K Number
K223890

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-06-16

(171 days)

Product Code
Regulation Number
876.5980
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NAJA Gastrointestinal Balloon Catheter is intended to administer contrast medium to the small intestine.

Device Description

The NAJA Gastrointestinal Balloon Catheter ("NAJA device") is a single-use, single-operator, over the-wire, double balloon catheter that is compatible with 0.035" guidewires. The device is introduced orally and designed for temporary occlusion of the gastrointestinal (GI) lumen (small intestine), allowing infusion of fluids for enteroclysis studies (i.e., imaging tests of the small intestine). The device is designed to fit inside the working channel of a gastroscope and reaches its intended location by tracking over a 0.035" guidewire.

The NAJA device features a separate hub connector that is attachable and detachable from the catheter. This feature allows removal of the gastroscope from the catheter once the balloons are in the correct location and before the hub connector is attached for balloon inflation and infusion of fluids into the GI lumen. The hub connector is supplied with 1-way stopcocks on the inflation and infusion ports. The NAJA device is also provided with a 30cc syringe for balloon inflation and infusion of fluids. The balloons are inflated independently using air.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the NAJA Gastrointestinal Balloon Catheter, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The FDA 510(k) summary details various design verification tests. The "Specification" column represents the acceptance criteria, and the "Test Result" column indicates if the device met that criterion (mostly "Pass" or a specific measurement).

Design Verification TestAcceptance CriteriaReported Device Performance
Biocompatibility Testing
Cytotoxicity StudyNo evidence of causing cell lysis or toxicity; grade less than 2 (mild reactivity) per ISO 10993-5:2009.Pass
ISO Intracutaneous StudyDifference between test article extract overall mean score and corresponding control extract overall mean score was 0.0 per ISO 10993-23:2021.Pass
ISO Maximization Sensitization StudyNo evidence of causing delayed dermal contact sensitization in guinea pig per ISO 10993-10:2021.Pass
Bench and Performance Testing (Package)
Legibility of MarkingsIFU, pouch label, and carton label are legible.Pass
Seal WidthPouch seal width is ≥ 0.19".Pass
Pouch Seal Visual InspectionPouch seals are free of channels or other damages that could impair sterility.Pass
Bubble Emission TestNo evidence of leakage through any surface of pouch.Pass
Pouch Seal StrengthPouch seal can withstand a force ≥ 3.35N over a seal length of 25.4mm.Pass
Pouch Ease of OpeningPouch is easily opened with gloved hands, while keeping internal contents stable. Components inside pouch are easily removable from backboard.Pass
Bench and Performance Testing (Component & Device)
Component Visual InspectionDevice components are free of damages that could affect performance of product (kinks, cracks, separated components, etc.).Pass
Surface DefectsAll surfaces and open channels of device are free of loose foreign matter, pores, cracks, and remainders of tooling agents.Pass
Atraumatic TipDistal tip is smooth, rounded, and free from sharp edges.Pass
Atraumatic Infusion HolesInfusion holes are smooth and free from sharp edges.Pass
Radiopaque MarkersFive (5) markers are present on device.Pass
0.035" GW Loading - Front/BackDevice accommodates front- and back-loading of a 0.035" GW.Pass
Proximal Extrusion DimensionsDimensions suitable for insertion into hub connector.Pass
10.5Fr Channel CompatibilityDistal 30cm of device fits through a 10.5Fr channel.Pass
Working LengthDevice working length is 2360 ± 20mm.Pass
Tip LengthTip length is 20 ± 2mm.Pass
Scaffold LengthScaffold length is 120 ± 2mm.Pass
Catheter Outer DiameterOuter diameter is ≤ 3.5mm for the catheter shaft, balloon welds, distal tip.Pass
Simulated Use Testing
Number of Balloons and SizeDevice has two balloons that can be inflated to a diameter of 50mm.Pass
Hub Connector DetachabilityHub connector is detachable from device and can be attached for fluid delivery.Pass
Shaft Insertion FeatureShaft marker is no longer visible once inserted into hub connector.Pass
Gastroscope CompatibilityDevice fits through working channel of gastroscope.Pass (implied, as other aspects of simulated use passed)
Simulated Use TestingThe device remains functional throughout the lifecycle of one clinical procedure.Pass (implied, as other aspects of simulated use passed)
Unique Hub Connector IdentificationProximal terminations of hub assembly are appropriately identified.Pass (implied, as other aspects of simulated use passed)
Legibility of Device MarkingsWording on hub assembly labels is legible.Pass (implied, as other aspects of simulated use passed)
Functional Testing
Hub Connector and Hub Assembly PressurePressure decay specification is met.Pass
Balloon Pressure and CrosstalkPressure decay specification is met.Pass
Balloon OD/L Ratio – unconfinedUnconfined balloon shape specification is met.Pass
Balloon Diameter and Inflation Volume RelationshipRelationship between balloon diameter and inflation volume shall be determined.Balloon Compliance: 9mL @ 20mm OD, 20mL @ 30mm OD, 44mL @ 40mm OD, 76mL @ 50mm (MAX)
Balloon OD/L Ratio – confinedConfined balloon shape specification is met.Pass
Inflation TimeInjection of 30mL of air in ≤ 4s.Pass
Deflation TimeRemoval of 30mL of air in ≤ 17s.Pass
Infusion TimeInfusion of 30mL of water in ≤ 19s.Pass
Channel OcclusionDevice is able to occlude channels 20-50mm.Pass
Destructive Testing
Rated Burst VolumeBalloons have a rated burst volume (RBV) ≥ 81.5mL.Pass
Bend RadiusNo shaft kink and/or fracture when wrapped 180° around a 20mm bend radius.Pass
Shaft Buckling ForceShaft buckle force specification is met.Pass
Hub to Shaft Slip ForceHub connector slip force specification is met.Pass
Tensile – Proximal Shaft to ExtrusionTensile force of proximal shaft to extrusion is ≥ 15N.Pass
Tensile - Distal Shaft to ExtrusionTensile force of distal shaft to extrusion is ≥ 15N.Pass
Tensile - Extrusion at ScaffoldTensile force of scaffold shall be ≥ 15N.Pass
Shear – Balloon Distal NeckShear force of distal balloon neck to extrusion bond is ≥ 15N.Pass
Peel - Balloon Proximal NeckPeel force of proximal balloon neck to extrusion bond is ≥ 15N.Pass
Tensile - Extension Tube to Hub ConnectorTensile force of extension tube to hub connector bond is ≥ 15N.Pass
Tensile - Luer to Extension TubeTensile force of luer to extension tube bond is ≥ 15N.Pass

