K Number
K223846

Validate with FDA (Live)

Device Name
iLet® ACE Pump
Manufacturer
Date Cleared
2023-05-19

(150 days)

Product Code
Regulation Number
880.5730
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The iLet ACE Pump is an alternate controller enabled (ACE) pump intended to deliver insulin under the skin based on input from an integrated continuous glucose monitor (iCGM) and an interoperable automated glycemic controller (iAGC), in people 6 years of age or older with diabetes mellitus. The iLet ACE Pump is intended for single-person use; it is not to be shared.

Device Description

The iLet ACE Pump is an alternate controller enabled (ACE) pump intended to deliver insulin under the skin based on input from an integrated continuous glucose monitor (iCGM) and an interoperable automated glycemic controller (iAGC) in people 6 years of age or older with diabetes mellitus.

The iLet ACE Pump provides a graphical user interface and alerts to interact with the iLet delivery system and an iAGC.

The iLet Bionic Pancreas System is a collection of wearable medical devices that work together to deliver insulin with minimal user oversight. The iLet System is made up of the iLet bionic pancreas (consisting of the iLet ACE Pump (with accessories) and iAGC which may reside on the ACE pump hardware), ACE pump disposables and accessories, iCGM and infusion set. The insulin is filled for iLet use by a user, in a ready-to-fill cartridge (from an insulin vial supplied by a drug manufacturer) with the use of the syringe and needle.

The iLet ACE Pump includes a motor-drivetrain pumping mechanism, which independently actuates the delivery of insulin from a cartridge that is separately loaded into the iLet. Insulin is injected under the skin via continuous infusion.

The iLet ACE Pump has a wirelessly rechargeable battery and is designed to be used by a single person and have a useful life of at least 4 years. The iLet is charged on a wireless charging pad which comes with the device. The Luer connector and drug cartridge need to be changed every 3 days. The insulin infusion set (including tubing and base) needs to be changed every 2-3 days for infusion sets with a teflon cannula (Convatec Inset™ and Inset™ 30) and every 1-2 days for infusion sets with a steel cannula (Convatec Contact™ Detach) as indicated in the infusion set manufacturers' labeling. The iCGM sensor needs to be changed every 10 days.

AI/ML Overview

The provided text does not describe the acceptance criteria and study that proves the device meets those criteria for an AI/ML powered device. The document pertains to the FDA clearance of the iLet ACE Pump, which is an alternate controller enabled infusion pump for insulin delivery. While it mentions the pump receives input from an integrated continuous glucose monitor (iCGM) and an interoperable automated glycemic controller (iAGC), and that the iAGC uses this information to calculate insulin needs, it does not describe an AI/ML algorithm that is being evaluated for its diagnostic or prognostic performance.

Therefore, I cannot extract the specific information requested in your bullet points related to AI/ML acceptance criteria and performance studies, such as:

  • A table of acceptance criteria and reported device performance (for an AI/ML algorithm)
  • Sample size used for the test set and data provenance (for AI/ML evaluation)
  • Number of experts used to establish ground truth and their qualifications (for AI/ML annotation)
  • Adjudication method (for AI/ML ground truth)
  • Multi-reader multi-case (MRMC) comparative effectiveness study and effect size (for AI/ML assistance)
  • Standalone performance (algorithm only without human-in-the-loop performance)
  • Type of ground truth used (for AI/ML)
  • Sample size for the training set (for AI/ML)
  • How ground truth for the training set was established (for AI/ML)

Instead, the document focuses on the regulatory clearance process for a medical device (insulin pump) and typically discusses:

  • Non-clinical performance data: This includes bench testing for dose delivery accuracy, occlusion detection, biocompatibility, sterility, electrical safety, EMC, CGM connectivity, packaging, shipping, and mechanical tests. All tests passed.
  • Data logging and Interoperability: The device was validated for logging timestamped events and interoperability with connected devices.
  • Cybersecurity: Mitigations for cybersecurity risks were incorporated, and an analysis was performed.
  • Human Factors: A human factors validation study was conducted in a simulated use condition, including training and documentation, and demonstrated the system's safety and effectiveness for intended users and uses.
  • Clinical Performance: A clinical study was not required for the iLet ACE Pump, as insulin delivery could be verified through bench testing and use validated through the human factors study.

