K Number
K223734

Validate with FDA (Live)

Device Name
ENT EM
Manufacturer
Date Cleared
2023-04-27

(135 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ENT EM is intended as an image-guided planning and navigation system to enable ENT procedures. The device is indicated for any medical condition in which a reference to a rigid anatomical structure can be identified relative to images (CT, CTA, X-Ray, MR, MRA and ultrasound) of the anatomy, such as:

  • Intranasal structures and Paranasal Sinus Surgery
    • Functional endoscopic sinus surgery (FESS)
    • Intranasal structures and paranasal sinus surgery, including revision and distorted anatomy
  • Anterior skull base procedures
Device Description

The Subject Device ENT EM is an image guided planning and navigation system to enable navigated surgery during ENT procedures. It offers guidance for setting up the EM equipment, different patient image registration methods and instrument selection and calibration to allow surgical navigation by using electromagnetic tracking (EM) technology. The device provides different workflows guiding the user through preoperative and intraoperative steps. To fulfill this purpose, it links patient anatomy (using a patient reference) and instruments in the real world or "patient space" to patient scan data or "image space". This allows for the continuous localization of medical instruments and patient anatomy for medical interventions in ENT procedures. The software is installed on a mobile Image Guided Surgery (IGS) platform (Kick 2 Navigation Station or Curve Navigation 17700) to support the surgeon in clinical procedures by displaying tracked instruments in patient's image data. The IGS platforms consist of a mobile Monitor Cart and an EM tracking unit for image guided surgery purposes. ENT EM consists of: Several software modules for registration, instrument handling, navigation and infrastructure tasks, IGS platforms and surgical instruments for navigation, patient referencing and registration.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the device meets those criteria for the Brainlab ENT EM system, which incorporates an AI/ML-based function for pre-registration in surface matching.

Here's the breakdown of the information requested:

1. Table of Acceptance Criteria and Reported Device Performance

Parameter/CharacteristicAcceptance CriteriaReported Device Performance
System Accuracy
Mean Positional Error≤ 2 mmAchieves the same accuracy performance (mean location error ≤ 2 mm) as both predicate and reference device.
Mean Angular Error≤ 2ºAchieves the same accuracy performance (mean trajectory angle error ≤ 2 degrees) as both predicate and reference device.
AI/ML Landmark DetectionEquivalent performance to conventional methodPerformance testing comparing conventional to machine learning based landmark detection were performed showing equivalent performance as in the reference device.
UsabilitySafe and effective for intended user groupSummative usability evaluation in a simulated clinical environment showed ENT EM is safe and effective for use by the intended user group.
Electrical Safety & EMCCompliance with standardsCompliance to IEC 60601-1, AIM 7351731, and IEC 60601-1-2. Tests showed the subject device performs as intended.
Instrument BiocompatibilityBiologically safeBiocompatibility assessment considering different endpoints provided.
Instrument ReprocessingAppropriateness of cleaning/disinfection/sterilizationCleaning and disinfection evaluation/reprocessing validation provided.
Instrument Mechanical PropertiesWithstand typical torsional strengths/torquesEvaluated considering typical torsional strengths, torques, and conditions instruments can be subject to during use.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample size for the test set used for the AI/ML algorithm's performance evaluation. It mentions that "The model's prediction and performance are then evaluated against the test pool. The test pool data is set aside at the beginning of the project."

The data provenance is not explicitly stated regarding country of origin or specific patient demographics. However, it indicates a "controlled internal process" for development and evaluation. It's a static algorithm (locked), suggesting it's developed and tested once rather than continuously learning. The context implies it's retrospective as data was "set aside."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not provided in the text. The document refers to "landmarks delivered by a ML based calculation" and compares its performance to a "conventional" method in the reference device, but it doesn't detail how the ground truth for these landmarks was established for testing.

4. Adjudication Method for the Test Set

The adjudication method for establishing ground truth for the test set is not explicitly described.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

There is no mention of a Multi Reader Multi Case (MRMC) comparative effectiveness study being performed with human readers to assess improvement with AI vs. without AI assistance. The testing primarily focuses on the AI/ML algorithm's performance equivalence to the predicate/reference device's conventional method, and overall system accuracy.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, a form of standalone performance evaluation was done for the AI/ML algorithm. The text states: "Performance testing comparing conventional to machine learning based landmark detection were performed showing equivalent performance as in the reference device." This implies an evaluation of the algorithm's output (landmark detection) without a human reader in the interpretation loop, by comparing its results directly to a "conventional" method.

