K Number
K230259

Validate with FDA (Live)

Date Cleared
2023-05-25

(114 days)

Product Code
Regulation Number
872.6660
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Soreal Press is indicated for single-unit anterior or posterior prostheses and for three-unit prostheses, such as veneers, inlay/onlay, partial crowns, anterior crowns, posterior crowns, using the hot press technique. Soreal CAD is indicated for single-unit anterior or posterior prostheses and for three-unit prostheses, such as veneers, inlay/onlay, partial crowns, anterior crowns, posterior crowns, using the CAD/CAM system.

Device Description

Our products are divided into Soreal Press and Soreal CAD. Soreal Press is put into the porcelain casting furnace and form it by hot pressing at 915-930 °C, and treatment by sandblasting and glazing. Soreal CAD is cut by grinding equipment with wet processing, and put into the porcelain casting furnace at 840-850°C for 30 min to secondary crystallization treatment. Then dye it. They are all composed of SiO2, Li2O, K2O, P2O5, Al2O3 and other oxides. They have a variety of colors, and users can choose the right color to match the patient's natural teeth. In general, our products are also called Dental glass ceramic.

AI/ML Overview

The provided document is a 510(k) Summary for a dental ceramic device (Soreal Press/Soreal CAD) and focuses on demonstrating substantial equivalence to a predicate device, rather than providing a performance study of an AI/ML powered device. Therefore, many of the requested criteria related to AI/ML device performance (like multi-reader multi-case studies, effect size of AI assistance, expert ground truth establishment for AI training/testing) are not applicable or findable in this document.

However, I can extract information related to the acceptance criteria for a non-AI/ML medical device and the study that proves it meets these criteria, as detailed in the document.

Here's a breakdown of the available information:

Acceptance Criteria and Reported Device Performance

The acceptance criteria for this dental ceramic device are based on the requirements of ISO 6872:2015 "Dentistry – Ceramic materials" and ISO 10993 "Biocompatibility." The document provides a table summarizing these, which I have reproduced below.

Table of Acceptance Criteria and Reported Device Performance

Test ItemRequirements (Acceptance Criteria)Reported Device Performance (Measured Data)Conclusion
UniformityThe colorants are dispersed uniformly throughout the dental ceramic material and in powdered ceramic products.The colorants are dispersed uniformly throughout the dental ceramic material and in powdered ceramic products.Pass
Freedom of extraneous materialsDevice shall be free from extraneous materials.Device shall be free from extraneous materials.Pass
Linear thermal expansion coefficient(10±1.5)×10⁻⁶K⁻¹11.1×10⁻⁶K⁻¹Pass
Glass transition temperature550±50 °CSoreal CAD: 518.4°C Soreal Press: 515.8°CPass
Chemical solubility<100µg/cm²Soreal CAD Range: 19.37 - 23.45 µg/cm² Soreal Press Range: 19.32 - 24.12 µg/cm² (average 21.28 µg/cm²)Pass
Radioactivity of dental ceramic< 1.0Bq·g⁻¹ of ²³⁸U<0.0073 Bq·g⁻¹ of ²³⁸UPass
Flexural Strength>300MPaSoreal CAD Range: 365 - 423 MPa Soreal Press Range: 379 - 436 MPaPass
In Vitro Cytotoxicity TestNo Cytotoxicity effectNo Cytotoxicity effectPass
Irritation Oral Mucosa IrritationNot a primary oral mucosa irritant under the conditions of the studyNot a primary oral mucosa irritant under the conditions of the studyPass
Skin SensitizationNot a sensitizer under the conditions of the studyNot a sensitizer under the conditions of the studyPass
Subacute and Subchronic ToxicityNo subacute and subchronic toxic effects observedNo subacute and subchronic toxic effects observedPass
GenotoxicityNo genotoxic effects observedNo genotoxic effects observedPass
Acute Systemic ToxicityNo acute Systemic ToxicityNo acute Systemic ToxicityPass
PyrogenMeets the requirements for the absence of pyrogensMeets the requirements for the absence of pyrogensPass

Study Details (Based on the Provided Document)

