K Number
K223755

Validate with FDA (Live)

Date Cleared
2023-09-11

(270 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The i-STAT G cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose in arterial. venous or capillary whole blood in point of care or clinical laboratory settings.

Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, and pancreatic islet cell carcinoma.

Device Description

The i-STAT G cartridge is used with the i-STAT 1 analyzer as part of the i-STAT 1 System to measure glucose in arterial, venous, or capillary whole blood for the diagnosis, monitoring, and treatment of metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

The i-STAT 1 System is an in vitro diagnostic (IVD) medical device intended for the quantitative determination of various clinical chemistry tests contained within i-STAT cartridges using whole blood. The i-STAT 1 System consists of a portable blood analyzer (i-STAT 1 analyzer), single-use disposable test cartridges (i-STAT cartridges), liquid quality control and calibration verification materials, and accessories (i-STAT 1 Downloader/Recharger, i-STAT Electronic Simulator and i-STAT 1 Printer). The i-STAT 1 Sustem, including the i-STAT G cartridge, is designed for use by trained medical professionals in point of care or clinical laboratory settings and is for prescription use only.

The i-STAT G cartridge contains the required sensors and a fluid pack (calibrant pouch), a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within the cartridge. The test is contained in a single-use, disposable cartridge. All the test steps and fluid movements occur within the i-STAT G cartridge. Cartridges require two to three drops of whole blood applied to the cartridge using a transfer device, by the trained user before the cartridge is placed within the analyzer.

The i-STAT 1 analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The analyzer interacts with the cartridge to move fluid across the sensors and generate a quantitative result (within approximately 2 minutes).

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the i-STAT G cartridge with the i-STAT 1 System, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" for each performance characteristic in a summarized table within the 510(k) summary. However, it indicates that the studies "met acceptance criteria" or "demonstrated equivalency" to a predicate device or allowable error.
For several tests (e.g., precision, linearity), the results are presented, and the implicit acceptance criterion is that these results fall within acceptable clinical or statistical limits, or demonstrate substantial equivalence to the predicate device.
For the purpose of this response, I will infer the acceptance criteria from the context of how the results are presented and the FDA's ultimate determination of "substantially equivalent."

Acceptance Criteria (Inferred)Reported Device Performance (Summary from Tables)
Precision
20-Day Precision (Aqueous Materials): Within laboratory precision (Total SD and %CV) within acceptable limits for a clinical glucose test.CV L1 (25.0 mg/dL): SD 0.55, %CV 2.19CV L2 (38.5 mg/dL): SD 0.49, %CV 1.27CV L3 (119.1 mg/dL): SD 0.78, %CV 0.66CV L4 (272.2 mg/dL): SD 1.66, %CV 0.61CV L5 (565.5 mg/dL): SD 5.41, %CV 0.96 (Table 1)
Multi-site/Operator Precision (Aqueous Materials): Overall precision (SD and %CV) within acceptable limits across multiple sites and operators.CV L1 (573.1 mg/dL): SD 2.89, %CV 0.50CV L2 (266.3 mg/dL): SD 0.82, %CV 0.31CV L3 (133.7 mg/dL): SD 0.61, %CV 0.46CV L4 (46.1 mg/dL): SD 0.48, %CV 1.04CV L5 (33.7 mg/dL): SD 0.58, %CV 1.73 (Table 2)
Whole Blood Precision: %CVs for different sample types (Venous, Arterial, Capillary) across various glucose ranges within acceptable clinical limits.Venous (20-700 mg/dL): %CVs 0.32-2.01Arterial (20-700 mg/dL): %CVs 0.26-0.46Capillary (20-700 mg/dL): %CVs 1.38-2.71 (Table 3)
Linearity/Reportable Range
Linearity: Slope ~1, Intercept ~0, R2 ~1 over the reportable range (20-700 mg/dL).Slope 1.002, Intercept -1.258, R2 0.999 for reportable range 20-700 mg/dL (Table 4).
LoQ: LoQ value must be less than or equal to the lower limit of the reportable range (20 mg/dL).LoQ for Glucose: 14 mg/dL. This is below the lower limit of the reportable range of 20 mg/dL (Table 5).
LoB/LoD: LoB and LoD values should be sufficiently low, ideally near zero, for a glucose test.LoB for Glucose: 0 mg/dLLoD for Glucose: 0.7 mg/dL (Table 6)
Analytical Specificity (Interference): Difference in means (or medians) between control and test samples for potentially interfering substances within allowable error (±Ea).Most tested substances showed "No" interference. Identified interferents (Bromide, Hydroxyurea, Isoniazid) resulted in a comment to "Use another method" or noted increased results (Table 7). This implies the device performed as expected by identifying interferents, potentially leading to warnings for users.
Oxygen Sensitivity: 95% confidence interval (CI) of the difference in means (or medians) between high and low oxygen conditions within the allowable error (±Ea).Insensitive to oxygen levels between 21 and 515 mmHg.
Hematocrit Sensitivity: Difference between low/high and mid hematocrit levels within the allowable error (±Ea).Insensitive to hematocrit levels between 15% to 75% packed cell volume (PCV).
Altitude: Correlation coefficient (r) and slope met acceptance criteria when compared to a comparator device at altitude.At approximately 10,000 feet above sea level, r = 0.97 (CI 0.964 to 0.972) and Slope = 1.00 (CI 1.000 to 1.000) (Table 8), demonstrating equivalent performance.
Method Comparison (vs. Comparator Device): Passing-Bablok regression with slope ~1, intercept ~0, and high correlation (r ~1).Arterial/Venous vs. i-STAT CHEM8+: N=571, Slope 1.00, Intercept 1.85, r 1.00 (Table 9)Capillary vs. epoc Blood Analysis System: Included in the pooled data above. (Table 9)By sample type:- Arterial vs. i-STAT CHEM8+: N=173, Slope 1.00, Intercept 1.00, r 1.00- Venous vs. i-STAT CHEM8+: N=164, Slope 1.00, Intercept 1.50, r 1.00- Capillary vs. epoc Blood Analysis System: N=234, Slope 1.00, Intercept 2.00, r 1.00 (Table 10)
Matrix Equivalence: Passing-Bablok regression (non-anticoagulated vs. heparinized) with slope ~1, intercept ~0, and high correlation (r ~1).N=158, r 1.00, Slope 1.00, Intercept 0.00 (Table 11).
EDTA Matrix Equivalence: Passing-Bablok regression (K2EDTA/K3EDTA vs. lithium heparin) with slope ~1, intercept ~0, and high correlation (r ~1).K2EDTA vs. LiHep: N=43, r 1.00, Slope 1.03, Intercept -1.037 (Table 12)K3EDTA vs. LiHep: N=43, r 1.00, Slope 1.03, Intercept 0.015 (Table 13)