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes bench and performance testing (also referred to as design verification testing), and biocompatibility testing. These are typical for medical device submissions and do not involve human subjects or clinical data in the form of a "test set" from patients.

  • Sample Size: Not explicitly stated for each individual bench or biocompatibility test, but device verification testing typically uses a statistically relevant number of samples to ensure robust performance across manufacturing batches.
  • Data Provenance: The tests are conducted in a laboratory setting using manufactured devices. There is no information regarding country of origin of data in terms of patient demographics, nor is it retrospective or prospective in a clinical sense.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This question is not applicable to the type of studies presented here.

  • The studies are engineering and laboratory tests (bench, performance, and biocompatibility). They do not involve interpretation of medical images or patient data that would require "experts" to establish "ground truth" in a clinical context. The "ground truth" for these tests is defined by the technical specifications and standards (e.g., ISO, internal design specifications).

4. Adjudication Method for the Test Set

This question is not applicable. As explained above, there is no "test set" in a clinical or diagnostic sense requiring human adjudication. The results are based on objective measurements against engineering specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The submission is a 510(k) for a physical medical device (catheter), not a diagnostic algorithm or AI-powered device. The studies focus on the physical and functional aspects of the catheter itself, not its impact on human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone algorithm performance study was not done. This device is a physical medical instrument, not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for the tests performed is defined by:

  • Engineering Specifications: Numerical ranges, physical properties, and functional requirements established during device design and development (e.g., "Pouch seal width is ≥ 0.19"").
  • International Standards: Specifically, ISO standards for biocompatibility (e.g., ISO 10993-5, ISO 10993-23, ISO 10993-10).
  • Observed Performance: Direct observation of device functionality and integrity (e.g., "Device components are free of damages").

These are objective, measurable criteria, not subjective interpretations requiring clinical ground truth like pathology, expert consensus, or outcomes data.

8. The Sample Size for the Training Set

This question is not applicable. The NAJA Gastrointestinal Balloon Catheter is a physical medical device, not an AI or machine learning model that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

This question is not applicable. As stated above, there is no "training set" for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square, the words "U.S. FOOD & DRUG" are stacked on top of the word "ADMINISTRATION".

