K Number
K223625

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-03-28

(113 days)

Product Code
Regulation Number
878.4800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Used for more accurate placement of pneumothorax decompression devices than the current standard of care auditory assessments. The Pneumothorax Decompression Indicator is a colorimetric capnography device indicated to detect the carbon dioxide in the gas passing through the device used to treat pneumothorax with suspected tension physiology.

Device Description

The Pneumeric Capnospot Pneumothorax Decompression Indicator is a capnography device that provides objective, real-time feedback during decompressive thoracostomy. Specifically, the device detects the presence of carbon dioxide in the gas from the decompressed pneumothorax, permitting more accurate placement of the needle or thoracostomy device compared to the current standard of care based upon auditory assessments.

The Pneumeric Capnospot is a therapeutic device intended to guide the placement of pneumothorax decompression needles and thoracostomy devices.

The product consists of a cylindrical polycarbonate tube containing a carbon dioxide detecting paper. The device has a male luer connector on the distal end of the device to connect to decompression needles and thoracostomy devices, a diaphragm to prevent backflow of gases, a color changing indicator and a female luer on the proximal end of the device that can be used to connect accessories.

The product is shipped non-sterile and labeled for single use only.

AI/ML Overview

The provided text describes a medical device called the "CapnoSpot™ Pneumothorax Decompression Indicator" and its 510(k) submission to the FDA. However, the document does not contain information about a study proving the device meets specific acceptance criteria based on performance metrics like sensitivity, specificity, or accuracy, particularly concerning an AI-driven component or human reader assistance.

The document discusses "Performance Data" in section 6.5, but this section focuses on design verification and shelf-life tests (visual inspection, dimensionals, backflow, cracking pressure, torque, luer leakage, animal data) to ensure the device meets specifications, rather than clinical performance metrics in humans or AI-driven acceptance criteria.

The 510(k) submission for this device is based on substantial equivalence to a predicate device (Turkel Pneumothorax Kit K923028), rather than demonstrating clinical efficacy or AI performance through a comparative study.

Therefore, based on the provided text, I cannot fulfill the request to describe acceptance criteria and a study proving the device meets them in the context of sensitivity, specificity, accuracy, or human reader improvement with AI. The document does not provide details on:

  1. A table of acceptance criteria and reported device performance related to diagnostic accuracy.
  2. Sample size used for a test set, its provenance, or how ground truth was established for such a test set.
  3. Number of experts and their qualifications for establishing ground truth.
  4. Adjudication method for a test set.
  5. Whether an MRMC comparative effectiveness study was done or its effect size.
  6. Whether a standalone AI performance study was done.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for diagnostic performance.
  8. Training set sample size or how its ground truth was established.

The "Performance Data" refers to engineering and durability tests, and an "Animal Data" point is mentioned, but no details of a study with acceptance criteria for a diagnostic/AI performance are present.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 28, 2023

Pneumeric, Inc. Johnathon Aho Chief Executive Officer 823 4th Street SW, c/o Johnathon Aho Rochester, Minnesota 55902

Re: K223625

Trade/Device Name: CapnoSpot™ Pneumothorax Decompression Indicator Regulation Number: 21 CFR 878.4800 Regulation Name: Manual Surgical Instrument For General Use Regulatory Class: Class I Product Code: GAA Dated: March 1, 2023 Received: March 2, 2023

Dear Johnathon Aho:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

MarkTrumbore -SDigitally signed byMark Trumbore -SDate: 2023.03.2813:01:59-04'00'
-----------------------------------------------------------------------------------------------------

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223625

Device Name

Pneumeric™ CapnoSpot™ Pneumothorax Decompression Indicator

Indications for Use (Describe)

Used for more accurate placement of pneumothorax decompression devices than the current standard of care auditory assessments. The Pneumothorax Decompression Indicator is a colorimetric capnography device indicated to detect the carbon dioxide in the gas passing through the device used to treat pneumothorax with suspected tension physiology.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the word "PNEUMERIC" in bold, black letters, with a trademark symbol next to it. To the left of the word is a gray image of lungs. The lungs are stylized and have a smooth, rounded shape. The image is set against a white background.

