K Number
K230148

Validate with FDA (Live)

Device Name
Vlab
Date Cleared
2023-10-11

(265 days)

Product Code
Regulation Number
882.1835
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DormoTech Vlab is a physiological data recorder intended to collect and record data from multiple physiological channels for use by clinical software used in polysomnography and sleep disorder studies. It is intended for use by or on the order of a physician is intended for use on adults in a supervised (hospital) or unsupervised (home) environment.

Device Description

The DormoTech Vlab is a physiological data recorder intended to collect and record data from multiple physiological channels for use by clinical software used in polysomnography and sleep disorder studies. It consists of: The Head Unit, The Body Unit, and The Central Unit.

AI/ML Overview

This FDA 510(k) summary describes the DormoTech Vlab, a physiological data recorder intended for polysomnography and sleep disorder studies. The acceptance criteria and the study proving it meets these criteria are detailed below.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document doesn't explicitly state "acceptance criteria" in a tabular format with specific thresholds. However, it presents a clinical study comparing the DormoTech Vlab to a "gold standard PSG study" by evaluating the agreement of various physiological parameters. The "conclusion" section of the clinical study acts as the implicit acceptance criteria, indicating that "Most parameters show good agreement between the devices, as indicated by the mean difference values close to zero and narrow limits of agreement." This implies that the device is considered acceptable if its measurements are statistically comparable to the gold standard.

The table below summarizes the device's performance based on the clinical study's agreement analysis (Bland-Altman statistics) between the DormoTech Vlab and a gold standard PSG (likely the NOX Sleep System, K192469, which is used as a reference).

ParameterMean Difference (Lower CI, Upper CI)Upper Limit of Agreement (Lower CI, Upper CI)Lower Limit of Agreement (Lower, Upper CI)
AHI (events/h)-0.1927 (-1.323, 0.9372)6.823 (4.866, 8.78)-7.209 (-9.166, -5.252)
ODI (events/h)-0.3244 (-1.108, 0.4597)4.544 (3.186, 5.902)-5.193 (-6.551, -3.835)
Snore (%)1.085 (-0.525, 2.523)10.01 (7.524, 12.5)-7.843 (-10.33, -5.353)
Sleep Latency (Minutes)4.653 (-0.9411, 10.25)38.01 (28.32, 47.7)-28.7 (-38.39, -19.01)
REM Latency-15.64 (-25.95, -5.327)44.98 (27.12, 62.85)-76.27 (-94.13, -58.4)
Wake after Sleep Onset (Minutes)-4.300 (-10.53, 1.926)31.77 (20.98, -42.55)-40.37 (-51.15, -29.58)
REM (%)0.4816 (-0.801, 1.764)8.129 (5.908, 10.35)-7.166 (-9.388, -4.945)
N1 (%)0.3263 (-1.839, 2.492)13.24 (9.488, 16.99)-12.59 (-16.34, -8.836)
N2 (%)-2.484 (-5.084, 0.1152)13.02 (8.513, 17.52)-17.98 (-22.49, -13.48)
N3 (%)1.011 (-0.07236, 2.093)7.468 (5.592, 9.343)-5.447 (-7.322, -3.571)
Wake (%)0.1972 (-1.301, 1.696)8.877 (6.282, 11.47)-8.483 (-11.08, -5.887)
Total Sleep Time (Minutes)0.72222 (-6.869, 8.313)44.69 (31.55, 57.84)-43.25 (-56.4, -30.1)
Sleep Efficiency (%)-0.03333 (-1.536, 1.47)8.673 (6.07, 11.28)-8.74 (-11.34, -6.136)
Position (Up) (%)0.01316 (-0.4649, 0.4913)2.864 (2.036, 3.692)-2.838 (-3.666, -2.01)
Position (Supine) (%)0.9974 (-0.3433, 2.338)8.991 (6.669, 11.31)-6.997 (-9.319, -4.675)
Position (Left) (%)0.3579 (-0.9967, 1.712)8.435 (6.089, 10.78)-7.719 (-10.07, -5.373)
Position (Right) (%)-0.3974 (-1.61, 0.8149)6.831 (4.732, 8.931)-7.626 (-9.726, -5.526)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 47 subjects.
  • Data Provenance: Prospective clinical study conducted in two sleep labs in Israel:
    • Shamir Medical Center – Be'er Ya'akov, Israel
    • Millenium Sleep Clinic - Be'er Sheva, Israel
      The study was "comparative, self-controlled, randomized, prospective study designed to assess the Vlab and compare its performance to a gold standard polysomnogram (PSG) conducted over 1 night in a sleep lab."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not explicitly state the number of experts used for ground truth establishment or their specific qualifications (e.g., "radiologist with 10 years of experience"). However:

  • The "gold standard polysomnogram (PSG)" implies scoring by trained sleep technologists or physicians, as PSG analysis typically requires specialized expertise.
  • The "Conclusion" section mentions "the role of human scoring," which suggests human experts were involved in generating the ground truth from the gold standard PSG data.

