K Number
K223856

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-08-11

(231 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NuEra Tight RF and NuEra Tight RF Plus are intended:

  • to provide topical heating for the purpose of elevating tissue temperature for the treatment of conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

  • to provide, with a massage device, a temporary reduction in the appearance of cellulite.

  • to provide, via heat-induced lipolysis at 1 and 2 MHz, abdominal circumference reduction with adjunctive improvement in the appearance of skin laxity.

Device Description

NuEra Tight RF and NuEra Tight RF Plus are designed to develop localized heat to warm the subcutaneous tissue by means of radio frequency energy, delivered through electrodes in contact with the patient. Specifically,

  • NuEra Tight RF radiofrequency generator with single RF electrode connector. -
  • NuEra Tight RF Plus radiofrequency generator with single RF electrode connector. In addition, the NuEra Tight RF Plus has a larger size to accommodate an additional electronic part that has previously been cleared (K201239) and classified under procode NGX (Stimulator, Muscle, Powered, For Muscle Conditioning, 21 CFR § 890.5850); hence, this model provides the functions classified under both procodes PBX and NGX.
    The treatment performed by the NuEra Tight RF models consists of increasing the temperature of the treated tissues up to maximum of 45°C. Therefore, depending on the treatment and intended use, different parts of the body can be treated.

The models use RF monopolar and bipolar capacitive electrodes. Monopolar capacitive RF electrodes have different sizes and plug into an RF handpiece that provides connection to the RF generator. Handpieces of different shapes are available to facilitate use by the operator on different body parts. Capacitive electrodes work in combination with a return plate that must be in contact with the patient's body during the treatment in order to close the circuit with the RF generator. Return plates are available as reusable or single use, with specific connectors on the panel below the front tray of the main control unit.

One bipolar capacitive electrode is provided fixed to a dedicated handpiece intended for the treatment of small body areas. Being bipolar, the electrode is not meant to work with a return plate.

One massage handpiece is provided to be used to add a mechanical treatment to the heat emission.

All the handpieces are intended to be used with a small amount (approximately 1 mm layer) of an RF conductive cream (Parker Redux cream, cleared under K782055, or similar cream available in the US). The purpose of the cream is to provide aid to transfer the heat.

A footswitch is provided as an optional user interface that allows to start and stop the medical treatment. It can be used as an alternative to the GUI start and stop button.

The pause handpiece is an optional accessory can be used to pause the treatment without using the GUI.

AI/ML Overview

The provided document describes a 510(k) premarket notification for the "NuEra Tight RF and NuEra Tight RF Plus" devices, focusing on the addition of an abdominal circumference reduction indication. The acceptance criteria and the study proving the device meets these criteria are detailed below.

1. Table of Acceptance Criteria and Reported Device Performance

Criteria CategoryAcceptance Criteria (Effectiveness)Reported Device Performance
Primary EffectivenessStatistically significant change in mean circumferential measurement at 12 weeks post last treatment compared to baseline.1 MHz Group (n=23): Statistically significant reduction in mean waist circumference at 12 weeks follow-up. 2 MHz Group (n=18): Statistically significant reduction in mean waist circumference at 12 weeks follow-up.
Secondary EffectivenessImprovement in skin laxity compared to baseline at 4 and 12 weeks post last treatment.Blinded assessors identified improvement in skin laxity in a majority of subjects at 4 and 12 weeks follow-up.
Aesthetic improvement compared to baseline at 4 and 12 weeks post last treatment.Investigators and subjects reported improvement in aesthetic appearance (GAIS score) in the majority of each group. Blinded reviewers correctly identified 'before' and 'after' pictures for most subjects.
SafetyAcceptable levels of subject discomfort and pain during treatment.Median VAS Score was 0, with 50% reporting no pain/discomfort. Average pain across all treatment arms was 1.6 (low-level).
Low incidence and severity of device-related adverse events.One mild burn (2 MHz arm) resolved with no action (0.3% of treatments). Two mild AEs (diarrhea, back pain) possibly related, resolved with no action. No serious AEs.

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Enrollment: 82 subjects were enrolled and randomized.
    • Effectiveness Analysis (Primary Endpoint): n=23 for the 1 MHz group and n=18 for the 2 MHz group. This reflects subjects who completed at least 3 of 4 treatment sessions.
  • Data Provenance: The study was a "multi-center" study, implying data from multiple locations, likely within one or more countries. The document does not explicitly state the country of origin. It was a prospective, randomized study.

