K Number
K223600

Validate with FDA (Live)

Date Cleared
2023-08-25

(266 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider via the following:
A) Pre-vacuum Steam: 134℃ for 4 minutes, 30 minutes dry time.
B) Ethylene Oxide Sterilization: 100% Ethylene Oxide (EO) with a concentration of 360-363 mg/L @ 54°C and 40% relative humidity for 480 minutes, 24 hours aeration time at 42 ℃
Model: ZSW1610, ZSW 1611. Material: 45gms SMS, 54gms SMS. Color of wrap: Pink/Blue (double layers). Size (inches): 1212, 1515, 1818, 2020, 2424, 3030, 3636, 4040, 4545, 4848, 5454, 7254.
The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) up to 90 days post sterilization.
Types of medical devices to be sterilized in the Pre-Vacuum Steam Sterilization and Ethylene Oxide Sterilization: General purpose non-lumened reusable metal and nonmetal devices with stainless steel diffusion restricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments etc. Models validated for both sterilization methods with stainless steel as part of the load, which with a maximum weight of each types, range from 21bs to 9lbs, which dependent on each model's size.
For size212, 1515, 1818, maximum weight is 216, for 2020, 3636, 4040, 4545, 4848, maximum weight is 6lb, and for 5454, 7254, maximum weight is 9lb.

Device Description

Sterilization Wrap is square or rectangular three-layer (SMS) non-woven sheet which manufactured with Polypropylene spunbond-meltblown-spunbond (SMS) fabric. Sterilization wrap provides a strong barrier which protects against cuts, tears with particularly device sets. The device is designed to be implemented as an outer sterilization wrap which allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) up to 90 days post Pre-Vacuum Steam Sterilization or Ethylene Oxide Sterilization.
Sterilization Wrap is two-layer sheet which ultrasonically bonded blue and pink nonwoven sheet together along two edges, thus for convenient simultaneous wrapping.
Sterilization Wrap separated into two distinct types depending on different grams of weight of material, for ZSW 1610, it is two 45gsm SMS non-woven sheet ultrasonically bonded together, and for model ZSW 1611, it is two 54gsm SMS sheet. Each model contains series specification, including 12 Sizes (inches): 1212, 1515, 1818, 2020, 2424, 3030, 3636, 4040, 4545, 4848, 5454, 7254. With non-lumened reusable metal and nonmetal devices as part of the load, which with a maximum weight of each type, range from 21bs to 9lbs, which dependent on each model's size.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the device meets the acceptance criteria for a Sterilization Wrap (K223600).

Here's the breakdown of the information requested:


1. A table of acceptance criteria and the reported device performance

The document presents the acceptance criteria and device performance as "Test Items" and "Test Result" (always "Pass"). The specific numerical/quantitative acceptance criteria are listed under "Standard and Acceptance Criteria".

