(170 days)
The device is a general purpose diagnostic ultrasound system intended for use by appropriately trained healthcare professionals in a hospital setting. The device is intended for ultrasound imaging, measurement, display and analysis of the human body and fluid.
This device is indicated for Abdominal; Fetal/Obstetrics; Gynecology;Transvaginal; Urology(including prostate); Transrectal; Cardiac(adult and pediatric); Peripheral Vascular; Small Organs/Parts(thyroid, breast, testicle, Musculo-skeletal Conventional and Superficial); Pediatrics(including neonatal cephalic); and Adult Cephalic diagnostic Ultrasound applications.
The operating modes supported by the device are B mode, Harmonic mode, M mode, Color Flow mode, Power-Doppler mode, Pulsed Wave Doppler mode, Continuous Wave Doppler, 3D mode, 4D mode, Spatio Temporal Image Correlation mode, Tissue Doppler mode, Tissue Velocity Imaging mode, Elastography mode and Contrast Agent Imaging mode.
The VINNO G50、VINNO G55、VINNO M50、VINNO E30、VINNO X5、VINNO X6、VINNO X7、VINNO M55、VINNO E35、VINNO X65、VINNO X55、VINNO X35、VINNO G55E、 VINNO G55P、VINNO M55E、VINNO M55P、VINNO E35E、VINNO E35P ultrasound devices are professional digital color ultrasonic diagnostic apparatus. It transmits ultrasound waves into the body tissues and displays the echo images of the tissues and blood flow accordingly.
The provided text does not contain detailed information about specific acceptance criteria and a study proving a device meets them in the context of AI performance or clinical efficacy. Instead, it describes a 510(k) premarket notification for a general-purpose diagnostic ultrasound system (VINNO G50, etc.) and focuses on demonstrating substantial equivalence to a predicate device (HS40 Diagnostic Ultrasound System).
The "Performance Data" section (VIII) refers to non-clinical testing to support the substantial equivalence determination, rather than clinical performance or AI-specific criteria.
Therefore, I cannot provide the requested information from the given text regarding:
- A table of acceptance criteria and reported device performance (for clinical/AI performance)
- Sample size and provenance for a test set
- Number and qualifications of experts for ground truth
- Adjudication method for the test set
- MRMC comparative effectiveness study results
- Standalone algorithm performance
- Type of ground truth used (for clinical/AI performance)
- Sample size for the training set
- How ground truth for the training set was established
The document primarily addresses electrical safety, EMC, software verification/validation (general to medical device software, not AI-specific), biocompatibility, and acoustic output, comparing these aspects to the predicate device and relevant standards. It concludes that the device is substantially equivalent, implying it is "as safe, as effective and performs as well as the legally marketed device" based on these non-clinical comparisons.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date June 17, 2023. The text is written in a clear, sans-serif font. The date is formatted with the month, day, and year. The text is black against a white background.
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
VINNO Technology (Suzhou) Co., Ltd. % Cordelia Liu Regulatory Registered Engineer 5F Building A, 4F Building C No. 27 Xinfa Rd. Suzhou Industrial Park Suzhou, Jiangsu 215123 CHINA
Re: K223920
Trade/Device Name: VINNO G50, VINNO G55, VINNO M50, VINNO E30, VINNO X5, VINNO X6, VINNO X7, VINNO M55, VINNO E35, VINNO X65, VINNO X55, VINNO X35, VINNO G55E, VINNO G55P, VINNO M55E, VINNO M55P, VINNO E35E. VINNO E35P Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: April 12, 2023 Received: May 17, 2023
Dear Cordelia Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K223920
Device Name
VINNO G50、VINNO G55、VINNO M50、VINNO E30、VINNO X5、VINNO X6、VINNO X7、VINNO M55、VINNO E35、VINNO X65、VINNO X55、VINNOX35、VINNO G55E、VINNO M55E、VINNO M55E、VINNO M55P、VINNO E35E、VINNOE35P
Indications for Use (Describe)
The device is a general purpose diagnostic ultrasound system intended for use by appropriately trained healthcare professionals in a hospital setting. The device is intended for ultrasound imaging, measurement, display and analysis of the human body and fluid.
This device is indicated for Abdominal; Fetal/Obstetrics; Gynecology;Transvaginal; Urology(including prostate); Transrectal; Cardiac(adult and pediatric); Peripheral Vascular; Small Organs/Parts(thyroid, breast, testicle, Musculoskeletal Conventional and Superficial); Pediatrics(including neonatal cephalic diagnostic Ultrasound applications.
