K Number
K223653

Validate with FDA (Live)

Device Name
BD Vaginal Panel
Date Cleared
2023-03-06

(90 days)

Product Code
Regulation Number
866.3975
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Vaginal Panel is an automated qualitative in vitro diagnostic test for the direct detection of DNA targets from bacteria associated with bacterial vaginosis (qualitative results reported based on detection and quantitation of targeted organism markers), Candida species associated with vulvovaginal candidiasis, and Trichomonas vaginalis from vaginal swabs in patients who are symptomatic for vaginosis. The test utilizes real-time polymerase chain reaction (PCR) for the amplification of specific DNA targets and utilizes fluorogenic target-specific hybridization probes to detect and differentiate DNA from:

  • · Bacterial vaginosis markers (Individual markers not reported) Lactobacillus spp. (L. crispatus and L. jensenii) Gardnerella vaginalis Atopobium vaginae Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) Megasphaera-1
  • · Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis)
  • · Candida glabrata
  • Candida krusei
  • Trichomonas vaginalis

The BD Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis.

The BD Vaginal Panel is available for use with the BD MAX™ System or the BD COR™ System.

Device Description

As with the existing BD Vaginal Panel for use with the BD MAX™ System (K201017), the BD COR™ PX/MX (BD COR) high throughput system conducts sample extraction steps to isolate and concentrate DNA which is then amplified to detect specific sequences for diagnostic purposes.

The BD COR™ System is designed to allow the user to place clinical specimens directly into designated transport racks to be loaded into the System. Once the specimens are loaded, the System will perform the necessary pre-analytical steps such as vortexing, aliquoting into a molecular tube with the correct diluent, sorting/grouping of the secondary samples for testing by assay, pre-warming and cooling of the sample (where required), and transport of the sample into a molecular analyzer, where extraction, amplification and detection will take place.

Additionally, the steps of ordering tests on the instrument for specific samples will be managed directly by the user interaction with the Laboratory Information System (LIS), which communicates drectly with the instrument.

Once the clinical specimens are received in the laboratory and loaded into the transport racks, the user will not be required to directly handle the specimen again reporting and removal from the system.

AI/ML Overview

The BD Vaginal Panel is an automated qualitative in vitro diagnostic test for the direct detection of DNA targets from bacteria associated with bacterial vaginosis (qualitative results reported based on detection and quantitation of targeted organism markers), Candida species associated with vulvovaginal candidiasis, and Trichomonas vaginalis from vaginal swabs in patients who are symptomatic for vaginosis.

The study aimed to demonstrate the equivalence of the BD Vaginal Panel on the BD COR™ System to its performance on the BD MAX™ System, which is the legally marketed predicate device.

  1. Table of Acceptance Criteria and Reported Device Performance:
Performance MetricAcceptance Criteria (Predicate: BD MAX™ System)Reported Device Performance (BD COR™ System)
Precision (Qualitative)Similar performance to BD MAX™ System.BD COR™ System (Example for some targets): Bacterial Vaginosis: True Negative: 100%, BV Negative: 100%, BV High Negative: 79.7%, Low Positive: 99.0%, Moderate Positive: 99.5%. Candida glabrata: True Negative: 100%, Low Positive: 100%. Candida krusei: True Negative: 100%, Low Positive: 100%. Candida albicans: True Negative: 100%, Low Positive: 100%, Moderate Positive: 100%. Trichomonas vaginalis: True Negative: 100%, Low Positive: 100%, Moderate Positive: 100%. (All 95% CI lower bounds generally above 90% for positives and negatives, with High Negative BV being an outlier which is acceptable given its definition.)
Reproducibility (Qualitative)Similar performance to BD MAX™ System across multiple sites and operators.BD COR™ System (Overall averages): Bacterial Vaginosis: True Negative: 100%, BV Negative: 100%, BV High Negative: 72.9%, Low Positive: 100%, Moderate Positive: 100%. Candida glabrata: True Negative: 100%, Low Positive: 100%. Candida krusei: True Negative: 100%, Low Positive: 100%. Candida albicans: True Negative: 99.6%, Low Positive: 100%, Moderate Positive: 100%. Trichomonas vaginalis: True Negative: 100%, Low Positive: 100%, Moderate Positive: 100%. (All 95% CI lower bounds generally high, indicating good reproducibility, with BV High Negative being an expected lower value.)
Analytical Sensitivity (LoD)Equivalence to BD MAX™ System with equivalence interval of [-6% of reference mean, +6% of reference mean] for mean Ct.score. For High Negative, overlapping 95% CIs.BD COR™ System: Equivalence established for all Vaginosis and Vaginitis targets at Low Positive and Moderate Positive levels based on the TOST analysis and equivalence intervals (all within the specified range). For High Negative Candida spp., overlapping 95% confidence intervals were observed, demonstrating equivalence. (See Tables 10 and 11 for detailed Ct.Score/SDPA differences and equivalence establishment for each target.)
Cross-Contamination RateNot explicitly stated as a numerical criterion for BD COR™, but demonstrated to meet acceptable levels.One false positive result out of 543 negative samples, resulting in a contamination rate of 0.18% (95% CI: 0.03-1.04%), which met the predefined study acceptance criteria.
Clinical Agreement (PPA & NPA)High percentage agreement (PPA and NPA) between BD COR™ and BD MAX™ results.BD COR™ System (Overall Averages): BV Contrived: Average PPA: 99.5%, Average NPA: 100%. BV Natural: Average PPA: 97.8%, Average NPA: 95.8%. C. glabrata: Average PPA: 100%, Average NPA: 100%. C. krusei: Average PPA: 100%, Average NPA: 100%. Candida Group: Average PPA: 99.4%, Average NPA: 98.9%. TV: Average PPA: 99.7%, Average NPA: 100%. (All bootstrap 95% CIs are high, demonstrating excellent clinical agreement.)
Non-Reportable RateDemonstrate a low non-reportable rate for combined targets.BD COR™ System (Overall Initial Rate): 0.6% (13/2047) (95% CI: 0.4, 1.1). BD COR™ System (Overall Final Rate after repeat testing): 0.0% (0/2044) (95% CI: 0.0, 0.2). This indicates effective resolution of initial non-reportable events.
  1. Sample sizes used for the test set and the data provenance:

    • Precision Study: Panel members (spiked in simulated vaginal matrix) were tested in 12 days, 2 runs/day, 2 replicates/panel, for a total of 48 runs. This resulted in varying numbers for total N for each target and level (e.g., 288 for BV True Negative, 48 for Candida Low Positive, etc.; see Table 3 for details). Data provenance is internal laboratory testing.
    • Reproducibility Study: Similar panel members to the precision study. Tested at 3 sites (2 external, 1 internal) over 8 days, with 2 operators performing 2 runs on alternate days, for a total of 48 runs. This resulted in varying numbers for total N for each target, level, and site (e.g., 192 for BV True Negative per site, 32 for BV High Negative per site; see Table 6 for details). Data provenance is internal and external laboratory testing.
    • Analytical Sensitivity Confirmation Study: 20 panels of Vaginosis and/or Vaginitis targets at varying concentration levels (1.99x LoD, 5x LoD, C5) with 48 replicates each. Samples were prepared by spiking organisms into simulated vaginal matrix. Data provenance is internal laboratory testing.
    • Cross-Contamination Study: 543 negative samples (interspersed with high positive samples). Data provenance is internal laboratory testing.
    • Clinical Agreement Study: 700 panel members. These included:
      • Clinical vaginal specimens from internal collections.
      • Pooled previously collected clinical specimens.
      • High positive clinical specimens spiked.
      • Contrived samples created by spiking organisms into negative vaginal matrix or simulated vaginal matrix.
      • BV Contrived panel members prepared with different BV marker combinations using simulated vaginal matrix.
      • BV Natural samples derived from Cgroup, TV, and negative vaginitis panel members in natural vaginal matrix.
      • Data provenance is a mix of internal collections and contrived samples.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    The document does not mention the use of experts or their qualifications for establishing ground truth. For the analytical studies (precision, reproducibility, analytical sensitivity, cross-contamination), ground truth was established by spiking known concentrations of target organisms/DNA into simulated or negative matrices. For the clinical agreement study, "BD MAX™ results served as the reference" meaning the predicate device's results were used as the comparator, not human expert consensus.