6. 510(K) SUMMARY

6.1 ADMINISTRATIVE INFORMATION

Date of Summary Preparation:

6.1.1 CONTACT INFORMATION

Submitter/ManufacturerJohnathon M.E. Aho M.D., Ph.D.Chief Executive Officer, Pneumeric, Inc.823 4th St SWRochester, MN 55902United States507-206-1942John.aho@pneumeric-medical.comjaho@mtu.edu
Primary Submission ContactJohnathon M.E. Aho M.D., Ph.D.Chief Executive Officer, Pneumeric, Inc.823 4th St SWRochester, MN 55902United States507-206-1942John.aho@pneumeric-medical.comjaho@mtu.edu
Secondary Submission ContactJonathan Sackner-Bernstein M.D.Chief Regulatory Officer, Pneumeric, Incjsb@jsbmd.com+1.212.888.2589Ming Cheng ChewRegulatory Consultant,Libra Medical Inc.8401 73rd Ave North, Suite 63

Minneapolis, MN 55428
Phone: 763-232-3701 Email: mcchew@libramed.com

Veronica McDougall

{4}------------------------------------------------

Regulatory Consultant, Nagl Medtech 3451 Commerce Pkwy Miramar F1 33025 Phone: 954-507-3830 Email: vmcdougall(@nmddo.com

6.1.2 DEVICE INFORMATION

Trade NameCapnoSpot Pneumothorax DecompressionIndicator
Common NameDecompression Indicator
Classification NameNeedle, Aspiration and Injection, Disposable
Classification Regulation878.4800
ClassI
PanelGeneral & Plastic Surgery
Product CodeGAA
FDA Documents Related toNone
Modified DeviceC210112

PREDICATE DEVICE 6.2

The Pneumeric CapnoSpot Pneumothorax Decompression Indicator ("CapnoSpot" or "device") is substantially equivalent to the Turkel Pneumothorax Kit (K923028).

DEVICE DESCRIPTION 6.3

The Pneumeric Capnospot Pneumothorax Decompression Indicator is a capnography device that provides objective, real-time feedback during decompressive thoracostomy. Specifically, the device detects the presence of carbon dioxide in the gas from the decompressed pneumothorax, permitting more accurate placement of the needle or thoracostomy device compared to the current standard of care based upon auditory assessments.

The Pneumeric Capnospot is a therapeutic device intended to guide the placement of pneumothorax decompression needles and thoracostomy devices.

The product consists of a cylindrical polycarbonate tube containing a carbon dioxide detecting paper. The device has a male luer connector on the distal end of the device to connect to decompression needles and thoracostomy devices, a diaphragm to prevent backflow of gases, a color changing indicator and a female luer on the proximal end of the device that can be used to connect accessories.

The product is shipped non-sterile and labeled for single use only.

6.4 INDICATIONS FOR USE/INTENDED USE

Used for more accurate placement of pneumothorax decompression devices than the current standard of care auditory assessments. The Pneumothorax Decompression Indicator is a

{5}------------------------------------------------

colorimetric capnography device indicated to detect the carbon dioxide in the gas passing through the device used to treat pneumothorax with suspected tension physiology.

PERFORMANCE DATA 6.5

Testing was performed before and after accelerated aging. The relevant design verification and shelf life tests included:

  • Visual inspection ●
  • Dimensionals ●
  • Backflow
  • Cracking Pressure ●
  • Torque ●
  • Luer leakage
  • Animal Data ●

The design verification and shelf life testing showed that the device meets specifications before and after aging. This indicates that the device is as safe and effective as the predicate device.

SUBSTANTIAL EQUIVALENCE 6.6

Table 6-1 Substantial Equivalence

Table 6-1 Substantial Equivalence
CharacteristicPredicate Device (K923028)CapnoSpot (Subject Device)
Intended UseEvacuate fluid or gas fromeither a body cavity or thepleural space while providingvisualization of needleplacement for safetyProvide a visual indication forthe accurate placement of apneumothorax needle
Indications forUseThe Turkel Safety Fluid Systemhas application in:-Thoracentesis Procedures –Diagnostic and Therapeutic-Paracentesis Procedures –Diagnostic and TherapeuticThe PneumothoraxDecompression Indicator is acolorimetric capnographydevice indicated to detect theCO2 in the gas passing throughthe device used to treatPneumothorax with suspectedtension physiology.
Regulation Class&Product CodeClass I, GCB878.4200Class I, GAA878.4800
IncludedComponents-Blunt cannula housed within asharp hollow needle-In-line stopcock-Color Change indicator-Ball Valve-One-way Valve- Color Change indicator- One-way valve
Presence of One-Way Valve toprevent backflowYesYes
Indicator Tubeand Luer MaterialUnknownPolycarbonate
One Way ValveMaterialUnknownSilicone

{6}------------------------------------------------

CharacteristicPredicate Device (K923028)CapnoSpot (Subject Device)
Indicator MethodMechanical SpringColorimetric Paper
SterilizationSAL of 10-6 via EtONon-sterile
BiocompatibilityISO 10993-1N/A
Single-use only?YesSame

The differences between this device and its predicate do not raise new questions of safety or efficacy.

6.7 CONCLUSION

The Pneumeric CapnoSpot Pneumothorax Decompression Indicator is substantially equivalent to the predicate device (K923028) in design, function, and intended use.

§ 878.4800 Manual surgical instrument for general use.

(a)
Identification. A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.