AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SKORE Male Natural Rubber Latex Condoms are used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

Device Description

The subject SKORE Male Natural Rubber Latex Condoms are made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane and used for contraception and for prophylactic purposes to help prevent pregnancy and the transmission of sexually transmitted infections. The subject device in this submission consists of SKORE Smooth, Dotted &Thin Condoms, SKORE Bulbous Condoms, and SKORE Benzocaine Condoms.

AI/ML Overview

The provided text describes the 510(k) summary for SKORE Male Natural Rubber Latex Condoms. It details the device, its intended use, comparison to a predicate device, and performance testing data.

Here's the requested information structured for acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

Test PerformedAcceptance Criteria (Standard)Reported Device Performance
Biocompatibility:ISO 10993-1:2017 (Biological Evaluation of Medical Devices)Demonstrates that the subject devices are biocompatible.
- Acute Systemic ToxicityISO 10993-11:2017Met specifications.
- In vitro CytotoxicityISO 10993-5:2009Met specifications.
- Skin SensitizationISO 10993-10:2021Met specifications.
- Vaginal Irritation TestISO 10993-23:2021Met specifications.
Mechanical Performance TestingISO 4074:2015 (Natural rubber latex male condoms - Requirements and test methods) & ASTM D3492-16 (Standard Specification for Rubber Contraceptives)The subject devices meet the standard specifications.
Air Burst Test Pressure> 1.0 kPa (as per predicate device comparison)> 1.0 kPa (as per predicate device comparison)
Air Burst Test Volume18 dm³ (as per predicate device comparison)18 dm³ (as per predicate device comparison)
Shelf Life5 Years (per ISO 4074:2015 & 21 CFR 801.435)Supports a shelf-life of five years as per the requirements.

2. Sample size used for the test set and the data provenance

  • Sample Size:
    • For Mechanical Performance Testing: "Three lots of subject devices, SKORE Male Natural Rubber Latex Condoms" were tested.
    • For Biocompatibility and Shelf Life: The sample sizes are not explicitly stated, but the tests were performed on "SKORE Male Natural Rubber Latex Condoms."
  • Data Provenance: Not explicitly stated (e.g., country of origin). The testing seems to be conducted by the manufacturer, TTK Healthcare Ltd., which is based in India. The studies are prospective in nature, as they are conducted to demonstrate compliance of the device before market entry.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to the evaluation of condoms. The ground truth for condom performance is established through adherence to recognized international and national standards (e.g., ISO 4074, ASTM D3492) and laboratory testing, rather than expert consensus on interpretive data.

4. Adjudication method for the test set

Not applicable. The tests are objective and quantitative, based on physical and chemical properties measured against established standards. There is no subjective interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a medical product (condom), not an AI-based diagnostic or assistive technology for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm-based device. Performance is determined through physical and biological testing.

7. The type of ground truth used

The ground truth used for this device is based on established international and national standards (ISO 4074:2015, ASTM D3492-16, ISO 10993 series, 21 CFR 801.435) for material properties (natural rubber latex), physical integrity (air burst pressure/volume), dimensions, biocompatibility, and shelf-life.

8. The sample size for the training set

Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 7, 2023

TTK Healthcare Limited Daniel J S Head - Corporate RA/OA [Medical Devices] 6. Cathedral Road Chennai, Tamil Nadu 600086 India

K230049 Re:

Trade/Device Name: SKORE Male Natural Rubber Latex Condoms Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: August 3, 2023 Received: August 8, 2023

Dear Daniel J S:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia -S

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230049

Device Name

SKORE Male Natural Rubber Latex Condoms

Indications for Use (Describe)

SKORE Male Natural Rubber Latex Condoms are used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY - K230049

TTK Healthcare Ltd.