7. The Type of Ground Truth Used for Performance Testing

The type of ground truth for the AI/ML landmark detection is implicitly based on the "conventional" landmark detection method used in the reference device. The document states "Performance testing comparing conventional to machine learning based landmark detection were performed showing equivalent performance as in the reference device." This suggests the conventional method's output serves as the reference ground truth, or there's an established "true" landmark position that both are compared against. For system accuracy, the ground truth is established through physical measurements of "Mean Positional Error" and "Mean Angular Error" against a known configuration.

8. The Sample Size for the Training Set

The document does not provide the numerical sample size for the training set. It mentions the algorithm was developed using a "Supervised Learning approach" and that "the training process begins with the model observing, learning, and optimizing its parameters based on the training pool data."

9. How the Ground Truth for the Training Set Was Established

The method for establishing ground truth for the training set is not explicitly detailed. It only states that the algorithm was developed using a "Supervised Learning approach" and a "controlled internal process" that defines activities from "inspection of input data to the training and verification." This implies that the training data included true labels or targets for the landmarks that the AI/ML algorithm was trained to detect, but the source or method of obtaining these true labels (e.g., expert annotation, manual registration results) is not specified.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.

April 27, 2023

Brainlab AG Esther Moreno Garcia QM Consultant Olof-Palme-Str. 9 Munich, 81829 Germany

Re: K223734

Trade/Device Name: Ent Em Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: March 28, 2023 Received: March 28, 2023

Dear Esther Moreno Garcia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be f ound in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shuchen Peng -S

Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223734

Device Name ENT EM

Indications for Use (Describe)

ENT EM is intended as an image-guided planning and navigation system to enable ENT procedures. The device is indicated for any medical condition in which a reference to a rigid anatomical structure can be identified relative to images (CT, CTA, X-Ray, MR, MRA and ultrasound) of the anatomy, such as:

· Intranasal structures and Paranasal Sinus Surgery

  • o Functional endoscopic sinus surgery (FESS)
    o Intranasal structures and paranasal sinus surgery, including revision and distorted anatomy

· Anterior skull base procedures

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink symbol on the left, resembling a medical symbol or a stylized brain. To the right of the symbol, the word "BRAINLAB" is written in large, bold, pink letters. The overall design is clean and modern.

510(k) Summary

April 27, 2023

General Information
ManufacturerBrainlab AG; Olof-Palme-Str.9, 81829, Munich, Germany
Establishment Registration8043933
Trade NameENT EM
Classification NameEar, nose, and throat stereotaxic instrument
Product CodePGW
Regulation Number882.4560
Regulatory ClassClass II
PanelNeurology - Ear Nose & Throat
Predicate Device(s)K200723 StealthStation FlexENT, StealthStation S8 ENT Software 1.3
Reference Device(s)K213989 Cranial EM System
Contact Information
Primary ContactAlternate Contact
Esther Moreno GarciaChiara Cunico
QM Consultant - Regulatory AffairsPhone: +49 89 99 15 68 0
Phone: +49 89 99 15 68 0Fax: +49 89 99 15 68 5033
Email: regulatory.affairs@brainlab.comEmail: chiara.cunico@brainlab.com

1. Indication for Use

ENT EM is intended as an image-guided planning and navigation system to enable ENT procedures.

The device is indicated for any medical condition in which a reference to a rigid anatomical structure can be identified relative to images (CT, CTA, X-Ray, MR, MRA and ultrasound) of the anatomy, such as:

  • . Intranasal structures and Paranasal Sinus Surgery
    • Functional endoscopic sinus surgery (FESS) o
    • Intranasal structures and paranasal sinus surgery, including revision and distorted o anatomy
  • . Anterior skull base procedures
    1. Device Description

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink symbol on the left, resembling a brain with interconnected pathways. To the right of the symbol, the word "BRAINLAB" is written in capital letters, also in pink. The overall design is clean and modern.

The Subject Device ENT EM is an image guided planning and navigation system to enable navigated surgery during ENT procedures. It offers guidance for setting up the EM equipment, different patient image registration methods and instrument selection and calibration to allow surgical navigation by using electromagnetic tracking (EM) technology. The device provides different workflows guiding the user through preoperative and intraoperative steps.

