K Number
K223567

Validate with FDA (Live)

Date Cleared
2023-07-31

(244 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CNS-2101 central monitor is a networked multi-patient monitoring system, that is intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters. The CNS-2101 does not perform any data processing on the Nihon Kohden compatible devices. When the CNS-2101 is to connect with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the CNS-2101 can:

  • · Admit and discharge patients on the Nihon Kohden network.
  • · Display and manage compatible devices' real-time patient clinical data.
  • Mimic the alarms of connected devices when a measured parameter falls outside a preset limits or when an arrhythmia is detected.
  • · Review and trend data calculated by connected Nihon Kohden devices.
  • · Store and transfer historical clinical data for the connected systems.
  • · Print patient data.

The CNS-2101 is intended for use in professional medical facilities by trained medical personnel.

Device Description

The CNS-2101 central monitor is a central monitoring device designed to support medical personnel to provide medical care to multiple patients at the same time. It acquires vital sign data from multiple monitoring devices such as bedside monitors and displays the acquired data such as ECG and pulse rate on the screen as well as informing alarms.

The CNS-2101 can communicate with other devices through a network connection. The CNS-2101 can acquire vital sign data directly from multiple monitoring devices (e.g., bedside monitors) connected to Nihon Kohen Monitoring device network or using multiple patient receivers and transmitters, or by a combination of both methods. The parameters to monitor on the central monitor can be changed as necessary by selecting a monitoring device such as a bedside monitor or transmitter and changing the parameter settings for that device.

The CNS-2101 is designed to be installed in a location outside the patient environment such as a nurse's station for central monitoring.

AI/ML Overview

The provided text is a 510(k) summary for the Nihon Kohden CNS-2101 Central Monitor. This document focuses on demonstrating substantial equivalence to a predicate device through comparison of features and adherence to recognized standards. It does not present specific acceptance criteria with numeric performance values or detailed performance study results in the way typically found for AI/ML-driven devices with diagnostic or prognostic claims.

Therefore, for the information requested:

  1. A table of acceptance criteria and the reported device performance: This information is not explicitly provided in the document. The document states that the device was tested according to international and FDA-recognized consensus standards to verify and validate its functionality and technical characteristics. However, specific performance metrics (e.g., sensitivity, specificity, accuracy) are not listed for the CNS-2101.

  2. Sample size used for the test set and the data provenance: This information is not provided. The testing described is verification and validation (V&V) against design specifications and consensus standards, not clinical performance testing with patient data in the context of diagnostic accuracy.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not applicable and therefore not provided. "Ground truth" in the context of diagnostic accuracy is not relevant to the described V&V testing of a central monitoring system.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable and therefore not provided.

  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable and therefore not provided. The CNS-2101 is a central monitoring system that displays data from other devices; it does not utilize AI to assist human readers in interpretation or diagnosis.

  6. If a standalone (i.e., algorithm-only without human-in-the-loop performance) was done: Not applicable and therefore not provided. The CNS-2101 is a networked multi-patient monitoring system, not a standalone diagnostic algorithm. It displays data from already cleared bedside monitors and telemetry systems. It "does not perform any data processing on the data received from the Nihon Kohden compatible devices."

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable and therefore not provided, as the testing relates to the functional performance and safety of the system rather than diagnostic accuracy against a specific condition.

  8. The sample size for the training set: Not applicable and therefore not provided. The document describes a central monitoring system, not an AI/ML device that requires training data.

