were established with the predicate device and are considered met by the Assert-IQ™ ICM due to substantial equivalence.

The document discusses "design verification and system validation activities" and "all necessary device and system verification testing," but these are engineering and functional tests, not clinical performance studies with specific statistical acceptance criteria for diagnostic accuracy.

Key takeaway: The acceptance criteria for clinical performance are implicitly assumed to be met through substantial equivalence to the predicate device, K212206, which already underwent clinical testing for its algorithms.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the document for the Assert-IQ™ ICM. The document explicitly states that "the clinical testing is not repeated" because the Assert-IQ™ ICM leverages the predicate's algorithms and functionality. Therefore, any clinical "test set" and its provenance would pertain to the predicate device, K212206, and details about that are not included here.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not provided for the Assert-IQ™ ICM, as clinical testing was not repeated. This detail would have been part of the predicate device's original clearance.

4. Adjudication Method for the Test Set

This information is not provided for the Assert-IQ™ ICM, as clinical testing was not repeated. This detail would have been part of the predicate device's original clearance.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

This information is not provided for the Assert-IQ™ ICM. MRMC studies typically assess human reader performance with and without AI assistance, which is not the focus of this substantial equivalence submission that explicitly states clinical testing was not repeated.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

This document does not describe a standalone study for the Assert-IQ™ ICM's algorithms. It states that the device "leverages the existing market cleared predicate... algorithms and functionality," implying that any standalone performance validation would have been done for the predicate device, K212206. The current submission focuses on demonstrating that the technological changes (e.g., longevity, size, programming features) do not alter the fundamental functionality or safety/effectiveness of the existing algorithms.

7. The Type of Ground Truth Used

This information is not provided for the Assert-IQ™ ICM, as clinical testing was not repeated. For the predicate device, ground truth for arrhythmias would typically involve electrophysiologist over-read of EGM data, potentially correlated with patient symptoms or other diagnostic tests, but these specifics are not in this document.

8. The Sample Size for the Training Set

This information is not provided for the Assert-IQ™ ICM. Training set details would pertain to the development of the algorithms, which were part of the predicate device.

9. How the Ground Truth for the Training Set Was Established

This information is not provided for the Assert-IQ™ ICM. Ground truth establishment for the training set would pertain to the development of the algorithms used in the predicate device.


Summary of what is provided regarding acceptance criteria indirectly:

The document states that the "Assert-IQ™ ICM leverages the existing market cleared predicate (Jot Dx ICM DM4500 (K212206)) algorithms and functionality, for which the clinical testing is not repeated."

The report concludes that the "modifications to the subject device do not alter the previously evaluated clinical acceptance criteria and were verified through design verification and system validation activities."

The document lists various engineering and safety tests that were performed to support substantial equivalence:

  • Software/Firmware Verification and System Verification
  • Cybersecurity
  • Bluetooth / coexistence
  • Preclinical Testing (Animal GLP Study)
  • Biocompatibility
  • Sterilization
  • Shelf Life
  • Packaging
  • Electromagnetic compatibility (EMC)
  • Electrical safety
  • MRI safety
  • Mechanical performance
  • System Validation
  • Usability testing

It states that "All verification and validation activities were successfully completed and did not raise new safety or performance issues." This indicates that for these listed engineering and safety aspects, the device met its internal design and regulatory requirements, which serve as a form of acceptance criteria for those specific aspects. However, quantitative acceptance criteria and performance metrics for these individual tests are not detailed in this summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

May 17, 2023

Abbott Medical Laura Sparks Project Manager, Regulatory Affairs 15900 Valley View Ct. Sylmar, California 91342

Re: K230286

Trade/Device Name: Assert-IQ™ Insertable Cardiac Monitor Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MXC. DSI Dated: April 14, 2023 Received: April 17, 2023

Dear Laura Sparks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed
Sara M. by Sara M. Royce
-S
Royce -S Date: 2023.05.17
10:43:48 -04'00'

for

Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230286

Device Name Assert-IOTM Insertable Cardiac Monitor

Indications for Use (Describe)

The Assert-IO™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms that may be cardiac-related such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias such as bradycardia, and sinus pauses. The Assert-IQ ICM is also indicated for patients who have been previously diagnosed with atrial fibrillation (AF) or who are susceptible to developing AF. The Assert-IQ ICM is intended to be inserted subcutaneously in the left pectoral region, also described as the left anterior chest wall. The Assert-IQ ICM has not been specifically tested for pediatric use.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Date Prepared:February 1, 2023
Submitter:Abbott Medical
Address:15900 Valley View Ct.Sylmar, CA 91342USA
Phone:(818) 362-6822
Establishment Registration:2017865
Contact Person:Laura SparksProject Manager, Regulatory(818) 493-2734laura.sparks@abbott.comJennifer DunhamAssoc. Director, Regulatory(818) 383-1630jennifer.dunham@abbott.com
Trade Name/ProprietaryAssert-IQTM Insertable Cardiac Monitor
Common Name:Insertable Cardiac Monitor
Model Numbers:DM5000, DM5300, DM5500
Classification Name:21 CFR 870.2800, Medical magnetic tape recorder
Product Code:MXC, DSI
Classification:Class II
Pediatric Use:The Assert-IQTM Insertable Cardiac Monitor has not been specificallytested for pediatric use.

