K Number
K223738

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-01-11

(28 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Alpha Control Liner System is to be used exclusively for exoprosthetic fittings of the upper limbs.

Device Description

The Alpha Control Liner System is an interface solution designed to provide the Complete Control System Gen2 with amputee muscle contraction signals needed for the pattern recognition algorithms. The Alpha Control System integrates electronics into a fabric covered elastomeric prosthetic liner to non-invasively detect muscle contractions and convert the signals into a digital format. The Alpha Control System is a substitute for the EMG Interface Cable currently provided with the Complete Control System Gen2. Patients can achieve control of their devices and benefit from quick donning and doffing of the prosthesis. The Alpha Control System simplifies electrode placement and allows a prosthetist to spend less time fabricating the prosthesis as well as adjusting system settings and configurations.

The Alpha Control System is used in conjunction with the Coapt Complete Control System Gen2 and does not require an additional battery.

The Alpha Control System contains the following components.

  • Alpha Control Liner (gel-lined user interface)
  • Alpha Control Module (re-packaged Complete CO-AMP consolidated EMG amplifier, Covered under 510(k) number K162891)
  • Module Cap (secures the Module inside the Liner)
  • Alpha Control Lock (socket suspension and connection to Complete Controller)
  • Electrodes (Covered under 510(k) number K190416)
  • Fabrication aids for the Lock
AI/ML Overview

The provided text does not contain detailed information about specific acceptance criteria and a study proving the device meets them in the format requested. The document is a 510(k) Premarket Notification from the FDA to Coapt Blair Lock for the Alpha Control Liner System. It primarily focuses on demonstrating substantial equivalence to a predicate device (Coapt Complete Control System Gen2) through non-clinical performance data.

Here's an analysis of the available information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance:

The document lists several non-clinical performance tests that the Alpha Control Liner System passed, which can be interpreted as demonstrating the device meets internal requirements, national, and international standards. However, specific numerical acceptance criteria (e.g., "force must be > X Newtons") are not provided. The results are simply reported as "Pass".

Test NameReported Device Performance
Electrical safety testing (IEC 60601-1)Pass
Cytotoxicity testing (ISO 10993-5)Pass
Skin irritation testing (ISO 10993-10)Pass
Sensitization testing (ISO 10993-10)Pass
Liner ThicknessPass
Liner Doffing ForcePass
Donning/Doffing FatiguePass
Liner Umbrella Static LoadingPass
Liner Umbrella Fatigue LoadingPass
Lead Impedance FatiguePass
Lead Impedance - CorrosionPass
Lock Body Static LoadingPass
Lock Body Fatigue LoadingPass
Liner High Pot TestPass
Module High Pot TestPass
Module Level TestPass
Coapt System Integration TestingPass

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the sample sizes used for any of the non-clinical tests. It states that "validation testing was performed on the finished device," but does not detail the number of units or iterations. The provenance of the data is "internal" and "in-house performance testing" by Coapt LLC and WillowWood Global LLC. The document does not mention external (e.g., country of origin) or retrospective/prospective data for these tests.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This information is not provided. The non-clinical tests are largely engineering and materials testing, which rely on established standards and protocols rather than expert medical interpretation for "ground truth" in the same way a clinical diagnostic test would.

4. Adjudication Method for the Test Set:

This information is not applicable and, therefore, not provided. The non-clinical tests do not involve human interpretation requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study was mentioned. The document explicitly states, "No human clinical testing was required to support the medical device as the indications for use is equivalent to the predicate device." This indicates that the evaluation was primarily based on non-clinical engineering and biocompatibility testing. The device is a "cutaneous electrode" for exoprosthetic fittings, not a diagnostic imaging or AI-driven decision support system that typically requires MRMC studies.

