K Number
K230079

Validate with FDA (Live)

Date Cleared
2023-08-23

(225 days)

Product Code
Regulation Number
878.4460
Age Range
0 - 5
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

These surgical gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with chemotherapy drugs, have low dermatitis potential, and are nonpyrogenic. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugs and Low Dermatitis Potential are sterile and disposable devices. Gloves are made of synthetic polyisoprene rubber, are white in color, and are available in sizes 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, and 9.0. A polyurethane polymer coating is applied to the inner surface of the glove to make donning easy.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device: "Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugs and Low Dermatitis Potential" (K230079). It compares the new device to a predicate device (K190018) and provides evidence of performance through various tests.

Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document presents a comparison of the subject device (K230079) against a predicate device (K190018) and also summarizes non-clinical testing. Below is a compilation of the acceptance criteria (standards/guidance/test criteria) and the reported device performance.

Test/CharacteristicAcceptance Criteria (Standard/Guidance/Criteria)Reported Device Performance (K230079)
Chemotherapy PermeationASTM D6978-05 (2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy DrugsAll drugs with BDT >240 minutes met criteria.Carmustine (BCNU) (3.3 mg/ml): 12.6 minutesThiotepa (10.0 mg/ml): 26.6 minutes (Note: For marketing, the device labels a warning against using with Carmustine and Thiotepa due to "extremely low permeation times".)
Dimensions (Length)ASTM D3577 requirements: Minimum 265mmPass
Dimensions (Palm Width)ASTM D3577 requirements (specific to each size 5.5-9.0, e.g., 5.5: 70±6mm, 9.0: 114±6mm)Pass for all sizes
Dimensions (Thickness - Finger)ASTM D3577 requirements: Minimum 0.10mmPass
Dimensions (Thickness - Palm)ASTM D3577 requirements: Minimum 0.10mmPass
Dimensions (Thickness - Cuff)ASTM D3577 requirements: Minimum 0.10mmPass
Physical PropertiesASTM D3577-19 requirements for tensile strength, ultimate elongation, and stress at 500% elongation before and after accelerated aging for synthetic surgical glovesPass
Freedom from holesASTM D3577-19 and ASTM D5151-06 requirements; AQL 1.5Pass
Powder-FreeASTM D3577-19, ASTM D6124-06 requirements: ≤ 2 mg per glovePass
SterilityANSI/AAMI/ISO 11137-1:2018 requirement of 10^-6 SALPass
ISO In Vitro CytotoxicityISO 10993-5:2009: Non-cytotoxic (as the general acceptance criteria)Under the conditions of the study, the device was found to be cytotoxic. However, "the device extracts did not elicit acute systemic response in the animal model" in the follow-up Acute Systemic Toxicity test (ISO 10993-11). This is a notable finding, indicating the device failed the initial cytotoxicity but was deemed safe due to subsequent testing.
ISO Skin Irritation StudyISO 10993-10:2010: Not an irritantPass
ISO Maximization Sensitization StudyISO 10993-10:2010: Not a sensitizerPass
ISO Acute Systemic Toxicity StudyISO 10993-11:2006: No mortality or evidence of acute systemic toxicityPass (addressed after initial cytotoxicity finding)
Human Skin SensitizationModified Draize-95 Test (per FDA guidance): Not a sensitizer (demonstrates low dermatitis potential)Pass: "The results showed low dermatitis potential for the human subject tested."
Endotoxin StudyLess than 20.0 EU/devicePass: "test results indicate low endotoxin level"
Material Mediated Pyrogenicity StudyISO 10993-11:2017: Meets the material mediated pyrogenicity requirementPass