2. Sample Sizes Used for the Test Set and Data Provenance

  • Precision (20-day, aqueous materials): N=80 for each of 5 fluid levels.
  • Precision (Multi-site and operator-to-operator, aqueous materials): N=89-90 for each of 5 fluid levels.
  • Precision (whole blood):
    • Venous whole blood: N=38 (20-90 mg/dL), N=67 (>90-150 mg/dL), N=32 (>150-250 mg/dL), N=15 (>250-400 mg/dL), N=12 (>400-700 mg/dL).
    • Arterial whole blood: N=9 (20-90 mg/dL), N=94 (>90-150 mg/dL), N=64 (>150-250 mg/dL), N=6 (>250-700 mg/dL).
    • Capillary whole blood: N=33 (20-90 mg/dL), N=53 (>90-150 mg/dL), N=37 (>150-250 mg/dL), N=16 (>250-700 mg/dL).
  • Linearity: Whole blood samples of varying glucose levels across the reportable range.
  • Limit of Quantitation (LoQ): Whole blood "altered to a low glucose level."
  • Limit of Blank/Detection (LoB/LoD): Whole blood "altered to a blank glucose level" and "two (2) low glucose levels."
  • Interference: Whole blood samples at low and high glucose levels.
  • Oxygen Sensitivity: Whole blood samples "altered to four (4) glucose levels."
  • Hematocrit Sensitivity: Whole blood samples at "three (3) hematocrit levels...at four (4) glucose levels."
  • Altitude: Whole blood samples "across the reportable range."
  • Method Comparison:
    • Arterial/Venous: N=571 (pooled data for i-STAT CHEM8+ comparison)
    • Arterial: N=173 (vs. i-STAT CHEM8+)
    • Venous: N=164 (vs. i-STAT CHEM8+)
    • Capillary: N=234 (vs. epoc Blood Analysis System)
  • Matrix Equivalence (non-anticoagulated vs. heparinized): N=158
  • EDTA Matrix Equivalence: N=43 for K2EDTA vs LiHep, N=43 for K3EDTA vs LiHep.