LEGALLY MARKETED DEVICE TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED

510(k) K212206 Jot Dx™ Insertable Cardiac Monitor

INDICATIONS FOR USE

The Indications for Use for the Assert-IO™ Insertable Cardiac Monitor (ICM) is as follows:

The Assert-IQ TM ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms that may be cardiac-related such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias such as bradycardia, and sinus pauses. The Assert-IQ ICM is also indicated for patients who have been previously diagnosed with atrial fibrillation (AF) or who are susceptible to developing AF. The Assert-IQ ICM is intended to be inserted subcutaneously in the left pectoral region, also described as the left anterior chest wall. The Assert-IQ ICM has not been specifically tested for pediatric use.

{4}------------------------------------------------

PRODUCT DESCRIPTION

The Assert-IO™ ICM system consists of Insertable Cardiac Monitor device models DM5000. DM5300, and DM5500 and is intended to help physicians monitor, diagnose, and document the rhythm in patients who are susceptible to cardiac arrhythmias and unexplained symptoms, as indicated.

Specific features include:

  • . Patient-initiated triggering of EGM storage using the myMerlin™ mobile application. This includes capability for the patient to identify symptoms, which are stored with the EGM for physician review.
  • Automated triggering of EGM storage when tachycardia, bradycardia, or pauses are detected; with physician-programmable values for pause duration, bradycardia rate, tachycardia rate, and number of tachycardia intervals.
  • 트 Automated triggering of EGM storage when atrial fibrillation (AF) is detected, with physician programmable values for AF duration.
  • 트 The ability to inhibit EGM storage due to noise and allow for detection and storage of AF and non-AF (pause, bradycardia, and tachycardia) arrhythmias after noise exit.
  • I Collection and display of diagnostic trends, including AF burden, PVC burden, and activity trends.
  • Remote monitoring capability
  • 트 Remote programming capability

The Assert-IQ ICM system includes the ICM device and the following key components, which are within the scope of this premarket submission.

  • . Delivery/Implant Tools (Incision Tool model DM5320 and Insertion Tool model DM5310 or DM5510) - used to implant the ICM device subcutaneously. The implantable device is pre-loaded into the associated insertion tool and packaged together with the incision tool.
  • . myMerlin™ mobile application (model APP1000 (Android) or APP1001 (iOS)) provides the means for the patient to activate EGM recording in the Assert-IQ™ ICM device, with data pass-through functionality to enable physician follow-up and remote programming via the Merlin.net Patient Care Network.
  • Merlin.net™ Software, model MN7000 Virtual Device Engine (VDE), which includes . services for user parameter management and updates to support remote programming of Assert-IQ™ ICM devices.

{5}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS

The fundamental technology of the Assert-IQ™ ICM, relative to the predicate Jot Dx™ ICM (K212206), is not changing.

An overview of some design/technological characteristics of the predicate Jot Dx ICM DM4500 (K212206) and the subject Assert-IQ ICM devices follows (refer to Section 11 for details).

  • Improved longevity-subject Assert-IQ ICM model DM5500 has a 6-year longevity, and ● subject Assert-IQ ICM models DM5000 and DM5300 have a 3-year longevity. The predicate device model DM4500 has a 2-year longevity.
  • The subject Assert-IQ ICM models DM5300 and DM5500 support remote programming, while the predicate device model DM4500 and subject Assert-IQ ICM model DM5000 do not.
  • The subject and predicate devices support remote monitoring. ●
  • 1.5T and 3T MR Conditional labeling (subject and predicate)
  • Assert-IO ICM subject device models DM5000 and DM5300 are slightly smaller in dimensions, volume, weight, and in electrode surface area than predicate device model DM4500, while Assert-IQ ICM subject device model DM5500 is slightly larger in dimensions, volume, weight, and in electrode surface area than predicate device model DM4500.
  • A suture hole is not present in the subject Assert-IQ ICM devices (present for optional ● use in predicate device model DM4500).
  • The header electrode is dual-sided (front and back) on the subject Assert-IQ ICM devices, while the predicate model DM4500 electrode is single-sided.
  • The can, electrode material, coating, and battery chemistry are the same on the subject Assert-IQ ICM and predicate model DM4500 devices, while the header material differs.
  • Sterilization and packaging are the same for subject and predicate devices. ●
  • Device telemetry using Bluetooth Low Energy is the same for subject and predicate devices.
  • The subject and predicate devices support automatic arrhythmia detection and manual (patient-activated) recording.
  • The subject Assert-IQ ICM devices include 3D accelerometers. Assert-IO ICM models DM5300 and DM5500 support 3D+HS and 1D modes of operation and support posture data at episode onset. Assert-IQ ICM model DM5000 supports 1D mode of operation without support for posture data at episode onset. The predicate model DM4500 device includes a 1D accelerometer without support for posture data at episode onset.
  • Additional diagnostic trends (PVC burden and activity) in the subject Assert-IQ ICM device models DM5300 and DM5500. All subject Assert-IQ ICM device models and the predicate DM4500 devices support AF burden diagnostic trends.

