(224 days)
Akunah Reflect Planning Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The software is used to assist in the positioning of shoulder components. Akunah Reflect Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. The software leads to the generation of a planning report. The software is to be used for adult patients only and should not be used for diagnostic purposes. Akunah Reflect should be used in conjunction with expert clinical judgement.
Akunah REFLECT Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The user interface software is connected to an online management system (OMS), and is intended to be used by orthopaedic surgeons as a preoperative planning software for shoulder surgery. The software is intended to allow surgeons to:
- Visualize, measure, reconstruct, and annotate anatomic data,
- Select and position shoulder implant devices,
- . Output pre-surgical planning reports
The software is to be used for adult patients only and should not be used for diagnostic purpose.
The acceptance criteria and study proving the device meets those criteria are described below.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| All functional requirements are met across all systems and modules. | Akunah REFLECT met the acceptance criteria stated for all functional verification testing across all systems and modules. |
| Geometric accuracy of virtual models created by the device. | Both Akunah Auto-segmentation and Premorbid Reconstruction demonstrated equivalent accuracy as established by the predicate devices. |
| Accuracy of measurement tools. | Measurement Tools were verified to achieve an acceptable accuracy. |
| Overall validation of the software against user needs (usability). | Usability testing demonstrated the validation of the software on the Desktop for surgical planning. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions that "Software verification and validation testing were performed."
3. Number of Experts and their Qualifications for Ground Truth
This information is not provided in the document.
4. Adjudication Method for the Test Set
This information is not provided in the document.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done to assess improvement with AI vs. without AI assistance. The study focuses on verifying the device's functional requirements, geometric accuracy, measurement tool accuracy, and usability.
6. Standalone Performance
The document primarily describes a standalone performance evaluation of the Akunah REFLECT software. It highlights that the software itself met its functional requirements, demonstrated geometric accuracy equivalent to predicate devices, and that its measurement tools achieved acceptable accuracy. Usability testing also validated the software's performance on its own for surgical planning. There is no mention of human-in-the-loop performance being evaluated in this section.
7. Type of Ground Truth Used
The document does not explicitly state the type of ground truth used for specific aspects like geometric accuracy or measurement tools. However, for geometric accuracy, it implicitly refers to "accuracy as established by the predicate devices," suggesting a comparison against a known standard or established method from those predicate devices.
8. Sample Size for the Training Set
The document does not provide information about the sample size used for any training set.
9. How Ground Truth for the Training Set Was Established
The document does not provide information about how ground truth was established for any training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Akunah PTY LTD Shanthan Pather Coo 131 Warriewood St QUEENSLAND, 4151 AUSTRALIA
Re: K222987
May 10, 2023
Trade/Device Name: Akunah REFLECT Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH, LLZ Dated: April 6, 2023 Received: April 6, 2023
Dear Shanthan Pather:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K22987
Device Name Akunah REFLECT
Indications for Use (Describe)
Akunah Reflect Planning Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The software is used to assist in the positioning of shoulder components. Akunah Reflect Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. The software leads to the generation of a planning report. The software is to be used for adult patients only and should not be used for diagnostic purposes. Akunah Reflect should be used in conjunction with expert clinical judgement.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Akunah. The logo consists of a circular design made up of many small orange circles. To the right of the circular design is the word "akunah" in orange lowercase letters.
510(k) Summary
K222987
Traditional 510(k) – Akunah REFLECT
May 10th 2023
| Device Name | |
|---|---|
| Trade name: | Akunah REFLECT |
| Common Name: | Image processing system |
| Classification Name: | Medical image management and processing system |
| Regulation Number: | §892.2050 |
| Submitter | |
| Name: | Akunah Pty Ltd |
| Address: | Suite 307, Nicholson Street Specialist CentreLevel 9, 121 Newdegate St, Greenslopes 4120Brisbane, Australia |
| Company Contact | |
| Contact Name: | Akunah Pty Ltd |
| Contact Person: | Dr. Shanthan Pather |
| Title: | COO, Design & Regulatory Affairs Engineer |
| Address: | Suite 307, Nicholson Street Specialist CentreLevel 9, 121 Newdegate St, Greenslopes 4120Brisbane, Australia |
| Tel: | +61 3445 1591 |
| Fax: | +61 3445 1592 |
| Email: | sha@akunah.com |
| Classification | |
| Device | Class: Class II |
| Classification Panel: | Radiology |
| Product Code: | QIH, LLZ |
21 CFR section 892.2050, Class II Radiology Devices – Panel: Radiology Office of Radiological Health (OHT8)
Equivalent / Predicate Device
| Trade Name | 510(k) number | Decision Date | Applicant |
|---|---|---|---|
| Mimics Medical (PRIMARY PREDICATE) | K183105 | March 27, 2019 | Materialise NV |
| BLUEPRINT™ Patient SpecificInstrumentation (REFERENCE) | K203315 | April 15, 2021 | Tornier SAS |
| Materialise Shoulder System,Materialise Shoulder Guide andModels, SurgiCase Shoulder Planner(REFERENCE) | K193560 | March 20, 2020 | Materialise NV |
Suite 307 Nicholson Street Specialist Centre Level 9, 121 Newdegate Street Greenslopes Private Hospital, QLD, 4127 AUSTRALIA
E: info@akunah.com P: +61 7 3445 1591 F: +61 7 3445 1592
akunah.com ABN 58 646 020 322
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Akunah. On the left side of the logo is a circular design made up of many small circles. The circles are arranged in a spiral pattern, with the circles getting smaller as they move towards the center. To the right of the circular design is the word "akunah" in lowercase letters. The word is written in a sans-serif font and is colored in a gradient of orange and yellow.
