K Number
K223780

Validate with FDA (Live)

Device Name
Lumos 3DX
Manufacturer
Date Cleared
2023-07-06

(202 days)

Product Code
Regulation Number
872.1800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Lumos 3DX ™ System is an extraoral X-ray source (intraoral X-ray detection) dental X-ray system for producing diagnostic dental radiographs of the teeth, jav, and other oral structures. The system provides 3D imaging for diagnostic purposes via tomosynthesis. For use on adult patients only.

Device Description

The Lumos 3DX system is a 3D dental chairside X-ray system on a mobile stand that uses tomosynthesis (limited-angle tomography) to generate 3D images of the teeth and surrounding structures. A custom digital sensor is placed in the patient's mouth and multiple images are acquired around an arc of rotation and then fed into a 3D reconstruction algorithm. The digital sensor provided with the system is specifically for use with the Lumos 3DX device. The sensor has a higher frame rate that enables the Lumos 3DX system to capture 30 images in a short period of time. A server is provided as part of the Lumos 3DX system and wirelessly connects to one or more Lumos 3DX systems. After capturing images, the Lumos 3DX wirelessly transfers them to the server for reconstruction. This server can then be connected to the local network for access of the reconstructed data.

AI/ML Overview

The Lumos 3DX System is a dental X-ray system that provides 3D imaging for diagnostic purposes via tomosynthesis, for use on adult patients only. Its acceptance criteria and performance are as follows:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Image Quality PerformanceDemonstrated high spatial resolution, good contrast, and low noise using image quality phantoms. The 3D volume voxel size was also verified.
Patient MotionEvaluation showed typical patient movements during imaging do not significantly affect the system or the resulting 3D volume. This justified the use of dental phantoms and cadaver subjects for clinical image quality analysis.
System Verification & ValidationHazard mitigation performed, demonstrating the Lumos 3DX meets design input and user needs.
Compliance to StandardsComplies with applicable IEC series of X-ray performance standards, including IEC60601-2-65. Meets all applicable 21CFR Subchapter J performance standards (radiation safety, dosimetry, leakage, stray radiation).
Clinical Diagnostic QualityImages obtained from dental phantoms with human teeth and simulated bone, as well as cadaver subjects, were evaluated by dental professionals and found to be of diagnostic quality for clinical use.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: "A number of 3D images" were obtained. Specific numerical sample sizes for the image quality phantoms, dental phantoms, and cadaver subjects are not provided in the document.
  • Data Provenance: The document does not explicitly state the country of origin. The test set primarily used dental phantoms with human teeth and simulated bone, as well as cadaver subjects for the clinical imaging evaluation. The study appears to be retrospective in the sense that existing phantoms and cadavers were used, but images were prospectively acquired using the Lumos 3DX system specifically for this evaluation.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Number of Experts: Not specified. The document mentions "dental professionals" were used for the evaluation.
  • Qualifications of Experts: The document states "dental professionals" were involved. Specific qualifications (e.g., years of experience, specialization like radiologist) are not provided.

4. Adjudication Method for the Test Set:

  • The document implies a consensus or evaluation by "dental professionals." However, a specific adjudication method (e.g., 2+1, 3+1) is not explicitly stated. The statement "Based on the evaluations made by the dental professionals, the images obtained with the Lumos 3DX were of diagnostic quality for clinical use" suggests a qualitative assessment.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with AI assistance versus without AI assistance was not conducted or reported in the provided text. The study primarily focuses on the standalone performance of the device's image quality for diagnostic use.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

  • Yes, the clinical utility of the Lumos 3DX was demonstrated through a "Clinical Imaging Evaluation." This evaluation assessed the images generated by the Lumos 3DX system itself (the algorithm/system only) for diagnostic quality, which dental professionals then reviewed. While human reviewers assessed the output, the performance being tested was that of the device generating the images, not an AI aiding human diagnosis.

7. The Type of Ground Truth Used:

  • Image Quality Phantoms: These phantoms inherently provide a known, objective ground truth for metrics like spatial resolution, contrast, and noise.
  • Dental Phantoms with Human Teeth and Simulated Bone, and Cadaver Subjects: For the clinical imaging evaluation, these served as the ground truth. The structures within these phantoms and cadavers represent actual anatomical conditions, and their diagnostic quality was assessed by dental professionals. This can be considered a form of "expert consensus" on the diagnostic utility relative to a known anatomical state.

