AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HOY A Illumina (sorafilcon A) Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -12.00D, exhibiting astigmatism of 1.00 diopter or less that does not interfere with visual acuity.

The HOY A Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -10.00D. The lenses may be worm by persons who exhibit astigmatism of up to 2.75 diopters that does not interfere with visual acuity.

The HOY A Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) and presbyopia in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -10.00D with add power ranging from +0.50D to +2.50D. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopter or less that does not interfere with visual acuity.

The sorafilcon A contact lenses are to be prescribed for single-use disposable wear and are to be discarded upon each removal.

Device Description

The HOY A sorafilcon A material is made from a hydrophilic siloxane copolymer of N-vinyl pyrrolidone and is 48% water by weight when immersed in a sterile phosphate buffered saline packaging solution. A UV-absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280 nm to 315 nm and less than 50% in the UVA range of 316 nm to 380 nm.

The HOYA sorafilcon A Contact Lens is to be prescribed for single-use disposable wear.

The physical properties of the lens are:
Refractive Index: 1.408
Light Transmittance: ≥ 95%
Water Content: 48%
Specific Gravity: 1.123
Surface Character: Hydrophilic
Oxygen Permeability: 112 x 10-11 (cm²/sec)(cm³O₂)/(mL x mmHg) @ 35 °C (polarographic method).

The lenses will be manufactured with the following parameters:
Diameter: 14.0 mm to 14.7 mm
Center Thickness: 0.062 mm to 0.330 mm
Base Curve: 8.40 mm to 9.20 mm
Power Range: +5.00D to -12.00D
Cylinder Power (Toric): -0.75D to -2.75D
Cylinder Axis: 0 o to 180 o in 10 o increments
Add Power Range: +0.50D to +2.50D

AI/ML Overview

The provided text describes the 510(k) premarket notification for HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens, HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens, and HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens. The document outlines acceptance criteria and the study conducted to prove the device meets these criteria, primarily focusing on demonstrating substantial equivalence to predicate devices.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly list "acceptance criteria" in a numerical table format for device performance metrics for the clinical study. Instead, it discusses the meeting of "primary safety hypotheses" and "primary efficacy hypotheses." The non-clinical testing section lists various tests conducted and states that "All test results met the pre-established acceptance criteria." However, the document does not provide the numerical values for these pre-established acceptance criteria for each test, nor the specific numerical results of the subject device for each non-clinical test.

For the clinical study, the acceptance criteria relate to non-inferiority for safety and visual acuity targets for efficacy.

Clinical Performance (Efficacy and Safety):

Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred/Stated)Reported Device Performance
SafetyNon-inferiority to control lens with respect to percentage of eyes with ocular adverse events across all follow-up visits, with a non-inferiority margin of 5%."The low incidence of ocular adverse events ensured that both of the primary safety hypotheses were met." "No serious or significant ocular adverse events were reported." (20 reported adverse events in 17 subjects, 5 ocular AEs in 4 subjects, 2 lens-related ocular AEs - bilateral redness, all classified as non-significant).
Efficacy (Visual Acuity)Proportion of eyes with VA with contact lenses and best sphere-correction corrected to 1.0 (0.0 logMAR, 20/20) or better to meet a target threshold of 90% at each visit.VA to the nearest line: 94% to 99% for test lens, 94% to 100% for control lens. ("Both lenses achieved the target threshold of 90% at each visit.")
Efficacy (Visual Acuity)VA to the nearest letter: 83% to 93% for test lens, 83% to 91% for control lens. (Note: "with habitual lenses, the proportion was also below the target threshold of 90%: 77% of test eyes and 91% of control eyes.")
Mean sphere-corrected VA at follow-up (13 weeks) to have upper and lower limit values of the 95% confidence intervals better than 0.0 logMAR (1.0 decimal, 20/20).Mean sphere-corrected VAs at follow-up were 0.07 logMAR for test lens and -0.06 logMAR for control lens. ("At each visit, the upper and lower limit values of the 95% confidence intervals were better than 0.0 logMAR... and, therefore, a primary efficacy hypothesis was met.")