  9. How the ground truth for the training set was established: Not applicable and therefore not provided.

Summary of available information regarding performance:

  • Acceptance Criteria: Not explicitly defined with performance metrics like accuracy, sensitivity, or specificity. Instead, the acceptance criteria are implied by adherence to recognized consensus standards for medical electrical equipment, software lifecycle, risk management, usability, electromagnetic compatibility, and alarm systems.
  • Study Proving Device Meets Acceptance Criteria: Non-clinical verification and validation testing was conducted on the CNS-2101 based on the following standards and guidance:
    • ISO 14971:2019 Medical devices - Application of risk management to medical devices
    • IEC 62304:2006/A1:2015 Medical device software - Software life cycle processes
    • ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment—Part 1: General requirements for basic safety and essential performance
    • IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
    • IEC 60601-1-8: 2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
    • IEC 60601-2-27:2011 Medical electrical equipment--Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
    • ANSI/AAMI/IEC 62366-1:2015 Medical devices -Part 1: Application of usability engineering to medical devices.
    • Various FDA guidance documents related to software validation, cybersecurity, interoperable medical devices, off-the-shelf software, EMC, and human factors.
  • Reported Device Performance: The document states that "All function in the CNS-2101 have been validated by the FDA-recognized consensus standards, therefore, the validity of the results can be ensured." It also notes that "Tests conducted within the software development cycle demonstrate that communication between the CNS-2101 and the Nihon Kohden 510(k)-cleared monitoring devices is robust and stable enough to exchange accurate physiological data in real time."

In essence, the document confirms that the CNS-2101 complies with relevant safety and performance standards for its intended function as a central monitoring system, without requiring a clinical validity study involving comparison to a "ground truth" or expert adjudication for diagnostic accuracy.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 31, 2023

Nihon Kohden Corporation % Sandra Gadeyne Sr. Director, Quality and Regulatory Affairs Nihon Kohden America 15353 Barranca Pkwy Irvine, California 92618

Re: K223567

Trade/Device Name: CNS-2101 Central Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX, MHX, DRQ, MWI Dated: November 29, 2022 Received: November 29, 2022

Dear Sandra Gadeyne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shruti N. Mistry -S

for

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223567

Device Name

CNS-2101 Central Monitor

The CNS-2101 central monitor is a networked multi-patient monitoring system, that is intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters. The CNS-2101 does not perform any data processing on the Nihon Kohden compatible devices. When the CNS-2101 is to connect with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the CNS-2101 can:

  • · Admit and discharge patients on the Nihon Kohden network.
  • · Display and manage compatible devices' real-time patient clinical data.
  • Mimic the alarms of connected devices when a measured parameter falls outside a preset limits or when an arrhythmia is detected.
  • · Review and trend data calculated by connected Nihon Kohden devices.
  • · Store and transfer historical clinical data for the connected systems.
  • · Print patient data.

The CNS-2101 is intended for use in professional medical facilities by trained medical personnel.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Nihon Kohden, a Japanese medical equipment manufacturer. The logo consists of a blue square with a white abstract design on the left and the company name "NIHON KOHDEN" in blue, sans-serif font on the right. The abstract design in the square resembles a stylized wave or a rising sun.

510(K) SUMMARY GENERAL PROVISIONS 1.

Table 1 Administrative Information
SponsorNihon Kohden Corporation1-31-4 Nishiochiai, Shinjuku-KuTokyo, Japan 161-8560
Initial ImporterNihon Kohden America15353 Barranca ParkwayIrvine, CA
Company Representative/U.S. AgentSandra Gadeyne, MBA, ASQ-CQASr. Director, Quality Assurance and Regulatory Assurance15353 Barranca ParkwayIrvine, CAOffice Phone: (949) 268-7708Mobile Phone: (949) 356-3401Email: Sandra Gadeyne@nihonkohden.com
Official Correspondent/Primary ContactHiroko HagiwaraGeneral Manager, Clinical Development and Regulatory AffairsDivision, NIHON KOHDEN CORPORATIONOffice Phone: +81-3-5996-8156Mobile Phone: +81-70-5541-9661Email: Hiroko Hagiwara@mb2.nkc.co.jp

Table 2 Submission Information

Submission TypeTraditional 510(k)
Common Device NameSystem, Network and Communication, Physiological Monitor
Regulation Medical SpecialtyCardiovascular
Classification PanelCardiovascular
Product CodesMSX
Premarket ReviewCardiovascular Devices (OHT2)
Cardiac Electrophysiology, Diagnostics, and Monitoring Devices
(DHT2A)
Regulation Number870.2300
ClassificationClass II
Decision Type510(k)
Proprietary NameCNS-2101 Central Monitor
Marketing NamesCNS-2101, CNS-2101 Central Monitoring System
Submission BasisNew Device;
Previous 510(k) SubmissionsNone
PredicatesBeneVision Central Monitoring System (K193391)
Date PreparedNovember 29, 2022

PRODUCT DESCRIPTIONS 2.

