K Number
K223171

Validate with FDA (Live)

Date Cleared
2022-11-10

(30 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 24 hour use AquaPulse® Irrigation Tubing (tubing and accessories to accommodate various GI endoscopes and irrigation pumps) is intended to provide irrigation via a sterile water bottle during gastrontestinal endoscopic procedures when used in conjunction with an irrigation pump.

Device Description

The AquaPulse® Irrigation Tubing is intended for 24-hour use and then discarded. The irrigation tubing is manufactured for use in conjunction with sterile water bottle, and together with irrigation pumps. The irrigation tubing is individually packed in sealed package, sold as a sterile device. The irrigation tubing is designed to the auxiliary water connector or biopsy irrigation accessory and to be inserted into pump head of the irrigation pump to provide irrigation through the auxiliary water channel to the distal end of endoscope.

AI/ML Overview

This document describes a 510(k) premarket notification for the AquaPulse® Irrigation Tubing. This is not a medical device that uses AI or machine learning.

Therefore, most of the requested information regarding acceptance criteria and study proving device performance related to AI/ML (e.g., sample size for training/test sets, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth for AI) is not applicable to this submission.

Here's a breakdown of the applicable information based on the provided text, and an explanation of why other aspects are not present:


Analysis of the Provided Document for Device Acceptance Criteria and Study

The provided document is a 510(k) premarket notification for a medical device called "AquaPulse® Irrigation Tubing." This type of submission is for demonstrating "substantial equivalence" to a legally marketed predicate device, rather than proving novel efficacy or performance through extensive clinical trials as would be required for a completely new device or a device with new technology (like AI/ML).

The core of this submission revolves around bench testing and biocompatibility testing to show that the modified device is as safe and effective as its predicate. It does not involve artificial intelligence or machine learning components.


1. Table of Acceptance Criteria and Reported Device Performance

Since this is not an AI/ML device, the "acceptance criteria" are related to manufacturing specifications, material properties, sterilization, and basic functional performance, rather than diagnostic accuracy or predictive power. The document summarizes these in comparison to a predicate device to establish substantial equivalence.

SpecificationPredicate Device CharacteristicSubject Device CharacteristicSubstantial Equivalence (Acceptance)
Product CodeOCXOCXIdentical (Met)
Regulatory Classification22Identical (Met)
Regulation No.21 CFR 876.150021 CFR 876.1500Identical (Met)
Regulation DescriptionEndoscope and accessoriesEndoscope and accessoriesIdentical (Met)
CompatibilityOlympus® OFP, Endo Status™ EGA-500 and Endogator® EGP-100 Irrigation PumpOlympus® OFP, Endo Status™ EGA-500 and Endogator® EGP-100 Irrigation PumpIdentical (Met)
Intended UseThe 24 hour use AquaPulse® Irrigation Tubing... is intended to provide irrigation via a sterile water bottle during gastrointestinal endoscopic procedures when used in conjunction with an irrigation pump.The 24 hour use AquaPulse® Irrigation Tubing... is intended to provide irrigation via a sterile water bottle during gastrointestinal endoscopic procedures when used in conjunction with an irrigation pump.Identical (Met)
Environment of UseHospital and/or clinicsHospital and/or clinicsIdentical (Met)
Single Use, Disposable24-hour use then discard24-hour use then discardIdentical (Met)
MaterialPolycarbonate and Polyvinyl ChloridePolycarbonate and Polyvinyl Chloride (Material not Made with DEHP)Substantial Equivalent (Met - notes specific material detail)
PackagingIrrigation tubing is packaged in a sealed pouchIrrigation tubing is packaged in a sealed pouchIdentical (Met)
Manufacturing MethodInjection molding and extrusionInjection molding and extrusionIdentical (Met)
Backflow Prevention DesignDiaphragm in the connector allow pressurized water passing through in one-way but not the other wayDiaphragm in the connector allow pressurized water passing through in one-way but not the other wayIdentical (Met)
SterilizationEO gasEO gasIdentical (Met)
Shelf Life1 year3 yearsSubstantial Equivalent (Met - based on accelerated aging test)

Reported Device Performance:

  • Flow Performance: "The bench testing was performed to support substantial equivalence such as verification tests on flow performances of modified devices. The performance data demonstrated that the subject devices met established specifications and it demonstrated substantial equivalence to the previously cleared predicate devices." (Section 9)
  • Sterilization: "The EO sterilization cycle has a Sterility Assurance Level (SAL) of 10⁻⁶. EO residuals on the components are below the maximum levels defined in ANSI/AAMI/ISO 10993-7 Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide sterilization residuals." (Section 10)
  • Shelf Life: "The modified devices have a three (3) year shelf life, based on the design and existing sterile barrier data from the existing packaging integrity test and performance test were conducted after accelerated aging test. The test result can imply that the modified device can provide and maintain a sterile barrier and its intended performance for at least the claimed shelf life." (Section 11)
  • Biocompatibility: "The test result shows that the modified devices are biocompatible" in accordance with ISO 10993-1 for surface devices with mucosal membrane contact for limited duration (not more than 24 hours). (Section 12)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not specified in terms of number of units tested. For bench tests, it typically refers to a number of manufactured units subjected to specific physical and chemical tests (e.g., flow rate, material composition, sterilization validation samples, accelerated aging samples, biocompatibility samples). The document states "The test result shows that the modified devices are biocompatible" and "The performance data demonstrated that the subject devices met established specifications," implying samples were tested, but not the specific quantity.
  • Data Provenance: Not explicitly stated as "country of origin of the data." These are likely lab/bench test results generated by the manufacturer (GA Health Company Limited, Hong Kong, CHINA) or a contracted testing facility.
  • Retrospective or Prospective: Not applicable. These are laboratory/bench test results of physical properties and performance, not patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not applicable. This device does not generate diagnostic interpretations or clinical assessments that would require human expert ground truth establishment. The "ground truth" for this device's performance is established by meeting engineering specifications through bench testing and validated laboratory methods (e.g., flow rate measurements, sterility testing, chemical analysis for residuals, biocompatibility assays).

4. Adjudication Method for the Test Set

  • Not applicable. As no expert interpretation or human assessment of results requiring consensus was involved, no adjudication method was used.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC study is designed to evaluate the impact of a diagnostic AI system on human reader performance. This device is an irrigation tubing, not a diagnostic imaging or AI system.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. This device is hardware and does not contain an algorithm.

7. The Type of Ground Truth Used

  • Functional/Performance Specifications & Regulatory Standards: The "ground truth" here is adherence to engineering design specifications, validated sterilization parameters (SAL 10⁻⁶, EO residuals), biocompatibility standards (ISO 10993-1), and demonstrated performance equivalence (flow rate, shelf life) to the predicate device. This is based on objective measurements rather than clinical outcomes or expert review of images/data.

8. The Sample Size for the Training Set

  • Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no training set for an AI/ML model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 10, 2022

GA Health Company Limited Lam Yu Assistant Regulatory Affairs Manager Unit 18, 21/F, Metropole Square, 2 On Yiu Street, Shatin Hong Kong, CHINA

Re: K223171

Trade/Device Name: AquaPulse® Irrigation Tubing Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OCX Dated: October 7, 2022 Received: October 11, 2022

Dear Lam Yu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223171

Device Name AquaPulse® Irrigation Tubing

Indications for Use (Describe)

The 24 hour use AquaPulse® Irrigation Tubing (tubing and accessories to accommodate various GI endoscopes and irrigation pumps) is intended to provide irrigation via a sterile water bottle during gastrontestinal endoscopic procedures when used in conjunction with an irrigation pump.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K223171 Page 1 of 4

510(k) Summary

The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.

1. Submission Sponsor

Submitter's Name: Submitter's Address:

Establishment Registration No.:

  1. Sponsor Contact

Cindy Ye Chief Executive Officer Telephone: +852 2833 9010 Email: cindy.ye@gahealth.com

GA Health Company Limited Unit 18, 21/F, Metropole Square

2 On Yiu Street, Shatin, N.T,

Hong Kong. CHINA

3014749926

Lam Wing Yu Senior Requlatory Affairs Specialist Telephone: +852 2833 9010 Email: rainy.lam@gahealth.com

3. Date Prepared

30 September 2022

4. Device Identification

  • Device Name: Common Name: Classification Number: Classification Name: Product Code: Product Code Name: Regulatory Class: Device Panel:
    AquaPulse® Irrigation Tubing Irrigation tubing for endoscope 21 CFR 876.1500 Endoscope and accessories ОСХ Endoscopic Irrigation/Suction System 2 Gastroenterology/Urology

5. Predicate Device Identification

Predicate Device 510(k) No.: Predicate Device Trade Name: Predicate Device Product Code: K202560 AquaPulse® Irrigation Tubing ОСХ

6. Device Description:

The AquaPulse® Irrigation Tubing is intended for 24-hour use and then discarded. The irrigation tubing is manufactured for use in conjunction with sterile water bottle, and together with irrigation pumps. The irrigation tubing is individually packed in sealed package, sold as a sterile device. The irrigation tubing is designed to the auxiliary water connector or biopsy irrigation accessory and to be inserted into pump head of the irrigation pump to provide irrigation through the auxiliary water channel to the distal end of endoscope.