The operating modes supported by the device are B mode, Harmonic mode, M mode, Color Flow mode, Power-Doppler mode, Pulsed Wave Doppler mode, Continuous Wave Doppler, 3D mode, Spatio Temporal Image Correlation mode, Tissue Doppler mode, Tissue Velocity Imaging mode, Elastography mode and Contrast Agent Imaging mode.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
l Submitter
Device submitter: VINNO Technology (Suzhou) Co., Ltd.
5F Building A. 4F Building C No. 27 XinFa Rd. Suzhou Industrial Park, Suzhou 215123 Jiangsu China
Contact person: Allen Liu
Regulatory Affairs Phone: +86 512 62873806 Fax: +86 512 62873801 Email: allen.liu@vinno.com
Date of written: 16 Jun 2023
II Device
Trade Name of Device: VINNO G50、VINNO G55、VINNO M50、VINNO E30、VINNO X5、 VINNO X6、VINNO X7、VINNO M55、VINNO E35、VINNO X65、VINNO X55、VINNO X35、VINNO G55E、VINNO G55P、VINNO M55E、VINNO M55P、VINNO E35E、VINNO E35P
Regulation name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1550 Regulatory Class: II Product code: IYN, IYO, ITX
III Predicate Devices
Trade name: HS40 Diagnostic Ultrasound System Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1550 Regulatory Class: II Product code: IYN, IYO, ITX Premarket Notification: K180409
IV Device description
The VINNO G50、VINNO G55、VINNO M50、VINNO E30、VINNO X5、VINNO X6、VINNO X7、VINNO M55、VINNO E35、VINNO X65、VINNO X55、VINNO X35、VINNO G55E、 VINNO G55P、VINNO M55E、VINNO M55P、VINNO E35E、VINNO E35P ultrasound devices are professional digital color ultrasonic diagnostic apparatus. It transmits ultrasound waves into the body tissues and displays the echo images of the tissues and
{4}------------------------------------------------
blood flow accordingly.
V Indications for use
The device is a general purpose diagnostic ultrasound system intended for use by appropriately trained healthcare professionals in a hospital setting. The device is intended for ultrasound imaging, measurement, display and analysis of the human body and fluid.
This device is indicated for Abdominal; Fetal/Obstetrics; Gynecology;Transvaginal; Urology(including prostate); Transrectal; Cardiac(adult and pediatric); Peripheral Vascular; Small Organs/Parts(thyroid, breast, testicle, Musculo-skeletal Conventional and Superficial); Pediatrics(including neonatal cephalic); and Adult Cephalic diagnostic Ultrasound applications.
The operating modes supported by the device are B mode. Harmonic mode, M mode, Color Flow mode, Power-Doppler mode, Pulsed Wave Doppler mode, Continuous Wave Doppler, 3D mode, 4D mode, Spatio Temporal Image Correlation mode, Tissue Doppler mode, Tissue Velocity Imaging mode, Elastography mode and Contrast Agent Imaging mode.
VI Comparison of technological characteristics with the predicate devices
The VINNO G50、VINNO G55、VINNO M50、VINNO E30、VINNO X5、VINNO X6、VINNO X7、VINNO M55、VINNO E35、VINNO X65、VINNO X55、VINNO X35、VINNO G55E、 VINNO G55P、VINNO M55E、VINNO M55P、VINNO E35E、VINNO E35P ultrasound devices have the same technological characteristics and fundamental design as the predicate devices. The VINNO G50、VINNO G55、VINNO M50、VINNO E30、VINNO X5、 VINNO X6、VINNO X7、VINNO M55、VINNO E35、VINNO X65、VINNO X55、VINNO X35、VINNO G55E、VINNO G55P、VINNO M55E、VINNO M55P、VINNO E35E、VINNO E35P ultrasound devices and the predicate device are all general purpose ultrasound devices designed to provide real-time images for diagnosis. The differences between The VINNO G50、VINNO G55、VINNO M50、VINNO E30、VINNO X5、VINNO X6、VINNO X7、VINNO M55、VINNO E35、VINNO X65、VINNO X65、VINNO X35、VINNO G55E、 VINNO G55P、VINNO M55E、VINNO M55P、VINNO E35E、VINNO E35P ultrasound devices and predicate devices do not alter suitability of the proposed device for its intended use.