June 16, 2023

Chess Medical Inc. % Christopher Sloan President Sloan Regulatory Consulting 322 Hart Road Gaithersburg, MD 20878

Re: K223890

Trade/Device Name: NAJA Gastrointestinal Catheter Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: FGD Dated: May 16, 2023 Received: May 16, 2023

Dear Christopher Sloan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling

{1}------------------------------------------------

(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sivakami Venkatachalam -S

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K223890

Device Name NAJA Gastrointestinal Balloon Catheter

Indications for Use (Describe)

The NAJA Gastrointestinal Balloon Catheter is intended to administer contrast medium to the small intestine.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary NAJA Gastrointestinal Balloon Catheter

Submitter

NameChess Medical
Address1650 Sherbrooke Street West, Suite 9EMontreal, Quebec H3H 1C9Canada
Contact PersonDr. Yen-I Chen
Phone(514) 291-3154
Email Addresscyen33@gmail.com
Date PreparedDecember 23, 2022

Device

Trade NameNAJA Gastrointestinal Balloon Catheter
Common NameGastrointestinal Balloon Catheter
Classification NameGastrointestinal Tube and Accessories
Classification Number21 CFR 876.5980
Product CodeFGD
Regulatory ClassII

Predicate Device

Name [510(k) Number]Maglinte Enteroclysis Balloon Catheter [K884379]
------------------------------------------------------------------------

Device Description

The NAJA Gastrointestinal Balloon Catheter ("NAJA device") is a single-use, single-operator, over the-wire, double balloon catheter that is compatible with 0.035" guidewires. The device is introduced orally and designed for temporary occlusion of the gastrointestinal (GI) lumen (small intestine), allowing infusion of fluids for enteroclysis studies (i.e., imaging tests of the small intestine). The device is designed to fit inside the working channel of a gastroscope and reaches its intended location by tracking over a 0.035" guidewire.

The NAJA device features a separate hub connector that is attachable and detachable from the catheter. This feature allows removal of the gastroscope from the catheter once the balloons are in the correct location and before the hub connector is attached for balloon inflation and infusion of fluids into the GI lumen. The hub connector is supplied with 1-way stopcocks on the inflation

{4}------------------------------------------------

and infusion ports. The NAJA device is also provided with a 30cc syringe for balloon inflation and infusion of fluids. The balloons are inflated independently using air.

Indications for Use

The NAJA Gastrointestinal Balloon Catheter is intended to administer contrast medium to the small intestine.

Comparison of Intended Use and Technological Characteristics with the Predicate Device

The proposed NAJA device has the same intended use and similar technological characteristics as the Maglinte Enteroclysis Balloon Catheter ("Maglinte Catheter"). The differences in the design between the devices do not raise different questions of safety and effectiveness. The table below summarizes the comparison between the NAJA device and Maglinte Catheter.

