K Number
K223810

Validate with FDA (Live)

Date Cleared
2023-02-24

(66 days)

Product Code
Regulation Number
862.1675
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Female luer type
· The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with syringes when transferring blood specimen from a syringe to blood collection tube or blood culture bottle.

Male luer type
· The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube.

Device Description

The Kawasumi Multiple Sample Adapter with Pre-attached Holder (MBCH) is a sterile, single use device consisting of a plastic holder with a non-patient contacting stainless steel cannula covered with a rubber sheath and a luer adapter. There are two types of the device with difference in luer adapter type; male luer adapter type and female luer adapter type. The predicate device (K190485) is the male luer type and the proposed device is the female luer type, and the subject of this special 510(k) premarket notification submission.

  • The male luer adapter type is used to collect blood specimen into blood collection tube by . connecting to female luer connectors of vascular access devices such as peripheral, central catheter, Huber needle, etc. (predicate [Item number MBCH-01])
  • The female luer adapter type is used to transfer blood from a syringe into blood collection tube . or blood culture bottle. After drawing patient blood into a syringe, connect the female luer adapter of the device to the syringe. When inserting a tube, the blood is transferred into the tube using tube's vacuum. (subject [Item number MBCH-02])

The device is sterilized using Ethylene Oxide. The device is a prescription-use device intended to be used in hospitals or healthcare facilities.

AI/ML Overview

The provided text describes a 510(k) summary for the Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH), specifically focusing on the newly introduced female luer type (MBCH-02) and comparing it to the legally marketed male luer type (MBCH-01) as the predicate device. The information details the non-clinical testing performed to establish substantial equivalence.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Test ItemAcceptance Criteria (Referenced Standard)Reported Device Performance
Luer connector evaluationISO 80369-7:2021PASS
Simulation testIn-housePASS
Physical testISO 1135-3:2016, in-housePASS
Particulate contaminationUSP <788>PASS
Chemical testISO 1135-3:2016PASS
Usability testISO 14971:2019, IEC 62366-1:2015 + Amd.1:2020PASS
EtO residualISO 10993-7:2008/Amd.1:2019PASS
Bacterial endotoxinUSP <85>PASS
Sterility testUSP <71>PASS
Transportation & PackageISO 11607-1:2019, ISO 11607-2:2019, ASTM D4169-16PASS

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes (number of devices tested) for each non-clinical test. The testing is described as "Non-Clinical Performance Testing," suggesting laboratory-based assessments rather than clinical studies with patient data. Therefore, there is no information on data provenance in terms of country of origin or retrospective/prospective nature.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to the provided document. The device is a blood specimen collection device, and the testing described is non-clinical performance testing against established standards (e.g., ISO, USP, ASTM, IEC). There is no mention of experts establishing a "ground truth" related to medical diagnosis or interpretation for this type of device and testing.

4. Adjudication method for the test set

This information is not applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where multiple readers or experts evaluate cases to establish a consensus-based ground truth. The presented tests are objective, laboratory-based performance assessments.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices, which is not the nature of the Kawasumi Multiple Sample Adapter.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical medical instrument (blood specimen collection adapter), not an algorithm or AI system.

7. The type of ground truth used

The "ground truth" for the non-clinical performance tests is defined by the referenced industry standards and in-house specifications (e.g., ISO 80369-7:2021 for luer connectors, USP <788> for particulate contamination). The device's performance is measured against the criteria established in these standards.

8. The sample size for the training set

This information is not applicable. The device is not an AI/ML model that requires a training set.

9. How the ground truth for the training set was established
This information is not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 24, 2023

SB-Kawasumi Laboratories, Inc. % Valerie Followell Regulatory Consultant Regulatory Compliance Associates, Inc. (RCA) 10411 Corporate Drive, Suite 102 Pleasant Prairie, Wisconsin 53158

Re: K223810

Trade/Device Name: Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: January 27, 2022 Received: January 27, 2022

Dear Valerie Followell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Danil Walloschek

David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223810

Device Name

Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH)

Indications for Use (Describe)

Female luer type

· The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with syringes when transferring blood specimen from a syringe to blood culture bottle.

Male luer type

· The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K223810 - 510(K) Summary

Manufacturer: SB-KAWASUMI LABORATORIES, INC.

Address:

3-25-4, Tonomachi, Kawasaki-ku Kawasaki-shi. Kanagawa 210-8602, JAPAN

Corresponding Official/Contact: Valerie Followell Regulatory Consultant Telephone Number: 847-400-6187 Email: v.followell@rcainc.com

Summary Date: Februarv 24, 2023

Trade Name: Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH)

Common or Usual Name: Blood Collection Tubes, Vials, Systems, Serum Separators

Regulation Number: 21 CFR 862.1675

Regulation Name: Blood specimen collection device

Product Code: JKA

Class: Class 2

Panel: Clinical Chemistry

Predicate Device: Kawasumi Multiple Sample Adapter with Pre-attached Holder has only one configuration with Male Luer Adapter (K190485)

Device Description:

The Kawasumi Multiple Sample Adapter with Pre-attached Holder (MBCH) is a sterile, single use device consisting of a plastic holder with a non-patient contacting stainless steel cannula covered with a rubber sheath and a luer adapter. There are two types of the device with difference in luer adapter type; male luer adapter type and female luer adapter type. The predicate device (K190485) is the male luer type and the proposed device is the female luer type, and the subject of this special 510(k) premarket notification submission.

