K Number
K222751

Validate with FDA (Live)

Date Cleared
2023-02-01

(142 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

LED Light Therapy Device use of the red, blue, Yellow and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions.

The red light (633±10mm wavelength) is generally indicated to treatment of superficial, benign vascular, and pigmented lesions.

The blue light (417±10 nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.

The Yellow light (599±10nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated for treatment of periorbital wrinkles and rhytides.

The infrared light (835±15nm wavelength) is generally use for the temporary relief of minor muscle and joint pain. arthritis and musele spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

Device Description

The LED Light Therapy Device uses specific wavelengths of light, produced by LEDs (Light emitting diodes), to manage aesthetic conditions. The device produces light in the red light region of the spectrum (633±10nm), in the blue light regions of the light spectrum (417±10mm), yellow light area (590 ± 10nm) and infrared light region of light spectrum (835±15nm).

The LED Light Therapy Device instrument is equipped with two irradiators. The RBY irradiator has five panels, each of which emits three different colors of light sources (red light, blue light, yellow light); RBI irradiator has five panels, each panel emits three different colors of light sources (red light, blue light, infrared light).

The instrument is mainly composed of a main frame, an irradiator, and a lifting stand.

AI/ML Overview

Based on the provided text, the document is a 510(k) Premarket Notification from the FDA regarding an LED Light Therapy Device (KN-7000L). It focuses on establishing substantial equivalence to previously cleared predicate devices, primarily through comparison of technical specifications and non-clinical testing.

Therefore, the document does NOT contain information regarding clinical studies, acceptance criteria for an AI/device performance, expert ground truth establishment, or multi-reader multi-case studies. The device in question is an LED light therapy device, not an AI-powered diagnostic or assistive device that would typically undergo such rigorous performance evaluation against patient data.

The document details the device's technical specifications and confirms compliance with electrical safety, electromagnetic compatibility, and photobiological safety standards.

As a result, I cannot provide the requested information for the following points as they are not present in the provided text:

  1. A table of acceptance criteria and the reported device performance: No performance metrics beyond technical specifications (wavelength, irradiance, etc.) and compliance with safety standards are mentioned.
  2. Sample sizes used for the test set and the data provenance: No clinical test sets or data provenance (country, retrospective/prospective) are discussed. The "test set" here refers to non-clinical tests (e.g., electrical, photobiological).
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as there's no ground truth established for clinical performance.
  4. Adjudication method for the test set: Not applicable.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No MRMC study was conducted or mentioned. This type of study is typical for AI devices assisting human readers in diagnostic tasks, which is not the function of this LED light therapy device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as there is no AI algorithm.
  7. The type of ground truth used: Not applicable.
  8. The sample size for the training set: Not applicable, as this is not an AI/machine learning device.
  9. How the ground truth for the training set was established: Not applicable.

However, I can extract the following information that is relevant to the type of acceptance criteria and "study" conducted for this device, which focuses on non-clinical performance and substantial equivalence:

1. Table of "Acceptance Criteria" (Non-Clinical Performance) and Reported Device Performance

For this type of device, "acceptance criteria" primarily relate to meeting established safety and performance standards rather than clinical efficacy metrics derived from patient data. The manufacturer asserts substantial equivalence based on these comparisons and non-clinical testing.

