(210 days)
The Access PCT assay is a paramagnetic, chemiluminescent immunoassay for in vitro quantitative determination of procalcitonin (PCT) levels in human serum and plasma (lithium heparin and EDTA) using the Access Immunoassay Systems. Measurement of PCT in conjunction with other laboratory findings and clinical assessments aids in the risk assessment of critically ill patients on their first day of Intensive Care Unit (ICU) admission for progression to severe sepsis and septic shock.
The Access PCT assay is a paramaqnetic, chemiluminescent immunoassay for in vitro quantitative determination of procalcitonin (PCT) levels in human serum and plasma (lithium heparin and EDTA) using the Access Immunoassay Systems. Measurement of PCT in conjunction with other laboratory findings and clinical assessments aids in the risk assessment of critically ill patients on their first day of Intensive Care Unit (ICU) admission for progression to severe sepsis and septic shock. A description of the reagent pack is provided below.
- R1a: Dynabeads* paramagnetic particles coated with mouse anti-. human Procalcitonin monoclonal antibody in a TRIS buffer with surfactant, protein (bovine), ≤ 0.1% sodium azide, and 0.1% ProClin**300
- R1b: 0.10 N Sodium Hydroxide ●
- R1c: MOPS Buffer with surfactant and protein (bovine, murine), ≤ ● 0.1 % sodium azide, and 0.1% ProClin 300
- R1d: Rat anti-Procalcitonin recombinant alkaline phosphatase . coniugate in a MOPS buffer with surfactant and protein (bovine, murine, recombinant).
≤ 0.1% sodium azide, and 0.1% ProClin 300
The device in question is the Access PCT Assay for the Dxl 9000 Access Immunoassay Analyzer, used for the in vitro quantitative determination of procalcitonin (PCT) levels. The study presented aims to demonstrate its substantial equivalence to the previously cleared Access PCT Assay on the Access 2 Immunoassay System (K192271).
Here's a breakdown of the acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance
| Test/Characteristic | Acceptance Criteria (New Device) | Reported Device Performance (New Device - Access PCT on Dxl 9000) |
|---|---|---|
| Method Comparison(vs. Predicate Access PCT on Access 2) | R² ≥ 0.95 | R² ≥ 0.95 (Met) |
| Slope | 1.00 ± 0.10 | 1.00 ± 0.10 (Met) |
| Method Concordance (at 0.5 ng/mL cutoff) | Not explicitly stated | 100.0% |
| Method Concordance (at 2.0 ng/mL cutoff) | Not explicitly stated | 98.3% |
| Imprecision (Within Laboratory) | SD ≤ 0.012 ng/mL for PCT < 0.150 ng/mL | SD between 0.006 - 0.008 ng/mL for PCT < 0.150 ng/mL (Met) |
| % CV ≤ 8.0% for PCT ≥ 0.150 ng/mL | % CV between 2.2% - 6.1% for PCT ≥ 0.150 ng/mL (Met) | |
| Imprecision (Within Run) | SD ≤ 0.009 ng/mL for PCT < 0.150 ng/mL | SD between 0.004 - 0.007 ng/mL for PCT < 0.150 ng/mL (Met) |
| % CV ≤ 6.0% for PCT ≥ 0.150 ng/mL | % CV between 1.9% - 4.7% for PCT ≥ 0.150 ng/mL (Met) | |
| Linearity | Detectable nonlinearity within ± 0.012 ng/mL for PCT ≤ 0.150 ng/mL | Met |
| Detectable nonlinearity within ± 10% for PCT > 0.150 ng/mL | Met | |
| Limit of Blank (LoB) | ≤ 0.005 ng/mL | 0.002 ng/mL (Met) |
| Limit of Detection (LoD) | ≤ 0.01 ng/mL | 0.003 ng/mL (Met) |
| Limit of Quantitation (LoQ) | ≤ 0.02 ng/mL (20% CV) | 0.002 ng/mL (reported as 0.003 ng/mL to align with LoD per CLSI EP17-A2 recommendation, Met) |
2. Sample Size and Data Provenance
The document does not explicitly state the specific sample sizes used for each test (e.g., method comparison, imprecision, linearity). It also does not specify the country of origin of the data or whether the data was retrospective or prospective. It states "The study" for imprecision and linearity, suggesting a single study per characteristic. The "Method Comparison" and "Method Concordance" imply patient samples were used to compare results between the new device and the predicate.
