K Number
K223464

Validate with FDA (Live)

Date Cleared
2023-03-14

(117 days)

Product Code
Regulation Number
890.5650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Leg and Foot Air Wave Pressure Therapy Device EP-1129 is intended for the temporary relief of minor muscle aches and pains and for the temporary increase in circulation to the treated areas in pood health, it can simulate kneading and stroking of tissues by using an inflatable garment.

Device Description

Leg and Foot Air Wave Pressure Therapy Device EP-1129 is a powered inflatable tube massager, comprised of an air compressor with intermittent pneumatic controller. 3-chamber sleeves covered with polyester fiber, and a connectable hose for connecting the device to the sleeves. The device can be powered by an external power supply, or be powered by an internal lithium ion battery. The sleeves have 3-chambers and different applicable body areas, such as the Foot Leg. The sleeves can be inflating and deflating sequentially to apply the circulating pressure on the target body areas which controlled by the main unit. The device simulates manual kneading and stroking of tissues by sequential inflated sleeves, to temporarily relieve minor muscle aches and/or pains and to temporarily increase circulation to the treated areas. It is to be used by adults who are in good health. The recommended treatment time is 15 minutes per time, and recommend to take one time treatment every two days.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, the "Leg and Foot Air Wave Pressure Therapy Device." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving novel safety and effectiveness. Therefore, the information provided does not include a detailed study with acceptance criteria and reported device performance in the way a clinical trial for a new therapeutic device would.

However, based on the non-clinical testing section and the comparison to predicate devices, we can infer the types of acceptance criteria that would have been used for the safety and performance evaluations.

Here's an attempt to structure the information according to your requested points, making inferences where the exact details are not explicitly stated in the provided documents:

1. A table of acceptance criteria and the reported device performance

Since this is a 510(k) submission, the "performance" is demonstrated by adherence to recognized standards and comparison to a predicate device, rather than specific clinical outcome metrics. The acceptance criteria for such a device are typically compliance with safety and performance standards.

Acceptance Criteria CategorySpecific Standard/Test (Implicit Acceptance)Reported Device Performance (Compliance)
Electrical SafetyANSI/AAMI ES60601-1: 2005+A1:2012; AMD2:2021Compliant with the standard.
Electromagnetic Compatibility (EMC)IEC 60601-1-2: 2014Compliant with the standard.
Home Healthcare Environment SafetyIEC 60601-1-11 Edition 2.0 2015-01Compliant with the standard.
Lithium Battery SafetyIEC 62133-2:2017Compliant with the standard.
Software ValidationFDA Guidance for the Content of Premarket Submissions for Software Contained in Medical DevicesCompliant with FDA guidance.
Product Service LifeProduct service life verification testsAll labeling claims supported.
Sleeve IntegrityBurst Strength TestAll labeling claims supported.
Overall Performance after ReliabilityProduct performance testing after reliability testingAll labeling claims supported.
Usability/Human FactorsHuman Factors Validation Testing (label comprehension/self-selection, usability/user interface studies)Human factors concerns evaluated and addressed.
Functional EquivalenceComparison to Predicate Device (K193354) and Reference Device (K220572) on design, intended use, indications, functions, material."The differences between subject device and predicate device do not raise any new questions of safety or effectiveness." (Substantially equivalent)
Operational ParametersDevice Pressure range, Inflation time, Keep time, Deflation time, Cycle time (compared to predicate)The subject device's parameters (e.g., 90-225mmHg pressure, 20-50s inflation, 1-3s keep, 5s deflation, 30s-3mins cycle) are within expected ranges and comparable to predicate devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not explicitly stated for specific non-clinical tests. These validations typically involve a set number of units for various environmental, electrical, and mechanical stress tests, which can vary depending on the standard. For the Usability Study, the sample size is also not specified, but such studies usually involve a small, representative group of users.
  • Data Provenance: The testing was "evaluated the safety and performance by lab bench testing" and included a "Usability Study." This indicates prospective, controlled testing. The submission is from China (manufacturer: ZHANGZHOU EASEPAL INNOVATION CO., LTD. in Fujian, China; correspondent: Guangzhou KEDA Biological Technology Co., Ltd in Guangzhou, China). Therefore, the testing likely occurred in China or at certified labs engaged by the manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • For the technical standards (electrical safety, EMC, etc.), the "ground truth" is established by the standard itself, and expert judgment is applied in performing and interpreting the tests according to the standard. The document does not specify the number or specific qualifications of experts involved in conducting these tests or reviewing the results for the submission. The presumption is that qualified engineers and technicians performed these analyses.
  • For the Usability Study, the "ground truth" would be the observed user interactions and feedback. The number and qualifications of evaluators (human factors experts) are not specified.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of medical images or symptoms where there is subjective assessment by multiple readers. This device is a physical therapy device with non-clinical testing for safety and performance against objective standards. Therefore, such adjudication methods would not be applicable or described in this type of submission. The compliance to standards is generally verifiable through a single, qualified test.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. This device is a physical therapy device (a powered inflatable tube massager) intended for direct user application for temporary relief of minor muscle aches and pains and increased circulation. It does not involve "human readers" or AI assistance in the interpretation of medical data, so this type of study is not relevant. The document explicitly states: "Clinical testing is not necessary for the subject device."