The closest thing to "acceptance criteria" presented are the successful completion of these various non-clinical and human factors tests, indicating the device meets its design specifications and regulatory requirements for safety and effectiveness, similar to its predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 19, 2023

Beta Bionics, Inc. Vikram Verma Sr. Director, Regulatory Affairs 300 Baker Avenue, Suite 301 Concord, MA 01742

Re: K223846

Trade/Device Name: iLet® ACE Pump Regulation Number: 21 CFR 880.5730 Regulation Name: Alternate Controller Enabled Infusion Pump Regulatory Class: Class II Product Code: OFG Dated: May 5, 2023 Received: May 5, 2023

Dear Vikram Verma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-torres -S

Marianela Perez-Torres, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223846

Device Name iLet® ACE Pump

Indications for Use (Describe)

The iLet ACE Pump is an alternate controller enabled (ACE) pump intended to deliver insulin under the skin based on input from an integrated continuous glucose monitor (iCGM) and an interoperable automated glycemic controller (iAGC), in people 6 years of age or older with diabetes mellitus. The iLet ACE Pump is intended for single-person use; it is not to be shared.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K223846 510(k) Summary

Manufacturer

510(k) Owner:Beta Bionics, Inc.
Address:300 Baker Avenue, Ste. 301, Concord, MA 01742
Phone:(978) 602-6239

Contact Person

Name of Contact Person:Vikram Verma
E-mail:vverma@betabionics.com
Date of Summary Preparation:May 17, 2023

Device Names and Classification

Trade Name of Device: iLet® ACE Pump

Common Name of Device: Alternate controller enabled infusion pump (ACE pump)

Classification Name: Alternate controller enabled infusion pump (21 CFR 880.5730, QFG)

Predicate Devices

t:slim X2 insulin pump with interoperable technology (Tandem Diabetes Care, Inc., K201214)

Description of the iLet ACE Pump

The iLet ACE Pump is an alternate controller enabled (ACE) pump intended to deliver insulin under the skin based on input from an integrated continuous glucose monitor (iCGM) and an interoperable automated glycemic controller (iAGC) in people 6 years of age or older with diabetes mellitus.

The iLet ACE Pump provides a graphical user interface and alerts to interact with the iLet delivery system and an iAGC.

The iLet Bionic Pancreas System is a collection of wearable medical devices that work together to deliver insulin with minimal user oversight. The iLet System is made up of the iLet bionic pancreas (consisting of the iLet ACE Pump (with accessories) and iAGC which may reside on the ACE pump hardware), ACE pump disposables and accessories, iCGM and infusion set. The insulin is filled for iLet use by a user, in a ready-to-fill cartridge (from an insulin vial supplied by a drug manufacturer) with the use of the syringe and needle. The iLet Bionic Pancreas System components are shown in Figure 1 below.

{4}------------------------------------------------

Image /page/4/Figure/0 description: The image shows the title of a figure. The title is "Figure 1: iLet Bionic Pancreas System". The title is written in a bold, sans-serif font. The figure number is 1.

Image /page/4/Figure/1 description: The image shows a diagram of a medical device and its components. The components are labeled with letters from a to i. Some of the components include a cannula, flexible tubing, a syringe, and a Dexcom G6 sensor.

The iLet Bionic Pancreas System: (a) iLet ACE Pump; (b) iLet Cartridge (ready to fill cartridge); (c) Filling Syringe (Becton Dickinson); (d) Needle (Becton Dickinson); (e) iLet Connect (Luer connection adapter); (f) Infusion Set Base and infusion tube (Convatec) (g) Insulin Infusion Set (Convatec), (h) iCGM (Dexcom G6 sensor and transmitter), and (i) iLet Charger (charging pad, micro-USB cable with power adapter)

For a better understanding of how the iLet bionic pancreas is used spatially on a person, it is shown applied to a human body in Figure 2 below. The iCGM is shown communicating with the iLet via Bluetooth. The iLet ACE Pump gets glucose readings from the iCGM every 5 minutes and the iAGC uses that information as one of the inputs to calculate the person's insulin needs.