6. Standalone (Algorithm Only) Performance:

The device (Alpha Control Liner System) is described as an "interface solution designed to provide the Complete Control System Gen2 with amputee muscle contraction signals needed for the pattern recognition algorithms." It integrates electronics to detect muscle contractions and convert signals into a digital format. While it converts signals, it is a component for a pattern recognition algorithm (part of the predicate device, Coapt Complete Control System Gen2), not the algorithm itself or a standalone algorithm. Therefore, "standalone (i.e. algorithm only without human-in-the loop performance)" is not applicable to this device description. The performance testing focuses on its physical and electrical integrity and its ability to integrate with the larger system.

7. Type of Ground Truth Used:

For the non-clinical tests, the "ground truth" refers to established engineering standards, material specifications, and functional requirements. For example, "Liner Thickness" would be compared against a specified design tolerance, and "Biocompatibility" against the criteria outlined in ISO 10993 standards. There is no mention of expert consensus, pathology, or outcomes data being used as ground truth for these non-clinical tests.

8. Sample Size for the Training Set:

This information is not applicable and not provided. The Alpha Control Liner System is a hardware component that collects and processes sEMG signals. It does not inherently learn or require a "training set" in the context of machine learning. The pattern recognition algorithms are part of the predicate device, and details about their training are not discussed in this submission for the Alpha Control Liner System.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable and not provided, as the device itself does not use a "training set."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

January 11, 2023

Coapt Blair Lock CEO 303 W. Institute Pl. Suite 200 Chicago, Illinois 60610

Re: K223738

Trade/Device Name: Alpha Control Liner System (ACLS) Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: Class II Product Code: GXY Dated: December 14, 2022 Received: December 14, 2022

Dear Blair Lock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tushar Bansal -S

for Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223738

Device Name Alpha Control Liner System (ACLS)

Indications for Use (Describe)

The Alpha Control Liner System is to be used exclusively for exoprosthetic fittings of the upper limbs.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for COAPT. The letters "COA" are in orange, and the letters "PT" are in gray. Below the letters is the phrase "COMPLETE CONTROL" in a smaller font.

Special 510(k) Summary Alpha Control Liner System

1. Submitter Information

Manufacturer:Coapt, LLC
303 W. Institute Pl. Suite 200
Chicago, IL 60610
Contact Person:Blair Lock
CEO
Contact Information:844-262-7800 x700
blair.lock@coaptengineering.com
Contract Manufacturer:WillowWood Global, LLC
15441 Scioto Darby Rd
Mt. Sterling, OH 43143
Contact Person:Matthew Wernke, PhD
Research and Development Manager
Contact Information:(740) 869-3377
matt.wernke@willowwood.com
Date Prepared:December 12, 2022

2. Disclaimer

The submitting company requests that they are notified should a Freedom of Information Request is made to obtain a copy of this 510k application so that the company can properly identify any confidential information and redact it prior to public release.

3. Device Identification

Trade/Proprietary Name:Alpha Control Liner System (ACLS)
Common/Usual Name:Cutaneous Electrode
Classification Name:Electrode, cutaneous
Regulation Number:21 CFR §882.1320
Product Code:GXY, Cutaneous electrode
Subsequent Product Code:Not applicable
Device Class:Class II
Classification Panel:Neurology

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for COAPT Complete Control. The letters "COA" are in orange, while the letters "PT" are in gray. Below the letters is the text "COMPLETE CONTROL" in gray.

4. Legally Marketed Predicate Device

510(k) NumberDevice Name
K191083Coapt Complete Control System Gen2

5. Device Description

The Alpha Control Liner System is an interface solution designed to provide the Complete Control System Gen2with amputee muscle contraction signals needed for the pattern recognition algorithms. The Alpha Control System integrates electronics into a fabric covered elastomeric prosthetic liner to non-invasively detect muscle contractions and convert the signals into a digital format. The Alpha Control System is a substitute for the EMG Interface Cable currently provided with the Complete Control System Gen2. Patients can achieve control of their devices and benefit from quick donning and doffing of the prosthesis. The Alpha Control System simplifies electrode placement and allows a prosthetist to spend less time fabricating the prosthesis as well as adjusting system settings and configurations.