3. Number of Experts and Qualifications for Ground Truth
Not applicable. This device is an in vitro diagnostic immunoassay measuring procalcitonin levels. Ground truth is established through analytical validation and comparison to an established predicate device, not through expert review of interpretations.
4. Adjudication Method
Not applicable. As this is an analytical performance study for an immunoassay, adjudication methods typically used for qualitative or imaging-based diagnostics by human readers are not relevant.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI-powered diagnostic tools where human interpretation is involved. The Access PCT assay is an automated immunoassay.
6. Standalone Performance (Algorithm Only without Human-in-the-Loop)
Yes, the studies described represent the standalone performance of the Access PCT assay on the Dxl 9000 Analyzer. The tests (method comparison, imprecision, linearity, LoB, LoD, LoQ) evaluate the analytical performance of the device itself, independent of human interpretation or interaction beyond standard laboratory procedures.
7. Type of Ground Truth Used
The ground truth or reference for comparison in this submission is the measurements obtained from the previously cleared Access PCT assay on the Access 2 Immunoassay System (K192271). For analytical precision, linearity, and limits of detection/quantitation, the "ground truth" is defined by established analytical methods and statistical calculations following guidelines like CLSI EP17-A2.
8. Sample Size for the Training Set
Not applicable. This device is an immunoassay, not an AI/Machine Learning algorithm that undergoes a "training" phase with a specific dataset. Its performance is based on the chemical and biological reactions designed within the assay.
9. How Ground Truth for the Training Set was Established
Not applicable, as there is no "training set" in the context of this immunoassay. The development of the assay involves optimization of reagents and protocols, but not in the sense of supervised learning from a labeled dataset. The reference for comparison is the predicate device's performance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
April 26, 2023
Beckman Coulter, Inc Neha Desai Staff Quality and Regulatory Affairs 1000 Lake Hazeltine Drive Chaska, Minnesota 55318
Re: K222996
Trade/Device Name: Access PCT Regulation Number: 21 CFR 866.3215 Regulation Name: Device To Detect And Measure Non-Microbial Analyte(s) In Human Clinical Specimens To Aid In Assessment Of Patients With Suspected Sepsis Regulatory Class: Class II Product Code: PTF Dated: September 27, 2022 Received: September 28, 2022
Dear Neha Desai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Noel J. Gerald -S
Noel J. Gerald, Ph.D. Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K222996
Device Name Access PCT
Indications for Use (Describe)
The Access PCT assay is a paramagnetic, chemiluminescent immunoassay for in vitro quantitative determination of procalcitonin (PCT) levels in human serum and plasma (lithium heparin and EDTA) using the Access Immunoassay Systems. Measurement of PCT in conjunction with other laboratory findings and clinical assessments aids in the risk assessment of critically ill patients on their first day of Intensive Care Unit (ICU) admission for progression to severe sepsis and septic shock.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circular shape with two curved lines inside, resembling a stylized wave or heartbeat. To the right of the circular shape, the words "BECKMAN" and "COULTER" are written in bold, black, sans-serif font, with "BECKMAN" on the top line and "COULTER" on the bottom line.
Access PCT 510(K) Summary
Immunodiagnostic Development Center
1000 Lake Hazeltine Drive Chaska, Minnesota 55318-1084
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a)(1).
