K Number
K223495

Validate with FDA (Live)

Date Cleared
2023-05-02

(162 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Prescription Use: Zeolite Hemostatic Cotton is intended for temporary external use to control traumatic bleeding.

Over-The-Counter Use: Zeolite Hemostatic Cotton is intended for temporary external use to stop bleeding of superficial wounds, minor cuts, and abrasions.

Device Description

The Zeolite Hemostatic Cotton consists of zeolite and cotton. It is provided in a sterile dressing format that conforms readily to the wound. There are 15 models of Zeolite Hemostatic Cotton, and the size ranges from 0.1g to 10g. The difference between each model is weight.

AI/ML Overview

The provided document, K223495, is a 510(k) Premarket Notification for the Zeolite Hemostatic Cotton. It details non-clinical tests conducted to demonstrate the device's substantial equivalence to a legally marketed predicate device (K211570, Zeolite Hemostatic Gauze).

It's important to note that this submission does not include clinical study data (as explicitly stated in Section 6: "No clinical study is included in this submission."). Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of this document refer to non-clinical performance testing and biocompatibility assessments, not to AI-driven diagnostic or assistive technologies, nor to comparative effectiveness studies involving human readers and AI. The device described is a medical device for managing bleeding, not an AI software/device.

Given this context, I will address your points where applicable based on the provided document, acknowledging that many of your questions are geared towards AI/software medical devices or clinical trials, which are not present here.


Acceptance Criteria and Reported Device Performance (Non-Clinical)

The acceptance criteria and reported performance are based on physical performance testing, sterile barrier packaging testing, sterilization and shelf-life testing, and biocompatibility testing.

Here's a table summarizing the acceptance criteria and reported performance specifically from the "Non-Clinical Test Conclusion" section (Section 5) of the document:

Table 1: Acceptance Criteria and Reported Device Performance (Non-Clinical)

Test CategorySpecific Test / ParameterAcceptance CriteriaReported Device Performance
Physical PerformanceWater Absorption≥10gTest results all meet the requirements (implies ≥10g)
Zeolite ContentNot less than 10%Test results all meet the requirements (implies ≥10%)
Hydration Temperature Rise≤3.0℃Test results all meet the requirements (implies ≤3.0℃)
Sterile Barrier & Shelf-LifeVisual Inspection (ASTM F1886/F1886M)Pass (integrity of seals)Test results show device package can maintain integrity for 3 years
Seal Strength (ASTM F88/F88M)Pass (seal integrity)Test results show device package can maintain integrity for 3 years
Dye Penetration Test (ASTM F3039)Pass (no leakage)Test results show device package can maintain integrity for 3 years
Vacuum Leak (ASTM D3078)Pass (no leaks)Test results show device package can maintain integrity for 3 years
Sterility (ISO 11737-2)SterileTest results show device package can maintain integrity for 3 years
Endotoxin Limit (USP <85>)Did not exceed 20 EU/10gDid not exceed 20 EU/10g
Shelf-Life Evaluation (Appearance, Water Absorption, Zeolite Content, Heat Release, Package Integrity)Maintain performance meet acceptance criteria for 3 yearsAccelerated stability test results showed device maintains performance for 3 years
BiocompatibilityCytotoxicityNo CytotoxicityNo Cytotoxicity (biocompatible)
SensitizationNo SensitizationNo Sensitization (biocompatible)
Intracutaneous ReactivityNo Intracutaneous ReactivityNo Intracutaneous Reactivity (biocompatible)
Acute Systemic ToxicityNo Acute Systemic ToxicityNo Acute Systemic Toxicity (biocompatible)
Pyrogen Testing (USP <151>)No PyrogenicityNo Pyrogen (biocompatible)

Responses to Specific Questions (based on the provided document):

