K Number
K230264

Validate with FDA (Live)

Device Name
Ezra Flash
Manufacturer
Date Cleared
2023-04-13

(72 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Ezra Flash is an image processing software used for image enhancement of MR images. It can be used to reduce image noise in images acquired as part of non-contrast MRI exams on 3-Tesla Siemens for Sagittal T1, Axial T2 and Axial Flair sequences within the head region for patients > 18 years of age.

Device Description

Ezra Flash is a Software as a Medical Device (SaMD) consisting of a software algorithm that enhances images of the head region taken by MRI scanners. As it only processes images for the end user, the device has no interface. It is intended to be used by radiologists in an imaging center, clinic, or hospital. The software can be used with MR images acquired as part of MRI exams on 3-Tesla Siemens and GE scanners. The outputs are images with enhanced image quality. Both the original non-enhanced studies and the Ezra Flash-enhanced studies are available to the end user. Ezra Flash receives DICOM-compliant non-contrast MR image inputs acquired on 3-Tesla scanners for Sagittal T1, Axial T2 and Axial Flair sequences within the head region. The software uses a convolutional neural network-based algorithm to improve image quality by reducing noise. The device outputs a DICOM-compliant copy of the images with improved image quality.

AI/ML Overview

Ezra Flash Acceptance Criteria and Performance Study

The Ezra Flash device for image enhancement of MR images (K230264) underwent performance testing to demonstrate its safety and effectiveness. The study focused on quantitative metrics (SNR, CNR) and a subjective image quality assessment by experts (Perceived Noise).

1. Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance CriteriaReported Device Performance
Signal-to-Noise Ratio (SNR)SNR of a selected region of interest (ROI) in each test dataset is on average improved by > 5% after Ezra Flash enhancement compared to original MR-acquired images (raw).The test results demonstrated that the Ezra Flash performs to its intended use and is deemed acceptable for clinical use. (Specific numerical improvement not provided, but implies criterion was met).
Contrast-to-Noise Ratio (CNR)CNR of a selected region of interest (ROI) in each test dataset is on average improved by > 0% after Ezra Flash enhancement compared to the MR-acquired raw images.The test results demonstrated that the Ezra Flash performs to its intended use and is deemed acceptable for clinical use. (Specific numerical improvement not provided, but implies criterion was met).
Image Quality Perceived NoiseThe mean Likert results for the Ezra Flash-enhanced images compared to the original MR-acquired images (raw) is greater than or equal to 0.5 Likert scale points.The test results demonstrated that the Ezra Flash performs to its intended use and is deemed acceptable for clinical use. (Specific numerical improvement not provided, but implies criterion was met).

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size for the test set or the data provenance (e.g., country of origin, retrospective or prospective). It mentions "each test dataset" for SNR and CNR, suggesting multiple datasets were used.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document mentions "Likert results for the Ezra Flash-enhanced images compared to the original MR-acquired images (raw)," which implies human expert assessment of image quality, likely for perceived noise. However, it does not specify:

  • The number of experts used.
  • The qualifications of those experts (e.g., specific years of experience, board certification).

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the expert ratings on the test set. It only refers to "mean Likert results."

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The provided text does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. The performance testing description focuses on objective image metrics (SNR, CNR) and a subjective image quality assessment, but not on human reader performance with and without AI assistance. Therefore, no effect size for human reader improvement is provided.

6. Standalone (Algorithm Only) Performance Study

Yes, a standalone performance study was done for the Ezra Flash algorithm. The described testing focuses on the algorithm's direct impact on image quality metrics (SNR, CNR, perceived noise) by processing raw MR images and comparing them to the enhanced outputs. The device "only processes images for the end user" and "has no interface," with outputs viewed on existing PACS workstations. This indicates evaluation of the algorithm's performance in isolation from a human-in-the-loop workflow.