Feature / CriterionAcceptance "Criterion" (via Predicate Comparison & Standards Compliance)Reported Device Performance (KN-7000L)
Product CodeSame as predicate (GEX)GEX
Indications for UseSame or substantially equivalent to predicate devices for treating dermatological conditions (superficial, benign vascular, pigmented lesions, moderate inflammatory acne vulgaris, periorbital wrinkles/rhytides, temporary relief of minor muscle/joint pain, arthritis, muscle spasm, stiffness, muscle tissue relaxation, increased local blood circulation).Same as predicates, encompassing red (633±10nm), blue (417±10nm), yellow (599±10nm), and infrared (835±15nm) regions for dermatological conditions and pain relief.
Wavelength(s)Within ranges of predicate devices.RBY irradiator: Red (633±10nm), Blue (417±10nm), Yellow (590±10nm). RBI irradiator: Red (633±10nm), Blue (417±10nm), Infrared (835±15nm). (Considered "basically the same" or "substantially equivalent" to predicates K190938 and K200104 despite minor numerical differences).
Maximum Power Density (mW/cm²)Within size ranges of predicate devices.Red: 20-96 mW/cm²; Blue: 10-120 mW/cm²; Yellow: 5-35 mW/cm²; Infrared: ≤ 70 mW/cm². Combined: Red/IR: 166 mW/cm², Blue/IR: 190 mW/cm². (Considered "substantially equivalent" as within predicate ranges).
Standard Dose (Joules)Within size ranges of predicate devices.Red: 155 J/cm²; Blue: 144 J/cm²; Yellow: 42 J/cm²; IR: 84 J/cm². Combined: Red/IR: 199 J/cm², Blue/IR: 228 J/cm². (Considered "substantially equivalent" as within predicate ranges).
Effective Irradiation Area (CM²)Within size ranges of predicate devices.900 cm² ±10% (Considered "substantially equivalent" compared to predicate ranges of 500-1008 cm²).
Electrical SafetyCompliance with IEC 60601-1, IEC 60601-1-2.Tested and meets IEC 60601-1: 2005+A1:2012, IEC 60601-1-2:2014.
Photobiological SafetyCompliance with IEC 62471.Tested and complies with IEC 62471:2006 (incorporating ANSI IESNA RP 27.1, 27.2, 27.3).
Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2.Tested and meets IEC 60601-1-2:2014. (Note: IEC 60601-2-57:2011 is also listed as an EMC test, but it is actually a particular standard for non-laser light sources for therapeutic and diagnostic use, which includes EMC aspects).
Software Verification & ValidationDocumentation consistent with moderate level of concern, software requirements met, hazards mitigated.Software documentation consistent with moderate level of concern submitted. System validation testing demonstrated all software requirement specifications met and hazards mitigated.
Additional Main Performance Indicators (e.g., power supply, noise, timing)Met for the device, and within acceptable ranges for medical devices of this type.Power supply: AC 100-240V±10%, 50/60 Hz±2%; Input power: 600VA; Working noise: ≤ 60dB (A); Timer error: ≤ ±2% of set value; Continuous working time ≤ 99min. Functions include manual stop, replaceable irradiator, preset treatments, calibration, time/dose modes, continuous/pulse output.

Summary of "Study" Proving Compliance:

The "study" in this context is a non-clinical testing and comparison study aiming to demonstrate substantial equivalence (SE) to predicate devices by verifying the new device shares fundamental technological characteristics and performs comparably in terms of safety and technical specifications.

  • Type of "Study": Non-clinical performance testing and comparative analysis to predicate devices.
  • Purpose: To demonstrate substantial equivalence (SE) for 510(k) clearance, indicating the device is as safe and effective as a legally marketed predicate device.
  • Ground Truth: The "ground truth" for this regulatory submission is adherence to recognized medical device standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, IEC 62471) and the established performance characteristics of the predicate devices. The device's measured technical specifications (e.g., wavelength, irradiance, power) were compared against these standards and predicate parameters.
  • Data Provenance/Sample Size: Not applicable in the traditional sense of patient data. The "sample size" refers to the tested units of the KN-7000L device undergoing the specified non-clinical tests. The data provenance is from the manufacturer's internal testing. The document does not specify the exact number of units tested, but it is implied that the testing performed on the device samples confirmed compliance.
  • Expert involvement for Ground Truth, Adjudication, MRMC, Standalone AI, Training Set: These concepts are not applicable to the non-clinical testing and substantial equivalence demonstration for a light therapy device. There is no AI component, and no human diagnostic performance is being evaluated.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 1, 2023

Xuzhou Kernel Medical Equipment Co., Ltd. % Shanfeng Jiang Regulation Control Manager Guangzhou Junyi Information Technology Co., Ltd. Room 304. Building A. No. 62 Nanvun 2nd Road. Science Town Huangpu District, Guangzhou City, Guangdong 510663 China

Re: K222751

Trade/Device Name: LED Light Therapy Device, KN-7000L Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 5, 2022 Received: December 5, 2022

Dear Shanfeng Jiang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang -S

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Device Name

LED Light Therapy Device

Indications for Use (Describe)

LED Light Therapy Device use of the red, blue, Yellow and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions.

The red light (633±10mm wavelength) is generally indicated to treatment of superficial, benign vascular, and pigmented lesions.

The blue light (417±10 nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.

The Yellow light (599±10nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated for treatment of periorbital wrinkles and rhytides.