Test ItemsStandard and Acceptance CriteriaReported Device Performance (Test Result)
Size and Gram weightASTM D3776M-20: Standard Test Methods for Mass Per Unit Area (Weight) of Fabric Size: Marking dimension tolerance ±10%; Weight: Marking gram tolerance ±2g/m²Pass
Bacterial Filtration Efficiency (BFE)ASTM F 2101-19 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus BFE ≥ 90%Pass
Hydrostatic PressureAATCC 127-03 Test Method for Water Resistance: Hydrostatic Pressure Hydrostatic Pressure ≥ 50 cm H2OPass
Tensile StrengthASTM D5034-09 Standard Test Method for Breaking Strength and Elongation of Textile Fabrics (Grab Test) MD≥ 90N; CD≥ 63NPass
Air PermeabilityASTM D737-18 Standard Test Method for Air Permeability of Textile Fabrics Air Permeability ≥ 30 cfmPass
Bursting StrengthASTM D3786-18 Standard Test Method for Bursting Strength of Textile Fabrics—Diaphragm Bursting Strength Tester Method Bursting Strength ≥ 130 kpa (18.86psi)Pass
Tearing StrengthASTM D5587-15 Standard Test Method for Tearing Strength of Fabrics by Trapezoid Procedure MD≥ 50N; CD≥ 30NPass
Lint GenerationISO9073-10: 2019 Textiles — Test methods for nonwovens — Part 10: Lint and other particles generation in the dry state Coefficient of linting ≤4.0Pass
Skin Irritation AND Skin SensitizationISO 10993-1, ISO 10993-5, ISO 10993-10:2010. "Under the conditions of the study, the test article (pouches and chemical indicator) extract did not show no significant evidence of causing skin sensitization in the guinea pig before and after sterilized."Pass
Shelf-Life ValidationShelf Life Validation Test: Real time aging method. Shelf Life: 2 Years; Shelf Life after Sterilized: 90 days. The device performance shall be meet the requirements of the device. SAL=10-6.Pass
Sterilization Process Validation for EOISO11135: 2014 Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process. Ethylene Oxide Sterilization: 100% Ethylene Oxide (EO) with a concentration of 360-363 mg/L @ 54°C and 40% relative humidity for 480 minutes, 24 hours aeration time at 42 °C. The device performance shall be meet the requirements of the device with the worst loading, SAL=10-6.Pass
Sterilization Process Validation for SteamAAMI / ANSI ST79: 2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Comprehensive Guide To Steam Sterilization And Sterility Assurance In Health Care Facilities and ISO 17665-1:2006. Pre-vacuum Steam: 134°C for 4 minutes, 30 minutes dry time. The device performance shall be meet the requirements of the device with the worst loading, SAL=10-6.Pass
Material compatibilityAfter Pre-vacuum Steam or 100% Ethylene Oxide (EO) sterilization, the materials were not degraded. Performance test including: Size and Gram weight, Bacterial Filtration Efficiency (BFE), Hydrostatic Pressure, Tensile Strength, Air Permeability, Bursting Strength, Tearing Strength, Lint Generation should meet requirements.Pass
EO/ECH residueISO 10993-7: 2008 Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals EO≤4mg/d, ECH ≤9mg/dPass

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes used for each of the non-clinical tests. It only states that the tests were conducted and "all results can meet standards requirements" or "Passed".

The data provenance is not explicitly stated in terms of country of origin or whether it's retrospective or prospective for each specific test. However, the submitter is "Wuhan Zonsen Medical Products Co., Ltd." located in Wuhan, Hubei, China, which implies the testing likely occurred in China.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as the device is a medical accessory (sterilization wrap) and the validation involves non-clinical laboratory and performance testing, not expert interpretation of medical images or data requiring ground truth establishment by medical experts. The "ground truth" for this device's performance is established by meeting the specified international and national standards for physical properties, sterilization efficacy, and biocompatibility.


4. Adjudication method for the test set

This information is not applicable for the type of testing performed for this device. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving multiple readers for diagnostic accuracy, which is not relevant here.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is a sterilization wrap, not an AI-powered diagnostic tool. Therefore, MRMC studies and human reader improvement with AI assistance are not relevant to its evaluation.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. As stated above, this is a physical medical accessory, not an algorithm or software device.


7. The type of ground truth used

The "ground truth" for this device's performance is based on established international and national standards for medical device physical properties, biocompatibility, and sterilization process validation (e.g., ASTM, AATCC, ISO standards). The "truth" is whether the device's measured properties and performance meet the predefined thresholds set by these standards.


8. The sample size for the training set

This information is not applicable. This is a physical medical device that undergoes manufacturing and quality control testing, not a machine learning model that requires a training set.


9. How the ground truth for the training set was established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Wuhan Zonsen Medical Products Co., Ltd. Linna Ye Registration Manager No 8 Jinchao Rd, Zhucheng Street Wuhan, Hubei China

Re: K223600

Trade/Device Name: Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: July 7, 2023 Received: July 21, 2023

Dear Linna Ye:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

August 25, 2023

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed Eileen by Eileen Cadel -Cadel -S Date: 2023.08.25 13:20:06 -04'00' for

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223600

Device Name Sterilization Wrap

Indications for Use (Describe)

Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a healtheare provider via the following:

A) Pre-vacuum Steam: 134℃ for 4 minutes, 30 minutes dry time. B) Ethylene Oxide Sterilization: 100% Ethylene Oxide (EO) with a concentration of 360-363 mg/L @ 54°C and 40% relative humidity for 480 minutes, 24 hours aeration time at 42 ℃

Model: ZSW1610, ZSW 1611. Material: 45gms SMS, 54gms SMS. Color of wrap: Pink/Blue (double layers). Size (inches): 1212, 1515, 1818, 2020, 2424, 3030, 3636, 4040, 4545, 4848, 5454, 7254.

