K Number
K223930

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-03-06

(66 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A1417MCW / A1717MCW / F1417MCW (Digital Flat Panel X-Ray Detector) are indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic systems. Not to be used for mammography.

Device Description

The A1717MCW/A1417MCW/F1417MCW is wired or wireless digital flat panel detectors that have been designed for faster, more streamlined approach to digital radiography systems. The A1717MCW/A1417MCW/F1417MCW detector utilize a combination of propriety TFT and scintillator (Csl), and those and electronics are housed in one package. The detectors support an auto-trigger signal sensing technology that allows the detectors to be used without generator integration. The flat panel sensors of the A1717MCW/A1417MCW/F1417MCW/ are fabricated using thin film technology based on amorphous silicon technology. Electronically, the sensors are much like conventional photodiode arrays. Each pixel in the array consists of a light-sensing photodiode and a switching Thin Film Transistor (TFT) in the same electronic circuit. Amorphous silicon photodiodes are sensitive to visible light, with a response curve roughly comparable to human vision. The sensitivity of amorphous sillcon photodiodes peaks in green wavelengths, well matched to scintillators such as Csl. The response has the excellent linearity of a charge-integrating-biased photodiode. SDK-MCW is the software of Detector that performs image acquisition, image correction, and preprocessing.

AI/ML Overview

The provided document is a 510(k) premarket notification for a Digital Flat Panel X-Ray Detector (A1417MCW/A1717MCW/F1417MCW). It does not describe an AI/ML device or its acceptance criteria and study as per your request. Instead, it focuses on demonstrating substantial equivalence to a predicate device (K213497) based on technological characteristics and non-clinical performance tests.

Therefore, I cannot extract the information you requested regarding acceptance criteria, study details for AI/ML device performance, sample sizes, expert qualifications, or adjudication methods from this document.

The document only states that:

  • Indications for use: "A1417MCW / A1717MCW / F1417MCW (Digital Flat Panel X-Ray Detector) are indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic systems. Not to be used for mammography."
  • Non-Clinical Test Summary: The device complies with voluntary standards for electrical safety, electromagnetic compatibility, and its software (SDK-MCW) has a "moderate level of concern" and was validated. Imaging performance tests were conducted according to IEC 62220-1.
  • Clinical Test Summary: "Clinical images were provided; these images were not necessary to establish substantial equivalence based on the differences from the predicate (note TFT technology with Csl scintillator that is identical to the predicate image plate) but they provide further evidence in addition to the laboratory performance data to show that the subject digital detector works as intended."

The document does not include specific acceptance criteria tables, sample sizes for test sets, data provenance, number or qualifications of experts, adjudication methods, MRMC studies, standalone algorithm performance, or details on how ground truth was established for training sets, as it is a submission for a hardware device (X-ray detector) and not an AI/ML diagnostic software.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

March 6, 2023

H&abyz Co., Ltd. % Sanglok Lee Manager Wise Company, Inc. #507, #508, 166 Gasan digital 2-ro Geumcheon-gu, Seoul 08503 KOREA

Re: K223930

Trade/Device Name: A1417MCW/A1717MCW/F1417MCW Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: December 30, 2022 Received: December 30, 2022

Dear Sanglok Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223930

Device Name A1417MCW / A1717MCW / F1417MCW

Indications for Use (Describe)

A1417MCW / A1717MCW / F1417MCW (Digital Flat Panel X-Ray Detector) are indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic systems. Not to be used for mammography.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K223930

[As Required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(a)]

March 3, 2023

2. Submitter's Information [21 CFR 807.92(a)(1)]

  • Name of Manufacturer: H&abyz Co. Ltd. ●
  • Address: 1F, 2-Dong, 41-16 Cheoinseong-Ro, Namsa-Myeon, Cheoin-Gu, Yongin-Si, Gyeonggi-Do, Republic of Korea [17118] ● Contact Name: Namkyu Hur / Quality Manager
  • Telephone No.: +82 070-4658-9300
  • Email Address: hnk@abyzr.com
  • Registration No.: 3016674851

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade/Device/Model NameA1417MCW/A1717MCW/F1417MCW
Common NameDigital Flat Panel X-ray Detector
Device Classification NameStationary X-ray System
Regulation Number21 CFR 892.1680
Classification Product CodeMQB
Device ClassII
510(k) Review PanelRadiology