Data Provenance:

  • Retrospective/Prospective: Primarily involves prospective testing of samples (aqueous materials, whole blood) under controlled conditions, or collection of patient samples for method comparisons.
  • Country of Origin: Not specified in the provided text, but implied to be within the scope of where Abbott Point of Care (US company) conducts its clinical evaluations, likely the US or other regions following CLSI guidelines.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not detailed in the provided document. The ground truth for analytical performance studies on diagnostic devices like this often relies on:

  • Reference methods using highly accurate laboratory analyzers.
  • Certified reference materials (like NIST SRM 965 mentioned for traceability).
  • Clinically established ranges and allowable errors.
  • The "experts" are primarily the laboratory scientists and statisticians who conduct and analyze the studies according to CLSI guidelines.

For method comparison studies, the "ground truth" is typically the result from a legally marketed predicate device (i-STAT CHEM8+ cartridge) or a well-established laboratory reference method (epoc Blood Analysis System). The document does not mention human expert annotation of samples for establishing ground truth, as it's an in-vitro diagnostic device not directly involving image interpretation or clinical decision-making by human experts for the output itself.

4. Adjudication Method for the Test Set

This information is not applicable in the traditional sense for an in-vitro diagnostic glucose test. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies where human readers or interpreters make subjective assessments that might differ, and a consensus or adjudication process is needed to establish a definitive ground truth. For quantitative measurements like glucose, the "adjudication" is inherent in the analytical process:

  • Reference methods provide the comparative "true" value.
  • Statistical analyses (e.g., Passing-Bablok regression, precision calculations) determine if the device's results are sufficiently close to this reference or internally consistent.
  • Discrepancies would be investigated through quality control and root cause analysis, not expert adjudication of contradictory readings from the device itself.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This type of study (comparing human readers with and without AI assistance on multiple cases) is relevant for AI-powered diagnostic imaging devices where human interpretation is a key component. The i-STAT G cartridge is an in-vitro diagnostic device that directly measures glucose levels, not an AI assistance tool for human interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the studies presented are essentially "standalone" performance evaluations. The i-STAT G cartridge with the i-STAT 1 System functions as an automated system for quantifying glucose. The performance data presented (precision, linearity, LoQ, LoB/LoD, interference, sensitivity, altitude, method comparison, and matrix equivalence) are all evaluations of the device's analytical performance on its own, without direct real-time human interpretation of the glucose value itself as the primary output to be compared. Human involvement is for sample collection, device operation, and quality control, but the reported glucose value is generated by the "algorithm" (the device's embedded processes).

7. The Type of Ground Truth Used

The ground truth used for performance validation includes:

  • Reference methods: For method comparison studies, the i-STAT CHEM8+ cartridge on the i-STAT 1 System (predicate device) and the epoc Blood Analysis System served as comparative methods to establish ground truth or reference values.
  • Certified reference materials/Calibrators: Mention of NIST SRM 965 for traceability and "i-STAT Calibration Verification set" for precision studies indicates the use of highly accurate, traceable materials.
  • Spiked samples: For linearity, LoQ, LoB/LoD, and interference studies, samples were "altered" or "spiked" to known concentrations of glucose or interfering substances.
  • Statistical models and clinical thresholds: The "allowable error (±Ea)" indicates that performance is judged against pre-defined clinical or statistical acceptance limits.

8. The Sample Size for the Training Set

This information is not provided in the document. The document describes performance validation studies (test sets). For an IVD device like this, the "training set" would refer to data used during the development and optimization phases before the formal validation for regulatory submission. Details on development-stage datasets are generally not included in 510(k) summaries.

9. How the Ground Truth for the Training Set was Established

This information is not provided for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 11, 2023

Mojgan Soleimani Associate Director Regulatory Affairs Abbott Point of Care Inc. 400 College Road East Princeton, New Jersey 08540

Re: K223755

Trade/Device Name: i-STAT G cartridge with the i-STAT 1 System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: CGA Dated: August 8, 2023 Received: August 9, 2023

Dear Mojgan Soleimani:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Paula V. Caposino -S

Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223755

Device Name i-STAT G cartridge on the i-STAT 1 System

Indications for Use (Describe)

The i-STAT G cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose in arterial. venous or capillary whole blood in point of care or clinical laboratory settings.

Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, and pancreatic islet cell carcinoma.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Abbott logo. The logo consists of a blue abstract shape on the left and the word "Abbott" in black on the right. The abstract shape is a stylized letter "A" with rounded corners. The word "Abbott" is in a bold, sans-serif font.

510(k) SUMMARY

The information in this 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.