2. Sample Size Used for the Test Set and Data Provenance:

  • Chemotherapy Drug Permeation (ASTM D6978-05): The document specifies the standard used but does not provide the exact sample size for the test sets (number of gloves or replicates per drug). The data is generated from laboratory testing ("under the conditions of the study"). Data provenance is not explicitly stated but implies laboratory testing for regulatory submission.
  • Dimensions, Physical Properties, Freedom from Holes, Powder-Free, Sterility: These tests refer to ASTM and ISO standards (e.g., ASTM D3577, ASTM D5151, ASTM D6124, ISO 11137-1). These standards typically define the sampling plans (e.g., AQL 1.5 for freedom from holes), but the specific number of units tested for this submission is not directly stated. Data provenance implies manufacturer's testing for regulatory submission.
  • Biocompatibility (ISO 10993 series): These are biological evaluations. The document refers to "the conditions of the study" and "animal model" for acute systemic toxicity but does not provide specific sample sizes (e.g., number of animals used). Data provenance implies laboratory testing for regulatory submission.
  • Human Skin Sensitization (Modified Draize-95 Test): This test was performed on "human subject tested." The sample size (number of human subjects) is not explicitly given, it ambiguously states "the human subject tested." Data provenance is from a clinical test as per FDA guidance.
  • Endotoxin Study & Material Mediated Pyrogenicity Study: The document refers to "the conditions of the study" but does not provide sample sizes. Data provenance implies laboratory testing for regulatory submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This submission is for a physical device (surgical gloves) and its material properties, not an AI/algorithm-driven device requiring expert image interpretation or similar ground truth establishment. Therefore, the concept of "experts" establishing a ground truth in that context is not applicable here. The "ground truth" for the performance tests (e.g., breakthrough time, tensile strength) is established by the validated methodologies of the referenced ASTM and ISO standards.

4. Adjudication Method for the Test Set:

Not applicable in the context of material and physical property testing of a device like surgical gloves. Test results are objective measurements based on defined standard protocols.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not done. This is not an AI device or a diagnostic imaging device where human reader performance is evaluated.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:

Not applicable as this is not an AI/algorithm device. The performance tests evaluate the physical gloves without any "human-in-the-loop" interaction in the testing methodology itself beyond conducting the tests.

7. The Type of Ground Truth Used:

  • Objective Measurement Standards: For most physical and chemical properties (dimensions, strength, freedom from holes, powder, sterility, chemotherapy permeation, endotoxin, pyrogenicity), the "ground truth" is defined by the objective measurement criteria and methodologies specified in the referenced ASTM and ISO standards. The results are quantitative measurements against these predefined thresholds.
  • Biological Response in Models: For biocompatibility tests (cytotoxicity, skin irritation, sensitization, acute systemic toxicity), the "ground truth" is established by observing biological responses in in vitro or in vivo (animal) models as per ISO 10993 standards and the Modified Draize-95 Test for human skin sensitization.

8. The Sample Size for the Training Set:

Not applicable. This is not an AI/machine learning device that requires a training set. The device is a manufactured physical product.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 23, 2023

Ansell Healthcare Products, LLC. Don Cronk Director, Regulatory Affairs for the Americas 2301 Robb Drive Reno, Nevada 89523

Re: K230079

Trade/Device Name: Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugs and Low Dermatitis Potential Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO, LZC, OPJ Dated: July 20, 2023 Received: July 24, 2023

Dear Don Cronk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bifeng Qian -S

BiFeng Qian, M.D., Ph.D Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230079

Device Name

Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugs and Low Dermatitis Potential

Indications for Use (Describe)

These surgical gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with chemotherapy drugs, have low dermatitis potential, and are nonpyrogenic. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Tested chemotherapy drugs are as follows:

Test Chemotherapy drug & ConcentrationBreakthrough Detection Time (Minutes)
Bleomycin (15 mg/ml)>240
Busulfan (6 mg/ml)>240
Carmustine (BCNU) (3.3 mg/ml)12.6
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (Cytoxan) (20.0 mg/ml)>240
Cytarabine (100 mg/ml)>240
Dacarbazine (DTIC) (10.0 mg/ml)>240
Daunorubicin (5 mg/ml)>240
Docetaxel (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (Toposar) (20.0 mg/ml)>240
Fludarabine (25 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Gemcitabine (Gemzar) (38 mg/ml)>240
Idarubicin (1 mg/ml)>240
Ifosfamide (50.0 mg/ml)>240
Irinotecan (20.0 mg/ml)>240
Mechlorethamine HCI (1.0 mg/ml)>240
Melphalan (5 mg/ml)>240
Methotrexate (25 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Mitoxantrone (2.0 mg/ml)>240
Oxaliplatin (2.0 mg/ml)>240
Paclitaxel (Taxol) (6.0 mg/ml)>240
Paraplatin (10 mg/ml)>240
Thiotepa (10.0 mg/ml)26.6
Vincristine Sulfate (1.0 mg/ml)>240
Ellence (2 mg/ml)>240
Rituximab (10 mg/ml)>240