{4}------------------------------------------------

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[

The identified predicate device within this submission is shown as follow;

Predicate Device

510(k) Number:K213497
Applicant:H&abyz Co., Ltd.
Detector Name:HAD1417MCW, HAD1717MCW
Common Name:Digital Diagnostic X-ray System
Classification Name:Stationary x-ray system
Regulation Number:21 CFR 892.1680
Classification ProductCodeMQB
Device Class:II
  • Radiology 510(k) Review Panel: ●

{5}------------------------------------------------

5. Description of the Device [21 CFR 807.92(a)(4)]

The A1717MCW/A1417MCW/F1417MCW is wired or wireless digital flat panel detectors that have been designed for faster, more streamlined approach to digital radiography systems. The A1717MCW/A1417MCW/F1417MCW detector utilize a combination of propriety TFT and scintillator (Csl), and those and electronics are housed in one package. The detectors support an auto-trigger signal sensing technology that allows the detectors to be used without generator integration.

The flat panel sensors of the A1717MCW/A1417MCW/F1417MCW/ are fabricated using thin film technology based on amorphous silicon technology. Electronically, the sensors are much like conventional photodiode arrays. Each pixel in the array consists of a light-sensing photodiode and a switching Thin Film Transistor (TFT) in the same electronic circuit. Amorphous silicon photodiodes are sensitive to visible light, with a response curve roughly comparable to human vision. The sensitivity of amorphous sillcon photodiodes peaks in green wavelengths, well matched to scintillators such as Csl. The response has the excellent linearity of a charge-integrating-biased photodiode.

SDK-MCW is the software of Detector that performs image acquisition, image correction, and preprocessing. According to the FDA guidance document entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," it has a moderate level of concern. Also, it does not base on previously-cleared software and was originally coded.

6. Indications for use [21 CFR 807.92(a)(5)]

A1417MCW / A1717MCW / F1417MCW (Digital Flat Panel X-Ray Detector) are indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic systems. Not to be used for mammography.

7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]

There are no significant differences in the technological characteristics of these devices compared to the predicate device which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the A1717MCW/A1417MCW/F1417MCW and the predicate device:

Proposed DevicePredicate DeviceNote
K Number-K213497-
ManufacturerH&abyz Co., Ltd.H&abyz Co., Ltd.-
Detector NameA1717MCWHAD1717MCW-
Common NameDigital Flat Panel X-ray DetectorDigital Flat Panel X-ray DetectorSame
Product CodeMQBMQBSame
RegulationNumber21 CFR 892.168021 CFR 892.1680Same
510(k) ReviewPanelRadiologyRadiologySame
Indications forUseA1417MCW / A1717MCW /F1417MCW (Digital Flat PanelX-Ray Detector) are indicated fordigital imaging solution designedfor providing generalradiographic diagnosis of humananatomy targeting both adult andchildren. It is intended to replacefilm based radiographicdiagnostic systems. Not to beused for mammography.ADD (Digital Flat Panel X-RayDetector) is indicated for digitalimaging solution designed forproviding general radiographicdiagnosis of human anatomytargeting both adult and children.It is intended to replace filmbased radiographic diagnosticsystems. Not to be used formammography.Same
[Table 1. Comparison of Proposed Device(A1717MCW) to Predicate Devices]
-----------------------------------------------------------------------------

{6}------------------------------------------------

Proposed DevicePredicate DeviceNote
ScintillatorCslCslSame
Effective PixelArea425.04 x 425.6 mm425.04 x 425.6 mmSame
Total PixelNumber3,072 x 3,072 pixels3,072 x 3,072 pixelsSame
Pixel Pitch140um140umSame
High ContrastLimitingResolution(LP/mm)Max. 3.57Max. 3.5Difference1)
CommunicationWired/WirelessWired/WirelessSame
DQE69% (0.5lp/mm, min.)43.8% (0.5lp/mm, min.)Difference2)
MTF97% (0.1lp/mm, min.)97% (0.1lp/mm, min.)Same
AnatomicalSitesGeneralGeneralSame
Exposure ModeNormal Mode (Manual),AED Mode (Auto ExposureDetection)Normal Mode (Manual),AED Mode (Auto ExposureDetection)Same
WirelessIEEE 802.11a/b/g/n/acIEEE 802.11a/b/g/nDifference3)