I. SUBMITTER INFORMATION

OwnerAbbott Point of Care Inc.400 College Road EastPrinceton, NJ 08540
ContactPrimary: Mojgan SoleimaniAssociate Director Regulatory AffairsPhone: +1 613-295-0932
Secondary: Robert GreggDirector Regulatory AffairsPhone: +1 609-454-9360
Date PreparedSeptember 11, 2023

II. DEVICE INFORMATION

Proprietary Namei-STAT G cartridge with the i-STAT 1 System
Common NameGlucose test, analyzer, handheld
510(k) Number:K223755
ProductCodeDevice Classification NameRegulationNumberClassPanel
CGAGlucose oxidase, Glucose862.1345IIClinical Chemistry

III. PREDICATE DEVICE

Proprietary Name

i-STAT CHEM8+ cartridge with the i-STAT 1 System

510(k) Number

K210958 (K183678)

ProductCodeDevice ClassificationNameRegulationNumberClassPanel
CGAGlucose oxidase, Glucose862.1345IIClinical Chemistry

{4}------------------------------------------------

IV. DEVICE DESCRIPTION

The i-STAT G cartridge is used with the i-STAT 1 analyzer as part of the i-STAT 1 System to measure glucose in arterial, venous, or capillary whole blood for the diagnosis, monitoring, and treatment of metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

The i-STAT 1 System is an in vitro diagnostic (IVD) medical device intended for the quantitative determination of various clinical chemistry tests contained within i-STAT cartridges using whole blood. The i-STAT 1 System consists of a portable blood analyzer (i-STAT 1 analyzer), single-use disposable test cartridges (i-STAT cartridges), liquid quality control and calibration verification materials, and accessories (i-STAT 1 Downloader/Recharger, i-STAT Electronic Simulator and i-STAT 1 Printer). The i-STAT 1 Sustem, including the i-STAT G cartridge, is designed for use by trained medical professionals in point of care or clinical laboratory settings and is for prescription use only.

The i-STAT G cartridge contains the required sensors and a fluid pack (calibrant pouch), a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within the cartridge. The test is contained in a single-use, disposable cartridge. All the test steps and fluid movements occur within the i-STAT G cartridge. Cartridges require two to three drops of whole blood applied to the cartridge using a transfer device, by the trained user before the cartridge is placed within the analyzer.

The i-STAT 1 analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The analyzer interacts with the cartridge to move fluid across the sensors and generate a quantitative result (within approximately 2 minutes).

V. INTENDED USE STATEMENT

The i-STAT G cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose in arterial, venous or capillary whole blood in point of care or clinical laboratory settings.

Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

Similarities and Differences (Test and Instrument)
Candidate DevicesPredicate Devices

VI. SUMMARY COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Similarities and Differences (Test and Instrument)
Feature orCharacteristicCandidate Device:Glucose test in the i-STAT G cartridgewith the i-STAT 1 SystemPredicate Device:Glucose test in the i-STAT CHEM8+cartridge with the i-STAT 1 SystemK210958 (K183678)
Intended UseThe i-STAT G cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of sodium, potassium, chloride, ionized calcium, glucose, blood urea nitrogen, creatinine, hematocrit, and total carbon dioxide in
Similarities and Differences (Test and Instrument)
Feature orCharacteristicCandidate Device:Glucose test in the i-STAT G cartridgewith the i-STAT 1 SystemPredicate Device:Glucose test in the i-STAT CHEM8+cartridge with the i-STAT 1 SystemK210958 (K183678)
Glucose measurements are used in thediagnosis, monitoring, and treatmentof carbohydrate metabolism disordersincluding, but not limited to, diabetesmellitus, neonatal hypoglycemia,idiopathic hypoglycemia, andpancreatic islet cell carcinoma.arterial or venous whole blood in point ofcare or clinical laboratory settings.The glucose test in the i-STAT CHEM8+cartridge with the i-STAT 1 System isintended for use in the in vitroquantification of glucose in arterial orvenous whole blood in point of care orclinical laboratory settings.Glucose measurements are used in thediagnosis, monitoring, and treatment ofcarbohydrate metabolism disordersincluding, but not limited to, diabetesmellitus, neonatal hypoglycemia,idiopathic hypoglycemia, and pancreaticislet cell carcinoma.
DeviceClassificationSameClass II
Product CodeSameCGA
RegulationNumberSame862.1345
ReportableRangeSame1.1 – 38.9 mmol/L20 – 700 mg/dL0.20 – 7.00 g/L
Sample TypeArterial, venous or capillary wholebloodArterial or venous whole blood
SampleVolume65 μL95 μL
SamplePreparationSameReady to use
SampleCollection• Without anticoagulant• With balanced heparin anticoagulantor lithium anticoagulant
Arterial andvenousWithout anticoagulant
Arterial andVenousWith balanced heparin,lithium, K2 or K3 EDTAanticoagulant
CapillaryWith balanced heparinor lithium anticoagulant
Similarities and Differences (Test and Instrument)
Feature or CharacteristicCandidate Device:Glucose test in the i-STAT G cartridgewith the i-STAT 1 SystemPredicate Device:Glucose test in the i-STAT CHEM8+ cartridgewith the i-STAT 1 SystemK210958 (K183678)
TraceabilitySameNIST SRM 965
CalibrationSame1-point on-board contained within cartridge
Time to Test / Sample Stability(Time from collection to sample fill)Without anticoagulant:Arterial and venous within 3 minutesWith anticoagulant:Capillary within 3 minutesArterial and venous within 30 minutesWithout anticoagulant:Arterial and venous within 3 minutesWith anticoagulant:Arterial and venous within 30 minutes
Principle of MeasurementSameAmperometric measurement of oxidized hydrogen peroxide produced by glucose oxidase activity
Reagent FormatSameCartridge
Reagent Storage and StabilitySameRefrigeration at 2°C to 8°C (35-46°F) until expiration date;Room Temperature at 18°C to 30°C (64–86 °F) for 14 days
Analyzer TypeSameHandheld