The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) up to 90 days post sterilization.

Types of medical devices to be sterilized in the Pre-Vacuum Steam Sterilization and Ethylene Oxide Sterilization: General purpose non-lumened reusable metal and nonmetal devices with stainless steel diffusionr estricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments etc. Models validated for both sterilization methods with stainless steel as part of the load, which with a maximum weight of each types, range from 21bs to 9lbs, which dependent on each model's size.

For size212, 1515, 1818, maximum weight is 216, for 2020, 3636, 4040, 4545, 4848, maximum weight is 6lb, and for 5454, 7254, maximum weight is 9lb.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

510 (K) Summary

This 510(K) Summary of safety and effectiveness information is being submitted in accordance with requirement of 21 CFR807.92

  1. Date of Submission: August, 23, 2023.

2. Submitter / 510(K) Holder

Wuhan Zonsen Medical Products Co., Ltd.

No. 8 Jinchao Road, Zhucheng Avenue, Xinzhou District, Wuhan City, Hubei Province, P.R. China 431400

Contact Person: Ms. Lisa Zhang Position: Registration Manager Email: registration01@zonsenmed.com

3. Proposed Device Name

Trade name: Sterilization Wrap Common name: Sterilization Wrap Classification Name: Wrap, Sterilization/Indicator, Physical/Chemical Sterilization Process Device Class: Class II Classification Panel: General Hospital Product Code: FRG Regulation Number: 21 CFR 880.6850

4. Predicate Devices

Primary Predicate Device: K160755

Product Name: Reliance® Solo Sterilization Wrap, Reliance® Tandem Sterilization Wrap

Submitter: Ahlstrom Nonwovens LLC

{4}------------------------------------------------

5. Device Description

Sterilization Wrap is square or rectangular three-layer (SMS) non-woven sheet which manufactured with Polypropylene spunbond-meltblown-spunbond (SMS) fabric. Sterilization wrap provides a strong barrier which protects against cuts, tears with particularly device sets. The device is designed to be implemented as an outer sterilization wrap which allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) up to 90 days post Pre-Vacuum Steam Sterilization or Ethylene Oxide Sterilization.

Sterilization Wrap is two-layer sheet which ultrasonically bonded blue and pink nonwoven sheet together along two edges, thus for convenient simultaneous wrapping.

Sterilization Wrap separated into two distinct types depending on different grams of weight of material, for ZSW 1610, it is two 45gsm SMS non-woven sheet ultrasonically bonded together, and for model ZSW 1611, it is two 54gsm SMS sheet. Each model contains series specification, including 12 Sizes (inches): 1212, 1515, 1818, 2020, 2424, 3030, 3636, 4040, 4545, 4848, 5454, 7254. With non-lumened reusable metal and nonmetal devices as part of the load, which with a maximum weight of each type, range from 21bs to 9lbs, which dependent on each model' size (table 1)

ModelMaterialLayersColorSizeMaximumWeight(lbs.)
Long(inches)Width(inches)
ZSW161045 gsmSMS2Pink and Bluedouble layers12122
15152
18182
20206
24246
30306
36366

Table 1 Device Specification

{5}------------------------------------------------

K223600510 (k) Summary
40406
45456
48486
54549
72549
12122
15152
18182
20206
24246
ZSW161154SMSgsm2Pink and Bluedouble layers30306
36366
40406
45456
48486
54549
72549

6. Indications for Use

Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider via the following:

A) Pre-vacuum Steam: 134°C for 4 minutes, 30 minutes dry time.

B) Ethylene Oxide Sterilization: 100% Ethylene Oxide (EO) with a concentration of 360~363 mg/L @ 54°C and 40% relative humidity for 480 minutes, 24 hours aeration time at 42 °C.

Model: ZSW1610, ZSW 1611.

Material (single layer): 45gms SMS, 54gms SMS.

Color of wrap: Pink/Blue.

Size (inches): 1212, 1515, 1818, 2020, 2424, 3030, 3636, 4040, 4545, 4848, 5454, 7254.