AttributeNAJA Device(Subject Device)Maglinte Catheter(Predicate Device)
510(k) NumberTo be assignedK884379
ApplicantChess MedicalLafayette Pharmacal
Classification Regulation;Device Class876.5980 (gastrointestinal tube andaccessories); class II876.5980 (gastrointestinal tube andaccessories); class II
Product Code(Device Type)FGD(Catheter, Retention, Barium EnemaWith Bag)FGD(Catheter, Retention, Barium EnemaWith Bag)
Intended Use / Indications for Use...intended to administer contrastmedium to the small intestine....intended to administer contrastmedium to the small intestine.
Sterile / Single UseYesYes
Sterilization MethodSterile (Ethylene Oxide)Sterile (Ethylene Oxide)
Principle of OperationInflation of one or two compliantballoons for occlusion of the smallintestine, followed by infusion ofcontrast medium in the smallintestine, distal to the proximalballoon or in the inter-balloonspace.Inflation of one compliant balloon forocclusion of the small intestine,followed by infusion of contrastmedium in the small intestine, distal tothe balloon.
Method of DeliveryOral insertion, through a compatibleworking channel (≥ 3.7 mm) of agastroscope.Nasal or oral insertion.
BalloonsNumber of BalloonsTwoOne
Balloon Inflation MediaAirAir
Balloon Diameter20mm - 50mm30 mm
AttributeNAJA Device(Subject Device)Maglinte Catheter(Predicate Device)
Balloon Length36 mm – 48 mm28 mm
Inflation Volume8 ml – 76 ml15 ml
Balloon MaterialPolyurethaneLatex
Marker BandsNumber of MarkerBandsFive(1) Distal tip, (4) proximal anddistal ends of the two balloonsOne(1) Distal tip
Marker Band LengthDistal tip - 4 mmProximal/Distal balloon ends – 2 mmDistal tip - 5 mm
Marker Band MaterialPolyether BlockAmide (Pebax®) 5533 SA 01 MEDw/ Tungsten 80%Metallic (material unknown)
Working Length260 cm155 cm
Number of LayersTwoInterior – Multi-Lumen ExtrusionExterior – Braided Shaft ExtrusionOneBi-Lumen Extrusion
Number of LumensFour(1) Guidewire lumen(1) Infusion lumen(2) Individual balloon inflationlumensTwo(1) Guidewire lumen,doubles as infusion lumen(1) Balloon inflation lumen
Number of Inflation ExitPortsTwoOne inflation exit port per balloonTwoTwo inflation exit ports to inflate asingle balloon
Number of Infusion ExitPortsThreeIn between the two balloonsFourDistal to the single balloon
ShaftMaterialsInterior – Multi-Lumen ExtrusionPolyether Block Amide (Pebax®)Exterior – Braided Shaft ExtrusionPolyether Block Amide (Pebax®)Radiopaque Polyvinyl Chloride
Outer DiameterDistal 30cm (Multi-LumenExtrusion)Outer Diameter: 7.5 FrProximal 206cm (Braided ShaftExtrusion)Outer Diameter: 10 FrOuter Diameter: 13 Fr
Distal TipOpen lumen at distal tip,guidewire can advanceClosed distal tip,guidewire cannot advance
Proximal EndEnclosed within a removable hubEnclosed within a permanent hub
AttributeNAJA Device(Subject Device)Maglinte Catheter(Predicate Device)
Proximal HubDetachableYesAble to attach/detach proximal hubto/from proximal end of shaftNoPermanently fixed to proximal end ofshaft
Number of PortsFour(1) Guidewire port(1) Infusion port(2) Individual balloon inflation portsTwo(1) Guidewire port, doubles as infusionport(1) Balloon inflation port
Mechanism to maintaininflation volumeOne-way StopcockCheck Valve
Guidewire Compatibility0.035in0.065in

Comparison of NAJA Device to Maglinte Catheter

{5}------------------------------------------------

{6}------------------------------------------------

Biocompatibility Testing

The following biocompatibility testing was performed on the NAJA device.

BiocompatibilityTestAcceptance CriteriaOriginResultConclusion
CytotoxicityStudy Using theISO ElutionMethodISO 10993-5: 2009,Biological Evaluation ofMedical Devices -Part 5: Tests for in vitrocytotoxicityThe test article extract showed no evidence ofcausing cell lysis or toxicity. The test articleextract met the requirements of the test sincethe grade was less than a grade 2 (mildreactivity).Pass
ISOIntracutaneousStudyISO 10993-23: 2021,Biological Evaluation ofMedical Devices -Part 23: Tests forirritationThe test article met the requirements of thetest since the difference between each testarticle extract overall mean score andcorresponding control extract overall meanscore was 0.0 and 0.0 for the SC and SO testarticle extracts, respectively.Pass
ISOMaximizationSensitizationStudyISO 10993-10: 2021,Biological Evaluation ofMedical Devices -Part 10: Tests for skinsensitizationThe test article extracts showed no evidenceof causing delayed dermal contactsensitization in the guinea pig. The test articlewas not considered a sensitizer in the guineapig maximization test.Pass

Bench and Performance Testing

The following bench and performance testing were conducted on the NAJA device.