  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    For the clinical agreement study, the positive or negative status of a panel member using the BD MAX™ System (the reference) was defined by ">2 out of 3 evaluable results obtained on the BD MAX™". This suggests a form of 3-reader consensus (with the BD MAX™ itself acting as a 'reader' in triplicate, or at least three independent runs determining the status). Equivocal BD MAX™ comparator results were defined as "one positive, one negative, and one non-evaluable result from the BD MAX™." No human adjudication is mentioned.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC comparative effectiveness study was done or reported in this document. The device is an automated in vitro diagnostic test (nucleic acid amplification test) and does not involve human readers interpreting images or other data with or without AI assistance. The comparison is between two automated systems (BD COR™ vs. BD MAX™).

  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Yes, the entire study focuses on the standalone performance of the BD Vaginal Panel on the BD COR™ System. The device is an automated diagnostic test, meaning it operates "algorithm only" without human-in-the-loop performance for result generation. The human role is in sample loading and result interpretation (based on the device's output), not in forming the initial diagnostic call.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Analytical Studies (Precision, Reproducibility, Analytical Sensitivity, Cross-Contamination): Laboratory-contrived ground truth through spiking known concentrations of target organisms/DNA into simulated vaginal matrix or negative matrix.
    • Clinical Agreement Study: The results from the predicate device (BD MAX™ System) were used as the reference ("ground truth") for comparison. For panel member status, a consensus method of ">2 out of 3 evaluable results obtained on the BD MAX™" was used.
  7. The sample size for the training set:

    The document does not explicitly mention a "training set" in the context of machine learning or AI algorithm development. The BD Vaginal Panel is a PCR-based assay. The performance studies described are for analytical and clinical validation of the device, not for training a new algorithm. The development of the assay (e.g., probe design, primer selection) would have involved extensive R&D, but the concept of a "training set" as typically used in AI/ML is not applicable here.

  8. How the ground truth for the training set was established:

    As a PCR-based diagnostic, it's not an AI system that relies on a "training set" in the conventional sense to learn to make predictions. The "ground truth" for developing such an assay typically relies on purified nucleic acid from known organisms, synthetic controls, and well-characterized clinical samples to ensure the primers and probes are specific and sensitive for the intended targets. The document does not describe the specific methods for establishing ground truth during the assay development phase.

{0}------------------------------------------------

March 6, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Becton, Dickinson and Company Joseph Basore Staff Regulatory Affairs Specialist 7 Loveton Circle Sparks, Maryland 21152

Re: K223653

Trade/Device Name: BD Vaginal Panel Regulation Number: 21 CFR 866.3975 Regulation Name: Device That Detects Nucleic Acid Sequences From Microorganisms Associated With Vaginitis And Bacterial Vaginosis Regulatory Class: Class II Product Code: PQA, OUY, OOI, NSU Dated: December 5, 2022 Received: December 6, 2022

Dear Joseph Basore:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Noel J. Gerald -S

Noel J. Gerald, Ph.D. Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223653

Device Name BD Vaginal Panel

Indications for Use (Describe)

The BD Vaginal Panel is an automated qualitative in vitro diagnostic test for the direct detection of DNA targets from bacteria associated with bacterial vaginosis (qualitative results reported based on detection and quantitation of targeted organism markers), Candida species associated with vulvovaginal candidiasis, and Trichomonas vaginalis from vaginal swabs in patients who are symptomatic for vaginosis. The test utilizes real-time polymerase chain reaction (PCR) for the amplification of specific DNA targets and utilizes fluorogenic target-specific hybridization probes to detect and differentiate DNA from:

  • · Bacterial vaginosis markers (Individual markers not reported) Lactobacillus spp. (L. crispatus and L. jensenii) Gardnerella vaginalis Atopobium vaginae Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) Megasphaera-1
  • · Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis)
  • · Candida glabrata
  • Candida krusei
  • Trichomonas vaginalis

The BD Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis.

The BD Vaginal Panel is available for use with the BD MAX™ System or the BD COR™ System.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

BD Vaginal Panel

Summary Preparation Date:

12/5/2022

Submitted by:

BD Integrated Diagnostic Solutions Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152

Contact:

Joseph Basore, Ph.D., RAC Staff Regulatory Affairs Specialist Tel: 616-301-4068 Email: Joseph.Basore@bd.com

Proprietary Names:

For the instrument: BD COR™ PX/MX System

For the assay: BD Vaginal Panel

Common Names:

For the instrument: High-throughput molecular system

For the assay: Bacterial Vaginosis Assay Vaginitis Assay TV Assay Candida Assay

Regulatory Information

Regulation section: 21 CFR 866.3975 – Device that detects nucleic acid sequences from microorganisms associated with vaginitis and bacterial vaginosis

Classification: Class II (Special Controls)

Panel: Microbiology (83)

Product Code(s):

{4}------------------------------------------------

Vaginitis and Bacterial Vaginosis Nucleic Acid Detection System PQA OUY Trichomonas vaginalis Nucleic Acid Amplification Test System

  • OOI Real Time Nucleic Acid Amplification System
  • Instrumentation for Clinical Multiplex Test Systems NSU

Predicate Device

BD MAX Vaginal Panel (DEN160001, K191957, K201017)

Device Establishment

Registration Number: 1119779

Intended Use

The BD Vaginal Panel is an automated qualitative in vitro diagnostic test for the direct detection of DNA targets from bacteria associated with bacterial vaginosis (qualitative results reported based on detection and quantitation of targeted organism markers), Candida species associated with vulvovaginal candidiasis, and Trichomonas vaginalis from vaginal swabs in patients who are symptomatic for vaginitis/vaginosis. The test utilizes real-time polymerase chain reaction (PCR) for the amplification of specific DNA targets and utilizes fluorogenic target-specific hybridization probes to detect and differentiate DNA from:

  • · Bacterial vaginosis markers (Individual markers not reported) Lactobacillus spp. (L. crispatus and L. jensenii) Gardnerella vaginalis Atopobium vaginae Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) Megasphaera-1
  • Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis)
  • · Candida glabrata
  • Candida krusei
  • Trichomonas vaginalis

The BD Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis.

The BD Vaginal Panel is available for use with the BD MAX™ System or the BD COR™ System.

Special Conditions for Use Statement: For Prescription Use Only

Special Instrument Requirements: BD CORTM PX/MX System

Device Description

As with the existing BD Vaginal Panel for use with the BD MAX™ System (K201017), the BD COR™ PX/MX (BD COR) high throughput system conducts sample extraction steps to isolate and concentrate DNA which is then amplified to detect specific sequences for diagnostic purposes.

The BD COR™ System is designed to allow the user to place clinical specimens directly into designated transport racks to be loaded into the System. Once the specimens are loaded, the System will perform the necessary pre-analytical steps such as vortexing, aliquoting into a molecular tube with the

CONFIDENTIAL AND PROPRIETARY

{5}------------------------------------------------

correct diluent, sorting/grouping of the secondary samples for testing by assay, pre-warming and cooling of the sample (where required), and transport of the sample into a molecular analyzer, where extraction, amplification and detection will take place.

Additionally, the steps of ordering tests on the instrument for specific samples will be managed directly by the user interaction with the Laboratory Information System (LIS), which communicates drectly with the instrument.

Once the clinical specimens are received in the laboratory and loaded into the transport racks, the user will not be required to directly handle the specimen again reporting and removal from the system.

Test Principle

The BD Vaginal Panel when performed on the BD CORTM System is designed for use with the BD Molecular Swab Collection kit. Samples are transported to the testing laboratory in BD Molecular Swab Sample Buffer Tubes. The BD COR™ MX Instrument, when combined with the BD COR™ PX Instrument, is to be used for automated sample preparation, extraction of nucleic acids from multiple specimen types, as well as the automated amplification of target nucleic acid sequences by fluorescence-based real-time PCR.

The BD Vaginal Panel extraction reagents are dried in 96-well microtiter plates that contain binding magnetic affinity beads and Sample Processing Control (SPC). Each well is capable of binding and eluting sample nucleic acids. The SPC monitors the integrity of the reagents and the process steps involved in DNA extraction, amplification and detection, as well as for the presence of potential assay inhibitors.