3. Number of Experts and Qualifications for Ground Truth

  • For Skin Laxity and Aesthetic Appearance: "Blinded assessors" and "blinded reviewers" were used to evaluate before/after pictures and identify improvements.
  • Qualifications: The document does not specify the exact number or qualifications of these assessors/reviewers (e.g., radiologist with 10 years of experience). It just indicates they were "blinded."

4. Adjudication Method for the Test Set

  • The document describes "blinded assessors" and "blinded reviewers" for evaluating photographic evidence for skin laxity and aesthetic appearance. It does not mention a specific adjudication method like 2+1 or 3+1 for resolving discrepancies, implying a consensus or majority opinion approach might have been used, or that the "majority" finding was sufficient. For quantitative measurements like circumference, direct measurement itself serves as the ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a Human-in-the-loop (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done. The device in question is a medical aesthetic device (radiofrequency energy for tissue heating, circumference reduction, cellulite, and skin laxity), not an AI-based diagnostic or interpretative tool that would typically involve human readers.

6. Standalone (Algorithm Only) Performance

  • Not applicable. This device is a physical RF energy delivery system, not an algorithm or software designed to operate without human intervention in a diagnostic or analytical capacity. Therefore, a standalone performance evaluation in the context of an algorithm would not be relevant. The "standalone" performance here relates to the device's ability to achieve the intended physiological effects.

7. Type of Ground Truth Used

  • Primary Effectiveness (Circumference Reduction): Objective measurements of waist circumference.
  • Secondary Effectiveness (Skin Laxity & Aesthetic Appearance): Expert assessment (blinded assessors/reviewers) of photographic evidence and subjective reports from investigators and subjects (GAIS score).
  • Safety: Directly observed adverse events and patient-reported discomfort/pain levels (VAS score).

8. Sample Size for the Training Set

  • The document does not explicitly mention a "training set" or a separate training phase for the device itself. The clinical study described is the primary efficacy and safety study. As this is not an AI/ML device in the traditional sense, a specific "training set" for an algorithm is not applicable. The device's operational parameters are based on established RF technology and prior predicate devices.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as a discrete "training set" for an algorithm, separate from the clinical validation study, is not described or relevant for this type of device. The "ground truth" for the device's design and operational parameters would have been established through engineering principles, pre-clinical testing, and knowledge gained from predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 11, 2023

BIOS s.r.l. Manuela Brambilla Regional Regulatory Affairs Leader Via Guido Rossa 10/12 Vimodrone, MI 20055 Italy

Re: K223856

Trade/Device Name: NuEra Tight RF Family Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: PBX, GEI Dated: July 14, 2023 Received: July 14, 2023

Dear Manuela Brambilla:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows a digital signature. The signature is from Mark Trumbore and was signed on August 11, 2023 at 11:59:57 -04'00'. The signature appears to be an official digital signature.

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K223856

Device Name NuEra Tight RF and NuEra Tight RF Plus

Indications for Use (Describe) The NuEra Tight RF and NuEra Tight RF Plus are intended:

  • to provide topical heating for the purpose of elevating tissue temperature for the treatment of conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

  • to provide, with a massage device, a temporary reduction in the appearance of cellulite.

  • to provide, via heat-induced lipolysis at 1 and 2 MHz, abdominal circumference reduction with adjunctive improvement in the appearance of skin laxity.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "bios" in a stylized font, with the "i" dotted with a small circle. Below the word "bios" is the phrase "by Lumenis", where the "L" in Lumenis is stylized with a blue circle around it. The text is in a sans-serif font and appears to be part of a logo or branding material.

510(k) Summary

510(k) NumberK223856
510(k) TypeTraditional
Applicant InformationBios s.r.l.Via Guido Rossa, 10/1220055 Vimodrone (MI) – Italy
ContactEng. Manuela BrambillaPhone: 0039-02-27304275Email: regulatory@biosgroup.eu
Date PreparedAugust 10, 2023
Device NameNuEra Tight RF and NuEra Tight RF Plus
Common NameElectrosurgical cutting and coagulation device and accessories
Regulatory ClassClass II (21CFR§878.4400)
Product CodesGEI, PBX
Regulation NameElectrosurgical cutting and coagulation device and accessories

Predicate and Reference Devices

510(k) RefPro Code/Reg NoTrade NameApplicant
Predicate Device
K210867PBX, 878.4400NuEra Tight RF and NuEra Tight RF PlusBios s.r.l.
Reference Device
K162512GEI, PBX, 878.4400truSculptCutera, Inc.
K163415GEI, 878.4400SlimShapeSyneron Medical Ltd.