The Alpha Control System is used in conjunction with the Coapt Complete Control System Gen2 and does not require an additional battery.

The Alpha Control System contains the following components.

  • Alpha Control Liner (gel-lined user interface) 1.
  1. Alpha Control Module (re-packaged Complete CO-AMP consolidated EMG amplifier, Covered under 510(k) number K162891)
    1. Module Cap (secures the Module inside the Liner)
    1. Alpha Control Lock (socket suspension and connection to Complete Controller)
    1. Electrodes (Covered under 510(k) number K190416)
    1. Fabrication aids for the Lock

6. Indication for Use Statement

The Alpha Control Liner System is to be used exclusively for exoprosthetic fittings of the upper limbs.

7. Comparison to Predicate Device

The following table compares the Alpha Control Liner System to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials used, and performance. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or efficacy based on the similarities to the predicate device.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for COAPT. The letters "CO" are in orange, while the letters "APT" are in gray. Below the letters is the phrase "COMPLETE CONTROL" in gray.

Manufacturer:Coapt, LLCWillowWood GlobalLLCDeviceComparison
Trade Name:COMPLETE CONTROLSystem Gen2Alpha Control LinerSystem
510(k) NumberK191083K223738Not applicable
Classification ProductCodeGXYGXYSame
Regulation Number21 CFR 882.132021 CFR 882.1320Same
Regulation NameCutaneous electrodeCutaneous electrodeSame
Subsequent ProductCodeIQZ (Hand, External LimbComponent, Powered)IQZ (Hand, ExternalLimb Component,Powered)Same
Indications for UseThe COMPLETECONTROL System Gen2is to be used exclusivelyfor exoprosthetic fittingsof the upper limbs.The Alpha Control LinerSystem is to be usedexclusively forexoprosthetic fittings ofthe upper limbs.Same
Use/Field ofApplicationThe COMPLETECONTROL System Gen2is suitable for unilateral orbilateral amputations.The Alpha Control LinerSystem is suitable forunilateral or bilateralamputations.Same
Conditions of UseThe COMPLETECONTROL System Gen2was developed foreveryday use and must notbe used for unusualactivities. These activitiesinclude, for example,sports with excessivestrain and/or shocks to thewrist unit (pushups,downhill, mountainbiking) or extreme sports(free climbing,paragliding, etc.)Furthermore, theCOMPLETE CONTROLSystem Gen2 should notbe used for the operationof motor vehicles ormotor-driven equipment.The Alpha Control LinerSystem was developedfor everyday use andmust not be used forunusual activities. Theseactivities include, forexample, sports withexcessive strain and/orshocks to the wrist unit(pushups, downhill,mountain biking) orextreme sports (freeclimbing, paragliding,etc.) Furthermore, theAlpha Control LinerSystem should not beused for the operation ofmotor vehicles or motor-driven equipment.Same
Mechanism of ActionThe components of theCOMPLETE CONTROLSystem Gen2 areassembled by a prosthetistaccording to the individualneeds of the amputee.The components of theAlpha Control LinerSystem are assembled bya prosthetist according tothe individual needs ofthe amputee.Same
Power Requirements5.0-20.0 VDC50 mA at 7.4V5.3-16.8 VDC115 mA at 7.4 VSimilar
Manufacturer:Coapt, LLCWillowWood GlobalLLCDeviceComparison
Trade Name:COMPLETE CONTROLSystem Gen2Alpha Control LinerSystem
User-InterfaceMaterialsElectrodes: 316L StainlessSteelElectrodes: 316LStainless SteelSame
Liner: Not applicableLiner: Tri-blockcopolymer gelDifferent;Improvementfor electrodecontact
Miscellaneous
Clinician SoftwareToolYesComplete ControlRoomSoftwareYesComplete ControlRoomSoftwareSimilar
Electrical SafetyTesting PassedIEC 60601-1IEC 61000-4-3IEC 61000-4-3IEC 61000-4-8IEC 60601-1-2CISPR 11FCC Part 15IEC 60601-1Similar
BiocompatibilityTesting PassedISO 10993-1ISO 10993-1ISO 10993-5ISO 10993-10Similar;additionalBio-compatibilitytestingperformed

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for COAPT Complete Control. The letters "COA" are in orange, and the letters "PT" are in gray. Below the letters is the text "COMPLETE CONTROL" in a smaller font.