See the following table for details on the comparison of technical features with the predicate device:
| Device feature | VINNO G50, VINNO G55, | SAMSUNG MEDISON HS40 |
|---|---|---|
| VINNO M50, VINNO E30, | K180409 (predicate device) | |
| VINNO X5, VINNO X6,VINNO X7, VINNO M55,VINNO E35, VINNO X65,VINNO X55, VINNO X35,VINNO G55E, VINNOG55P, VINNO M55E,VINNO M55P, VINNOE35E, VINNO E35P(subject device) | ||
| Indications for use | The device is a general purpose diagnostic ultrasound system intended for use by appropriately trained healthcare professionals in a hospital setting. The device is intended for ultrasound imaging, measurement, display and analysis of the human body and fluid.This device is indicated for Abdominal; Fetal/Obstetrics; Gynecology;Transvaginal; Urology(including prostate); Transrectal; Cardiac(adult and pediatric); Peripheral Vascular; Small Organs/Parts(thyroid, breast, testicle, Musculo-skeletal Conventional and Superficial); Pediatrics(including neonatal cephalic); and Adult Cephalic diagnostic Ultrasound applications | The HS40 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal/Obstetric,Abdominal,Gynecology,Intra-operative,Pediatric,Small Organ,Neonatal Cephalic,Adult Cephalic,Trans-rectal, Trans-vaginal,Muscular-Skeletal(Conventional,Superficial), Urology,Cardiac Adult,Cardiac Pediatric and Peripheral vessel |
{5}------------------------------------------------
{6}------------------------------------------------
| The operating modessupported by the deviceare B mode, Harmonicmode, M mode, ColorFlow mode,Power-Doppler mode,Pulsed Wave Dopplermode, Continuous WaveDoppler, 3D mode, 4Dmode, Spatio TemporalImage Correlation mode,Tissue Doppler mode,Tissue Velocity Imagingmode, Elastographymode and ContrastAgent Imaging mode. | ||
|---|---|---|
| User qualification | Qualified healthcareprofessionals | Qualified physicians orsonographers |
| Physical Specification | Width: 605mm;Depth: 875mm;Height: 1260mm;Weight: Basic unit withoutaccessories approx:60kg;Control panel: Ergonomicdesigned control panelwith interactiveback-lighting;+60°to - 60°horizontalswivel together with touchpanel and monitor;Number of Probes:4 | Height: 1354~1620mm;Width: 520mm;Depth: 730mm;Weight:51Kg(withoutaccessories);Weight: Approx. 57Kg(with safeworking load) |
| Patient contactmaterials | Probe housing: ABS,Probe lens: Silicon rubberComply with ISO10993series | Probe housing: ABSProbe lens: Silicon rubberComply with ISO10993 series |
| Operating modes | B mode, Harmonic mode,M mode, Color Flowmode, Power-Dopplermode, Pulsed Wave | Mode (2D), Color Doppler Mode(C), Pulse Wave Doppler (PWD),Continuous WaveDoppler(CWD): Steered / Static |
| Dopplermode,Continuous WaveDoppler, 3D mode, 4Dmode, Spatio TemporalImage Correlation mode,Tissue Doppler mode,Tissue Velocity Imagingmode, Elastographymode, Contrast AgentImaging mode. | Power Doppler Mode (PD),S-Flow ™ Mode, M-Mode (M),Anatomical MMode,Single/Dual/Quad Mode,Volume ModeTDI/TDW, ElastoScan Mode | |
| Operating controls | Gain, Frequency, FocusPosition, TView, L/R, Fullscreen, VT issue,Dynamic Range, Linedensity, VSharpen, ImageAngle, Persistence, GrayFilter, Smooth, FocusWidth, Acoustic Power,Biopsy, VNear,EdgeEnhance,NeedleEnhance, TI,SGC, Depth,TGC slidescontrols, B Steer, 2DAutomatic Optimization,HarmonicImaging,VFusion,Vspeckle, VSharpen,U/D,Trapezoid,PView,Map, Mmode, Sweep Speed,Curve MAM, PRF, WallFilter, Packet Size, ColorLevel, Invert, Color Map,Sync Display, RadiantFlow, Sync ROI,BaseLine Flash | Dynamic Range,Frame Average,Frequency,Gain , Harmonic,Pulse Inversion Harmonic (Probedependent),Line Density, Power,Reject,Scan Area, TGC, Write Zoom, -MultiVision (Probe Dependent),Beam Steering (ProbeDependent), Trapezoid (ProbeDependent), Free Angle Plane,Filter, PRF (Scale), SampleVolume Angle, Sample VolumePosition, Sample Rate, SweepSpeed, Chroma Map, Gray Map,Image Size, ClearVision, M ModeMap, Alpha Blending, BlendingLevel, Enhancement, Hide Color,Balance, Baseline,TraceDirection, Trace Method, Sound |
