K Number
K230249

Validate with FDA (Live)

Device Name
Ikshana
Manufacturer
Date Cleared
2023-10-16

(259 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Ikshana is a software device to display medical images. It includes functions for image review, image manipulation, measurements, and 3D visualization.

Medical images may only be interpreted using an FDA-cleared display monitor that meets technical specifications that are reviewed and accepted by the FDA.

Ikshana is intended to be used as an adjunct to the interpretation of images performed using diagnostic imaging systems and is not intended for primary diagnosis. Display monitors used for reading medical images for diagnostic purposes must be FDA-approved radiology monitors.

Ikshana software is indicated for use by qualified healthcare professionals, including, but not restricted to, radiologists, non-radiology specialists, physicians, and technologists.

When accessing the Ikshana software from a wireless stereoscopic head-mounted display (HMD) or mobile device, the images viewed are for informational purposes only and are not intended for diagnostic use.

Device Description

Ikshana is a stand-alone modular software platform to be used by clinicians for the visualization of medical images in 3D to allow for surgical planning activities. The device takes 2D medical images and creates accurate 3D representations that clinicians can then view on a stereoscopic display. This modular package is used to

  • · Load patient CT/MR DICOM data
  • . View DICOM data using a traditional computer monitor or in Augmented Reality (AR) using a head-mounted display, HMD (Microsoft HoloLens 2).
AI/ML Overview

The provided document describes the acceptance criteria and the study conducted for the Ikshana device, particularly focusing on its measurement and segmentation capabilities.

1. Table of Acceptance Criteria and Reported Device Performance:

Feature/MetricAcceptance CriteriaReported Device Performance
Measurement StudyInter and intra-user variability within set acceptable limits. Paired t-tests comparing Ikshana to Medical Mimics measurements (p > 0.05). Bland-Altman plots showing 95% of differences within acceptable limits.All inter and intra-user measurements fell within the set acceptance criteria. Paired t-tests resulted in p-values > 0.05, indicating no significant difference between Ikshana and Medical Mimics measurements. Bland-Altman plots confirmed high equivalence with 95% of differences within acceptable limits.
Segmentation StudyVisual comparison with reference device showing high level of equivalence. Average DICE coefficient representing high agreement. Paired t-test comparing volume measurements (p > 0.05).Visual comparison showed a high level of equivalence between Ikshana and Mimics Medical. Average DICE coefficient of approximately 96% for 60 trials. Paired t-test resulted in a p-value above 0.05, suggesting likely equivalence between the two methods.

2. Sample Size Used for the Test Set and Data Provenance:

  • Measurement Study: A combination of orthopedic and maxillofacial models was used. The specific number of cases for the test set is not explicitly stated, but the study evaluated "multiple users."
  • Segmentation Study: 60 trials were conducted using a combination of cardiovascular, orthopedic, and maxillofacial models. The specific number of cases is not explicitly stated beyond "60 trials."
  • Data Provenance: Not specified in the provided text (e.g., country of origin or whether it was retrospective/prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Segmentation Study: The results were "validated by subject matter experts." The specific number and qualifications of these experts are not mentioned.
  • Measurement Study: The ground truth for the measurement study seems to be derived from a comparison with the reference device, Medical Mimics, which implies its measurements are considered a reference standard, rather than expert consensus on individual cases.

4. Adjudication Method for the Test Set:

  • The document does not explicitly describe an adjudication method like 2+1 or 3+1. For the segmentation study, it states results were "validated by subject matter experts," which suggests some form of expert review, but the specific process is not detailed. For the measurement study, the comparison is directly with the reference device's measurements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described for assessing the improvement of human readers with AI assistance. The studies focused on the performance of the device itself (measurement and segmentation accuracy/equivalence).

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, the performance studies described (Measurement Study and Segmentation Study) appear to be standalone evaluations of the Ikshana software's capabilities. The measurement study compared Ikshana's measurements to another device, and the segmentation study compared Ikshana's segmentation models to a reference device, with expert validation. This assesses the algorithm's performance directly.