  1. Sample size used for the test set and the data provenance:

    • The document does not explicitly state the sample sizes for each specific test (e.g., how many samples were used for flexural strength, chemical solubility, etc.). It only provides the measured data points obtained from these tests.
    • Data Provenance: The tests were "Non-clinical data" performed according to ISO standards. The origin (country) of the test data is not specified, but the applicant company is Shenzhen UP3d Technology Co., Ltd. in China. The data is retrospective, as it was generated to support a 510(k) submission.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as the device is a dental ceramic, not an AI/ML-powered diagnostic device. "Ground truth" in this context refers to physical and chemical properties measured in a lab, not expert interpretations of medical images or data.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable for physical and chemical property tests of a material. The tests follow standardized protocols (ISO standards).
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI-powered device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI-powered device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device is the physical, chemical, and biological properties measured according to international standards (ISO 6872:2015 for dental ceramic materials and ISO 10993 for biocompatibility). This involves laboratory measurements and assessments, not clinical "ground truth" derived from patient outcomes or expert consensus on medical images.
  7. The sample size for the training set:

    • Not applicable in the context of an AI/ML training set, as this is a non-AI/ML medical device. For material development, internal R&D would involve extensive material science work, but a specific "training set" in the AI sense is not relevant here.
  8. How the ground truth for the training set was established:

    • Not applicable for a non-AI/ML medical device. For material development, the "ground truth" of material properties is established through adherence to material science principles, experimental methodology, and international standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 25, 2023

Shenzhen UP3d Technology Co., Ltd. Judy Zhang Head of Firm 401, Block B, Nanshan Yungu Nanfeng Tower 4093 Liuxian Avenue, Nanshan District Shenzhen, Guangdong 518055 CHINA

Re: K230259

Trade/Device Name: Soreal Press, Soreal CAD Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: April 13, 2023 Received: April 25, 2023

Dear Judy Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael E. Adjodha -S

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K230259

Device Name Soreal CAD Soreal Press

Indications for Use (Describe)

Soreal Press is indicated for single-unit anterior or posterior prostheses and for three-unit prostheses, such as veneers, inlay/onlay, partial crowns, anterior crowns, posterior crowns, using the hot press technique. Soreal CAD is indicated for single-unit anterior or posterior prostheses and for three-unit prostheses, such as veneers,

inlay/onlay, partial crowns, anterior crowns, posterior crowns, using the CAD/CAM system.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo with an orange geometric shape on the left and the text "云甲科技" on the right. The geometric shape appears to be a stylized representation of a cube or a series of interconnected boxes. The text is in Chinese characters, and the overall design is simple and modern.

510(K) Summary - K230259

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date:

1. Submitter's Information

The submitter of this pre-market notification is:

Name:Shenzhen UP3d Technology Co., Ltd.
Address:401, Block B, Nanshan Yungu Nanfeng Tower4093 Liuxian Avenue, Nanshan District, Shenzhen, China
Contact person:Judy Zhang
Title:Head of Firm
E-mail:zhangwanlin@up3d.cn
Tel:086-13713634397

2. Device Identification

Device/Trade Name:Soreal Press/Soreal CAD
Regulation Number:872.6660
Regulation Class:Class 2
Product Code:EIH
Common Name:Powder, Porcelain

3. Predicate Device

510(K) number:K192231
Device Name:Dental Glass Ceramics Blocks
Manufacturer:Aidite (Qinhuangdao) Technology Co., Ltd.
Regulation Number:872.6660
Regulation Class:Class 2
Product Code:EIH
Common Name:Powder, Porcelain

4. Device Description

Our products are divided into Soreal Press and Soreal CAD.

Soreal Press is put into the porcelain casting furnace and form it by hot pressing at 915-930 °C, and

treatment by sandblasting and glazing.

Soreal CAD is cut by grinding equipment with wet processing, and put into the porcelain casting furnace at 840-850°C for 30 min to secondary crystallization treatment. Then dye it.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains a logo with an abstract geometric shape on the left and Chinese characters on the right. The logo and characters are both in orange color. The characters appear to be the name of a company or organization.