SKORE Male Natural Rubber Latex Condoms

ADMINISTRATIVE INFORMATION

Manufacturer:TTK Healthcare Ltd.
12, K P Natham Road, Thiruvandar Koil,
Puducherry, INDIA 605 102
Telephone: +91 413 2261 401/402
Official Contact:J.S. Daniel
General Manager & Head - Corporate RA/QA
TTK Healthcare Ltd.
No.6, Cathedral Road,
Chennai, Tamilnadu, INDIA – 600 086
Ph: +91 44 4200 8200
Email: daniel.js@ttkhealthcare.com
Date prepared:September 6, 2023

DEVICE NAME AND CLASSIFICATION

SKORE Male Natural Rubber Latex Condoms Trade/Proprietary Name: Common Name: Male Natural Rubber Latex Condom Regulation Number: 21 CFR 884.5300 Regulation Name Condom Regulatory Class: II Product Code: HIS (condom)

PREDICATE DEVICE

K211152: Durex Condom with Benzocaine

The predicate device has not been subject to a design-related recall.

DEVICE DESCRIPTION

The subject SKORE Male Natural Rubber Latex Condoms are made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane and used

{4}------------------------------------------------

for contraception and for prophylactic purposes to help prevent pregnancy and the transmission of sexually transmitted infections. The subject device in this submission consists of

. SKORE Smooth, Dotted &Thin Condoms

SKORE Smooth Condoms -SKORE Blues Vanilla Scented. These condoms are straight walled, smooth, blue colored, teat ended, natural rubber latex having a length of 190 ± 10 mm, width of 53±2mm and a thickness of 0.060 to 0.074 mm, meeting the requirements as specified in ASTM D3492 and ISO 4074. The subject device includes the color blue and flavor vanilla.

SKORE Dotted Condoms -SKORE Warm Belgian Chocolate & SKORE Dots Vanilla Scented; These condoms are straight walled, raised dots, teat ended, natural rubber latex condoms having a length of 190 ± 10 mm, width of 53 ±2 mm and a thickness of 0.060 to 0.080 mm, meeting the requirements as specified in ASTM D3492 and ISO 4074. The subject device includes the color brown and flavors Belgium chocolate and vanilla.

SKORE Thin Condoms - SKORE NOTHING Chocolate, SKORE NOTHING Strawberry & SKORE SKINTHIN Ultra fine Condoms. These condoms are straight walled, smooth, thin, teat ended, natural rubber latex condoms having a length of 190 ± 10 mm. width 53±2mm and a thickness of 0.045 to 0.055 mm. meeting the requirements as specified in ASTM D3492 and ISO 4074. The subject device includes the flavors strawberry and chocolate.

● SKORE Bulbous Condoms

SKORE WARM and SKORE COOL Condoms - These condoms are bulbous dotted, ribbed, and colored natural rubber latex condoms having a length of 190 ± 10 mm, width 56±2mm, meeting the requirements as specified in ASTM D3492 and ISO 4074. The subject device includes the colors orange and blue and flavors warming lubricant and cool tropic mint.

SKORE Benzocaine Condoms .

SKORE NOTHING Climax Delay - These condoms are straight walled, thin, teat ended, natural rubber latex condoms having a length of 190 ± 10 mm, width 53 ±2 mm and a thickness of 0.050 to 0.060, mm meeting the requirements as specified in ASTM D3492 and ISO 4074.

{5}------------------------------------------------

SKORE Duo Max - These condoms are bulbous transparent dotted & ribbed, teat ended, natural rubber latex condoms having a length of 190 ± 10 mm, width of 56±2 mm meeting the requirements as specified in ASTM D3492 and ISO 4074.

SKORE NOT OUT - These condoms are straight walled, raised dots, red colored, teat ended, natural rubber latex condoms having a length of 190 ± 10 mm, width of 53 ±2 mm and a thickness of 0.060 to 0.080 mm meeting the requirements as specified in ASTM D3492 and ISO 4074. The subject device includes the color red.

INDICATIONS FOR USE

SKORE Male Natural Rubber Latex Condoms are used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

PREDICATE DEVICE COMPARISON

The table below compares the intended use and technological characteristics of the subject and predicate device.