7. The Type of Ground Truth Used:

  • Measurement Study: The ground truth for the measurement study appears to be established through measurements obtained using the referenced predicate device, Mimics Medical. The study aimed to show "equivalence" with the predicate, implying the predicate's measurements serve as the reference.
  • Segmentation Study: The ground truth for the segmentation study was established by comparing Ikshana's segmentation models with those from the previously cleared reference device, Mimics Medical (K183105), and these results were "validated by subject matter experts." This indicates a hybrid approach, using a cleared device as a primary reference and expert review for validation.

8. The Sample Size for the Training Set:

  • The document does not provide information about the sample size used for the training set. The focus is on the performance evaluation of the final device.

9. How the Ground Truth for the Training Set Was Established:

  • The document does not provide information on how the ground truth for the training set was established, as details about the training process are not included.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

ImmersiveTouch Prashant Banerjee CEO 910 W Van Buren Street, Suite # 715 CHICAGO, ILLINOIS 60607

October 16, 2023

Re: K230249

Trade/Device Name: Ikshana Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management and Processing System Regulatory Class: Class II Product Code: LLZ Dated: September 11, 2023 Received: September 14, 2023

Dear Prashant Banerjee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Jessica Lamb

Jessica Lamb Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230249

Device Name Ikshana

Indications for Use (Describe)

Ikshana is a software device to display medical images. It includes functions for image manipulation, measurements, and 3D visualization.

Medical images may only be interpreted using an FDA-cleared display monitor that meets technical specifications that are reviewed and accepted by the FDA.

Ikshana is intended to be used as an adjunct to the interpretation of images performed using diagnostic imaging systems and is not intended for primary diagnosis. Display monitors used for reading medical images for diagnostic purposes must be FDA-approved radiology monitors.

Ikshana software is indicated for use by qualified healthcare professionals, including, but not restricted to, radiologists, non-radiology specialists, physicians, and technologists.

When accessing the Ikshana software from a wireless stereoscopic head-mounted display (HMD) or mobile device, the images viewed are for informational purposes only and are not intended for diagnostic use.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. SUBMITTER

ImmersiveTouch, Inc. 910 W Van Buren Street, Suite #715, Chicago, IL 60607

PHONE: +1 630 570 5943 Fax: +1 908 634 1219

Contact Person: Dr. P. Pat Banerjee Date Prepared: January 15th, 2023

II. DEVICE

Device Name: Ikshana Regulation Name: Medical Image Management and Processing System Regulation Number: 21 CFR 892.2050 Regulation Class: II Product Code: LLZ

III. PREDICATE DEVICE

Ikshana is claimed to be substantially equivalent to predicate device SurgicalAR (K190764).

Ikshana is also substantially equivalent to reference device Mimics Medical (K183105) in its ability to perform image segmentation based on input data.

IV. DEVICE DESCRIPTION

Ikshana is a stand-alone modular software platform to be used by clinicians for the visualization of medical images in 3D to allow for surgical planning activities. The device takes 2D medical images and creates accurate 3D representations that clinicians can then view on a stereoscopic display. This modular package is used to

  • · Load patient CT/MR DICOM data
  • . View DICOM data using a traditional computer monitor or in Augmented Reality (AR) using a head-mounted display, HMD (Microsoft HoloLens 2).
S. No.SpecificationsHoloLens 2 (Head Mounted Display)
1The manufacturer and model of the HMDManufacturer - MicrosoftHMD Model - Microsoft HoloLens 2
S. No.SpecificationsHoloLens 2 (Head Mounted Display)
2The number of horizontal andvertical pixels of the display per eyeResolution 1440×936 per eye
3Horizontal and vertical field of viewGiven that Microsoft said the HoloLens 2 FoVhas a 3:2 "screen" ratio, some basic math showedthe device had a horizontal FoV of 43° and avertical of 29°.
4Number of pixels per degree33 pixels per degree of field of view
5The minimum and maximumluminance of the displayThe maximum luminance of the HoloLens 2display is reported at 500 cd/m2, and theminimum luminance is 0 (for an empty pixel).
6Unity depth buffer sizeUnity depth textures at the HoloLens 2 end are 16bits. 24-bit textures are used before images aresent to the HoloLens for the purposes of accuratepixel sorting, but this is down sampled to 16-bitsbefore sending over the network, where it is usedfor latency compensation. The far planes of thevirtual cameras used for rendering are set to under10 meters, down from Unity's default value of1000 meters, to maximize the accuracy of thesedepth values.
7Maximum temperature andrecommended temperature of theHMDThe devices operate in a temperature range of10°C to 35°C.
8HMD's interpupillary distance(IPD), the adjustment range of IPD:The HoloLens 2 display actively color-correctsimages based on the position of the user's eyes.Eye calibration provides two critical inputs:(1) the user's interpupillary distance (IPD) and(2) the direction each eye is looking. Without eyecalibration, the system defaults to a nominal eyeposition without eye movement.
9Eye relief and compatibility withcorrective lenses18mm to 23mm
10Distance of the virtual content fromthe user in the software designHoloLens displays are fixed at an optical distanceof approximately 2.0 m away from the user. Usersmust always accommodate nearly 2.0 m tomaintain a clear image in the device.