Please note that the following drugs have extremely low permeation times: Carmustine (BCNU) 12.6 minutes and Thiotepa: 26.6 minutes. Warning Do not use with Carmustine and Thiotepa

{3}------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510K Summary

K230079 Submitter: Ansell Healthcare Products LLC. 2301 Robb Drive Reno, NV 89523 USA

Contact Person(s):

Don Cronk Assoc. Director, Regulatory Phone: 775-750-1723 don.cronk@ansell.com

Carson Delaloye Sr. Administrator, Quality Phone: 530-401-8977 carson.delaloye@ansell.com

Date Prepared: August 23, 2023

Name of Device

Trade Names:Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugsand Low Dermatitis Potential
Common Name:Surgeon's Gloves
Classification Name:Non-Powdered Surgeon's Glove
Classification Regulation:21 CFR 878.4460
Device Class: ProductI
Code: ClassificationKGO, LZC, OPJ
Panel: 510k NumberGeneral and Plastic Surgery
Assigned:K230079

Legally Marketed Predicate Device

K190018-Gammex Non-Latex PI White Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugs

Device Description

Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugs and Low Dermatitis Potential are sterile and disposable devices. Gloves are made of synthetic polyisoprene rubber, are white in color, and are available in sizes 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, and 9.0. A polyurethane polymer coating is applied to the inner surface of the glove to make donning easy.

Indication for Use Statement

These surgical gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with chemotherapy drugs, have low dermatitis potential, and are non-pyrogenic. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Please note that the following drugs have extremely low permeation times: Carmustine (BCNU) 12.6 minutes and Thiotepa: 26.6 minutes. Warning Do not use with Carmustine and Thiotepa.

{5}------------------------------------------------

Chemotherapy Drug Permeation

(Minimum breakthrough detection time in minutes) (ASTM D6978-05 2019)

TEST CHEMOTHERAPY DRUG ANDCONCENTRATIONMINIMUMBREAKTHROUGHTIME (Minutes)
Blenoxane (15 mg/ml)>240
Busulfan (6 mg/ml)>240
Carmustine (BCNU) (3.3 mg/ml)12.6
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (Cytoxan) (20.0 mg/ml)>240
Cytarabine (100 mg/ml)>240
Dacarbazine (DTIC) (10.0 mg/ml)>240
Daunorubicin (5 mg/ml)>240
Docetaxel (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (Toposar) (20.0 mg/ml)>240
Fludarabine (25 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Gemcitabine (Gemzar) (38mg/ml)>240
Idarubicin (1 mg/ml)>240
Ifosfamide (50.0 mg/ml)>240
Irinotecan (20.0 mg/ml)>240
Mechlorethamine HCI (1.0 mg/ml)>240
Melphalan (5 mg/ml)>240
Methotrexate (25 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Mitoxantrone (2.0 mg/ml)>240
Oxaliplatin (2.0 mg/ml)>240
Paclitaxel (Taxol) (6.0 mg/ml)>240
Paraplatin (10 mg/ml)>240
Thiotepa (10.0 mg/ml)26.6
Vincristine Sulfate (1.0 mg/ml)>240
Ellence (2 mg/ml)>240
Rituximab (10 mg/ml)>240