  1. A table of acceptance criteria and the reported device performance:
    (See Table 1 above)

  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: The document mentions "Three discrete batches of subject device were tested for water absorption, Zeolite content, and heat release." For other tests (e.g., packaging integrity, biocompatibility), the sample sizes are not explicitly stated within the public summary, but it's implied that sufficient samples were tested to meet the requirements of the listed ISO and ASTM standards.
    • Data Provenance: The tests were non-clinical, likely performed in laboratories affiliated with the manufacturer or their testing partners. The manufacturer is Hangzhou Zeo-Innov Life Technology Co., Ltd. in China. The data would be prospective, as these are tests performed on newly manufactured devices for regulatory submission.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • This question is not applicable. The "ground truth" here is defined by meeting predefined engineering, material science, and biological safety standards (e.g., specific thresholds for absorption, temperature, endotoxins, or qualitative assessments like "no cytotoxicity"). These are determined by lab analyses and standard protocols, not by expert consensus on clinical interpretation. There are no human "experts" establishing "ground truth" in the sense of clinical diagnoses in this context.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies or studies where human readers interpret data, often with conflicting interpretations requiring a tie-breaker. This document focuses on non-clinical laboratory testing where results are quantitative or qualitative against established thresholds, not subjective interpretations requiring adjudication.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. The document explicitly states: "No clinical study is included in this submission." This device is not an AI-assisted diagnostic tool, and no MRMC study was performed or required for its 510(k) clearance based on substantial equivalence to a predicate hemostatic device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. This device is a physical medical product (hemostatic cotton), not an algorithm or software. Therefore, there is no "standalone performance" of an algorithm.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the non-clinical tests, the "ground truth" is based on pre-defined quantitative and qualitative performance specifications outlined in recognized international and national standards (e.g., ASTM, ISO, USP). For example, water absorption must be ≥10g, or certain biocompatibility tests must show "no cytotoxicity." This is a technical ground truth based on scientific and engineering principles rather than clinical outcomes or expert consensus.
  8. The sample size for the training set:

    • This is not applicable. This device is a physical product, not an AI/ML model that requires a training set.
  9. How the ground truth for the training set was established:

    • This is not applicable as there is no training set for a physical medical device.

In summary: The provided FDA 510(k) summary for the Zeolite Hemostatic Cotton focuses entirely on non-clinical testing (physical properties, packaging, sterilization, biocompatibility) to demonstrate substantial equivalence. It explicitly states that no clinical studies were performed. Therefore, questions pertaining to AI/ML, human reader studies, clinical ground truth establishment, or training sets are not relevant to this document. The "acceptance criteria" discussed are for the device's physical and biological performance as per established standards for medical devices of this type.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 2, 2023

Hangzhou Zeo-Innov Life Technology Co., Ltd Zhao Lijuan Chief Executive Officer 101, Building 2, No.291 Fucheng Road Hangzhou, Zhejiang 310018 China

Re: K223495

Trade/Device Name: Zeolite Hemostatic Cotton Regulatory Class: Unclassified Product Code: FRO Dated: March 31, 2023 Received: March 31, 2023

Dear Zhao Lijuan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Julie A. Morabito -S

Julie A. Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223495

Device Name Zeolite Hemostatic Cotton

Indications for Use (Describe)

Prescription Use: Zeolite Hemostatic Cotton is intended for temporary external use to control traumatic bleeding.

Over-The-Counter Use: Zeolite Hemostatic Cotton is intended for temporary external use to stop bleeding of superficial wounds, minor cuts, and abrasions.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K223495

1. Date of Preparation: 04/24/2023

2. Sponsor Identification

Hangzhou Zeo-Innov Life Technology Co., Ltd.

101, building 2, No.291, Fucheng Road, Hangzhou, 310018, China

Establishment Registration Number: 600558

Contact Person: Lijuan Zhao Position: Chief Executive Officer Tel: +86- 571-86651470 Fax: +86- 571-86651470 Email: zhaolj@zeo-innov.com

3. Identification of Proposed Device

Trade Name: Zeolite Hemostatic Cotton Common Name: Hemostatic Cotton

Regulatory Information Product code: FRO Classification Name: Dressing, Wound, Drug Classification: Unclassified;

Review Panel: General and Plastic Surgery

Indication for Use:

Prescription Use: Zeolite Hemostatic Cotton is intended for temporary external use to control traumatic bleeding.

Over-The-Counter Use: Zeolite Hemostatic Cotton is intended for temporary external use to stop bleeding of superficial wounds, minor cuts, and abrasions.

Device Description

The Zeolite Hemostatic Cotton consists of zeolite and cotton. It is provided in a sterile dressing format that conforms readily to the wound. There are 15 models of Zeolite Hemostatic Cotton, and the size ranges from 0.1g to 10g. The difference between each model is weight.

