K Number
K222753

Validate with FDA (Live)

Date Cleared
2022-12-09

(88 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SeaSpine 7D Navigation Instruments are intended to be used during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The SeaSpine 7D Navigation Instruments are specifically designed for use with the 7D Surgical System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure can be identified.

Device Description

The SeaSpine 7D Navigation Instruments are used during the preparation and placement of SeaSpine screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The SeaSpine 7D Navigation Instruments are wireless optically tracked and specifically designed for use with the 7D Surgical system.

The SeaSpine 7D Navigation Instruments utilize commercially available passive reflective marker spheres to determine the position and orientation of instruments. Each tracked Instruments requires a unique marker position configuration to enable the tracking system to distinguish the tools from one to the other. Position and orientation data of tracked Spinal Instruments are linked to the preoperative scan data using the 7D Surgical System workstation.

AI/ML Overview

The provided text concerns the FDA 510(k) premarket notification for the "SeaSpine 7D Navigation Instruments." This submission focuses on the mechanical compatibility and accuracy of newly designed instruments within an existing navigated surgical system (7D Surgical System), rather than an AI/ML-driven diagnostic or prognostic device.

Therefore, many of the requested criteria related to AI/ML device validation (e.g., number of experts for ground truth, adjudication methods, MRMC studies, training set details) are not applicable to this type of device and the information is not provided in the submission. The device is a surgical instrument that is optically tracked and mechanically connects to an existing navigation system.

Here's an analysis of the provided information relative to your request:

Acceptance Criteria and Device Performance (Mechanical & Positional Accuracy)

The acceptance criteria are primarily focused on system compatibility, mechanical design, and positional accuracy of the instruments.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryDescription from DocumentReported Device Performance / Conclusion
System VerificationScope of the test is to verify the design requirement specifications of the SeaSpine Navigation Adapter, Reference Clamp, Pedicle Probe compatibility with the 7D Surgical System under test case protocols.Verification successful, all design requirements have been fulfilled.
System ValidationScope of the test is to validate the Indications for Use and Customer Requirements of the SeaSpine Navigation Adapter, Reference Clamp, Pedicle Probe compatibility with the 7D Surgical System under simulated use case situations.Validation successful, all user needs met.
Safety Regarding Risk AnalysisImplementation and effectiveness of all risk control requirements specified in the SeaSpine Navigation Adapter, Reference Clamp, Pedicle Probe risk analysis are tested and verified.Risk Control requirements are effective and mitigate the associated risks to an acceptable level.
Product Safety StandardsThe SeaSpine Navigation Adapter, Reference Clamp, Pedicle Probe was tested to the following recognized standards: ISO 10993-1, and ISO 17665-1. (Note: These standards relate to biocompatibility and sterilization, not directly positional accuracy).Compliance with recognized standards have been verified in the previous application. Previous test results have not been affected by this change.
Non-Clinical Accuracy (Positional)System's accuracy is tested using the SeaSpine Navigation Adapter, Reference Clamp, Pedicle Probe compatibility with the 7D Surgical System on phantom models following the ASTM F2554-10 Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems in addition to Target Registration Error (TRE). Also, Angular Trajectory Error (ATE) was used.All accuracy specifications have been met for the SeaSpine® Navigation Adapter, Reference Clamp, and Pedicle Probe. Accuracy testing for Universal Tracking / non-tracked surgical probe Array/Navigation Adapter assembly have been verified in previous application K192140 and K162375. (Specific numerical values for TRE/ATE or other accuracy metrics from ASTM F2554-10 are not explicitly provided in this summary, only that "all accuracy specifications have been met").

2. Sample Size for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated as a number of distinct "cases" or "patients" for positional accuracy testing. The document states "phantom models" were used. It implies repeatable tests were run to demonstrate accuracy, but the number of phantom models or repetitions is not specified.
  • Data Provenance:
    • Country of Origin: Not specified but likely conducted in North America (given the submitter's location in Canada and the FDA clearance).
    • Retrospective/Prospective: Neither. The studies were non-clinical performance surgical simulations conducted on phantom models. This is a prospective testing methodology for device performance on inanimate objects.

