K Number
K223080

Validate with FDA (Live)

Date Cleared
2022-11-22

(53 days)

Product Code
Regulation Number
876.1500
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. Itis intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. Itis intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Device Description

The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are software-controlled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

AI/ML Overview

The provided text describes a 510(k) submission for a labeling modification to the da Vinci Xi and X Surgical Systems, not a study to prove a device meets acceptance criteria regarding active performance metrics. The submission asserts substantial equivalence based on the current device's performance being equivalent to predicate devices for new labeled uses.

Here's an analysis of the provided information, focusing on how the request's points relate to the document:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state acceptance criteria in the typical sense of numerical thresholds for device performance. Instead, it aims to demonstrate substantial equivalence for new labeled uses by comparing the real-world performance of the da Vinci systems (robotic-assisted, "R") with traditional laparoscopic procedures ("L") for duodenal switch bariatric surgeries. This comparison functions as the "performance" data.

Table of Device Performance (from the document, Table 1):

OutcomesL (Unmatched)R (Unmatched)p-value (Unmatched)L (PS Matched)R (PS Matched)p-value (PS Matched)
1. Operative Time (min), median (IQR)180 (150, 240)284 (240, 345)<.001210 (165, 270)270 (240, 330)<.001¶
2. Length of Hospital Stay (days), median (IQR)2 (1, 3)2 (1, 3)0.5182 (1, 3)2 (1, 3)0.372§§
3. Conversion to Open Surgery, n (%)5 (0.3)0 (0.0)0.5901 (0.3)0 (0.0)1.000§§
4. Transfusion Rate, n (%)9 (0.6)4 (1.1)0.3083 (1.0)3 (1.0)1.000
5. Intraoperative Complication Rate, n (%)2 (0.1)1 (0.3)0.4901 (0.3)1 (0.3)1.000
6. 30 Day Post-Operative Complication Rate, n (%)309 (20.6)86 (22.8)0.34385 (27.1)67 (21.3)0.094
7. 30 Day Readmission84 (5.6)14 (3.7)0.14228 (8.9)10 (3.2)0.003
8. 30 Day Reoperation30 (2.0)7 (1.9)0.85914 (4.5)5 (1.6)0.036
9. 30 Day Mortality2 (0.1)0 (0.0)1.0001 (0.3)0 (0.0)1.000

Acceptance Criteria Interpretation: The "acceptance criteria" here are implied by the substantial equivalence argument: the da Vinci system performance for these new indications should be comparable to or better than traditional laparoscopic methods, without raising new questions of safety or effectiveness. The p-values are used to assess statistical differences between the robotic and laparoscopic cohorts. While the robotic operative time is significantly longer (p < 0.001), some key safety outcomes (conversion to open, transfusion, intraoperative complications, and mortality) show no significant difference or favorable trends for robotic assistance (e.g., lower reoperation rate and readmission rate in the PS matched data). The submission concludes that the RWE demonstrates the safety and effectiveness of robotic-assisted duodenal switch bariatric surgical procedures and that da Vinci-assisted procedures are substantially equivalent to laparoscopic procedures based on these outcomes.

2. Sample size used for the test set and the data provenance

  • Sample Size (Test Set):
    • Unmatched: 1,878 patients in total (377 robotic-assisted, 1501 laparoscopic).
    • Propensity Score Matched: 628 patients (314 robotic-assisted, 314 laparoscopic).
  • Data Provenance:
    • Country of Origin: USA.
    • Nature of Data: Retrospective Real World Evidence (RWE) from the Premier Health Database (PHD). The data covers the period from 2016 to 2020.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not describe a process of individual expert ground truth establishment for each case. Instead, the "ground truth" is derived from real-world clinical outcomes recorded in the Premier Health Database using ICD-10 and CPT codes. This implicitly relies on the medical professionals (e.g., surgeons, diagnosticians, hospital staff) who documented these codes and outcomes in clinical practice, rather than a separate panel of experts being convened to review the data for the purpose of this study.

4. Adjudication method for the test set

There is no mention of an adjudication method in the context of expert review for ground truth, as the data provenance is real-world clinical records rather than expert interpretation of medical images or other subjective assessments. The "adjudication" in this context is the inherent process of medical documentation and coding within the healthcare system.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This study compares the outcomes of robotic-assisted surgery (where the surgeon uses the da Vinci system) versus traditional laparoscopic surgery. It is not comparing human readers' performance with and without AI assistance in reading or diagnosis. The da Vinci system is a surgical tool, not an AI diagnostic assistant. Therefore, there's no "effect size of how much human readers improve with AI vs without AI assistance" to report here.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

No, a standalone (algorithm only) performance was not done. The da Vinci system is an "Endoscopic Instrument Control System" that "is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments." It is fundamentally a human-in-the-loop system, where the surgeon controls the robotic arms. The study evaluates the outcomes of surgery performed by surgeons with the assistance of the da Vinci system, not an autonomous algorithm.

