K Number
K223879

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-04-13

(107 days)

Product Code
Regulation Number
870.4240
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedure up to 6 hours in duration. Devices are intended for adult patients.

Device Description

The subject of the present Special 510(k) is the Quantum PureFlow Standard Heat Exchanger product family for inclusion of three additional variants with respect to the currently cleared (K220110) portfolio.

Quantum PureFlow Standard Heat Exchanger devices are designed to manage the temperature of blood during surgical procedures requiring cardiopulmonary bypass (CPB) for periods lasting less than 6 hours.

HX-S devices are designed to:

  • keep circulating blood at a specific temperature, depending on the type of surgery being performed;
  • maintain blood/patient thermoregulation during the CPB;
  • rewarm blood at the conclusion of the CPB in order to restore normothermic patient condition.

Quantum PureFlow Standard Heat Exchangers have been designed to be powered by heatercooler systems that use

  • Water
    or
  • Glycol-based solution
    as Heat Transfer Fluid (HTF).
AI/ML Overview

The provided text is an FDA 510(k) summary for a medical device called the "Quantum PureFlow Standard Heat Exchanger." It outlines the device's purpose, regulatory classification, and the comparison to an existing predicate device to demonstrate substantial equivalence.

However, the information requested in your prompt (Acceptance Criteria, Study Details, Sample Sizes, Expert Qualifications, Adjudication Methods, MRMC study, Standalone Performance, Ground Truth, Training Set Details) is typically related to the performance evaluation of AI/ML-based medical devices, particularly regarding diagnostic accuracy or clinical effectiveness.

The Quantum PureFlow Standard Heat Exchanger is a physical medical device (a cardiopulmonary bypass heat exchanger), not an AI/ML software device. Therefore, the types of studies and acceptance criteria described in your prompt (e.g., accuracy metrics, human reader improvement with AI, ground truth establishment by experts, training set details) are not applicable to this device.

The "Performance Data" section in the document (Page 5) does mention "NON-CLINICAL TESTING" and lists:

  • Updating labeling and IFU according to ISO 15223-1:2021.
  • Performance tests for new REF codes according to ISO 7199:2016.

This indicates that the acceptance criteria for this device would be related to meeting the standards outlined in ISO 7199:2016 for cardiopulmonary bypass heat exchangers (e.g., heat exchange efficiency, pressure drop, biocompatibility), as well as general safety and performance requirements for medical devices. The "study that proves the device meets the acceptance criteria" would be the non-clinical testing performed according to these standards.

In summary, none of the specific details you requested regarding AI/ML performance evaluation (such as number of experts for ground truth, MRMC study, sample sizes for test and training sets for an algorithm, etc.) are present in this document because it describes a hardware medical device, not an AI/ML software.

If this were an AI/ML device, the requested information would be crucial for understanding its validation and demonstrating its safety and effectiveness. Since it's a heat exchanger, the relevant "acceptance criteria" revolve around its physical performance (e.g., heat transfer rate, flow resistance) as per recognized standards (like ISO 7199), and the "study" would be bench testing to verify compliance with those standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 13, 2023 Quara S.r.1. Raffaella Tommasini QA&RA Director Via di Mezzo, 23 Mirandola, Modena 41037 Italy

Re: K223879

Trade/Device Name: Quantum PureFlow Standard Heat Exchanger (HX11W-S2, HX11W-S1M and HX11W-S2M) Regulation Number: 21 CFR 870.4240 Regulation Name: Cardiopulmonary bypass heat exchanger Regulatory Class: Class II Product Code: DTR Dated: March 15, 2023 Received: March 15, 2023

Dear Raffaella Tommasini:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Kathleen M. Grunder -S for Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223879

Device Name

Quantum PureFlow Standard Heat Exchanger (HX11W-S2, HX11W-S1M and HX11W-S2M)

Indications for Use (Describe)

The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedure up to 6 hours in duration. Devices are intended for adult patients.

Type of Use (Select one or both, as applicable)
✓ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for QURA advanced medical solutions. The logo features a stylized human figure with arms outstretched, enclosed within a curved line. To the right of the figure is the word "QURA" in a bold, sans-serif font. Below "QURA" are the words "advanced medical solutions" in a smaller, lighter font.

510(K) SUMMARY

SUBMITTER l.

Submitter Name:Qura S.r.l.
Submitter Address:Via di Mezzo, 23 41037 Mirandola (MO) Italy
Contact Person:Raffaella Tommasini, QA&RA Director – Qura s.r.l.
Phone:+39 0535 1803050
e-mail:raffaella.tommasini@quramed.com
Fax:+39 0535 1803051
Date Summary Prepared:23 December, 2022

DEVICE II.

