K Number
K222888

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-08-11

(322 days)

Product Code
Regulation Number
862.1358
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BlueStar® CGM insulin dose calculator is software intended for the management of type 1 or type 2 diabetes in persons aged 18 years and older requiring fast-acting insulin. The BlueStar CGM insulin dose calculator allows patients to calculate a dose of bolus insulin for a given amount of carbohydrates, the most recent CGM glucose reading and rate of change, activity, and, optionally, insulin on board (IOB). The BlueStar CGM insulin dose calculator requires a prescription.

Device Description

The BlueStar® CGM insulin dose calculator is a software device module existing in the same mobile medical application as BlueStar® Rx (K203434), which is intended for the management of diabetes. When connected to a compatible integrated continuous glucose monitor (iCGM) and under authorization from a qualified healthcare provider, the BlueStar CGM insulin dose calculator allows patients to calculate a dose of bolus insulin for a given amount of carbohydrates, the most recent iCGM glucose reading and its rate of change, activity, and, optionally, insulin on board (IOB). Other patient-specific inputs from BlueStar Rx are used in the calculation of the recommended dose- specifically, duration of insulin to carb ratio, correction factor, and target glucose. In addition to calculating specific dosing recommendations, the BlueStar CGM insulin dose calculator also provides coaching messages to assist the user in maintaining glucose within the target range.

The use of CGM inputs differentiates the BlueStar CGM insulin dose calculator from the insulin dose calculator included in the previously cleared BlueStar Rx, which uses blood glucose (BG) values from a BG meter using a "fingerstick" method. The CGM insulin dose calculator is intended to coexist with the BG insulin calculator function in the BlueStar Rx software application as the BG calculation may be necessary when CGM is unavailable or the CGM estimated blood glucose does not match how the user feels.

AI/ML Overview

Here's the information about the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with specific numerical targets. However, the primary clinical endpoint serves as the de facto acceptance criterion for efficacy, while safety measures address other key concerns.

Acceptance Criteria (Inferred)Reported Device Performance
Efficacy: Mean CGM Time in Range (TIR) when using the BlueStar® CGM insulin dose calculator is not inferior to baseline TIR (percentage of time spent between 70 and 180 mg/dL).The statistical analysis of the study data showed that the mean TIR when using BlueStar® with the CGM insulin dose calculator is not inferior to the baseline mean TIR prior to using the device.
Safety: No increase in time spent in hypoglycemia (glucose below 70 mg/dL).There was no increase in time spent with glucose below 70 mg/dL.
Safety: No increase in time spent in severe hypoglycemia (glucose below 54 mg/dL).There was no increase in time spent with glucose below 54 mg/dL in the subject populations.
Safety: Acceptable adverse events profile, including hypoglycemic events.Adverse events including hypoglycemic events were recorded. (The document states they were recorded, and the conclusion mentions "These data support the safety of the Bluestar CGM insulin dose calculator," implying an acceptable safety profile, though specific numbers are not provided in this summary.)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: 27 adult subjects.
  • Data Provenance: The document does not specify the country of origin of the data. The study was prospective as data was "collected for each subject for 30 days while using the BlueStar mobile app with the CGM insulin dose calculator."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The document does not mention the use of experts to establish ground truth for the test set. The ground truth was established by the CGM data itself, as it measured glucose time in range.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe any adjudication method for the test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. The study focused on the effectiveness of the device (BlueStar® CGM insulin dose calculator) in improving or maintaining Time in Range compared to baseline, rather than comparing human reader performance with and without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the clinical study assessed the performance of the "BlueStar mobile app with the CGM insulin dose calculator" as used by the subjects. While patients are "human-in-the-loop" by entering carbohydrates, accepting recommendations, and performing activities, the primary performance metric (TIR) evaluates the algorithm's impact on glucose management, implying a standalone assessment of its functional impact when integrated into patient self-management. The device itself is software that calculates a dose, so the clinical study assesses the impact of these calculations on patient outcomes.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth used was outcomes data, specifically CGM glucose time in range (TIR) and time spent in hypoglycemia, measured directly from Continuous Glucose Monitoring (CGM) devices.

8. The sample size for the training set

The document does not provide information about a separate training set or its sample size. The clinical study described appears to be a validation study.