HoloLens 2 Specifications

{4}------------------------------------------------

{5}------------------------------------------------

V. INDICATIONS FOR USE

Ikshana is a software device to display medical images. It includes functions for image review, image manipulation, measurements, and 3D visualization.

Medical images may only be interpreted using an FDA-cleared display monitor that meets technical specifications that are reviewed and accepted by the FDA.

Ikshana is intended to be used as an adjunct to the interpretation of images performed using diagnostic imaging systems and is not intended for primary diagnosis. Display monitors used for reading medical images for diagnostic purposes must be FDA-approved radiology monitors.

Ikshana software is indicated for use by qualified healthcare professionals, including, but not restricted to, radiologists, non-radiology specialists, physicians, and technologists.

When accessing the Ikshana software from a wireless stereoscopic head-mounted display (HMD) or mobile device, the images viewed are for informational purposes only and are not intended for diagnostic use.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Ikshana employs similar fundamental technologies as the identified predicate devices, including:

  • The subject and predicate device both have the same intended use.
  • . The subject and predicate device both have similar abilities to process, review and analyze medical imaging data.
  • The subject and predicate device both have similar functionalities to perform measurements · and aid in surgical planning activities.

Differences

Ikshana can perform segmentation, which is different from the predicate listed.

Comparison with reference device of K183105 - Mimics Medical

Intended use and Functionality - Both Ikshana and Mimics Medical offer advanced image processing capabilities, including segmentation and 3D visualization. Ikshana and Mimics Medical can perform image segmentation based on the input data. However, Mimics Medical has a broader range of functionalities. Ikshana is primarily designed for 3D visualization and pre-surgical planning using diagnostic imaging systems and is not intended for primary diagnosis. In contrast, Mimics Medical encompasses a wider range of applications, including treatment planning, and the creation of output files for additive manufacturing. Ikshana specifies that medical images must be interpreted on FDAapproved radiology monitors, while Mimics Medical does not include specific display requirements in its intended use.

The results for Segmentation study have been compared with a previously cleared reference device Mimics Medical (K183105). Mimics Medical (K183105), a software interface and image segmentation system that is intended for measuring and treatment planning was used in the Performance Testing to

{6}------------------------------------------------

compare the results of the subject device. All of the results showed under this specific chapter show a high level of equivalence between Ikshana and Mimics Medical.

FeaturesDescriptionSubject DevicePredicate Device
IkshanaSurgicalAR
Measurement ToolLineAngleYesYes
User Installation RequirementsAs per IFUYesYes
Data Type SupportedDICOMYesYes
Image View/ManipulationImage ZoomPanWindow LevelAuto WindowResetImage RotateImage FlipMagnify/ZoomYesYes
Data EncryptionHTTPSYesYes
Patient Information DisplayCapable of displaying patient informationYesYes
LinkingCo-planar linking:Auto linkManualYesYes
User and Password ControlYesYesYes
Data SecurityStored on serverYesYes
Audit TrailAudit trail loggedYesYes
User ManagementDatabase structure allows mapping users to group internally or mapping external groupsYesYes
FeaturesDescriptionSubject DevicePredicate Device
IkshanaSurgicalAR
Transmission ModesVia the web with Internet browsersYesYes
MPR ViewingThis viewing feature enables the display ofreformatted CT and MR images into axial,coronal and sagittal orientationsYesYes
3D Volume RenderedViewingThis viewing feature enables the display of3D perspective views of CT and MR imagesets that have been transformed intovolumes. It also provides presets to enableusers to alter the visualization parametersof the 3D views to highlight features.YesYes
Active Target ToolThis viewing feature provides a facility toview a single target locationYesYes
Crosshair Navigationand SynchronizationThis viewing feature provides a facility tosynchronize and scroll through multipleviews at the same time. (2D slice view)YesYes
Ability to close animage by clicking an"X" in the upper-leftportion of the viewportAbility to close an image by clicking an"X" in the upper-left portion of theviewport.YesYes
HMD support forvisualization purposeonly (not for diagnosticuse)This viewing feature makes Ikshanacompatible with off-the shelf-wireless, Wi-Fi enabled, stereoscopic head-mounteddisplayYesYes
Diagnostic qualitymedical image reviewAbility to provide diagnostic qualitymedical image review for multi-dimensional digital images acquired from avariety of imaging devices- via FDAapproved diagnostic monitorsYesYes
Image SegmentationAbility to perform segmentation based onthe input dataYesNo
Intended UseEnvironmentIntraoperative useNoNo