The BD Vaginal Panel liquid reagent plate includes Wash, Elution and Neutralization buffers. The beads (described above), together with the bound nucleic acids, are washed and the nucleic acids are eluted by a combination of heat and pH. Eluted DNA is neutralized and transferred to the Amplification reagent (described below) to rehydrate the PCR reagents. After reconstitution, the BD COR™ PX/MX System dispenses a fixed volume of PCR-ready solution containing extracted nucleic acids into the BD PCR Cartridge.

Microvalves in the BD PCR Cartridge are sealed by the system prior to initiating PCR in order to contain the amplification mixture and thus prevent evaporation and contamination. The amplified DNA targets are detected using hydrolysis (TaqMan®) probes, labeled at one end with a fluorescent reporter dye (fluorophore), and at the other end, with a quencher moiety. Probes labeled with different fluorophores are used to detect the target analytes in different optical channels of the BD COR™ PX/MX System. When the probes are in their native state, the fluorescence of the fluorophore is quenched due to its proximity to the quencher. However, in the presence of target DNA, the probes hybridize to their complementary sequences and are hydrolyzed by the 5'-3' exonuclease activity of the DNA polymerase as it synthesizes the nascent strand along the DNA template. As a result, the fluorophores are separated from the quencher molecules and fluorescence is emitted. The BD COR™ PX/MX System monitors these signals at each cycle of the PCR and interprets the data at the end of the reaction to provide qualitative test results for each vaginitis analyte as well as qualitative results for bacterial vaginosis based on detection and quantitation of targeted bacterial vaginosis markers.

Substantial Equivalence1

1 The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or ther application by the courts.

{6}------------------------------------------------

Table 3 provides the similarities and differences between the submitted device and the legally marketed predicate device.

Table 1: Comparison to Predicate Device
-----------------------------------------
ItemsBD Vaginal PanelPredicate - BD MAXTM Vaginal Panel,BD MAXTM System (K201017)
RegulationSame866.3975
Product CodeSamePQA, OUY, OOI, NSU
Device ClassSameII
Intended UseThe BD Vaginal Panel is an automatedqualitative in vitro diagnostic test for thedirect detection of DNA targets frombacteria associated with bacterialvaginosis (qualitative results reportedbased on detection and quantitation oftargeted organism markers), Candidaspecies associated with vulvovaginalcandidiasis, and Trichomonas vaginalisfrom vaginal swabs in patients who aresymptomatic for vaginitis/vaginosis. Thetest utilizes real-time polymerase chainreaction (PCR) for the amplification ofspecific DNA targets and utilizesfluorogenic target-specific hybridizationprobes to detect and differentiate DNAfrom: Bacterial vaginosis markers(Individual markers notreported) Lactobacillus spp. ( L.crispatus and L. jensenii ) Gardnerella vaginalis Atopobium vaginae Bacterial VaginosisAssociated Bacteria-2(BVAB-2) Megasphaera-1 Candida spp. ( C. albicans, C.tropicalis, C. parapsilosis, C.dubliniensis ) Candida glabrata Candida krusei Trichomonas vaginalis The BD Vaginal Panel is intended to aidin the diagnosis of vaginal infections inThe BD MAX Vaginal Panel performedon the BD MAX System is an automatedqualitative in vitro diagnostic test for thedirect detection of DNA targets frombacteria associated with bacterialvaginosis (qualitative results reportedbased on detection and quantitation oftargeted organism markers), Candidaspecies associated with vulvovaginalcandidiasis, and Trichomonas vaginalisfrom vaginal swabs in patients who aresymptomatic for vaginitis/vaginosis. Thetest utilizes real-time polymerase chainreaction (PCR) for the amplification ofspecific DNA targets and utilizesfluorogenic target-specific hybridizationprobes to detect and differentiate DNAfrom: Bacterial vaginosis markers(Individual markers notreported) Lactobacillus spp. ( L.crispatus and L. jensenii ) Gardnerella vaginalis Atopobium vaginae Bacterial VaginosisAssociated Bacteria-2(BVAB-2) Megasphaera-1 Candida spp. ( C. albicans, C.tropicalis, C. parapsilosis, C.dubliniensis ) Candida glabrata Candida krusei Trichomonas vaginalis
ItemsBD Vaginal PanelPredicate - BD MAXTM Vaginal Panel,BD MAXTM System (K201017)
women with a clinical presentationconsistent with bacterial vaginosis,vulvovaginal candidiasis andtrichomoniasis.The BD Vaginal Panel is available foruse with the BD MAXTM System or theBD CORTM System.The BD MAX Vaginal Panel is intendedto aid in the diagnosis of vaginalinfections in women with a clinicalpresentation consistent with bacterialvaginosis, vulvovaginal candidiasis andtrichomoniasis.
Indications for UseSameSymptomatic patients
Specimen TypeSameClinician and patient-collected femalevaginal swab
Collection/ TransportDeviceBD Molecular Swab Collection KitBD MAX UVE Specimen Collection KitBD Molecular Swab Collection Kit
TechnologySamePCR
Organisms DetectedSameLactobacillus spp. (L. crispatusand L.jensenii) Gardnerella vaginalis Atopobium vaginae Bacterial Vaginosis AssociatedBacteria-2 (BVAB-2) Megasphaera-1 Candida spp. (C. albicans, C.tropicalis, C. parapsilosis, C.dubliniensis) Candida glabrata Candida krusei Trichomonas vaginalis
Sample Prep/ResultsAutomated by BD CORTM SystemPartially Automated by BD MAXTMSystem

{7}------------------------------------------------

Analytical Performance

Analytical performance of the BD Vaginal Panel was evaluated on the BD MAX™ System and the results may be found under DEN160001 with updates described in subsequent submissions K191957 and K201017. As the formulation of the BD Vaginal Panel reagents for use on the BD COR™ System has not changed from those used with the BD MAX™ System, certain analytical studies performed and documented in the package insert on BD MAX™ are equally applicable to the BD COR™ System (specimen stability, analytical sensitivity, inclusivity, cross-reactivity, and interfering substances). The following sections describe the analytical studies that were performed to demonstrate that the assay performance, when used on

CONFIDENTIAL AND PROPRIETARY

{8}------------------------------------------------

the BD COR™ System, is unchanged from the performance demonstrated on the BD MAX™ System. The new analytical studies included: within-laboratory precision and multi-site reproducibility, confirmation of the analytical sensitivity and a cross-contamination study, all performed on the BD COR™ System.

Precision for BD COR™ System

The precision of the BD Vaginal Panel on the BD COR™ System was confirmed to be equivalent to that of the BD MAX™ System. Within-laboratory precision was evaluated for the BD Vaginal Panel on both the BD MAX™ System and the BD COR™ System at one site with one reagent lot. Testing was performed over twelve (12) days, two (2) runs per day, two (2) replicates per panel, for a total of 48 runs. Panel members were made of target organisms (or plasmid DNA for Megasphaera-1 and BVAB-2) spiked in simulated vaginal matrix. Bacterial vaginosis panel members were prepared at varying concentrations of multiple targeted species with sample compositions designed to generate low positive, moderate positive, high negative, or negative results for bacterial vaginosis. For Candida and Trichomonas vaginalis panel members, the target organisms were spiked at concentrations based on the assay LoD. Table 4 describes the panel members evaluated.

The qualitative and quantitative precision results for BD COR™ System are presented in Table 5. Second Derivative Peak Abscissa (SDPA), an internal criterion used to determine a final assay result, was selected as a means of assessing quantitative assay reproducibility. Mean SDPA values with variance components (SD and % CV) are shown in Table 6 and Table 7.

Concentration DesignationBacterial Vaginosis(% of positive results expectedbased on the organism composition)Candida and Trichomonas vaginalis(x LoD)
ModeratePositive~100≥2 to ≤5
Low Positive~95<2
High BVNegative~20-80
BV Negative~5
True Negative0 (No Target)No Target

Table 2: Precision Study Spiking Concentrations

{9}------------------------------------------------

TargetLevelBD CORTMBD MAXTM
N TotalN Correct%Correct95% CI LCL95% CI UCLN TotalN Correct%Correct95% CI LCL95% CI UCL
BacterialVaginosisTrueNegativea28828810098.710028828810098.7100
BV Negativea484810092.6100484810092.6100
BV HighNegativeb,d19215379.773.484.819216887.582.191.5
Low Positivec28828599.097.099.628828810098.7100
ModeratePositived19219199.597.199.919219210098.0100
CandidaglabrataTrueNegativea72072010099.510072071999.999.2100
Low Positive484810092.6100484810092.6100
CandidakruseiTrueNegativea72072010099.510072072010099.5100
Low Positive484810092.6100484810092.6100
CandidaalbicansTrueNegativea72072010099.510072071899.799.099
Low Positive484810092.6100484810092.6100
ModeratePositive484810092.6100484810092.6100
TrichomonasvaginalisTrueNegativea72072010099.510072071999.999.2100
Low Positive484810092.6100484810092.6100
ModeratePositive484810092.6100484810092.6100

Table 3: Qualitative Precision Study Results Summary for Vaginosis and Vaginitis Targets for BD COR™ and BD MAX™ Systems

a For the True Negative and BV Negative levels, the reported agreement indicates the percent of negative results.

b For the High Negative category, the reported agreement indicates the percent of positive results.