The CNS-2101 central monitor is a central monitoring device designed to support medical personnel to provide medical care to multiple patients at the same time. It acquires vital sign data from multiple

{4}------------------------------------------------

Traditional 510(k) - CNS-2101 Central Monitor 510(k) Summary

monitoring devices such as bedside monitors and displays the acquired data such as ECG and pulse rate on the screen as well as informing alarms.

The CNS-2101 can communicate with other devices through a network connection. The CNS-2101 can acquire vital sign data directly from multiple monitoring devices (e.g., bedside monitors) connected to Nihon Kohen Monitoring device network or using multiple patient receivers and transmitters, or by a combination of both methods. The parameters to monitor on the central monitor can be changed as necessary by selecting a monitoring device such as a bedside monitor or transmitter and changing the parameter settings for that device.

The CNS-2101 is designed to be installed in a location outside the patient environment such as a nurse's station for central monitoring.

3. INDICATIONS FOR USE/INTENDED USE

The CNS-2101 central monitor is a networked multi-patient monitoring system, that is intended to display, record and print monitored physiological data from the Nihon Kohden bedside monitors, telemetry receiver and/or transmitters.

The CNS-2101 does not perform any data processing on the data received from the Nihon Kohden compatible devices. When the CNS-2101 is to connect with the Nihon Kohden bedside monitors and telemetry receivers/transmitters, the CNS-2101 can:

  • . Admit and discharge patients on the Nihon Kohden network.
  • o Display and manage compatible devices' real-time patient clinical data.
  • . Mimic the alarms of connected devices when a measured parameter falls outside a preset limit or when an arrhythmia is detected.
  • . Review and trend data calculated by connected Nihon Kohden devices.
  • Store and transfer historical clinical data for the connected systems. ●
  • Print patient data.

The CNS-2101 is intended for use in professional medical facilities by trained medical personnel.

SUBMISSION SCOPE 4.

Nihon Kohden (NK) is requesting market clearance for the CNS-2101 and accessories. The accessories include software licenses, peripheral devices and networking materials. The CNS-2101 communicates with other NK devices on the NK network (NET-9). The previously cleared devices compatible with the CNS-2101 are provided in Table 3.

Table 5 NK Compatible Devices
Device nameProductcode510(k)numberClearance date
Life Scope® BSM-1700 Series Bedside MonitorMHXK220976July 21, 2022
Life Scope® BSM-6000 Bedside Monitoring SystemMHXK213316Dec 29, 2021
Life Scope® BSM-3000 Bedside MonitorMHXK213316Dec 29, 2021
Life Scope® CSM-1901 Bedside Monitoring SystemMHXK213316Dec 29, 2021
Life Scope® G5 Bedside Monitoring SystemMHXK203435July 16, 2021
Life Scope® G7 Bedside Monitoring SystemMHXK203435July 16, 2021
Nihon Kohden CNS-6200 Series CNS Mode1-6201/6801MHXK102376Dec. 7, 2010

Table 3 NK Compatible Devices

Nihon Kohden Corporation

{5}------------------------------------------------

Device nameProductcode510(k)numberClearance date
Nihon Kohden SVM-7200 series Vital Signs MonitorMWIK190468May 28, 2019
Nihon Kohden Vital Sign Telemeter GZ-140PMHX/DRGK163459Apr. 4, 2017
Nihon Kohden Vital Sign Telemeter, GZ-120P/GZ-130PMHX/DRGK153707Sept. 2, 2016
Transmitter ZS-940PA Series, ZM 520/521/530/531DRTK043517Feb. 3, 2005
Multiple Patient Receiver, Model ORG-9700/9100ADRGK071058June 29, 2007