4. Adjudication Method for the Test Set

The document does not specify the adjudication method used for the test set (e.g., 2+1, 3+1, none). It only mentions that the study compared the Vlab's performance to a "gold standard PSG study" with implied human scoring.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study, as typically described (measuring human reader improvement with AI vs. without AI assistance), was not done. The study focused on the standalone performance of the DormoTech Vlab device in comparison to a gold standard PSG system, not on how the Vlab system assists human interpretation.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

Yes, a standalone performance study was done. The clinical study directly compared the measurements obtained from the DormoTech Vlab ("the device") to those from a "gold standard PSG study." The Vlab device collects and records physiological data for use by "clinical software," and the performance reported is of the device's data collection compared to the gold standard, rather than evaluating the accuracy of any integrated AI for interpretation or how it assists a human.
The 510(k) summary states: "The subject and predicate device are sensor arrays for use in collecting and transmitting data from sleep studies that is analyzed by automated, FDA-cleared software, which is not part of this submission." This reinforces that the Vlab's performance study is focused on its ability to acquire signals accurately, which is then fed into other (cleared) software for analysis.

7. Type of Ground Truth Used

The ground truth used was established via a "gold standard PSG study." This gold standard involves the comprehensive recording of physiological signals during sleep, which are then typically scored and interpreted by trained professionals according to established clinical guidelines (e.g., AASM rules). The measurements from these gold standard PSG studies (e.g., AHI, sleep stages) served as the reference against which the DormoTech Vlab's measurements were compared.

8. Sample Size for the Training Set

The document does not provide any information about a training set or its sample size. This is a performance study comparing the device to a gold standard, not a study describing the development or training of an AI algorithm within the DormoTech Vlab itself. The Vlab is described as a "physiological data recorder," and any software for analysis is mentioned as "clinical software used in polysomnography and sleep disorder studies" that is "FDA-cleared software, which is not part of this submission."

9. How the Ground Truth for the Training Set Was Established

As no information about a training set is provided, there is no description of how ground truth for a training set was established.

{0}------------------------------------------------

October 11, 2023

FDA U.S. FOOD & DRUG
ADMINISTRATION

DormoTech Medical Ltd. % Paul Dryden Consultant ProMedic Consulting, LLC 131 Bay Point Dr. NE Saint Petersburg, Florida 33704

Re: K230148

Trade/Device Name: Vlab Regulation Number: 21 CFR 882.1835 Regulation Name: Physiological Signal Amplifier Regulatory Class: Class II Product Code: GWL, MNR Dated: September 11, 2023 Received: September 11, 2023

Dear Paul Dryden:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Patrick Antkowiak -S

for Jay Gupta Assistant Director

{2}------------------------------------------------

DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K230148

Device Name

DormoTech Vlab

Indications for Use (Describe)

The DormoTech Vlab is a physiological data recorder intended to collect and record data from multiple physiological channels for use by clinical software used in polysomnography and sleep disorder studies. It is intended for use by or on the order of a physician is intended for use on adults in a supervised (hospital) or unsupervised (home) environment.

Type of Use (Select one or both, as applicable)

7XX Prescription Use (Part 21 CFR 801 Subpart D)

█ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20) 443-6740

Page 1 of 1

SC Publishing Services (301)

{4}------------------------------------------------

Date Prepared:11-Sep-2023
Sponsor:DormoTech Medical Ltd. YitzhakRabin 21Afula, Israel+ 972-4-7799729
Sponsor Contact:Abed Nassir, Head of Firm
Submission Contact:Paul DrydenProMedic, LLC
Proprietary or Trade Name:Common/Usual Name:Classification Name:Product CodeRegulation NumberDormoTech VlabPhysiological Signal AmplifierPhysiological Signal AmplifierGWL, MNR21 CFR 882.1835
Predicate Device:Reference Device:Respironics, Alice PDx, K090484NOX, Nox Sleep System,K102469

Device Description:

The DormoTech Vlab is a physiological data recorder intended to collect and record data from multiple physiological channels for use by clinical software used in polysomnography and sleep disorder studies.