The design of the Alpha Control Liner System is similar to the predicate currently on the market. The Alpha Control Liner System is a substitute for one of the components of the predicate device and therefore is intended to work in conjunction with the predicate device. Together, they can be used for external prosthetic fittings of the upper limbs and are suitable for unilateral and bilateral amputations. The subject and predicate device also share the same conditions for use and must only be used for normal daily activities, as listed in the labeling.

The Alpha Control Liner System and the listed predicate device are classified as product code: GXY, cutaneous electrodes under 21 CFR 882.1320.

The Alpha Control Liner System contains hardware that is functionally equivalent to the predicate device. Both devices contain electronics that use various algorithms to process surface electromyography signals into a digital signal that can be used by the pattern recognition algorithm to control a prosthetic limb. The Alpha Control Liner System is designed to operate with the predicate device. The Alpha Control Liner System and the predicate system are the same in that they are not provided with a power source. They are both rated to functioning at similar voltages. The predicate system can operate using a power source between 5.0-20.0VDC. The subject Alpha Control Liner System can operate using a slightly narrower voltage range, from 5.3-16.8 VDC. The slightly narrower operating voltage does not raise any concerns of safety or effectiveness, as the

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for COAPT Complete Control. The word "COAPT" is in large, bold letters, with "CO" in orange and "APT" in gray. Below the word "COAPT" is the phrase "COMPLETE CONTROL" in smaller, gray letters. The logo is simple and modern, with a focus on the company's name.

subject device has passed electrical safety testing under IEC 60601-1, and in-house performance testing.

Another slight difference between the subject device and the predicate, Complete Control System Gen2 is the increased power consumption. The Complete Control System Gen2 uses less than half of the energy as the subject Alpha Control Liner System. This does not raise any concerns of safety and efficacy as the subject passed all required testing for electrical safety under IEC 60601-1, and inhouse performance testing.

The subject and predicate device differ slightly in their interface materials. The predicate device used an EMG Interface Cable to connect the electrodes to the controller electronics. This cable is routed on the outside of the socket by the prosthetist treating the patient. The subject Alpha Control Liner System suspends the cable pathways into a tri-block copolymer gel, fabric covered liner which is worn by the patient. This is an improvement to allow more consistent contact of the electrodes and skin within an exoprosthesis socket. This improvement is possible because the elastic properties of the elastomer and fabric allow the liner to stretch over the patient's limb and therefore have a constant pressure keeping the electrodes in contact with the skin. Since the Alpha Control Liner is worn by the patient, it must pass the signal to the socket. To best preserve the sEMG signals, the Alpha Control liner includes electronics that convert the sEMG signals to a digital signal before passing to the socket. The functions for signal amplification, filtering, and EMG (electromyography) signal transmission of the Alpha Control Liner System are based on the same design of the predicate device. This does not add or modify any existing risks as the subject device has passed all applicable safety and performance testing.

The predicate, Complete Control System Gen2 is used with passivated 316L stainless steel electrodes (K190416). This is the electrodes (K190416) embedded in the Alpha Control Liner System. The use of the same material for the electrodes for predicate and subject device do not raise any additional questions of safety, as the Alpha Control Liner System has been tested for biocompatibility to ISO 10993-1, 10993-5, and 10993-10. The predicate, Complete Control System Gen2 does not include a liner. The liner material of the Alpha Control Liner System does not raise any additional questions of safety, as the Alpha Control Liner System has been tested for biocompatibility to ISO 10993-1, 10993-5, and 10993-10.