5. COMPARISON WITH THE PREDICATE DEVICE

Subject Device Information 5.1

Table 4 provides the regulations that the CNS-2101 is the subject device. with the product code MSX

Table 4 Regulatory Information on CNS-2101
Regulation NoProduct codeDevice classificationClassification
§ 870.2300MSXSystem, Network andCommunication, PhysiologicalMonitorsClass II (performance standards)

Table 4, Regulatory Information on CNS-2101

5.2 Predicate Device Information

Table 5 lists the basic information about the predicate device including 510(k) number, device trade name, 510(k) holder, and clearance date. Regulatory information used for comparison is provided in Table 6.

Table 5 Predicate Device General Information

510(k) product510(k) holderClearance date
BeneVision CentralMonitoring System (K193391)Shenzhen MindrayBio-medicalElectronics Co., LTD.22 April, 2020

Table 6 Predicate Device Regulatory Information

Regulation NoProduct codeDevice classificationClassification
§870.2300MSXSystem, Network andCommunication,Physiological MonitorsClass II(performance standards)

{6}------------------------------------------------

Comparison Table 5.3

Table 7 Comparison Table of CNS-2101 Central Monitor and its Predicate
------------------------------------------------------------------------
CharacteristicsCNS-2101 Central Monitor(Proposed device)BeneVision Central Monitoring System(Predicate Device: K193391)Comparison
Regulation items
Classification PanelCardiovascularCardiovascularSame
Device NameSystem, Network And Communication,Physiological MonitorsSystem, Network And Communication,Physiological MonitorsSame
Regulatory ClassClass II (performance standards)Class II (performance standards)Same
Regulatory Number21CFR 870.2300- Cardiac monitor (includingcardiotachometer and rate alarm)21CFR 870.2300 -Cardiac monitor (includingcardiotachometer and rate alarm)Same
Product CodePrimary Product Codes:MSX - System, network andcommunication, physiological monitorsSubsequent Product Codes:MHX- Monitor, physiological, patient(witharrhythmia detection or alarms)DRQ- Amplifier and signalconditioner, transducer signalMWI- Monitor, Physiological, Patient(Without Arrhythmia DetectionOr Alarms)Primary Product Codes:MSX - System, network and communication,physiological monitorsSubsequent Product Codes:DRQ- amplifier and signal conditioner,transducer signalMHX- monitor, physiological, patient(witharrhythmia detection or alarms)DRT- monitor, cardiac (incl.cardiotachometer & rate alarm)DXN- system, measurement, bloodpressure, non-invasiveDQA- OximeterDSB- Impedance plethysmographBoth devices share the sameMSX primary code andperformance specifications.The subsequent(secondary) product codes arerelated to the type of the devicesthe CNS-2101 can connect to.The NK systems were clearedwith the additional secondarycodes.
CharacteristicsCNS-2101 Central Monitor(Proposed device)BeneVision Central Monitoring System(Predicate Device: K193391)Comparison
Indications for useThe CNS-2101 central monitor is a networked multi-patient monitoring system, that is intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters. The CNS-2101 does not perform any data processing on the data received from the Nihon Kohden compatible devices. When the CNS-2101 is to connect with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the CNS-2101 can:Admit and discharge patients on the Nihon Kohden network. Display and manage compatible devices' real-time patient clinical data. Mimic the alarms of connected devices when a measured parameter falls outside a preset limits or when an arrhythmia is detected. Review