7. The type of ground truth used

The ground truth is based on outcomes data and clinical diagnoses/procedures (documented via ICD-10 and CPT codes) from real-world patient records in the Premier Health Database. These outcomes include operative times, length of hospital stays, conversion rates, various complication rates, readmission rates, reoperation rates, and mortality rates.

8. The sample size for the training set

The document does not mention a separate "training set" in the context of an AI/machine learning model being developed or trained. This submission is for a labeling modification of an existing surgical system, not the initial clearance of a new AI algorithm. The real-world evidence presented is used to support the safety and effectiveness of the existing device for new labeled uses, but not to train a model in the typical sense.

9. How the ground truth for the training set was established

As there is no mention of a "training set" for an AI model, this question is not applicable to the provided document. The real-world data itself serves as the basis for demonstrating equivalence for the new indications, not as data for training an algorithm where ground truth needs to be established by experts.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it. The full name reads "U.S. Food & Drug Administration".

November 22, 2022

Intuitive Surgical, Inc Connor McCarty Senior Regulatory Engineer 1266 Kifer Road Sunnyvale, California 94086

Re: K223080

Trade/Device Name: da Vinci X Surgical System (IS4200), da Vinci Xi Surgical System (IS4000) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: NAY Dated: September 30, 2022 Received: September 30, 2022

Dear Connor McCarty:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2022.11.22
13:42:23 -05'00'

Mark Trumbore Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223080

Device Name

Intuitive Surgical® da Vinci® X Endoscopic Instrument Control System (da Vinci X System, Model IS4200) and Endoscopic Instruments and Accessories

Indications for Use (Describe)

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. Itis intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K223080

Device Name

Intuitive Surgical® da Vinci® Xi Endoscopic Instrument Control System (da Vinci Xi System, Model IS4000) and Endoscopic Instruments and Accessories

Indications for Use (Describe)

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. Itis intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary (21 CFR § 807.92(c))

I. SUBMITTER INFORMATION

Submitter:Intuitive Surgical, Inc.
1266 Kifer Road
Sunnyvale, CA 94086
Contact:Connor McCarty
Senior Regulatory Engineer
Connor.McCarty@intusurg.com
805.798.4205
Date Summary Prepared:September 30, 2022
II. SUBJECT DEVICE INFORMATION
Device Trade Name:da Vinci® Xi and X Surgical Systems, Model IS4000 and Model IS4200
Common Name:System, Surgical, Computer Controlled Instrument
Classification Name:Endoscope and Accessories (21 CFR §876.1500)
Regulatory Class:II
Product Code:NAY

III. PREDICATE DEVICE INFORMATION:

Predicate Devices:Intuitive Surgical da Vinci Xi and X Surgical Systems, Models IS4000 and
IS4200 (K131861, K152578, K153276, K161178, K170713, K171632, K171294
K172643, K173842, K173585, K182140, K183086, K202834, K211784)
Intuitive Surgical da Vinci Si Surgical System, Model IS3000 (K081137,
K123463, K090993)

IV. DEVICE DESCRIPTION:

Submission Type:

This 510(k) is for a labeling modification only, to include the following additional representative, specific "duodenal switch bariatric surgical procedures" under the cleared "general laparoscopic surgical procedures" Indications for Use of the da Vinci Xi Surgical System, Model IS4000 (K131861) and the da Vinci X Surgical System, Model IS4200 (K171294):

    1. Biliopancreatic Diversion-Duodenal-Switch/BPD-DS,
    1. Single Anastomosis-Duodeno Ileal Bypass with Sleeve/SADI-S,

Traditional 510(k)

    1. One Anastomosis Duodenal Switch/OADS, and
    1. Transit Bipartition/TB.

There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are software-controlled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

{5}------------------------------------------------

V. INDICATIONS FOR USE

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Models: IS4000 and IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Precaution for Representative Uses

The demonstration of safety and effectiveness for the representative specific procedures did not include evaluation of outcomes related to the treatment of cancer (overall survival, disease-free survival, local recurrence) or treatment of the patient's underlying disease/condition. Device usage in all surgical procedures should be guided by the clinical judgment of an adequately trained surgeon.

VI. COMPARISON OF INTENDED USE, INDICATIONS FOR USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

There are no changes to the technological characteristics for the subject devices compared to the cleared predicate devices, da Vinci Xi Surgical System, Model IS4000 (K131861) and the da Vinci X Surgical System, Model IS4200 (K171294). This 510(k) is for a labeling modification to include the representative, specific "duodenal switch bariatric surgical procedures" as a labeled use under the cleared "general laparoscopic surgical procedures" Indications for Use of the cleared predicate devices, da Vinci Xi Surgical System, Model IS4000 and the da Vinci X Surgical System, Model IS4200. The subject devices differ from the predicate devices by this modification to the labeling. Results of clinical data from the literature demonstrated that the subject devices have intended use as the predicate devices.