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not applicable. The device is a physical therapy device, not an AI algorithm. Its performance is inherent to its mechanical and electrical function as controlled by an internal microprocessor, not through a standalone AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's evaluation is primarily:

  • Compliance with recognized international and national safety and performance standards (e.g., ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62133-2).
  • Demonstrated functional parameters (e.g., pressure range, cycle times, inflation rates) that are comparable to legally marketed predicate devices, implying similar therapeutic effect without requiring new clinical outcome data.
  • Validation of software according to FDA guidance.
  • Usability testing to ensure the device can be used safely and effectively by its intended lay operators.

8. The sample size for the training set

This question is not applicable. The device is not an AI/ML algorithm that requires a "training set." It's a hardware device with an embedded microprocessor and software for control.

9. How the ground truth for the training set was established

This question is not applicable, as there is no "training set" for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 14, 2023

Zhangzhou Easepal Innovation Co Ltd. % Jet Li Regulation Manager Guangzhou KEDA Biological Tech Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District GuangZhou City, China Guangzhou, Guangdong 510000 China

Re: K223464

Trade/Device Name: Leg and Foot Air Wave Pressure Therapy Device Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: March 3, 2023 Received: March 3, 2023

Dear Jet Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Heather L. Dean -S

Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223464

Device Name

Leg and Foot Air Wave Pressure Therapy Device

Indications for Use (Describe)

Leg and Foot Air Wave Pressure Therapy Device EP-1129 is intended for the temporary relief of minor muscle aches and pains and for the temporary increase in circulation to the treated areas in pood health, it can simulate kneading and stroking of tissues by using an inflatable garment.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K223464.

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements according to 21 CFR 807.92 (c), and there were no prior submissions for the subject device.

1. Submitter Information

Sponsor: ZHANGZHOU EASEPALINNOVATION CO., LTD.

Address: No.36 Longkun Road, Hongjian Village, Jiaomei Town, Zhangzhou Taiwanese Investment

Zone, Fujian CHINA 363000

Contact Person: Xu Zibing

Title: Quality Manager

Phone: 13275000278

E-mail: eqc-oca@easepal.com.cn

Application Correspondent: Jet Li Company: Guangzhou KEDA Biological Technology Co., Ltd E-mail: med-jl@foxmail.com Phone: 86-18588874857 Address: 6F, No.1 TianTairoad, Science City, LuoGang District, GuangZhou City, China

2. Subject Device Information

Type of 510(k) submission: Traditional Common Name: Leg and Foot Air Wave Pressure Therapy Device Model: EP-1129 Classification Name: Powered inflatable tube massager Review Panel: Physical Medicine Product Code: IRP

{4}------------------------------------------------

Regulation Number: 21 CFR 890.5650

Regulation Class: 2

ItemPredicate DeviceReference Device
SponsorShenzhen Dongjilian ElectronicsCo., Ltd.Ceragem International, Inc.
Device NameAir Compression TherapyDeviceCeragem Automatic ThermalMassager, Model CGM MB-1701 &CGM MB-1702
510(k) NumberK193354K220572
Product CodeIRPIRP
Regulation Number21 CFR 890.565021 CFR 890.5880,21 CFR 890.5650
Regulation Class22

3. Predicate and Reference Device Information

4. Device Description

Leg and Foot Air Wave Pressure Therapy Device EP-1129 is a powered inflatable tube massager, comprised of an air compressor with intermittent pneumatic controller. 3-chamber sleeves covered with polyester fiber, and a connectable hose for connecting the device to the sleeves.

The device can be powered by an external power supply, or be powered by an internal lithium ion battery. The sleeves have 3-chambers and different applicable body areas, such as the Foot Leg. The sleeves can be inflating and deflating sequentially to apply the circulating pressure on the target body areas which controlled by the main unit.

The device simulates manual kneading and stroking of tissues by sequential inflated sleeves, to temporarily relieve minor muscle aches and/or pains and to temporarily increase circulation to the treated areas. It is to be used by adults who are in good health.

The recommended treatment time is 15 minutes per time, and recommend to take one time treatment every two days.