and trend data calculated by connected Nihon Kohden devices. Store and transfer historical clinical data for the connected systems. Print patient data. The CNS-2101 is intended for use in professional medical facilities by trained medical personnel.The indications for use of the BeneVision Central Monitoring System include: Real time viewing of patient clinical data and alarms Storage and historical review of patient clinical data and alarms Printing of real time and historical patient data Configuration of local settings as well as synchronizing settings across the network to a remote device Transfer of patient clinical data and settings between several CentralStations The BeneVision Central Monitoring System is a networked patient monitoring system intended for use in a fixed location, installed in professional healthcare facilities to provide clinicians remote patient monitoring. The network connections between the various devices can be any combination of Ethernet (Wired), Wireless WIFI (WLAN), and Wireless WMTS. The BeneVision Central Monitoring System supports one or more Mindray compatible physiological monitors and will display, store, print, and transfer information received from the compatible monitors; The BeneVision Central Monitoring System supports bi-directional configuration of the compatible monitors. No data processing is done by the BeneVision Central Monitoring System for data received from compatible monitors.Same - Both devices are compatible with their commercial monitoring devices. The BeneVision includes additional devices; the NK's systems are cleared under a separate 510(k) and is equivalent in terms of functions.
CharacteristicsCNS-2101 Central Monitor(Proposed device)BeneVision Central Monitoring System(Predicate Device: K193391)Comparison
The telemetry monitoring systems are designed toacquire and monitor physiological data forambulating patients within a defined coveragearea. The BeneVision Central Monitoring Systemsupports Telemetry Systems: TMS-6016,Telepack-608, TMS60, TM80, and TM70.The TMS-6016 transmitter is intended foruse on Adult and Pediatric patients tomonitor ECG and SpO2 physiologicaldata. The Panorama Telepack-608 transmitteris intended for use on Adult patients tomonitor ECG and SpO2 physiologicaldata. The TMS60 transmitter is intended foruse on Adult and Pediatric patients overthree years old to monitor ECG, SpO2,NIBP and Resp physiological data. Thephysiological data can be reviewedlocally on the display of the transmitter. The CentralStation will support ECG,Heart Rate, SpO2, NIBP, Resp, PulseRate, Arrhythmia analysis, QTmonitoring, andST Segment Analysis for the TMS60. The TM80/TM70 telemetry monitor isintended for use on Adult and Pediatricpatients over three years old to monitorECG, SpO2, NIBP and Respphysiological data. The physiologicaldata can be analyzed, alarmed, stored,reviewed locally on the display of themonitor, and the CentralStation canconfigure and display the physiologicalparameters from the TM80/TM70. The BeneVision Central Monitoring System isintended for use in professional healthcarefacilities under the direct supervision of a licensedhealthcare practitioner.
CharacteristicsCNS-2101 Central Monitor(Proposed device)BeneVision Central Monitoring System(Predicate Device: K193391)Comparison
Hardware
Display1,920×1,080 resolution23.8-inch color TFT type LCD1920 x 1080 resolution,19" 16:9 widescreenLCD display, 21.5" and 23"The BeneVision offers 2 sizes:the CNS-2101 is offered as 1size that is equivalent to theBeneVisions larger size and is anLCD screen. The difference isminor.