4. Identification of Predicate Device

510(k) number: K211570 Product Name: Zeolite Hemostatic Gauze

{4}------------------------------------------------

5. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

  • ASTM F1886/F1886M: 2016 Standard Test Method for Determining Integrity of Seals for A Flexible Packaging by Visual Inspection
  • A ASTM F88/F88M: 2021 Standard Test Method for Seal Strength of Flexible Barrier Materials
  • ASTM F1980: 2021 Standard Guide for Accelerated Aging of Sterile Barrier Systems for A Medical Devices
  • ASTM F3039: 2015 Standard Test Method for Leakage of Nonporous Packaging or Flexible Barrier Materials by Dye Penetration

  • A ASTM D3078:2002 (Reapproved 2021)e1 Standard Test Method for Determination of Leaks in Flexible Packaging by Bubble Emission
  • ISO 11737-2:2019 Sterilization of health care products - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process

  • USP <85> Bacterial Endotoxins Test >
  • USP <151> Pyrogen Test (USP Rabbit Test)

  • ISO 10993-5: 2009 Biological Evaluation of Medical Devices-Part 5: Tests for in Vitro Cytotoxicitv

  • ISO 10993-10: 2021 Biological Evaluation of Medical Devices-Part 10: Tests for Skin A Sensitization
  • ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

  • ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation

The following performance data were provided in support of the substantial equivalence determination.

Physical performance testing

Three discrete batches of subject device were tested for water absorption, Zeolite content, and heat release. The water absorption of the devices is ≥10g, and the hydration temperature rise is ≤3.0℃. The Zeolite amount of the proposed device is not less than 10%. The physical performance test results all meet the requirements of the acceptance criteria.

Sterile barrier packaging testing were performed on the proposed device, which include visual inspection (ASTM F1886/F1886M: 2016), seal strength (ASTM F88/F88M: 2021), dye penetration test (ASTM F3039: 2015), vacuum leak (ASTM D3078:2002 (Reapproved 2021)e1) and sterility (ISO 11737-2:2019 ). The test result showed that the device package can maintain its integrity during a shelf life of 3 years.

{5}------------------------------------------------

Sterilization and shelf-life testing listed in following table were performed on the proposed device. Endotoxin limit did not exceed 20EU/10g. Accelerated stability test results showed that the device can maintain its performance meet the acceptance criteria during a shelf life of 3 years.

Bacteria Endotoxin LimitUSP <85>
Shelf-Life EvaluationAppearance, water absorption, Zeolite content,heat release, and package integrity tests wereperformed on samples of after accelerated agingto verify the claimed shelf life of the device.

Biocompatibility testing

The contact level of the proposed device is breached or compromised surfaces, and the contact duration is limited contact (≤24 hours). The proposed device was evaluated for the following tests. The results for the biocompatibility testing showed that the proposed device is biocompatible.

  • A Cytotoxicity,
  • Sensitization,

  • Intracutaneous Reactivity,

  • Acute Systemic Toxicity,

  • Pyrogen testing.

6. Clinical Test

No clinical study is included in this submission.

7. Substantially Equivalent (SE) Comparison

Table 1 Comparison of Technology Characteristics
ItemProposed DevicePredicate DeviceComparison
K223495K211570
Product CodeFROFROSame
ClassUnclassifiedUnclassifiedSame
Indication for UsePrescription Use:Zeolite Hemostatic Cotton isintended for temporary externaluse to control traumatic bleeding.Over-The-Counter Use:Zeolite Hemostatic Cotton isintended for temporary externaluse to stop bleeding of superficialPrescription Use:Zeolite Hemostatic Gauze isintended for temporary externaluse to control traumatic bleeding.Over-The-Counter Use:Zeolite Hemostatic Gauze isintended for temporary externaluse to stop bleeding of superficialSame

