K Number
K230074

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-07-27

(198 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Rapid Aneurysm Triage and Notification (ANRTN) is a radiological computer-assisted triage and notification software device for analysis of CT images of the head. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and prioritizing studies with suspected saccular aneurysms during routine patient care. Rapid ANRTN uses an artificial intelligence algorithm to analyze images and highlight studies with suspected saccular aneurysms in a standalone application for study list prioritization or triage in parallel to ongoing standard of care. The device generates compressed preview images that are meant for informational purposes only and not intended for diagnostic use. The device does not alter the original medical image and is not intended to be used as a diagnostic device. Analyzed images are available for review through the PACS, email and mobile application. When viewed the images are for informational purposes only and not for diagnostic use. The results of Rapid ANRTN, in conjunction with other clinical information and professional judgment, are to be used to assist with triage/prioritization of saccular aneurysm cases. Radiologists who read the original medical images are responsible for the diagnostic decision. Rapid ANRTN is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis.

Rapid ANRT is limited to detecting saccular aneurysms at least 4mm in diameter in adults.

Device Description

Rapid ANRTN software device is a radiological computer-assisted image processing software device. The Rapid ANRTN device is a CTA processing module which operates within the integrated Rapid Platform to determine the suspicion of head saccular aneurysm(s). The ANRTN software analyzes input CTA images that are provided in DICOM format and provides notification of suspected saccular aneurysm(s) and a non-diagnostic, compressed image for preview. Rapid ANRTN is an AI/ML image processing module which integrates within the Rapid Platform.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the device (Rapid Aneurysm Triage and Notification - Rapid ANRTN) meets these criteria.

Here's the breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

MetricAcceptance Criteria (Product Code QFM Definition)Reported Device Performance
AUC (for overall performance)> 0.95 (for high performance)> 0.95
SensitivityNot explicitly defined as a threshold, but reported as a key metric.0.933
SpecificityNot explicitly defined as a threshold, but reported as a key metric.0.868

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: 266 CTA cases (151 positive for aneurysm, 115 negative).
  • Data Provenance:
    • Country of Origin: Not explicitly stated in the provided text.
    • Retrospective or Prospective: Not explicitly stated, but the mention of cases "obtained from Siemens, GE, Toshiba, and Philips scanners" and "698 (633 training, 65 validation) CTA cases from multiple sites" suggests a retrospective collection of existing imaging data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: 3 experts.
  • Qualifications of Experts: Not explicitly stated beyond "experts." It is typically assumed these are trained medical professionals (e.g., radiologists) with relevant experience, but specific qualifications are not detailed in the provided text.

4. Adjudication method for the test set

  • Adjudication Method: "Ground truth established by 3 experts." This implies a consensus-based approach, but the specific adjudication method (e.g., majority vote, specific tie-breaking rules, or if all 3 had to agree) is not explicitly detailed (e.g., 2+1, 3+1). It likely refers to a consensus reading among the three experts.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly stated or described. The study focused on the standalone performance of the algorithm. The device's intended use is to "assist hospital networks and trained radiologists in workflow triage," implying an assistive role to humans, but the provided data only shows the algorithm's performance, not human performance with and without assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Yes, a standalone performance validation was done. The text explicitly states: "Final device validation included standalone performance validation." and "This performance validation testing demonstrated the Rapid ANRTN device provides accurate representation of key processing parameters under a range of clinically relevant perturbations associated with the intended use of the software."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: Expert consensus. The text states, "ground truth established by 3 experts."

8. The sample size for the training set

  • Training Set Sample Size: 633 CTA cases. (The broader algorithm development dataset included 698 total, split into 633 training and 65 validation cases, with the 266 cases being the final performance validation set).