It consists of:

The Head Unit

The head unit acquires electric signals indicative of EEG and eye movement located in the upper part of the unit (on the patient's forehead). Head relative position are also measured using accelerometer sensors located in the upper part of the unit. The middle part of the unit is located below the mouth, it contains 2 nasal (one in each nostril) and 1 oral airflow sensors, 1 EMG sensor, along with snoring sensor.

The Body Unit

The body unit is made of 2 belts, the upper belt sits on the chest, and the stomach of the patient. Both belts contain respiratory effort and accelerometer sensors, in addition, the upper belt contains an accelerometer to measure body position.

The Central Unit

Both the head and chest units communicate with the central units via Bluetooth, the wearable units send the measured data to the central unit. The central unit receives the data, it stores it within an internal Flash drive and then transmits the data via Wi-Fi to online servers for further diagnosis. Central unit, located in the test room (up to 10 meters from the patient). No contact with the patient.

Principle of Operation:

The device acquires and digitizes signals from sensors and sends them to a polysomnography system. Sensors are:

  • Direct measurement of electrical potential at the skin (EEG, EMG, EOG).
  • Thermistors (flow)
  • Accelerometers (movement)
  • Respiratory Effort ●
  • Snoring Sensor ●

{5}------------------------------------------------

Indications for Use:

The DormoTech Vlab is a physiological data recorder intended to collect and record-data from multiple physiological channels for use by clinical software used in polysomnography and sleep disorder studies. It is intended for use by or on the order of a physician is intended for use on adults in a supervised (hospital) or unsupervised (home) environment.

Patient Population:

Adults, 22 years and older

Substantial Equivalence:

The subject device, DormoTech Vlab, is substantially equivalent to the Respironics Alice PDx cleared under K090484 when compared to a cleared PSG system, the NOX Sleep system, K192496.

Performance Testing:

Non-clinical

Substantial equivalence to the predicate is based upon performance testing on collecting, recording and transferring data as well as the following.

  • Biocompatibility
  • Software Verification and Validation
  • Electrical Safety
  • Electromagnetic Compatibility Compliance
  • . Product Requirements Verification
  • . Consumables Verification

Clinical

We performed a prospective clinical study compared to a gold standard PSG study.

Summarizing the clinical study:

Study design:

An IEC approved, comparative, self-controlled, randomized, prospective study designed to assess the Vlab and compare its performance to a gold standard polysomnogram (PSG) conducted over 1 night in a sleep lab.

Study Locations:

Two sleep labs were used:

  • Shamir Medical Center – Be'er Ya'akov, Israel
  • . Millenium Sleep Clinic - Be'er Sheva, Israel

Patient Population:

  • Adults 22 years and older ●
  • Male and female .
  • 47 subjects .

The participants were patients who were referred by physicians for a full sleep study at the sleep laboratory.

Primary Endpoints Measures:

  • AHI Apnea-Hypopnea Index Score and Classification ●

Secondary Endpoints Measures:

  • ODI Oxygen Desaturation as measured by separate SpO2 sensor ●
  • Snore-Total Snore % ●
  • Sleep - Total Sleep Time (min), Sleep Efficiency (%), Sleep Stages [Wake, N1, N2, N3, REM] (%), Sleep Latency (min), Wake After Sleep Onset (min), REM Latency (min)
  • . Position: Supine position, Left, Right, Up [% from TST]