5. Intended Use/Indication for use

Leg and Foot Air Wave Pressure Therapy Device EP-1129 is intended for the temporary relief of minor muscle aches and pains and for the temporary increase in circulation to the treated areas in people who are in good heath, it can simulate kneading and stroking of tissues by using an inflatable garment.

6. TestSummary

{5}------------------------------------------------

Non Clinical testing:

The Leg and Foot Air Wave Pressure Therapy Device has been evaluated the safety and performance by lab bench testing according to the following standards:

  • -ANSI/AAMI ES60601-1: 2005+A1:2012; AMD2:2021, Medical Electrical Equipment - Part 1: General Requirements for Safety,
  • -IEC 60601-1-2: 2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
  • -IEC 60601-1-11 Edition 2.0 2015-01, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment,
  • -IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications- Part 2: Lithium systems
  • -The software was validated according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

The device also was verified with the following benching testing and usability study:

  • Product service life verification -
  • -Burst Strength Test
  • Product performance testing after reliability testing -

Tests results are supporting all labeling claims in order to establish substantial equivalency.

Usability Study was completed in the subject device:

A human Factors Validation Testing was done to evaluate the HFE/ UE concerns, including label comprehension/self-selection and usability/user interface studies.

Clinical testing: Clinical testing is not necessary for the subject device.

7. Comparison to Predicate Device

Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise any new questions of safety or effectiveness. See the below form.

