K Number
K230134

Validate with FDA (Live)

Date Cleared
2023-05-23

(126 days)

Product Code
Regulation Number
890.5500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Laser Therapy Hair Growth Comb is indicated to treat Androgenetic Alopecia, and promote hair growth in females who have Ludwig (Savin) I-4, II-1, II-2, or frontal patterns of hair loss and in males who have Norwood Hamilton Classifications of IIa to V and who both have Fitzpatrick Skin Types I to IV.

Device Description

Laser Therapy Hair Growth Comb is a hand-held comb-shaped low level laser therapy device that emits laser light designed to promote hair growth in women and men. The device provides distributed laser light to the scalp while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp. Lasercomb-001 contains 7 laser diodes, and Lasercomb-002 contains 9 laser diodes.

AI/ML Overview

Here's an analysis of the provided FDA 510(k) summary, specifically addressing the acceptance criteria and supporting studies for the Laser Therapy Hair Growth Comb.

It's important to note that this document is a 510(k) summary for a medical device and typically focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving independent effectiveness through a novel clinical trial with pre-defined acceptance criteria for performance metrics like sensitivity or specificity. The "acceptance criteria" in this context refer to demonstrating equivalence in safety and fundamental operating characteristics.

1. A table of acceptance criteria and the reported device performance

The acceptance criteria for this 510(k) submission are not explicitly stated as numerical performance targets (e.g., "sensitivity must be >X%"). Instead, the acceptance criteria are implicitly met by demonstrating substantial equivalence to the predicate devices in terms of intended use, technological characteristics, and safety/performance standards. The reported device performance is primarily confirmed through adherence to recognized electrical safety, EMC, and biocompatibility standards, and by matching the technical specifications of the predicate devices.

Acceptance Criteria (Implicit for 510(k) Substantial Equivalence)Reported Device Performance
Intended Use Equivalence:The Laser Therapy Hair Growth Comb (models Lasercomb-001 & Lasercomb-002) shares the same intended use: to treat Androgenetic Alopecia and promote hair growth in specific populations (females with Ludwig (Savin) I-4, II-1, II-2, or frontal patterns of hair loss; males with Norwood Hamilton Classifications of IIa to V; both with Fitzpatrick Skin Types I to IV) as the predicate devices (HairMax LaserComb Advanced 7 - K103368 & K112524, and HairMax LaserComb Lux 9 - K110233).
Technological Characteristics Equivalence:The device's technological characteristics (Type of laser, Wavelength, Amount of laser diodes, Energy per laser diode, Classification according to IEC60825-1, Treatment time, Treatment frequency, Applicable people, Applicable skin, Shape design) are either identical to or within acceptable variations of the predicate devices. For instance, both the subject device and predicates use visible red light-emitting diodes at 650nm ± 10nm. The number of diodes (7 for -001, 9 for -002) matches the respective predicate models.
Safety and EMC Performance:Complies with IEC 60601-1 (medical electrical equipment general safety), IEC 60601-1-11 (home healthcare environment), IEC 60601-1-2 (electromagnetic compatibility), and IEC 60825-1 (laser safety). The battery conforms to IEC 62133. These are the same standards met by the predicate devices.
Biocompatibility:All body-contacting materials comply with ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation and skin sensitization), and ISO 10993-23 (irritation). These are the same standards met by the predicate devices.
Clinical Performance Equivalence:The submission concludes that "Laser Therapy Hair Growth Comb was found to have a safety and effectiveness profile that is same as the predicate device," based on the comparative analysis of intended use, design, materials, and performance (safety data). This implies that since the predicate device was shown to be effective, and the subject device is substantially equivalent, its effectiveness is inferred.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not describe any clinical test set or patient data (retrospective or prospective) used to directly assess the performance of the subject device the "Laser Therapy Hair Growth Comb" for its primary intended use (treating alopecia and promoting hair growth).