The infrared light (835±15nm wavelength) is generally use for the temporary relief of minor muscle and joint pain. arthritis and musele spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 - 510(k) Summary

K222751

Date of Summary Preparation: September 12, 2022 Update to current: January 31, 2023

1. Submitter's Identifications

Submitter's Name: Xuzhou Kernel Medical Equipment Co., Ltd. Address: Kernel Mansion, Economic Development District, Xuzhou City, Jiangsu Province, China Contact Person: Cuiling Xi Contact Title: Management representative Contact E-mail Address: mananger@kernelmed.com Telephone: +86-516-87732208

2. Correspondent's Identifications

Correspondent's Name: Guangzhou Junyi Information Technology Co., Ltd. Address: Room 304, Building A, No. 62 Nanyun 2nd Road, Science Town, Huangpu District, Guangzhou City, Guangdong, 510663, China Contact Person: Shanfeng Jiang Contact Title: Regulation Control Manager Contact E-mail Address: Jiang13620586569@126.com Telephone: +86-20-82329549 Fax: +86-20-82329549

3. Name of the Device

Device Classification Name: Laser surgical instrument for use in general and plastic surgery and

in dermatology Trade Name: LED Light Therapy Device Model: KN-7000L Classification Panel: General & Plastic Surgery Product Code: GEX Device Classification: Class II Regulation Number:21 CFR878.4810

4. The Predicate Devices

Phototherapy Systems K190938 K200104 Oxylight

5. Device Description

5.1 Device introduction:

The LED Light Therapy Device uses specific wavelengths of light, produced by LEDs (Light

{4}------------------------------------------------

emitting diodes), to manage aesthetic conditions. The device produces light in the red light region of the spectrum (633±10nm), in the blue light regions of the light spectrum (417±10mm), yellow

light area (590 ± 10nm) and infrared light region of light spectrum (835±15nm).

The LED Light Therapy Device instrument is equipped with two irradiators. The RBY irradiator has five panels, each of which emits three different colors of light sources (red light, blue light, yellow light); RBI irradiator has five panels, each panel emits three different colors of light sources (red light, blue light, infrared light).

5.2 Composition : The instrument is mainly composed of a main frame, an irradiator, and a lifting stand.

5.3 Equipment functions:

  • Light source is extensible according to the areas to be treated to realize large-area irradiation.
  • Light source is arranged in matrix, making it more suitable for facial treatment.
  • Fixing and electrical connection of the light source is realized through a quick connector, making the replacement more convenient and fast.
  • Design of free-lifting cantilever allows the light source to stay at any position and angle.
  • 10.4" touch screen simplifies the operation and eliminates the need for professional training.
  • 5 commonly used treatment schemes can be stored to avoid repeated setting.
  • Warm voice guide makes it easy for you to operate.
  • Effective irradiance of the light source can be adjusted according to the treatment need.
  • Double switch protection of power switch and power-on password, no need to worry about being misused.
  • Non-invasive operation, no damage to skin cells.
  • No special care is required after treatment, and makeup can be applied normally.

5.4 Instrument type:

Protection against electric shock: Class I.

Operating mode: continuous operation.

An enclosed apparatus that is not protected against liquid ingress.

Equipment that cannot be used in the presence of flammable anesthetic gas mixed with air or with oxygen or nitrous oxide.

5.5 Main performance indicators

Power supply: AC 100-240V±10%, 50/60 Hz土2% Input power: 600V A Fuse specification, model & rating:T8.0AL/250V Φ5*20 Structure: wheeled Display mode: LCD display Effective irradiation area: 900cm2 ± 10% Irradiation distance: 6cm±1cm Spectral peak wavelength: Red light: 633 nm士 10 nm; Blue light: 417 nm±10 nm;

{5}------------------------------------------------

Yellow light: 590 nm±10 nm;

Infrared: 835nm±15nm.

Effective irradiance: The error between the effective irradiance and the nominal value indicated in the nameplate shall be no more than ±25%, and the effective irradiance shall be not more than 200mW/cm2.

Red light: 20~96mW/cm2;

Blue light: 10~120mW/cm2;

Yellow light: 5~35mW/cm2;

Infrared ≤ 70mW/cm2.

Red/IR: 20~166mW/cm²,

Blue/IR: 10~190mW/cm²

Working noise:

Under normal working conditions, the noise generated by the instrument shall not exceed 60dB (A).

Stand adjustment:

Under normal working condition, the stand can be adjusted up, down, left and right, and the irradiator can be fixed at any angle.

Timing and functions

The device has a timer with an error not greater than ±2% of the set value, Continuous working time ≤ 99min;

The function of manually stopping radiation output can be realized through the software pause and stop buttons.

Replaceable irradiator.

Treatment can be preset.

The instrument has a calibration function.

The instrument has a time mode.

The instrument has a dose mode.

The instrument has two functions of continuous output and pulse output.