7. The Type of Ground Truth Used

The ground truth for the quantitative metrics (SNR, CNR) is inherently derived from the raw MR-acquired images themselves, serving as the baseline for improvement. For the "Image Quality Perceived Noise" metric, the ground truth is based on expert consensus (or at least expert scores) derived from Likert scale ratings comparing enhanced images to raw images. This is a form of subjective expert evaluation.

8. The Sample Size for the Training Set

The document does not specify the sample size for the training set. It mentions that the device uses a "convolutional neural network-based algorithm" whose "parameters of the filters were obtained through an image-guided optimization process," implying a training phase. However, the specific number of images or cases used for training is not disclosed.

9. How the Ground Truth for the Training Set was Established

The document states that the "parameters of the filters were obtained through an image-guided optimization process." While this indicates that the model was trained, it does not explicitly describe how the ground truth for the training set was established. Typically, for image enhancement, training ground truth might involve:

  • Paired low-quality/high-quality images.
  • Synthetic noise addition to high-quality images.
  • Expert-annotated "ideal" images or metrics.

However, the provided text does not elaborate on this aspect.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 13, 2023

Ezra AI, Inc. % David Girard VP of Operations, Quality and Regulatory 419 Park Ave S, Suite 600 New York, NY 10016

Re: K230264

Trade/Device Name: Ezra Flash Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: April 6, 2023 Received: April 6, 2023

Dear David Girard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230264

Device Name Ezra Flash

Indications for Use (Describe)

Ezra Flash is an image processing software used for image enhancement of MR images. It can be used to reduce image noise in images acquired as part of non-contrast MRI exams on 3-Tesla Siemens for Sagittal T1, Axial T2 and Axial Flair sequences within the head region for patients > 18 years of age.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary 1

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.

Submitter Information 1.1

Submitter's NameEzra Al, Inc.
Address419 Park Ave S, Suite 600, New York, NY 10016 USA
Telephone+1 (646) 402-5751
Contact PersonDavid Girard
Date of Summary PreparationJanuary 31st, 2023

Subject Device Information 1.2

Device NameEzra Flash
Model NumberFLHAI00PL00IF00IT00LB00
Common NameEzra Flash
ClassificationII
Review PanelRadiology
Product CodeLLZ
Regulations21 CFR 892.2050- System, Image Processing, Radiological

Predicate Device Information 1.3

Device Name SubtleMR 510(k) Number K203182

{4}------------------------------------------------

Device Description 2

Ezra Flash is a Software as a Medical Device (SaMD) consisting of a software algorithm that enhances images of the head region taken by MRI scanners. As it only processes images for the end user, the device has no interface. It is intended to be used by radiologists in an imaging center, clinic, or hospital. The software can be used with MR images acquired as part of MRI exams on 3-Tesla Siemens and GE scanners.

The outputs are images with enhanced image quality. Both the original non-enhanced studies and the Ezra Flash-enhanced studies are available to the end user.

Ezra Flash receives DICOM-compliant non-contrast MR image inputs acquired on 3-Tesla scanners for Sagittal T1, Axial T2 and Axial Flair sequences within the head region. The software uses a convolutional neural network-based algorithm to improve image quality by reducing noise. The device outputs a DICOM-compliant copy of the images with improved image quality.

3 Indications for Use

Ezra Flash is an image processing software used for image enhancement of MR images. It can be used to reduce image noise in images acquired as part of non-contrast MRI exams on 3-Tesla Siemens and GE scanners for Sagittal T1, Axial T2 and Axial Flair sequences within the head region for patients > 18 years of age.

{5}------------------------------------------------

Summary of Technological 4 Characteristics Comparison

Table 1 shows that Ezra Flash and the Predicate Device (K203182) are equivalent in technological characteristics.