8. The Sample Size for the Training Set:

  • The document does not provide any information regarding a training set size. The context is a 510(k) summary for a medical device (an X-ray system), not an AI/ML algorithm submission that typically requires specific training set details. The "reconstruction algorithm" mentioned is likely a traditional image processing algorithm rather than a machine learning model requiring a distinct training phase.

9. How the Ground Truth for the Training Set Was Established:

  • As no training set is mentioned for an AI/ML algorithm, this information is not applicable and not provided in the document.

{0}------------------------------------------------

July 6, 2023

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

3Dio % Mr. Douglas Hansen President 1119 South 1680 West OREM UT 84058

Re: K223780

Trade/Device Name: Lumos 3DXTM Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: EHD Dated: June 1, 2023 Received: June 5, 2023

Dear Mr. Hansen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223780

Device Name Lumos 3DX тм

Indications for Use (Describe)

The Lumos 3DX ™ System is an extraoral X-ray source (intraoral X-ray detection) dental X-ray system for producing diagnostic dental radiographs of the teeth, jav, and other oral structures. The system provides 3D imaging for diagnostic purposes via tomosynthesis. For use on adult patients only.

Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)[ ] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for 3Dio, a Turner Company. The logo consists of a stylized tooth inside of a circle on the left, followed by the text "3Dio" in a sans-serif font. Below the text "3Dio" is the text "a Turner Company" in a smaller font. The tooth and the text "3Dio" are in a light teal color, while the circle and the text "a Turner Company" are in a light gray color.

510k #: K223780

Lumos 3DX X-ray Imaging System

12 June 2023

Submitter:

Name:3Dio, LLC.
Address:1119 S 1680 WOrem, UT 84058
Official Correspondent:Doug Hansen, CEO and President
Telephone No:801-796-2951
Email:dhansen@3dteeth.com
Proposed Device:
Trade Name:Lumos 3DX™
Class:II
Common/Usual Name:Dental X-ray System, Mobile
Classification Name:Extraoral source x-ray system
Primary Product Code:EHD
Regulatory Standard:21CFR 872.1800
Predicate Device:
Manufacturer:Surround Medical Systems
Trade Name:Portray System
510(k):K211014
Class:II
Common/Usual Name:Extraoral source x-ray system
Primary Product Code:EHD
Regulatory Standard:21CFR 872.1800

Description: The Lumos 3DX system is a 3D dental chairside X-ray system on a mobile stand that
uses tomosynthesis (limited-angle tomography) to generate 3D images of the teeth and
surrounding structures. A custom digital sensor is placed in the patient's mouth and
multiple images are acquired around an arc of rotation and then fed into a 3D
reconstruction algorithm.

The digital sensor provided with the system is specifically for use with the Lumos 3DX
device. The sensor has a higher frame rate that enables the Lumos 3DX system to

capture 30 images in a short period of time.

A server is provided as part of the Lumos 3DX system and wirelessly connects to one or more Lumos 3DX systems. After capturing images, the Lumos 3DX wirelessly transfers them to the server for reconstruction. This server can then be connected to the local network for access of the reconstructed data.

  • Indications for Use: The Lumos 3DX™ System is an extraoral X-ray source (intraoral X-ray detection) dental X-ray system for producing diagnostic dental radiographs of the teeth, jaw, and other oral structures. The system provides 3D imaging for diagnostic purposes via tomosynthesis. For use on adult patients only.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for 3Dio, a Turner Company. The logo consists of a circular icon with a tooth inside, colored in shades of gray and teal. To the right of the icon, the text "3Dio" is written in a teal sans-serif font, with the words "a Turner Company" appearing in a smaller font size below.

Document Title: Lumos 3DX 510(k) Summary

510k #: K223780

Page 2 of 5

Table 1. Comparison with predicate device:

Characteristicor PropertySurround Medical, Portray System,K2110143Dio Lumos 3DX™
ClassificationRegulation number: 21 CFR 872.1800Regulation name: Extraoral source x-ray systemRegulatory Class: IIProduct Code: EHDRegulation number: 21 CFR 872.1800Regulation name: Extraoral source x-raysystemRegulatory Class: IIProduct Code: EHD
Intended UseThe Portray System is an extraoral X-ray source (intraoral X-ray detection)dental X-ray system for producingdiagnostic dental radiographs of theteeth, jaw and other oral structures. Thesystem provides 2D imaging fordiagnostic purposes and 3D imaging asan adjunctive tool.The Lumos 3DX™ System is an extraoralX-ray source (intraoral X-ray detection)dental X-ray system for producingdiagnostic dental radiographs of the teeth,jaw, and other oral structures. The systemprovides 3D imaging for diagnosticpurposes via tomosynthesis. For use onadult patients only.
TargetAnatomical SiteOral CavityOral Cavity
Principle of UseX-ray tubeX-ray tube
Electrical:
PowerMust be plugged in to AC MainsMust be plugged in to AC Mains
X-ray SourceAnode: TungstenCathode: Carbon NanotubeFocal spot: 0.7mmTube Voltage: 60 or 70 kVCurrent: 7 mAExposure time: 0.6 Sec (2D & 3D)Tube Anode Angle: 12°Anode: TungstenCathode: Tungsten FilamentFocal spot: 0.4mmTube Voltage range: 60 to 70 kVCurrent range: 3 mA to 7mAExposure Time: 0.063 – 1.0 sec (3D)Tube Anode Angle: 12°
LeakageRadiation< 0.25 mGy/h (@ 1 m)< 0.25 mGy/h (@ 1 m)
Exposure TypeMulti-beam stationarySingle beam pulsed
Exposure TimesMicroprocessor controlled exposuretimesMicroprocessor controlled exposure times
Exposure ModesPreset loading factors or manual modePreset loading factors or manual mode
SelectableparameterPatient type (adult/child), anatomicalpositions, 2D or 3D mode, x-ray tubevoltage, exposure timeAnatomical positions, 3D mode, x-ray tubevoltage & current
Image type2D diagnostic and/or 3D adjunct3D Diagnostic with 2D slices
Image dataacquired7 images30 images
DetectorDigital SensorFrame Rate: < 2 images per secondXray Sensitivity: Standard dentaldetector X-ray sensitivityCustom #2 intra-oral digital sensor (19.5μm x 19.5 μm pixel size) Frame Rate: ≥ 9.0images per secondXray Sensitivity: High-sensitivity for low-dose rapid 3D data acquisition
Characteristicor PropertySurround Medical, Portray System,K2110143Dio Lumos 3DX
StandardsIEC 60601-1IEC 60601-1-2IEC 60601-1-3IEC 60601-1-6IEC 60601-2-65IEC 61233-3-4IEC 62304IEC 60601-1IEC 60601-1-2IEC 60601-1-3IEC 60601-1-6IEC 60601-2-65IEC 61223-3-4IEC 62304IEC 62366-1
Mechanical/Physical:
PhysicalDimensionsHead:Length: ~49 cm (~19 in)Width: ~20 cm (~8 in)Height: ~18 cm (~7 in)Arm:Vertical Reach: ±19.5" from neutralHorizontal Reach: 41.1" extensionarm reach with 177° rotation.The articulating arm has 42.7”of additional extension with205° of rotation.X-Ray Head:Length: 34.5 cm (13.6 in)Width: 38.4 cm (15.1 in)Height: 31.8 cm (12.5 in)Arm:Vertical Reach: ±18" from neutralHorizontal Reach: The articulating armhas 40" of extension with 90° ofrotation.Stand:Height with folded arm: 200 cm (78.7 in)Width of base: 54.0 cm (21.2 in)Length of base: 50.8 cm (20.0 in)
Source toDetectordistance (SDD)400 mm313 mm
MinimumSource to SkinDistance (SSD)308 mm200 mm
Weight112.5 lbs117.0 lbs
Imaging, Display, and Software:
Sensor PhysicalDimensionsExterior Size: 41.76 mm x 30.42 mmImaging Size: 35.92 mm x 25.82 mmPixel Size: 19.5 µm x 19.5 µmImage Resolution: 1324 x 1842Exterior Size: 41.82 mm x 30.50 mmImaging Size: 35.98 mm x 26.25 mmPixel Size: 19.5 µm x 19.5 µmImage Resolution: 1346 x 1845
SensorTechnologyCMOS with Cesium Iodide ScintillatorCMOS with Cesium Iodide Scintillator
Dimensions ofX-ray beam atImaging Plate36.6 mm x 33.6 mm Rectangle63.6 mm x 59.7 mm Ellipse
Data TransferRate<0.5 Frames/sec< 10.2 Frames/sec
Characteristicor PropertySurround Medical, Portray System,K2110143Dio Lumos 3DX
X-ray signalGain (WRTstandard)1XAbout 9X
X-ray emissioncontrolWired ControlWired control
3D dataacquisition time6 sec (approximately)3 sec (approximately)
3D slicethickness0.5 mm fixedVariable down to <0.1 mm
3D volumerenderingNoYes
2D ImageSynthetic2D Slices
AcquisitionGeometryArc segment360° Conical
InstallationConfigurationWall MountMobile Stand
SoftwareWindows operating system andWindows-like user interface.Windows operating system withtouchscreen user interface.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for 3Dio, a Turner Company. The logo consists of a circular icon on the left, featuring a stylized tooth shape in white against a background of gray and light blue. To the right of the icon is the text "3Dio" in a light blue sans-serif font, with the words "a Turner Company" appearing in a smaller font size below.