They are all composed of SiO2, Li2O, K2O, P2O5, Al2O3 and other oxides. They have a variety of colors, and users can choose the right color to match the patient's natural teeth. In general, our products are also called Dental glass ceramic.

5. Indication for use

Soreal Press is indicated for single-unit anterior or posterior prostheses and for three-unit prostheses, such as veneers, inlay/onlay, partial crowns, posterior crowns, using the hot press technique.

Soreal CAD is indicated for single-unit anterior or posterior prostheses and for three-unit prostheses, such as veneers, inlay/onlay, partial crowns, anterior crowns, using the CAD/CAM system.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains a logo and text. The logo is an abstract shape in orange. To the right of the logo is the text "云甲科技", also in orange. The text appears to be in Chinese.

6. Comparison to Predicate Device

Compared to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following table

SEComparisonsProposed DevicesK230259Predicate DeviceK192231Similarities/Differences
NameSoreal PressDental Glass CeramicsBlocks/
ClassificationClass 2Class 2Same
Indications forUseSoreal Press isindicated forsingle-unitanterior orposteriorprostheses andfor three-unitprostheses, such asveneers, inlay/onlay,partial crowns,anterior crowns,posteriorcrowns, using thehot press technique.Soreal CAD isindicated forsingle-unitanterior orposteriorprostheses andfor three-unitprostheses, such asveneers, inlay/onlay,partial crowns,anterior crowns,posteriorcrowns, using theCAD/CAM system.Dental Glass CeramicsBlocks are indicated forfabricating all-ceramicrestorations such asveneers, inlay/onlay, partialcrowns, anterior crowns,posterior crowns, using thehot press technique orCAD/CAM systemSimilar Note 1
MaterialsSiO2, Li2O,K2O, P2O5,Al2O3 and otheroxidesSiO2, Li2O, K2O, Al2O3 andother oxidesSimilar Note 2
Processing atDental labHot Press(Soreal Press)Hot Press (Up. PressSeries)CAD/CAM (Up. CAD Series)Same
GeometryCylinderBlocksSimilar Note 3
DimensionVariousVariousVariousSame
Single useYesYesYesSame
AvailablecolorVariousVariousVariousSame
SterileNon-sterileNon-sterileNon-sterileSame
UniformityThe colorantsare disperseduniformlythroughout thedental ceramicmaterial and inpowderedceramicproducts.The colorantsare disperseduniformlythroughout thedental ceramicmaterial and inpowderedceramicproducts.Meet the requirements ofISO 6872:2015Same
Freedom ofextraneousmaterialsDevice shall befree fromextraneousmaterialsDevice shall befree fromextraneousmaterialsMeet the requirements ofISO 6872:2015Same
Radioactivityof dentalceramic< 1.0Bq·g-1 of238U< 1.0Bq·g-1 of238UMeet the requirements ofISO 6872:2015Same
FlexuralStrength>300MPa>300MPaMeet the requirements ofISO 6872:2015Similar Note 4
Linear thermalexpansioncoefficient(10±1.5)×10-6K-1(10±1.5)×10-6K-1Meet the requirements ofISO 6872:2015Similar Note 5
Glasstransitiontemperature550±50 °C550±50 °CMeet the requirements ofISO 6872:2015Similar Note 6
Chemicalsolubility<100µg/cm²<100µg/cm²Meet the requirements ofISO 6872:2015Similar Note 7
DensityCeramic blockdensity: 2.4 ~2.6 g/cm³Ceramic blockDensity: 2.3 ~2.6 g/cm³Density ofporcelain blockafter sintering:2.4 ~ 2.7 g/cm³Meet the requirements ofISO 6872:2015Similar Note 8
In VitroCytotoxicityTest (ISO10993-5:2009)No Cytotoxicity effectNo Cytotoxicity effectNo Cytotoxicity effectSame
Irritation OralMucosaIrritation (ISO10993-10:2010)Not a primary oral mucosa irritantunder the conditions of the studyNot a primary oral mucosa irritantunder the conditions of the studyNot a primary oral mucosairritant under the conditionsof the studySame
SkinSensitization(ISO 10993-10:2010)Not a sensitizer under theconditions of the studyNot a sensitizer under theconditions of the studyNot a sensitizer under theconditions of the studySame
Subacute andSubchronicToxicity (ISO10993-11:2017)No subacute and subchronictoxic effects observedNo subacute andsubchronictoxic effects observedNo subacute andsubchronictoxic effects observedSame
Genotoxicity(ISO10993-3:2014)No genotoxic effects observedNo genotoxic effectsobservedNo genotoxic effectsobservedSame
AcuteSystemicToxicity (ISO10993-11:2017)No acute Systemic Toxicity//Different Note1
Pyrogen (ISO10993-11:2017)Meets the requirements for theabsence of pyrogens//Different Note2