Subject Device& PredicateDeviceSKORE Male Natural Rubber LatexCondomsK211152 (Durex Condomswith Benzocaine)
Indications forUseSKORE Male Natural Rubber LatexCondoms are used for contraceptionand for prophylactic purposes (to helpprevent pregnancy and thetransmission of sexually transmitteddiseases).The Durex Condom withBenzocaine is used forcontraception and forprophylactic purposes ( tohelp reduce the risk ofpregnancy and thetransmission of sexuallytransmitted infections,STIs)
Dusting PowderMagnesium Carbonate & CalciumCarbonate-
CondomMaterialNatural Rubber LatexNatural Rubber Latex
Nominal WidthThin$53 \pm 2$ mm
Bulbous$56 \pm 2$ mm
Dotted$53 \pm 2$ mm
Smooth$53 \pm 2$ mm
$56\pm2$ mm
NominalLength190±10 mm195±10 mm
NominalThicknessThin0.045- 0.055mm0.065±0.010 mm
Dotted0.060 - 0.080mm
Smooth0.060- 0.074mm
Nothing ClimaxDelay0.050-0.060mm
NOT OUT0.060-0.080mm
DUO MAX0.075-0.095mm
LubricantSilicone Oil/ Zeus Odor Maskerflavoring, benzocaine cream, orwarming.
LubricantQuantitySmooth, dotted, thin, and vanillacondoms500 ± 50 mg480 ± 50 mg
Belgian chocolate,strawberry,chocolate, andskinthin400 ± 50mg
Warm600 ± 50 mg
Cool tropical mint600 ± 50mg
Benzocainecondoms400 ± 50 mg &480 ± 50 mg
BenzocaineQuantity120- 160 mg (4.75 - 5.25 %)Benzocaine cream / Retardant solution
Air Burst TestPressure> 1.0 kPa> 1.0 kPa
Air Burst TestVolume18 dm318 dm3
TextureSmooth, dotted & thin-
Shelf Life5 Years5 Years
Color AdditivesBlue, Brown, Orange, Red & Blue-
Flavor AdditivesVanilla, Belgian Chocolate, Vanilla, Strawberry Chocolate, Warming Lubricant (Orange), & Cool Tropic Mint flavor (Blue)-
Prescription orOver-The-Counter useOver- The- CounterOver- The- Counter
PackagingMaterialLaminate consisting of a layer of suitable impermeable flexible aluminum foil, and layers of plastic materials suitable for the mechanical protection of the metal foil and or printing and sealing.Laminate consisting of a layer of suitable impermeable flexible aluminum foil, and layers of plastic materials suitable for the mechanical protection of the metal foil and or printing and sealing.

Substantial Equivalence Table:

{6}------------------------------------------------

{7}------------------------------------------------

The subject and predicate device have similar indications for use and have the same intended use. The technological characteristics of the subject device and predicate device are similar in that they are natural rubber latex-based, are lubricated with silicone, have the same shelf-life duration, and for the SKORE Benzocaine Condoms, they have the same amount of benzocaine. The subject and predicate devices have different technological characteristics, including different lubricant quantities, dimensions, and specifications. However, the different technological characteristics of the subject devices do not raise different questions of safety and effectiveness.

PERFORMANCE TESTING

Biocompatibility:

Biocompatibility testing was performed on the SKORE Male Natural Rubber Latex Condoms in accordance with ISO 10993-1: Biological Evaluation of Medical Devices.

Test PerformedStandard
Acute Systemic Toxicity ReactivityISO 10993-11:2017
In vitro CytotoxicityISO 10993-5:2009
Skin SensitizationISO 10993-10:2021
Vaginal Irritation TestISO 10993-23:2021

The results of demonstrate that the subject devices are biocompatible.

{8}------------------------------------------------

Mechanical Performance Testing:

Three lots of subject devices, SKORE Male Natural Rubber Latex Condoms were tested at baseline and met specifications of ISO 4074:2015 - Natural rubber latex male condoms -Requirements and test methods and ASTM D3492-16- Standard Specification for Rubber Contraceptives

The results demonstrate that the subject devices meet the standard specifications.

Shelf Life:

Shelf-life testing was conducted per ISO 4074:201521 and 21 CFR 801.435. The accelerated (7 days at 70°c, 90 days at 40-50°c) and real-time stability (5 Years at 28°C to 35° C) of the SKORE Male Natural Rubber Latex Condoms supports a shelf-life of five years as per the requirements.

CONCLUSION

The results of the performance testing described above demonstrate that the SKORE Male Natural Rubber Latex Condoms are as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.5300 Condom.

(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.