{7}------------------------------------------------

{8}------------------------------------------------

Operator profile

Qualified healthcare professionals, including but not restricted to surgeons, radiologists, nonradiology specialists, physicians, and technologists.

Patient population

The device is software that allows for viewing of DICOM data. Therefore, this is not restricted to a specific patient population.

Operating principle

There are different operating principles such as viewing:

  • On desktop PCs with a traditional monitor the interaction with the software is performed . with a mouse and/or keyboard.
  • On a head-mounted display, the interaction with the software is performed using hand . gestures.

VII. PERFORMANCE DATA

Software Verification and Validation Testing

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

This includes verification against defined requirements, and validation against user needs. Both enduser validation and bench testing were performed.

Software verification and validation includes:

  • · Verification of each independent software subsystem against defined requirements.
  • . Verification of interfaces between software subsystems against defined requirements.
  • . Validation of fully integrated systems including all subsystems against overall system requirements.

Measurements Study and Segmentation Study

The purpose of the Measurement study was to evaluate, measure and compare the inter and intra user variability between measurements taken by multiple users in the subject device. Comparison of the inter and intra user measurements showed that all measurements fell within the set acceptance criteria. A combination of orthopedic, and maxillofacial models was used for this study.

For the measurement study the paired t-tests comparing Immersive Touch Ikshana measurements to Medical Mimics measurements resulted in p-values greater than 0.05, indicating a lack of significant difference between the two. Bland Altman plots also confirmed a high level of equivalence for both linear and angular measurements, with 95 percent of differences falling within acceptable limits. Overall, these findings suggested that Immersive Touch Ikshana and Medical Mimics measurements are likely equivalent.

The purpose of the Seementation study was to visually compare segmentation models created by the subject device with a previously cleared reference device Mimics Medical (K183105). The results were validated by subject matter experts. The comparison showed a high level of equivalence between Ikshana and Mimics Medical for all the anatomical models. A combination of cardiovascular, orthopedic, and maxillofacial models was used. For the segmentation study, an average DICE

{9}------------------------------------------------

coefficient of approximately 96% was achieved for 60 trials comparing ImmersiveTouch and Materialize Mimics volumes. The paired t-test resulted in a p-value above 0.05, suggesting that the null hypothesis of no significant difference between the two volume measurements cannot be rejected, indicating likely equivalence between the two methods.

The results for Segmentation study have been compared with a previously cleared reference device Mimics Medical (K183105). Mimics Medical (K183105), a software interface and image segmentation system that is intended for measuring and treatment planning was used in the Performance Testing to compare the results of the subject device. A measurement accuracy study was conducted to analyze and validate the reproducibility and the accuracy of measurements in subject device in comparison to measurements made in Materialize Mimics. All of the results showed a high level of equivalence between Ikshana and Mimics Medical.

AR HMD Testing

The functioning of the AR HMD was tested through validation tests and the 3D anatomical representations were successfully visualized via the Head mounted display.

Summary

The performance data indicates that the verification and validation testing performed on Ikshana successfully demonstrates conformity to pre-established specifications and acceptance criteria. The acceptance criteria were established in order to demonstrate device performance and substantial equivalence of the software to the predicate device.

VII. CONCLUSIONS

Ikshana demonstrates substantial equivalence to its predicate devices incorporate similar inputs, operations, and outputs.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).