{6}------------------------------------------------

wounds, minor cuts, andabrasions.wounds, minor cuts, andabrasions.
MaterialCottonZeoliteGauzeZeoliteAnalysis 1
Single UseYesYesSame
SizeP-2cm×2cm-2,Analysis 2
P-2cm×2cm-4,
P-3cm×25cm-2,
P-3cm×25cm-4,
P-5cm×5cm-4,
P-5cm×5cm-8,
P-5cm×5cm-12,
0.1g,P-8cm×8cm-4,
0.2g,P-8cm×8cm-8,
0.3g,P-8cm×8cm-12,
0.4g,P-8cm×8cm-16,
0.5g,P-8cm×8cm-24,
0.6g,P-9cm×10cm-4,
0.7g,P-9cm×10cm-8,
0.8g,P-9cm×10cm-12,
0.9g,P-9cm×10cm-16,
1.0g,P-9cm×10cm-24,
1.2g,P-10cm×15cm-4,
1.5g,P-10cm×15cm-8,
3g,P-10cm×15cm-12,
5g,P-10cm×15cm-16,
10gP-10cm×15cm-24,
P-12cm×12cm-4,
P-12cm×12cm-8,
P-12cm×12cm-12,
P-12cm×12cm-16,
P-12cm×12cm-24,
P-15cm×15cm-4,
P-15cm×15cm-8,
P-15cm×15cm-12,
P-15cm×15cm-16,
P-15cm×15cm-24,
P-20cm×20cm-4,

{7}------------------------------------------------

P-20cm×20cm-8,
P-20cm×20cm-12,
P-20cm×20cm-16,
P-20cm×20cm-24,
P-30cm×30cm-4,
P-30cm×30cm-8,
P-30cm×30cm-12,
P-30cm×30cm-16,
P-30cm×30cm-24;
J-5cm×150cm,
J-5cm×300cm,
J-8cm×150cm,
J-8cm×300cm,
J-10cm×150cm,
J-10cm×300cm;
Z-5cm×150cm×4cm,
Z-5cm×150cm×6cm,
Z-5cm×150cm×8cm,
Z-5cm×350cm×6cm,
Z-5cm×350cm×8cm,
Z-8cm×150cm×4cm,
Z-8cm×150cm×6cm,
Z-8cm×150cm×8cm,
Z-8cm×150cm×10cm,
Z-8cm×150cm×12cm,
Z-8cm×350cm×4cm,
Z-8cm×350cm×6cm,
Z-8cm×350cm×8cm,
Z-8cm×350cm×10cm,
Z-8cm×350cm×12cm,
Z-10cm×150cm×4cm,
Z-10cm×150cm×6cm,
Z-10cm×150cm×8cm,
Z-10cm×150cm×10cm,
Z-10cm×150cm×12cm,
Z-10cm×350cm×4cm,
Z-10cm×350cm×6cm,
Z-10cm×350cm×8cm,

{8}------------------------------------------------

Z-10cm×350cm×10cm,Z-10cm×350cm×12cm;L-0.8cm×3.8cm,L-1cm×3.8cm,L-1.2cm×3.8cm,L-1.5cm×3.0cm,L-1.8cm×3.0cm,L-2cm×2.5cm
CytotoxicityNo CytotoxicityNo CytotoxicitySame
PyrogenNo PyrogenNo PyrogenSame
SensitizationNo SensitizationNo SensitizationSame
IntracutaneousReactivityNo Intracutaneous ReactivityNo Intracutaneous ReactivitySame
Acute SystemicToxicityNo Acute Systemic ToxicityNo Acute Systemic ToxicitySame
SterilizationCo60 Radiation sterilizationCo60 Radiation sterilizationSame
SAL10-610-6Same
Label and LabelingConform with 21CFR Part 801Conform with 21CFR Part 801Same

Analysis 1-Material

The hemostatic material of the proposed device is the predicate device. The proposed device materials are cotton and Zeolite, and the predicate device materials are gauze and Zeolite. The gauze is also composed of cotton fiber.

Analysis 2-Size

The configuration of the subject device and the predicate device are different in representation; the subject device is shown in mass, whereas the predicate device is shown based on length, width, and height. The mass of the subject device is similar to the predicate device, and the amount of Zeolite is the same, ≥10% by weight.

8. Substantially Equivalent Conclusion

The conclusion drawn from the non-clinical tests demonstrate that the proposed device in 510(k) submission, Zeolite Hemostatic Cotton, substantial equivalence to the device cleared under K211570, Zeolite Hemostatic Gauze.

N/A