C Performance includes combined results from replicates of six panel members containing different organism compositions.

d Performance includes combined results from replicates of four panel members containing different organism compositions.

{10}------------------------------------------------

Within Run(Residual)RunDayTotal
TargetLevelNMeanSD%CVSD%CVSD%CVSD%CV
Candida glabrataLowPositive4830.880.832.681.003.23001.304.20
Candida kruseiLowPositive4828.870.230.790.130.45000.260.91
Candida albicansLowPositive4827.840.260.94000.100.350.281.00
Candida albicansModeratePositive4827.120.200.750.090.32000.220.82
TrichomonasvaginalisLowPositive4832.700.341.040.200.60000.391.20
TrichomonasvaginalisModeratePositive4831.630.300.960.050.15000.310.98

Table 4: Quantitative Precision Summary of Variance Components by Vaginitis Target for BD COR™ System

Table 5: Quantitative Precision Summary of Variance Components by Vaginitis Target for BD
MAXTM Svstem
Within Run(Residual)RunDayTotal
TargetLevelNMeanSD%CVSD%CVSD%CVSD%CV
Candida glabrataLowPositive4831.330.882.810.732.340.361.151.203.83
Candida kruseiLowPositive4828.870.321.110.160.560.170.580.401.37
Candida albicansLowPositive4828.240.250.900.090.33000.270.96
Candida albicansModeratePositive4827.260.260.95000.170.620.311.14
TrichomonasvaginalisLowPositive4832.640.461.4200000.461.42
TrichomonasvaginalisModeratePositive4831.630.240.7600000.240.76

Reproducibility for BD COR™ System

The reproducibility of the BD Vaginal Panel on the BD COR™ System was confirmed to be equivalent to that of the BD MAX™ System using the same sample categories as defined above for the precision study (Table 4). To evaluate the site-to-site reproducibility, testing was performed at 3 sites (2 external and 1 internal) over 8 days. At each site, 2 operators performed 2 runs on alternate days, for a total of 48 runs. Reproducibility results are summarized in Table 8.

The qualitative and quantitative reproducibility results across sites and by target are presented in Table 9. Second Derivative Peak Abscissa (SDPA), an internal criterion used to determine a final assay result, was selected as a means of assessing quantitative assay reproducibility. Mean SDP A values with variance components (SD and % CV) for BD COR™ System are shown in Table 10 and Table 11.

{11}------------------------------------------------

TargetLevelTest SiteN TotalN CorrectBD CORTMBD MAXTM
%Correct95% CI LCL95% CI UCLN Correct% Correct95% CI LCL95% CI UCL
True Negativea119219210098.010019210098.0100
219219210098.010019210098.0100
319219210098.010019210098.0100
Overall57657610099.310057610099.3100
BV Negativea1323210089.31003210089.3100
2323210089.31003210089.3100
3323210089.31003210089.3100
Overall969610096.21009610096.2100
Bacterial VaginosisBV High Negativeb,c1322578.161.289.01340.625.557.7
2321753.136.469.12990.675.896.8
3322887.571.995.03093.879.998.3
Overall967072.963.380.87275.065.582.6
Low Positived1646410094.31006398.491.799.7
2646410094.31006296.989.399.1
3636310094.31006310094.3100
Overall19119110098.010018898.495.599.5
Moderate Positived1323210089.31003210089.3100
2323210089.31003210089.3100
3323210089.31003210089.3100
Overall969610096.21009610096.2100
Candida glabrataTrue Negativea116016010097.710016010097.7100
216016010097.710016010097.7100
315915910097.610015597.593.799.0
Overall47947910099.210047599.297.999.7
Low Positive1323210089.31003210089.3100
2323210089.31003210089.3100
3323210089.31003210089.3100
Overall969610096.21009610096.2100
Candida kruseiTrue Negativea116016010097.710016010097.7100
216016010097.710016010097.7100
315915910097.610015910097.6100
Overall47947910099.210047910099.2100
Low Positive1323210089.31003210089.3100
2323210089.31003210089.3100
3323210089.31003210089.3100
Overall969610096.21009610096.2100
Candida albicansTrue Negativea116016010097.710016010097.7100
216015999.496.599.916010097.7100
315915899.496.599.915698.194.699.4
Overall47947799.698.599.947699.498.299.8
Low Positive1323210089.31003210089.3100
2323210089.31003210089.3100
3323210089.31003210089.3100
Overall969610096.21009610096.2100
TargetLevelTest SiteN TotalN CorrectBD CORTMBD MAXTM
%Correct95% CI LCL95% CI UCLN Correct% Correct95% CI LCL95% CI UCL
Moderate Positive2323210089.31003210089.3100
3323210089.31003210089.3100
Overall969610096.21009610096.2100
True Negativea116016010097.710016010097.7100
216016010097.710016010097.7100
315915910097.610015910097.6100
Overall47947910099.210047910099.2100
Trichomonas vaginalisLow Positive1323210089.31003210089.3100
2323210089.31003210089.3100
3323210089.31003210089.3100
Overall969610096.21009610096.2100
Moderate Positive1323210089.31003210089.3100
2323210089.31003210089.3100
3323210089.31003210089.3100
Overall969610096.21009610096.2100

Table 6: Qualitative Reproducibility by Target and Site for BD COR™ System and BD MAX™ System

CONFIDENTIAL AND PROPRIETARY

{12}------------------------------------------------

a For the True Negative and BV Negative levels, the reported agreement indicates the percent of negative results.

b For the BV High Negative category, the reported agreement indicates the percent of positive results.

© Performance includes results from replicates of a single panel member.

d Performance includes combined results from replicates of two panel members containing different organism compositions.

Table 7: Quantitative Reproducibility Site-to-Site Summary by Vaginitis Target for BD COR™ System

Within Run(Residual)Between RunBetween DayBetweenSiteTotal
TargetLevelNMeanSD%CVSD%CVSD%CVSD%CVSD%CV
CandidaglabrataLow Positive9630.300.531.7600000.260.870.591.96
Candida kruseiLow Positive9628.930.200.710.070.2300000.220.74
CandidaLow Positive9626.690.381.44000.160.590.070.250.421.57
albicansModeratePositive9626.080.301.140.040.170.060.240.120.480.331.27
TrichomonasLow Positive9632.850.381.1700000.150.450.411.25
vaginalisModeratePositive9631.660.300.93000.040.13000.300.94

{13}------------------------------------------------

Within Run(Residual)Between RunBetween DayBetweenSiteTotal
TargetLevelNMeanSD%CVSD%CVSD%CVSD%CVSD%CV
CandidaglabrataLow Positive9630.950.712.28000.200.640.511.660.892.89
Candida kruseiLow Positive9629.090.421.440.150.5300000.451.54
Candida albicansLow Positive9627.340.421.550.200.7200000.471.71
ModeratePositive9626.490.421.59000.050.18000.431.60
TrichomonasvaginalisLow Positive9632.730.461.40000.210.650.240.730.561.71
ModeratePositive9631.690.421.3100000.080.250.421.34

Table 8: Quantitative Reproducibility Site-to-Site Summary by Vaginitis Target for BD MAX™ System

Quality Controls

External Control materials are not provided by BD; however, Quality Control procedures are included in the package insert. Various types of External Controls are recommended to allow the user to select the most appropriate for their laboratory quality control program:

  • Commercially available positive control materials ●
  • Trichomonas vaginalis (ATCC 30001) ●
  • . Candida albicans (ATCC 10231)
  • Candida glabrata (ATCC 2001) ●
  • . Candida krusei (ATCC 6258)

The assay includes a Specimen Processing Control (SPC) that is present in the Extraction Tube. The SPC monitors DNA extraction steps, thermal cycling steps, reagent integrity and the presence of inhibitory substances.