9. How the ground truth for the training set was established

  • Ground Truth Establishment for Training Set: The text states, "Algorithm development was performed using 698 (633training, 65 validation) CTA cases from multiple sites." While it mentions the cases were "selected [to] covered a wide range of suspected saccular aneurysms," the specific method for establishing ground truth for the training set (e.g., expert review, clinical reports, or a combination) is not explicitly detailed in the provided document. It is implied, but not stated, that a similar expert review process would have been used as for the test set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.

iSchemaView Inc. James Rosa SVP Regulatory and Quality 1120 Washington Ave., Ste 200 GOLDEN, COLORADO 80401

July 27, 2023

Re: K230074

Trade/Device Name: Rapid Aneurysm Triage and Notification Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological Computer Aided Triage And Notification Software Regulatory Class: Class II Product Code: QFM Dated: July 10, 2023 Received: July 11, 2023

Dear James Rosa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230074

Device Name Rapid Aneurysm Triage and Notification

Indications for Use (Describe)

Rapid Aneurysm Triage and Notification (ANRTN) is a radiological computer-assisted triage and notification software device for analysis of CT images of the head. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and prioritizing studies with suspected saccular aneurysms during routine patient care. Rapid ANRTN uses an artificial intelligence algorithm to analyze images and highlight studies with suspected saccular aneurysms in a standalone application for study list prioritization or triage in parallel to ongoing standard of care. The device generates compressed preview images that are meant for informational purposes only and not intended for diagnostic use. The device does not alter the original medical image and is not intended to be used as a diagnostic device. Analyzed images are available for review through the PACS, email and mobile application. When viewed the images are for informational purposes only and not for diagnostic use. The results of Rapid ANRTN, in conjunction with other clinical information and professional judgment, are to be used to assist with triage/prioritization of saccular aneurysm cases. Radiologists who read the original medical images are responsible for the diagnostic decision. Rapid ANRTN is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis.

Rapid ANRT is limited to detecting saccular aneurysms at least 4mm in diameter in adults. Inclusion Criteria:

The module shall detect unruptured saccular brain aneurysms greater than or equal to 4 mm in size within the following anatomical region:

  • · Internal Carotid Artery (ICA): ophthalmic segment to ICA terminus
  • · Middle Cerebral Artery (MCA): M1 segment through M1/M2 bifurcation
  • · Anterior Cerebral Artery (ACA): A1 and A2 segments
  • · Anterior Communicating Artery (ACOM)
  • · Posterior Communicating Artery (PCOM)
  • · Posterior Cerebral Artery (PCA): P1 segment
  • · Vertebral Artery (VA): intracranial portion of the VA through the vertebrobasilar (VB) junction
  • · Basilar Artery (BA)
  • Contraindications/Exclusions/Cautions:
  • · The module shall process only CTA datasets with a patient position of 'headfirst supine'.
  • · The module shall process only a series within CTA datasets comprised only of axial slices.

· The module shall process only CTA datasets with Z FOV (cranio-caudal transverse anatomical coverage) greater than or equal to 90 mm

  • · The module shall process only CTA datasets with in-plane pixel spacing (X & Y resolution): 0.2 1.0 mm
  • · The module shall process only CTA datasets with in-plane X FOV: 160-400mm.
  • · The module shall process only CTA datasets with in-plane Y FOV: 160-400 mm
  • · The module shall process only CTA datasets with Z slice spacing of 0.2 1.25 mm

· The module shall process only CTA data acquired at x-ray tube voltage 80kVp-140kVp, as reported in the input data DICOM header

  • · The module shall process only CTA datasets consisting of continuous order without gaps.
  • · The module shall process only CTA datasets where the images belong to a single series (i.e., identical SeriesInstanceUID DICOM tag value for all images).
  • · The module shall not process CTA datasets where the DICOM tags "ImagePositionPatient" (IPP),

{3}------------------------------------------------

  • · The module shall only process a single series per study, being the largest available that passes the above requirements.
  • · The module shall accept input files within a parent folder named by a unique identifier for the case.
  • · The module shall accept axial-oblique CTA datasets with an oblique plane of < 35 degrees.
  • · Cases with significant motion and/or imaging artifacts.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K230074 510(k) Summary

iSchemaView, Inc.'s Rapid Aneurysm Triage and Notification

This document contains the 510(k) summary for the iSchemaView Rapid Aneurysm Triage and Notification Device (ANRTN). The content of this summary is based on the requirements of 21 CFR Section 807.92(c).