| Reduction,Persistence,Transparency, Steer, Biplane probe | ||
| Measurements | Abdominal, Obstetrics, Gynecology, Urology, Pediatrics, Cardiac, Vessel, Small organs(e.g. breast, testes, thyroid) , Trans-cranial Doppler, Fetal heart | Caliper, Abdomen, Cardiac, Gynecology, Vascular, Obstetrics, Fetal Heart, Urology, MSK, Small Parts, Pediatric |
| Comments | Comments and bodymarks | Comments and bodymarks |
| Probe types | Convex arrayLinear arrayPhased array4D ProbeEndocavityMicro ConvexVolume Convex | Convex arrayLinear arrayPhased arrayEndo-CavityMicro-Convex ArrayPencilVolume Probe |
| Display monitor | 21.5/23.8 inch LCD monitor | 21.5 inch LED monitor |
| Acoustic output | Comply with Track 3 limits:Ispta.3≤720mW/cm²MI≤1.9 | Comply with Track 3 limits:Ispta.3≤720mW/cm²MI≤1.9 |
| Conformity standards | IEC6061-1IEC60601-1-2IEC60601-2-37IEC60601-1-6IEC62366NEMA UD 2 | IEC6061-1IEC60601-1-2IEC60601-2-37IEC60601-1-6IEC62366NEMA UD 2 |
| Peripherals | Printers, USB DVDRW, ECG, S-Video output Cable, Buletooth/Wireless adapter, Foot switch, HDMI Monitor | S-Video Output, HDMI, Printers, DVD, USB, Foot switch, ECG, Monitor |
{7}------------------------------------------------
{8}------------------------------------------------
VII Performance data
The following performance data were provided in support of the substantial equivalence
{9}------------------------------------------------
determination.
Biocompatibility testing
Biocompatibility of the probes was evaluated in accordance with ISO 10993-1:2009. All evaluation acceptance criteria were met
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Ultrasound System. The system complies with the IEC 60601-1 and IEC 60601-2-37 for safety and the IEC 60601-1-2 for EMC.
Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.
Acoustic output testing
Acoustic output testing was performed according to NEMA UD2 and IEC60601-2-37.
VIII Conclusion
The VINNO G50、VINNO G55、VINNO M50、VINNO E30、VINNO X5、VINNO X6、VINNO X7、VINNO M55、VINNO E35、VINNO X65、VINNO X55、VINNO X35、VINNO G55E、 VINNO G55P、VINNO M55E、VINNO M55P、VINNO E35E、VINNO E35P ultrasound devices are substantially equivalent to their predicate device. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the leqally marketed device.
Comparison:
-Both the subject and predicate device are used for the diagnostic ultrasonic imaging and fluid flow analysis. The indications for use of the subject devices - VINNO G50, VINNO G55, VINNO M50, VINNO E30, VINNO X5, VINNO X6, VINNO X7, VINNO M55, VINNO E35, VINNO X65, VINNO X55, VINNO X35, VINNO G55E, VINNO G55P, VINNO M55E, VINNO M55P, VINNO E35E, VINNO E35P are covered by the indications for use of the predicate device. Ophthalmic and Trans-esoph are not intended to be used.
-The subject and predicate device employ the same fundamental scientific technology and have similar physical designs.
-The operating modes of the subject devices are covered by the operating modes of the predicate device Diagnostic Ultrasound System.
-The operating controls are similar as the predicate device.
-The measurement part are also similar with the the predicate device.
-The probe types are the same as the predicate device HS40 Diagnostic Ultrasound
{10}------------------------------------------------
System (K180409).
-The display monitor are almost similar with the predicate device.
-The contact materials are similar as those of the predicate device, and all passed the biocompatibility test respectively.
-The subject and predicate device are all track 3 devices meeting the maximum limits of the acoustic output parameters.
-The subject and predicate device conform to the same electrical safety standards.
-The subject and predicate device support similar interfaces with accessories.
N/A