9. How the ground truth for the training set was established

Since information about a training set is not provided, how its ground truth was established is also not available in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 11, 2023

Welldoc, Inc. Ian Cadieux Head of Regulatory Affairs 10221 Wincopin Circle, Ste #150 Columbia, Maryland 21044

Re: K222888

Trade/Device Name: BlueStar® CGM insulin dose calculator Regulation Number: 21 CFR 862.1358 Regulation Name: Insulin Therapy Adjustment Device Regulatory Class: Class II Product Code: QRX Dated: September 22, 2022 Received: September 23, 2022

Dear Ian Cadieux:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Joshua Balsam -S

Joshua M. Balsam, PhD Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222888

Device Name BlueStar® CGM Insulin Dose Calculator

Indications for Use (Describe)

The BlueStar® CGM insulin dose calculator is software intended for the management of type 1 or type 2 diabetes in persons aged 18 years and older requiring fast-acting insulin dose calculator allows patients to calculate a dose of bolus insulin for a given amount of carbohydrates, the most recent CGM glucose reading and rate of change, activity, and, optionally, insulin on board (IOB). The BlueStar CGM insulin dose calculator requires a prescription.

Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date Prepared:May 25, 2023
Name of Manufacturer:WellDoc, Inc.
Address:10221 Wincopin Circle, Suite 150Columbia, MD 21044
Contact Person:Ian CadieuxHead of Regulatory Affairs
Phone:(619) 894-0873
Fax:(443) 692-3099
Trade or Proprietary Name:BlueStar® CGM insulin dose calculator
Common or Usual Name:Continuous glucose monitor informed insulin dose calculator
Product Codes:Classification: QRX
Regulation:21 CFR 862.1358
Regulatory Class:II
Classification Panel:Clinical Chemistry
Predicate Device:Omnipod 5 SmartBolus Calculator (K203772)
510k Number:K222888

{4}------------------------------------------------

Device Description

The BlueStar® CGM insulin dose calculator is a software device module existing in the same mobile medical application as BlueStar® Rx (K203434), which is intended for the management of diabetes. When connected to a compatible integrated continuous glucose monitor (iCGM) and under authorization from a qualified healthcare provider, the BlueStar CGM insulin dose calculator allows patients to calculate a dose of bolus insulin for a given amount of carbohydrates, the most recent iCGM glucose reading and its rate of change, activity, and, optionally, insulin on board (IOB). Other patient-specific inputs from BlueStar Rx are used in the calculation of the recommended dose- specifically, duration of insulin to carb ratio, correction factor, and target glucose. In addition to calculating specific dosing recommendations, the BlueStar CGM insulin dose calculator also provides coaching messages to assist the user in maintaining glucose within the target range.

The use of CGM inputs differentiates the BlueStar CGM insulin dose calculator from the insulin dose calculator included in the previously cleared BlueStar Rx, which uses blood glucose (BG) values from a BG meter using a "fingerstick" method. The CGM insulin dose calculator is intended to coexist with the BG insulin calculator function in the BlueStar Rx software application as the BG calculation may be necessary when CGM is unavailable or the CGM estimated blood glucose does not match how the user feels.

Indications for Use

The BlueStar® CGM insulin dose calculator is software intended for the management of type 1 or type 2 diabetes in persons aged 18 years and older requiring fast-acting insulin. The BlueStar CGM insulin dose calculator allows patients to calculate a dose of bolus insulin for a given amount of carbohydrates, the most recent CGM glucose reading and rate of change, activity, and, optionally, insulin on board (IOB). The BlueStar CGM insulin dose calculator requires a prescription.

Comparison to Predicate Device

The subject device matches the description of a Continuous Glucose Monitor Informed Insulin Dose Calculator – Insulin Therapy Adjustment Device described in 21 CFR 862.1358. A predicate device, the Omnipod 5 SmartBolus Calculator (K203772) of the same device type, shares substantially equivalent characteristics as the subject device. Both devices are intended for the management of diabetes. The indications for use of the subject device are within the indications of the predicate device, differing with respect to a narrower age indication in the subject device. Both devices are algorithmic software devices that calculate a recommended bolus insulin dose based on user-entered carbohydrates, most recent sensor glucose value, rate of change of the sensor glucose, insulin on board, and programmable correction factor, insulin to carbohydrate ratio, and target glucose value; more insulin is calculated when the sensor glucose trend is increasing, and less insulin is calculated when the sensor glucose trend is decreasing. The predicate device is a software component of a mobile medical application. Likewise, the subject device is a software component of a mobile medical application. Minor differences exist in the device settings that allow user-specific adjustments to the calculations and minor differences exist in the algorithms, which may result in minor differences in recommended insulin doses. The differences in device

{5}------------------------------------------------

settings and algorithms do not raise new questions regarding safety and effectiveness. Therefore, the BlueStar CGM insulin dose calculator is substantially equivalent to the predicate device.

