K Number
K223327

Validate with FDA (Live)

Date Cleared
2023-02-16

(108 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The product is intended to use for injecting fluids intradermally. It is for adult only (greater than 21 years old).

Device Description

The Sterile Aesthetic Cannula and Hypodermic Needle is a combined medical device, including a blunt needle-cannula and a hypodermic needle. The sharp needle consists of a needle hub, a needle, and a protective cap. The blunt needle-cannula consists of a needle hub, a blunt cannula, and a protective cap. The device is provided for single-used, sterilized, and prescription use only. It offers various model configurations, including multiple needle gauge sizes, lengths, and tips, to fit most usage scenarios.

AI/ML Overview

This document does not contain an AI/ML device. Therefore, the requested information elements related to AI/ML device performance, such as acceptance criteria, sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, and standalone performance, are not applicable and cannot be extracted.

The document describes a Sterile Aesthetic Cannula and Hypodermic Needle (K223327), which is a conventional medical device. The information provided is primarily about its physical and biological performance testing to demonstrate substantial equivalence to a predicate device.

{0}------------------------------------------------

February 16, 2023

Shanghai Kindly Enterprise Development Group Co., Ltd % Evan Hu Marketing & Technical Manager Shanghai Mind-link Consulting Co., Ltd. 1399 Jiangyue Road, Minhang Shanghai, Shanghai 201114 China

Re: K223327

Trade/Device Name: Sterile Aesthetic Cannula and Hypodermic Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: FMI Dated: February 10, 2023 Received: February 13, 2023

Dear Evan Hu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Alan M.
Stevens -
S3

CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223327

Device Name

Sterile Aesthetic Cannula and Hypodermic Needle

Indications for Use (Describe)

The product is intended to use for injecting fluids intradermally. It is for adult only (greater than 21 years old).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. Preparation date: 02/13/2023

2. Submitter

Manufacturer: Shanghai Kindly Enterprise Development Group Co., Ltd. Address: No.658 Gaochao Road, 201803, Shanghai, China Contact person: Liu Hualong, 86-02169118232, henry liu@kdlchina.net Submission correspondent: Evan Hu, 86-18616124827, Evan.hu@mind-link.net

3. Device

Trading name: Sterile Aesthetic Cannula and Hypodermic Needle Common name: Aesthetic needle Regulation No.: 21 CFR 880.5570 Classification name: Needle, Hypodermic, Single Lumen Classification: Class II Product code: FMI

4. Predicate device

Predicate device: K200017-Eclipse DermaFlex Cannula

5. Device description

The Sterile Aesthetic Cannula and Hypodermic Needle is a combined medical device, including a blunt needle-cannula and a hypodermic needle. The sharp needle consists of a needle hub, a needle, and a protective cap. The blunt needle-cannula consists of a needle hub, a blunt cannula, and a protective cap.

The device is provided for single-used, sterilized, and prescription use only. It offers various model configurations, including multiple needle gauge sizes, lengths, and tips, to fit most usage scenarios. Details of the specification refer to Table 1.

Cannula (Blunt needle)Sharp needle (Normal needle)
Gauge(G)Length(mm)Wall typeGauge(G)Length(mm)Wall type
Group A: Only cannula
2240, 50, 60, 70, 90ETWN/A
2330, 40, 50, 60ETW
2530, 40, 50, 60ETW
2613, 25, 30, 40ETW
2713, 25, 30, 40ETW
3013, 25, 30ETW
Group B: Cannula and sharp needle with the same gauge
2240ETW2225TW
2250ETW2225TW
2270ETW2225TW
2330ETW2325TW
2340ETW2325TW
2350ETW2325TW
2530ETW2516TW
2540ETW2516TW
2550ETW2516TW
2613ETW2616RW
2625ETW2616RW
2713ETW2713RW
2725ETW2713RW
2740ETW2713RW
2750ETW2713RW
3013ETW3013RW
3025ETW3013RW
Group C: Cannula and sharp needle with different gauges
2220ETW2125TW
2225ETW2125TW
2240ETW2125TW
2250ETW2125TW
2270ETW2125TW
2340ETW2225TW
2350ETW2225TW
2370ETW2225TW
2440ETW2325TW
2450ETW2325TW
2538ETW2425TW
2550ETW2425TW
2570ETW2425TW
2613ETW2525TW
2625ETW2525TW
2635ETW2525TW
2640ETW2525TW
2650ETW2525TW
2713ETW2625RW
2725ETW2625RW
2740ETW2625RW
2750ETW2625RW
3013ETW2913RW
3025ETW2913RW
3038ETW2913RW

Table 1. Device specifications

{4}------------------------------------------------

{5}------------------------------------------------

6. Indications for use/Intended use

The product is intended to use for injecting fluids intradermally. It is for adult only (greater than 21 years old).

