K Number
K230155

Validate with FDA (Live)

Date Cleared
2023-08-30

(223 days)

Product Code
Regulation Number
892.5700
Age Range
0 - 21
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the MICK Valencia Applicator is to facilitate the placement of a radioactive source near the target area on intact skin for remote afterloading brachytherapy treatment.

Device Description

The Mick Valencia Applicator is used for remote afterloading treatment on intact skin/surface. The applicator with associated plastic cap is placed onto the patient, locked into position, and connected to the remote afterloading equipment (via transfer tube) to facilitate the brachytherapy treatment in accordance with the physician approved treatment parameters. The radioactive source travels from the machine into the applicator and deliver the desired dose distribution to the defined treatment volume. When the treatment is completed, the source returns to the machine. The applicator is disconnected from the treatment unit and removed from the patient.

The applicator does not control the treatment unit, it strictly provides a treatment path for the radioactive source. The Mick Valencia applicator is temporarily placed during the treatments designed as a closed system to prevent the radioactive source from coming in contact with bodily fluids.

The device is the same as the legally marketed device cited, both in design and materials.

AI/ML Overview

The provided document (K230155) is a 510(k) premarket notification for the Mick Valencia Applicator Set. It does not contain information about a study proving device performance against acceptance criteria in the way one would expect for an AI/ML medical device.

This document describes a medical device used for brachytherapy, which is a physical applicator for radioactive sources. The determination of substantial equivalence for this type of device relies on demonstrating that its design, materials, and intended use are similar to a legally marketed predicate device, and that it performs similarly. The "non-clinical tests" mentioned are bench tests for connection compatibility, not performance trials for diagnostic accuracy or clinical outcomes.

Therefore, I cannot provide the requested information regarding acceptance criteria and study details because the document does not report on a study that would typically have such components. There is no mention of a "device" in the context of an AI/ML algorithm that would require ground truth, expert consensus, MRMC studies, or training/test sets.

Here's why the requested information cannot be extracted:

  • No AI/ML Component: The device, the Mick Valencia Applicator Set, is a physical medical instrument (an applicator for a radioactive source). It's not an AI or software device.
  • Substantial Equivalence, Not Performance Study: The entire submission is built around demonstrating "substantial equivalence" to a predicate device (Nucletron/Elekta Valencia Skin Applicator K073107). This is a comparative assessment, not an independent performance study of a novel device
  • No "Ground Truth" or "Experts": For a physical applicator, "ground truth" and "experts" in the context of diagnostic performance (e.g., radiologists interpreting images) are not relevant. The "ground truth" for this device's function would be its physical specifications and compatibility with afterloaders, which are verified through engineering tests and comparison to the predicate.
  • No "Training/Test Sets": These concepts apply to machine learning models, which are absent here.
  • No "MRMC Study": A Multiple Reader Multiple Case study is for evaluating human interpretation with and without AI assistance, which is not applicable to a physical brachytherapy applicator.

The document states:

  • "Non-clinical testing is performed on the Mick Valencia Applicator connection to the transfer tube (the transfer tubes are supplied by the Afterloader manufacturer). These are bench top tests, performed on every applicator produced."
  • "The Mick Valencia Applicator Sets are substantially equivalent in design, manufacture, construction and materials and have the same intended use and performance characteristics to the predicate device."

These statements indicate that the "proof" of the device meeting its acceptance criteria (which are implicitly its functional specifications, material properties, and compatibility) comes from bench testing and the demonstration of substantial equivalence to a device already on the market. There's no performance study in the sense of a clinical trial or algorithm validation study described.

{0}------------------------------------------------

August 30, 2023

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Mick Radio-Nuclear Instruments, Inc. % James Hurlman Director, Quality Management & Regulatory Affairs 521 Homestead Avenue MOUNT VERNON NY 10550

Re: K230155

Trade/Device Name: Mick Valencia Applicator Set Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: Class II Product Code: JAQ Dated: August 2, 2023 Received: August 3, 2023

Dear James Hurlman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Locon Werder

Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230155

Device Name

MICK Valencia Applicator Sets

Indications for Use (Describe)

The intended use of the MICK Valencia Applicator is to facilitate the placement of a radioactive source near the target area on intact skin for remote afterloading brachytherapy treatment.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MICK, a BEBIG Medical company. The word "MICK" is written in large, blue, sans-serif font. Below the word "MICK" is the phrase "A BEBIG Medical company" written in a smaller, orange, sans-serif font.