{6}------------------------------------------------

Technical Characteristic Comparison Table
Predicate DeviceSubject DeviceComparison
Trade NameGammex Non-Latex PI WhitePolyisoprene Surgical Gloves Tested forUse with Chemotherapy DrugsPolyisoprene Surgical Glove Testedfor Use with Chemotherapy Drugs andLow Dermatitis PotentialDifferent
510(k) NumberK190018K230079Different
SubmitterAnsell Healthcare Products LLCAnsell Healthcare Products LLCSame
Product CodeKGOKGO, LZC, OPJSame
RegulationNumber21 CFR 878.446021 CFR 878.4460Same
RegulationNameNon-powdered Surgeon's gloveNon-powdered Surgeon's gloveSame
Indicationsfor UseGammex Non-Latex PI White Surgical Gloves Testedfor Use with Chemotherapy Drugs are intended to beworn by operating room personnel to protect asurgical wound from contamination. These gloveswere tested for use with Chemotherapy Drugs as perASTM D6978-05 Standard Practice for Assessment ofMedical Gloves to Permeation by ChemotherapyDrugs.These surgical gloves are intended to be worn byoperating room personnel to protect a surgical woundfrom contamination. These gloves were tested for usewith chemotherapy drugs, have low dermatitispotential, and are non-pyrogenic. These gloves weretested for use with Chemotherapy Drugs as per ASTMD6978-05 Standard Practice for Assessment of MedicalGloves to Permeation by Chemotherapy Drugs.Similar
Tested chemotherapy drugs are as follows:TESTCHEMOTHERAPYDRUGANDCONCENTRATIONMINIMUMBREAKTHROUGHDETECTIONTIME (Minutes)TEST CHEMOTHERAPYDRUGANDCONCENTRATIONMINIMUMBREAKTHROUGHTIME (Minutes)
Blenoxane (15mg/ml)>240Blenoxane (15mg/ml)>240
Busulfan (6 mg/ml)>240Busulfan (6 mg/ml)>240
Carmustine(BCNU)(3.3 mg/ml)10.2Carmustine (BCNU)(3.3 mg/ml)12.6
Cisplatin (1.0 mg/ml)>240Cisplatin (1.0 mg/ml)>240
Cyclophosphamide(Cytoxan) (20.0mg/ml)>240Cyclophosphamide(Cytoxan) (20.0mg/ml)>240
Cytarabine (100mg/ml)>240Cytarabine (100mg/ml)>240
Dacarbazine (DTIC)(10.0 mg/ml)>240Dacarbazine (DTIC)(10.0 mg/ml)>240
Daunorubicin (5mg/ml)>240Daunorubicin (5mg/ml)>240
Docetaxel (10.0mg/ml)>240Docetaxel (10.0mg/ml)>240
DoxorubicinHydrochloride (2.0mg/ml)>240DoxorubicinHydrochloride (2.0mg/ml)>240
Etoposide (Toposar)(20.0 mg/ml)>240Etoposide (Toposar)(20.0 mg/ml)>240
Fludarabine (25mg/ml)>240Fludarabine (25mg/ml)>240
Fluorouracil (50.0mg/ml)>240Fluorouracil (50.0mg/ml)>240
Gemcitabine(Gemzar) (38 mg/ml)>240Gemcitabine(Gemzar) (38 mg/ml)>240
Idarubicin (1 mg/ml)>240Idarubicin (1 mg/ml)>240
Ifosfamide (50.0mg/ml)>240Ifosfamide (50.0mg/ml)>240
Irinotecan (20.0mg/ml)>240Irinotecan (20.0mg/ml)>240
Mechlorethamine HCI(1.0 mg/ml)>240Mechlorethamine HCI(1.0 mg/ml)>240
Melphalan (5 mg/ml)>240Melphalan (5 mg/ml)>240
Methotrexate (25mg/ml)>240Methotrexate (25mg/ml)>240
Mitomycin C (0.5mg/ml)>240Mitomycin C (0.5mg/ml)>240
Mitoxantrone (2.0mg/ml)>240Mitoxantrone (2.0mg/ml)>240
Oxaliplatin (2.0mg/ml)>240Oxaliplatin (2.0mg/ml)>240
Paclitaxel (Taxol) (6.0mg/ml)>240Paclitaxel (Taxol) (6.0mg/ml)>240
Paraplatin (10 mg/ml)>240Paraplatin (10 mg/ml)>240
Thiotepa (10.0 mg/ml)11.5Thiotepa (10.0 mg/ml)26.6
Vincristine Sulfate (1.0mg/ml)>240Vincristine Sulfate (1.0mg/ml)>240
Ellence (2 mg/ml)>240Ellence (2 mg/ml)>240
Rituximab (10 mg/ml)>240Rituximab (10 mg/ml)>240
Please note that the following drugs haveextremely low permeation times:Carmustine (BCNU) 10.2 minutes andThiotepa: 11.5 minutes. Warning Do notuse with Carmustine and ThiotepaPlease note that the following drugs haveextremely low permeation times:Carmustine (BCNU) 12.6 minutes andThiotepa: 26.6 minutes. Warning Do notuse with Carmustine and Thiotepa
Prescriptionor Over-TheCounter-UseOver-The-Counter-UseOver-The-Counter-UseSame
MaterialsSynthetic polyisoprene rubberSynthetic polyisoprene rubberSame
CoatingPolyurethane polymer inner coating toaid donningPolyurethane polymer inner coating toaid donningSame
DesignSingle useSingle useSame
Powder-freePowder-freeSame
Hand SpecificHand SpecificSame
Beaded cuffBeaded cuffSame
ColorWhiteWhiteSame
Sizes5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.05.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0Same
Dimensions andphysicalpropertiesMeets ASTM D3577Meets ASTM D3577Same
SterilitySterileSterileSame
SterilizationmethodIrradiationIrradiationSame