Proprietary Name:Quantum PureFlow Standard Heat Exchangers (including HX11W-S2, HX11W-S1M and HX11W-S2M, object of present submission)
Common Name:Standard Heat Exchanger
Classification Name:Cardiopulmonary bypass heat exchanger
Regulatory Class:II
Product Code:DTR
Panel:Cardiovascular Devices, Office of Health Technology 2 (OHT2) / Division of Health Technology 2 B (Circulatory Support, Structural and Vascular Devices)

III. PREDICATE DEVICE

Proprietary Name:Quantum PureFlow Standard Heat Exchangers
Common Name:Standard Heat Exchanger
Classification Name:Cardiopulmonary bypass heat exchanger
Regulatory Class:II
Product Code:DTR
Panel:Cardiovascular Devices, Office of Health Technology 2 (OHT2) /Division of Health Technology 2 B (Circulatory Support,Structural and Vascular Devices)
510(k) Number:K220110

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a logo for Qura Advanced Medical Solutions. The logo features a stylized human figure inside of an oval shape on the left side of the image. To the right of the figure is the word "QURA" in a bold, sans-serif font. Below the word "QURA" is the phrase "advanced medical solutions" in a smaller, sans-serif font.

IV. DEVICE DESCRIPTION

The subject of the present Special 510(k) is the Quantum PureFlow Standard Heat Exchanger product family for inclusion of three additional variants with respect to the currently cleared (K220110) portfolio.

Quantum PureFlow Standard Heat Exchanger devices are designed to manage the temperature of blood during surgical procedures requiring cardiopulmonary bypass (CPB) for periods lasting less than 6 hours.

HX-S devices are designed to:

  • keep circulating blood at a specific temperature, depending on the type of surgery being performed;
  • maintain blood/patient thermoregulation during the CPB;
  • . rewarm blood at the conclusion of the CPB in order to restore normothermic patient condition.

Quantum PureFlow Standard Heat Exchangers have been designed to be powered by heatercooler systems that use

  • . Water
    or

  • Glycol-based solution
    as Heat Transfer Fluid (HTF).

V. INTENDED USE / INDICATIONS FOR USE

The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration. Devices are intended for adult patients.

No changes have been made to the devices' intended use with respect to the original K220110 submission.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed within a partial circular shape. The text "QURA" is written in a bold, sans-serif font, with the words "advanced medical solutions" appearing in a smaller font size beneath it. The color scheme is primarily blue.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Additional device models have been introduced in the HX-S product family in order to provide to clinicians/users an additional option to perform cardiopulmonary bypass procedures.

New REF codes introduced in the portfolio for Quantum PureFlow Standard Heat Exchanger product family are aligned with information provided in original submission K220110 in terms of general structure (including packaging), principle of operation, intended use, manufacturing and sterilization processes.

Technical specifications have been verified through testing activities performed according to the same internal applicable standards/protocols of the original cleared devices.

Applicable testing activities demonstrated that the proposed device model does not raise any new issues of safety and effectiveness as compared to the currently cleared predicate and reference products.

VII. PERFORMANCE DATA

NON-CLINICAL TESTING

The following activities were performed to demonstrate product safety and effectiveness, considering the proposed change and related impact:

  • । update of labeling and Instructions for Use (IFU) according to ISO 15223-1:2021 Medical Devices - Symbols to Be Used with Medical Device Labels, Labelling, And Information to Be Supplied - Part 1: General Requirements [Recognition Nr. 5-134];
  • -Performance tests for new REF codes introduced in the portfolio, according to ISO 7199:2016 [Recognition Nr. 3-150].

Animal Study

No animal studies have been performed to support changes object of the present Special 510(k).

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with multiple arms and legs, enclosed within a partial oval shape. To the right of the figure, the word "QURA" is written in a bold, sans-serif font, with the words "advanced medical solutions" written in a smaller font underneath.

CLINICAL TESTING

No clinical data on Quantum PureFlow Standard Heat Exchanger have been included in the current 510(k) submission.

VIII. CONCLUSIONS

Considering all changes performed on original devices cleared by K220110, it could be stated that devices under evaluation are identical in terms of intended use and applicable medical technique.

Based on the testing activities, technological characteristics and the indications for use, the proposed devices have been demonstrated to be appropriate for their intended use and are considered substantially equivalent to Qura's own original devices.

§ 870.4240 Cardiopulmonary bypass heat exchanger.

(a)
Identification. A cardiopulmonary bypass heat exchanger is a device, consisting of a heat exchange system used in extracorporeal circulation to warm or cool the blood or perfusion fluid flowing through the device.(b)
Classification. Class II (performance standards).