Device Description

NuEra Tight RF and NuEra Tight RF Plus are designed to develop localized heat to warm the subcutaneous tissue by means of radio frequency energy, delivered through electrodes in contact with the patient. Specifically,

  • NuEra Tight RF radiofrequency generator with single RF electrode connector. -

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "bios by Lumenis". The word "bios" is written in a sans-serif font, with the "i" having a curved top. Below "bios" is the text "by Lumenis", with the "L" in "Lumenis" being a stylized blue "U" shape.

  • NuEra Tight RF Plus radiofrequency generator with single RF electrode connector. In addition, the NuEra Tight RF Plus has a larger size to accommodate an additional electronic part that has previously been cleared (K201239) and classified under procode NGX (Stimulator, Muscle, Powered, For Muscle Conditioning, 21 CFR § 890.5850); hence, this model provides the functions classified under both procodes PBX and NGX.
    The treatment performed by the NuEra Tight RF models consists of increasing the temperature of the treated tissues up to maximum of 45°C. Therefore, depending on the treatment and intended use, different parts of the body can be treated.

The models use RF monopolar and bipolar capacitive electrodes. Monopolar capacitive RF electrodes have different sizes and plug into an RF handpiece that provides connection to the RF generator. Handpieces of different shapes are available to facilitate use by the operator on different body parts. Capacitive electrodes work in combination with a return plate that must be in contact with the patient's body during the treatment in order to close the circuit with the RF generator. Return plates are available as reusable or single use, with specific connectors on the panel below the front tray of the main control unit.

One bipolar capacitive electrode is provided fixed to a dedicated handpiece intended for the treatment of small body areas. Being bipolar, the electrode is not meant to work with a return plate.

One massage handpiece is provided to be used to add a mechanical treatment to the heat emission.

All the handpieces are intended to be used with a small amount (approximately 1 mm layer) of an RF conductive cream (Parker Redux cream, cleared under K782055, or similar cream available in the US). The purpose of the cream is to provide aid to transfer the heat.

A footswitch is provided as an optional user interface that allows to start and stop the medical treatment. It can be used as an alternative to the GUI start and stop button.

The pause handpiece is an optional accessory can be used to pause the treatment without using the GUI.

Indications for Use

The NuEra Tight RF and NuEra Tight RF Plus are intended:

  • . to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.
  • to provide, with a massage device, a temporary reduction in the appearance of cellulite.
  • to provide, via heat-induced lipolysis at 1 and 2 MHz, abdominal circumference 트 reduction with adjunctive improvement in the appearance of skin laxity.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for "bios by Lumenis". The word "bios" is written in a modern, sans-serif font, with the letters connected in a flowing manner. Below "bios" is the text "by Lumenis", with the "Lumenis" part featuring a stylized "U" logo. The logo is clean and professional, suggesting a brand identity focused on technology and innovation.

Predicate Device Comparison

The purpose of the 510(k) submission is to add the abdominal circumference reduction indications for the NuEra Tight RF models that were last cleared under K210867. The subject, predicate and reference devices, as shown in Table 1 below, all share the same operating principle, which is the delivery of RF energy by skin-contact electrodes to heat the underlying tissues to a certain temperature for a time period to achieve the intended effect. The reference devices are cleared with a circumference reduction indication.

TABLE 1. PREDICATE COMPARISON
FeatureSubject DevicesPredicate DevicesReference Device #1Reference Device #2
Device nameNuEra Tight RF andNuEra Tight RF PlusNuEra Tight RF andNuEra Tight RF PlustruSculptSlimShape
ManufacturerBios s.r.l., ItalyBios s.r.l., ItalyCutera, USASyneron Medical Ltd.,Israel
510(k)K223856K210867K162512K163415
FDA Product CodeGEI, PBXPBXGEI, PBXGEI
FDA RegulationNameElectrosurgical Cutting andCoagulation Device andAccessoriesElectrosurgical Cutting andCoagulation Device andAccessoriesElectrosurgical Cutting andCoagulation Device andAccessoriesElectrosurgical Cuttingand Coagulation Deviceand Accessories
FDA RegulationNumber21 CFR 878.440021 CFR 878.440021 CFR 878.440021 CFR 878.4400
OperatingPrincipleIdentical to PredicateRadio frequency energydelivered to the tissue toinduce heat and increase thetemperature of the treatedarea for therapeuticpurposesRadio frequency energydelivered to the tissue toinduce heat and increase thetemperature of the treatedarea for therapeuticpurposes.Radio frequency energydelivered to the tissue toinduce heat and increasethe temperature of thetreated area andconcomitant vacuumsuction of the treated area.
Indications for useThe NuEra Tight RF andNuEra Tight RF Plus areintended:- to provide topical heatingfor the purpose of elevatingtissue temperature for thetreatment of selectedmedical conditions such astemporary relief of pain,muscle spasms, andincrease in local circulation.- to provide, with a massagedevice, a temporaryreduction in the appearanceof cellulite.NuEra Tight RF Family isintended:- to provide topical heatingfor the purpose of elevatingtissue temperature for thetreatment of selectedmedical conditions such astemporary relief of pain,muscle spasms, andincrease in local circulation.- to provide, with a massagedevice, a temporaryreduction in the appearanceof cellulite.-truSculpt RF energy isintended to provide topicalheating for the purpose ofelevating tissue temperaturefor the treatment of selectedmedical conditions, such asrelief of pain and musclespasms and increase in localcirculation.- truSculpt massage device isintended to provide atemporary reduction in theappearance of celluliteDoes not have theseindications.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for "bios by Lumenis". The word "bios" is written in a sans-serif font, with the letters in a dark blue color. Below the word "bios", the words "by Lumenis" are written in a smaller font size. The "Lumenis" part of the logo is also in dark blue, and there is a stylized "L" symbol next to it.