The assigned 510(k) number is K222996
Submitted By:
Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318 Telephone: (952) 368-1142 Fax: (952) 368-7610
Contact Person:
Neha Desai 1000 Lake Hazeltine Drive Chaska, MN 55318 Telephone: (612) 244-9788 Email: nhdesai@beckman.com
Alternate Contact:
Kulieet Kaur Office Phone: (952) 465-1914 Email: kkaur@beckman.com
Date Prepared:
February 17th, 2023
Device Name:
Proprietary / Trade Name: Access PCT Reagent Common Name: Procalcitonin Immunoassay Classification Name: Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis Classification Regulation: 21 CFR 866.3215 Classification Product Code: PTF
{4}------------------------------------------------
Predicate Device:
The Access PCT Assay/Calibrators claims substantial equivalence to previously cleared Access PCT Assay FDA 510(k) Number K192271.
Device Description:
The Access PCT assay is a paramaqnetic, chemiluminescent immunoassay for in vitro quantitative determination of procalcitonin (PCT) levels in human serum and plasma (lithium heparin and EDTA) using the Access Immunoassay Systems. Measurement of PCT in conjunction with other laboratory findings and clinical assessments aids in the risk assessment of critically ill patients on their first day of Intensive Care Unit (ICU) admission for progression to severe
sepsis and septic shock.A description of the reagent pack is provided below.
- R1a: Dynabeads* paramagnetic particles coated with mouse anti-. human Procalcitonin monoclonal antibody in a TRIS buffer with surfactant, protein (bovine), ≤ 0.1% sodium azide, and 0.1% ProClin**300
- R1b: 0.10 N Sodium Hydroxide ●
- R1c: MOPS Buffer with surfactant and protein (bovine, murine), ≤ ● 0.1 % sodium azide, and 0.1% ProClin 300
- R1d: Rat anti-Procalcitonin recombinant alkaline phosphatase . coniugate in a MOPS buffer with surfactant and protein (bovine, murine, recombinant).
≤ 0.1% sodium azide, and 0.1% ProClin 300
*Dynabead® is a registered trademark of Dynal A.S., Oslo, Norway **ProClin™ is a trademark of The Dow Chemical Company ("Dow") or an affiliate company of Dow.
Intended Use:
The Access PCT assay is a paramagnetic, chemiluminescent immunoassay for in vitro quantitative determination of procalcitonin (PCT) levels in human serum and plasma (lithium heparin and EDTA) using the Access Immunoassay Systems. Measurement of PCT in conjunction with other laboratory findings and clinical assessments aids in the risk assessment of critically ill patients on their first day of Intensive Care Unit (ICU) admission for progression to severe sepsis and septic shock.
{5}------------------------------------------------
Comparison of Technological Characteristics to the Predicate (Assay)
| Predicate Access PCT on | Access PCT on Dxl 9000 | |
|---|---|---|
| System | Access 2 Immunoassay | Access Immunoassay |
| Attribute/Characteristic | System (K192271) | Analyzer |
| Intended Use/Indications for Use | The Access PCT assay is aparamagnetic particle,chemiluminescentimmunoassay for in vitroquantitative determination ofprocalcitonin (PCT) levels inhuman serum and plasma(lithium heparin and EDTA)using the AccessImmunoassay System.Measurement of PCT inconjunction with otherlaboratory findings and clinicalassessment aids in the riskassessment of critically illpatients on their first day ofIntensive Care Unit (ICU)admission for progression tosevere sepsis and septicshock. | Same |
| Analyte Measured | Procalcitonin (PCT) | Same |
| Technology | Two-step sandwich | Same |
| Format | Chemiluminescent | Same |
| Method | Automated | Same |
| Calibration | Utilizes a stored calibrationcurve | Same |
| Sample Type | Human Serum or Plasma(LiHep and EDTA) | Same |
| Measuring Range | 0.05 to 100 ng/mL | Same |
| LoB | ≤ 0.005 ng/mL | Same |
| LoD | ≤ 0.01 ng/mL | Same |
| LoQ | 20% CV at ≤ 0.02 ng/mL | Same |