Operation systemMicrosoft Windows 10 IoT(Embedded)Microsoft Windows 7 ProfessionalEmbedded SP1Microsoft Windows 10Both devices support the currentversion of Windows 10. MS isno longer supporting Window 7.The difference does not affecttheir substantial equivalence.
Storage media480 GB SSD (Solid State Drive)500 GB Hard Disk DriveBoth devices have storagecapacity. The difference does notaffect their substantialequivalence.
Monitor designPC Integrated main unit and displayInto 1 devicePC unit and display are separatedThe CNS-2101 has an integrateddesign. Both devices arecompliant with IEC 60601-1:2005+A1:2012. The differencedoes not affect their substantialequivalence.
USB ports52The CNS-2101 has an integrateddesign. Both devices arecompliant with IEC 60601-1:2005+A1:2012. The differencedoes not affect their substantialequivalence.
USB lockoutYesUnknownThe additional USB port lockoutis a cybersecurity enhancementand does not affect theperformance of the device.
AudioBuilt-in speakersBuilt-in speakersSame
Operation Methods
TouchscreenYesYesSame
KeyboardYesYesSame
MouseYesYesSame
Thermal array recordercompatibleYes-Nihon Kohden WS-140P Recorder UnitYes-Mindray thermal array module productSame - Both devices use thermalrecorders.
CharacteristicsCNS-2101 Central MonitorBeneVision Central Monitoring SystemComparison
(Proposed device)(Predicate Device: K193391)
Installation methodStand / Wall mountStand onlyThe CNS-2101 can be wallmounted or be free standing on adesk. The BeneVision is onlyoffered as a free standing device.The difference does not affecttheir substantial equivalence.
Power sourceAC Power SuppliedAC Power SuppliedSame
Power interruptionsupportYes- Internal battery (3-minute backup)Yes- UPSBoth devices have a batterybackup system.
NetworkYes (NK Network)YesSame
Compatible monitoringsystemsNK Bedside Monitors (MHX):BSM: 1700, 3000, 6000, G9, G5, G7Vital Signs Monitor (MWI): SVM-7200NK Telemetry (MHX/DRG): GZ-120/130/140Multiple Patient Receiver and Transmitters(DRG): ORG-9700/9100(DRT): ZS-940, ZM-520/521/530/531Central Monitor (MHX): CNS- 6201 and 6801Supports the following telemetry systems thatincluded its premarket notification scope:- TMS-6016 (K162607)- TMS60 (K162607)- Telepack-608 (K162607)- TM80(K162607)- TM70(K193391)Same - The CNS-2101communicates with othermonitoring systems within themanufacturer's productportfolio.
Extended displayYesYesSame
Max connections aresupported for onecentral stationUp to 32Up to 32Same
Communicationprotocol (andcompatiblemonitors)NET-9/LS-NET communicationCMS+ protocolELAN protocolMD2 protocolSame - Both devices use acommunication protocol tocommunicate between deviceson the network.
Output to EMRYesYesSame
Serial data outputcapableYesYesSame
Display
Wave displayYesYesSame
Numeric displayYesYesSame
Patient Admit/Discharge
Patientadmit/discharge/transferFollowing functions are supported• Admit, discharge, pause, and transferfunctions• Patient transfer within one central station• Patient transfer between central stations• Entering patient information (Manual,Auto)Following functions are supported• Admit, discharge, pause, and transferfunctions• Patient transfer within one central• Patient transfer between central stations• Entering patient information (Manual, Auto)Same
Review window
Review windows• Trend window• Tabular Trend window• Full Disclosure window• Expanded Waveform window• Arrhythmia Recall window• ST Recall window• Event List window• Alarm Events window• ECG 12 Lead Analysis window• Hemodynamics List window• SpO2 Trend window• Minitrend• Trend review• Full disclosure• C.O. review• NIBP review• Event review• OxyCRG review• Historic review• 12-lead Review• ST reviewSame - Both devices can connectand display monitoredparameters. The viewable datapoints are similar on each of themonitoring system.