Applicant Name and Address:

Name:iSchemaView, Inc.
Address:1120 Washington St., Suite 200Golden, CO 80401
Official Contact:Jim RosaPhone: (303) 704-3374Email: rosa@ischemaview.com

July 25, 2023 Summary Preparation Date:

Device Name and Classification:

Trade Name:Rapid Aneurysm Triage and Notification
ClassificationName:Radiological computer aided triage andnotification software
Classification:II
Product Code:QFM
Regulation No:21 C.F.R. §892.2080
ClassificationPanel:Radiology Devices

Predicate Devices:

The iSchemaView Rapid ANRTN device is claimed to be substantially equivalent to the following legally marketed predicate devices:

Viz.AI's Aneurysm Software (K213319)

Device Description:

Rapid ANRTN software device is a radiological computer-assisted image processing software device. The Rapid ANRTN device is a CTA processing module which operates within the integrated Rapid Platform to determine the suspicion of head saccular aneurysm(s). The ANRTN software analyzes input CTA images that are provided in DICOM format and provides notification of suspected saccular aneurysm(s) and a non-diagnostic, compressed image for preview. Rapid ANRTN is an AI/ML image processing module which integrates within the Rapid Platform.

Indications for Use:

Rapid Aneurysm Triage and Notification (ANRTN) is a radiological computer-assisted triage and notification software device for analysis of CT images of the head. The device is

{5}------------------------------------------------

K230074 510(k) Summary

intended to assist hospital networks and trained radiologists in workflow triage by flagging and prioritizing studies with suspected saccular aneurysms during routine patient care.

Rapid ANRTN uses an artificial intelligence algorithm to analyze images and highlight studies with suspected saccular aneurysms in a standalone application for study list prioritization or triage in parallel to ongoing standard of care. The device generates compressed preview images that are meant for informational purposes only and not intended for diagnostic use. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

Analyzed images are available for review through the PACS, email and mobile application. When viewed the images are for informational purposes only and not for diagnostic use. The results of Rapid ANRTN, in conjunction with other clinical information and professional judgment, are to be used to assist with triage/prioritization of saccular aneurysm cases. Radiologists who read the original medical images are responsible for the diagnostic decision. Rapid ANRTN is limited to analysis of imaging data and should not be used inlieu of full patient evaluation or relied upon to make or confirm diagnosis.

Rapid ANRT is limited to detecting saccular aneurysms at least 4mm in diameter in adults.

Inclusion Criteria:

The module shall detect unruptured saccular brain aneurysms greater than or equal to 4 mm in size within the following anatomical region:

  • Internal Carotid Artery (ICA): ophthalmic segment to ICA terminus .
  • . Middle Cerebral Artery (MCA): M1 segment through M1/M2 bifurcation
  • Anterior Cerebral Artery (ACA): A1 and A2 segments .
  • Anterior Communicating Artery (ACOM) .
  • Posterior Communicating Artery (PCOM) .
  • Posterior Cerebral Artery (PCA): P1 segment .
  • Vertebral Artery (VA): intracranial portion of the VA through the vertebrobasilar (VB) . junction
  • Basilar Artery (BA) .

Contraindications/Exclusions/Cautions:

  • . The module shall process only CTA datasets with a patient position of 'headfirst supine'.
  • The module shall process only a series within CTA datasets comprised only of axial slices.
  • . The module shall process only CTA datasets with Z FOV (cranio-caudal transverse anatomical coverage) greater than or equal to 90 mm
  • . The module shall process only CTA datasets with in-plane pixel spacing (X & Y resolution): 0.2 – 1.0 mm
  • The module shall process only CTA datasets with in-plane X FOV: 160-400mm.
  • . The module shall process only CTA datasets with in-plane Y FOV: 160-400 mm
  • . The module shall process only CTA datasets with Z slice spacing of 0.2 – 1.25 mm
  • . The module shall process only CTA datasets with data acquired at x-ray tube voltage 80kVp-140kVp, as reported in the input data DICOM header
  • . The module shall process only CTA datasets consisting of input slices of continuous order without gaps.
  • . The module shall process only CTA datasets where the images belong to a single series (i.e., identical SeriesInstanceUID DICOM tag value for all images).

{6}------------------------------------------------

K230074 510(k) Summary

  • The module shall not process CTA datasets where the DICOM tags "ImagePositionPatient" ● (IPP), "ImageOrientationPatient" (IOP), or "PixelSpacing" are missing
  • The module shall only process a single series per study, being the largest available that passes the • above requirements.
  • The module shall accept input files within a parent folder named by a unique identifier for the case.
  • The module shall accept axial-oblique CTA datasets with an oblique plane of ≤ 35 degrees. •
  • Cases with significant motion and/or imaging artifacts.