  • The male luer adapter type is used to collect blood specimen into blood collection tube by . connecting to female luer connectors of vascular access devices such as peripheral, central catheter, Huber needle, etc. (predicate [Item number MBCH-01])
  • The female luer adapter type is used to transfer blood from a syringe into blood collection tube . or blood culture bottle. After drawing patient blood into a syringe, connect the female luer adapter of the device to the syringe. When inserting a tube, the blood is transferred into the tube using tube's vacuum. (subject [Item number MBCH-02])

The device is sterilized using Ethylene Oxide. The device is a prescription-use device intended to be used in hospitals or healthcare facilities.

{4}------------------------------------------------

Indications for Use:

Female luer type

· The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with syringes when transferring blood specimen from a syringe to blood collection tube or blood culture bottle.

Male luer type

· The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube.

510(k) Summary Device Comparison Table
Feature of the DeviceSubject Device:Predicate Device:Discussion/Comment:
SB-KAWASUMI LABORATORIES, INC.Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH)Kawasumi Laboratories, Inc.Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH)The company name has been changed. Statement included in submission.
Item number (Model number)MBCH-01 and MBCH-02MBCH-01DifferentThe new item number MBCH-02 is added.
510(k) NumberK223810K190485-
Product CodeJKAJKASame
Classification21 CFR 862.167521 CFR 862.1675Same
Classification NameTubes, Vials, Systems, Serum Separators, Blood CollectionTubes, Vials, Systems, Serum Separators, Blood CollectionSame
Common NameBlood Specimen Collection DeviceBlood Specimen Collection DeviceSame
Rx or OTCRx onlyRx OnlySame
Number of UsesSingle-UseSingle-UseSame
Use EnvironmentHospital and/or Healthcare FacilityHospital and/or Healthcare FacilitySame
Indications for UseFemale luer type The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with syringes when transferring blood specimen from a syringe to blood collection tube or blood culture bottle. Male luer type The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube.The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube.DifferentAlthough there is a difference in where to connect the device, both types are used for collecting blood specimen to blood collection tube or blood culture bottle.
Intended UseSame as the Indications for UseSame as the Indications for UseSame
Design FeatureGauge: 20GLength (mm): $64.1 \pm 0.3$Width (mm): $31 \pm 0.3 \times 23.5 \pm 0.2$Gauge: 20GLength(mm): $64.1 \pm 0.3$Width(mm): $31 \pm 0.3 \times 23.5 \pm 0.2$Same
Luer AdapterTypeFemaleMaleDifferentDifferences do notchange assertion ofequivalence of devices
Number of UsesSingle-Use Rx OnlySingle-Use Rx OnlySame
MaterialsHubPolycarbonate (PC)Polycarbonate (PC)DifferentMaterial is identical withthe exception of its color.Supportive informationincluded in submission.
CannulaStainless SteelStainless SteelSame
GlueEpoxyEpoxySame
SiliconeSilicone OilSilicone OilDifferentSupportive informationincluded in submission.
SheathIsoprene RubberNot made with Natural RubberLatexIsoprene RubberNot made with Natural RubberLatexSame
HolderPolypropylene (PP)Polypropylene (PP)Same
BiocompatibilityComplies with ISO 10993Complies with ISO 10993Same
SterilizationMethodEthylene OxideEthylene OxideSame
Shelf Life3 years3 yearsSame

{5}------------------------------------------------

{6}------------------------------------------------

Technological Characteristics:

The Subject Device and the Predicate Device have similar technological characteristics. Both devices are sterile, single-use, non-invasive devices with a Pre-Attached Holder used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube.

The Kawasumi Multiple Sample Adaptor with Pre-Attached Holder (Subject Device [MBCH-02]) is substantially equivalent to the Predicate Device [MBCH-01] with regard to technologic characteristics, materials, performance and intended use.

Non-Clinical Testing:

Non-Clinical Performance Testing was performed to ensure that the device meets design requirements and specifications and to confirm performance of the Kawasumi Multiple Sample Adaptor with Pre-Attached Holder [MBCH-02].

Test ItemReferenced standardResult
Performance & Chemical Testing
Luer connector evaluationISO 80369-7:2021PASS
Simulation testIn-housePASS
Physical testISO 1135-3:2016, in-housePASS
Particulate contaminationUSP<788>PASS
Chemical testISO 1135-3:2016PASS
Usability testISO 14971:2019IEC 62366-1:2015 + Amd.1:2020PASS
Sterilization, Transportation & Packaging
EtO residualISO 10993-7:2008/Amd.1:2019PASS
Bacterial endotoxinUSP <85>PASS
Sterility testUSP <71>PASS
Transportation & PackageISO 11607-1:2019ISO 11607-2:2019ASTM D4169-16PASS

Conclusion:

Based on the Non-Clinical Performance Testing conducted on the subject device, intended use, and principles of operation it may be concluded that the Kawasumi Multiple Sample Adapter with Pre-Attached Holder (Subject Device [MBCH-02]) is substantially equivalent to the legally marketed Predicate Device [MBCH-01].

§ 862.1675 Blood specimen collection device.

(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.