{5}------------------------------------------------

{6}------------------------------------------------

VII. PERFORMANCE CHARACTERISTICS

A. Analytical Performance

a. Precision/Reproducibility:

i. Precision 20 days (aqueous materials)

The precision of the i-STAT Glucose test in the i-STAT G cartridge with the i-STAT 1 System was evaluated using five (5) levels of aqueous material. This single-site 20-day multi-day precision testing was based on CLSI document EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition. The study was conducted using multiple analyzers and one (1) test cartridge lot over at least 20 days at one (1) site. Repeatability, between-run, between-day, and within-laboratory precision were estimated for each level. The results of the 20-day precision study are shown in Table 1.

{7}------------------------------------------------

Table 1: Results of 20-Day precision of the i-STAT Glucose test on the i-STAT 1 analyzer using i-STAT Calibration Verification set (mg/dL)

Fluid LevelNMean(mg/dL)RepeatabilityBetween-runBetween-dayWithin-Laboratory
SD%CVSD%CVSD%CVSD%CV
CV L18025.00.431.710.060.230.341.360.552.19
CV L28038.50.380.990.220.580.210.540.491.27
CV L380119.10.690.580.210.180.310.260.780.66
CV L480272.21.420.520.400.150.760.281.660.61
CV L580565.54.330.772.710.481.780.325.410.96

ii. Multi-site and operator-to-operator precision (aqueous materials)

Multi-day precision testing was performed at three (3) sites using a panel of aqueous material containing five (5) levels of glucose. At each site, testing was performed once per day by two (2) operators for five (5) days on six (6) i-STAT 1 analyzers using one (1) lot of i-STAT G cartridges. Within-run, between-day, between-operator and within-site (total) variance components were calculated by site. These components were also calculated for all sites combined and provided in the Table 2 below.

Table 2: Results of point of care multi-day precision of i-STAT Glucose test on the i-STAT 1 analyzer using i-STAT TriControls Calibration Verification set (mg/dL)

Fluid LevelNMeanWithin-RunBetween-DayBetween-OperatorWithin-Site (Total)Between-SiteOverall
SD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
CV L189573.12.630.461.200.210.000.002.890.500.000.002.890.50
CV L290266.30.710.270.370.140.160.060.820.310.050.020.820.31
CV L390133.70.570.430.120.090.180.130.610.460.000.000.610.46
CV L49046.10.330.720.310.670.040.090.460.990.150.320.481.04
CV L59033.70.571.680.000.000.140.400.581.730.000.000.581.73

iii. Precision (whole blood)

Whole blood precision of the i-STAT Glucose test in the i-STAT G cartridge on the i-STAT 1 System was evaluated using arterial, venous, and capillary whole blood specimens collected with lithium heparin. The whole blood precision was assessed using the duplicate test results collected across multiple point of care sites. For each sample type, samples were grouped into subintervals based on their mean values. The results are summarized in Table 3.

Test(units)Sample TypeSample RangeNMeanSD%CV
Glucose(mg/dL)Venous whole blood20-903875.00.320.43
>90-15067109.60.390.35
>150-25032195.80.730.37

{8}------------------------------------------------

>250-40015315.01.170.37
>400-70012559.02.010.36
Arterial whole blood20-90982.40.330.40
>90-15094125.00.570.46
>150-25064182.00.540.30
>250-7006357.00.910.26
Capillary whole blood20-903370.91.922.71
>90-15053116.02.442.10
>150-25037196.64.402.24
>250-70016297.14.091.38

b. Linearity/assay reportable range:

i. Linearity

The study was designed based on CLSI EP06-Ed2: Evaluation of Linearity of Quantitative Measurement Procedures - Second Edition.