{5}------------------------------------------------

Image /page/5/Figure/0 description: The image shows a diagram of a continuous glucose monitoring (CGM) system. The system consists of a CGM sensor, an insulin set cannula, and an iLet device. The CGM sensor continuously measures blood glucose levels, and the iLet device receives input from the CGM and uses a control algorithm to analyze and regulate blood glucose levels. The insulin set cannula is a tiny tube that is inserted under the skin to deliver insulin.

Figure 2: iLet Bionic Pancreas on a Person with Diabetes

The iLet ACE Pump includes a motor-drivetrain pumping mechanism, which independently actuates the delivery of insulin from a cartridge that is separately loaded into the iLet. Insulin is injected under the skin via continuous infusion. The figure above shows insulin being injected from the iLet through an infusion set. The infusion set must be placed at least 3 inches away from the iCGM sensor.

The iLet ACE Pump has a wirelessly rechargeable battery and is designed to be used by a single person and have a useful life of at least 4 years. The iLet is charged on a wireless charging pad which comes with the device. The Luer connector and drug cartridge need to be changed every 3 days. The insulin infusion set (including tubing and base) needs to be changed every 2-3 days for infusion sets with a teflon cannula (Convatec Inset™ and Inset™ 30) and every 1-2 days for infusion sets with a steel cannula (Convatec Contact™ Detach) as indicated in the infusion set manufacturers' labeling. The iCGM sensor needs to be changed every 10 days.

Indications for Use for iLet ACE Pump

The iLet ACE Pump is an alternate controller enabled (ACE) pump intended to deliver insulin under the skin based on input from an integrated continuous glucose monitor (iCGM) and an interoperable automated glycemic controller (iAGC), in people 6 years of age or older with diabetes mellitus. The iLet ACE Pump is intended for single-person use; it is not to be shared.

{6}------------------------------------------------

Intended Use, Technological and Performance Characteristics of Subject Device Compared with Predicate Device

Element ofComparisonTandem t:slim X2 Pump(Predicate Device) (K201214)iLet ACE Pump(Subject Device)Comment
Intended UseAn ACE pump which is intendedto work with an iCGM and iAGCto deliver insulin subcutaneouslyfor the management of diabetesmellitusAn ACE pump which is intendedto work with an iCGM and iAGCto deliver insulin subcutaneouslyfor the management of diabetesmellitusIdentical
Pump TypeAlternate controller enabled(ACE) infusion pumpAlternate controller enabled(ACE) infusion pumpIdentical
Specific Drug /Biologic UseU-100 InsulinIndicated for use with NovoLogor HumalogU-100 InsulinSystem tested with NovoLog,HumalogSimilar
PrescriptionStatusPrescription DevicePrescription DeviceIdentical
Size7.95 cm L x 5.08 cm W x 1.52 cmH9.10 cm L x 5.90 cm W x 1.50cm HSimilar
Weight3.95 ounces (112 grams)110 grams (without infusion set)Similar
OperatingConditionsTemperature: 41°F (5°C) to98.6°F (37°C)Humidity: 20% to 90% RH non-condensingTemperature: 41°F (5°C) to104°F (40°C)Humidity: 15% to 90% RH non-condensingSimilar
AtmosphericPressure-1,300 feet to 10,000 feet (-396meters to 3,048 meters)15.4 to 10.2 psia (Relativealtitude -1300 feet to 10,000 feet)Identical
MoistureProtectionIPX7: Watertight to a depth of 3feet (0.91 meters) for up to 30minutesIPX8: Protected againstimmersion in water for up to 12feet for 30 minutesSimilar
Maximum BasalRate15 units/hr.0 - 11.5 units/hrBasal rate in subjectdevice is internal andcannot be set.Effective Rates areprovidedcorresponding toisolated basal doses
PowerRequirementsRechargeable lithium batterypowered device, USB-A wiredcharging through a DC AdapterRechargeable lithium batterypowered device, wirelesscharging through a charging padconnected to a DC AdapterSimilar

Table 1: iLet ACE Pump Comparison with Predicate Device

Hazard Analysis, Risk Mitigation

A hazard analysis was conducted to account for the unique design elements, intended use, and risks of the iLet bionic pancreas. The hazard analysis accounted for the risks associated with interoperability between the device and other third-party digital devices which met predefined criteria but were not specifically identified, including scenarios in which the device was put into an environment in which both compatible and incompatible digital devices attempted to communicate with the device and deliver commands. This analysis identified hazards which could reasonably be anticipated to impact the proper use of the device, traced all identified risks

{7}------------------------------------------------

to adequate design controls, and demonstrated that design features were appropriately implemented and validated.