{7}------------------------------------------------

{8}------------------------------------------------

SterilityAssurance Level(SAL)10-6 SAL10-6 SALSame
Freedom fromholesMeets ASTM D3577-09(2015)Inspection level/AQL: GI/AQL 1.5Meets ASTM D3577-09(2015)Inspection level/AQL: GI/AQL 1.5Similar
Powder-FreeMeets ASTM D 6124-06. MeetsApplicable Definition for Powder Free; ≤ 2mg per gloveMeets ASTM D 6124-06. MeetsApplicable Definition for Powder Free; ≤ 2 mg per gloveSame
Protein ContentNot ApplicableNot ApplicableSame
In VitroCytotoxicityFailed. Cytotoxicity Grade 4.Failed. Cytotoxicity Grade 4.Same
BiocompatibilityPrimary SkinIrritationISO 10993-10:2010Under the conditions of the study (perISO 10993-10), the device is not anirritantUnder the conditions of the study (perISO 10993-10), the device is not anirritantSame
BiocompatibilityDermalSensitization -ISO 10993-10:2010Under the conditions of the study(per ISO 10993-10), not a sensitizerUnder the conditions of the study(per ISO 10993-10), not a sensitizerSame
BiocompatibilityAcute SystemicToxicity - ISO10993-11: 2006Under the conditions of the study, therewas no mortality or evidence of acutesystemic toxicityUnder the conditions of the study,there was no mortality or evidenceof acute systemic toxicitySame
Material-MediatedPyrogenicityNo pyrogenicity claims madeUnder the conditions of the study,both inner and outer surface isnon-pyrogenicDifferent
BacterialEndotoxinNo endotoxin claimsUnder the conditions of the study,test results indicate low endotoxinlevelDifferent
Low DermatitisPotentialNo low dermatitis potential claims madeUnder the conditions of the study,the test results demonstrated lowdermatitis potential for the subjectglove.Different

{9}------------------------------------------------

The subject device is manufactured from synthetic polyisoprene rubber with polyurethane polymer inner coating to aid donning. The predicate device is manufactured from synthetic polyisoprene rubber with polyurethane polymer inner coating to aid donning.

The subject device meets the applicable requirements for surgeon's gloves with regards to dimensions and sizes, physical properties, freedom from holes, and powder residues, as found in the following standards: ASTM D3577, ASTM D5151 and ASTM D6124. The subject device passes biological reactivity testing for dermal sensitization, acute systemic toxicity, and material-mediated pyrogenicity in accord with the ISO 10993-10 and ISO 10993-11.