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains a logo and Chinese text. The logo is an abstract orange shape, possibly representing a stylized letter or symbol. To the right of the logo, there are four Chinese characters, also in orange, which appear to be the name of a company or organization. The text is crisp and clear, suggesting a professional design.

Traditional 510(k) Submission of Soreal Press & Soreal CAD

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a logo with an abstract orange shape on the left and the text "云甲科技" on the right. The abstract shape is composed of several lines that form a geometric pattern. The text is in Chinese characters and appears to be the name of a company or organization.

Traditional 510(k) Submission of Soreal Press & Soreal CAD

Tips: "/" means that it cannot be applicable.

Similar Note 1: Although their Indications for Use is not same, they all meet the requirement of ISO 6872. So, this difference will not increase risk and affect the effectiveness and safety of the device. Similar Note 2: Although their composition may be slightly different, their main composition is SiO2, Li2O and K2O, and both meet the requirements of ISO 6872:2015. Their chemical and physical properties are basically the same and will not affect the safety and effectiveness of the device.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for YunJia Technology. The logo consists of an orange abstract shape on the left, followed by the company name in Chinese characters on the right. The characters are also orange and appear to be slightly blurred.

Similar Note 3: Since the shape of the equipment can be customized in subsequent processing, the shape will not affect the safety and effectiveness of the device.

Similar Note 4: The Flexural Strength of devices is similar. But both products meet the requirements of ISO 6872:2015. The standard requires that the Flexural Strength should be higher than 300MPa. The data gap does not affect the actual safety and effectiveness of the product. This similarity does not affect the substantial equivalence of the product.

Similar Note 5: According to ISO 6872:2015, the linear thermal expansion coefficient of the ceramics shall not deviate by more than 0.5 × 10-6 K-1 from the value stated by the manufacturer. The linear thermal expansion coefficient of ceramics specified by us is (10±1.5)×10ºK-1, and the test results of ISO 6872 show that the deviation does not exceed 0.5 × 10-6 K-1. The linear thermal expansion coefficient declared by the predicate product also meets the requirements of ISO 6872:2015. So, the differences will not affect performance and safety of devices. This similarity does not affect the substantial equivalence of the product.

Similar Note 6: We cannot get the glass transition temperature of predicate device, so our device may be different from predicate device on glass transition temperature. According to ISO 6872:2015, the glass transition temperature of the ceramics shall not deviate by more than 20 ℃ from the value stated by the manufacturer. The glass transition temperature of ceramics specified by us is 550±50 °C, and the test results of ISO 6872 show that the deviation does not exceed 20 °C. The glass transition temperature declared by the predicate product also meets the requirements of ISO 6872:2015. So, the differences will not affect performance and safety of devices. This similarity does not affect the substantial equivalence of the product.

Similar Note 7: The Chemical solubility of devices is similar. But both products meet the requirements of ISO 6872:2015. The standard requires that the Chemical solubility should be less than 100 uq/cm². The data qap does not affect the actual safety and effectiveness of the product. This similarity does not affect the substantial equivalence of the product.

Similar Note 8: According to 6872:2015, the density is not necessary. We test the density. And the different density will not affect performance and safety of devices. This similarity does not affect the substantial equivalence of the product.

Different Note 1: Predicate Device did not test the Acute Systemic Toxicity. But we test the Acute Systemic Toxicity. The difference will not affect performance and safety of devices.