Review history• Trendgraph data: 120 hours• Tabular trend data: 120 hours• Full disclosure and expanded waveformdata: 120 hours• Dynamic short trend: 8 hours• Trend review: 240 hours• Wave review: 240 hours of full-disclosurewaveforms and compressed waveforms• NIBP review: Most recent 3000 NIBPmeasurements• Event review: 3000 events• 12-lead review: 720 12-lead analysis results,12 analysis waveforms for each analysisresult• ST review: Most recent 240 hours of STsegments• Cardiac output review: 720 measurementsSame - Both devices providereview history data on thecentral monitor system each ofthe system can store data. Thedifferences do not affect theirsubstantial equivalence.
Alarm
Compliance standardIEC 60601-1-8: 2012IEC 60601-1-8: 2012Same
Setting alarm prioritiesBedside monitor: NoTelemetry system and transmitter: YesYes (For telemetry system)Same
Silence AlarmYesYesSame
Printing and Recording
PrintPatient information (Hospital Name, Bedname, Patient name)Recording start time or file creation dateand timeRecording type (Multiple Waveform,Expanded Waveform, Trend, TabularTrend, Full Disclosure, Arrhythmia Recall,ST Recall, Event List, Alarm Event, ECG12 Lead Analysis, Hemodynamics List)and Review window parameters• Trend window• Tabular Trend window• Full Disclosure window• Expanded Waveform window• Arrhythmia Recall window• ST Recall window• Event List window• Alarm Events window• ECG 12 Lead Analysis window• Hemodynamics List window• SpO2 Trend windowPatient information, real-time waveform, real-timealarm, Alarm Settings, Multilead ECG Report, CSAReport, waveform review, Arrhythmia StatisticResult, Trend Review, C.O. measurement, events, 12-lead Review, ST review, QT View Report, drugcalculations, hemodynamics calculations,oxygenation calculations, ventilation calculations,renal calculations, ICG hemodynamic parameter,CCO hemodynamic parameter, SvO2/ScvO2oxygenation parametersSame - Both devices are capableof printing to an externalnetwork printer.
RecordingPatient information (Hospital name, Bedname, Patient name)Recording start time or file creation dateand timeRecording type (Multiple Waveform,Expanded Waveform, Trend, TabularTrend, Full Disclosure, Arrhythmia Recall,ST Recall, Event List, Alarm Event, ECG12 Lead Analysis, Hemodynamics List)Recording duration, Recording speed (25or 50 mm/s), Numeric data, Waveforms(up to 16 channels), Sensitivity, LeadWaveform PrintingPatient information, real-time waveform,real-time alarm, waveform review, C.O.measurement, events, 12-lead Review, STreview, drug calculation, hemodynamicscalculations, oxygenation calculations,ventilation calculations, renalcalculations, ICG hemodynamicparameter, CCO hemodynamic parameter,SvO2/ScvO2 oxygenation parametersSame - Both devices are capableof printing on a thermal printerwhen the parameters fall outsidethe limits.
ECG features
ECG functions displayYesYesThis item does not affect the
HR displayYesYessubstantial equivalence at the
ST alarm page settingsUpper Limit: OFF, -1.99 to 2.00 mVLower Limit: OFF, -2.00 to 1.99 mVRange: -2.0 to 2.0 mVpoint that the central monitoritself does not perform vital signmeasurements or dataprocessing.
ST segment analysisdisplayYesYes
Arrhythmia detectiondisplayYesYes
Arrhythmia analysissettingYesYes
QTc/QRSd displaySpot check and continuousWhen connected to BSM-1700, G5 and G7Yes
QRS Detection displayYesYes
Pace pulse rejection(Display)YesYes