Document Title: Lumos 3DX 510(k) Summary

510k #: K223780

Page 3 of 5

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the 3Dio logo. The logo consists of a tooth inside of a circle on the left and the text "3Dio a Turner Company" on the right. The tooth is white and the circle is gray and teal. The text is teal.

Document Title: Lumos 3DX 510(k) Summary

510k #: K223780

Page 4 of 5

The Lumos 3DX system and the Portray System are both intra-oral x-ray systems that Discussion: take multiple images and render a finished image for viewing. The systems view these images in slices that allow the dental professionals to see the teeth at various depths. The Portray system uses carbon nanotubes in its X-ray source that are positioned along an arc to get their seven shots from different angles. The Lumos 3DX uses a conventional X-ray Tube that pulses 30 times while rotating along a 360° circular path.

The Lumos 3DX and predicate device both allow for viewing slices, but the thickness is much smaller in the Lumos 3DX system which allows for a more detailed viewing process for the dental professional. In addition, because the Lumos 3DX takes images around a circle rather than along a single plane it provides more spatial information for reconstruction. The Portray System is a wall-mounted system while the Lumos 3DX system is on a stand, but both have articulating arms. In both systems, an alignment aid is used to hold the detector in the correct position relative to the x-ray source and that alignment aid is magnetically attached to the front of the nose cone.

In dental imaging Panoramic and Cone Beam CT systems are used to obtain 3D images of patient's oral structures. This is possible because the x-ray source and detector are external and rotate around the patient's head. These systems are large and expose the patient to a significant amount of radiation. The Lumos 3DX system is a smaller system that can take images of a more focused area using less radiation. Its portability also allows it to be used for unconscious patients while they lay in the chair rather than needing to move them to the x-ray device.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the title of a document. The title is "Document Title: Lumos 3DX 510(k) Summary". The text is in a clear, sans-serif font and is easy to read. The title suggests that the document is a summary of the Lumos 3DX 510(k) application.

Image /page/7/Picture/1 description: The image is a logo for 3Dio, a Turner Company. The logo features a stylized tooth inside of a circle that is split into two colors, gray and light blue. To the right of the circle is the text "3Dio" in a light blue color, with the words "a Turner Company" underneath in a smaller font.

510k #: K223780

Bench and Radiation Safety

Testing Summary: An Image Quality Performance test was completed using image quality phantoms for spatial resolution, contrast, and noise. The 3D volume voxel size was also verified.

The effect of patient motion was evaluated to verify that typical movements induced by a patient during imaging do not significantly affect the system or the resulting 3D volume. The results of the patient motion studies justify the use of dental phantoms and cadaver subjects for the analysis of clinical image quality.

In addition to the image quality bench studies, system verification and validation testing including hazard mitigation has been performed to demonstrate the Lumos 3DX meets design input and user needs.

The Lumos 3DX has been tested to show compliance with the applicable IEC series of x-ray performance standards, including IEC60601-2-65. It also meets all applicable 21CFR Subchapter J performance standards including those for radiation safety, such as dosimetry, leakage, and stray radiation.

Clinical Image Quality

Testing Summary: The clinical utility of the Lumos 3DX was demonstrated by performing a Clinical Imaging Evaluation with dental professionals. Dental phantoms with human teeth and simulated bone as well as Cadaver subjects were selected to represent typical use cases for the Lumos 3DX. A number of 3D images were obtained for analysis and review.

Based on the evaluations made by the dental professionals, the images obtained with the Lumos 3DX were of diagnostic quality for clinical use.

  • Based on successful verification and validation testing, conformance to recognized Conclusion: performance standards and FDA guidance, and development under the 3Dio Quality Management System, we conclude that the Lumos 3DX Imaging System is substantially equivalent to the predicate device Surround Medical, Portray System (K211014) and is safe and effective for its intended use.

§ 872.1800 Extraoral source x-ray system.

(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.