Mick Radio-Nuclear Instruments, Inc. A BEBIG Medical Company

521 Homestead Ave Mount Vernon, NY 10550 USA www.micknuclear.com

Main: +1 914 667 3999 Fax: +1 914 665 8834 Toll Free: +1 877 597 MRNI E-Mail: isaac.strickland@micknuclear.com

Traditional 510(k) Summary of Safety and Effectiveness Information As required by section 807.92(c) K230155

August 28, 2023

1. General Provisions

Common Name:Remote controlled radionuclide brachytherapy source (JAQ
Proprietary Name:Mick Valencia Applicator Set
Owner NameMick Radio-Nuclear Instruments, Inc.
Address521 Homestead AvenueMount Vernon, New York 10550
Telephone(914) 667-3999
Fax(914) 665-8834
Contact NameIsaac Strickland
Contact emailisaac.strickland@micknuclear.com

Name of Predicate Device: 2.

Original510(k) #Trade/ Device NameRegulationNumberRegulatoryClassificationProductCodeReview Panel
K073107Valencia Skin Applicator Set21 CFR 892.5700Class IIJAQRadiology

3. Classification

Class 2, Product Code: JAQ Regulation Number: 21 CFR 892.5700

{4}------------------------------------------------

4. Intended Use

The intended use of the Mick Valencia Applicator is to facilitate the placement of a radioactive source near the target area on intact skin for remote afterloading brachytherapy treatment.

Device Description 5.

The Mick Valencia Applicator is used for remote afterloading treatment on intact skin/surface. The applicator with associated plastic cap is placed onto the patient, locked into position, and connected to the remote afterloading equipment (via transfer tube) to facilitate the brachytherapy treatment in accordance with the physician approved treatment parameters. The radioactive source travels from the machine into the applicator and deliver the desired dose distribution to the defined treatment volume. When the treatment is completed, the source returns to the machine. The applicator is disconnected from the treatment unit and removed from the patient.

The applicator does not control the treatment unit, it strictly provides a treatment path for the radioactive source. The Mick Valencia applicator is temporarily placed during the treatments designed as a closed system to prevent the radioactive source from coming in contact with bodily fluids.

The device is the same as the legally marketed device cited, both in design and materials.

Image /page/4/Picture/6 description: The image shows two sets of metal objects. Each set contains two objects that appear to be probes or sensors. The objects have a cylindrical base and a rounded or cylindrical top, with a thin metal rod extending from the side.

MICK Valencia Skin Applicator

Elekta Valencia Skin Applicator

Image /page/4/Figure/9 description: The image shows a technical drawing with several dimensions labeled. The dimensions include Dim. A, Dim. B, Dim. C, Dim. D, and Dim. E. The drawing also labels the source cable and shielding of the device. Additionally, a measurement of 3.75 is shown on the right side of the image.

MICK Valencia Skin Applicator

{5}------------------------------------------------

SetModelDescriptionDim ADim BDim CDim DDim E
1503(BEBIG)1503-20Valencia applicator20mm20289591.25105
1503-30Valencia applicator30mm3036115111.25129
1504(Varian)1504-20Valencia applicator20mm2028113109.25123
1504-30Valencia applicator30mm3036113109.25127

6. Manufacturing Process

This device is manufactured according to Good Manufacturing Practices (GMPs) as defined in 21CFR part 820. The processes used to fabricate these devices are identical to those used for the predicate device described in this 510(k) notification.

7. Non-Clinical Tests

Non-clinical testing is performed on the Mick Valencia Applicator connection to the transfer tube (the transfer tubes are supplied by the Afterloader manufacturer). These are bench top tests, performed on every applicator produced.