In addition, a Use Related Risk Analysis (URRA) was conducted that identifies all User Groups, User Tasks, Possible Use Errors, Potential Clinical Harm to Patient, Severity of Harm, Risk Mitigations and Validation Methods associated with use of the iLet device. Critical tasks per the FDA's guidance document Applying Human Factors and Usability Engineering to Medical Devices 2016 are identified within the URRA. The URRA includes all warnings, cautions, and contraindications.

Summary of Non-Clinical Performance Data and Conclusions including compliance with Special Controls

Analytical Performance

The iLet ACE Pump was tested for dose delivery accuracy and occlusion detection. All tests passed.

Other Supportive Test Data

  • Biocompatibility, Sterility, Insulin Compatibility and Stability, Electrical EMC and . Safety, CGM connectivity, Packaging/ Shipping Integrity and Mechanical Tests:
    The iLet ACE Pump and accessories were subjected to the above tests as applicable. All tests passed.

  • Data logging and Interoperability: ●
    The iLet ACE Pump has been validated for logging timestamped events, including information related to its state, user inputs, and device settings, as required by special controls. All tests passed.

The iLet ACE Pump software has been validated to be interoperable with all connected devices.

  • Cybersecurity:
    The iLet ACE Pump has incorporated adequate mitigations for cybersecurity risks. A cybersecurity analysis was performed using the draft FDA guidance, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices – Issued October 18, 2020, and the principles outlined in the FDA guidance. Postmarket Management of Cybersecurity in Medical Devices - Issued December 28, 2016. Beta Bionics has provided a software bill of materials and penetration testing.

  • Labeling and Training: ●
    The iLet bionic pancreas device labeling and training for users and healthcare

{8}------------------------------------------------

practitioners was reviewed by the FDA. Labeling is sufficient and satisfies applicable requirements of 21 CFR 801 and 21 CFR 809.

Special Controls

The iLet bionic pancreas was found to be compliant with all Special Controls for Alternate controller enabled infusion pump (21 CFR 880.5730, QFG).

Summary of Human Factors and Clinical Performance Data and Conclusions

Human Factors

Beta Bionics executed a human factors and usability engineering process that followed and complied with FDA-recognized standards IEC62366:2015-1 and HE75:2009 as well as the FDA's guidance document. Applying Human Factors and Usability Engineering to Medical Devices - Issued February 3, 2016. Human Factors validation study testing was conducted with the iLet bionic pancreas (iLet ACE Pump with the iLet Dosing Decision Software installed) in a simulated use condition, including associated training and accompanying documentation. All critical tasks were tested in the validation study. The results of the validation study demonstrated the iLet system has been found to be safe and effective for the intended users for its intended uses in its intended use environment.

Clinical Performance

A clinical study was not required for the iLet ACE Pump as insulin delivery from the device can be verified through bench performance testing and use of the device was validated through a human factors validation study (described above).

Conclusions drawn from non-clinical tests that demonstrate safety, effectiveness, and performance to be as well as or better than predicate device

The preceding summary of predicate device comparison, non-clinical bench testing which supports the subject device design, and human factors study testing, demonstrate that the subject device performance in terms of safety and effectiveness is comparable to the predicate device.

§ 880.5730 Alternate controller enabled infusion pump.

(a)
Identification. An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions.
(A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes (
e.g., head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices.
(ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration (
e.g., occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability (
e.g., temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low.
(v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance.
(vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment.
(vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated.
(viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing.
(2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented.
(ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to external devices.
(ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices.
(iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers (
e.g., battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified.
(v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol.
(4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption (
e.g., reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (e.g., commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include:
(i) A record of all drug delivery
(ii) Commands issued to the pump and pump confirmations
(iii) Device malfunctions
(iv) Alarms and alerts and associated acknowledgements
(v) Connectivity events (
e.g., establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(7) Device labeling must include the following:
(i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate.
(ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters.
(iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is 250 percent of the commanded amount.
(iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point.
(v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps.
(vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.