{6}------------------------------------------------

510(k) Summary Page 3 of 8

ParameterMean Difference(Lower CI, UpperCI)Upper Limit ofAgreement (Lower CI,Upper CI)Lower Limit ofAgreement(LowerUpper CI)
AHI(events/h)-0.1927(-1.323, 0.9372)6.823(4.866, 8.78)-7.209(-9.166, -5.252)
ODI(events/h)-0.3244(-1.108, 0.4597)4.544(3.186, 5.902)-5.193(-6.551, -3.835)
Snore (%)1.085(-0.525, 2.523)10.01(7.524, 12.5)-7.843(-10.33, -5.353)
Sleep Latency (Minutes)4.653(-0.9411, 10.25)38.01(28.32, 47.7)-28.7(-38.39, -19.01)
REM Latency-15.64(-25.95, -5.327)44.98(27.12, 62.85)-76.27(-94.13, -58.4)
Wake after Sleep Onset(Minutes)-4.300(-10.53, 1.926)31.77(20.98, -42.55)-40.37(-51.15, -29.58)
REM (%)0.4816(-0.801, 1.764)8.129(5.908, 10.35)-7.166(-9.388, -4.945)
N1 (%)0.3263(-1.839, 2.492)13.24(9.488, 16.99)-12.59(-16.34, -8.836)
N2 (%)-2.484(-5.084, 0.1152)13.02(8.513, 17.52)-17.98(-22.49, -13.48)
N3 (%)1.011(-0.07236, 2.093)7.468(5.592, 9.343)-5.447(-7.322, -3.571)
Wake (%)0.1972(-1.301, 1.696)8.877(6.282, 11.47)-8.483(-11.08, -5.887)
Total Sleep Time (Minutes)0.72222(-6.869, 8.313)44.69(31.55, 57.84)-43.25(-56.4, -30.1)
Sleep Efficiency (%)-0.03333(-1.536, 1.47)8.673(6.07, 11.28)-8.74(-11.34, -6.136)
Position (Up) (%)0.01316(-0.4649, 0.4913)2.864(2.036, 3.692)-2.838(-3.666, -2.01)
Position (Supine) (%)0.9974(-0.3433, 2.338)8.991(6.669, 11.31)-6.997(-9.319, -4.675)
Position (Left) (%)0.3579(-0.9967, 1.712)8.435(6.089, 10.78)-7.719(-10.07, -5.373)
Position (Right) (%)-0.3974(-1.61, 0.8149)6.831(4.732, 8.931)-7.626(-9.726, -5.526)

a d

Conclusion:

The table above provides a comprehensive comparison between two devices across multiple sleep-related parameters. Most parameters show good agreement between the devices, as indicated by the mean difference values close to zero and narrow limits of agreement. This suggests that for many measures, such as AHi, ODI, and Snore, the devices are interchangeable. By considering both the statistical measures and the role of human scoring, this table offers a nuanced understanding of the performance and reliability of the Dormotech Vlab device when compared to the gold-standard PSG.

{7}------------------------------------------------

510(k) Summary Page 4 of 8

Table 1 – Comparison of Predicate

FeaturesSubject DevicePredicateComments
DormoTech VlabAlice PDx - K090484
ClassificationGWL, MNRPhysiological Signal Amplifier882.1835GWL, MNR, DQAPhysiological Signal Amplifier882.1835
Indications for useThe DormoTech Vlab is a physiological datarecorder intended to collect and record-data frommultiple physiological channels for use by clinicalsoftware used in polysomnography and sleepdisorder studies. It is intended for use by or on theorder of a physician. It is intended for use on adultsin a supervised (hospital) or unsupervised (home)environment.The Alice PDx is a physiological data recorderintended to collect and record data from multiplephysiological channels for use by clinicalsoftware used in polysomnography and sleepdisorder studies. It is intended for use by or onthe order of a physician. It is intended for use onadults in a supervised (hospital) or unsupervised(home) environment.Similar
PopulationAdultsAdultsSimilar
Environment of UseSupervised (hospital) or unsupervised (home)environmentSupervised (hospital) or unsupervised (home)environmentSimilar
Principle of OperationUses sensors to collect signals to be processed bycleared analysis softwareUses sensors to collect signals to be processedby cleared analysis softwareSimilar
Physiological signalscollectedEEG, EOG, EMG, ECGAirflowSnoreThoracic and Abdominal EffortBody PositionRequires separate SpO2 sensor and Heart (pulse)Rate from PPGEEG, EOG, EMG, ECGAirflowSnoreThoracic and Abdominal EffortBody PositionRequires separate SpO2 sensor and Heart (pulse)Rate from PPGSimilar
Device isWearable and portableWearable and portableSimilar
PrescriptiveYesYesSimilar
Non-clinical TestingRespiratory SignalsBench testing was performedagainst the reference whichconfirmed the signal output.Then all measured parameterswere compared in clinical study toNOX Sleep System - K192469
Total Airflow
Snore
Respiratory Effort (Thoracic & Abdominal)
FeaturesSubject DevicePredicateComments
DormoTech VlabAlice PDx - K090484
EXG Signals
EEG
EOG
EMG
Position Signals
Body Position
Head Position
From recommended Nonin SpO2 sensorFrom recommended Nonin SpO2 sensor
BiocompatibilitySurface contact and Externally communicataingLimited duration < 24 hoursSurface contact and Externally communicataingLimited duration < 24 hoursSimilar