TABLE 1. PREDICATE COMPARISON
FeatureSubject DevicesPredicate DevicesReference Device #1Reference Device #2
Device nameNuEra Tight RF andNuEra Tight RF PlusNuEra Tight RF andNuEra Tight RF PlustruSculptSlimShape
- to provide, via heat-induced lipolysis at 1 and 2MHz, abdominalcircumference reductionwith adjunctiveimprovement in theappearance of skin laxity.Do not have this proposedindication.-2 MHz setting for the 40cm² handpiece is indicatedfor temporary reduction incircumference of theabdomen.The SlimShape System isindicated for non-invasivelipolysis (breakdown offat) of the abdomen. Thedevice is indicated forreduction in circumferenceof the abdomen.
RF Frequencies470 kHz; 1 MHz; 2 MHz;4MHz; 6 MHz.For the proposed additionalabdominal circumferencereduction indication: 1 and2 MHz470 kHz; 1 MHz; 2 MHz;4MHz; 6 MHz2 MHz1 MHz
RF output powerMax 250 WMax 250 WMax 300 WNot Provided in 510(k)Summary
TemperatureMonitoringYesYesYesYes
ElectrodesCapacitive electrodes of20, 30, 40, 40+massage, 60,70, 80, and 100 mm indiameter.Capacitive electrodes of20, 30, 40, 40+massage, 60,70, 80, and 100 mm indiameterCapacitive electrode of 40cm² areaNot described
How UsedElectrode is manually andcontinuously moved overthe treated area by theoperator to maintain thespecified temperature of thetreated tissue over thespecified duration.Electrode is manually andcontinuously moved overthe treated area by theoperator to maintain thespecified temperature of thetreated tissue over thespecified duration.Electrode is manually andcontinuously moved overthe treated area by theoperator to maintain thespecified temperature of thetreated tissue over thespecified duration.SlimShape is described asfollows,"the SlimShape ApplicatorBelt incorporates an arrayof vacuum cavities; eachcavity includes a pair ofRF electrodes and asuction pad."
Prescription useRx onlyRx onlyRx onlyRx only
Clinical data tosupport theabdominalcircumferencereductionindicationNCT04406935Not applicableNCT02873104NCT02999763

Performance Bench Testing

Bench testing to the following standards were repeated to confirm that the devices continue to meet their specifications:

  • . IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012; A2:2021Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for "bios by Lumenis". The word "bios" is written in a modern, sans-serif font, with the letters in a dark blue color. Below "bios" is the text "by Lumenis", with the "Lumenis" part of the text being in a smaller font size. To the left of "Lumenis" is a small, stylized "U" shape, also in dark blue.

  • EN 60601-1-2:2015 Medical electrical equipment General requirements for basic . safety and essential performance. Collateral Standard: Electromagnetic disturbances. Requirements and tests
    No performance bench testing of the NuEra Tight RF models was conducted specifically for the additional circumference reduction indications because no device changes were needed to support this additional indication.

Performance Clinical Testing

A clinical study was performed to demonstrate the safety and effectiveness of the NuEra Tight RF Models for the proposed additional circumference reduction indications.

The study endpoints were as follows:

Primary Effectiveness Endpoint: Change in mean circumferential measurement at 12 weeks post last treatment compared to baseline within each treatment group.

Secondary Effectiveness Endpoints:

  • . Change in skin laxity compared to baseline at 4 and 12 weeks post last treatment.
  • Aesthetic improvement compared to baseline at 4 and 12 weeks post last treatment.