3. Number of Experts Used to Establish Ground Truth and Qualifications of those Experts

  • Not Applicable. This is not an AI/ML diagnostic device requiring expert interpretation for ground truth. Ground truth for positional accuracy testing (TRE/ATE) would be established by precise physical measurements using metrology equipment on the phantom models, not human expert consensus.

4. Adjudication Method for the Test Set

  • Not Applicable. As above, no human expert consensus or adjudication was involved in establishing ground truth for mechanical accuracy on phantom models.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This is not an AI/ML diagnostic or image-based assistance device. It's a suite of surgical instruments used with a navigation system. MRMC studies are not relevant for this type of device.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Partially Applicable / Adapted Interpretation: While not an AI algorithm, the device's accuracy (its "standalone" performance within the larger navigation system) was assessed through "Non-Clinical Accuracy" tests on phantom models. This involved evaluating the positional and angular accuracy of the SeaSpine instruments when used with the 7D Surgical System. The ASTM F2554-10 standard is for measuring the positional accuracy of computer-assisted surgical systems. The "standalone" performance here refers to the system's ability to accurately track and report the position of the instruments relative to the "ground truth" on the phantom.

7. The Type of Ground Truth Used

  • Precision Measurement/Metrology on Phantom Models: Ground truth for Target Registration Error (TRE) and Angular Trajectory Error (ATE) is established by precise physical measurements of the actual position of fiducials or targets on the phantom model, compared against the position reported by the navigation system. This is a form of engineered ground truth for spatial accuracy.

8. The Sample Size for the Training Set

  • Not Applicable. This is not a machine learning device that requires a "training set." The instruments are designed and verified based on engineering principles and compatibility with an existing optical tracking system.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As above, no training set for a machine learning model.

In summary, the provided document details the non-clinical testing of a surgical navigation instrument for its compatibility, safety, and mechanical positional accuracy with an existing stereotactic system. It does not describe the validation of an AI/ML-driven diagnostic or prognostic device, hence many of the requested criteria are not present or applicable. The key validation data provided relates to meeting ASTM standards for positional accuracy and demonstrating compatibility with the predicate device's system.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

December 9, 2022

SeaSpine Orthopedics Corporation % Elena Marenny Regulatory Specialist 7D Surgical ULC 60 Scarsdale Road, Unit 118 Toronto, ON M3B 2R7 Canada

Re: K222753

Trade/Device Name: SeaSpine 7D Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: September 7, 2022 Received: September 12, 2022

Dear Elena Marenny:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jesse Muir -S

For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222753

Device Name SeaSpine 7D Navigation Instruments

Indications for Use (Describe)

The SeaSpine 7D Navigation Instruments are intended to be used during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The SeaSpine 7D Navigation Instruments are specifically designed for use with the 7D Surgical System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure can be identified.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for SeaSpine. The logo features three curved lines in blue on the left side. To the right of the lines is the word "SeaSpine" in orange font. There is a period after the word "SeaSpine".

510(k) Summary of Safety and Effectiveness

The SeaSpine 7D Navigation Instruments

This summary of safety and effectiveness information is submitted in accordance with 21CFR §807.92.

    1. Submitter's name, address, telephone number, contact person. SeaSpine Orthopedics Corporation 5770 Armada Drive Carlsbad, CA 92008 Contact person: Jesse Albright Associate Manager, Regulatory Affairs Phone: (760) 216-5176 Fax: (760) 683-6874 Email: jesse.albright@seaspine.com
      Date prepared: December 09, 2022
    1. Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
Common/usual name:Stereotactic Instrument
Proprietary name:SeaSpine 7D Navigation Instruments

These devices are classified as follows:

Classification Name21 CFR SectionProduct Code
Stereotaxic instrument21 CFR §882.4560OLO

3. Substantially Equivalent Devices

SeaSpine Orthopedics Corporation believes the SeaSpine Navigation Adapter is substantially equivalent to the following currently marketed devices:

Product510(k)
7D Surgical System- Universal Tracking Clamp(Primary)K192140
Envision 3D™: Image Guidance System (7D SurgicalSystem was initially named “Envision 3D “)K162375

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for SeaSpine. The logo consists of a stylized wave graphic in shades of blue on the left, followed by the word "SeaSpine" in orange. The wave graphic has three curved lines that start thick and taper off to the right. The word "SeaSpine" is in a sans-serif font and is slightly blurred.