The subject, Alpha Control Liner System is intended to be used with the Complete Control System Gen2. Therefore, the Alpha Control Liner System will use the same software tool, the Complete ControlRoom. The use of the same software tool does not raise any questions of safety and effectiveness as the Alpha Control Liner System has passed validation testing showing compatibility with the Complete Control System Gen2 and Complete ControlRoom software.

Both the subject device, Alpha Control Liner System, and predicate device, Complete Control System Gen2, underwent electrical safety testing for IEC 60601-1. Both devices received passing results. Additionally, the predicate device, Complete Control System Gen2, underwent electromagnetic compatibility testing in accordance with IEC 60601-1-2, CISPR 11, and FCC Part 15. These standards are not applicable to the subject device. Alpha Control Liner System, as it does not contain any wireless capabilities.

The predicate device, Complete Control System Gen2, does not come into contact with the user. Therefore, evaluation to the standard ISO 10993-1 for biocompatibility resulted in no tests being applicable. The subject device, Alpha Control Liner, comes into contact with healthy skin of the user. Therefore, evaluation to the standard ISO 10993-1 for biocompatibility resulted in testing to ISO 10993-5 for cytotoxicity, ISO 10993-10 for skin irritation, and ISO 10993-10 for sensitization. The Alpha Control Liner System received passing results for all three biocompatibility tests.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for COAPT Complete Control. The letters "COA" are in orange, while the letters "PT" are in gray. Below the letters is the text "COMPLETE CONTROL" in gray.

The subject device provides muscle contraction signals similar to the EMG Interface Cable of the predicate device. As with the predicate the subject device does not include prosthetic limbs.

8. Non-Clinical Performance Data

As part of demonstrating safety and effectiveness of the Alpha Control Liner System and in showing substantial equivalence to the predicate device that is subject to this Special 510(k) submission, Coapt LLC and WillowWood Global LLC completed a number of non-clinical performance tests. The Alpha Control Liner System meets all the requirements for overall design and electrical safety results, confirming that the design output meets the design inputs and specifications for the device.

The Alpha Control Liner System passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:

  • . Electrical safety testing per IEC 60601-1
  • Cytotoxicity testing per ISO 10993-5 ●
  • Skin irritation testing per ISO 10993-10
  • . Sensitization testing per ISO 10993-10

The Alpha Control Liner System was also tested internally to ensure that it meets device specifications & requirements and operates as intended. The following validation testing was performed on the finished device:

Test NameResult
Liner ThicknessPass
Liner Doffing ForcePass
Donning/Doffing FatiguePass
Liner Umbrella Static LoadingPass
Liner Umbrella Fatigue LoadingPass
Lead Impedance FatiguePass
Lead Impedance - CorrosionPass
Lock Body Static LoadingPass
Lock Body Fatigue LoadingPass
Liner High Pot TestPass
Module High Pot TestPass
Module Level TestPass
Coapt System Integration TestingPass
    1. Clinical Performance Data

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for COAPT Complete Control. The letters "CO" are in orange, while the letters "APT" are in gray. Below the letters is the phrase "COMPLETE CONTROL" in gray.

No human clinical testing was required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate device, have been on the market for many years with proven safety and efficacy for the use of the device. The nonclinical testing detailed in this submission supports the substantial equivalence of the device.

10. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device(s).

The technological characteristics between the Alpha Control Liner System and predicate are remarkably similar and therefore substantially equivalent. The slight differences between devices do not raise new questions of safety and effectiveness as compared to the predicate as the Alpha Control liner System has received passing results for safety testing for electrical safety, ingress protection, and mechanical strength; passing results for performance and usability testing for cabling connection, power on and boot, inputs, and outputs; and validation testing with compatible prosthetic devices.

The Alpha Control Liner System, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).