Different Note 2: Predicate Device did not test the Pyrogen. But we test the Pyrogen. The difference will not affect performance and safety of devices.

All the differences don't affect the safety and effectiveness which is concluded after all the required testing, so no safety and effectiveness issues relating to the system come into conclusion.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for YunJia Technology. The logo consists of an abstract geometric shape on the left, followed by the company name in Chinese characters on the right. The logo and text are both in orange color.

7. Performance Data

Clinical test:

Clinical testing is not required.

Non-clinical data

The Performance tests were performed according to ISO 6872:2015 Dentistry - Ceramic materials, and the test results showed that the proposed device meets the requirements specified in the standard (see the following Table)

Test ItemRequirementsMeasured dataConclusion
UniformityThe colorants aredispersed uniformlythroughout the dentalceramic material and inpowdered ceramicproducts.The colorants are disperseduniformly throughout thedental ceramic material and inpowdered ceramic products.Pass
Freedom ofextraneousmaterialsDevice shall be freefrom extraneousmaterialsDevice shall be free fromextraneous materialsPass
Linear thermalexpansioncoefficient$(10±1.5)×10^{-6}K^{-1}$$11.1×10^{-6}K^{-1}$Pass
Glass transitiontemperature550±50 °CSoreal CAD: 518.4°CSoreal Press: 515.8°CPass
Chemical solubility<100µg/cm²Soreal CAD(µg/cm²)Pass
22.37
21.56
22.38
19.98
23.20
19.37
20.68
23.45
21.76
Soreal Press(µg/cm²)
21.87
22.56
21.33
19.32
21.21
19.34
19.63
22.41
24.12
22.2421.28
Radioactivity ofdental ceramic< 1.0Bq·g⁻¹ of ²³⁸U<0.0073 Bq·g⁻¹ of ²³⁸UPass
Flexural Strength>300MPaSoreal CAD(MPa)Soreal Press(MPa)Pass
389410
367398
388386
365405
378421
386436
377397
398392
423427
385379

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a logo with an abstract orange design on the left and Chinese characters on the right. The design on the left appears to be a stylized geometric shape. The Chinese characters on the right are arranged horizontally and are also in orange. The logo is simple and modern, with a focus on the design and the Chinese characters.

Traditional 510(k) Submission of Soreal Press & Soreal CAD

Biocompatibility testing was performed to verify the equivalent safety of the materials that are used.

The Biocompatibility tests were performed according to ISO 10933 Biocompatibility, and the test results showed that the proposed device meets the requirements specified in the standard (see the following Table)

Test ItemRequirementsMeasured dataConclusion
In VitroCytotoxicity Test(ISO 10993-5:2009)No Cytotoxicity effectNo Cytotoxicity effectPass
Irritation OralMucosa Irritation(ISO 10993-10:2010)Not a primary oral mucosairritant under theconditions of the studyNot a primary oral mucosairritant under the conditionsof the studyPass
Skin Sensitization(ISO 10993-10:2010)Not a sensitizer under theconditions of the studyNot a sensitizer under theconditions of the studyPass

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for YunJia Technology. The logo consists of an orange abstract shape on the left, followed by the company name in Chinese characters on the right. The overall design is simple and modern, with a focus on the company's name and brand identity.

Subacute andSubchronicToxicity (ISO10993-11:2017)No subacute andsubchronic toxic effectsobservedNo subacute andsubchronic toxic effectsobservedPass
Genotoxicity(ISO10993-3:2014)No genotoxic effectsobservedNo genotoxic effectsobservedPass
Acute SystemicToxicity (ISO10993-11:2017)No acute SystemicToxicityNo acute Systemic ToxicityPass
Pyrogen (ISO10993-11:2017)Meets the requirementsfor the absence ofpyrogensMeets the requirements forthe absence of pyrogensPass

8. Conclusion

Information included in this premarket notification supports the substantial equivalence of the proposed Soreal Press/Soreal CAD. The proposed device has the identical intended use, identical indication for us, identical performance, identical fundamental technology, and identical biocompatibility as the predicate device K192231. The results of the testing support a determination of substantial equivalence.

§ 872.6660 Porcelain powder for clinical use.

(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.