Design Verification TestSpecificationTest Result
Package Performance Testing
Legibility of MarkingsIFU, pouch label, and carton label are legible.Pass
Design Verification TestSpecificationTest Result
Seal WidthPouch seal width is $\ge$ 0.19".Pass
Pouch Seal VisualInspectionPouch seals are free of channels or otherdamages that could impair sterility.Pass
Bubble Emission TestNo evidence of leakage through any surface ofpouch.Pass
Pouch Seal StrengthPouch seal can withstand a force $\ge$ 3.35N overa seal length of 25.4mm.Pass
Pouch Ease of OpeningPouch is easily opened with gloved hands,while keeping internal contents stable.Components inside pouch are easilyremovable from backboard.Pass
Component VisualInspectionDevice components are free of damages thatcould affect performance of product, such as:kinks, cracks, separated components, etc.Pass
Surface DefectsAll surfaces and open channels of device arefree of loose foreign matter, pores, cracks, andremainders of tooling agents.Pass
Atraumatic TipDistal tip is smooth, rounded, and free fromsharp edges.Pass
Atraumatic Infusion HolesInfusion holes are smooth and free from sharpedges.Pass
Radiopaque MarkersFive (5) markers are present on device.Pass
0.035" GW Loading -Front/BackDevice accommodates front- and back-loadingof a 0.035" GW.Pass
Proximal ExtrusionDimensionsDimensions suitable for insertion into hubconnector.Pass
10.5Fr ChannelCompatibilityDistal 30cm of device fits through a 10.5Frchannel.Pass
Working LengthDevice working length is 2360 $\pm$ 20mm.Pass
Tip LengthTip length is 20 $\pm$ 2mm.Pass
Scaffold LengthScaffold length is 120 $\pm$ 2mm.Pass
Catheter Outer DiameterOuter diameter is $\le$ 3.5mm for the cathetershaft, balloon welds, distal tip.Pass
Simulated Use
Number of Balloonsand SizeDevice has two balloons that can be inflated toa diameter of 50mm.Pass
Hub ConnectorDetachabilityHub connector is detachable from device andcan be attached for fluid delivery.Pass
Shaft Insertion FeatureShaft marker is no longer visible once insertedinto hub connector.Pass

{7}------------------------------------------------

{8}------------------------------------------------

Design Verification TestSpecificationTest Result
Gastroscope CompatibilityDevice fits through working channel ofgastroscope.
Simulated Use TestingThe device remains functional throughout thelifecycle of one clinical procedure.
Unique Hub ConnectorIdentificationProximal terminations of hub assembly areappropriately identified. Proximalterminations of the hub assembly areappropriately identified.
Legibility of DeviceMarkingsWording on hub assembly labels is legible.
Functional Testing
Hub Connector and HubAssembly PressurePressure decay specification is met.Pass
Balloon Pressure andCrosstalkPressure decay specification is met.Pass
Balloon OD/L Ratio –unconfinedUnconfined balloon shape specification is met.Pass
Balloon Diameter andInflation VolumeRelationshipRelationship between balloon diameter andinflation volume shall be determined.Balloon ComplianceVolume [mL]OD [mm]920203044407650 (MAX)
Balloon OD/L Ratio –confinedConfined balloon shape specification is met.Pass
Inflation TimeInjection of 30mL of air in $\le$ 4s.Pass
Deflation TimeRemoval of 30mL of air in $\le$ 17s.Pass
Infusion TimeInfusion of 30mL of water in $\le$ 19s.Pass
Channel OcclusionDevice is able to occlude channels 20-50mm.Pass
Destructive Testing
Rated Burst VolumeBalloons have a rated burst volume (RBV)$\ge$ 81.5mL.Pass
Bend RadiusNo shaft kink and/or fracture when wrapped180° around a 20mm bend radius.Pass
Shaft Buckling ForceShaft buckle force specification is met.Pass
Hub to Shaft Slip ForceHub connector slip force specification is met.Pass
Tensile – Proximal Shaft toExtrusionTensile force of proximal shaft to extrusion is$\ge$ 15N.Pass
Tensile - Distal Shaft toExtrusionTensile force of distal shaft to extrusion is$\ge$ 15N.Pass

{9}------------------------------------------------

Design Verification TestSpecificationTest Result
Tensile -Extrusion atScaffoldTensile force of scaffold shall be ≥ 15N.Pass
Shear – Balloon Distal NeckShear force of distal balloon neck to extrusionbond is ≥ 15N.Pass
Peel - Balloon ProximalNeckPeel force of proximal balloon neck toextrusion bond is ≥ 15N.Pass
Tensile - Extension Tube toHub ConnectorTensile force of extension tube to hubconnector bond is ≥ 15N.Pass
Tensile - Luer to ExtensionTubeTensile force of luer to extension tube bond is≥ 15N.Pass

Conclusions

The NAJA Gastrointestinal Balloon Catheter (subject device) is substantially equivalent to the legally marketed Maglinte Enteroclysis Balloon Catheter (predicate device) for administering contrast medium to the small intestine. The two devices have intended use and similar technological characteristics. The differences in the design between the devices do not raise different questions of safety and effectiveness. The results of biocompatibility testing confirm that the materials of construction of the NAJA device are safe, and the results of bench and performance (design verification) testing demonstrate that the NAJA device can function adequately in accordance with its intended use.

Chess Medical Inc. concludes that the NAJA Gastrointestinal Balloon Catheter (subject device) is substantially equivalent to the Maglinte Enteroclysis Balloon Catheter (predicate device).

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.