Non-Clinical Testing: The document states that "All test results met the pre-established acceptance criteria," for various tests (e.g., lens shape, water content, light transmittance, biocompatibility). However, the specific numerical acceptance criteria and the actual numerical results are not provided in the text.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Clinical Test Set): A total of 133 subjects were enrolled. 80 subjects completed the study as per protocol (59 subjects wearing test lenses, 21 subjects wearing control lenses).
  • Data Provenance: The clinical trial was a multicenter study conducted in USA and UK. It was a prospective trial, described as a "three-month, double-masked (subject and investigator), bilateral, randomized clinical trial."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not provided in the document. The study design does not appear to involve human experts interpreting data for a "ground truth" in the way it might for an AI diagnostic device. Instead, the ground truth for safety and efficacy in this clinical trial would be derived from objective measurements (e.g., visual acuity, recorded adverse events) and clinical assessments by the investigators and subjects themselves, rather than a consensus of independent experts adjudicating outcomes.

4. Adjudication Method for the Test Set

The document describes a clinical trial for contact lenses. The concept of "adjudication method (e.g., 2+1, 3+1)" is typically specific to image-based diagnostic studies where multiple readers interpret images to establish a consensus ground truth. This type of adjudication method is not applicable and therefore not mentioned in the context of this contact lens clinical trial. Safety and efficacy outcomes were assessed via standard clinical trial methodologies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. An MRMC study is relevant for evaluating the performance of diagnostic devices, especially those involving human interpretation of images, and often assesses the impact of AI assistance on human readers. This study evaluates the safety and efficacy of a contact lens itself. The "control lens" served as a comparator for the device's performance directly, not as an unassisted human reading condition in an MRMC setting.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This question is not applicable for a contact lens. The device is a physical contact lens, not a software algorithm. Therefore, there is no "standalone" algorithm performance to evaluate. The study assessed the device's performance in human subjects.

7. The Type of Ground Truth Used

The ground truth used for this clinical study was based on:

  • Clinical Measurements and Observations: Visual acuity measurements (logMAR, decimal, 20/20), and objective and subjective reporting of ocular adverse events by study investigators and subjects.
  • Comparison to Predicate Device: The study's conclusion of "substantial equivalence" implies that the performance of the HOYA Illumina lenses (test device) was compared directly to that of the MyDay® (stenfilcon A) Soft Daily Disposable Contact Lens (control/predicate device) as the de facto "ground truth" for proving equivalence in a regulatory context.

8. The Sample Size for the Training Set

This information is not applicable. Contact lenses are physical medical devices, not AI/ML algorithms that require a "training set" of data in the computational sense. The clinical study was a product evaluation, not an algorithm development process.

9. How the Ground Truth for the Training Set was Established

This information is not applicable as there is no "training set" for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 03, 2024

HOYA Lamphun Ltd. Timothy Ting Global QA/RA Director Northern Region Industrial Estate, 75/2 Moo 4, Tambol Banklang, Amphur Muang, Lamphun 51000 Thailand

Re: K223397

Trade/Device Name: HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens, HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens, HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens.

Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL. MVN Dated: December 2, 2023 Received: December 5, 2023

Dear Timothy Ting:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

J Angelo Green -S

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K223397

Device Name

HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens

Indications for Use (Describe)

The HOY A Illumina (sorafilcon A) Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -12.00D, exhibiting astigmatism of 1.00 diopter or less that does not interfere with visual acuity.

The HOY A Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -10.00D. The lenses may be worm by persons who exhibit astigmatism of up to 2.75 diopters that does not interfere with visual acuity.

The HOY A Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) and presbyopia in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -10.00D with add power ranging from +0.50D to +2.50D. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopter or less that does not interfere with visual acuity.

The sorafilcon A contact lenses are to be prescribed for single-use disposable wear and are to be discarded upon each removal.

Type of Use (Select one or both, as applicable)
❌ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."


{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "HOYA" in blue font. A horizontal line extends from the left of the "H" to the right of the "O", bisecting the "O" in the middle. The font is sans-serif and the letters are evenly spaced.