The linearity of the i-STAT Glucose test in the i-STAT G cartridge with the i-STAT 1 System was evaluated by preparing whole blood samples of varying glucose levels across the reportable range for the test. The i-STAT Glucose test in the i-STAT G cartridge on the i-STAT 1 System demonstrated linearity over the reportable range of 20 - 700 mg/dL. Regression summary of the response for the i-STAT Glucose test versus the concentration of the whole blood samples of varying glucose levels is provided in Table 4.

TestUnitsReportable RangeRange TestedSlopeInterceptR2
Glucosemg/dL20-70015.3 – 793.31.002-1.2580.999

c. Detection Limit

Limit of Quantitation (LoQ) i.

The study was based on the CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition.

The LoQ of the i-STAT Glucose test in the i-STAT G cartridge was evaluated on the i-STAT 1 analyzer using two (2) cartridge lots and whole blood that was altered to a low glucose level. The LoQ for the i-STAT Glucose test was determined to be 14 mg/dL, which is below the lower limit of the reportable range for the i-STAT Glucose test as shown in Table 5.

Table 5: LoQ result for the i-STAT Glu test in the i-STAT G cartridge
TestLower limit of the reportable rangeLoQ
Glucose20 mg/dL14 mg/dL

{9}------------------------------------------------

ii. Limit of Blank and Detection (LoB/LoD)

The study was based on CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition.

The LoB and LoD of the i-STAT Glucose test in the i-STAT G cartridge were evaluated on the i-STAT 1 analyzer using two (2) cartridge lots. Whole blood was altered to a blank glucose level for LoB testing and two (2) low glucose levels for LoD testing. The LoB and LoD were determined based on the maximal LoB or LoD value obtained for each lot tested. The LoB and LoD for the i-STAT Glucose test on the i-STAT 1 analyzer was determined as shown in the Table 6 below.

Table 6: Summary of LoB and LoD results for the i-STAT G cartridge
TestLoBLoD
Glucose0 mg/dL0.7 mg/dL

d. Analytical Specificity

i. Interference

The study was based on CLSI EP07-ED3: Interference Testing in Clinical Chemistry - Third Edition.

The interference performance of the i-STAT Glucose test in the i-STAT G cartridge on the i STAT 1 analyzer with the i-STAT 1 System was evaluated using whole blood samples at low and high glucose levels for all substances. The effect of each substance was evaluated by comparing the test results of a control sample, spiked with blank solvent solution, with the test results from a test sample spiked with the potentially interfering substance at the toxic or pathological concentration based on CLSI EP37-ED1: Supplemental Tables for Interference Testing in Clinical Chemistry, First Edition, as applicable. A substance was identified as an interferent if the difference in means (or medians) between the control and test samples was outside of the glucose allowable error (±Ea). For an identified interferent, a dose-response analysis was performed to determine the degree of interference as a function of the substance concentration.

Table 7 contain the lists of potentially interfering substances tested for the i-STAT Glucose test and the interference results.

Table 7: Potentially interfering substances and test concentrations for the i-STAT Glucose
test in the i-STAT G cartridge
SubstanceSubstanceConcentration(mmol/L)SubstanceConcentration(mg/dL)TestInterference(Yes/No)Comment
Acetaldehyde0.0450.2GluNo
Acetaminophen1.0315.6GluNo
Acetoacetate(Lithium Acetoacetate)2.020GluNo
Table 7: Potentially interfering substances and test concentrations for the i-STAT Glucose
test in the i-STAT G cartridge
SubstanceSubstanceConcentration(mmol/L)SubstanceConcentration(mg/dL)TestInterference(Yes/No)Comment
Acetyl Cysteine(N-Acetyl-L-Cysteine)0.9215.0GluNo
Ammonium(AmmoniumChloride)2.010.7GluNo
Ascorbic Acid(L-AscorbicAcid)0.2985.25GluNo
β-HydroxybutyricAcid6.062.46GluNo
Bilirubin0.68440GluNo
Bromide2.521.7GluNo
(LithiumBromide)37.5325.7GluYesUse another method.
Cholesterol10.3400GluNo
Creatinine1.32615GluNo
Dopamine(DopamineHydrochloride)4.06 µmol/L0.0621GluNo
Ethanol130600GluNo
Fluoride(LithiumFluoride)0.06320.12GluNo
Formaldehyde0.1330.399GluNo
Fructose118GluNo
Galactose3.3360GluNo
Gentamicin(GentamicinSulfate)0.06283GluNo
Gentisic Acid0.09731.5GluNo
Glucosamine(GlucosamineHydrochloride)0.0300.647GluNo
Glutathione,reduced33 mEq/LGluNo
Glycolic Acid10.076.05GluNo
Guaifenesin0.02270.45GluNo
Hemoglobin10 g/L1000GluNo
Heparin(SodiumHeparin)3.30 U/mL330 U/dLGluNo
Hydroxyurea0.4053.08GluYesIncreased results ≥ 0.08mmol/L.
Ibuprofen1.0621.9GluNo
Intralipid 20%N/A3151GluNo