[Table 2. Comparison of Proposed Device(A1417MCW) to Predicate Devices]

Proposed DevicePredicate DeviceNote
K Number-K213497-
ManufacturerH&abyz Co., Ltd.H&abyz Co., Ltd.-
Detector NameA1417MCWHAD1417MCW-
Common NameDigital Flat Panel X-ray DetectorDigital Flat Panel X-ray DetectorSame
Product CodeMQBMQBSame
RegulationNumber21 CFR 892.168021 CFR 892.1680Same
510(k) ReviewPanelRadiologyRadiologySame
Indications forUseA1417MCW / A1717MCW /F1417MCW (Digital Flat PanelX-Ray Detector) are indicated fordigital imaging solution designedfor providing generalradiographic diagnosis of humananatomy targeting both adult andchildren. It is intended to replacefilm based radiographicdiagnostic systems. Not to beused for mammography.ADD (Digital Flat Panel X-RayDetector) is indicated for digitalimaging solution designed forproviding general radiographicdiagnosis of human anatomytargeting both adult and children.It is intended to replace filmbased radiographic diagnosticsystems. Not to be used formammography.Same
ScintillatorCslCslSame
Effective PixelArea345.24 x 425.6 mm345.24 x 425.6 mmSame
Total PixelNumber2,560 x 3,072 pixels2,560 x 3,072 pixelsSame

{7}------------------------------------------------

Proposed DevicePredicate DeviceNote
Pixel Pitch140um140umSame
High ContrastLimitingResolution(LP/mm)Max. 3.57Max. 3.5Difference4)
CommunicationWired/WirelessWired/WirelessSame
DQE69% (0.5lp/mm, min.)44.6% (0.5lp/mm, min.)Difference5)
MTF97% (0.1lp/mm, min.)97% (0.1lp/mm, min.)Same
AnatomicalSitesGeneralGeneralSame
Exposure ModeNormal Mode (Manual),AED Mode (Auto ExposureDetection)Normal Mode (Manual),AED Mode (Auto ExposureDetection)Same
WirelessIEEE 802.11a/b/g/n/acIEEE 802.11a/b/g/nDifference6)

[Table 3. Comparison of Proposed Device(F1417MCW) to Predicate Devices]

Proposed DevicePredicate DeviceNote
K Number-K213497-
ManufacturerH&abyz Co., Ltd.H&abyz Co., Ltd.-
Detector NameF1417MCWHAD1417MCW-
Common NameDigital Flat Panel X-ray DetectorDigital Flat Panel X-ray DetectorSame
Product CodeMQBMQBSame
RegulationNumber21 CFR 892.168021 CFR 892.1680Same
510(k) ReviewPanelRadiologyRadiologySame
Indications forUseA1417MCW / A1717MCW /F1417MCW (Digital Flat PanelX-Ray Detector) are indicated fordigital imaging solution designedfor providing generalradiographic diagnosis of humananatomy targeting both adult andchildren. It is intended to replacefilm based radiographicdiagnostic systems. Not to beused for mammography.ADD (Digital Flat Panel X-RayDetector) is indicated for digitalimaging solution designed forproviding general radiographicdiagnosis of human anatomytargeting both adult and children.It is intended to replace filmbased radiographic diagnosticsystems. Not to be used formammography.Same
ScintillatorCslCslSame
Effective PixelArea345.24 x 425.6 mm345.24 x 425.6 mmSame
Total PixelNumber2,560 x 3,072 pixels2,560 x 3,072 pixelsSame
Pixel Pitch140um140umSame
High ContrastLimitingResolutionMax. 3.57Max. 3.5Difference7)