Submission No.: K223397

510(k) Summary K223397

SUBMITTER INFORMATION

Date Prepared: November 10, 2022 Name: HOYA Lamphun Ltd. Address: Northern Region Industrial Estate, 75/2 Moo 4, Tambol Banklang, Amphur Muang, Lamphun 51000 Thailand. Contact Person: Timothy Ting Global Quality Assurance/Regulatory Affairs Director Phone Number: (66) 93-132-1699 Email: chongchai.ting@hoya.com

DEVICE INFORMATION

Trade Name: HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens

Brand Name: ILLUMINA 1 DAY Soft silicone hydrogel contact lens ILLUMINA 1 DAY Toric Soft silicone hydrogel contact lens ILLUMINA 1 DAY Multifocal Soft silicone hydrogel contact lens

Common Name: Soft contact lens, daily disposable wear. Device Classification: Class II (21 CFR 886.5925) Classification Name: Soft (hydrophilic) contact lens Product Code: LPL, MVN. Classification Panel: Ophthalmic.

PREDICATE DEVICES

HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens is substantially equivalent to the following predicate devices:

Primary predicate devices MyDay® (stenfilcon A) ASPHERE Soft Contact Lens, MyDay® (stenfilcon A) TORIC Soft Contact Lens, and MyDay® (stenfilcon A) MULTIFOCAL Soft Contact Lens. Cleared under K191763.

Secondary predicate devices

Bausch + Lomb® (samfilcon A) Soft (hydrophilic) Contact Lens Bausch + Lomb® (samfilcon A) Soft (hydrophilic) Contact Lens for Astigmatism

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "HOYA" in a sans-serif font. The letters are a bright blue color. The "O" in HOYA has a horizontal line going through it, which resembles a horizon line.

Bausch + Lomb® (samfilcon A) Soft (hydrophilic) Contact Lens for Presbyopia Cleared under K131208.

DEVICE DESCRIPTION

The HOY A sorafilcon A material is made from a hydrophilic siloxane copolymer of N-vinyl pyrrolidone and is 48% water by weight when immersed in a sterile phosphate buffered saline packaging solution. A UV-absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280 nm to 315 nm and less than 50% in the UVA range of 316 nm to 380 nm.

The HOYA sorafilcon A Contact Lens is to be prescribed for single-use disposable wear.

The physical properties of the lens are:

Refractive Index1.408
Light Transmittance≥ 95%
Water Content48%
Specific Gravity1.123
Surface CharacterHydrophilic
Oxygen Permeability112 x 10-11 (cm²/sec)(cm³O₂)/(mL x mmHg) @ 35 °C(polarographic method).

The lenses will be manufactured with the following parameters:

Diameter14.0 mm to 14.7 mm
Center Thickness0.062 mm to 0.330 mm
Base Curve8.40 mm to 9.20 mm
Power Range+5.00D to -12.00D
Cylinder Power (Toric)-0.75D to -2.75D
Cylinder Axis0 o to 180 o in 10 o increments
Add Power Range+0.50D to +2.50D

INDICATIONS FOR USE

HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens

The HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) in non-aphakic person with non-diseased eyes in spherical powers ranging from +5.00D to -12.00D, exhibiting astigmatism of 1.00 diopter or less that does not interfere with visual acuity.

HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens

The HOYA Illumina (sorafilcon A) Toric Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in non-aphakic person with non-

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "HOYA" in a sans-serif font. The letters are a bright blue color. There is a horizontal line that extends from the left side of the "H" to the right side of the "A", passing through the center of the "O".

diseased eyes in spherical powers ranging from +5.00D to -10.00D. The lenses may be worn by persons who exhibit astigmatism of up to 2.75 diopters that does not interfere with visual acuity.

HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens

The HOYA Illumina (sorafilcon A) Multifocal Daily Disposable Soft Contact Lens is indicated for the correction of refractive ametropia (myopia or hyperopia) and presbyopia in non-aphakic person with nondiseased eyes in spherical powers ranging from +5.00D to -10.00D with add power ranging from +0.50D to +2.50D. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopter or less that does not interfere with visual acuity.