{10}------------------------------------------------

{11}------------------------------------------------

Table 7: Potentially interfering substances and test concentrations for the i-STAT Glucose
test in the i-STAT G cartridge
SubstanceSubstanceConcentration(mmol/L)SubstanceConcentration(mg/dL)TestInterference(Yes/No)Comment
Isoniazid0.4386GluYesIncreased results ≥ 0.29mmol/L.
Lactate(LithiumLactate)1090GluNo
Maltose10.5360GluNo
Mannose118.02GluNo
Nithiodote(SodiumThiosulfate)16.7264.04GluNo
pH8.0 pH unitsN/AGluNo
Pyruvate(LithiumPyruvate)0.5705GluNo
Salicylate(LithiumSalicylate)0.2072.86GluNo
Thiocyanate(LithiumThiocyanate)0.8985.22GluNo
Triglyceride16.941500GluNo
Uric Acid1.423.5GluNo
Xylose345.04GluNo

ii. Other sensitivity studies

1) Oxygen Sensitivity

The effect of oxygen on the i-STAT Glucose test in the i-STAT G cartridge on the i-STAT 1 System was evaluated with low and high oxygen levels using whole blood samples altered to four (4) glucose levels across the reportable range of the i-STAT Glucose test. The equivalency between the high and low oxygen conditions was determined if the 95% confidence interval (CI) of the difference in means (or medians) was within the allowable error (+ Ea).

The study demonstrated that i-STAT Glucose test in the i-STAT G cartridge with the i-STAT 1 System is insensitive to oxygen levels between 21 and 515 mmHg.

2) Hematocrit Sensitivity

The effect of hematocrit on the i-STAT Glucose test in the i-STAT G cartridge with the i-STAT 1 System was assessed. Three (3) hematocrit levels (low, mid and high) were evaluated at four (4) glucose levels across the reportable range of the i-STAT Glucose test in the i-STAT G cartridge. The hematocrit sensitivity at each glucose level was assessed by comparing the results at the low and high hematocrit level to the mid hematocrit level. Equivalency was assessed by determining whether the difference between the low and high hematocrit level

{12}------------------------------------------------

and the mid hematocrit level was within the allowable error (±Ea). The study demonstrated that i-STAT Glucose test in the i-STAT G cartridge with the i-STAT 1 System is insensitive to hematocrit levels between 15% to 75% packed cell volume (PCV).

3) Altitude

The performance of the i-STAT Glucose test in the i-STAT G cartridge on the i-STAT 1 analyzer was evaluated at an altitude of approximately 10,000 feet above sea level using whole blood samples across the reportable range. The glucose test results obtained from the i-STAT G cartridges (candidate device) were compared to the glucose test results obtained from the i-STAT CHEM8+ cartridges on the i-STAT 1 analyzer (comparator device). Passing-Bablok regression analyses between the 1st replicate of the candidate device (v-axis) and mean of the comparator device (x-axis) were performed based on the CLSI cc: Measurement Procedure Comparison and Bias Estimation using Patient Samples, Third Edition. The results of the correlation coefficient and slope met acceptance criteria and demonstrate equivalent performance between the candidate and comparator condition at approximately 10,000 feet above sea level. The results are summarized in Table 8 below.

Table 8: Summary of altitude study results for the i-STAT G cartridge Correlation Coefficient (r) Slope Test 95% Cl 95% Cl r Slope

0.97

0.964 to 0.972

B. Comparison Studies

Glucose

a. Method Comparison with Comparator Device

1.00

Method comparison was demonstrated in studies based on CLSI EP09c-ED3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples -Third Edition. Lithium heparin arterial and venous whole blood specimens collected across multiple point of care sites were evaluated using the first replicate result from the i-STAT G cartridge on the i-STAT 1 analyzer versus the mean result from the i-STAT CHEM8+ cartridge on the i-STAT 1 analyzer.