| Hook | No hook effect up toprocalcitonin concentrations of5,000 ng/mL | Same |
| Stability | Stable at 2 to 10°C for 42 daysafter initial use | Same |
| Reagent Pack materials | Mouse anti-humanprocalcitonin monoclonalantibody | Same |
| Assay Duration | Approximately 20 minutes | Approximately 14 minutes |
| SystemAttribute/Characteristic | Predicate Access PCT onAccess 2 ImmunoassaySystem (K192271) | Access PCT on Dxl 9000Access ImmunoassayAnalyzer |
| Sample Volume | 35 µL | 15 µL |
| Instrument | Access 2 Immunoassaysystem | Dxl 9000 AccessImmunoassay Analyzer |
| Substrate | Access Substrate | Lumi-Phos PRO substrate |
{6}------------------------------------------------
{7}------------------------------------------------
Summary of Studies
Method Comparison: The results of the method comparison study met the acceptance criteria of R2 ≥ 0.95 and slope 1.00 ± 0.10 and supports the equivalence of the Access PCT assay on Dxl 9000 to the Access PCT assay on Access 2 Instrument. The bias data supports the reference intervals defined on the Access 2 system have not changed appreciably on the Dxl 9000 system.
Method Concordance: The percentage of concordance between the Access 2 lmmunoassay System and the Dxl 9000 Access Immunoassay Analyzer for the cut-offs of 0.5 and 2.0 ng/mL were 100.0% and 98.3%, respectively.
lmprecision: The Within Laboratory (total) % CV was between 2.2% and 6.1% for PCT concentrations ≥ 0.150 ng/mL. The Within laboratory (total) SD was between 0.006 - 0.008 for PCT concentrations < 0.150 ng/mL. The study met the acceptance criteria of SD ≤ 0.012 ng/mL for values < 0.150 nq/mL and CV ≤ 8.0% for values ≥ 0.150 nq/mL.
The Within Run (total) % CV was between 1.9% and 4.7% for PCT concentrations ≥ 0.150 ng/mL. The Within Run (total) SD was between 0.004 -0.007 for PCT concentrations < 0.150 ng/mL. The study met the acceptance criteria of SD ≤ 0.009 ng/mL for values < 0.150 ng/mL and CV ≤ 6.0% for values ≥ 0.150 nq/mL.
Linearity: The study shows that acceptance criteria was met for detectable nonlinearity within ± 0.012 ng/mL for values ≤ 0.150 ng/mL and ± 10% for values > 0.150 ng/mL.
Limit of Blank (LoB): The data demonstrated the LoB estimate of the PCT assay is 0.002 ng/mL. which met the acceptance criteria of ≤ 0.005 ng/mL.
Limit of Detection (LoD): The data demonstrated the LoD estimate of PCT is 0.003 ng/mL which met the acceptance criteria of ≤ 0.01 ng/mL.
Limit of Quantitation (LoQ): The results demonstrated the 20% CV LoQ estimate for the Access PCT assay to be 0.002 ng/mL which meets the acceptance criteria of less than or equal to 0.02 ng/mL
The maximum observed LoQ estimate for Access PCT assay on the Dxl 9000 immunoassay System is less than the reported LoD value (0.003ng/mL). Following the CLSI EP17-A2 recommendation that the LoQ must be greater than or equal to LoD, the LoQ value is reported as 0.003 ng/mL to align with LoD.
{8}------------------------------------------------
Conclusion:
The information provided in this submission supports a substantial equivalence determination, and therefore 510(k) premarket notification clearance of the PCT assay on Dxl 9000 Access Immunoassay Analyzer.
§ 866.3215 Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis.
(a)
Identification. A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended.
(2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection.
(3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability.
(4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(i) Results must demonstrate adequate device performance relative to a well-accepted comparator.
(ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population.
(iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses.
(5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
e.g., age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples (
e.g., collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.