The sorafilcon A contact lenses are to be prescribed for single-use disposable wear and are to be discarded upon each removal.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

PropertyPrimary Predicate DeviceSecondary Predicate DeviceSubject Device
MyDay® (stenfilcon A) SoftDaily Disposable ContactLens Product FamilyK191763Bausch + Lomb (samfilcon A)Soft (hydrophilic) ContactLens Product FamilyK131208HOYA Illumina(sorafilcon A) DailyDisposable Soft ContactLens Product Family
Intended UseDaily wear and dailydisposable cornea softcontact lens for myopia,hyperopia, astigmatism,presbyopiaDaily wear and dailydisposable cornea soft contactlens for myopia, hyperopia,astigmatism, presbyopiaSame as predicate devices
FunctionalityThe contact lenses act as arefractive medium thatfocus light rays from nearand distant objects on theretina.The contact lenses act as arefractive medium that focuslight rays from near and distantobjects on the retina.Same as predicate devices
ModalityDaily disposable wearcontact lensDaily disposable wear contactlensSame as predicate devices
ManufacturingMethodCast MoldedCast MoldedSame as predicate devices
Material GroupSilicone Hydrogel FDAGroup II (high watercontent, nonionic polymer)Silicone Hydrogel FDA GroupI (low water content, nonionicpolymer)Silicone Hydrogel FDAGroup I (low watercontent, nonionic polymer)
FDA Category(Group)content, nonionic polymer)
USAN Namestenfilcon Asamfilcon Asorafilcon A
Water Content54%46%48%
OxygenPermeability*80114112
Refractive Index1.4011.4111.408

A side-by-side comparison of technological characteristics of the predicate devices and the new device:

{7}------------------------------------------------

HOYAHOYA Lamphun Ltd.Traditional 510(k) Premarket Notification HOYA Illumina(sorafilcon A) Daily DisposableSoft Contact LensSubmission Date:02-December-2023
Submission No.:K223397
PropertyPrimary Predicate DeviceMyDay® (stenfilcon A) SoftDaily Disposable ContactLens Product FamilyK191763Secondary Predicate DeviceBausch + Lomb (samfilcon A)Soft (hydrophilic) ContactLens Product FamilyK131208Subject DeviceHOYA Illumina(sorafilcon A) DailyDisposable Soft ContactLens Product Family
Specific Gravity1.0331.0481.123
Visibility TintReactive Blue 246Reactive Blue 246No color pigment wasadded
SterilizationMethodMoist HeatAutoclaveSame as predicatedevices

Note: *Oxygen Permeability shown was determine by polarographic method:

x 10-11 [(cm2/sec)x(cm3O2)/ (ml x mmHg)] @ 35 ℃

SUMMARY OF NON-CLINICAL TESTING

Non-clinical studies for HOYA sorafilcon A Contact Lenses were conducted in accordance with FDA's Class II Daily Wear Contact Lenses - Premarket Notification [510(k)] Guidance Document issued date June 27, 1994 and 21 CFR Part 58 Good Laboratory Practice for Nonclinical Laboratory Studies. Each test was conducted based on the ANSI, ISO, and ASTM standard as indicated below:

  • · Lens shape, edge thickness, inclusion and surface imperfection per ANSI Z80.20:2016 and ISO 18369-3:2017.
  • · Lens parameters per ANSI Z80.20: 2016 and ISO 18369-3:2017.
  • · Water content per ANSI Z80.20-2016 and ISO 18369-4:2017.
  • · Refractive index per ANSI Z80.20-2016 and ISO 18369-4:2017.
  • · Light transmittance per ANSI Z80.20-2016 and ISO 18369-4:2017.
  • · UV radiation transmittance per ANSI Z80.20-2016 and ISO 18369-4:2017.
  • Extractables per ANSI Z80.20-2016 and ISO 18369-4:2017. .
  • · Contact angle per ANSI Z80.20-2016.
  • · Mechanical properties per ANSI Z80.20-2016 and ASTM D1708-02a.
  • · Sterile blister packaging integrity per ASTM 1929-98.

All test results met the pre-established acceptance criteria which demonstrated the lenses are safe and effective.

SUMMARY OF BIOCOMPATIBILITY TESTING

Biocompatibility testing for HOY A sorafilcon A Contact Lenses was conducted on in accordance with the FDA's Class II Daily Wear Contact Lenses - Premarket Notification [510(k)] Guidance Document issued date June 27, 1994. ISO 10993-1:2018 Biological Evaluation of Medical Devices and 21 CFR Part 58 Good Laboratory Practice for Nonclinical Laboratory Studies as listed below:

  • · Cytotoxicity per ISO 10993-5: 2009

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "HOYA" in blue font. The font is a sans-serif font, and the letters are all capitalized. The "O" in HOYA has a horizontal line going through it. The background of the image is white.