To fulfill this purpose, it links patient anatomy (using a patient reference) and instruments in the real world or "patient space" to patient scan data or "image space". This allows for the continuous localization of medical instruments and patient anatomy for medical interventions in ENT procedures.

The software is installed on a mobile Image Guided Surgery (IGS) platform (Kick 2 Navigation Station or Curve Navigation 17700) to support the surgeon in clinical procedures by displaying tracked instruments in patient's image data. The IGS platforms consist of a mobile Monitor Cart and an EM tracking unit for image guided surgery purposes.

ENT EM consists of: Several software modules for registration, instrument handling, navigation and infrastructure tasks, IGS platforms and surgical instruments for navigation, patient referencing and registration.

With this submission, an already existing feature is now performed introducing a new algorithm using artificial intelligence and machine learning (AI/ML). This ML based functionality is used as an aid in the registration step (in surface matching) by allowing a pre-registration based on quide points. These guide points or landmarks are delivered by a ML based calculation. The AI/ML algorithm is a Convolutional Network (CNN) developed using a Supervised Learning approach. The algorithm was developed using a controlled internal process that defines activities from the inspection of input data to the training and verification of the training process begins with the model observing, learning, and optimizing its parameters based on the training pool data. The model's prediction and performance are then evaluated against the test pool. The test pool data is set aside at the beginning of the project. This is a static algorithm (locked).

3. Substantial Equivalence

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE AND REFERENCE DEVICE

CharacteristicPredicate Device K200723Reference Device K213989Subject deviceENT EM
Indicationsfor useThe StealthStationFlexENT™ system with theStealthStation™ ENTsoftware, is intended as anaid for precisely locatinganatomical structures inCranial EM is intended as animage-guided planning andnavigation system to enableneurosurgery procedures.The device is indicated forany medical condition inAll devices are intendedfor the localization ofanatomical structures(navigation). Compared topredicate device, sameindications for use (Sinus
either open or percutaneous ENT procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy. This can include, but is not limited to, the following procedures:- Functional Endoscopic Sinus Surgery (FESS)- Endoscopic Skull Base procedures- Lateral Skull Base procedureswhich a reference to a rigid anatomical structure can be identified relative to images (CT, CTA, X-Ray, MR, MRA and ultrasound) of the anatomy, such as:• Cranial Resection-Resection of tumors and other lesions-Resection of skull-base tumor or other lesions•Intracranial catheter placementsurgery and anterior skull base procedures).
Localization techniqueElectromagnetic tracking: The Side Emitter emits low intensity and varying electromagnetic field which induce small currents in the sensors embedded EM instruments. The position and spatial orientation of the sensors integrated in the EM instruments are calculated in the Instrument Interface Box.Electromagnetic tracking: The Field generator emits low intensity and varying electromagnetic field which induce small currents in the sensors embedded in the EM instruments. The position and spatial orientation of the sensors integrated in the EM instruments are calculated in the Base station.Same localization technique compared to both predicate and reference device. The EM tracking device used is the same as in the reference device.
System accuracyUnder representative worst-case configuration, the StealthStation FlexENTt and S8 Systems With StealthStation S8 ENT vl.3 Software, has demonstrated performance in 3D positional accuracy with a mean error <= 2.0 mm and in trajectory angle accuracy with a mean error <= 2.0 degrees.Under representative worst case configuration, the Cranial EM System is tested to ensure that its mean location error is ≤ 2 mm and its mean trajectory angle error is ≤ 2 degrees.Under representative worst case configuration the subject device achieves the same system accuracy performance (mean location error is ≤ 2 mm and its mean trajectory angle error is ≤ 2 degrees) as both predicate and reference device.
Programming languageC++C++Same programming language compared to both predicate and reference device.
RegistrationmethodsExam-to-Exam RegistrationFeatures:Identify Merge RegistrationManual Merge RegistrationAutomatic Merge RegistrationPatient registrationFeatures:PointMerge registrationTracer registrationTouch registrationLandmark registrationSurface matchingregistrationRegistration methods aresimilar compared to thepredicate device and sameas reference device.
Detection ofanatomicallandmarks/guide pointsN/ADetection ofanatomicallandmarks used for a pre-registration step within theSurfacematchingregistration based on anatlas of the human anatomy.Compared tothereference device, overallfunctionality is the same,but landmarks aredelivered by an AI/MLbased method.Testing showedequivalent performancewith and without AI/ML.
NavigationfeaturesNavigation ofprecalibrated and non-precalibrated instrumentson different viewsenriched with additionalplanning content.Navigation of precalibratedand non-precalibratedinstruments on differentviews enriched withadditional planning content.Same navigation conceptas both predicate andreference device.
IGS PlatformsIGS platforms consisting ofmobile monitor cart and anEM tracking unit:•StealthStation FlexENTIGS platforms consisting ofmobile monitor cart and anEM tracking unit:•Curve Navigation 17700(17700)•Kick 2 NavigationStation EM (18202)Similar to primarypredicate and same IGSplatforms compared toreference device (withminor modifications tostrengthene.g.cybersecurity)
InstrumentsN/AInstruments for patientreferencing, registration andnavigation.Compatible 3rd partyinstruments from KLSMartin.Same instruments asreference device.Former KLS MartinInstruments are nowlegally manufactured byBrainlab