FeatureSubject DevicePredicate Device (K203182)
Regulation NumberCFR 892.2050Same
Product CodeLLZSame
Indications for UseEzra Flash is an image processing soft-ware used for image enhancement ofMR images. It can be used to reduceimage noise for non-contrast 3T headMRI scans.SubtleMR is an image processing soft-ware that can be used for image en-hancement in MRI images. It canbe used to reduce image noise forhead, spine, neck, abdomen, pelvis,prostate, breast and musculoskeletalMRI, or increase image sharpness forhead MRI.
Physical Character-isticsSoftware package that operates on off-the-shelf hardwareSame
ComputerLinux CompatibleSame
DICOMStandardComplianceThe software processes DICOM com-pliant image dataSame
Operating SystemLinuxSame
ModalitiesMRISame
User InterfaceNone - enhanced images are viewedon existing PACS workstationsSame

Table 1: Subject and Predicate Device Comparison.

{6}------------------------------------------------

Image EnhancementAlgorithm DescriptionEzra Flash software implements an im-age enhancement algorithm using aconvolutional neural network-based fil-tering. Original images are enhancedby running through a cascade of filterbanks, where thresholding and scal-ing operations are applied. These fil-ters result in a single machine learn-ing model that reduces noise. The pa-rameters of the filters were obtainedthrough an image-guided optimizationprocess.SubtleMR software implements an im-age enhancement algorithm using con-volutional neural network based filter-ing. Original images are enhancedby running through a cascade of filterbanks, where thresholding and scalingoperations are applied. Separate neu-ral network based filters are obtainedfor noise reduction and sharpness in-crease. The parameters of the fil-ters were obtained through an image-guided optimization process.
WorkflowThe software operates on DICOM fileson the file system, enhances the im-ages, and stores the enhanced imageson the file system. The receipt of orig-inal DICOM image files and delivery ofenhanced images as DICOM files de-pends on other software systems. En-hanced images co-exist with the origi-nal images.Same

{7}------------------------------------------------

5 Performance Testing

The Ezra Flash has been developed in a manner consistent with accepted standards for software development and evaluated in accordance with design specifications and applicable performance standards through software verification, validation, and usability testing.

Ezra performed the following main performance testing for Ezra Flash:

    1. Signal-to-Noise Ratio (SNR)
    • SNR of a selected region of interest (ROI) in each test dataset is on average improved by > 5% after Ezra Flash enhancement compared to original MR-acquired images (raw).
    1. Contrast-to-Noise Ratio (CNR)
    • CNR of a selected region of interest (ROI) in each test dataset is on average improved by > 0% after Ezra Flash enhancement compared to the MR-acquired raw images.
    1. Image Quality Perceived Noise
    • The mean Likert results for the Ezra Flash-enhanced images compared to the original MR-acquired images (raw) is greater than or equal to 0.5 Likert scale points.

The test results demonstrated that the Ezra Flash performs to its intended use, is deemed acceptable for clinical use, and does not introduce new questions of safety or efficacy. The testing was conducted in accordance to the software validation/verification plans and protocols. A full description of the software functionality, device hazard analysis, software requirements, verification, validation and usability study is provided in this submission.

{8}------------------------------------------------

Conformance Standards б

There are no applicable FDA mandated performance standards for this device. However, voluntary standards have been utilized in the production of the software. The device was designed and developed in accordance to the following conformance standards:

  • · ISO 14971:2019 Medical Devices Application of risk management to medical devices
  • · IEC 62304 Edition 1.1: 2015 Medical device software Software life cycle processes
  • NEMA PS 3.1-3.20 (2021e) Digital Imaging and Communications in Medicine (DICOM) set

Substantial Equivalence Conclusion 7

The Ezra Flash has the same intended use and similar technological characteristics as the predicate SubtleMR (K203182) device. As demonstrated in this submission, the subject and predicate device are identical in indications for use and intended use for the reduction of image noise in 3T head MRI scans. The subject Ezra Flash is substantially equivalent to the predicate SubtleMR (K203182) device, and the minor differences in the technological characteristics of the subject and predicate device do not raise any new or different questions of safety and effectiveness.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).