Non-clinical Performance Testing

The following non-clinical testing was performed and reviewed to support the substantial equivalence of the subject device:

SoftwareSoftware verification and validation per: the FDA Guidance for the Content ofPremarket Submissions for Software Contained in Medical Devices (May 11,2005) for a Major Level of Concern and FDA Guidance for General Principlesof Software Validation (Jan 11, 2002).
CybersecurityCybersecurity was evaluated per the FDA Guidance FDA Guidance forCybersecurity in Medical Devices: Quality System Considerations and Content ofPremarket Submissions (April 2022).
Human FactorsA human factors and usability engineering process was followed consistent with theFDA's guidance document, Applying Human Factors and Usability Engineering toMedical Devices (Issued February 3, 2016). Human factors validation wasconducted with the intended user population and demonstrated safe and effective useof the BlueStar CGM insulin dose calculator.
ClinicalSimulationClinical simulations were performed to evaluate the clinical validity of the doserecommendations across the range of input parameter combinations includinginaccuracies from an interoperable input device.
Special ControlsThe subject device was evaluated to ensure the Special Controls in 21 CFR862.1358 are adequately addressed.

Clinical Testing

A clinical study was conducted on 27 adult subjects with either type 1 or type 2 diabetes who used a CGM and whose insulin regimen consisted of basal and mealtime (bolus) insulin. The primary endpoint of the study was to establish that the mean CGM glucose time in range (TIR) (defined as the percentage of time spent between 70 and180 mg/dL) of study participants using a version of the Bluestar mobile application integrated with the BlueStar CGM insulin dose calculator is not inferior to their baseline TIR determined from CGM data from 30 days prior. In addition, safety measures including time in hypoglycemia (obtained with CGM) and adverse events including hypoglycemic events were to be recorded. Data was collected for each subject for 30 days while using the BlueStar mobile app with the CGM insulin dose calculator. The statistical analysis of the study data showed that the mean TIR when using BlueStar with the CGM insulin dose calculator is not inferior to the baseline mean TIR prior to using the device. In

{6}------------------------------------------------

addition, there was no increase in time spent with glucose below 70 mg/dL and below 54 mg/dL in the subject populations. These data support the safety of the Bluestar CGM insulin dose calculator.

Substantial Equivalence Conclusion

The subject device in this premarket notification, the Bluestar CGM insulin dose calculator, has the same intended use and similar indications for use and technological characteristics as those of the predicate device. Performance testing demonstrated that the device performs as intended. The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The BlueStar CGM insulin dose calculator is substantially equivalent to the predicate device.

§ 862.1358 Insulin therapy adjustment device.

(a)
Identification. An insulin therapy adjustment device is a device intended to incorporate biological inputs, including glucose measurement data from a continuous glucose monitor, to recommend insulin therapy adjustments as an aid in optimizing insulin therapy regimens for patients with diabetes mellitus.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) A complete description of the required data inputs, including timeframe over which data inputs must be collected and number of data points required for accurate recommendations;
(ii) A complete description of the types of device outputs and insulin therapy adjustment recommendations, including how the recommendations are generated;
(iii) Robust data demonstrating the clinical validity of the device outputs and insulin therapy recommendations;
(iv) A robust assessment of all input data specifications, including accuracy requirements for continuous glucose monitors and other devices generating data inputs, to ensure accurate and reliable therapy adjustment recommendations. This assessment must include adequate clinical justification for each specification;
(v) A detailed strategy to ensure secure and reliable means of data transmission to and from the device, including data integrity checks, accuracy checks, reliability checks, and security measures;
(vi) Robust data demonstrating that users can understand and appropriately interpret recommendations generated by the device; and
(vii) An appropriate mitigation strategy to minimize the occurrence of dosing recommendation errors, and to mitigate the risk to patients of any residual dosing recommendation errors to a clinically acceptable level.
(2) The device must not be intended for use in implementing automated insulin dosing.
(3) Your 21 CFR 809.10(b) labeling must include:
(i) The identification of specific insulin formulations that have been demonstrated to be compatible with use of the device;
(ii) A detailed description of the specifications of compatible devices that provide acceptable input data (e.g., continuous glucose monitors, insulin pumps) used to provide accurate and reliable therapy adjustment recommendations;
(iii) A detailed description of all types of required data (inputs) and dosing recommendations (outputs) that are provided by the device; and
(iv) A description of device limitations, and instructions to prevent possible disruption of accurate therapy adjustment recommendations (e.g., time zone changes due to travel).