The performance data provided (biocompatibility, electrical/EMC safety) are results from laboratory testing of the device components and system, not clinical trials on patients. The "sample size" for these tests would refer to the number of device units or material samples tested, not a patient population.

The clinical effectiveness claimed for the subject device is established by its substantial equivalence to predicate devices that presumably had clinical data supporting their original clearance. This document does not detail the clinical studies of the predicate devices. Therefore, there's no information on a specific test set, sample size, or data provenance relevant to the subject device's clinical performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as no clinical test set for performance (e.g., hair growth assessment) of the subject device is described in this 510(k) summary. The "ground truth" for the technical safety and biocompatibility tests is established by recognized international standards (IEC, ISO).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable as no clinical test set requiring expert adjudication for a ground truth is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is a "Laser Therapy Hair Growth Comb," which is a direct treatment device, not an AI-powered diagnostic or assistive tool for human interpretation. Therefore, MRMC studies and "human readers improving with AI assistance" are irrelevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical therapeutic product, not a software algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the technical and safety performance data provided:

  • Safety (Electrical & EMC): Ground truth is defined by compliance with specified international standards (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60825-1, IEC 62133). Testing against these standards determines compliance.
  • Biocompatibility: Ground truth is defined by compliance with specified international standards (ISO 10993-5, ISO 10993-10, ISO 10993-23). Laboratory tests are conducted to assess cell viability, irritation, and sensitization effects.

For the clinical effectiveness as argued for substantial equivalence, the ground truth would have been established by clinical studies of the predicate devices (HairMax LaserComb models), likely involving:

  • Outcomes data: Objective measurements of hair count, hair density, hair thickness, or photographic assessments over a specified treatment period.
  • Expert assessment: Dermatologists or trichologists assessing the degree of hair loss and improvement.
  • (Potentially) Patient reported outcomes: Surveys or questionnaires on perceived hair growth and satisfaction.

However, details about these ground truth methods for the predicate devices are not provided in this specific summary.

8. The sample size for the training set

This information is not applicable. The device is a physical product, not an AI algorithm that requires a training set in the machine learning sense. The manufacturing processes and quality control systems might involve historical data, but not a "training set" for AI.

9. How the ground truth for the training set was established

This information is not applicable as there is no "training set" for the device in the context of AI/machine learning.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health and Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 12, 2023

Hubei YJT Technology Co.,Ltd % Gamma Zhang RA Manager Tacro Guangzhou Branch Rm. 501, No.55 West Tiyu Rd., Tianhe Dist., Guangzhou Guangdong Guangzhou, Guangdong 510000 China

Re: K230134

Trade/Device Name: Laser Therapy Hair Growth Comb, Model: Lasercomb-001 & Lasercomb-002 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP

Dear Gamma Zhang:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 23,2023. Specifically, FDA is updating this SE Letter as an administrative correction of incorrect trade name and model name in the 510(k) summary.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jianting Wang, OHT4: Office of Surgical and Infection Control Devices, Acting Assistant Director, jianting.wang@fda.hhs.gov

Sincerely,

Image /page/0/Picture/9 description: The image shows the name "Jianting Wang -S" in a large, sans-serif font. The text is black and appears to be the main subject of the image. The background is a light blue color, which provides contrast to the black text.

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, featuring a stylized graphic of a human figure. To the right is the FDA logo, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 23, 2023

Hubei YJT Technology Co.,Ltd % Gamma Zhang RA Manager Tacro Guangzhou Branch Rm. 501, No.55 West Tiyu Rd., Tianhe Dist., Guangzhou Guangdong Guangzhou, Guangdong 510000 China

Re: K230134

Trade/Device Name: Laser Therapy Hair Growth Comb, Model: Lasercomb-001 & Lasercomb-002 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: April 7, 2023 Received: April 7, 2023

Dear Gamma Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang -S

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K230134

Device Name Laser Therapy Hair Growth Comb

Indications for Use (Describe)

Laser Therapy Hair Growth Comb is indicated to treat Androgenetic Alopecia, and promote hair growth in females who have Ludwig (Savin) I-4, II-1, II-2, or frontal patterns of hair loss and in males who have Norwood Hamilton Classifications of IIa to V and who both have Fitzpatrick Skin Types I to IV.

