K Number
K223768

Validate with FDA (Live)

Device Name
RadiForce RX270
Manufacturer
Date Cleared
2023-04-20

(126 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This Product is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

Device Description

RadiForce RX270 is a color LCD monitor for viewing medical images other than those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 1,200 x 1,600 pixels (2MP) with a pixel pitch of 0.270 mm.

Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used.

There are two model variations. RX270 and RX270-AR. The difference of the two variations is the surface treatment of the display screens; the surface treatment of the RX270 is Anti-Glare (AG) treatment and that of the RX270-AR is Anti-Reflection (AR) coating.

RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including the RadiForce RX270 based on the OC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS and its subset, RadiCS LE, are included in this 510(k) submission as an accessory to the RadiForce RX270.

RadiCS is of Minor level of concern and that it's being used unchanged from the predicate software. RadiCS supports the functions of the monitor RadiForce RX270 and it's not a medical imaging software.

AI/ML Overview

This document does not contain information about an AI/ML device or its performance criteria. The device described, RadiForce RX270 and RX270-AR, is a medical grade display monitor. The submission states that the device is "indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography."

The performance testing section (Section 7) describes bench tests performed on the display monitor to evaluate its characteristics, such as spatial resolution (MTF), pixel defects, artifacts, temporal response, luminance, and conformance to DICOM GSDF. These tests are standard for evaluating medical displays and ensure the monitor can accurately present radiological images.

Therefore, I cannot provide the requested information regarding acceptance criteria and study details for an AI-powered device, a multi-reader multi-case (MRMC) study, or standalone algorithm performance. The document focuses on the technical specifications and performance validation of a display device, not an AI or imaging analysis product.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.

EIZO Corporation % Hiroaki Hashimoto Senior Manager 153 Shimokashiwano Hakusan, Ishikawa 924-8566 Japan

April 20, 2023

Re: K223768

Trade/Device Name: RadiForce RX270 and RX270-AR Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: PGY Dated: January 25, 2023 Received: January 30, 2023

Dear Hiroaki Hashimoto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223768

Device Name RadiForce RX270, RX270-AR

Indications for Use (Describe)

This Product is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

Type of Use (Select one or both, as applicable)

Recreational (Day Use Only, Subject to Closure) Overnight Camping (Not Subject to Closure)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for EIZO. The logo consists of a stylized, geometric shape on the left and the word "EIZO" in bold, sans-serif font on the right. The geometric shape is a square with a checkered pattern and jagged edges. The "R" in a circle is located to the right of the letter "O".

EIZO Corporation, 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan

U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002

NameHiroaki Hashimoto
DepartmentRegulatory Compliance andSafety
Telephone+81 (76) 274-2468
E-Mailhiroaki.hashimoto@eizo.com

510(k) Summary K223768

1. Submitter

EIZO Corporation 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan Phone: +81 (76) 274-2468

Contact Person: Hiroaki Hashimoto Date of Prepared: December 9th, 2022

2. Device

  • Name of Device: RadiForce RX270, RX270-AR
  • Common or Usual Name: 21.3 inch (54.0 cm) Color LCD Monitor
  • Classification Name: Medical image management and processing system ● (21 CFR 892.2050)
  • Regulatory Class: ● II
  • Product Code: PGY ●

3. Predicate Device

  • Name of Device: RadiForce RX250, RX250-AR (K160247) ●
  • Common or Usual Name: 2MP Color LCD Monitor ●
  • Classification Name: Medical image management and processing system ● (21 CFR 892.2050)
  • Regulatory Class: II
  • . Product Code: PGY
  • RadiForce RX360, RX360-AR (K182591) ● Name of Device:
  • Common or Usual Name: 54.1 cm (21.3 inch) class Color LCD Monitor
  • Classification Name: Picture archiving and communications system ● (21 CFR 892.2050) Regulatory Class: ● II
  • Product Code: PGY ●

{4}------------------------------------------------

4. Device Description

RadiForce RX270 is a color LCD monitor for viewing medical images other than those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 1,200 x 1,600 pixels (2MP) with a pixel pitch of 0.270 mm.

Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used.

There are two model variations. RX270 and RX270-AR. The difference of the two variations is the surface treatment of the display screens; the surface treatment of the RX270 is Anti-Glare (AG) treatment and that of the RX270-AR is Anti-Reflection (AR) coating.

RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including the RadiForce RX270 based on the OC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS and its subset, RadiCS LE, are included in this 510(k) submission as an accessory to the RadiForce RX270.

RadiCS is of Minor level of concern and that it's being used unchanged from the predicate software. RadiCS supports the functions of the monitor RadiForce RX270 and it's not a medical imaging software.

5. Indications for use

This Product is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

{5}------------------------------------------------

6. Comparison of Technological Characteristics with the predicate device

The comparison table below enumerates information derived from the product brochure and measured values of the each device and different technological characteristics are discussed in it:

AttributesProposed Device:RadiForce RX270Predicate Device:RadiForce RX250
Indications for UseThis Product is indicated for use indisplaying radiological images forreview, analysis, and diagnosis bytrained medical practitioners. Thedisplay is not intended formammography.This product is intended to be used indisplaying and viewing digitalimages for review, analysis anddiagnosis by trained medicalpractitioners. It does not support thedisplay of mammography images fordiagnosis.
Display TechnologyTFT Color LCD Panel (IPS)TFT Color LCD Panel (IPS)
Screen size54.0cm / 21.3"Aspect ratio: 3 : 454.0cm / 21.3"Aspect ratio: 3 : 4
Backlight typeLEDLED
Frame rate and refresh rate31 - 100 kHz / 59 - 61 Hz(VGA TEXT: 69 - 71 Hz)Frame synchronous mode: 59 - 61Hz31 - 100 kHz / 59 – 61 Hz (VGAText: 69-71 Hz)Frame synchronous mode: 59 - 61Hz
Display Interface
Input video signalsDVI-D x 1,DisplayPort x 2DVI-D x 1,DisplayPort x 1
Output video signalsDisplayPort x 1 (daisy chain)DisplayPort x 1 (daisy chain)
Video bandwidth
DVI : 25 - 164.5MHzDisplayPort : 25 - 164.5MHzDVI : 25 - 164.5MHzDisplayPort : 25 - 164.5MHz
Ambient light sensing
Ambient light sensorPhoto Diode Position: In the upperbezel of the screenSoftware tool: RadiCSPhoto Diode Position: In the upperbezel of the screenSoftware tool: RadiCS
Luminance calibration tools
Integrated optical sensorExternal optical sensorIntegrated optical sensorExternal optical sensor
Quality-control procedures
Software: RadiCSSoftware: RadiCS
New functions
Hybrid Gamma PXLThis function is implemented on the510(k) cleared RadiForce MX216:K190153 or later.This function is user selectable anddefault: off.Even turn on, this function does notaffect monochrome images.-
LEA (LifeExpectancyAnalyzer)This function is implemented on the510(k) cleared RadiForce RX350:K153354 or later.This function is to calculate theproduct life and it does not affectimages.-
USB Selection forKVM SwitchThis function is implemented on the510(k) cleared RadiForce RX660:K163335 or later.This function is a USB switchingfunction and it does not affectimages.-

{6}------------------------------------------------

It is clear that the technological characteristics differences discussed above are substantially equivalent and new functions do not affect the safety and the effectiveness of the RX270.

{7}------------------------------------------------

7. Performance Testing

The bench tests below were performed on the RadiForce RX270 following the instructions in "Guidance for Industry and FDA Staff: Display Devices for Diagnostic Radiology" issued on September 28,2022:

  • . Measurement of spatial resolution expressed as modulation transfer function (MTF)
  • The maximum number allowed for each type of pixel defects/faults ●
  • . Visual check of presence or absence of miscellaneous artifacts on the display screen as specified in TG18 guideline
  • Measurement of temporal response
  • Measurement of Luminance
  • . Verification of the conformance to DICOM GSDF as specified in Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline)
  • Measurement of Color tracking ●

The test results showed that the RadiForce RX270 has display characteristics equivalent to those of the predicate device, RadiForce RX250.

Besides, the display characteristics of the RadiForce RX270 meet the pre-defined criteria when criteria are set.

No animal or clinical testing was performed on the RadiForce RX270.

8. Conclusion

The RadiForce RX270 was determined to be substantially equivalent to the predicate device due to the following reasons:

  • . The stated intended use is substantially the same as that of the predicate device.
  • . It was confirmed that the technological characteristics differences from those of the predicate device do not affect the safety or the effectiveness.
  • . The bench tests demonstrated that the display characteristics are equivalent to those of the predicate device.

N/A