7. Comparison of technological characters between proposed and predicate devices

CharactersProposed device(K223327- Sterile AestheticCannula and Hypodermic Needle)Predicate device(K200017-Eclipse DermaFlexCannula)Comments
Indications foruseThe product is intended to use forinjecting fluids intradermally. It isfor adult only (greater than 21years old).The Eclipse DermaFlex Cannula isintended to inject fluidsintradermally.Same
Product codeFMIFMISame
Regulation No.21 CFR 880.557021 CFR 880.5570Same
Structure andmaterialsNeedle hub: PPProtective cap: PP (cannula andsharp needle), PE (cannula)Needle tube: SUS 304Lubricant: Silicone oilAdhesive: Epoxy resin, UV resinNeedle hub: PPProtective cap: PPNeedle tube: SUS 304Lubricant: Silicone oil#1
Color codingMeet requirements of ISO 6009Meet requirements of ISO 6009Same
Needle gaugeCannula: 22, 23, 24, 25, 26, 27,30GSharp needle: 21, 22, 23, 24, 25,26, 27, 29, 30GCannula: 21, 22, 23, 25, 26, 27,30GSharp needle: 21, 22, 23, 25, 26,27, 30G#2
Needle lengthCannula: 13, 25, 35, 38, 40, 50,70mmSharp needle: 13, 16, 25mmCannula: 25, 38, 40, 50, 60,70mmSharp needle: 25, 38, 40, 50, 60,70mm#3
Tip configurationSharp needle: sharpened tipCannula: closed blunt tip withlateral openingSharp needle: sharpened tipCannula: closed blunt tip withlateral openingSame
Connection tosyringeLuer taperLuer taperSame
PhysicalperformanceMeet requirements of ISO 9626,ISO 7864 and ISO 80369-7Meet requirements of ISO 9626,ISO 7864 and ISO 80369-7Same
BiocompatibilityperformanceMeet requirements of ISO 10993-1, -4, -5, -10, -11Meet requirements of ISO 10993-1, -4, -5, -10, -11Same
SterilizationmethodSterilized by EO gasSAL 10-6Sterilized by EO gasSAL 10-6Same

Table 2. Characters comparison

{6}------------------------------------------------

#1: The materials used for manufacturing protective cap are different. The function of the protective cap is for needle protection before use. The different materials do not raise new questions of safety or effectiveness. Performance data results also showed all testing met the requirements of standards or guidance.

#2 and #3: The proposed device's needle gauge specification range is within the predicate device's range. However, the proposed device has shorter needle lengths (13mm and 16mm) than the predicate device. Shorter needle lengths offer options for end users' choice. The performance of needles with shorter lengths was tested in compliance with ISO 9626 and ISO 7864. Performance data results showed all testing met the requirements of standards or guidance. The different needle lengths do not raise new questions of safety and effectiveness.

8. Non-clinical testing

PERFORMANCE TESTING

The proposed device was tested in compliance with ISO 7864:2016 and ISO 9626:2016 for evaluating the overall non-clinical performance. Additionally, luer connector testing in compliance with ISO 80369-7:2016 and ISO 80360-20:2015 was completed. The performance and design testing results met the standards' requirements.

BIOCOMPATIBILITY TESTING:

The proposed device was tested in compliance with the 2020 FDA Guidance document Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process", as the Externally Communicating Device, Blood Path Indirect, Limited Contact (< 24hrs). Biocompatibility testing was conducted as following items.

  • Acute systemic toxicity -
  • -Pyrogenicity
  • -Skin sensitization
  • Intracutaneous reactivity -
  • -In-vitro hemolytic
  • -In-vitro cytotoxicity

Residual particles and Endotoxin were tested in compliance with USP <788> and USP <85>/USP <161>, respectively.

STERILE, PACKAGE, AND SHELF-LIFE VALIDATION:

The sterilization method by using Ethylene Oxide (EO) gas was validated in compliance with ISO 11135:2014 to achieve the sterility assurance level (SAL) of 10-6.

EO and Epoxy Chlorohydrin (ECH) residuals met the limit requirements of ISO 10993-7:2008, and Pyrogen testing met the requirements of USP <151> 0.9% saline injection method.

The Shelf-Life validation study was conducted under accelerated aging condition in compliance with ASTM F1980-16 to verify the claimed shelf-life of 5 years.

{7}------------------------------------------------

Package integrity testing under simulated shipping conditions was conducted to satisfy the requirements in ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems. All packaging was deemed acceptable for protection of product and sterility maintenance.

Sterile barrier testing was conducted in compliance with the following FDA recognized consensus standards.

  • Seal Strength ASTM F88/F88-15 –
  • Dye penetration ASTM F1929-15 ।
  • -Internal Pressurization ASTM F1140/F1140M:2013
  • -Sterility USP <71>

9. Clinical testing

Not applicable for this submission.

10. Conclusion

The differences between the predicate and the proposed device do not raise any new or different questions of safety or effectiveness. The proposed device is substantially equivalent to the predicate device with respect to indications for use, technological characteristics, and performance.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).