8. Afterloader Compatibility

The Mick Valencia Applicator Set is compatible with the following HDR afterloaders, BEBIG Medical GmbH model Saginova®. Varian model GammaMedplus™ and Varian model Bravos afterloaders. The Mick Valencia Applicator sets are not compatible with Elekta afterloaders.

9. Biocompatibility

The materials in the subject device are the same as the previously cleared predicate device and thus, there are no new biocompatibility concerns.

10. Summary of Similarities and Differences

The Mick Valencia Applicator Sets are substantially equivalent in design, manufacture, construction and materials and have the same intended use and performance characteristics to the predicate device. The fundamental scientific technology is unchanged from the original predicate device. The connectors used on the Mick Valencia Applicator sets, to attach to the afterloaders transfer tubes, are substantially equivalent to the predicate device in their fundamental design for a standard click fit type connector.

Some differences from the predicate device are as follows;

  1. Our device has more of a dome shape then the predicate device,

{6}------------------------------------------------

    1. the shaft and tube diameters are slightly thicker, and
    1. the Mick Valencia Applicator set comes with two marker caps; one of the marker caps is used for applicator placement, while the other cap is placed on the target area for treatment.
    1. The Mick Valencia Applicator set comes in two sizes, in 20 mm & 30 mm diameters and both sizes are compatible with the BEBIG Medical GmbH model Saginova®, the Varian model GammaMedplus™ and the Varian model Bravos afterloaders.

11. Comparison Table

The MICK Valencia Applicator Set has the same functionality, performance and design as the predicate device the Nucletron/Elekta Valencia Skin Applicator. The equivalence is based on a comparison of clinical, technical and biological aspects of the two devices. The following table identifies the clinical, technical and biological aspects of the predicate device and the MICK Valencia Skin Applicator Set:

Equivalence TablePredicate DeviceNucletron/ElektaValencia Skin ApplicatorK073107MICK Valencia SkinApplicator SetsK230155
Clinical Aspects
Intended Purpose: HDR Brachytherapy ofskin/surface lesionsYesYes
Suitable for any patient with skin/surface lesions(benign or malignant) regardless of their age,gender, ethnicity or predisposition.YesYes
Can be used for any site on the body surface thathas been prescribed HDR BrachytherapytreatmentYesYes
Applicator is placed in direct contact with thetreatment area/skin surface.YesYes
Long term side effects are limited to a slight markor discoloration of the skin and/or hair loss in thetreatment area and is dependent on theprescription doseYesYes
Patient is under the care of a RadiationOncologist, physicists, dosimetrists and radiationtherapists trained in Radiation Oncology and,specifically, brachytherapy techniquesYesYes
Technical Aspects
Available in 20 mm and 30mm diameterYesYes
Only one source position is used for thetreatmentYesYes
Source delivery through a single channel (transferYesYes
tube)
Tungsten shielded applicator with reinforcedshielding in the dome to reduce dose from thebackside of the applicatorYesYes
Includes flattening filter for homogeneous dosedistributionYesYes
Plastic caps available: providing a flat surfaceimproved homogeneous doseYesYes
Filtering low dose scatter radiationYesYes
Improved applicator positioning byimprinting skin surfaceYesYes
2 types of caps available (with and withouta marker)YesYes
The device is reusableYesYes
The applicator can be cleaned and disinfectedYesYes
Applicator is MR unsafeYesYes
Provides and unobstructed path for theradioactive sourceYesYes
Connects securely to remote afterloadingbrachytherapy equipment via transfer tubesYesYes
Biological Aspects
The applicators are made from the samematerials. (tungsten, plastic caps, stainless steel)YesYes
The applicators have the same manufacturingprocessYesYes
The applicators come in direct contact with skinsurfaceYesYes
Caps are provided for treatment to providefiltration of scatter radiation.YesYes
The applicators are made from materials that arebiocompatibleYesYes
Provided Non-SterileYesYes

{7}------------------------------------------------

Summary of technological considerations: 12.

The Mick Valencia Applicator Set is substantially equivalent to the cleared predicate device Valencia Skin Applicator Set, 510(k): K073107

§ 892.5700 Remote controlled radionuclide applicator system.

(a)
Identification. A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.(b)
Classification. Class II.