{6}------------------------------------------------

The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) up to 90 days post sterilization.

Types of medical devices to be sterilized in the Pre-Vacuum Steam Sterilization and Ethylene Oxide Sterilization:

General purpose non-lumened reusable metal and nonmetal devices including devices with stainless steel diffusionr estricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments etc.

Models validated for both sterilization methods with stainless steel as part of the load, which with a maximum weight of each type, range from 21bs to 91bs, which dependent on each model's size.

For size: 212, 1515, 1818, maximum weight is 21b, for 2020, 2424, 3030, 3636, 4040. 4545. 4848. maximum weight is 6lb. and for 5454. 7254. maximum weight is 9lb.

7. Technological Characteristics Comparison Table

The Sterilization Wrap is compared with the predicate device Reliance® Solo Sterilization Wrap, Reliance® Tandem Sterilization Wrap (K160755). The results are shown below in the Technological Characteristics Comparison Table:

DevicecharacteristicProposed DevicePredicate Device510(K) No. K160755Comparison Result
ManufacturerWuhan Zonsen MedicalProducts Co., Ltd.Ahlstrom Nonwovens,LLC--
Product nameSterilization wrapReliance® SoloSterilizationWrap, Reliance®TandemSterilization Wrap--
510(k) ReferenceK223600K160755--

Table 2 General Comparison between proposed and predicate device

{7}------------------------------------------------

510 (k) Summary

ClassClass IIClass IISameK223600510 (k) Summary
Product CodeFRGFRGSamedays post sterilization.device(s) until used.
RegulationNumber21 CFR PART 880.685021 CFR PART 880.6850SameMaterialCompositionPolypropylenespunbondmeltblown-spunbond (SMS)fabricPolypropylenespunbondmeltblown-spunbond (SMS)fabricSame
Intended UseSterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider.Reliance® Tandem and Solo Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care Provider.SameDesign featureZonsen Sterilization Wrap issquare or rectangular three-layer (SMS) non-wovensheet which manufacturedwith Polypropylenespunbond-meltblown-spunbond (SMS) fabric.The device has two-layersheet which ultrasonicallybonded blue and pink non-woven sheet together alongtwo edges, thus forconvenient simultaneouswrapping.The Reliance® Solo andReliance® TandemSterilization Wraps aresquare or rectangular.nonwoven sheetsproduced using a three-layer SMS (spunbond-meltblown-spunbond)process. The Reliance®SMS Sterilization Wrapsare separated into twodistinct product offerings:Reliance® Solo: Consistsof two sheets of SMSwrap, ultrasonicallybonded together alongtwo edges for convenientsimultaneous wrapping ofone or a collection ofmedical devices that willbe sterilized followingstandard healthcarepractices.Reliance® Tandem:Consists of single sheetsSame
Indication for useSterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider via the following:Pre-vacuum Steam: 134°C for 4 minutes, 30 minutes dry time.B) 100% Ethylene Oxide (EO) with a concentration of 360~363 mg/L @ 54°C and 40% relative humidity for 480 minutes, 24 hours aeration time at 42 °C.Ahlstrom Reliance® Tandem and Solo Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider via the following:A) Pre-vacuum Steam 270°F/132°C for 4 minutes, validated for dry time of 20 minutes for Models T100/S100, T200/S200, T300/S300, Validated for dry time of 30 minutes for Models T400/S400, T500/S500, T600/S600Similar
B) 100% Ethylene Oxide (EO) with a concentration of 725~735 mg/L @ 131°F/55°C and 40%~80% relative humidity for 60 minutes.C) Gravity Steam 250°F/121°C for 30 minutesD)Advanced Sterilization Products (ASP) STERRAD® 100SE) Advanced Sterilization Products (ASP) STERRAD® 100NX (Standard, Express and Flex cycles)F) STERIS Amsco® V-PRO 1, STERIS Amsco® V-PRO 1 Plus, STERIS Amsco® V-PRO maX Low Temperature Sterilization Systems (Lumen, Non-Lumen and Flexible Cycles)
The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) up to 90The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosedSimilar

{8}------------------------------------------------

{9}------------------------------------------------

510 (k) Summary

{10}------------------------------------------------

of SMS wrap, where twosheets are used togetherfor the sequentialwrapping of one or acollection of medical.devices that will besterilized followingstandard healthcarepractices.
Size and colorPink and blueBlue and greenSimilar
Sterilization Wrap separatedinto two distinct typesdepending on differentgrams of weight of material,DimensionalSpecifications ofReliance® Tandem andReliance® Solo Wrap:six type Models dependon various grams ofweight,
Size (inches): 1212, 1515,Size (inches): 99, 1212,
1818, 2020, 24*24,1515, 1818, 20*20,
3030, 3636, 40*40,2424, 3030, 36*36,
4545, 4848, 54*54,4040, 4545, 48*48,
72*54.5454, 6060, 72*54.
Single Use vs.ReusableSingle useSingle useSame
Disposable vs.Non-DisposableDisposableDisposableSame
MaximumRecommendedWrapped PackageTypes of medical devices tobe sterilized in the Pre-Vacuum Steam SterilizationAll models of Reliance®Tandem and Solovalidated for pre-vacuumSimilar

{11}------------------------------------------------

Content WeightsEthylene Oxide Sterilization:steam Sterilization and ethylene oxideS10 (k) Summary
General purpose non-lumened reusable metal and nonmetal devices including devices with stainless steel diffusion restricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments etc.Models validated for both sterilization methods with stainless steel as part of the load, which with a maximum weight of each type, range from 2lbs to 9lbs, which dependent on each model's size.sterilization with stainless steel lumens (3 mm in diameter or larger and 400 mm in length or less) as part of the load with a maximum weight of 25lbs dependent on the model.
Size and Gram weightASTM D3776/ D3776M-20Size: Marking dimension tolerance ±10%; Weight : Marking gram tolerance ±2g/m2 the inspection result is PASSASTM D3776/ D3776M-20Size: Marking dimension tolerance ±10%; Weight : Marking gram tolerance ±2g/m2 the inspection result is PASSSame
Bacterial Filtration Efficiency (BFE)ASTM F2101-19 Bacterial Filtration Efficiency (BFE) Using a Biological AerosolASTM F2101-19 Bacterial Filtration Efficiency (BFE) Using a Biological AerosolSame
K223600510 (k) Summary
of Staphylococcus aureus ≥90%the inspection result isPASSBiological Aerosol ofStaphylococcus aureus ≥90%the inspection result isPASS
HydrostaticPressureAATCC 127-03, WaterResistance: HydrostaticPressure ≥ 50 cm H2Othe inspection result isPASSAATCC 127-03, WaterResistance: HydrostaticPressure ≥ 50 cm H2Othe inspection result isPASSSame
Tensile StrengthASTM D5034-09 Standard,Breaking Strength andElongation of TextileFabrics (Grab Test)MD≥ 90N; CD≥ 63Nthe inspection result isPASSASTM D5034-09Standard, BreakingStrength and Elongationof Textile Fabrics (GrabTest)MD≥ 90N; CD≥ 63Nthe inspection result isPASSSame
Air PermeabilityASTM D737-18Standard,Air Permeability of TextileFabrics ≥ 30 cfmthe inspection result isPASSASTM D737-18Standard,Air Permeability ofTextile Fabrics ≥ 30 cfmthe inspection result isPASSSame
Bursting StrengthASTM D3786-18Standard Test Method forBursting Strength of TextileFabrics—DiaphragmBursting Strength TesterMethod ≥ 130 kpa(18.86psi)the inspection result isASTM D3786-18Standard Test Method forBursting Strength ofTextile Fabrics—Diaphragm BurstingStrength Tester Method ≥130 kpa (18.86psi)the inspection result isSame
PASSPASS
Tearing StrengthASTM D5587-15, TearingStrength of Fabrics, MD≥50N, CD≥ 30N.the inspection result isPASSASTM D5587-15,Tearing Strength ofFabrics, MD≥ 50N, CD≥30N.the inspection result isPASSSame
Lint GenerationISO9073-10, Textiles Lint,Cofficient of linting ≤4.0the inspection result isPASSISO9073-10,TextilesLint, Cofficient of linting≤4.0the inspection result isPASSSame
MaterialCompatibilityAfter Pre-vacuum Steam or100% Ethylene Oxide (EO)sterilization, the materialswere not degraded.Performance test includingSize and Gram weightBacterial FiltrationEfficiency (BFE)Hydrostatic PressureTensile StrengthAir PermeabilityBursting StrengthTearing StrengthLint Generationcan meet requirementsAfter Pre-vacuum Steamor 100% Ethylene Oxide(EO) sterilization, thematerials were notdegraded.Performance testincludingSize and Gram weightBacterial FiltrationEfficiency (BFE)Hydrostatic PressureTensile StrengthAir PermeabilityBursting StrengthTearing StrengthLint Generationcan meet requirementsSame
BiocompatibilityISO 10993-5 non-cytotoxicpost sterilizationISO 10993-5, non-cytotoxic postSame
sterilization
Maintenance ofSterility90days90daysSame
Shelf Life2 years from date of manufactureunknownSimilar