Analytical Sensitivity Confirmation for BD COR™ System

The analytical sensitivity/Limit of Detection (LoD) of the BD Vaginal Panel on the BD COR™ System was confirmed to be equivalent to that of the BD MAX System with use of the BD Molecular Swab Collection Kit. The study design included using 20 panels of Vaginosis and/or Vaginitis targets at varying concentration levels created using the LoD previously determined on the BD MAX™ System at the target levels identified in Table 11. Samples were prepared by spiking of representative vaginosis target and/or vaginitis target in the presence of simulated vaginal matrix (SVM). Testing was conducted over more than 3 days using qualified BD Vaginal Panel reagents and included two BD COR™ PX/MX Systems and four BD MAX™ Systems.

A Two, One-Sided Test (TOST) of Equivalence was performed for the low positive (1.99x LoD) and moderate positive (5x LoD) target levels for each strain. The 90% confidence intervals for the difference in mean Ct.score between the BD COR™ System and BD MAX™ Systems were determined for each strain at each target level. Equivalence of the two systems is established when it is contained within the equivalence margin of [-6% of the reference mean, +6% of the reference mean]. TOST analysis was not performed on High Negative (C5) Candida spp. panel members based on the proportion positivity being less than 95% at C5, a predicate requirement. SDPA and 95% confidence intervals were

{14}------------------------------------------------

generated. Equivalence between the BD COR™ System and BD MAX™ Systems is demonstrated at the High Negative level by overlapping 95% confidence intervals.

Results from the study demonstrate that the analytical sensitivity of the BD Vaginal Panel on BD COR™ is equivalent to the analytical sensitivity (LoD) demonstrated on the BD MAX™ System as shown in Table 12 and Table 13.

Low Positive (1.99x C95)Moderate Positive (5x C95)High Negative (C5)
PanelTargetConcentrationPanelTargetConcentrationPanelTargetConcentration
1Cgla402 CFU/mL8Cgla1010 CFU/mL15Cgla10 CFU/mL
2GV1914 CFU/mL9GV4810 CFU/mL16Ckru6 CFU/mL
2Ckru2060 CFU/mL9Ckru5175 CFU/mL16Ckru6 CFU/mL
3BVAB923 Copies/mL10BVAB2320 Copies/mL17Calb53 CFU/mL
3Calb35396 CFU/mL10Calb88935 CFU/mL17Calb53 CFU/mL
4Mega4507 Copies/mL11Mega11325 Copies/mL18Cpara399 CFU/mL
4Cpara61013 CFU/mL11Cpara153300 CFU/mL18Cpara399 CFU/mL
5Ljens1015 CFU/mL12Ljens2550 CFU/mL19Cdub52 CFU/mL
5Cdub7964 CFU/mL12Cdub20010 CFU/mL19Cdub52 CFU/mL
6Lcrisp109 CFU/mL13Lcrisp275 CFU/mL20Ctrop16 CFU/mL
6Ctrop623 CFU/mL13Ctrop1565 CFU/mL20Ctrop16 CFU/mL
7Ato253 CFU/mL14Ato635 CFU/mL
7TV44 Cells/mL14TV110 Cells/mL

Table 9: Analytical Sensitivity Confirmation Panel Members

Table 10: Analytical Sensitivity Confirmation in Vaginal Swab for BD CORT™ System for the
Vaginosis Targets
TargetTargetLevelSystemNProportionPositivityMean Ct.ScoreDifference inMean Ct.Score[COR - MAX] and90% CIEquivalenceIntervalEquivalenceEstablished
AtopobiumvaginaeLowPositiveCOR48100%26.230.21(-1.56, 1.56)Yes
LowPositiveMAX48100%26.02(0.11, 0.31)
AtopobiumvaginaeModeratePositiveCOR48100%25.020.12(-1.49, 1.49)Yes
ModeratePositiveMAX48100%24.90(-0.02, 0.26)
BVAB-2*LowPositiveCOR48100%29.790.19(-1.78, 1.78)Yes
LowPositiveMAX48100%29.60(0.06, 0.33)
BVAB-2*ModeratePositiveCOR48100%28.520.21(-1.70, 1.70)Yes
ModeratePositiveMAX48100%28.31(0.10, 0.32)
Megasphaera*LowPositiveCOR4898%30.350.34(-1.80, 1.80)Yes
LowPositiveMAX48100%30.01(0.15, 0.52)
Megasphaera*ModeratePositiveCOR48100%29.180.27(-1.73, 1.73)Yes
ModeratePositiveMAX48100%28.91(0.13, 0.41)
GardnerellavaginallisLowPositiveCOR48100%28.690.23(-1.71, 1.71)Yes
LowPositiveMAX4898%28.45(0.12, 0.34)
ModerateCOR48100%27.470.20

{15}------------------------------------------------

TargetTargetLevelSystemNProportionPositivityMean Ct.ScoreDifference inMean Ct.Score[COR - MAX] and90% CIEquivalenceIntervalEquivalenceEstablished
PositiveMAX48100%27.27(0.10, 0.30)
LactobacillusjenseniiLowPositiveCOR48100%24.650.50Yes
MAX48100%24.15(0.39, 0.61)(-1.45, 1.45)
ModeratePositiveCOR48100%23.410.48Yes
MAX48100%22.93(0.39, 0.57)(-1.38, 1.38)
LowPositiveCOR48100%26.520.40
LactobacilluscrispatusPositiveMAX48100%26.12(0.24, 0.56)(-1.57, 1.57)Yes
ModeratePositiveCOR48100%25.180.27Yes
PositiveMAX48100%24.91(0.15, 0.39)(-1.49, 1.49)

*BVAB and Megasheara on non-cultivatable organisms. Plasmids were used for the LoD confirmation study.

Table 11: Analytical Sensitivity Confirmation in Vaginal Swab for BD COR™ System for the Vaginitis Targets

TargetTarget LevelSystemNProportionPositivity(95% CI forHighNegatives)MeanSDPADifference inMean SDPA[COR - MAX]with 90% CIEquivalenceIntervalEquivalenceEstablished
Trichomonas vaginalisLow PositiveCOR48100%32.310.17(-1.93, 1.93)Yes
Low PositiveMAX48100%32.14(0.06, 0.28)
Trichomonas vaginalisModerate PositiveCOR48100%31.34-0.10(-1.89, 1.89)Yes
Moderate PositiveMAX48100%31.44(-0.21, 0.01)
Candida albicans*High NegativeCOR4890%(78%, 95%)34.33
*High NegativeMAX4890%(78%, 95%)34.30
Candida albicansLow PositiveCOR48100%27.750.15(-1.66, 1.66)Yes
Low PositiveMAX48100%27.61(0.00, 0.29)
Candida albicansModerate PositiveCOR48100%26.890.07(-1.61, 1.61)Yes
Moderate PositiveMAX48100%26.82(-0.12, 0.26)
Candida parapsilosis*High NegativeCOR4871%(57%, 82%)35.39
*High NegativeMAX4867%(53%, 78%)35.17
Candida parapsilosisLow PositiveCOR48100%29.390.43(-1.74, 1.74)Yes
Low PositiveMAX48100%28.96(0.21, 0.65)
Candida parapsilosisModerate PositiveCOR48100%28.150.31(-1.67, 1.67)Yes
Moderate PositiveMAX48100%27.85(0.08, 0.53)
Candida tropicalis*High NegativeCOR4885%(73%, 93%)34.69
*High NegativeMAX4885%(73%, 93%)34.45
Candida tropicalisLow PositiveCOR48100%30.830.28(-1.83, 1.83)Yes
Low PositiveMAX48100%30.55(0.12, 0.44)
Candida tropicalisModerate PositiveCOR48100%29.920.22(-1.78, 1.78)