{8}------------------------------------------------

K230148

510(k) Summary Page 5 of 8

{9}------------------------------------------------

510(k) Summary Page 6 of 8

FeaturesSubject DeviceDormoTech VlabReferenceNox Sleep System – K192469Comments
ClassificationGWL, MNRPhysiological Signal Amplifier882.1835OLZ, KZMElectroencephalograph882.1400This system includes morehardware than just sensorscollecting data. It is considered agold standard for comparativeclinical studies.
Indications for useThe DormoTech Vlab is a physiological datarecorder intended to collect and record data frommultiple physiological channels, for use byclinical software used in polysomnography andsleep disorder studies. It is intended for use by oron the order of a physician is intended for use onadults in a supervised (hospital) or unsupervised(home) environment.The Nox Sleep System is used as an aid in thediagnosis of different sleep disorders and for theassessment of sleep. The Nox Sleep System isused to measure, record, display, organize,analyze, summarize, and retrieve physiologicalparameters during sleep and wake. The Nox SleepSystem allows the user to decide on thecomplexity of the study by varying the number andtypes of physiological signals measured. The NoxSleep System allows for generation of user/pre-defined reports based on subject's data. The userof the Nox Sleep System are medical professionalswho have received training in the areas ofhospital/clinical procedures, physiologicalmonitoring of human subjects, or sleep disorderinvestigation. The intended environments arehospitals, institutions, sleep centers, sleep clinics,or other test environments, includingpatient's home.The intended use of collecting dataand the population and useenvironments are similar. Thesubject device does not analyzethe signals.
PopulationAdults, 22 years and olderAdultsSimilar
Environment of UseSupervised (hospital) or unsupervised (home)environmentHospital) and homeSimilar
Principle of OperationUses sensors to collect signals to be processed bycleared analysis softwareUses sensors to collect signals to be processed bycleared analysis softwareSimilar

{10}------------------------------------------------

510(k) Summary Page 7 of 8

FeaturesSubject DeviceDormoTech VlabReferenceNox Sleep System – K192469Comments
Physiological signalscollectedEEG, EOG, EMG, ECGAirflowSnoreThoracic and Abdominal EffortBody PositionPulse Rate (PPG) with a separate SpO2 sensorEEG, EOG, EMG, ECGAirflow/PressureSnoreThoracic and Abdominal EffortBody PositionPulse Rate (PPG) with a separate SpO2 sensorThe subject device includessensors which are attached and thedata collected and analyzed as partof a PSG sleep study.
Device isWearable and portableWearable and portableSimilar, yes the sensors for thereference are also wearable andportable.
PrescriptiveYesYesSimilar
Performance TestingRespiratory SignalsTotal AirflowSnoreRespiratory Effort (Thoracic & Abdominal)EXG SignalsEEGEOGEMGPosition SignalsBody PositionHead PositionPlethysmography signalFrom recommended Nonin SpO2 sensorRespiratory SignalsTotal AirflowSnoreRespiratory Effort (Thoracic & Abdominal)EXG SignalsEEGEOGEMGPosition SignalsBody PositionHead PositionPlethysmography signalFrom recommended Nonin SpO2 sensorIn a comparative clinical study theresults demonstrated that the Vlabwas substantially equivalent inmeasuring the specifiedparameters.

{11}------------------------------------------------

Discussion of Differences:

The subject and predicate device are sensor arrays for use in collecting and transmitting data from sleep studies that is analyzed by automated, FDA-cleared software, which is not part of this submission.

The reference, NOX Sleep system. K192469, is a complete system that includes the sensors and software to analyze the collected data. It includes additional sensors and measured parameters which the subject device does not include, however, what is measured is similar to the predicate device.

We chose the reference as we needed to demonstrate that the signal acquisition quality for all parameters were substantially equivalent and in order to do that we had to process the collected signals through FDAcleared analytic software for comparison.

The design of the sensor array is similar to both the predicate and reference as they are placed on the head and thoracic.

The technology of the sensors is similar to both the predicate and reference.

The differences do not raise different concerns or risk for safety or effectiveness when compared to both the predicate and reference.

Substantial Equivalence Discussion:

As presented in the tables above, the subject device is substantially equivalent to the predicate and the reference devices for indications for use, technological characteristics, environments of use, population and performance.

Conclusion:

The comparison of features, non-clinical and clinical testing support substantial equivalence to the predicate, Respironics Alice PDx, K090484.

§ 882.1835 Physiological signal amplifier.

(a)
Identification. A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).(b)
Classification. Class II (performance standards).