Type of Use (Select one or both, as applicable)

Research involving OFF-Label Substance Use
Investigational use of OFF-Label Substance

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

K230134

"510(k) Summary" as required by 21 CFR Part 807.92.

Date: 2023-05-23

I. Submitter

Hubei YJT Technology Co.,Ltd. Room1-4 , 8F,Block7,Guannan Fuxing Pharmaceutical Park, No.62 Optical Valley Ave, East Lake High-tech Development Zone, Wuhan, Hubei Province, China Tel.: 027-87771565 Hua Xiang (RA engineer) Tel: +86 17364042492 Email: 528149645@qq.com

II. Correspondent Consultant

Tacro Guangzhou Branch Rm. 501, No.55 West Tiyu Rd., Tianhe Dist., Guangzhou, Guangdong Gamma Zhang(RA manager) Tel: +86 13433933949 Email: wbzhang@tacro.cn

III. Device

Type of 510(k): Traditional Common Name: Laser Therapy Hair Growth Comb Trade Name: Laser Therapy Hair Growth Comb Model: Lasercomb-001 & Lasercomb-002 Classification Name: Laser, Comb, Hair Regulation Number: 21 CFR 890.5500 Review Panel: General & Plastic Surgery Regulatory Class: II Product Code: OAP

IV. Predicate Device

ApplicantPredicate Device510(k) Number
Lexington International, LLCHairMax LaserComb Advanced 7, HairMax LaserComb Lux 9, Hairmax lasercomb professional 12K103368

{5}------------------------------------------------

510(k)s –Section 6. 510(k) Summary

Lexington International,LLCHairMax LaserComb (Model:HairMax Advanced 7)K112524
Lexington International,LLCHairMax LaserComb (Model:HairMax Lux 9)K110233

V. Device Description

Laser Therapy Hair Growth Comb is a hand-held comb-shaped low level laser therapy device that emits laser light designed to promote hair growth in women and men. The device provides distributed laser light to the scalp while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp. Lasercomb-001 contains 7 laser diodes, and Lasercomb-002 contains 9 laser diodes.

VI. Indications for Use

Laser Therapy Hair Growth Comb is indicated to treat Androgenetic Alopecia, and promote hair growth in females who have Ludwig (Savin) 1-4, II-2, or frontal patterns of hair loss and in males who have Norwood Hamilton Classifications of IIa to V and who both have Fitzpatrick Skin Types I to IV..

{6}------------------------------------------------

VII.Comparison of Technological Characteristics With the Predicate Device

Laser Therapy Hair Growth Comb raises no safety or efficacy concerns when compared to the predicate devices.

A technical comparison to the predicate is provided below:

ComparisonElementsSubject DevicePredicate Device IPredicate Device II
K NumberK230134K103368K112524
Trade nameLaser Therapy HairGrowth CombHairMax LaserComb Advanced7HairMax LaserComb Advanced7
ModelLasercomb-001HairMax LaserComb Advanced7HairMax LaserComb Advanced7
ClassificationnameInfrared LampInfrared LampInfrared Lamp
Product codeOAPOAPOAP
Intendeduse/Indicationsfor UseLaser Therapy HairGrowth Comb is indicatedto treat AndrogeneticAlopecia, and promotehair growth in femaleswho have Ludwig (Savin)I-4, II-1, II-2, or frontalpatterns of hair loss and inmales who have NorwoodHamilton Classificationsof Ila to V and who bothhave Fitzpatrick SkinTypes I to IV.The HairMax LaserCombAdvanced 7, Lux9, and theProfessional 12 models areindicated to treat androgeneticalopecia, promote hair growthand help prevent further hairloss in males who haveNorwood HamiltonClassifications of Ila to V andwho both have Fitzpatrick SkinTypes I to IV.The HairMax LaserCombAdvanced 7 is indicated to treatandrogenetic alopeciaand promote hair growth infemales who have Ludwig(Savin) Scale I-4, II-1, II-2, orfrontal and Fitzpatrick SkinTypes I to IV.
Location for useOTC applicationOTC applicationOTC application
Type of laserVisible red light-emittingdiodesVisible red light-emitting diodesVisible red light-emitting diodes
Wavelength650nm ± 10nm650nm ± 10nm650nm ± 10nm
Amount of laserdiodes777
Energy of perlaser diode4.63 mW, 4.56 mW, 4.66mW, 4.78 mW, 4.86 mW,4.78 mW, 4.89 mW <5mW4.71mW, 4.68mW, 4.77mW,4.73mW, 4.89mW, 4.88mW,4.90mW < 5mW4.71mW, 4.68mW, 4.77mW,4.73mW, 4.89mW, 4.88mW,4.90mW < 5mW
Classificationaccording toIEC60825-1Class 3RClass 3RClass 3R
Treatment time15 min per treatment15 min per treatment15 min per treatment
ComparisonElementsSubject DevicePredicate Device IPredicate Device II
Treatmentfrequency3 times per week, spacedout every other day, aslittle as 16 weeks3 times per week, spaced outevery other day, as little as 16weeks3 times per week, spaced outevery other day, as little as 16weeks
ApplicablepeopleNorwood-HamiltonIIaV (males)Ludwig-Savin III(females)Norwood-Hamilton IIa~V(males)Ludwig-Savin I~II(females)
Applicable skinFitzpatrick SkinPhototypes I-IVFitzpatrick Skin PhototypesI-IVFitzpatrick Skin PhototypesI-IV
Shape designCombCombComb
Safety featureComplied withIEC60601-1,IEC60601-1-11,IEC60601-1-2and IEC60825-1Complied withIEC 62133(Battery pack)Complied withIEC60601-1,IEC60601-1-11,IEC60601-1-2 andIEC60825-1Complied withIEC62133 (Batterypack)Complied withIEC60601-1,IEC60601-1-11,IEC60601-1-2 andIEC60825-1Complied withIEC62133 (Batterypack)
BiocompatibilityAll body-contactingmaterials arecomplied withISO10993-5 andISO 10993-10,ISO 10993-23.All body-contactingmaterials arecomplied withISO10993-5 and ISO10993-10, ISO10993-23.All body-contactingmaterials are compliedwith ISO10993-5 andISO 10993-10, ISO10993-23.