CONFIDENTIAL AND PROPRIETARY

{16}------------------------------------------------

TargetTarget LevelSystemNProportionPositivity(95% CI forHighNegatives)MeanSDPADifference inMean SDPA[COR - MAX]with 90% CIEquivalenceIntervalEquivalenceEstablished
Candida dubliniensisMAX48100%29.70(0.06, 0.37)
*High NegativeCOR4858%(44%, 71%)34.20
*High NegativeMAX4865%(50%, 77%)34.54
Candida dubliniensisLow PositiveCOR48100%28.920.39(-1.71, 1.71)Yes
Low PositiveMAX48100%28.53(0.19, 0.59)
Moderate PositiveCOR48100%27.660.22(-1.65, 1.65)Yes
Moderate PositiveMAX48100%27.43(0.02, 0.43)
Candida glabrata*High NegativeCOR4873%(59%, 83%)36.04
*High NegativeMAX4869%(55%, 80%)37.83
Low PositiveCOR48100%29.640.05Yes
Low PositiveMAX48100%29.59(-0.12, 0.22)(-1.78, 1.78)
Moderate PositiveCOR48100%28.590.26Yes
Moderate PositiveMAX48100%28.33(0.12, 0.40)(-1.70, 1.70)
Candida krusei*High NegativeCOR4844%(31%, 58%)36.48
*High NegativeMAX4854%(40%, 67%)36.55
Low PositiveCOR48100%29.350.32Yes
Low PositiveMAX48100%29.04(0.14, 0.50)(-1.74, 1.74)
Moderate PositiveCOR48100%28.170.28(-1.67, 1.67)
Moderate PositiveMAX48100%27.89(0.16, 0.40)Yes

*TOST analysis was not performed on High Negative Candida spp. (CS) panel members based on the proportion positivity being less than 95% at C5, a predicate requirement. Overlapping 95% confidence intervals between the both the BD MAX and COR platforms indicates equivalency at the High Negative level.

Cross-Contamination for BD CORTM System

A study was conducted to investigate cross-contamination while processing samples with high bacterial loads of BD Vaginal Panel targets. A panel made of alternating high positive and negative members was used at a prevalence of 50% high positive samples to simulate the most sensitive case for cross-contamination. In the high positive panel member, vaginosis markers were represented by a combination of L. jensenii (5.57E+07 CFU/mL), G. vaginalis (4.29E+07 CFU/mL), A. vaginae (1.65E+08 CFU/mL), and BVAB-2 (1.00E+09 copies/mL) while vaginitis targets were represented by T. vaginalis (8.0E+03 cells/mL). The negative member did not contain any target analytes. Of the 543 negative samples tested, 1 false positive result was observed, demonstrating a contamination rate of 0.18% (95% CI: 0.03- 1.04%), which met the predefined study acceptance criteria. Clinical Agreement Study between BD MAX™ and BD COR™ Systems

The performance of the BD Vaginal Panel on the BD COR™ System was evaluated in a clinical agreement study by comparing the assay results obtained on the BD COR™ System to the results obtained on the BD MAX™ System. BD MAX™ results served as the reference in the clinical agreement study.

Clinical vaginal specimens obtained from both internal collections were used for this comparison study. Clinical panels were created by pooling the previously collected clinical specimens and, where necessary, spiking in a high positive clinical specimen or pooling positive specimens at specific Secondary Derivative Peak Abscissa (SDPAs) to reach the necessary analyte

{17}------------------------------------------------

level(s) for Cgroup and TV. In addition, contrived samples were created by spiking organisms into simulated matrix. For C. glabrata and C. krusei, contrived samples were created by spiking organisms into negative vaginal matrix or in simulated vaginal matrix because of their very low prevalence. For BV, contrived panel members with different BV marker combinations were prepared using simulated vaginal matrix. These samples are denoted as BV Contrived. Additionally, the Cgroup, TV, and negative vaginitis panel members in natural vaginal matrix were analyzed for BV targets. These samples are denoted as BV Natural.

The clinical agreement study included 700 panel members. The panel members were prepared so the majority of positive specimens for each analyte were at the low positive and moderate positive level. Three aliquots were prepared for each panel member. Each aliquot was tested on the BD COR™ System and BD MAX ™ System with the order of system testing randomized. Testing occurred at two external sites and one internal site.

To demonstrate that the performance of the BD Vaginal Panel on the BD CORTM System is equivalent to the performance on BD MAX™, both positive/negative percent agreement analysis and Deming regression analysis of the Ct.Score or SDPA values were performed. Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) between the BD MAX™ and the BD COR™ Systems were calculated separately for each target, the PPA and NPA were calculated for each of the three sites where BD COR™ testing occurred, against composite comparator results where the positive or negative status of a panel member is defined by >2 out of 3 evaluable results obtained on the BD MAX™. PPA and NPA estimates were also averaged across the three BD COR™ testing sites. The PPA and NPA results as well as the corresponding 95% confidence interval at each BD COR™ testing site and the average across all three BD COR™ testing sites are summarized in Tables 14-19 for each target. The denominator for PPA and NPA calculations includes panel members with equivocal comparator results from BD MAX™, as indicated at the bottom of the tables. Equivocal BD MAX™ comparator result is defined as one positive, one negative, and one non-evaluable result from the BD MAXTM.

The systematic differences in numeric value between Ct.Score or SDPA results from BD COR™ and BD MAX™ were evaluated by the Weighted Deming regression analysis based on the average Ct.Score or SDPA of BD COR™ results and the average Ct.Score or SDPA of BD MAX™ results of a given panel member across all corresponding testing sites. The results from the Deming regression analysis are provided in Figure 1 - Figure 4-4 for C. albicans, C. glabrata, C. krusei, and TV. Due to the aggregate BV result reporting, the Deming results are not provided for the Vaginosis targets (Atopobium vaginae, Gardnerella vaginalis, Lactobacillus spp., BVAB-2 and Megasphera-1). The point estimate of intercept and slope as well as the corresponding 95% Confidence Interval of each Deming regression line are provided in Table 20. Additionally, the Weighted Deming regression bias estimate along with 95% confidence interval at different analyte levels are presented in Table 21.

{18}------------------------------------------------

Table 12: Percent Agreement of BD COR™ versus BD MAX™ Result by Test Site for BV
Contrived Specimens
Targeted Organism: BV ContrivedBD MAX™ Result
BD COR™ Test SiteBD COR™ ResultPositiveNegative
1Positive720
Negative0100
Total72100
PPA: 100% (72/72); 95% CI: (94.9, 100)NPA: 100% (100/100); 95% CI: (96.3, 100)OPA: 100% (172/172); 95% CI: (97.8, 100)Number of non-evaluable samples: 0
2Positive710
Negative1100
Total72100
PPA: 98.6% (71/72); 95% CI: (92.5, 99.8)NPA: 100% (100/100); 95% CI: (96.3, 100)OPA: 99.4% (171/172); 95% CI: (96.8, 99.9)Number of non-evaluable samples: 0
3Positive720
Negative0100
Total72100
PPA: 100% (72/72); 95% CI: (94.9, 100)NPA: 100% (100/100); 95% CI: (96.3 100)OPA: 100% (172/172); 95% CI: (97.8, 100)Number of non-evaluable samples: 0
Overall:
Average PPA: 99.5%; Bootstrap 95% CI: (98.4, 100)
Average NPA: 100%; Bootstrap 95% CI: N/A
Average OPA: 99.8%; Bootstrap 95% CI: (99.3, 100)
Average number of BD MAX™ equivocal results: 0
Note 1: Confidence intervals for point estimates at each site are calculated by a score method.
Note 2: Confidence intervals for point estimates averaged over 3 sites are calculated by a bootstrap method.

{19}------------------------------------------------

Table 13: Percent Agreement of BD COR™ versus BD MAX™ Result by Test Site for BV Natural Specimens

Targeted Organism: BV NaturalBD MAX™ Result
BD COR™ Test SiteBD COR™ ResultPositiveNegative
1Positive2266
Negative6114
Total232120
PPA: 97.4% (226/232); 95% CI: (94.5, 98.8)NPA: 95.0% (114/120); 95% CI: (89.5, 97.7)OPA: 96.6% (340/352); 95% CI: (94.1, 98.0)Number of non-evaluable samples: 2
2Positive2284
Negative5116
Total233120
PPA: 97.9% (228/233); 95% CI: (95.1, 99.1)NPA: 96.7% (116/120); 95% CI: (91.7, 98.7)OPA: 97.5% (344/353); 95% CI: (95.2, 98.7)Number of non-evaluable samples: 1
3Positive2285
Negative4113
Total232118
PPA: 98.3% (228/232); 95% CI: (95.7, 99.3)NPA: 95.8% (113/118); 95% CI: (90.5, 98.2)OPA: 97.4% (341/350); 95% CI: (95.2, 98.6)Number of non-evaluable samples: 4
Overall:
Average PPA: 97.8%; Bootstrap 95% CI: (96.3, 99.1)
Average NPA: 95.8%; Bootstrap 95% CI: (93.1, 98.2)
Average OPA: 97.2%; Bootstrap 95% CI: (95.8, 98.4)
Average number of BD MAX™ equivocal results: 0
Note 1: Confidence intervals for point estimates at each site are calculated by a score method.
Note 2: Confidence intervals for point estimates averaged over 3 sites are calculated by a bootstrap

| method.