VII. PERFORMANCE DATA

There were no technological changes to the subject devices, thus no bench testing, electromagnetic compatibility testing, sterilization testing or biocompatibility testing was required.

Real World Evidence (RWE) from the Premier Health Database (PHD)

Real world evidence (RWE) from the Premier Health Database (PHD) support use of the da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200) in "duodenal switch bariatric surgical procedures" that fall under the cleared "general laparoscopic surgical procedures" Indication for Use. Real-world data (RWD) were not provided for all of the representative, specific procedures. Instead, RWD were provided only for the most complex/highest risk representative, duodenal switch bariatric surgical procedures (referred to as the "umbrella" procedures). The RWE for these "umbrella" procedures (i.e.,

{6}------------------------------------------------

Billopancreatic Diversion-Duodenal-Switch/BPD-DS, Single Anastomosis-Duodeno Ileal Bypass with Sleeve/SADI-S, and, One Anastomosis Duodenal Switch/OADS) were deemed sufficient to cover the less complex/lower risk procedure of "Transit Bipartition" (referred to as "covered" procedure), so RWD on the noted covered procedure was not provided.

Using ICD-10 and CPT codes, 1,878 patients, including 377 robotic-assisted and 1501 laparoscopic, were identified in PHD that underwent primary DS Bariatric surgical procedures for an obesity indication at 73 hospitals in the USA from 2016 to 2020. After propensity score matching, 314 patients were evaluated in each cohort. Table 1 includes a summary of the unmatched and propensity matched comparative data from the robotic-assisted and laparoscopic duodenal switch bariatric surgical procedure cohorts. Outcomes were assessed through 30 days. The results demonstrated the safety and effectiveness of robotic-assisted duodenal switch bariatric surgical procedures. The findings from this analysis demonstrate that da Vinci-assisted procedures as compared to laparoscopic procedures are substantially equivalent based on the following nine (9) outcomes of interest:

    1. Operative Times
    1. Lengths of Hospital Stay (LOS)
    1. Conversion Rates
    1. Transfusion Rates
    1. Intraoperative Complication Rates
    1. 30 Day Post-operative Complication Rates
    1. 30 Day Readmission Rates
    1. 30 Day Reoperation Rates
    1. 30 Day Mortality Rates

VIII. CONCLUSION

The do Vinci Xi and X Surgical Systems (models IS4000 and IS4200) have the same intended use as the predicate devices, as demonstrated by the RWE to support the safety and effectiveness for the new labeled use of the representative, specific duodenal switch surgical procedures of: Billopancreatic Diversion-Duodenal-Switch/BPD-DS, Single Anastomosis-Duodeno Ileal Bypass with Sleeve/SADI-S, One Anastomosis Duodenal Switch/OADS, and Transit Bipartition under the "general laparoscopic surgical procedure" Indications for Use as compared to the predicate devices. In addition, the subject devices have the same technological characteristics as the predicate devices. Therefore, the da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200) are substantially equivalent to the cleared predicate devices.

{7}------------------------------------------------

K223080

UnmatchedPS Matched
OutcomesLRp-valueLRp-value
1.Operative Time (min),median (IQR)180(150, 240)284(240, 345)<.001210(165, 270)270(240, 330)<.001¶
2. Length of Hospital Stay(days), median (IQR)2 (1, 3)2 (1, 3)0.5182 (1, 3)2 (1, 3)0.372§§
3.Conversion to OpenSurgery, n (%)5 (0.3)0 (0.0)0.5901 (0.3)0 (0.0)1.000§§
4.Transfusion Rate, n (%)9 (0.6)4 (1.1)0.3083 (1.0)3 (1.0)1.000
5.IntraoperativeComplication Rate, n (%)2 (0.1)1 (0.3)0.4901 (0.3)1 (0.3)1.000
6.30 Day Post-OperativeComplication Rate, n (%)309 (20.6)86 (22.8)0.34385 (27.1)67 (21.3)0.094
7.30 Day Readmission84 (5.6)14 (3.7)0.14228 (8.9)10 (3.2)0.003
8.30 Day Reoperation30 (2.0)7 (1.9)0.85914 (4.5)5 (1.6)0.036
9.30 Day Mortality2 (0.1)0 (0.0)1.0001 (0.3)0 (0.0)1.000

TABLE 1: Unmatched and Propensity Score Matched da Vinci and Laparosocopic Duodenal Switch Bariatric Surgical Procedures

§§ No adjustment to the p-value according to the step-down approach described in section 5.2.1. We started with the p-value for OR time first and then LOS, p-value of LOS greater that the significance level was observed therefore the adjustment method stopped.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.