For Model: Lasercomb-001

{7}------------------------------------------------

For Model: Lasercomb-002

ComparisonElementsSubject DevicePredicate Device IPredicate Device II
K NumberK230134K103368K110233
Trade nameLaser Therapy HairGrowth CombHairMax LaserComb Lux 9HairMax LaserComb Lux 9
ModelLasercomb-002HairMax LaserComb Lux 9HairMax LaserComb Lux 9
ComparisonElementsﺭﺍ ﺭﺍﻳﻮﻧﻼﺭ ﺭەﮬﯩﻴﯩﺴﻰ (). ﺭﺍﻳﻮﻧﻼﺭ ﺭﺍﻳﻮﻧﻠﯩﺮﯨﻨﯩﯔSubject DevicePredicate Device IPredicate Device II
ClassificationnameLaser, Comb, HairLaser, Comb, HairLaser, Comb, Hair
Product codeOAPOAPOAP
Intendeduse/Indicationsfor UseLaser Therapy HairGrowth Comb is indicatedto treat AndrogeneticAlopecia, and promotehair growth in femaleswho have Ludwig (Savin)I-4, II-1, II-2, or frontalpatterns of hair loss and inmales who have NorwoodHamilton Classificationsof IIa to V and who bothhave Fitzpatrick SkinTypes I to IV.The HairMax LaserCombAdvanced 7, Lux9, and theProfessional 12 models areindicated to treat androgeneticalopecia, promote hair growthand help prevent further hairloss in males who haveNorwood HamiltonClassifications of IIa to V andwho both have Fitzpatrick SkinTypes I to IV.The HairMax LaserComb Lux 9is indicated to treat androgeneticalopecia and promote hairgrowth in females who haveLudwig (Savin) Scale I-4, II-1,II-2, or frontal and FitzpatrickSkin Types I to IV.
Location for useOTC applicationOTC applicationOTC application
Type of laserVisible red light-emittingdiodesVisible red light-emitting diodesVisible red light-emitting diodes
Wavelength650nm±10nm650nm±10nm650nm±10nm
Amount of laserdiodes999
Energy of perlaser diode4.59 mW, 4.67 mW, 4.67mW, 4.77 mW, 4.78 mW,4.81 mW, 4.84 mW, 4.86mW,4.91 mW < 5mW4.53mW,4.82mW,4.66mW,4.68mW, 4.87mW, 4.70mW,4.66mW,4.59mW,4.70mW<5mW4.53mW,4.82mW,4.66mW,4.68mW,4.87mW,4.70mW,4.66mW,4.59mW,4.70mW<5mW
Classificationaccording toIEC60825-1Class 3RClass 3RClass 3R
Treatment time11 min per treatment11 min per treatment11 min per treatment
Treatmentfrequency3 times per week, spacedout every other day, aslittle as 16 weeks3 times per week, spaced outevery other day, as little as 16weeks3 times per week, spaced outevery other day, as little as 16weeks
ApplicablepeopleNorwood-HamiltonIIaV (males)Ludwig-Savin III(females)Norwood-Hamilton IIa~V(males)Ludwig-Savin I~II(females)
Applicable skinFitzpatrick SkinPhototypes I-IVFitzpatrick Skin PhototypesI-IVFitzpatrick Skin PhototypesI-IV
Shape designCombCombComb
510(k)s -Section 6. 510(k) Summary
ComparisonElementsSubject DevicePredicate Device IPredicate Device II
Safety featureComplied withIEC60601-1,IEC60601-1-11,IEC60601-1-2and IEC60825-1Complied withIEC62133(Battery pack)Complied withIEC60601-1,IEC60601-1-11,IEC60601-1-2 andIEC60825-1Complied withIEC62133 (Batterypack)Complied withIEC60601-1,IEC60601-1-11,IEC60601-1-2 andIEC60825-1Complied withIEC62133 (Batterypack)
BiocompatibilityAll body-contactingmaterials arecomplied withISO10993-5 andISO 10993-10,ISO 10993-23.All body-contactingmaterials arecomplied withISO10993-5 and ISO10993-10, ISO10993-23.All body-contactingmaterials are compliedwith ISO10993-5 andISO 10993-10, ISO10993-23.

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

VIII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

1)Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of Laser Therapy Hair Growth Comb was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on June 16, 2016", as recognized by FDA. The testing was performed to, and passed, including:

  • ISO 10993-5:2009/(R)2014, Biological Evaluation of Medical Devices –Part 5: Tests for In Vitro Cytotoxicity

  • ISO 10993-10:2010. Biological Evaluation of Medical Devices -Part 10: Tests for Irritation and Skin Sensitization

  • ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices Part 23: Tests for irritation

2)Electrical and EMC Safety

Electrical safety and EMC safety testing was performed to, and passed, the following standards:

  • IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance

  • IEC 60601-1-11 Medical electrical equipment -Part 1-11: General requirements for basic safety and essential performance -Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

  • IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility -Requirements and tests

In addition, testing to IEC 60825-1 certifies the laser system to classification 3R, which is the same as the predicate devices.

The battery conforms to IEC 62133.

Summary

Based on the above performance as documented in this application, Laser Therapy Hair Growth Comb was found to have a safety and effectiveness profile that is same as the predicate device.

VIII. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, Laser Therapy Hair Growth Comb is to be concluded same to its predicate devices.

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.