Device Description
Akunah REFLECT Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The user interface software is connected to an online management system (OMS), and is intended to be used by orthopaedic surgeons as a preoperative planning software for shoulder surgery. The software is intended to allow surgeons to:
- Visualize, measure, reconstruct, and annotate anatomic data,
- Select and position shoulder implant devices,
- . Output pre-surgical planning reports
The software is to be used for adult patients only and should not be used for diagnostic purpose.
Intended Use
Akunah REFLECT Software is intended to be used as a medical software to assist in pre-operative surgical planning for shoulder surgery.
Indication For Use
Akunah Reflect Planning Software is intended to be used as a pre-surgical planner for shoulder arthroplasty, fracture, deformity correction and stabilisation procedures. The software is used to assist in the positioning of shoulder components. Akunah Reflect Planning Software allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. The software leads to the generation of a planning report. The software is to be used for adult patients only and should not be used for diagnostic purposes. Akunah Reflect should be used in conjunction with expert clinical judgement.
Comparison to Predicate Device
Akunah Reflect has been compared to legally marketed predicate device Materialise – Mimics Medical (K183105), and references devices:- Tornier - BLUEPRINT Patient Specific Instrumentation (K203315), and Materialise - Surgicase Shoulder Planner (K212569). The predicate device shares the same intended use for the software, specific to treatment planning for surgical procedures, allowing visualisation, reconstruction, measurements, annotations and editing of anatomic data. To achieve these features Akunah REFLECT employs similar fundamental technologies as the predicates.
The following technological differences exists between the subject device and the predicates:
- Akunah REFLECT does not offer a hardware component – Patient Specific Guide (PSI) or Anatomic Models, as it is a software-only device
- . Akunah REFLECT possess a unique user interface to optimise workflow and user experience.
The differences outlined above do not raise new questions of safety and effectiveness over the predicate device as demonstrated in verification and validation testing.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Akunah. The logo consists of two parts: a circular pattern of orange dots on the left and the word "akunah" in orange on the right. The dots are arranged in concentric circles, creating a visually appealing design. The word "akunah" is written in a sans-serif font, with a gradient of orange color that transitions from a darker shade to a lighter shade.
Performance Data
Software verification and validation testing were performed, and documentation provided as per the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (2005). Verification testing was performed to address all functional requirements, and usability testing was performed to demonstrate overall validation of the software against the user needs.
- Akunah REFLECT met the acceptance criteria stated for all functional verification testing across all systems and modules.
- . Geometric accuracy of the virtual models created in the subject device were assessed. Both Akunah Auto-segmentation and Premorbid Reconstruction demonstrated equivalent accuracy as established by the predicate devices.
- . Measurement Tools were verified to achieve an acceptable accuracy
- . Usability testing demonstrated the validation of the software on the Desktop for surgical planning
In conclusion, all performance testing conducted demonstrated device performance and substantial equivalence to the predicate device.
Substantial Equivalence Conclusion
The results of the verification and validation testing raise no concerns regarding the safety and effectiveness of the product. Minor differences in the non-fundamental technological characteristics raise no question of safety and effectiveness of the software. Performance data support the demonstration of substantial equivalence and usability. Therefore, the intended use, fundamental technological characteristics, and principle of operation of the Akunah REFLECT software are substantially equivalent to the predicate device.
N/A