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

The design and technological characteristics of the Sterilization wrap is similar to the predicate chosen. There are minor differences between the devices including Indication for use (sterilization parameter). Size and color. Maximum Recommended Wrapped Package Content Weights,

Note 1 : Indication for use (sterilization parameter)

The proposed device and the predicate device have same sterilizaiton method application, including Pre-vacuum Steam or 100% Ethylene Oxide (EO) sterilization. For those two devices, parameter recommend are different. However, the proposed device has been conducted verification in accordance with ISO11135 and ISO 17665, performance test has been done before and after sterilizaiton, all results can meet standards requirements. Therefore, there are no different questions of safety and effectiveness questions pertaining to performance of the proposed device.

Note 2: Size and color

The proposed device and the predicate device have same dimention, but they are differences in color. However, the proposed device meets the requirements of the standard ISO 10993-5, articles after the steam and EO sterilized were used for the biocompatibility validation testing. Test result shows that the device has non-cytotoxicity. Therefore, the different technological specifications of the proposed device do not raise different questions of safety.

Note 3 Maximum Recommended Wrapped Package Content Weights

Although the maximum recommended wrap package loading are different from the predicate device,the proposed device has been conducted verification in accordance with ISO11135 and ISO 17665, Types of medical devices to be sterilized in the Pre-Vacuum 12/17

{15}------------------------------------------------

Steam Sterilization and Ethylene Oxide Sterilization including general purpose nonlumened reusable metal and nonmetal devices including devices with stainless steel diffusionr estricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments etc. All models validated for both sterilization methods with stainless steel as part of the load, which with a maximum weight of each types, range from 21bs to 9lbs, which dependent on each model's size.

Performance test has been done before and after sterilizaiton, all results can meet standards requirements. There are no different questions of safety and effectiveness questions pertaining to performance of the proposed device.

Note 4 Performance test

Both the predicate device and the proposed device has been tested including. Size and Gram weight, Bacterial Filtration Efficiency (BFE), Hydrostatic Pressure, Tensile Strength, Air Permeability, Bursting Strength, Tearing Strength, Lint Generation, Result shows that there are no different questions of safety and effectiveness questions related to performance.

Based on above analysis mentioned above, we can conclude that subject device substantial equivalence with the predicate device.

8. Performance Data

Non-Clinical Performance Test Conclusion

Non- clinical tests were conducted to verify that the proposed device met all design specifications was Substantially Equivalent (SE) to the predicate Sterilization Wraps complies with the following standards:

Test ItemsStandard and Acceptance CriteriaTest Result
Size and Gram weightASTM D3776M-20: Standard Test Methods for Mass Per Unit Area (Weight) of FabricSize : Marking dimension tolerance ±10%; Weight : Marking gram tolerance ±2g/m²Pass
BacterialASTM F 2101- 19 Standard Test Method for EvaluatingPass
FiltrationEfficiency(BFE)the Bacterial Filtration Efficiency (BFE) of MedicalFace Mask Materials, Using a Biological Aerosol ofStaphylococcus aureusBFE ≥ 90%
HydrostaticPressureAATCC 127-03 Test Method for Water Resistance:Hydrostatic PressureHydrostatic Pressure ≥ 50 cm H2OPass
TensileStrengthASTM D5034-09 Standard Test Method for BreakingStrength and Elongation of Textile Fabrics (Grab Test)MD≥ 90N; CD≥ 63NPass
AirPermeabilityASTM D737-18Standard Test Method for Air Permeabilityof Textile FabricsAir Permeability≥ 30 cfmPass
BurstingStrengthASTM D3786-18 Standard Test Method for BurstingStrength of Textile Fabrics—Diaphragm Bursting StrengthTester MethodBursting Strength≥ 130 kpa (18.86psi)Pass
TearingStrengthASTM D5587-15 Standard Test Method for Tearing Strengthof Fabrics by Trapezoid ProcedureMD≥ 50N; CD≥ 30NPass
Lint GenerationISO9073-10: 2019 Textiles — Test methods for nonwovens— Part 10: Lint and other particles generation in the dry stateCoefficient of linting ≤4.0Pass
Skin IrritationAND SkinSensitizationISO 10993- 1 Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk managementprocess.ISO 10993-5 Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicity of medical devices.ISO 10993-10:2010 Biological Evaluation of MedicalDevices – Part 10: Tests for irritation and skin sensitization.Under the conditions of the study, the test article (pouchesPass
K223600510 (k) Summary
and chemical indicator) extract did not show no significantevidence of causing skin sensitization in the guinea pigbefore and after sterilized.
Shelf-LifeValidationShelf Life Validation Test;Conducted the shelf-life testing as real time aging method.After the shelf life indicated as following,Shelf Life: 2 Years; Shelf Life after Sterilized: 90 days;The device performance shall be meet the requirements of thedevice. SAL=10-6Pass
SterilizationProcessValidation forEOEO Sterilization Process Validation Test.Conducted the EO sterilization process validation as themethod/principle of ISO11135: 2014 Sterilization of healthcare products - Ethylene oxide - Part 1: Requirements fordevelopment, validation and routine control of a sterilizationprocess for medical devices (Healthcare Product Sterilization- Ethylene Oxide-Medical Device Sterilization ProcessDevelopment. Validation and Routine ControlRequirements).Use the half cycle method to validate the EO sterilizationcycle claimed in indication for use is effective.Ethylene Oxide Sterilization: 100% Ethylene Oxide (EO)with a concentration of 360-363 mg/L @ 54°C and 40%relative humidity for 480 minutes, 24 hours aeration time at42 °CThe device performance shall be meet the requirements of thedevice with the worst loading, SAL=10-6Pass
K223600510 (k) Summary
SterilizationProcessValidation forsteamSteam Sterilization Process Validation Test.Conducted the Steam sterilization process validation as themethod/principle of AAMI / ANSI ST79: 2010 & A1:2010 &A2:2011 & A3:2012 & A4:2013, Comprehensive Guide ToSteam Sterilization And Sterility Assurance In Health CareFacilities and ISO 17665-1:2006 Sterilization of health careproducts Moist heat Part 1: Requirements for thedevelopment, validation and routine control of asterilization process for medical devices 's requirements.Pre-vacuum Steam: 134°C for 4 minutes, 30 minutes drytime.The device performance shall be meet the requirements of thedevice with the worst loading, SAL=10-6Pass
MaterialcompatibilityAfter Pre-vacuum Steam or 100% Ethylene Oxide (EO)sterilization, the materials were not degraded.Performance test including:Size and Gram weightBacterial Filtration Efficiency (BFE)Hydrostatic PressureTensile StrengthAir PermeabilityBursting StrengthTearing StrengthLint Generationshould meet requirements.Pass
EO/ECH residueISO 10993-7: 2008 Biological evaluation of medical devices— Part 7: Ethylene oxide sterilization residuals$EO≤4mg/d$ , $ECH ≤9mg/d$Pass

{16}------------------------------------------------

{17}------------------------------------------------

{18}------------------------------------------------

Non-clinical tests including physical characteristics shown in table above, shelf life validation, biocompatibility evaluation, EO Sterilization Process Validation and Steam

{19}------------------------------------------------

Sterilization Process Validation has been taken to shown all results can meet the standards. Aged device were used for material compatibility research, after Pre-vacuum Steam or 100% Ethylene Oxide (EO) sterilization, the materials were not degraded and all physical characteristics test result can still meet requirements.

Clinical Test Conclusion

No clinical study is included in this submission.

9. Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as the predicate device K160755.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).