{6}------------------------------------------------

Elements ofcomparisonSubject DevicePredicate DeviceReference Device
ManufacturerZHANGZHOU EASEPALINNOVATION CO., LTD.Shenzhen Dongjilian Electronics Co., Ltd.Ceragem International, Inc.
510K numberK223464K193354K220572
Product NameLeg and Foot Air Wave PressureTherapy DeviceModel EP-1129Air Compression Therapy Device (S9019)Ceragem Automatic Thermal Massager, ModelCGM MB-1701 & CGM MB-1702
Regulation Number21 CFR 890.565021 CFR 890.565021 CFR 890.5880,21 CFR 890.5650
Classification NamePowered inflatable tube massagerPowered inflatable tube massagerPowered inflatable tube massager
Regulation Class222
OTC & RxOTCOTCOTC
Indications for UseLeg and Foot Air Wave PressureTherapy Device EP-1129 isintended for the temporary relief ofminor muscle aches and pains andfor the temporary increase incirculation to the treated areas inpeople who are in good health, itcan simulate kneading andstroking of tissues by using aninflatable garment.The Air Compression Therapy Device isindicated for the temporary relief of minormuscle aches and pains and fortemporary increase in circulation to thetreated areas in people who are in goodhealth. The Air Compression TherapyDevice simulates kneading and stroking oftissues by using an inflatable garment.The intended use of the Ceragem AutomaticThermal Massager, Model CGM MB-1701 & CGMMB-1702 is to provide muscle relaxation therapy bydelivering heat and soothing massage.Additionally, the product provides topical radiantinfrared heat for:-Temporary relief of minor muscle and joint painstiffness-Temporary relief of minor joint pain associated witharthritis-Temporary increase in local circulation whereapplied- Relaxation of musclesThe Air Cell Massager (only CGM MB-1701) isindicated for the temporary relief of minor muscleaches and pains and for temporary increase in
Elements ofcomparisonSubject DevicePredicate DeviceReference Device
Power SourceAC100-240V, 50/60Hz100~240V 50/60Hz100-127Vac 50/60Hz
Power consumption11.1W12W480VA
Dimensions (WHD)340210850mm10.25.925.6 (in)/
PhotoImage: Subject Device RemoteImage: Predicate Device Remote/
Weight2.0 Kg(4.4pounds)4.6 pounds/
Size and appearanceof sleeves (leg part)Image: Subject Device Leg SleevesOne size: 85*34 cmImage: Predicate Device Leg SleevesOne size: 73*26 cmImage: Reference Device Leg SleeveLeg:One size: 28*58.4cm
Inflation time20-50 s3-30 s1 min ~ 1 min 30 s
Keep time1-3 s1-5 s1 ~ 13 s
Elements ofcomparisonSubject DevicePredicate DeviceReference Device
Deflation time5 s1-5 s1 ~ 10 s
Cycle time30 seconds to 3 minsRange of 25 sec to 3 min 40 secRange of 28 sec to 39 sec
Number of Chambers3 Chambers3 Chambers4 Chambers
Sleeve MaterialsTribute satin and Oxford clothNylon with a Polyurethane laminateOxford and Nylon
Mode of CompressionSequentialSequentialSequential
Device Pressure range90-225mmHg0-240 mmHg48~240 mmHg
Air pressure level/Compression levels4 levels settings:Level 1:90mmHg;Level 2:135mmHg;Level 3: 188mmHgLevel 4: 225mmHg3 levels settings:low level:150mmHg;Mid level:185mmHg;High Level: 215mmHg9 levels settings:Level 1 : 48mmHgLevel 2 : 72mmHgLevel 3 : 96mmHgLevel 4 : 120mmHgLevel 5 : 144mmHgLevel 6 : 168mmHgLevel 7 : 192mmHgLevel 8 : 216mmHgLevel 9 : 240mmHg
Treatment TimeDefault as 15min20 min18 minutes
Work ModeImage: Compression ModeMode 1:Starting with the foot chamber andprogressing up the thigh chamber, eachImage: Diagram of foot and leg
Elements ofcomparisonSubject DevicePredicate DeviceReference Device
Image: Subject Device
M1: ① → ② → ③ (sole air bagsinflate - deflate - calf air baginflates - deflate - thigh airbags inflate - deflate - cycle inturn)M2: ① → ①② → ①②③ (Soleair bags inflate - pressuremaintaining-- calf air bag inflate--pressure maintaining- thigh airbags inflate - sole, calf, thigh airbags deflate simultaneously --cycle in turn)M3: ① → ② → ③ → ① → ①②→ ①②③ (Sole air bags inflatedeflate - calf air bag inflates -deflate - thigh air bags inflate -deflate - sole air bags inflate -pressure maintaining - calf airbag inflates - pressuremaintaining - thigh air bagsinflate- sole, calf, thigh air bagsdeflate simultaneously- cycle inturn)Cycle in turn.section compresses and thepressure gradually rises to the pre-determined air pressure level, thendecompresses and the air pressure drops.Once the thigh section decompresses, thecycle begins again.Mode 1 follows this pressure sequence:Image: Predicate Device Mode 1Mode 2:Starting with the foot chamber andprogressing up the thigh, each sectioncompresses and the pressure graduallyrises to the pre-determined air pressurelevel, holds the air until the entire garmentis compressed. All three sections thendecompress simultaneously and the airpressure drops, then cycle begins again.Mode 2 follows this pressure sequence:Image: Predicate Device Mode 2Mode 1:Starting with the foot chamber and progressing upthe thigh chamber, each section compresses andthe pressure gradually rises to the pre-determinedair pressure level, then decompresses and the airpressure drops. Once the thigh sectiondecompresses, the cycle begins again.Mode 2:Starting with the foot chamber and progressing upthe thigh, each section compresses, and thepressure gradually rises to the pre-determined airpressure level, holds the air until the entire garmentis compressed. All four sections then decompresssimultaneously, and the air pressure drops, thencycle begins again.Mode 3:Start between the foot chamber to the thighchamber except for No. 4 chamber. And No.4chamber is compressed while the other chambersare maintained. After that all four sectionsdecompress simultaneously and the air pressuredrops, then cycle begins again. This sequence issimilar to the Mode 2.
Elements ofcomparisonSubject DevicePredicate DeviceReference Device
Mode 3:include two stage, stage 1: it worksaccording to the method of mode 1, afterthe stage 1 is completed, it goes to stage2(working according to the method of mode2) without interruption time until finish thestage 2, then enter next cycle withoutinterruption.Mode1 $\rightleftharpoons$ Mode2The pressure sequence of mode 3combines mode 1 and mode 2
Patient contactNon-conductive appliancesNon-conductive appliancesNon-conductive attachments
Software/Firmware/Microprocessor ControlMicroprocessorMicroprocessorMicroprocessor
TechnologyCompressor and valve systemwhich sequentially inflates cells ofapplianceCompressor and valve system whichsequentially inflates cells of applianceCompressor and valve system which sequentiallyinflates cells of appliance
Electrical safety, EMCANSI/AAMI ES60601-1IEC 60601-1-2IEC 60601-1-11IEC 62133ES 60601-1;IEC 60601-1-2;ISO 10993-5;ISO 10993-10;IEC 60601-1-11ANSI AAMI ES60601-1:2005IEC 60601-1-2 Edition 4.0IEC 60601-1-6 Edition 3.1IEC 60601-1-11 Edition 2.0ISO 10993-5 Third editionISO 10993-10 Third Edition

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

8. Conclusion

The subject device Leg and Foot Air Wave Pressure Therapy Device has all features of the predicate devices for intended use. Thus, the subject device is substantially equivalent to the predicate devices.

9. Summary Prepared Date

3 March 2023

§ 890.5650 Powered inflatable tube massager.

(a)
Identification. A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.(b)
Classification. Class II (performance standards).