  • . Sensitization per ISO 10993-10: 2021
  • 22 day contact lens wear study in rabbits per ISO 9394: 2012 .
  • · Acute Ocular Irritation per ISO 10993-23:2021
  • Systemic Toxicity per ISO 10993-11: 2017 .
  • · Genotoxicity (bacterial reverse mutation and chromosomal aberration assays studies) per ISO 10993-3: 2014
  • · Chemical characterization of lens materials per ISO 10993-18: 2020

The biological compatibility testing was conducted on the subject device phosphate buffered saline packaging solution as listed below:

  • · Cytotoxicity per ISO 10993-5: 2009
  • · Sensitization per ISO 10993-10: 2021
  • · Acute Ocular Irritation per ISO 10993-23:2021

The biological compatibility testing was also conducted on the subject device final packaging materials as listed below:

  • · Cytotoxicity per ISO 10993-5: 2009
  • · Acute systemic toxicity per ISO 10993-11:2017
  • · Acute Ocular Irritation per ISO 10993-23:2021

HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens was demonstrated to be biocompatible with the mucosal membrane (cornea) and the device performs as expected and is equivalent to the predicates without creating additional risk to the user.

SUMMARY OF CLINICAL PERFORMANCE DATA

HOYA conducted a three-month, double-masked (subject and investigator), bilateral, randomized, multicenter (9 centers) clinical trial was conducted in USA and UK with a total of 133 subjects enrolled. Eighty subjects completed the study as per protocol: 59 subjects wearing test lenses and 21 subjects wearing control lenses. No serious or significant ocular adverse events were reported.

The primary safety endpoint was HOYA sorafilcon A daily disposable contact lens was non-inferior to the control lens with respect to the percentage of eyes with ocular adverse events across all follow-up visits using a non-inferiority margin of 5%. No serious or significant ocular adverse events were reported. There were 20 reported adverse events in 17 subjects. Five of these were ocular adverse events in four subjects, all of which were classified as non-significant. Two of the ocular adverse events were considered lensrelated: bilateral redness. The low incidence of ocular adverse events ensured that both of the primary safety hypotheses were met.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows information about HOYA Lamphun Ltd. It mentions a Traditional 510(k) Premarket Notification for HOYA Illumina (sorafilcon A) Daily Disposable Soft Contact Lens. The submission date is 02-December-2023, and the submission number is K223397.

The primary efficacy hypotheses related to visual acuity (VA). The proportion of eyes with VA with contact lenses and best sphere-correction corrected to 1.0 (0.0 logMAR, 20/20) or better was calculated at each visit. With VA calculated to the nearest line, the proportions ranged from 94% to 99% for the test lens and 94% to 100% for the control lens. Using this criterion, both lenses achieved the target threshold of 90% at each visit. With VA calculated to the nearest letter, these proportions ranged from 83% to 93% for the test lens and 83% to 91% for the control lens. It is notable, however, that with habitual lenses, the proportion was also below the target threshold of 90%: 77% of test eyes and 91% of control eyes. The mean spherecorrected VAs at follow-up (13 weeks) were 0.07 and -0.06 logMAR for the test and control lenses respectively. At each visit, the upper and lower limit values of the 95% confidence intervals were better than 0.0 logMAR (1.0 decimal, 20/20) and, therefore, a primary efficacy hypothesis was met.

The results of the study demonstrated that HOYA Illumina (sorafilcon A) Soft Contact Lens is safe and has effectiveness substantially equivalent to the predicate device, MyDay® (stenfilcon A) Soft Daily Disposable Contact Lens.

SUBSTANTIAL EQUIVALENCE CONCLUSION

The cumulative results of the bench performance testing, in vitro and in vivo testing, biocompatibility testing, and clinical study sponsored by HOYA Lamphun Ltd. demonstrated that the safety, effectiveness, and performance of the subject device is substantially equivalent to the predicate devices.

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.