{20}------------------------------------------------

Targeted Organism: C. glabrataBD MAXTM Result
BD CORTM Test SiteBD CORTM ResultPositiveNegative
1Positive580
Negative0125
Total58125
PPA: 100% (58/58); 95% CI: (93.8, 100)
NPA: 100% (125/125); 95% CI: (97.0, 100)
OPA: 100% (183/183); 95% CI: (97.9, 100)
Number of non-evaluable samples: 1
2Positive570
Negative0125
Total57125
PPA: 100% (57/57); 95% CI: (93.7, 100)
NPA: 100% (125/125); 95% CI: (97.0, 100)
OPA: 100% (182/182); 95% CI: (97.9, 100)
Number of non-evaluable samples: 2
3Positive570
Negative0122
Total57122
PPA: 100% (57/57); 95% CI: (93.7, 100)
NPA: 100% (122/122); 95% CI: (96.9, 100)
OPA: 100% (179/179); 95% CI: (97.9, 100)
Number of non-evaluable samples: 5
Overall:
Average PPA: 100%; Bootstrap 95% CI: N/A
Average NPA: 100%; Bootstrap 95% CI: N/A
Average OPA: 100%; Bootstrap 95% CI: N/A
Average number of BD MAXTM equivocal results: 0
Note 1: Confidence intervals for point estimates at each site are calculated by a score method.
Note 2: Confidence intervals for point estimates averaged over 3 sites are calculated by a bootstrap method

Table 14: Percent Agreement of BD COR™ versus BD MAX™ Result by Test Site for C. glabrata

{21}------------------------------------------------

Targeted Organism: C. kruseiBD MAX™ Result
BD CORT™ Test SiteBD CORT™ ResultPositiveNegative
1Positive500
Negative0125
Total50125
PPA: 100% (50/50); 95% CI: (92.9, 100)NPA: 100% (125/125); 95% CI: (97.0, 100)OPA: 100% (175/175); 95% CI: (97.9, 100)Number of non-evaluable samples: 2
2Positive490
Negative0125
Total49125
PPA: 100% (49/49); 95% CI: (92.7, 100)NPA: 100% (125/125); 95% CI: (97.0, 100)OPA: 100% (174/174); 95% CI: (97.8, 100)Number of non-evaluable samples: 3
3Positive510
Negative0122
Total51122
PPA: 100% (51/51); 95% CI: (93.0, 100)NPA: 100% (122/122); 95% CI: (96.9, 100)OPA: 100% (173/173); 95% CI: (97.8, 100)Number of non-evaluable samples: 4
Overall:
Average PPA: 100%; Bootstrap 95% CI: N/A
Average NPA: 100%; Bootstrap 95% CI: N/A
Average OPA: 100%; Bootstrap 95% CI: N/A
Average number of BD MAX™ equivocal results: 0
Note 1: Confidence intervals for point estimates at each site are calculated by a score method.
Note 2: Confidence intervals for point estimates averaged over 3 sites are calculated by a bootstrap method.

Table 15: Percent Agreement of BD COR™ versus BD MAX™ Result by Test Site for C. krusei

{22}------------------------------------------------

Targeted Organism: Candida GroupBD MAX™ Result
BD CORT™ Test SiteBD COR™ ResultPositiveNegative
1Positive1151
Negative2124
Total117125
PPA: 98.3% (115/117); 95% CI: (94.0, 99.5)
NPA: 99.2% (124/125); 95% CI: (95.6, 99.9)
OPA: 98.8% (239/242); 95% CI: (96.4, 99.6)
Number of non-evaluable samples: 2
2Positive1172
Negative0123
Total117125
PPA: 100% (117/117); 95% CI: (96.8, 100)
NPA: 98.4% (123/125); 95% CI: (94.4, 99.6)
OPA: 99.2% (240/242); 95% CI: (97.0, 99.8)
Number of non-evaluable samples: 2
3Positive1181
Negative0121
Total118122
PPA: 100% (118/118); 95% CI: (96.8, 100)
NPA: 99.2% (121/122); 95% CI: (95.5, 99.9)
OPA: 99.6% (239/240); 95% CI: (97.7, 99.9)
Number of non-evaluable samples: 4
Overall:
Average PPA: 99.4%; Bootstrap 95% CI: (98.5, 100)
Average NPA: 98.9%; Bootstrap 95% CI: (97.4, 100)
Average OPA: 99.2%; Bootstrap 95% CI: (98.3, 99.9)
Average number of BD MAX™ equivocal results: 0
Note 1: Confidence intervals for point estimates at each site are calculated by a score method.
Note 2: Confidence intervals for point estimates averaged over 3 sites are calculated by a bootstrap method

Table 16: Percent Agreement of BD COR™ versus BD MAX™ Result by Test Site for Cgroup

{23}------------------------------------------------

Targeted Organism: TVBD MAX Result
BD COR Test SiteBD COR ResultPositiveNegative
1Positive1100
Negative0125
Total110125
PPA: 100% (110/110); 95% CI: (96.6, 100)
NPA: 100% (125/125); 95% CI: (97.0, 100)
OPA: 100% (235/235); 95% CI: (98.4, 100)
Number of non-evaluable samples: 1
2Positive1090
Negative1125
Total110125
PPA: 99.1% (109/110); 95% CI: (95.0, 99.8)
NPA: 100% (125/125); 95% CI: (97.0, 100)
OPA: 99.6% (234/235); 95% CI: (97.6, 99.9)
Number of non-evaluable samples: 1
3Positive1100
Negative0122
Total110122
PPA: 100% (110/110); 95% CI: (96.6, 100)
NPA: 100% (122/122); 95% CI: (96.9, 100)
OPA: 100% (232/232); 95% CI: (98.4, 100)
Number of non-evaluable samples: 4
Overall:
Average PPA: 99.7%; Bootstrap 95% CI: (99.0, 100)
Average NPA: 100%; Bootstrap 95% CI: N/A
Average OPA: 99.9%; Bootstrap 95% CI: (99.5, 100)
Average number of BD MAX™ equivocal results: 0
Note 1: Confidence intervals for point estimates at each site are calculated by a score method.
Note 2: Confidence intervals for point estimates averaged over 3 sites are calculated by a bootstrap method.

Table 17: Percent Agreement of BD COR™ versus BD MAX™ Result by Test Site for TV

Table 18: Weighted Deming Regression Coefficients for Ct.Score for BD COR™ vs. BD MAX™

Master MixTargetSiteIntercept (95%CI)Slope (95%CI)
MM2(SDPA)C. glabrata1.00 (-4.13, 6.12)0.98 (0.80, 1.16)
CgroupOverall-5.44 (-8.29, -2.59)1.19 (1.09, 1.30)
C. krusei3.08 (-1.53, 7.68)0.90 (0.73, 1.06)
TV-2.77 (-4.63, -0.91)1.09 (1.02, 1.15)

{24}------------------------------------------------

MasterMixTargetActual LevelCt.score/SDPAof BD MAXTMBiasEstimate95% CI
MM2(SDPA)C. glabrataHigh Positive27.680.38(0.15, 0.62)
Moderate28.960.36(0.19, 0.53)
Positive29.870.34(0.06, 0.61)
Low Positive40.000.11(-1.92, 2.15)
Negative24.05-0.83(-1.21, -0.45)
CgroupHigh Positive27.16-0.23(-0.42, -0.04)
Moderate30.120.34(-0.03, 0.70)
Positive40.002.23(0.87, 3.59)
Low Positive27.000.32(0.09, 0.55)
C. kruseiNegative28.330.19(0.06, 0.31)
High Positive29.590.06(-0.20, 0.32)
Moderate40.00-1.00(-2.93, 0.93)
Positive26.36-0.48(-0.76, -0.21)
TVLow Positive30.24-0.14(-0.31, 0.03)
Negative32.830.08(-0.17, 0.33)
High Positive40.000.70(0.05, 1.35)
Moderate

Table 19: Weighted Deming Regression Bias Estimate for BD COR™ vs. BD MAX™

{25}------------------------------------------------

Image /page/25/Figure/1 description: This image is a figure titled "Figure 1: Deming Regression for the BD Vaginal Panel on the BD CORT™ System versus the BD MAX™ System – C. glabrata". The figure is comparing the BD CORT system to the BD MAX system. The figure is for C. glabrata.