The indications for use of the subject device are equivalent to the predicate device K192140 and K162375. Furthermore, the technological characteristics of the SeaSpine Navigation Adapter are substantially equivalent. The differences in the technological characteristics do not raise new questions of safety and effectiveness. Consequently, the subject is substantially equivalent to the predicate device.

4. Indications for Use

The SeaSpine 7D Navigation Instruments are intended to be used during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The SeaSpine 7D Navigation Instruments are specifically designed for use with the 7D Surgical System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure can be identified.

5. Device Description and Technical Comparison to Predicate Devices

The SeaSpine 7D Navigation Instruments are used during the preparation and placement of SeaSpine screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The SeaSpine 7D Navigation Instruments are wireless optically tracked and specifically designed for use with the 7D Surgical system.

The SeaSpine 7D Navigation Instruments utilize commercially available passive reflective marker spheres to determine the position and orientation of instruments. Each tracked Instruments requires a unique marker position configuration to enable the tracking system to distinguish the tools from one to the other. Position and orientation data of tracked Spinal Instruments are linked to the preoperative scan data using the 7D Surgical System workstation.

SeaSpine Navigation Adapter (K192140) l.

The SeaSpine Navigation Adapter designed to connect to a 7D Surgical Universal Array (1 or 2) allows for rotation and continuous visualization of compatible SeaSpine Insert (SeaSpine Mariner Nav Insert) in the navigated 7D Surgical System. As with predicate Universal Tracking Clamp 510(K) 192140 device, calibration of the 7D Surgical Integration Tools is required for the surgeon to proceed with he intended workflow using the 7D Surgical System.

II. SeaSpine Reference Clamp (K162375)

The SeaSpine Reference Clamp (Radiolucent) is designed to be used when rigid fixation to the spinous process is required. As with predicate 7D Surgical Reference Clamp K162375, the SeaSpine Reference Clamp attaches rigidly to the spinous process in the region where

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for SeaSpine. The logo consists of a blue graphic on the left and the word "SeaSpine" in orange on the right. The graphic is made up of three curved lines that resemble waves or arcs.

the user plans to perform surgery. Once fixated, the user attaches the 7D Surgical Flex Array which holds the spherical markers which are tracked by the 7D Surgical System.

SeaSpine Pedicle Probe (K162375) III.

The SeaSpine Pedicle Probe is designed to extend the initial hole through the cancellous (soft) bone of the vertebra. As with predicate 7D Surgical Pedicle Probe K162375, the SeaSpine Pedicle probe Tracked Navigation Tool is used by the surgeon to perform imageguided surgery via 7D Surgical System integration, guiding the pedicle screws length selection and insertion into the vertebra.

6. Safety Considerations

The SeaSpine 7D Navigation Instruments add compatibility to the 7D System to include navigation and tracking of the SeaSpine compatible insert and tools did not impact conformity to regulatory compliance standards. Mechanical Design, and User Instructions risk control measures have been implemented to ensure all new risks associated with use of the SeaSpine 7D Navigation Instruments have been adequately controlled.

7. Technological Characteristics

The literature research and the comparison to the predicate devices show that the device makes use of equivalent technological characteristics and functionality and is intended for equivalent surgical procedures as compared to the predicate devices.

8. Nonclinical Performance Data

Verification and Validation activities have been conducted to provide assurance that the device meets the performance requirements under the indications for use conditions.