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized symbol on the left and the word "BRAINLAB" in capital letters on the right. The symbol appears to be a stylized representation of the brain. The color of the logo is pink.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized symbol on the left and the word "BRAINLAB" on the right. The symbol is a pink abstract design, and the word "BRAINLAB" is written in pink capital letters. The logo is simple and modern.

4. Performance Data

The following testing was conducted on the Subject Device to establish substantial equivalence with the predicate device:

Software Verification and Validation Testing

To address differences in terms of software development and the specific implementation of features, software verification and validation testing were conducted and documentation was

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized caduceus symbol on the left, followed by the word "BRAINLAB" in capital letters. The color of the logo is a bright pink.

provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." This included product specifications, risk analysis and software verification testing. The software (ENT EM 1.2) for this device was considered as a "major" level of concern.

In particular, ENT EM implements a feature for landmark detection in the registration software using AI/ML. The same feature without AI/ML was already used in the reference device. Performance testing comparing conventional to machine learning based landmark detection were performed showing equivalent performance as in the reference device.

System accuracy testing

The positional and angular navigation accuracy for ENT procedures of the Subject Device including the software, the platforms and the instruments was evaluated considering a realistic clinical setup and representative worst case scenarios. The results show the following acceptance criteria are fulfilled:

  • . Mean Positional Error of the placed instrument's tip ≤ 2 mm
  • . Mean Anqular Error of the placed instrument's axis ≤ 2º

Therefore, the Subject Device achieves the same accuracy performance as both predicate and reference device.

Usability

To account for differences regarding the intended user group compared to the reference device and the different workflow steps and GUI implementation compared to the predicate device a summative usability evaluation in a simulated clinical environment was carried out, where the user had to go through all main steps and frequently used functions. This showed ENT EM is safe and effective for use by the intended user group.

Electrical safety and electromagnetic compatibility (EMC)

Compliance to electrical safety, RFID and EMC was evaluated on the Subject device according to the standards: IEC 60601-1, AIM 7351731 and IEC 60601-1-2. The tests have shown that the subject device performs as intended.

Instruments

Due to the change in legal manufacturer of the skull reference instruments, the biocompatibility testing and reprocessing validation for the instruments was provided in order to demonstrate their biological safety and appropriateness of the cleaning, disinfection and sterilization methods. This included:

  • Biocompatibility assessment considering different end points .
  • . Cleaning and disinfection evaluation/reprocessing validation

The mechanical properties of instruments were also evaluated considering the typical torsional strengths, torques and conditions the instruments can be subject to during their use.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink symbol on the left, resembling a medical symbol or a stylized brain. To the right of the symbol, the word "BRAINLAB" is written in large, bold, pink letters. The overall design is clean and modern.

No clinical testing was needed for the Subject Device since the EM tracking technology in the scope of image guided surgery for the included indications for use is well established in the market. Bench testing demonstrated that the device performs as the predicate and that no different questions on safety or effectiveness were raised.

5. Conclusion

The comparison of the Subject Device with the predicate and reference device shows that ENT EM has similar functionality, intended use and technological characteristics as the predicate and reference devices. Based on this comparison and the performance testing conducted, the Subject Device is considered substantially equivalent to the predicate device.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).