{8}------------------------------------------------

Proposed DevicePredicate DeviceNote
(LP/mm)
CommunicationWired/WirelessWired/WirelessSame
DQE71% (0.5lp/mm, min.)44.6% (0.5lp/mm, min.)Difference8)
MTF98% (0.1lp/mm, min.)97% (0.1lp/mm, min.)Difference9)
AnatomicalSitesGeneralGeneralSame
Exposure ModeNormal Mode (Manual),AED Mode (Auto ExposureDetection)Normal Mode (Manual),AED Mode (Auto ExposureDetection)Same
WirelessIEEE 802.11a/b/g/n/acIEEE 802.11a/b/g/nDifference10)
NoDifferencesExplanation
1, 4, 7High Contrast LimitingResolutionHigh contrast limiting resolution is the value at which thesignal detects lines or holes at fine intervals that differconsiderably from the background. This is the result valuecalculated by the pixel pitch value and does not contributeany adverse impact to the device's safety andeffectiveness.
2, 5, 8DQEThe detective quantum efficiency (DQE) is one of thefundamental physical variables related to image quality inradiography and refers to the efficiency of a detector inconverting incident x-ray energy into an image signal. Thenew detectors are higher than that of predicate devices.These differences do not contribute any adverse impact tothe device's safety and effectiveness.
3, 6 10WirelessThis is related to the wireless standards applied to theproduct. This change does not contribute any adverseimpact to the device's safety and effectiveness.
9MTFThe modulation transfer function (MTF) describes thefrequency behavior of the system and is a curve that haslower values for high frequencies which represent the smallimage structures. The proposed device(F1417MCW) is 98%(0.1lp/mm, min.) and the predicate device is 97%(0.1lp/mm, min.). These differences do not contribute anyadverse impact to the device's safety and effectiveness.

8. Non-Clinical Test summary

The A1717MCW/A1417MCW/F1417MCW complies with voluntary standards for electrical safety, electromagnetic compatibility. The following data were provided in support of the substantial equivalence determination:

1) Electrical Safety, Electromagnetic Compatibility and Performance:

The A1717MCW/A1417MCW/F1417MCW complies with the electrical safety and electromagnetic compatibility requirements established by the standards.

StandardsNo.StandardsOrganizationStandard TitleVersionPublicationYear
ES60601-1AAMI ANSIMedical Electrical Equipment - Part 1: GeneralRequirements for Basic Safety and EssentialPerformance (IEC 60601-1:2005, MOD)ES60601-1:2005(R)2012 andA1:20122014
60601-1-2IECMedical Electrical Equipment - Part 1-2: GeneralRequirements for Safety – Collateral Standard:60601-1-2Edition 4.12020

{9}------------------------------------------------

Electromagnetic Compatibility - Requirementsand Tests2014-02
62220-1-1IECMedical electrical equipment-Characteristics ofdigital X-ray imaging devices Part 1-1:Determination of the detective quantumefficiency Detectors used in radiographicimaging62220-1-1 Edition1.02015-032015

2) Software Validation

The A1717MCW/A1417MCW/F1417MCW contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA quidance:

  • · The content of premarket submissions for software contained in medical devices, on May 11, 2005

3) Biocompatibility

  • . ISO 10993-1 and series, biological evaluation of medical devices

4) Performance Test

Imaging performance test has been conducted according to:

  • . IEC 62220-1, Medical Electrical Equipment – Characteristics of Digital X-ray Imaging Devices – Part
    • 1-1: Determination of the Detective Quantum Efficiency Detectors Used in Radiographic Imaging.

We select the predicate device in order to demonstrate adequate DQE performance of the A1717MCW/A1417MCW/F1417MCW/detector. According to the above comparison table, subject device shows better DQE.

5) Cybersecurity

  • Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, on . October2, 2014

6) Label

  • CFR Part 801 ●
  • . Pediatric Information for X-ray Imaging Device Premarket Notifications, on November 28, 2017

9. Clinical Test Summary

Clinical data has been provided according to FDA guidance document "Guidance for the Submission of 510(k)s for Solid Sate X-ray Imaging Devices". Clinical images were provided; these images were not necessary to establish substantial equivalence based on the differences from the predicate (note TFT technology with Csl scintillator that is identical to the predicate image plate) but they provide further evidence in addition to the laboratory performance data to show that the subject digital detector works as intended

10. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]

There are no significant differences between the A1717MCW/A1417MCW/F1417MCW and the predicate device, K213497 that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.

11. Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, concludes that A1717MCW/A1417MCW/F1417MCW is substantially equivalent in safety and effectiveness to the predicate device as described herein.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.