Image /page/25/Figure/2 description: The image is a scatter plot that compares COR SDPA Score and MAX SDPA Score. The plot includes a Deming Regression Fit line with the equation 1 + 0.98 * MAX SDPA Score, and an identity line. The Pearson's r correlation coefficient is 0.708, indicating a strong positive correlation between the two variables. The Deming Regression Fit is based on a sample size of n=58.

The 0.95-confidence bounds are calculated with the jackknife method.

{26}------------------------------------------------

Image /page/26/Figure/1 description: This image is titled "Figure 2: Deming Regression for the BD Vaginal Panel on the BD CORT™ System versus the BD MAXT™ System - Cgroup". The image is a title for a figure. The title describes a Deming Regression analysis performed on the BD Vaginal Panel. The analysis compares the BD CORT™ System with the BD MAXT™ System within the Cgroup.

Image /page/26/Figure/2 description: The image is a scatter plot that compares COR SDPA Score and MAX SDPA Score. The plot includes a Deming Regression Fit line, represented by the equation -5.44 + 1.19 * MAX SDPA Score, along with an identity line. The Pearson's r correlation coefficient is 0.938, indicating a strong positive correlation between the two scores. The Deming Regression Fit is based on a sample size of 114.

The 0.95-confidence bounds are calculated with the jackknife method.

{27}------------------------------------------------

Image /page/27/Figure/1 description: The image is a title for a figure. The title states, "Figure 3: Deming Regression for the BD Vaginal Panel on the BD CORT™ System versus the BD MAX™ System – C. krusei".

Image /page/27/Figure/2 description: The image is a scatter plot that shows the relationship between COR SDPA Score and MAX SDPA Score. The plot includes a Deming Regression Fit line, represented by the equation 3.08 + 0.9 * MAX SDPA Score, along with a shaded confidence interval. There is also a dashed line representing the identity line. The Pearson's r correlation coefficient is 0.881, indicating a strong positive correlation between the two variables.

MAX SDPA Score The 0.95-confidence bounds are calculated with the jackknife method.

{28}------------------------------------------------

Figure 4: Deming Regression for the BD Vaginal Panel on the BD COR™ System versus the BD MAXTM System - TV

Image /page/28/Figure/2 description: The image is a scatter plot that compares COR SDPA Score and MAX SDPA Score. The x-axis represents MAX SDPA Score, ranging from 15 to 40, while the y-axis represents COR SDPA Score, also ranging from 15 to 40. A Deming Regression Fit line is plotted with the equation -2.77 + 1.09 * MAX SDPA Score, along with an identity line, and the Pearson's r correlation coefficient is 0.962.

The 0.95-confidence bounds are calculated with the jackknife method.

BD Vaginal Panel Non-Reportable Results for BD COR™ System

As with BD MAX™, BD COR™ System results that were due to an internal control failure, extraction transfer failure, or liquid level failure are considered non-reportable. Additionally, on BD COR™ System, incomplete (INC) results may occur on the MX instrument after the specimen has been aspirated for extraction (i.e., aborted runs due to Inventory Robot (IR) error). INC results that occur on MX were also included in the BD COR™ System non-reportable rate calculation. Error results on BD COR™ System were marked noncompliant if they were due to an operator error and were not included in the Non-reportable rate calculation. Non-reportable rates on BD COR™ System are shown in Table 22.

{29}------------------------------------------------

Table 20: Summary of BD COR™ Total Non-Reportable Rate for Combined Target by BD COR™ Test Site

CombinedTargetBD CORTMTest SiteUnresolvedb RateIndeterminatec RateIncompleted RateTotal UNR+IND+INCRate
Initial(95% CI)Finala(95% CI)Initial(95% CI)Finala(95% CI)Initial(95% CI)Finala(95% CI)Initial(95% CI)Finala(95% CI)
10.6%(4/681)(0.2, 1.5)0.0%(0/681)(0.0, 0.6)0.0%(0/681)(0.0, 0.6)0.0%(0/681)(0.0, 0.6)0.0%(0/681)(0.0, 0.6)0.0%(0/681)(0.0, 0.6)0.6%(4/681)(0.2, 1.5)0.0%(0/681)(0.0, 0.6)
20.6%(4/685)(0.2, 1.5)0.0%(0/683)(0.0, 0.6)0.4%(3/685)(0.1, 1.3)0.0%(0/683)(0.0, 0.6)0.0%(0/685)(0.0, 0.6)0.0%(0/683)(0.0, 0.6)1.0%(7/685)(0.5, 2.1)0.0%(0/683)(0.0, 0.6)
30.3%(2/681)(0.1, 1.1)0.0%(0/680)(0.0, 0.6)0.0%(0/681)(0.0, 0.6)0.0%(0/680)(0.0, 0.6)0.0%(0/681)(0.0, 0.6)0.0%(0/680)(0.0, 0.6)0.3%(2/681)(0.1, 1.1)0.0%(0/680)(0.0, 0.6)
Overall0.5%(10/2047)(0.3, 0.9)0.0%(0/2044)(0.0, 0.2)0.1%(3/2047)(0.0, 0.4)0.0%(0/2044)(0.0, 0.2)0.0%(0/2047)(0.0, 0.2)0.0%(0/2044)(0.0, 0.2)0.6%(13/2047)(0.4, 1.1)0.0%(0/2044)(0.0, 0.2)

a The final rate is calculated with the number of remaining non-reportable events after repeat testing.

b Unresolved, invalid SPC due to presence of inhibitory substances or reagent failure.

c Indeterminate, BD COR™ System failure (with Warning or Error Codes).

d Incomplete Run, aborted run or BD COR™ System failure that halts robot operations (with Warning or Error Codes).

§ 866.3975 Device that detects nucleic acid sequences from microorganisms associated with vaginitis and bacterial vaginosis.

(a)
Identification. A device that detects nucleic acid sequences from microorganisms associated with vaginitis and bacterial vaginosis is a qualitative in vitro diagnostic device intended for the detection of microbial nucleic acid sequences in vaginal specimens collected from patients with signs and symptoms of vaginitis or bacterial vaginosis. This device is intended to aid in the diagnosis of vaginitis or bacterial vaginosis when used in conjunction with clinical signs and symptoms and other laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) Documentation with a detailed device description of device components; ancillary reagents required but not provided; and explanation of the methodology including primer/probe sequence, design, and rationale for sequence selection.
(ii) Documentation with information that demonstrates the performance characteristics of the device, including:
(A) Limit of Detection;
(B) Precision (reproductivity);
(C) Analytical specificity;
(D) Analytical reactivity (inclusivity);
(E) Specimen stability; and
(F) Effects of interfering substances.
(iii) Detailed documentation from a prospective clinical study. As appropriate to the intended use, the prospective clinical study must be performed on an appropriate study population, including women of various ages and ethnicities. The prospective clinical study must compare the device performance to results obtained from well-accepted comparator methods.
(iv) Detailed documentation for device software, including software applications and hardware-based devices that incorporate software.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed explanation of the interpretation of results and acceptance criteria;
(ii) For devices with an intended use that includes detection of nucleic acid sequences from bacteria associated with bacterial vaginosis, clinical performance stratified by patient demographics such as race, ethnicity, age, and pregnancy status.
(iii) For devices with an intended use that includes detection of nucleic acid sequences from bacteria associated with bacterial vaginosis, a summary of device results in an asymptomatic population with demographic characteristics appropriate to the intended use population.
(iv) For devices with an intended use that includes detection of either Candida species or bacteria associated with bacterial vaginosis, a limitation that
Candida species and bacterial compositions associated with bacterial vaginosis can be present as part of normal vaginal flora and results should be considered in conjunction with available clinical information.