{4}------------------------------------------------

7. Indications for Use:

The 24 hour use AquaPulse® Irrigation Tubing (tubing and accessories to accommodate various GI endoscopes and irrigation pumps) is intended to provide irrigation via a sterile water bottle during gastrointestinal endoscopic procedures when used in conjunction with an irrigation pump.

8. Technological Characteristics

Table 2 summarizes a comparison of the device features, indications for use, and other information necessary to demonstrate that the modified device is substantially equivalent to the predicate device. The difference does not affect the substantial equivalence of the device.

The predicate and subject devices are both irrigation tubing with same indications for use. There are some minor differences in the technological characteristics. These differences are material for tubing and shelf-life.

Table 2 Summary of design, features and principles of operation and technological characteristics between the subject device and predicate devices

SpecificationPredicate DeviceSubject DeviceSubstantialEquivalence
Product codeOCXOCXIdentical
RegulatoryClassification22Identical
Regulation No21 CFR 876.150021 CFR 876.1500Identical
RegulationDescriptionEndoscope and accessoriesEndoscope and accessoriesIdentical
CompatibilityOlympus® OFP, Endo Status™EGA-500 and Endogator® EGP-100 Irrigation PumpOlympus® OFP, Endo Status™EGA-500 and Endogator® EGP-100 Irrigation PumpIdentical
Intended UseThe 24 hour use AquaPulse®Irrigation Tubing (tubing andaccessories to accommodatevarious GI endoscopes andirrigation pumps) is intended toprovide irrigation via a sterilewaterbottleduringgastrointestinalendoscopicprocedures when used inconjunctionwith an irrigationpump.The 24 hour use AquaPulse®Irrigation Tubing (tubing andaccessories to accommodatevarious GI endoscopes andirrigation pumps) is intended toprovide irrigation via a sterilewaterbottleduringgastrointestinalendoscopicprocedures when used inconjunctionwith an irrigationpump.Identical
Environmentof UseHospital and or clinicsHospital and or clinicsIdentical
Single Use,Disposable24-hour use then discard24-hour use then discardIdentical
MaterialPolycarbonate and PolyvinylChloridePolycarbonate and PolyvinylChloride (Material not Made withDEHP)SubstantialEquivalent

{5}------------------------------------------------

SpecificationPredicate DeviceSubject DeviceSubstantialEquivalence
PackagingIrrigation tubing is packaged in asealed pouchIrrigation tubing is packaged in asealed pouchIdentical
ManufacturingmethodInjection molding and extrusionInjection molding and extrusionIdentical
BackflowPreventionDesignDiaphragm in the connectorallow pressurized water passingthrough in one-way but not theother wayDiaphragm in the connectorallow pressurized water passingthrough in one-way but not theother wayIdentical
SterilizationEO gasEO gasIdentical
Shelf Life1 year3 yearsSubstantialEquivalent

9. Performance Test

The bench testing was performed to support substantial equivalence such as verification tests on flow performances of modified devices. The performance data demonstrated that the subject devices met established specifications and it demonstrated substantial equivalence to the previously cleared predicate devices.

10. Sterilization

All the subject devices are sold in a sterile package. The subject devices have been sterilized in a validated EO sterilization cycle. The EO sterilization cycle has a Sterility Assurance Level (SAL) of 10°. EO residuals on the components are below the maximum levels defined in ANSI/AAMI/ISO 10993-7 Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide sterilization residuals.

11. Shelf Life

The modified devices have a three (3) year shelf life, based on the design and existing sterile barrier data from the existing packaging integrity test and performance test were conducted after accelerated aging test. The test result can imply that the modified device can provide and maintain a sterile barrier and its intended performance for at least the claimed shelf life.

12. Biocompatibility

The biocompatibility of the modified devices was conducted in accordance with the FDA quideline "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The modified devices are classified as surface devices with mucosal membrane contact for a limited duration (not more than 24 hours). The test result shows that the modified devices are biocompatible.

13. Clinical Testing

No clinical data is being submitted to support this Special 510(k).

14. Conclusion

The subject devices have the same intended use as the predicate devices. Based on the technological characteristics and overall performance of the devices in bench testing and

{6}------------------------------------------------

biocompatibility test, there are no significant differences exist between the subject devices and the predicate devices. The subject device is as safe and as effective as the predicate device. From a clinical perspective and comparing design specifications, the subject devices and the predicate device are substantially equivalent.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.