Safety Endpoints:

  • . Subject Discomfort and Pain Levels during Treatment.
  • Incidence and severity of device related adverse events during the study period. .

The study was a multi-arm, prospective, randomized, and multi-center. Initially, three arms were included in the study design:

  • . Arm 1: NuEra treatment using 470 kHz
  • . Arm 2: NuEra treatment using 1 MHz
  • Arm 3: NuEra treatment using 2 MHz

An equal 1:1:1 randomization allocation was planned. However, due to the COVID-19 pandemic that led to restrictions on non-essential medical procedures, quarantine, or reticence by subjects, the study experienced a high drop-rate. To maintain study power due to the availability of subjects during the COVID-19 pandemic and resource constraints, the sponsor revised the protocol to stop enrolling Arm 1. The second phase of the study randomized the remaining participants into only two arms: 1 MHz treatment or 2 MHz treatment, with a similar 1:1 randomized equal allocation. As each arm will be analyzed separately, the statistical power for each arm was unaffected.

The treatment regimen in the revised protocol remained the same as in the original protocol. Each subject received 4 treatments at 1 week intervals. Follow-up evaluations took place at 4 weeks (Follow-up 1) and 12 weeks (Follow-up 2) after the last treatment.

The results of the study were as follows:

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for "bios by Lumenis". The word "bios" is written in a modern, sans-serif font, with the letters in a dark blue color. Below the word "bios", the phrase "by Lumenis" is written in a smaller font size. The "L" in "Lumenis" is stylized with a blue circle around it.

For the 82 enrolled and randomized subjects:

  • mean age of 46.8±11.3 years, with a range of 21.7-65.4 years; ●
  • . seventy-two (72) females and 10 males;
  • self-identified to be: White (79%); Other (10%); Black or African American (6%); or ● Asian (5%), and to be of Hispanic or Latino ethnicity (13%); and,
  • mean circumference of 85.93±8.51 cm, with a range of 67.93-104.6 cm.

Due to the COVID-19 pandemic, 18 subjects did not complete at least 3 of 4 treatment sessions. Also, 32 subjects had a major protocol deviation, through either not having the second followup visit (13 subjects) or had the second follow-up visit more than 4 weeks late (19 subjects).

The data supporting the additional indications are from subjects who completed at least 3 of the 4 treatment sessions. They include:

Primary Effectiveness Endpoint: The mean change (reduction) from baseline in waist circumference at 12 weeks follow up was found to be statistically significant in the 1 MHz (n=23) and 2 MHz (n=18) groups:

Secondarv Effectiveness Endpoints:

  • . Skin laxity: Blinded assessors evaluated before/after pictures and identified improvement in skin laxity in a majority of subjects at the 4 and 12 week follow up visitsAesthetic appearance: Investigators and subjects were asked to assess aesthetic appearance after each visit (GAIS score). The majority in each group reported improvement in appearance.
    In addition, blinded reviewers were asked to identify which photos of subjects (per visit and treatment arm in subjects who completed the study course and appeared for all followup visits) were 'before treatment' and 'after treatment' pictures. The results were that most pictures were correctly identified by the blinded reviewers. Safety:

  • The median VAS Score was 0, with 50% of the subjects reported no pain or discomfort ● during any of the treatments and the average pain for all treatment arms was 1.6 (lowlevel of pain).

  • . One subject from the 2 MHz arm had a mild burn that was deemed related to the procedure, the adverse event (AE) resolved with no action taken (1/276 treatments, 0.3%).

  • There were 2 AEs considered by the investigators as possibly related to the device or the procedure (diarrhea and back pain). The events were mild and resolved with no action.

  • None of the AEs was considered serious. ●

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for "bios by Lumenis". The word "bios" is written in a modern, sans-serif font, with the letters connected in a flowing manner. Below "bios" is the text "by" followed by the Lumenis logo, which is a stylized "U" shape.

Conclusion

The proposed additional indication for the subject devices has already been cleared for the reference devices at treatment frequencies of 1 MHz and 2 MHz and using the same RF induction technology as the subject devices, which show that the proposed additional indication does not present a new intended use for RF induction devices. The provided clinical data for the subject devices, operating at 1 MHz frequencies, demonstrate that they are safe and effective for the proposed additional abdominal circumference reduction indication. The subject devices have the identical intended use and technological characteristics as their predicates, and the proposed additional indication does not create a new intended use or raise a new safety or effectiveness question. Thus, the subject devices are substantially equivalent to the predicate devices.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.