Technological Characteristics and Substantial Equivalence:

Rapid ANRTN does not raise new questions of safety or effectiveness compared to the previously cleared Viz.AI ANX Software (K213319). There are minor differences in technical characteristics with the predicate device; with the minor changes the clinical use for Rapid ANRTN device is the same with no additional risks. Thus, the Rapid ANRTN device is substantially equivalent.

The following table summarizes and compares data on the Viz.AI ANX Software (K213319) to the Rapid ANRTN device that is the subject of this Traditional 510(k) submission.

ParameterRapid ANRTN (Subject Device)Viz Aneurysm (Predicate Device)
Product CodeQFMQFM
Regulation21 CFR §892.208021 CFR §892.2080
Intended Use/Indications forUseRapid Aneurysm Triage andNotification (ANRTN) is aradiological computer-assistedtriage and notification softwaredevice for analysis of CT imagesof the head. The device isintended to assist hospitalnetworks and trained radiologistsin workflow triage by flagging andprioritizing studies with suspectedsaccular aneurysms during routinepatient care.Rapid ANRTN uses an artificialintelligence algorithm to analyzeimages and highlight studies withsuspected saccular aneurysms in astandalone application for studylist prioritization or triage inparallel to ongoing standard ofcare. The device generatescompressed preview images thatare meant for informationalpurposes only and not intended forViz ANEURYSM (Viz ANX)is a radiological computer-assisted triage and notificationsoftware device for analysis ofCT images of the head. Thedevice is intended to assisthospital networks and trainedradiologists in workflow triageby flagging and prioritizingstudies with suspectedaneurysms during routinepatient care.Viz ANEURYSM uses anartificial intelligence algorithmto analyze images and highlightstudies with suspectedaneurysms in a standaloneapplication for study listprioritization or triage inparallel to ongoing standard ofcare. The device generatescompressed preview imagesthat are meant for informational
ParameterRapid ANRTN (Subject Device)Viz Aneurysm (Predicate Device)
diagnostic use. The device doesnot alter the original medicalimage and is not intended to beused as a diagnostic device.Analyzed images are available forreview through the PACS, emailand mobile application. Whenviewed the images are forinformational purposes only andnot for diagnostic use. The resultsof Rapid ANRTN, in conjunctionwith other clinical information andprofessional judgment, are to beused to assist withtriage/prioritization of saccularaneurysm cases. Radiologists whoread the original medical imagesare responsible for the diagnosticdecision. Rapid ANRT is limitedto analysis of imaging data andshould not be used in-lieu of fullpatient evaluation or relied upon tomake or confirm diagnosis.Rapid ANRT is limited todetecting saccular aneurysms atleast 4mm in diameter in adults.<inclusions and="" limitations<br="">provided above>purposes only and not intendedfor diagnostic use. The devicedoes not alter the originalmedical image and is notintended to be used as adiagnostic device. Analyzedimages are available for reviewthrough the standaloneapplication. When viewedthrough the standaloneapplication the images are forinformational purposes onlyand not for diagnostic use. Theresults of Viz ANEURYSM, inconjunction with other clinicalinformation and professionaljudgment, are to be used toassist with triage/prioritizationof medical images.Radiologists who read theoriginal medical images areresponsible for the diagnosticdecision. Viz ANEURYSM islimited to analysis of imagingdata and should not be used in-lieu of full patient evaluation orrelied upon to make or confirmdiagnosis.Viz ANEURYSM is limited todetecting aneurysms at least4mm in diameter.
AnatomicalRegionHeadHead
IndependentStandard of CareYesYes
Notification/PrioritizationYesYes
Identify patientswith pre-specifiedclinical conditionYesYes
Clinical ConditionAneurysmAneurysm
Intended UserRadiologistRadiologist
ParameterRapid ANRTN (Subject Device)Viz Aneurysm (Predicate Device)
DICOMYesYes
Imaging ModalityCTACTA
Alteration of ImageNoNo
Preview ImagesNon-diagnosticNon-diagnostic

{7}------------------------------------------------

K230074 510(k) Summary

{8}------------------------------------------------

K230074 510(k) Summary

AI/ML Module Development:

Algorithm development was performed using 698 (633training, 65 validation) CTA cases from multiple sites. Cases selected covered a wide range of suspected saccular aneurysms. Cases were obtained from Siemens, GE, Toshiba, and Philips scanners.