No.NameQuantityUnit
1Main frame1Unit
2Irradiator (Irradiator BRY or Irradiator BRI for option)2Pcs
3Protective goggles (for doctors, patients)2Pieces
4Power cord1Pcs.
5Fuse2Pcs.
6Tools (hex wrench (8mm, 4mm, 5mm and 2mm) andscrewdriver, 1 for each)1Set
7M3*10 screw4Pcs.
8M6*15 hexagon socket head cap screw6Pcs.
9Upper hole plug, lower hole plug, seal plug2 for eachPcs.
10Instructions for Use1Copy

5.6 The list of accessories

{6}------------------------------------------------

11Certificate of Conformity1Copy
----------------------------------------

6. Intended Use of Device

LED Light Therapy Device use of the red, blue, Yellow and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions.

The red light (633±10nm wavelength) is generally indicated to treatment of superficial, benign vascular, and pigmented lesions.

The blue light (417±10 nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.

The Yellow light (599±10nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated for treatment of periorbital wrinkles and rhytides.

The infrared light (835±15nm wavelength) is generally use for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

{7}------------------------------------------------

7. Summary of Substantial Equivalence

Proposed DevicePredicate devicePredicate deviceComparison
510k NumberK222751K190938K200104------
Product CodeGEXGEXGEXSame
Proprietary NameLED Light Therapy DevicePhototherapy SystemOxylight-------
ModelKN-7000LHS-770/-------
ManufacturerXuzhou Kernel Medical Equipment Co., Ltd.Shanghai Apolo Medical Technology Co., Ltd.RAJA Trading Company, Inc.------
Indications for UseLED Light Therapy Device use of the red, blue, Yellow and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions. The red light(633±10nm wavelength) is generally indicated to treatment of superficial, benign vascular, and pigmented lesions. The blue light(417±10 nm wavelength); is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acnePhototherapy Systems use of the red, blue and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions. The blue light (415nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris The red light (630nm wavelength) is generally indicated to treatment of superficial, benign vascular,The Oxylight is intended for dermatological use by physicians and healthcare professionals for the following: LED Technology is intended for: -Blue LED – 465nm – to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris. -Red LED 625nm- for treatment of superficial, benign vascular and pigmented lesions. -Yellow LED 590nm - treatment of periorbitalSame
vulgaris.and pigmented lesionswrinkles and rhytides.
The Yellow light(599±10nmwavelength)isgenerallyindicated to treatdermatological conditions andspecifically indicated fortreatment of periorbitalwrinkles and rhytides.The infrared light (835±15nmwavelength)is generally usefor the temporary relief ofminor muscle and joint pain,arthritis and muscle spasm;relieving stiffness; promotingthe relaxation of muscletissue; and to temporarilyincrease local bloodcirculation where applied.The infrared light (835nmwavelength) is generally usefor the temporary relief ofminor muscle and joint pain,arthritis and muscle spasm;relieving stiffness; promotingthe relaxation of muscletissue; and to temporarilyincrease local bloodcirculation where applied.
Wavelength(s)(nm)RBY irradiator : (Red light633 ± 10nm, blue light 417 ±10nm, yellow light 590 ±10nm).RBI irradiator: (Red 633 ±10nm, blue 417 ± 10nm,infrared 835 ± 15nm)Red light 630 ± 15nmBlue light 415 ± 15nmInfrared 835 ± 15nmRed light 625nmBlue light 465nmYellow light 590nmSubstantially equivalentKn-7000L Wavelength(s)is basically the same asK190938 and K200104.
Panels TypeRBY Irradiator has 5 panels:Red light:465EA LEDs;blue light: 470EA LEDS;3 panel: 180EA LEDS4 Panel: 240 EA LEDs.The panels may emit the threeThree type, each head typehasonly one light.Substantially equivalentKn-7000L Panels Typeis basically the same as
yellow light: 465EA LEDs;RBI irradiator: has 5 panels:Red light:465EA LEDs;blue light: 470EA LEDs;infrared: 465EA LEDs;The panels may emit the threelight (red, blue infrared)individual or in combinationlight (red, blue, infrared)individual or in combinationRed, Blue, yellow.K190938 and K200104.
Output Powerindividual or in combination.Each panel has three differentkinds of light-emitting diodes,and the energy power of thediode is 0.5WEach LED lamp bead has 4diodes that emit differentcolors, the energy power of adiode is 3W.unknownThe Output Power of theproposed device is differentfrom the subject predicateK190938.Theproposeddevice has passed the safetytest. This difference will notimprovethesafetyandeffectiveness of the proposeddevice.
Maximum powerdensity in mW(mW/CM²)Red light: 2096 mW/cm2Blue light: 10120 mW/cm2Yellow light: 5~35 mW/cm2Infrared: ≤ 70 mW/cm2Red/IR: 166mW/cm2,Blue/IR: 190mW/cm2(1) Red light: 115mW/cm2,(2) Blue light: 120mW/cm2,(3) IR: 70mW/cm2,(4) Red/IR: 120mW/cm2,(5) Blue/IR: 150mW/cm2.(1) Red: 100mW/cm2(2) Blue: 45mW/cm2(3) Yellow 35W/cm2Substantially equivalentSubstantiallyequivalent-within size rangesof predicate devices.
Standard does inJoulesRed light: 155J/cm2,blue light: 144J/cm2,Yellow light: 42J/cm2,IR: 84J/cm2,Red/IR: 199J/cm2(1) Red light: 138J/cm2,(2) blue light: 144J/cm2,(3) IR: 84J/cm2,(4) Red/IR: 144J/cm2(5) Blue/IR: 180J/cm2(1) Red: 120J/cm2(2) Blue: 54J/cm2(3) Yellow: 42J/cm2Substantially equivalentSubstantiallyequivalent-within size rangesof predicate devices.
Blue/IR: 228J/cm2
Adjustable doserangeRed light: 2096 J/cm2Blue light: 10120 J/cm2 ²Yellow light: 535 J/cm2Infrared: ≤ 70 J/cm2Red/IR: 20166 J/cm2,Blue/IR: 10~190 J/cm2(1) Red light: 1-242J/cm2,(2) blue light: 1-180J/cm2,(3) IR:1-147J/cm2,(4) Red/IR: 1-144J/cm2,(5) Blue/IR:1-180J/cm2(1) Red light: 1-240J/cm2,(2) blue light: 1-108J/cm2,(3) Yellow IR:1-84J/cm2,Substantially equivalentSubstantiallyequivalent-within size rangesof predicate devices.
Numbers of LEDsRed light:465EA LEDs;Blue light: 470EA LEDs;Yellow light: 465EA LEDs;Infrared: 465EA LEDs;3 panels: 180EA,4 panels: 240EAunknownSubstantially equivalentSubstantiallyequivalent-within size rangesof predicate devices.
Effective irradiationarea: (CM²)900 cm² ±10%756cm2 and 1008cm2500cm² and 860cm²Substantiallyequivalent-within size rangesof predicate devices.
Structural styleWheeledWheeledWheeledSame
Structure compositionMain frame, irradiator, liftingframe.Main frame, irradiator, liftingframeIrradiator and supportSame
Power supplyAC 100-240V 50/60HzAC 100-240V 50/60Hz 10AAC 100-240V 50/60HzSame
Treatment time20min(Recommendedtreatment time)20min(Recommendedtreatment time)20min(Recommendedtreatment time)Same
Operation interfaceDisplay ScreenDisplay ScreenDisplay ScreenSame
SoftwareYesYesYesSame
Safety classificationClass IClass IClass ISame
StandardIEC 60601-1IEC 60601-1-2IEC 60601-2-57IEC 60601-1IEC 60601-1-2IEC 60601-2-57IEC 60601-1IEC 60601-1-2IEC 60601-2-57SameAlldevicesmeettherequirements of effectiveness
IEC 62471IEC 62471IEC 62471safety.