SUBSTANTIAL EQUIVALENCE

The subject Assert-IQ™ ICM system is substantially equivalent to the predicate Jot DxTM ICM system (K212206).

{6}------------------------------------------------

The underlying indications for use are not impacted by the changes within the scope of this premarket submission, though minor updates have been made to the language of the indications for use of the Assert-IQ™ ICM system as compared to the predicate Jot Dx™ ICM system (K212206) for clarity and for harmonization with global labeling.

Predicate Device Jot Dx™ ICM (K212206)Subject Device Assert-IQ™ ICM
Indications for UseIndications for Use
The Jot Dx™ ICM is indicated for the monitoring anddiagnostic evaluation of patients who experienceunexplained symptoms such as the following:dizziness, palpitations, chest pain, syncope, andshortness of breath, as well as patients who are at riskfor cardiac arrhythmias. It is also indicated for patientswho have been previously diagnosed with atrialfibrillation or who are susceptible to developing atrialfibrillation. The Jot Dx ICM has not been specificallytested for pediatric use.The Assert-IQ™ ICM is indicated for the monitoringand diagnostic evaluation of patients who experienceunexplained symptoms that may be cardiac-related suchas: dizziness, palpitations, chest pain, syncope, andshortness of breath, as well as patients who are at riskfor cardiac arrhythmias such as bradycardia,tachycardia, and sinus pauses. The Assert-IQ ICM isalso indicated for patients who have been previouslydiagnosed with atrial fibrillation (AF) or who aresusceptible to developing AF. The Assert-IQ ICM isintended to be inserted subcutaneously in the leftpectoral region, also described as the left anterior chestwall. The Assert-IQ ICM has not been specificallytested for pediatric use.

Both the subject and the predicate device (K212206) systems have the same fundamental functionality and technological characteristics. The changes within scope of this premarket submission do not raise new issues of safety and effectiveness. The modifications to the subject device do not alter the previously evaluated clinical acceptance criteria and were verified through design verification and system validation activities.

TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION

All necessary device and system verification testing was conducted on the subject Assert-IQ™ ICM to support a determination of substantial equivalence to the predicate device, including:

  • Software/Firmware Verification and System Verification
  • Cybersecurity ●
  • Bluetooth / coexistence
  • Preclinical Testing (Animal GLP Study)
  • Biocompatibility
  • Sterilization
  • Shelf Life ●
  • Packaging
  • Electromagnetic compatibility (EMC) ●
  • Electrical safety ●
  • MRI safety ●

{7}------------------------------------------------

  • Mechanical performance ●
  • System Validation
  • Usability testing

The Assert-IQ™ ICM leverages the existing market cleared predicate (Jot Dx ICM DM4500 (K212206)) algorithms and functionality, for which the clinical testing is not repeated.

All verification and validation activities were successfully completed and did not raise new safety or performance issues; the Assert-IQ™ ICM system has been shown to be appropriate and safe for its intended use and is substantially equivalent to the predicate Jot Dx™ ICM system (K212206).

CONCLUSION

The Assert-IQ™ ICM is an implantable cardiovascular monitoring device, for which the subject device shares its fundamental design and mechanism of action, as well as the underlying indications for use, with the identified predicate Jot Dx™ ICM (K212206). The results of the verification and validation tests have demonstrated the subject Assert-IQ™ ICM in accordance with product specifications. The fundamental technology of the Assert-IQ™ ICM, relative to the predicate Jot DxTM ICM (K212206), remains unchanged. The subject Assert-IQ™ ICM is substantially equivalent to the predicate Jot Dx™ ICM (K212206) in terms of technology, intended use, and performance.

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).