Clinical Characteristics:

The primary users of Rapid ANRTN software are medical professionals including radiologists, neuroradiologists, emergency physicians, neurologists, neurosurgeons, etc., who may review CTA scans of the head for indication of aneurysm.

Performance Standards:

Rapid ANRTN has been developed in conformance with the following standards, as applicable:

EN ISO 14971:2019 (R2021)Application of Risk Management to Medical Devices
IEC 62304:2006 (R2015)Medical device software – Software lifecycle processes
IEC 62366:2015 (R2020)Application of Usability Engineering to Medical Devices
NEMA PS 3.1 - 3.20Digital Imaging and Communications in Medicine (DICOM)

Performance Data:

Rapid ANRTN complies with DICOM (Digital Imaging and Communications in Medicine) -Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20.

iSchemaView conducted extensive performance validation testing and software verification and validation testing of the Rapid ANRTN device. Final device validation included standalone performance validation. This performance validation testing demonstrated the Rapid ANRTN device provides accurate representation of key processing parameters under a range of clinically relevant perturbations associated with the intended use of the software. Software performance, validation and verification testing demonstrated that the Rapid ANRTN device met all design requirements and specifications. Final performance validation included 266 (151 pos, 115 neg) CTA cases with ground truth established by 3 experts. The primary endpoint passed with AUC > 0.95 for high performance per the OFM product code definition. Additionally, Sensitivity (0.933) and Specificity(0.868) supported the finding. Age range evaluated was 21-99 with mean 64. Sub-segmented performance by scanners have different performance levels which are affected by sample sizes as seen in the confidence interval spread.

{9}------------------------------------------------

iSchemaView - Traditional 510(k) Rapid Aneurysm Triage and Notification K230074 510(k) Summary

Image /page/9/Figure/1 description: This image is a plot of sensitivity versus 1-specificity. The y-axis is labeled "Sensitivity True Positive" and ranges from 0 to 1.00 in increments of 0.10. The x-axis is labeled "1-Specificity False Positive" and ranges from 0 to 1.00 in increments of 0.10. A black curve is plotted on the graph, and a yellow line is plotted from approximately (0, 0.8) to (0.2, 1.0).

The following performance is seen across the indicated sub analysis:

ManufacturerMeasureNEstimateLower 95% CIUpper 95% CI
GESensitivity370.9730.8620.995
Specificity320.8750.7190.950
PHILIPSSensitivity190.8420.6240.945
Specificity210.8570.6540.950
SIEMENSSensitivity360.8890.7470.956
Specificity330.8180.6560.914
TOSHIBASensitivity420.9760.8770.996
Specificity280.9290.7740.980
GenderMeasureNEstimateLower 95% CIUpper 95% CI
FemaleSensitivity960.9580.8980.984
Specificity530.8680.7520.935
MaleSensitivity360.8610.7130.939
Specificity510.9020.7900.957
ANR SizeMeasureNEstimateLower 95% CIUpper 95% CI
[4, 5)Sensitivity410.9270.8060.975
[5, 7)Sensitivity540.9440.8490.981
[7, 10)Sensitivity280.8930.7280.963
[10, 25)Sensitivity111.0000.7411.000
ANR LocationMeasureNEstimateLower 95% CIUpper 95% CI
ACASensitivity81.0000.6761.000
ACOMSensitivity290.9310.7800.981
Basilar/PCASensitivity180.9440.7420.990
ICASensitivity400.9000.7690.960
MCASensitivity330.9390.8040.983
VA/PICASensitivity61.0000.6101.000

{10}------------------------------------------------

K230074 510(k) Summary

Prescriptive Statement:

Caution: Federal law restricts this device to sale by or on the order of a physician.

Safety & Effectiveness:

Rapid ANRTN has been designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with EN ISO 14971:2019 (risk management). The Rapid ANRTN performance has been validated with case data.

Rapid ANRTN has been developed using the cybersecurity framework defined within the FDA Draft Guidance: Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions

Conclusion:

In conclusion, the iSchemaView Rapid ANRTN software device is substantially equivalent in intended use, technological characteristics, safety and performance characteristics to the legally marketed predicate device, Viz.AI ANX Software (K213319).

N/A