Table 1 Comparison to Predicate Device for KN-7000L

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

8. Substantial Equivalence discussion:

The indication for use of the KN-7000L LED Light Therapy Device are the same as that for the precifications of the KN-7000L LED Light Therapy Device are either the same or substantially equivalent as compared to the predicate devices. There are no technological differences that raise new or different questions of safety or effectiveness.

{12}------------------------------------------------

9. Non-Clinical Tests Performed:

Electrical Safety and Electromagnetic Compatibility Testing - KN-7000L LED Light therapy equipment has been tested and meets the following standard requirements of medical equipment:

● IEC60601-1: 2005+2005+CORR.1:2006+CORR.2:2007+A1:2012 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance.

· IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests.

· IEC 60601-2-57:2011 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests.

Photobiological Safety Testing - The KN-7000L LED Light Therapy Device have been tested and comply with IEC 62471:2006, Photobiological safety of lamps and lamp systems, 1st edition. This IEC standard incorporates the principles of the following ANSI IESNA recommended practices:

· RP 27.1:2005 Recommended practice for photobiological safety for lamps and lamp systems -General requirements.

· RP 27.2:2000 Recommended practice for photobiological safety for lamps and lamp systems -Measurement techniques.

· RP-27.3:2007 Recommended practice for photobiological safety for lamps and lamp systems -Risk group classification and labeling.

Software Verification and Validation - Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

10. Conclusion:

Based on comparing to predicate device, the proposed device of KN-7000L are determined to be Substantially Equivalent (SE) to the predicate device, in respect of safety and effectiveness.

--- End of this section ---

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.