The following tests were performed to ensure the safety and effectiveness of the SeaSpine 7D Navigation Instruments:

  • Non-Clinical System, Software, and Instrumentation Verification and Validation
  • Non-Clinical Performance Surgical Simulations Conducted on Phantom Models
  • Compliance Conformity Assessments
    • ASTM F2554-10 Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems

Device performance tests were performed to verify the absolute accuracy and repeatability of the accuracy of the device, and the navigation accuracy according to ASTM F2554-10. In addition, Target Registration Error (TRE) and Angular Trajectory Error (ATE) has been used to evaluate the clinical accuracy of the system on phantom models in a clinical simulated environment. TRE and ATE evaluates the error discrepancy between the position of the SeaSpine Navigation Adapter, Reference Clamp, and Pedicle Probe reported by the image guided surgery system and the ground truth position measured physically or otherwise.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for SeaSpine. The logo features a stylized wave graphic in shades of blue on the left. To the right of the wave is the company name "SeaSpine" in orange lettering.

The following Table 1 contains a summary of verification and validation performed on the 7D Surgical System:

Verification and ValidationDescriptionConclusion
System VerificationScope of the test is to verify thedesign requirement specificationsof the SeaSpine NavigationAdapter, Reference Clamp,Pedicle Probe compatibility withthe 7D Surgical System under testcase protocols.Verification successful, all designrequirements have been fulfilled.
System ValidationScope of the test is to validate theIndications for Use and CustomerRequirements of the SeaSpineNavigation Adapter, ReferenceClamp, Pedicle Probecompatibility with the 7D SurgicalSystem under simulated use casesituations.Validation successful, all userneeds met.
Safety regarding risk analysisImplementation and effectivenessof all risk control requirementsspecified in the SeaSpineNavigation Adapter, ReferenceClamp, Pedicle Probe riskanalysis are tested and verified.Risk Control requirements areeffective and mitigate theassociated risks to an acceptablelevel.
Product Safety standardsThe SeaSpine NavigationAdapter, Reference Clamp,Pedicle Probe was tested to thefollowing recognized standards:ISO 10993-1, and ISO 17665-1.Compliance with recognizedstandards have been verified inthe previous application. Previoustest results have not been affectedby this change.
Non-Clinical AccuracySystem's accuracy is tested usingthe SeaSpine Navigation Adapter,Reference Clamp, Pedicle Probecompatibility with the 7D SurgicalSystem on phantom modelsfollowing the ASTM F2554-10Standard Practice forMeasurement of PositionalAccuracy of Computer AssistedSurgical Systems in addition toTarget Registration Error.All accuracy specifications havebeen met for the SeaSpine®Navigation Adapter, ReferenceClamp, and Pedicle Probe.Accuracy testing for UniversalTracking / non-tracked surgicalprobe Array/Navigation Adapterassembly have been verified inprevious application K192140 andK162375.

All non-clinical tests successfully passed demonstrating that the subject device performs as safely and effectively as the predicate device and supporting substantial equivalence.

9. Clinical Data

A clinical trial was not required to demonstrate that the SeaSpine 7D Navigation Instruments perform as safely and effectively as the predicate device. Clinical validation is unnecessary as the SeaSpine 7D Navigation Instruments introduce no new indications for use, and device features are equivalent to the previously cleared predicate device identified. The clinical safety and effectiveness of the SeaSpine 7D Navigation Instruments are historically accepted for both the predicate and subject device.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains the logo for SeaSpine. The logo consists of a blue graphic on the left and the word "SeaSpine" in orange on the right. The graphic is made up of three curved lines that are stacked on top of each other. The word "SeaSpine" is written in a sans-serif font.

10. Conclusion

The SeaSpine 7D Navigation Instruments is substantially equivalent in safety and effectiveness to the predicate devices identified above:

  • . The predicate devices and The SeaSpine 7D Navigation Instruments use equivalent technologies.
    The non-clinical verification and validation performed support the safety and effectiveness of The SeaSpine 7D Navigation Instruments compatibility and useability with the 7D Surgical System. The conclusions drawn from the non-clinical tests demonstrate that The SeaSpine 7D Navigation Instruments perform as safely and effectively as the legally marketed device according to the comparison based on the requirements of 21 CFR §882.4560 and the information provided herein. It is concluded that The SeaSpine 7D Navigation Instruments is substantially equivalent to the predicate device with respect to its indications for use, technological characteristics, and performance characteristics.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).