1.000 to 1.000

In addition, two (2) capillary whole blood specimens collected from skin punctures with balanced heparin capillary tubes from each study subject across multiple point of care sites were evaluated using the singlicate result from the i-STAT G cartridge on the i-STAT 1 analyzer versus the singlicate result from the epoc Blood Analysis System.

The arterial, venous, and capillary whole blood data were pooled, and a Passing-Bablok linear regression analysis was performed using the results from the i-STAT G cartridges on the i-STAT 1 analyzer versus the comparative method results. Method comparison results for arterial, venous, and capillary whole blood specimens are shown in Table 9. In the table, N is the number of specimens in the data set, and r is the correlation coefficient.

Table 9: Method comparison results for i-STAT G cartridge with i-STAT 1 System

{13}------------------------------------------------

TestComparative MethodNSlopeInterceptr
Arterial/VenousCapillary
Glui-STAT CHEM8+epoc Blood AnalysisSystem5711.001.851.00

A Passing-Bablok linear regression analysis was performed using the results of each sample from the i-STAT G cartridges on the i-STAT 1 analyzer versus the comparative method results. Method comparison results for arterial, venous and capillary whole blood specimens are shown in Table 10. In the table, N is the number of specimens in the data set, and r is the correlation coefficient.

Table 10: Method comparison results for i-STAT G cartridge with i-STAT 1 System bysample type
TestSample TypeComparative MethodNSlopeIntercept
Sample TypeComparative MethodNSlopeInterceptr
GluArteriali-STAT CHEM8+1731.001.001.00
Venousi-STAT CHEM8+1641.001.501.00
Capillaryepoc Blood Analysis System2341.002.001.00

b. Matrix Equivalence

A matrix equivalence study was conducted to evaluate the performance of the i-STAT Glucose test in the i-STAT G cartridge on the i-STAT 1 analyzer using nonanticoagulated arterial or venous whole blood specimens compared to heparinized whole blood specimens. The study design and analysis method were based on CLSI EP35: Assessment of Equivalence or Suitability of Specimen Types for Medical Laboratory Measurement Procedures – First Edition. The matrix equivalence was assessed by comparing arterial or venous specimens collected without anticoagulant (candidate specimen type) to samples collected with balanced heparin or lithium heparin anticoagulant (primary specimen type). Each specimen was tested in duplicate using two (2) i-STAT G cartridges with two (2) i-STAT 1 analyzers. A Passing-Bablok linear regression analysis was performed using the first replicate result from the candidate (y-axis) versus the mean result from the primary specimen (x-axis). The regression analysis results are summarized in Table 11. In the table. N is the number of specimens in the data set, and r is the correlation coefficient.

Table 11: Matrix equivalence results
NCandidate Specimen Range(mg/dL)Primary Specimen Range(mg/dL)rSlopeIntercept
15842-67942-6811.001.000.00

c. EDTA Matrix Equivalence

A study was conducted to evaluate matrix equivalency between whole blood sample collected in lithium heparin (primary specimen type) and ethylenediaminetetraacetic acid (K2EDTA and K3EDTA) anticoagulants (candidate specimen types) and tested with the i-STAT Glucose test in the i-STAT G cartridge on the i-STAT 1 analyzer. The

{14}------------------------------------------------

study design was based on CLSI EP35: Assessment of Equivalence or Suitability of Specimen Types for Medical Laboratory Measurement Procedures – First Edition. Each specimen was tested in duplicate using two (2) i-STAT G cartridges with i-STAT 1 analyzers. A Passing-Bablok regression analysis was performed using the first valid result of the K2EDTA and K3EDTA anticoagulated specimen types versus the first valid result of the specimens with lithium heparin. The regression analysis results are summarized in Table 12 and Table 13.

Table 12: Passing-Bablok regression summary for K₂EDTA vs lithium heparin
NK₂EDTACandidate Specimen Range(mg/dL)LiHepPrimary Specimen Range(mg/dL)rSlopeIntercept
4330.3 – 522.929.5 – 510.91.001.03-1.037
Table 13: Passing-Bablok regression summary for K3EDTA vs lithium heparin
NK3EDTACandidate Specimen Range(mg/dL)LiHepPrimary Specimen Range(mg/dL)rSlopeIntercept
4330.4 – 520.629.5 – 510.91.001.030.015

CONCLUSION VIII.

The results of these studies demonstrate that performance of the i-STAT Glucose test in the i-STAT G cartridge with the i-STAT 1 System is substantially equivalent to the predicate device.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.