{10}------------------------------------------------

Summary of Non-clinical Testing

TechnologicalCharacteristicsPurposeStandard/Test/FDA GuidanceCriteriaResult Summary
DimensionsTo evaluate thedimension of the gloveASTM D3577Meets ASTM D3577 requirementsfor length, width and thicknessPass
LengthMinimum 265mmPass
Palm Width (size)(mm)Pass
5.570±6Pass
6.076±6Pass
6.583±6Pass
7.089±6Pass
7.595±6Pass
8.0102±6Pass
8.5108±6Pass
9.0114±6Pass
ThicknessTo evaluate the thicknessof the glove(mm)Pass
FingerMinimum 0.10Pass
PalmMinimum 0.10Pass
CuffMinimum 0.10Pass
PhysicalPropertiesTo evaluate the tensilestrength, ultimateelongation, and stress at500% elongationASTM D3577-19Meets ASTMD3577-19requirements for tensile strength,ultimate elongation and stress at500% elongation before and afteraccelerated aging for syntheticsurgical glovesPass
Freedom fromholesTo evaluate the presenceof holes in the glovesASTM D3577-19 ASTM D5151-06Meets ASTM D3577-19andASTM D5151-06 requirementsAQL 1.5Pass
Powder-FreeTo evaluate the level ofpowder on the glovesASTM D3577-19 ASTM D6124-06Meets applicable requirement forPowder Free; ≤ 2 mg per glovePass
SterilityTo demonstrate thesterilization performanceANSI/AAMI/ISO 11137-1:2018Meets ANSI/AAMI/ISO 11137-1:2018 requirement of 106 SALPass
TechnologicalCharacteristicsTo EvaluateStandard/Test/FDA GuidanceResult Summary
Biocompatibility:
ISO in vitrocytotoxicityTo evaluatecytotoxicityISO 10993-5:2009Non-cytotoxicUnder the conditions othe study, the devicewas found to becytotoxic and thereforethe device extractswere evaluated by ISO10993-11 – Test forsystemic toxicity. FromAcute Systemic Toxicitydevice extracts, thedevice extracts did notelicit acute systemicresponse in the animalmodel.
ISO Skin IrritationStudyTo demonstrate low skinirritation potentialISO 10993-10:2010Under the conditions of the study,not an irritantPass
ISO MaximizationSensitization StudyTo demonstrate lowsensitization potentialISO 10993-10:2010Under the conditions of the studynot a sensitizerPass
ISO AcuteSystemicToxicity StudyTo demonstrate noacute systemictoxicityISO 10993-11:2006Under the conditions of the study,there was no morality or evidenceof acute systemic toxicityPass
ChemotherapyPermeationStandardTo demonstratechemotherapy drugbarrier performanceASTM D6978 - 05(2019)Under the conditions of the studythe permeation is acceptablefor the drugs tested for use withthe subject glovesPass
Human SkinSensitizationTo demonstrate lowdermatitis potentialModified Draize-95 TestUnder the conditions of the study,not a sensitizer.Pass
Endotoxin StudyTo demonstrate lowendotoxin levelsLess than 20.0 EU/devicePass
Material MediatedPyrogenicity StudyTo demonstrate lowpyrogenicity potentialISO 10993-11:2017Under the conditions of the study,has met the material mediatedpyrogenicity requirementPass

{11}------------------------------------------------

{12}------------------------------------------------

Clinical Summary:

Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugs and Low Dermatitis potential were tested in accordance with Modified Draize-95 Test, per FDA's guidance document "Guidance for Industry and FDA Reviewers/Staff: Premarket Notification [510k] Submissions for Testing Skin Sensitization Chemicals in Natural Rubber Products 1999". Both the inner and the outer surfaces of the subject glove were tested. The results showed low dermatitis potential for the human subject tested.

Conclusion

The conclusions drawn from the clinical and non-clinical tests performed demonstrate that the subject device, Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugs and Low Dermatitis Potential, is as safe, as effective, and performs as well as or better than the predicate device K190018.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).