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

Traditional 510(k) - CNS-2101 Central Monitor 510(k) Summary

{14}------------------------------------------------

6. PERFORMANCE DATA

6.1 Summary of Non-Clinical Performance Testing

NK conducted non-clinical verification and validation testing for the CNS-2101. These tests were performed on the CNS-2101 according to the international and FDA-recognized consensus standards listed below. All function in the CNS-2101 have been validated by the FDA-recognized consensus standards, therefore, the validity of the results can be ensured.

RecognitionnumberDescription
Standard
5-125ISO 14971 :2019 Medical devices - Application of risk management to medical devices
13-79IEC 62304:2006/A1:2015 Medical device software -Software life cycle processes
19-4ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment—Part 1: General requirements for basic safety and essential performance
19-8IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
5-76IEC 60601-1-8: 2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
3-126IEC 60601-2-27:2011 Medical electrical equipment--Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
5-114ANSI/AAMI/IEC 62366-1:2015, Medical devices -Part 1: Application of usability engineering to medical devices.
Guidance
"General Principles of Software Validation" Document issued on: January 11, 2002
"Content of Premarket Submissions for Software Contained in Medical Devices."Document Issued on May 11, 2005
Draft Guidance "Content of Premarket Submissions for Device Software Functions"Document Issued on November 4, 2021
"Content of Premarket Submissions for Management of Cybersecurity in Medical Devices"Document Issued on October 2, 2014.
Draft Guidance "Cybersecurity in Medical devices: Quality System Considerations andContents of Premarket Submissions." Document Issued on April 8, 2022.
"Postmarket Management of Cybersecurity in Medical Devices" Document Issued onDecember 28, 2016
"Design Considerations and Pre-market Submission Recommendations for InteroperableMedical Devices" Document issued on September 6, 2017
"Off-The-Shelf Software Use in Medical Devices" Document Issued on September 27,2019.
"Electromagnetic Compatibility (EMC) of Medical Devices" Document issued on June 6,2022.
"Applying Human Factors and Usability Engineering to Medical Devices" Document issuedon: February 3, 2016

Table 8 Applied Standards and Guidance List for CNS-2101

{15}------------------------------------------------

Traditional 510(k) - CNS-2101 Central Monitor 510(k) Summary

The functionality and technical characteristics of the CNS-2101 have been verified and validated by testing based on the above FDA-recognized standards and FDA guidance.

6.2 Substantial Equivalence Discussion

The comparison of the CNS-2101 central monitor to its predicate device (BeneVision / K193391) and the discussion of the key aspects are presented in the following sections.

6.2.1 Discussion for Intended Use/Indications for Use

The predicate and subject devices share the following general functions.

  • Real-time viewing of measured patient clinical data.
  • Receiving alarm indications and generating an alarm including vital sign alarm, arrhythmia alarm and technical alarm.
  • Review and trend application data calculated by connected monitoring devices.
  • Configuration settings for connected monitoring devices.
  • o Connections to other systems not associated with active patient monitoring. such as information systems.
  • Storage and Historical review of patient clinical data and alarms.
  • Printing of real time and historical patient data.

The design verification and validation procedure confirmed for the operation of the software and hardware of the device is in accordance to the design specifications.

Therefore, based on the above, Nihon Kohden believes that the CNS-2101 central monitor is substantially equivalent to the predicate device.

6.2.2 Discussion for Technological Difference

This section provides description for two minor technical differences between the CNS-2101 and the BeneVision.

The following two minor hardware differences between the CNS-2101 and the BeneVision are not considered significant.

  • The CNS-2101 PC main unit and display are integrated into one unit.
  • The CNS-2101 operates on AC power, but has a built-in sub-battery for use in the event of a sudden loss of power, such as a power interruption.

Electrical safety and Electromagnetic compatibility tests etc. were conducted in consideration of the CNS-2101's built-in battery of the CNS-2101. It was concluded that their hardware differences do not raise concerns about the effectiveness and safety of the CNS-2101 as compared to the predicate device.

{16}------------------------------------------------

6.2.3 Submission Scope (Presence or Absence of Monitoring Device)

In contrast to the predicate device, the CNS-2101 does not include a vital sign measuring device in the scope of its premarket notification. The CNS-2101 is a networked device that receives and displays vital signs measured by the Nihon Kohden 510(k)-cleared vital signs monitoring devices through the network. Tests conducted within the software development cycle demonstrate that communication between the CNS-2101 and the Nihon Kohden 510(k)-cleared monitoring devices is robust and stable enough to exchange accurate physiological data in real time. Therefore, this difference in the scope of submission does not affect its effectiveness and safety as compared to the predicate device.

CONCLUSION 6.3

The results of the substantial equivalence assessment, taken together with non-clinical bench testing, electrical safety, and electromagnetic compatibility, software verification, and validation demonstrate that the CNS-2101 does not raise concerns regarding its safety and effectiveness compared to its predicate device and operates in accordance with claimed indications for use.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).